Using the Serial Communications Port on the MC-1000
1JPt100 (3926)
The MC-1000 calibrator is equipped with a DB-15 RS-232 serial communications port, located on the top of the calibrator.
Not e: The user will require a special RS-232 cable to connect
the calibrator to a computer or terminal.
Communication
Once the cable is attached, set the communications program set­tings as follows:
Stop Bits = 1 Parity = None
Flow Control = Xon/Xoff
The MC-1000 calibrator can be used with virtually any terminal emulation software. (i.e. Procomm or HyperTerminal).
Serial Communications Procedure for HyperTerminal Program
1. From the Windows® desktop, click on “Start”.
2. Click on “All Programs”. Click on “Accessories.”
3. Under the category “Communications”, click on “Hyper Terminal”.
4. Next appears a screen labeled: CONNECTION DESCRIP­TION. In the box titled: Name, enter a name for your program settings... example: “MC Connection”. In a box titled: Icon,
click on the you want to use. Click on the OK box to accept these entries.
5. Next appears a screen labeled: CONNECT TO. In the box titled: Connect Using, select the serial port that corresponds to
the computer serial port that you’ll be using to communicate to the MC-1000 calibrator. Click on the OK box to accept this
entry.
6. Next appears a screen labeled: COM X PROPERTIES (where X corresponds to the serial port selected in step # 6). Enter port
settings as follows:
(A.) Bits per second: 9600 (B.) Data bits: 8
(C.) Parity: None (D.) Stop Bits: 1
(E.) Flow Control: Xon/Xoff Click OK box to accept these entries.
7. Hook up the MC-1000 calibrator to the PC utilizing the special Martel Electronics RS-232 cable as follows:
(A.) Plug the 9 pin connector end of the cable to the
appropriate serial port on the computer.
(B.) With the MC-1000 calibrator switched off, plug the
other end of the cable to the DB-15 RS-232 port on the top of the MC-1000 calibrator.
8. Power up the MC-1000 calibrator per the instruction manual. You can now interact with the calibrator using the commands
shown in the table on the right. Note: You may wish to save the communications settings when exiting HyperTerminal. If so, click on the OK box when
prompted “Do you want to save X?”, where X corresponds to the name entered in Step # 4.
® Microsoft, Windows, Windows NT, and MS-DOS are registered trademarks of Microsoft Corporation. All other product names or services identified throughout this manual are registered trademarks or trademarks of their respective companies.
Computer Controlled Operation
The MC-1000 calibrator can be operated remotely via a computer or terminal. When connected to a computer, via the RS-232 serial
data port and appropriate cabling, the user may control the calibrator functions remotely by utilizing simple computer keyboard commands.
The following chart details which computer keys control which MC-1000 calibrator functions:
Serial Input Description
A mA measurement a mA source I mA loop
i mA 2W Sim V Volts measurement v Volts source
M mV measurement m mV source
E HVAC measure e HVDC measure
K Khz measurement k Khz source
H Hz measurement h Hz source
P CPM measurement p CPM source
W 2-wire measurement (Ohms and RTDs) X 3-wire measurement (Ohms and RTDs) Y 4-wire measurement (Ohms and RTDs)
T Thermocouple measurement (default Type
J) use "S" command to select sensor type
t Thermocouple source (default Type J) use
"S" command to select sensor type
C Selects Centigrade ( T/C - RTD ) F Selects Fahrenheit ( T/C - RTD ) R RTD measurement mode (default Pt 100
385) use "S" command to select sensor type
r RTD source mode (default Pt 100 385) use
"S" command to select sensor type
0-9 Enter a source value using ascii characters
-, <CR> 0,1,2,..,9,-,.terminated by < CR> (carriage return)
D Single Broadcast of most recent display
value and units
d Single Broadcast of most recent display
value
"S" COMMANDS SELECT SENSOR TYPE
Serial Input Selection Entry Description S No. Thermocouple Type RTD Type
2 K Pt100 (385) 3 T Pt100 (3916) 4 E Cu10 5 R Pt200 (385)
6 S Pt500 (385) 7 B Pt1000 (385) 8 L NI120 9 U YSI
A B
N mV
OHMS