Operating Temperature-10 to 50°C
Storage Temperature-20 to 70°C
Power(4) AAAlkaline or optional rechargeable batteries
Low Battery WarningYes, on display
ClearBrite™ DisplayHigh contrast 128 x 64 pixel addressable graphic
LCD w/daylight backlight
2.4 x 1.8 in. • 63 x 44 mm
Serial CommunicationsYes, ASCII, RS-232, requires optional Martel
1919069 serial cable or 1919896 USB cable
CE – EMCEN50082-1 and EN55022: 1994 Class B
SafetyCSA C22.2 No. 1010-1: 1992
Weight (with batteries)1.8 lb (0.82 kg)
Size8.5 x 4.8 x 2.1 in. (22 x 12 x 5.3 cm)
OtherIP54
Maximum current output in voltage ranges is 1 mA with an output
impedance of ≤ 1 Ω
TC TypeRange (°C)Accuracy (°C)
With CJ OFFWith CJ ON
J-210.0 – 0.00.40.6
0.0 – 800.00.20.4
800.0 – 1200.00.30.5
K-200.0 – 0.00.60.8
0.0 – 1000.00.30.5
1000.0 – 1372.00.50.7
T-250.0 – 0.00.60.8
0.0 – 400.00.20.4
E-250.0 – -100.00.60.8
-100.0 – 1000.00.20.4
R0.0 – 1767.01.21.4
S0.0 – 1767.01.21.4
B600.0 – 800.01.21.0
800.0 – 1000.01.31.5
1000.0 – 1820.01.51.7
C0.0 – 1000.00.60.8
1000.0 – 2316.02.32.5
XK-200.0 – 800.00.20.4
BP0.0 – 800.00.91.1
800.0 – 2500.02.32.5
L-200.0 – 0.00.30.5
0.0 – 900.00.20.4
U-200.0 – 0.00.50.7
0.0 – 600.00.30.5
N-200.0 – 0.00.81.0
0.0 – 1300.00.40.6
Notes
All specifications apply at 23°C ± 5°C unless otherwise stated. Outside of this range add ±0.005% of reading/°C.
Specifications are for battery operation only.
Accuracy is % of reading ± floor spec.
Maximum current output in voltage ranges is 1 mA with an output impedance of ≤ 1Ω.
Maximum load on mA source is 1000Ω. Voltage input range on simulate mode 5 – 30 V.
Frequency input voltage amplitude range is 1V to 20V zero based square wave only. Output amplitude is adjustable from 1V to 20V, and is a square wave
with 50% duty cycle. For output frequency, a negative offset of approximately -0.1V is present to assure zero crossing.
In Ohms source and RTD source modes, units are compatible with smart transmitters and PLCs that use a strobing excitation current. Frequency
0
0 to 500 (0 to 35 Bar)±0.025 %200 %
0 to1000 (0 to 70 Bar)±0.025 %200 %
0 to1500 (0 to 100 Bar)±0.035 %200 %
0 to3000 (0 to 200 Bar)±0.05 %200 %
0 to5000 (0 to340 Bar)±0.05 %200 %
0 to10000 (0 to 700 Bar)±0.01 %120 %
Non Isolated Compound (PSIG):
-0.4 to 0.4 (-20 to 20 mBar)
-1 to 1 (-70 to 70 mBar)
-5 to 5 (-350 to 350 mBar)
-7.2 to 7.2 (-500 to 500 mBar)
-10 to 10 (-700 to 700 mBar)
-15 to 15 (-1 to 1 Bar)
-15 to 30 (-1 to 2 Bar)
Isolated Compound (PSIG):
-12 to 50 (-0.8 to 3.5 Bar)±0.03 %±0.03 %200 %
-12 to 100 (-0.8 to 7 Bar)±0.025 %±0.025 % 200 %
-12 to 150 (-0.8 to 10 Bar)±0.03 %±0.03 %200 %
-12 to 300 (-0.8 to 20 Bar)±0.025 %±0.025 % 150 %
Isolated Absolute (PSIA):
0 to 15 (0 to 1 Bar)±0.04 %300 %
0 to 30 (0 to 2 Bar)±0.025 %300 %
0 to 50 (0 to 3.5 Bar)±0.03 %300 %
0 to 100 (0 to 7 Bar)±0.025 %300 %
0 to 300 (0 to 20 Bar)±0.025 %200 %
Differential (PSID):
0 to 5 (0 to 350 mBar)±0.075%400 %
0 to 30 (0 to 2 Bar)±0.025%300 %
0 to 50 (0 to 3.5 Bar)±0.03%300 %
Martel Electronics offers 27 standard pressure
modules, covering gauge, vacuum, absolute,
compound, and differential measurements. All
modules are directly compatible with the BetaGauge II. With the Model BPPA-100 Pressure
Module Adapter, all modules (with the exception
of the DC measurement model) are fully compatible
with the Martel MC-1210 and MC-1010 MultiFunction Calibrators, the BetaGauge 330, 321A,
311A and 301 Pressure Calibrators, the DMC-1410
Documenting Multi-Function Calibrator, and the
Martel Electronics M2001 and 3001 Laboratory/Bench
Standards.
Pressure ranges may be displayed in any of 13 user
selectable units. Water density correction factors of 4 °C,
20 °C, or 60 °F can be selected for either water column
unit. The choice of pressure unit may be restricted by
limitations on resolution of the instrument display of the
particular calibrator the module is used with. For optimum
mechanical strength, external pressure connection is made
by a 1/8” FNPT 316SS connector welded to a stainless steel
metal plate.
General Features
.
27 standard ranges
.
Gauge, vacuum, absolute, compound, and
differential measurements
.
Accuracy specified over 15 °C to 35 °C range
.
Isolated and non-isolated measurements,
range dependant
Need similar performance
for voltage and current only?
If your calibration needs are limited to voltage and current, the
Martel M2000 Bench Calibrator provides all of the performance,
functionality, and ease of use of the 3001.