Calibration Technology Starts Here
-NEW-
FEATURES
Precision Current Loop Calibrator with
HART Communications/Diagnostics
The Standard of Loop Calibration
just got better with additional
Knob). This combination dramatically reduces the time it takes to measure, or source
HART capability!
The LC-110H differs from the standard LC-110 in that it incorporates HART communications
and supports a select set of the HART universal and common practice commands. This unique
feature allows the LC-110H to be used as both a loop calibrator and communication tool.
In the communicator mode the user will be able to read basic device information, perform diagnostic
tests, and trim the calibration on most HART enabled transmitters. In the past, this could only be done with a
dedicated communicator, high-end multifunction calibrator costing thousands of dollars, or a laptop computer
with HART modem. The LC-110H will allow many more technicians to service and support HART devices.
In addition to HART communications, we also gave the user the ability to get information out of the LC-110H.
Need to quickly document the parameters of all the HART transmitters in your plant? Just add the
BetaLOG HART software/cable to capture and upload to twenty configurations in either
LC-110 & LC-110H
The Martel LC-110 and LC-110H are mA (loop) calibrators designed to
take the loop calibrator class to the next level. The new LC-110 series features
a user friendly interface with dedicated buttons and a rotary encoder (Quick-Set
voltage or current and power up a loop. The rugged case is contoured to easily fit a
technician’s hand and the large back lit graphics LCD is best in class.
(.csv) or (.txt) format.
We did not stop there! How about the ability to data log or record data on a
particular transmitter for troubleshooting? The data log tool features
selectable capture interval from 1 to 60 seconds and a logging
capacity of 9800 records or 99 individual sessions. Each data
sample contains the LC110H measurement, all four process
variables, and the standard status conditions
.
General Features
.
Best in Class Accuracy at 0.01% Reading
.
Small Rugged design operates on (6) standard AAA batteries
.
Intuitive interface that features a Quick-Set Knob
.
24VDC loop power with mA Measure Mode (-25% to 125%)
.
Resolution of 1µA on mA ranges and 1mV on voltages ranges
.
Built in selectable 250Ω Resistor for HART communications
.
Simple two wire connection for all measurements
.
Auto Shutdown to conserve battery life (adjustable up to
30 minutes)
.
Variable step & ramp time in seconds
.
Adjustable span selection (0 to 20mA or 4 to 20mA)
.
Valve Test (simulate defined mA values with % keys)
For more information:
- Call us toll free at 800-821-0023 or 603-434-1433
-
Fax us at 603-434-1653
- Or go to website, www. Martelcalibrators.com and click
on LC-110 & LC-110H.
LC-110 & LC-110H
Specifications
Model LC-110 & LC-110H
Functions: mA source, mA simulate, mA read, mA read/loop power,
and volts read.
Ranges: mA (0 to 24mA) and Volts (0 to 30VDC)
Resolution: 1uA on mA ranges and 1mV on voltage range
Accuracy: 0.01% +/- 2LSD all ranges (@23° +/- 5°C)
Operating Temp range: -10°C to 55°C
Humidity range: 10 to 95% non-condensing
Stability: 20ppm of F.S. /°C from -10°C to 18°C and 28°C to 55°C
Display: 128 x 64 pixels, LCD Graphic w/backlight, .34” high digits
Power: 6AAA alkaline, lithium, or NiMH batteries
Battery Life: ≥ 40 hours continuous use (measure mode using alkaline)
Loop Compliance Voltage: 24VDC @ 20mA
Over-Voltage Protection: 240VAC
Overload Current Protection: 28mA DC
EMC: EN61326 Annex A (Portable Instruments)
Dimensions (L x W x D): 6” x 3.6” x 1.3” (15 cm x 9 cm x 3 cm)
Weight: 9.5 ounces (0.3 kg)
Included Accessories: NIST traceable calibration certificate with data,
batteries, test leads, and manual
Additional Features (LC110H)
.
Built in HART modem for communication capability to
perform the following commands:
- Read Message, Descriptor,
Cal Date, Short and Long Tag
- Read Sensor PV and Output Information
- Write Short and Long Tag
- Write Descriptor and Message
- Select PV Unit
- Write PV Ranges (Upper and Lower)
- Enter/Exit Fixed Current Mode
- Set Zero Offset
- Trim D/A at 4mA & 20mA
.
The ability to store up to twenty HART device configuration files for
uploading via the BetaLOG Hart software. Configurations can be stored
as .csv or .txt files. This allows the end user to document the entire plant
for HART devices without spending thousands of dollars for plant
assetmanagement software.
NEW
Operation
The LC-110 & LC-110H utilizes a rotary encoder (Quick-Set
Knob) to set the output (mA) and also to select the function to
implement. The Quick-Set Knob features an integral push button
that you press to enter a selection. A dedicated menu key brings up
a list of functions. The Quick-Set interface highlights an item on that
list, and then pressing the knob (clicking it) selects that item to drive
further in the menu. The “Menu/Exit” key is used to go back to the
previous menu or exit from menu structure into measure/source mode.
All functions and HART commands are accessed in this manner. When
used as mA calibrator only, there are dedicated keys for automatic step
and ramp generation as well as keys to perform a quick zero and span
(4mA, 20mA and 25% steps).
LC110 Series Ordering Information
*Each unit comes standard with batteries, manual, test leads (banana to
alligator clips), and NIST Cert with Data
Part Number Description
1919974 LC110 Loop Calibrator
1920024 LC110H Loop Calibrator with HART™ Communications
1920050 BetaLOG HART Software with Lemo to USB Cable
6161102 Soft Carrying Case
1920051 LC110H Loop Calibrator Kit:
BetaLOG HART Software with Lemo to USB Cable
Soft Carrying Case
Banana to Mini-grabber® Test Leads
1920049 Lemo to USB Cable
5353050 Test Leads (1 Pair – Banana to Alligator)
5353093 Test Leads (1 Pair – Banana to Minigrabber®)
Martel Electronics Corporation
3 Corporate Park Drive, Unit 1
Derry, NH 03038
Phone: +1 (603) 434-1433
Fax: +1 (603) 434-1653
Copyright 2012 Martel Electronics;
All trademarks are the property of their respective owners. 1205R1CD-LC110
Specifications subject to change without notice