7. Manipulation of the needle valve, vernier control and the
handles can produce any pressure desired within the
MECP10K hand pump operating range (0 to 10,000 PSI).
8. CAUTION: To prevent personnel injury, the pressure must be
reduced to zero before disconnecting any element of the
system. When the calibration is completed, carefully open the
needle valve to release the pressure. NOTE: Opening the
needle valve releases the pressure media into the reservoir.
Do not allow the reservoir to overfill while venting. Store the
unit in a carrying case or another safe location. The MECP10K
may be stored with the operating fluid in the reservoir. It is
recommended that the fluid being utilized be identified so that
another potential operator of the MECP10K will be advised.
WARNING:
• Never use the MECP10K to generate pressure
greater than 10,000 PSI/700 Bar
• Never connect the MECP10K to a pressure source
• The 1/8” FNPT Port should only be used with optional
Relief Valve P/N 1010050
WARRANTY
Martel Electronics Corporation warrants the MECP10K hydraulic
pump against defects in materials and workmanship for a period of
twelve (12) months after date of sale. An invoice may be required as
proof of purchase for warranty claims. Damage due to
contamination or other misuse is not covered under this warranty.
If any product is to be returned, a “Return Material Authorization”
number may be obtained by calling our Customer Service
Department toll free at 1-800-821-0023 in the US or Canada or
+1-603-434-8179 outside the US. Access our RMA Form by using
www.martelcorp.com, under Customer Service.
The MECP10K Hydraulic Hand Pump provides a hydraulic
pressure source up to 10,000 PSI. A wide variety of pressure media
may be used but distilled water is the preferred fluid. The seals are
Fluorocarbon and the only other non-metallic materials are plastic.
Viscous oils should not be used.
CAUTION: Hydrocarbon fluids should never be used in oxygen
service applications.
1. Isolate the device to be calibrated. CAUTION: Do not connect
to high pressure sources. Safe operating procedures must be
exercised to avoid personal injury and damage to the
calibration system.
2. Remove the fill cap and fill the reservoir approximately
1
/2 - 2/3
full. Select a fluid compatible with the system and the
MECP10K. Do not allow the fluid level to drop below the
reservoir. (Doing so will allow air to be introduced into the test
system).
3. Care should be exercised to minimize the total test system
volume. Connect the device(s) to be calibrated to the pump
port(s) using fittings or braided hose with the appropriate
pressure ratings.
4. Adjust the vernier control to the mid position and close the
needle bleed valve.
5. NOTE: Air in the system should be minimized because it can
effect calibration and high pressure gases can be dangerous.
It may be necessary to "prime" the test system to eliminate
gases. Repeat until bubbles are not seen in the reservoir.
6. Close the needle valve and squeeze the handles until the
approximate desired pressure is reached. The vernier control
can be used to "fine tune" the pressure. Systems with large
volume may require additional bleeding. Continue until all
pressure points are calibrated. As higher pressures are
achieved, pumping force can be reduced by reducing the
stroke length using the continuously variable stroke
adjustment.