Martel Electronics MC-1200 Data sheet

Multi-Function Calibrator
Multi-Function Calibrator
The Martel MC-1200 Multi-Function Calibrator provides a feature
set unmatched in high accuracy, hand-held calibrators in its price range. The
MC-1200 Calibrator provides the functions and accuracy associated with
fixed-installation, laboratory instruments, and has everything needed for
virtually any calibration task. Measure and source thermocouples, RTDs,
current, voltage, and frequency, and source pulse trains. A communications
is an isolated mA/V read-back circuit. Arrow keys, direct numeric keypad
entry, and three software-driven function buttons, plus a large backlit, menu-
driven graphics display combine to provide a highly intuitive, simple yet
powerful operator interface. Built-in 250 Ohm resistor for Hart™ compatibility,
compatibility with smart transmitters and PLCs, full fuseless protection, and a
serial communications port for full control with ASCII commands, are just some
MC-1200
MC-1200
of the additional features that make the MC-1200 the single, most indispensable
tool available for virtually any calibration task. The MC-1200 is supplied in a
tough, rubber boot; a carrying case is also available as an option.
Features:
Measure and source T/Cs (13 types), RTDs (13 types), Ohms, current, voltage, frequency; source pulse trains
Isolated mA/V read-back circuit for complete transmitter calibration
Communication port compatible with Fluke 700 Series pressure modules
Built-in 24 V supply can drive 4 – 20 mA loops up to 1000 Ohms
Direct entry of custom RTD coefficients (RO, A, B, C)
All source modes can be programmed with dedi­cated setpoints to speed calibration and linearity tests
Highest accuracy in class to 0.015% of reading
Meets CE requirements and is designed to IEC 1010 safety standards
Specifications (23 °C ±5 °C unless otherwise noted)
Voltage Read and Source
Source 0.000 to 20.000 VDC Read Isolated 0.000 to 30.000 VDC Non-isolated 0.000 to 20.000 VDC
Thermocouple mV
Read and Source -10.000 to +75.000 mV
Current (mA)
Source 0.000 to 24.000 mA Read Isolated 0.000 to 24.000 mA Non-isolated 0.000 to 24.000 mA
Frequency (1 to 20 V selectable amplitude)
CPM Source and Read 2.0 to 600.0 CPM Hz Source and Read 1.0 to 1000.0 Hz kHz Source and Read 1.0 to 10.00 kHz
Pulse (Source only; 1 to 20 V selectable amplitude)
Pulses 1 to 30,000.0 2 CPM to 10 kHz
Ohms
Source 5.0 to 4000 Ohms Read 0.00 to 4000.0 Ohms
Thermocouple Read and Source
J Thermocouple -200.0 to +1200.0 °C K Thermocouple -200.0 to +1370.0 °C T Thermocouple -200.0 to +400.0 °C E Thermocouple -200.0 to +950.0 °C R Thermocouple -20.0 to +1750.0 °C S Thermocouple -20.0 to +1750.0 °C B Thermocouple +600.0 to +1800.0 °C C Thermocouple 0 to +2316.0 °C XK Thermocouple -200.0 to +800.0 °C BP Thermocouple 0 to +2500.0 °C L Thermocouple -200.0 to +900.0 °C U Thermocouple -200.0 to +400.0 °C N Thermocouple -200.0 to +1300.0 °C
RTD Read and Source
Ni120 (672) -80.0 to +260.0 °C Pt100 (385) -200.0 to +800.0 °C Pt100 (3926) -200.0 to +630.0 °C Pt100 (3916) -200.0 to +630.0 °C Pt200 (385) -200.0 to +630.0 °C Pt500 (385) -200.0 to +630.0 °C Pt1000 (385) -200.0 to +630.0 °C Cu10 -100.0 to +260.0 °C YSI400 +15.00 to +50.00 °C Cu50 -180 to +200 °C Cu100 -180 to +200 °C Pt385-10 -200 to +800.0 °C Pt385-50 -200 to +800.0 °C
©Copyright 2003 Martel Electronics All 
Environmental
Operating Temperature -10 °C to +50 °C Storage Temperature -20 °C to +70 °C Stability ±0.005% of reading/ °C outside of 23 °C ±5 °C Power Requirements 6 VDC Batteries 4 AA; alkaline or optional rechargeable Physical (includes rubber boot) Dimensions 8.7”H x 4.2”W x 2.3”D (220.9 x 106.6 x 58.4 mm) Weight 30.5 ounces (863 gms) Accessories Included Test leads, 4 AA alkaline batteries, NIST Certificate, and instruction manual Optional Accessories Carrying Case, Model CC572
Accuracy
Voltage ±0.015% of reading, ±2 mV Thermocouple mV ±0.02% of reading, ±10 µV Thermocouple Errors (in °C; add 0.2 for Cold Junction Compensation error) Type J 0.2 °C Type K 0.3 °C Type T 0.2 °C Type E 0.2 °C Type R 1.2 °C Type S 1.2 °C Type B 1.2 °C Type C 0.6 °C Type XK 0.2 °C Type BP 0.9 °C Type L 0.2 °C Type U 0.25 °C Type N 0.4 °C
Read Source Current (mA) ±0.015% of reading, ±2 µA ±0.015% of reading, ±2 µA CPM ±0.05% of reading, ±1 LSD ±0.05% of reading Hz ±0.05% of reading, ±1 LSD ±0.05% of reading kHz ±0.05% of reading, ±1 LSD ±0.125% of reading Ohms works with all pulsed transmitters down to 5 ms 400 Ohm Range ±0.025% of reading, ±0.05 Ω ±0.025% of reading, ±0.05
Ω
RTD Read and Source Ni120 (672) 0.2 °C Pt100 (385) 0.2 °C Pt100 (3926) 0.2 °C Pt100 (3916) 0.2 °C Pt200 (385) 0.8 °C Pt500 (385) 0.4 °C Pt1000 (385) 0.2 °C Cu10 1.4 °C YSI400 0.1 °C Cu50 0.4 °C Cu100 0.3 °C Pt385-10 1.4 °C Pt385-50 0.4 °C
Fluke® Pressure Modules
The MC-1200 is compatible with Fluke 700 Series pressure modules. Fluke pressure modules are available from Martel Electronics. Please refer to the separate Pressure Modules data sheet for ranges and model numbers. (Pressure modules
require accessory Model 700MA Module Adapter.)
Loading...