Martel Electronics MS-420 User Manual

MINI-SOURCE PORTABLE CALIBRATOR
SPECIFICATIONS:
Output: 5 current steps Battery Life: 50 hours @ 4 mA out (4, 8, 12, 16, 20 mA)
Accuracy:
±0.075 of full scale
Voltage Compliance:
Source: Internal 9 volt battery Simulate: External supply - up to 30 Vdc clips on 8" leads.
Temperature Stability: 100 ppm/°C Size: 3.8" x 2.4" x 1.0" Power Supply: 9 volt alkaline battery
TYPICAL APPLICATIONS:
• CHART RECORDER
• PANEL METER
• DATA ACQUISITION
• OTHER PROCESS INSTRUMENTS
Temperature Range:
Operating: 0° to 50°C
Storage: –25° to 60°C Output Terminations: Two alligator
Unit furnished with standard belt clip.
+
(+) outputs (–)
4
8 12 16 20
milliamps
batt
ovl
simulate source off
MINI-SOURCE
MINIATURE CURRENT CALIBRATOR
output select
CURRENT SIMULATE MODE (2 Wire Transmitter)
+
(+) outputs (–)
4
8 12 16 20
milliamps
batt
ovl
simulate source off
MINI-SOURCE
MINIATURE CURRENT CALIBRATOR
output select
CURRENT SOURCE MODE
4-20 mA LOOP
Mini-Source Acting as Current Source
4-20 mA LOOP
Mini-Source Acting as 2 Wire Transmitter
Device Under Calibration
+
+
Device Under Calibration
+
300 ohm Max. Load Resistance
DC Power Supply
30 VDC Max.
“The Ultimate Process Loop Tool”
Martel Electronics
P.O. Box 770, 1F Commons Dr., Londonderry, NH 03087
Tel: 1-800-821-0023 1-603-434-8179 E-mail: sales@martelcorp.com
Loading...