Martel Electronics MC-1010 Operating Manual

MC1010
Reference Manual
MC1010 Reference Manual
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.1 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
1.3 Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2. Calibrator Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.1 Main Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
2.2 Menu Bar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
3. Using Measure Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3.1 Measuring volts and frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3.2 Measuring mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .12
3.3 Measuring mA with Loop Power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3.4 Measuring Temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3.5 Measuring Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4. Using Source Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
4.1 Setting 0% and 100% Output Parameters . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
4.2 Using the Automatic Output Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
4.3 Sourcing mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
4.4 Simulating a Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
4.5 Sourcing volts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
4.6 Sourcing frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19
4.7 Sourcing a pulse train . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
4.8 Sourcing Thermocouples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
4.9 Sourcing Ohms/RTDs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
5. Remote Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
5.1 Setting up the RS-232 Port for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
5.2 Changing Between Remote and Local Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5.3 Using Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5.4 Remote Commands and Error Codes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
5.5 Entering Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
6. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
7. Maintenance / Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
7.1 Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
7.2 Cleaning the Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
7.3 Service Center Calibration or Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
7.5 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .43
2
3
1. Introduction
The Martel MC1010 Multifunction Process Calibrator is a handheld, battery-operated instrument that measures and sources electrical and physical parameters. The calibrator has the following features and functions:
• A graphics display. The display can be used to source and measure volts, current, pressure, resistance temperature detectors (RTDs), thermocouples, frequency, and resistance, and to source pulse trains
• A thermocouple (TC) input/output terminal with automatic reference-junction temperature compensation.
• Five setpoints in each range for increasing/decreasing output
• An interactive menu
• Complete RS232 interface for remote control
1.1 Customer Service
Corporate Office:
www.martelcorp.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Dr.
Derry, NH 03038
1.2 Standard Equipment
Check to see if your calibrator is complete. It should include:
MC1010 Calibrator, Instruction Manual, Test Leads, Rubber Boot, NIST Certificate
1.3 Safety information
Symbols Used
The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Symbol Description
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
Canadian Standards Association
The following definitions apply to the terms “Warning” and “Caution”.
• “Warning” identifies conditions and actions that may pose hazards to the user.
• “Caution” identifies conditions and actions that may damage the instrument being used.
Use the calibrator only as specified in this manual, otherwise injury and damage to the calibrator may occur.
4
Warning
To avoid possible electric shock or personal injury:
• Do not apply more than the rated voltage. See specifications for supported ranges.
• Follow all equipment safety procedures.
• Never touch the probe to a voltage source when the test leads are plugged into the current terminals.
• Do not use the calibrator if it is damaged. Before you use the calibrator, inspect the case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
• Select the proper function and range for your measurement.
• Make sure the battery cover is closed and latched before you operate the calibrator.
• Remove test leads from the calibrator before you open the battery door.
• Inspect the test leads for damaged insulation or exposed metal. Check test leads continuity. Replace damaged test leads before you use the calibrator.
• When using the probes, keep your fingers away from the probe contacts. Keep your fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you disconnect test leads, disconnect the live test lead first.
• Do not use the calibrator if it operates abnormally. Protection may be impaired. When in doubt, have the calibrator serviced.
• Do not operate the calibrator around explosive gas, vapor, or dust.
• When using a pressure module, make sure the process pressure line is shut off and depressurized before you connect it or disconnect it from the pressure module.
• Disconnect test leads before changing to another measure or source function.
• When servicing the calibrator, use only specified replacement parts.
• To avoid false readings, which could lead to possible electric shock or personal injury, replace the battery as soon as the battery indicator appears.
• To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.
Caution
To avoid possible damage to calibrator or to equipment under test:
• Use the proper jacks, function, and range for your measurement or sourcing application.
• To avoid mechanically damaging the pressure module, never apply more than 10 ft-lb. of torque between the pressure module fittings, or between the fittings an the body of the module.
• To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the module.
• To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the pressure module documentation for material compatibility.
5
2. Calibrator Interface
Figure 1 shows the location of the input and output connections on the calibrator, while Table 1 describes their use.
Table 1: Input and Output Terminals
No. Name Description
1 TC input/output Terminal for measuring, or simulating thermocouples. Accepts
miniature polarized thermocouple plugs with flat in-line blades spaced 7.9 mm (0.312 in) center to center.
2,3 Source/Measure Terminals for sourcing and measuring voltage, frequency,
V,RTD 2W, Hz, pulse train, and RTDs
4,5 Source/Measure Terminals for sourcing and measuring current, and performing
mA terminals, 3W 4W RTD measurements with 3-wire or 4-wire setups.
6 Pressure module connector Connects calibrator to a pressure module for pressure meas-
urements.
