LC-100 Precision Loop Calibrator
Description
The LC-100 Precision Loop Calibrator provides
signifi cantly extended performance when compared
to any competitive calibrator. With an accuracy of
0.015% of reading, with 0.001 mA resolution, the LC100 has the highest accuracy in its class. Features such
as a “% Error” function, which eliminates manual error
calculation and allows the display of the actual versus
ideal error at any calibration point, put the LC-100 way
beyond similar instruments. The LC-100 can simulate,
power, and measure two-wire transmitters. With automatic Step and Ramp functions, the LC-100 enables
remote calibration of 4-20 mA devices.
Intuitive Interface
The LC-100 Loop Calibrator provides a highly intuitive user interface. A pair of cursor keys allows rapid
selection of the decade to be modifi ed, while a large
knob allows precise and rapid adjustment to any value
needed, with immediate feedback from the LCD display. A single MODE key cycles through and selects
the source, simulate, and measure modes.
Powerful Features
The LC-100 provides innovative, powerful features to
make your calibration job easier. A handful of clearly
marked keys provide access to the full functionality of
the unit. To enable remote calibration of devices and
processes, the LC-100 features automatic STEP and
RAMP output modes, which automate either a discrete
4-8-12-16-20-16-12-8-4 mA stair step output, or two
speeds of a linear 4-20-4 mA ramp output, respectively.
One or more simple pushes of the AUTO key (1) are
all that are needed to access this time-saving function.
To further aid remote calibration, the LC-100 features
a Minimum/Maximum value recall function. In the
measure mode, holding the AUTO(/MAX) key (1) provides a direct readout of the maximum value achieved,
while holding the 25%(/MIN) key (4) provides a direct
readout of the minimum value achieved. Holding
the % ERROR key (2) while turning the large knob
provides instant feedback of the difference between
the ideal and actual current values required for a given
device response. Instant verifi cation of whether the device is within its accuracy range is provided. No more
calculations are necessary!
Features
• Very high accuracy – ±0.015% of reading
• Very high resolution – 0.001 mA
• % Error function – eliminates manual error calculation
• Extended adjustment range—digital knob and
user-selectable decade output allow both large
and incremental (0.001 mA step) output changes
• Reads 0 to 28.000 VDC
• Built-in 250 Ohm resistor facilitates calibration of
Hart™ devices
• Remote calibration through Automatic Step and
Ramp functions
• Loop power measurement capability
• Simulates, powers, and measures two-wire transmitters
• Five preset outputs (4, 8, 12, 16, 20 mA)
• Fuseless input protection up to 250 VAC
• Supplied in full rubber boot
Martel Electronics Corporation LC-100 Precision Loop Calibrator
LC-100 Specifi cations (25 °C unless otherwise noted)
Input
Current Range 0.000 to 24.000 mA
-25.00 to +125.00 %
Voltage (Reading) 0 to 28 VDC
Input Protection fuseless, up to 250 VAC
Output
Source/Simulate Range 0.000 to 24.000 mA
-25.00 to +125.00 %
Drive Capability
Without Hart™ 1200 Ohm
With Hart™ 950 Ohm
Loop Supply Voltage 24 VDC
Range Select Decade; incremental to
0.001 mA steps
Resolution
Accuracy
1 µA
±0.015% of reading ± 2 µA
Environmental
Operating Temperature -10 °C to +55 °C
Storage Temperature -20 °C to +70 °C
Power Requirements
Battery 9V, one required; included
AC Adapter optional
9 VDC
Mechanical
Dimensions 5.7” H x 3.15” W x 1.43” D
(144.7 x 80.0 x 3.63 mm);
without boot
Weight 12 ounces (340 gms)
Accessories Included
Optional Accessories
Test leads, 9V battery,
NIST Certifi cate, &
instruction manual
Carrying case,
Model LCA-05A;
Rechargeable battery
Companion all-in-one instrument for
multi-function calibration
requirements.
Need a single instrument for current, voltage, frequency, temperature, RTD and pressure calibration? Martel offers its MC-1200
Multi-Function Calibrator for
virtually any high performance
calibration requirement.
Powerful Features (continued)
A LOOP POWER key (3) provides instant 24 VDC
loop power output.
The 25% (4) key provides instant outputs of 4, 8, 12,
16 and 20 mA loop current, which increments each
time the key is pushed.
Powerful features and an intuitive interface come together to provide a Loop Calibrator like no other in its
class. The LC-100.
Companion products for specifi c calibration needs.
Martel Electronics offers a dedicated instrument for thermocouple or pressure calibration.
Please request information on
either of these instruments.
Pressure
T140
Thermocouple
TC-100
Martel Electronics Corporation
PO Box 770
Londonderry, NH 03053 USA
Tel: 800-821-0023
Email: sales@martelcorp.com
Web: www.martelcorp.com
© Copyright 2003-2005 Martel Eectronics Corporation All trademarks are the property of their respective owners. 050502R0/OEMM/2.5M
BETA Calibrators is a Martel Electronics Corporation company brand. P/N 6800043, 5/05 Rev. A
Distributed by: