Martel Electronics EL-USB-4 User Manual

EL-USB-4 4-20mA Current Loop USB Data Logger
This data logger measures and stores up to 32,000 current loop readings over a 4-20mA d.c. measurement range. The user can easily set up the logging rate and start time, and download the stored data by plugging the module straight into a PC's USB port and running the purpose designed software under Windows 98, 2000 or XP. The data can then be graphed, printed and exported to other applications. The data logger is supplied complete with a long-life lithium battery. Correct functioning of the unit is indicated by a flashing red and green LED. The data logger features a pair of screw terminals and is is supplied complete with a set of measurement leads terminating in crocodile clips.
FEATURES
• 4-20mA d.c. Current Loop Measurement Range  Logging Rates between 1s and 12hr  Stores 32,000 readings  Connection via two screw terminals  USB Interface for Set-up and Data Download  User-Programmable Alarm Thresholds  Red and Green LED Status Indication  Replaceable Internal Lithium Battery
WINDOWS CONTROL SOFTWARE
Easy to install and use, the control software runs under Windows 98, 2000 and XP (Home and Professional Editions). It allows the user to set up and download any EL­USB-4.
ORDERING INFORMATION
Stock Number Standard Data Logger EL-USB-4 (Data Logger, Measurement Leads, Software on CD and Battery)
DATA LOGGER SET-UPS
 Logger Name  Logging Rate (1s, 10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)  High and Low Alarms  Start Date and Start Time
Replacement Battery BAT 3V6
SPECIFICATIONS
Specification Min. Typ. Max. Unit
4-20mA d.c. measurement range 0 4-20 mA d.c.
Internal resolution 0.05 mA d.c. Accuracy (overall error) 1 % 1 count Logging rate every 1s every 12hr ­Operating temperature range -35 (-31) +80 (176) °C (°F) 1/2AA 3.6V Lithium Battery Life* 1 Year
* depending on ambient temperature, logging rate and use of alarm LEDs.
Specifications liable to change without prior warning
LASCAR ELECTRONICS INC. 3750 West 26th Street Erie PA 16506 USA TEL: +1 (814) 835 0621
PO Box 770, Londonderry, NH 03053 1-800-821-0023
FAX: +1 (814) 838 8141 E-mail: us-sales@lascarelectronics.com
www.martelcorp.com
EL-USB-4 Issue 1 July/2005 M.C. Applies to EL-USB-4/1
+
+ +
LASCAR ELECTRONICS (HK) LIMITED FLAT C, 5/FL., LUCKY FTY. BLDG. 63-65 HUNG TO ROAD KWUN TONG KOWLOON HONG KONG TEL: +852 2797 3219 FAX: +852 2343 6187 E-mail: b4lascar@samsongroup.com.hk
EL-USB-4 4-20mA Current Loop USB Data Logger
LED FLASHING MODES
EL-USB-4 features a red and a green LED.
The LEDs on an EL-USB-4 will flash in one of the following ways, assuming alarm latching is turned off (exact timings to be confirmed).
- Green single flash, Logger primed and ready to start. every 30 seconds
- Green single flash, Logging in progress, last stored value OK every 10 seconds
- Red single flash, Logging in progress, the last stored reading is equal to or exceeds the low alarm level every 10 seconds (If latching is turned on, then a stored value is equal to or exceeds the low alarm level)
- Red double flash Logging in progress, the last stored reading is equal to or exceeds the high alarm level every 10 seconds (If latching is turned on, then a stored value is equal to or exceeds the high alarm level)
- Green single flash, Low battery, logging in progress, the last stored value OK every 20 seconds
- Red single flash, Low battery, logging in progress, the last stored value is equal to or exceeds the low alarm level every 20 seconds
- Red double flash, Low battery, logging in progress, the last stored value is equal to or exceeds the high alarm level every 20 seconds
- Green double flash, Logger is full, last stored value OK every 20 seconds
- Red/Green single flash, Logger is full, last stored value is equal to or exceeds the low or high alarm level every 20 seconds, alarm latching turned on
- No LEDs flash LEDs disabled during configuration to save battery power and increase logging duration.
or
Battery dead.
By default latching is disabled. The red LED will then no longer continue to flash after the logged curent has returned to normal. Instead, the green LED will then flash.
Latching can be turned on via the control software, so the red LED will continue to flash, even after the logged current has returned to normal. The red LED will effectively have latched into its alarm condition. This feature ensures that the user is notified that an alarm level has been exceeded, without the need to download the data from the logger.
PO Box 770, Londonderry, NH 03053 1-800-821-0023
Specifications liable to change without prior warning
www.martelcorp.com
EL-USB-4 Issue 1 July/2005 M.C. Applies to EL-USB-4/1