7 Serial port Connects calibrator to a PC for remote control.
6
3
1
2
7
6
5
4
F1 F2 F3
CE ENTER
89
4
1
5
2
0
6
3
.
7
HOME
Figure 1. Input/Output Terminals
Figure 2 shows the location of the keys on the calibrator. Table 2 lists the functions of each key.
Table 2. Key Functions
No. Name Function
1 Function Keys F1, F2, F3 Used to operate the menu bar at the bottom of the calibrator
display. F1 is used for selecting options in the left box, F2 for the center box, and F3 for the right box.
2 Home Returns to home menu on the menu bar.
3 Power Turns calibrator on and off.
4 Cursor Control Key Left and right arrow keys are used to select which decade to
be changed in output value. Up and down arrow keys are used to increase, decrease, or ramp output value.
5 Numeric Keypad Allows user to enter Numeric values.
2.1 Main Display
Figure 3. Display
7
F1
89
4
1
5
2
0
6
3
.
F2 F3
7
CE ENTER
5
3
4
2
1
HOME
Figure 2. Keypad
RTD IN 4W P100-385
MENU
0.0 °C
1
4
5
7
6
2
3
LIGHT
The display of the calibrator is shown in Figure 3.
The menu bar is used to setup the display to perform the desired function.
Table 3 explains the different parts of the display:
Table 3: Display Functions
No. Name Description
1 Primary Parameters Determine what parameter is going to be measured or
sourced. The available options are: VOLTS, TC (thermocouple), RTD, FREQ (frequency), PULSE, PRESSURE, mA, and mA 2W SIM.
2 Input/Output control Switches between input mode (read), and output mode
(source).
3 Additional Settings Available only for TC (thermocouple), and RTD measurements.
For TC this setting turns the CJC (Cold Junction Connection) on and off. For RTD measure [RTD IN], this setting sets the number of wires used in the measurement (2-wire, 3-wire, or 4-wire)
4 Units Shows what unit the measurement or source value is in.
Available options are for RTD and TC (°C or °F), and for FREQ and PULSE (CPM, Hz, or KHz)
5 Sensor Types Allow for measurements to be made for different types of RTDs
and TCs. All types are shown in the Specifications. Also, dis­plays the amplitude of the pulse and frequency source, and pressure units.
6 Numeric Displays Display the numeric values of the signal being measured, or
sourced. An “OL” reading indicates an out of range or overload condition.
7 Secondary Valve Shows the base unit Ω for RTD or mV for TC.
2.2 Menu Bar
The parameters on the display are controlled by the menu bar, which is located at the bottom of the LCD. The function keys (F1, F2, and F3) are used to navigate through all the levels and choices of the menu bar. Refer to the menu tree for a clarification on the layout of all the levels.
The top level of the menu is the home menu. It can be accessed anytime by pressing the HOME key. There are three variations of the home menu: the input home menu, the output home menu, and the pulse home menu.
In the input home menu the only active options are [MENU] and [LIGHT]. The [MENU] option is used to enter the next level of the menu bar, the main menu. Press the corresponding function key (F1) to enter the main menu. The [LIGHT] option is used to turn on the LCD back light. Press the corresponding function key (F2) to turn on the back light.
8
9
In the output home menu there are three active options, [MENU], [LIGHT] and [STEP] or [RAMP]. The first two options work the same as in the input home menu. The third option is selectable in the Auto Function Menu and is used to turn on and off the selected auto function. See Section 4.2, Using the Automatic Output Functions. Also leaving this menu or pressing the Home button will stop the auto functions.
The pulse home menu also has three active options, [MENU], [TRIG], and [COUNTS]. The [TRIG] and [COUNTS] options are used for pulse simulation. The function of these options is explained in Section 4.2-6 (Sourcing a Pulse).
The next level of the menu bar is the main menu. The levels under the main menu depend on what mode the calibrator is in.
The main menu has two active options [CONFIG] and [MORE].
Choosing [CONFIG] calls up the parameter selection menu. [MORE] enters the next menu level.
The Auto Function Menu is the next menu if you are in source mode. Its options are [AUTO FUNC], [NEXT] and [DONE]. [AUTO FUNC] allows you to adjust the Automatic Output Function parameters. [NEXT] proceeds to the next menu level and [DONE] returns to the home menu. See Section 4.2,Using the Automatic Output Functions.
The contrast menu is usually the next menu level. Its options are [CONTRAST], [NEXT], and [DONE]. The [CONTRAST] option is used to adjust contrast. [NEXT] proceeds to the auto off main menu, and [DONE] returns to home menu. Contrast is adjusted using the arrow options, which are available after choosing [CONTRAST].
