EL-USB-3
This data logger measures and stores up to 32,510 voltage readings over a 0-30V d.c. measurement range. The user can easily set up
the logging rate and start time, and download the stored data by plugging the module straight into a PC's USB port and running the
purpose designed software under Windows 98, 2000 or XP. The data can then be graphed, printed and exported to other
applications. The data logger is supplied complete with a long-life lithium battery. Correct functioning of the unit is indicated by
flashing red and green LED. The data logger features a pair of screw terminals and is is supplied complete with a set of measurement
leads terminating in crocodile clips.
Voltage USB Data Logger
FEATURES
• 0-30V d.c. Measurement Range
• Logging Rates between 1s and 12hr
• Stores 32,510 readings
• Connection via two screw terminals
• USB Interface for Set-up and Data Download
• User-Programmable Alarm Thresholds
• Red and Green LED Status Indication
• Replaceable Internal Lithium Battery
WINDOWS CONTROL SOFTWARE
Easy to install and use, the control software runs under
Windows 98, 2000 and XP (Home and Professional
Editions). It allows the user to set up and download any ELUSB-3. The latest version of the control software may be
downloaded from .www.lascarelectronics.com
ORDERING INFORMATION
Stock Number
Standard Data Logger EL-USB-3
(Data Logger, Measurement Leads,
Software on CD and Battery)
DATA LOGGER SET-UPS
• Logger Name
• Logging Rate (1s, 10s, 1m, 5m, 30m, 1hr, 6hr, 12hr)
• High and Low Alarms
• Start Date and Start Time
Replacement Battery BAT 3V6
SPECIFICATIONS
Specification Min. Typ. Max. Unit
0-30V d.c. measurement range 0 30 V d.c.
Internal resolution 50 mV d.c.
Accuracy (overall error) 1 %
Logging rate every 1s every 12hr Operating temperature range -25 (-13) +80 (176) °C (°F)
1/2AA 3.6V Lithium Battery Life* 1 Year
* depending on ambient temperature, logging rate and use of alarm LED.
Martel Electronics Corporation Tel: 800-821-0023
PO Box 770 Email: sales@martelcorp.com
Londonderry, NH 03053 USA Web: www.martelcorp.com
© Copyright 2005 Martel Eectronics Corporation All trademarks are the property of their respective owners. P/N 000000000 Rev. x
+
EL-USB-3
Voltage USB Data Logger
LED FLASHING MODES
EL-USB-3 features a red and a green LED.
The LEDs on an EL-USB-3 will flash in one of the following ways, assuming alarm latching is turned off (exact
timings to be confirmed).
- Green single flash, Logger primed and ready to start.
every 30 seconds
- Green single flash, Logging in progress, last stored value OK
every 10 seconds
- Red single flash, Logging in progress, the last stored reading is equal to or the low alarm level
every 10 seconds (If latching is turned on, then a stored value is equal to or exceeds the low alarm level)
- Red double flash Logging in progress, the last stored reading is equal to or exceeds the high alarm level
every 10 seconds (If latching is turned on, then a stored value is equal to or exceeds the high alarm level)
- Green single flash, Low battery, logging in progress, the last stored value OK
every 20 seconds
- Red single flash, Low battery, logging in progress, the last stored value is equal to or exceeds the low alarm level
every 20 seconds
- Red double flash, Low battery, logging in progress, the last stored value is equal to or exceeds the high alarm level
every 20 seconds
- Green double flash, Logger is full, last stored value OK
every 20 seconds
- Red/Green single flash, Logger is full, last stored value is equal to or exceeds the low or high alarm level
every 20 seconds,
alarm latching turned on
- No LEDs flash LEDs disabled during configuration to save battery power and increase logging duration.
or
Battery dead.
exceeds
By default latching is disabled, so the red LED will continue to flash, even after the logged voltage has returned to
normal. The red LED will effectively have latched into its alarm condition. This feature ensures that the user is
notified that an alarm level has been exceeded, without the need to download the data from the logger.
Latching can be turned on via the control software. The red LED will then no longer continue to flash after the
logged voltage has returned to normal. Instead, the green LED will then flash.
2 Martel Electronics Corporation DT-EL-USB3