Martel Electronics LC-100 Data sheet

LC-100 Precision Loop Calibrator
Description
The LC-100 Precision Loop Calibrator provides signifi cantly extended performance when compared to any competitive calibrator. With an accuracy of
0.015% of reading, with 0.001 mA resolution, the LC­100 has the highest accuracy in its class. Features such as a “% Error” function, which eliminates manual error calculation and allows the display of the actual versus ideal error at any calibration point, put the LC-100 way beyond similar instruments. The LC-100 can simulate, power, and measure two-wire transmitters. With auto­matic Step and Ramp functions, the LC-100 enables remote calibration of 4-20 mA devices.
Intuitive Interface
The LC-100 Loop Calibrator provides a highly intui­tive user interface. A pair of cursor keys allows rapid selection of the decade to be modifi ed, while a large knob allows precise and rapid adjustment to any value needed, with immediate feedback from the LCD dis­play. A single MODE key cycles through and selects the source, simulate, and measure modes.
Powerful Features
The LC-100 provides innovative, powerful features to make your calibration job easier. A handful of clearly marked keys provide access to the full functionality of the unit. To enable remote calibration of devices and processes, the LC-100 features automatic STEP and RAMP output modes, which automate either a discrete 4-8-12-16-20-16-12-8-4 mA stair step output, or two speeds of a linear 4-20-4 mA ramp output, respectively. One or more simple pushes of the AUTO key (1) are all that are needed to access this time-saving function. To further aid remote calibration, the LC-100 features a Minimum/Maximum value recall function. In the measure mode, holding the AUTO(/MAX) key (1) pro­vides a direct readout of the maximum value achieved, while holding the 25%(/MIN) key (4) provides a direct readout of the minimum value achieved. Holding the % ERROR key (2) while turning the large knob provides instant feedback of the difference between the ideal and actual current values required for a given device response. Instant verifi cation of whether the de­vice is within its accuracy range is provided. No more calculations are necessary!
Features
• Very high accuracy – ±0.015% of reading
• Very high resolution – 0.001 mA
• % Error function – eliminates manual error calcu­lation
• Extended adjustment range—digital knob and user-selectable decade output allow both large and incremental (0.001 mA step) output changes
• Reads 0 to 28.000 VDC
• Built-in 250 Ohm resistor facilitates calibration of Hart™ devices
• Remote calibration through Automatic Step and Ramp functions
• Loop power measurement capability
• Simulates, powers, and measures two-wire trans­mitters
• Five preset outputs (4, 8, 12, 16, 20 mA)
• Fuseless input protection up to 250 VAC
• Supplied in full rubber boot
Martel Electronics Corporation LC-100 Precision Loop Calibrator
LC-100 Specifi cations (25 °C unless otherwise noted)
Input
Current Range 0.000 to 24.000 mA
-25.00 to +125.00 % Voltage (Reading) 0 to 28 VDC Input Protection fuseless, up to 250 VAC
Output
Source/Simulate Range 0.000 to 24.000 mA
-25.00 to +125.00 % Drive Capability Without Hart™ 1200 Ohm With Hart™ 950 Ohm Loop Supply Voltage 24 VDC Range Select Decade; incremental to
0.001 mA steps
Resolution Accuracy
1 µA ±0.015% of reading ± 2 µA
Environmental
Operating Temperature -10 °C to +55 °C Storage Temperature -20 °C to +70 °C
Power Requirements
Battery 9V, one required; included AC Adapter optional
9 VDC
Mechanical
Dimensions 5.7” H x 3.15” W x 1.43” D
(144.7 x 80.0 x 3.63 mm); without boot
Weight 12 ounces (340 gms)
Accessories Included
Optional Accessories
Test leads, 9V battery, NIST Certifi cate, & instruction manual
Carrying case, Model LCA-05A; Rechargeable battery
Companion all-in-one instrument for multi-function calibration requirements.
Need a single instrument for cur­rent, voltage, frequency, temper­ature, RTD and pressure calibra­tion? Martel offers its MC-1200 Multi-Function Calibrator for virtually any high performance calibration requirement.
Powerful Features (continued)
A LOOP POWER key (3) provides instant 24 VDC loop power output.
The 25% (4) key provides instant outputs of 4, 8, 12, 16 and 20 mA loop current, which increments each time the key is pushed.
Powerful features and an intuitive interface come to­gether to provide a Loop Calibrator like no other in its class. The LC-100.
Companion products for specifi c calibra­tion needs.
Martel Electronics offers a dedi­cated instrument for thermo­couple or pressure calibration. Please request information on either of these instruments.
Pressure
T140
Thermocouple
TC-100
Martel Electronics Corporation
PO Box 770 Londonderry, NH 03053 USA Tel: 800-821-0023 Email: sales@martelcorp.com Web: www.martelcorp.com
© Copyright 2003-2005 Martel Eectronics Corporation All trademarks are the property of their respective owners. 050502R0/OEMM/2.5M BETA Calibrators is a Martel Electronics Corporation company brand. P/N 6800043, 5/05 Rev. A
Distributed by:
Loading...