NOTE: The MC1010 calibrator offers a wide range contrast adjustment feature to accommodate operation in extreme temperatures.
In certain cases making large changes in contrast may render the display difficult to read under normal conditions. If this occurs and the display is too dim or dark to read, proceed with the following process to set the contrast back to a default setting.
1. Turn on the unit while holding down the “HOME” key.
2. Hold the key down for a count of 10 seconds to restore contrast default settings.
If the display is so dim that you cannot tell if the unit is on or off, use the backlight key to determine if the power is on or off.
CONFIG MORE
10
The auto off main menu contains the options [AUTO OFF], [NEXT], and [DONE].
The [AUTO OFF] option is used to turn the automatic shutoff on and off and set the amount of time the unit needs to stay dormant to shut off. [NEXT] and [DONE] both return to home menu.
When the calibrator is in the frequency or pulse mode, the frequency level menu is added after the main menu. The options available in this menu are [FREQ LEVEL], [NEXT], and [DONE]. The [FREQ LEVEL] option is used to adjust the amplitude of the wave. [NEXT] is used to access the contrast main menu, and [DONE] returns to the home menu.
When the calibrator is in RTD CUSTOM mode, the RTD custom setup menu, is inserted after the main menu. Options [SET CUSTOM], [NEXT], and [DONE] are available. [SET CUSTOM] is used to enter a custom PRT into the calibrator. Refer to Section 4.1-8a for instructions. [NEXT] is used to enter the contrast main menu, and [DONE] to return to the home menu.
The pressure zeroing main menu is the final variation to choosing [MORE] in the main menu.
It has the options [ZERO ], used to zero pressure, [NEXT] and [DONE], which have the same function as above. Refer to the Section 5.3 for instructions on zeroing.
The parameter selection menu is called up when [CONFIG] is selected from the main menu. It contains the following options: [SELECT], [NEXT], and [DONE]. When the display is selected, a parameter will start to flash. Use the [SELECT] option to change the parameter, and the [NEXT] option to switch to another variable. [DONE] returns to the home menu and enables the selected mode.
11
2.3 Cursor control / Setpoint control
The output value can be controlled by the four cursor control arrows on the keypad. By pressing one of the arrows a cursor will be added to the display under the last digit of the output value. The left and right arrow keys are used to select which decade to be changed in the output value. The up and down arrow keys are used to increase, decrease, or ramp the output value.
The menu bar will change to the setpoint menu with the touch of any one of the four arrow keys.
The three function keys are associated with 0, 25, and 100% values, respectively. 0 and 100% values can be stored by entering a value and then holding down the corresponding function key. The 25% key will then automatically step through the 25% values.
Figure 4. The Menu Tree
MENU
LIGHT
MENU
MENU
MENU
MENU
MENU
Home Menu
Home Menu
Home Menu
Home Menu
Home Menu
Home Menu
CONFIG
Selection Menu
MORE
SELECT
Parameter Selection
NEXT DONE
FREQ LEVEL
Frequency Level
NEXT DONE
SET CUSTOM
RTD Custom Menu
If calibrator is in Frequency Out or Pulse mode
NEXT DONE
ZERO
Pressure Zeroing
NEXT DONE
CONTRAST
Contrast Menu
NEXT DONE
AUTO OFF
Auto Off Menu
NEXT DONE
If calibrator is in RTD CUSTOM mode
If calibrator is in Pressure mode
LIGHT LIGHT LIGHT
LIGHT
LIGHT
AUTO FUNC
Auto Function Menu
NEXT DONE
If calibrator is in Source mode
12
3. Using Measure Modes
3.1 Measuring volts and frequency
The calibrator can measure volts or frequency on these jacks. To make the desired measurements, follow these steps:
1. Select the [CONFIG] option from the main menu.
2. Select the desired parameter for measurement.
3. Connect leads as shown in Figure 5.
Figure 5. Measuring Volts and Frequency with Input/Output Terminals
3.2 Measuring mA
To measure mA follow these steps:
1. Select the [CONFIG] option from the main menu.
2. Choose [mA] from the primary parameters.
3. Make sure the input/output control is set to IN.
4. Connect leads as shown in Figure 6.
Figure 6. Measuring mA with Input/Output Terminals
VOLTS IN
MENU
-0.001 V
LIGHT
MEASURE / SOURCE
3W mA+
TC
+ V
Hz –
Ω
4W mA–
+
MEASURE / SOURCE
3W mA+
TC
+ V
Hz
Ω
4W mA–
+
mA IN -25.00%
MENU
0.000 mA
LIGHT
3.3 Measuring mA with Loop Power
To test a 2-wire, loop powered transmitter that is disconnected from wiring, use the loop power function. This function activates a 24V supply in series with the current measuring circuit. To use this option proceed as follows:
To measure mA with loop power follow these steps:
1. Use the [CONFIG] option from the main menu to select mA IN LOOP.
2. Connect leads as shown in Figure 7.
Figure 7. Connection Using Current Loop
3.4 Measuring Temperature
3.4-1 Using Thermocouples
The calibrator supports the following thermocouple types: B, C, E, J, K, L, N, R, S, T, U, BP, and XK. The characteristics of all the types are described in Specifications section. The calibrator also has a Cold Junction Compensation (CJC) function. Normally this function should be ON and the actual temperature of the thermocouple will be measured. With CJC OFF, the calibrator will measure the difference between the thermocouple at the junction and at its TC input terminal.
Note: CJC off mode should only be used when calibration is being done using an external ice bath.
To use the thermocouple to measure temperature, follow these steps:
1. Attach the thermocouple leads to the TC miniplug, and insert the plug into the
input/output of the calibrator, as in Figure 8.
Note: For best accuracy wait 2 to 5 minutes for the temperature between the miniplug and the calibrator to stabilize before any measurements are taken.
2. Select the [CONFIG] option from the main menu.
3. Select TC from the primary parameters. Choose [IN] in the input/output control, and than
the thermocouple type from the sensor types. The temperature unit may also be changed from Celsius to Fahrenheit.
13
TC IN
CJ=24.9 °C
CJC ON
TYPE K
MENU
50.0 °C
LIGHT
+
MEASURE / SOURCE
3W mA+
TC
+ V
Hz
Ω
4W mA–
The calibrator can also measure the mV of a Thermocouple, which can be used along with a table in case the corresponding TC type is not supported by the calibrator. To do so, proceed as above and choose mV from sensor types.
Figure 8. Measuring Temperature Using Thermocouple Terminals
3.4-2 Using Resistance-Temperature-Detectors (RTDs)
The supported types of RTDs are shown in Section 8. Specifications. RTDs are characterized by their 0°C resistance, R0. The calibrator accepts two, three, and four wire inputs, with four wire input being the most accurate.
To use the RTD option, apply the following steps:
1. Select the [CONFIG] option from the main menu.
2. Select RTD from the primary parameters. Select [IN] from input/output control.
3. Choose 2, 3, or 4-wire connection [2W, 3W, 4W]. (4-wire allows for the most precise measurement)
4. Select RTD type from the sensor types.
5. Attach RTD leads as shown in Figure 9.
Figure 9. Measuring Temperature with RTD Connection
Resistance can also be measured using this function. To do so, use the above procedure and choose OHMS from the sensor types. This option can be used along with a table to measure an RTD which is not programmed into the calibrator.
14
MEASURE / SOURCE
3W mA+
TC
+ V
Hz –
Ω
4W mA–
+
+
+
+
3W
4W
RTD
RTD
RTD
2W
RTD IN 4W P100-385
100.00 Ω
MENU
0.0 °C
LIGHT
MEASURE / SOURCE
3W mA+
TC
+ V
Hz –
Ω
4W mA–
+
TC PLUG
Note: The TC wire used must match the thermocouple type being calibrated.
15
3.5 Measuring Pressure
Note: The MC1010 is compatible with BETA Calibrator Pressure Modules. The accessory BPPA-100 is required for pressure measurement.
Note: Pressure is not read from modules with frequency or pulse train mode enabled.
Note: On high pressure modules engineering units normally associated with low pressure
ranges such as, inH
2
O, cmH2O, etc. are not valid selections. Selecting one of this units with a
high pressure module attached will cause the display to read "----".
Warning!
To avoid a violent release of pressure in a pressurized system, shut off the valve and slowly bleed off the pressure before you attach the pressure module to the pressure line.
Caution
To avoid mechanically damaging the pressure module, never apply more than 10 ft-lb. of torque between the pressure module fittings, or between the fittings an the body of the module.
To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the module.
To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the pressure module documentation for material compatibility.
To measure pressure, follow these steps:
1. Connect the pressure module to the calibrator as shown in Figure 10.
Note: Make sure the calibrator is on before you plug in the pressure module.
2. Select the [CONFIG] option from the main menu.
3. Select [PRESSURE] from the primary parameters.
4. Select the desired measuring unit.
5. Zero the pressure module. The zero function on the calibrator can be found in the pressure zeroing menu.
Figure 10. Connections for Measuring Pressure
PRESSURE
MODULE
PRESSURE
MODULE
ADAPTER
(700mA)
VALVE
F1 F2 F3
CE ENTER
89
4
1
5
2
0
6
3
.
7
HOME
Loading...
+ 33 hidden pages