Martel Electronics DPM970 User Manual

DPM 970
Digital AC Voltmeter Module
A 3 ½ digit LCD panel meter, featuring 19mm (0.75“) digit height and LED backlighting, the DPM 970 is a true RMS voltage and current meter with 500V a.c.
(2kV )and 2mA full scale readings as standard. By changing the value of scaling resistors or adding a shunt, this module will measure most a.c. voltages and currents. The DPM 970 features auto-zero and suppression of d.c. offset voltage. The digital hold function freezes the last displayed reading. Calibration is achieved via two multi-turn potentiometers for offset and scale. Connection to the meter is via screw terminals. The module is housed in a robust carrier which can be boltedinplace or panel mountedusingthe bezel, window andclips provided.
19mm (0.75“) Digit Height
Screw Terminal Connections
Digital Hold
500V a.c. (2kV ) Full Scale Reading
Measures Voltage and Current
LED Backlighting
SCALING
The value of the three scaling resistors RA1, RA2 and RB may be altered in order to change the full scalereading(F.S.R.)ofthe meter
- see table. voltage/current, short circuit ACV and COM andadjust the ZERO potentiometer. CAL potentiometer.
Required F.S.R. RA2 RBRA1
200mV 0R 0R Open 2V 910k 100k 20V 1M 10k 200V 1M 1k 2kV (factory default) 464K 536K 100R 2mA (factory default) - - 100R 20mA - - 10R 200mA - - 1R 2A - - 0.1R 20A - - 0.01R
NOTE
➀ ➁
➂ ➃
Other scaling factors can be achieved via the correct selection of RA1, RA2 and RB values.
To ensure that the meter reads "000" with zero input
The meter will need re-calibration by adjusting the
➀ ➀ ➀ ➀
0R 0R 0R
➀ ➁
➂ ➂ ➂ ➂
➂ ➃
Use inputs ACV and COM for measuring voltage. For measuring voltages above 500V RMS, external scaling resistors must be used. Use inputs ACI and COM for measuring current. We recommend the use of a current transformer when measuring in excess of 10 Amps. Consult current transformer manufacturer ’s dataforconnectionsand recommendations.
Stock Number
Standard Meter DPM 970
Specification Min. Typ. Max. Unit
Accuracy (overall error) * 0.5 1.5 % (±2 count) Full scale reading ** 0 1999 V Resolution 1 V Sample rate 3 samples/sec Operating temperature range 0 50 °C Temperature s 200 ppm/°C
tability Frequency range (input voltage) 15 15k Hz Input impedance (2kV range) 1 M DC Supply voltage 7.5 9 14 V d.c. Supply current (Backlighting off) 1.5 mA d.c. Supply current (Backlighting on) 50 mA d.c.
@ 5Vd.c.
* To ensure maximum accuracy, re-calibrate periodically. ** Maximum safe input voltage across the ACV and COM terminals is 500V .
If voltages to be measured exceed 500V , then fit scaling resistors
RMS
RMS
RMS
RMS
externally to the module.
DIMENSIONS
40.5 (1.59)
44.0 (1.73)
24.0 (0.94)
39.0 (1.54)
All dimensions in mm (inches)
76.0 (2.99)
72.5 (2.85)
51.0 (2.01)
61.5 (2.42)
51.0 (2.01)
71.5 (2.81)
24.0 (0.94)
Ø5.7
(0.22)
24.5 (0.96)
Ø3.0
(0.12)
a
b
c
d
a. 1.0 (0.04) b. 2.0 (0.08) c. 6.5 (0.26) d. 11.5 (0.45)
51 (2.01)
24 (0.94)
Display in Test Mode
Panel cut-out 72 x 40 (2.83 x 1.57) Maximum panel thickness 3mm (0.12).
TST
XDP
ALM
RB
RA2
RA1
ACV ACI COM COM V+ V- LP- LP+
WEBSITE: http://www.lascarelectronics.com/
REAR VIEW
V+
DP1
DP2
CAL ZERO
DP3
LASCAR ELECTRONICS LTD.
MODULE HOUSE, WHITEPARISH,
TEL: +44 (0)1794 884567 FAX: +44 (0)1794 884616
WILTSHIRE SP5 2SJ UK
E-mail: sales@lascar.co.uk
PO Box 770, Londonderry, NH 03053 1-800-821-0023 www.martelcorp.com
LASCAR ELECTRONICS INC.
3750 West 26th Street,
TEL: +1 (814) 835 0621 FAX: +1 (814) 838 8141
Erie, PA 16506 USA
E-mail: us-sales@lascarelectronics.com
LASCAR ELECTRONICS (HK) LIMITED
FLAT C, 5/FL., LUCKY FTY. bldg., 63-65 HUNG TO ROAD
KWUN TONG, KOWLOON, HONG KONG
TEL: +852 2797 3219 FAX: +852 2343 6187
E-mail: b4lascar@samsongroup.com.hk
SCREW TERMINAL FUNCTIONS
ACV Alternating voltage measurement input, acceptsup to500Va.c. By changing the values ofscalingresistors RA1, RA2 and RB, other voltage and
COM Return line forvoltage measurementinput(ACV) or currentmeasurement (ACI)input. COM Return line forvoltage measurementinput(ACV) or currentmeasurement (ACI)input.
LP+ Positivesupplyfor LED backlighting.
HLD Digital Hold option. Connect toV+to hold last displayed reading indefinitely. ALM Alarm “!” annunciator.Connect to XDP to display.
XDP Annunciator drive wavefor m. Used to drive alar m annunciator (ALM) ifrequired.
current ranges can be achieved.Tomeasure in excess of 500V a.c., remove RA1, RA2 and RB from the meter and externally;linkRA1 and RA2 on the meter. See scaling table for suitable valuesforRA1, RA2 and RB.
ACI Alternating current measurement input.
See Scaling table for suitable values for shunt resistor RB.
V+ Positivepowersupply input.
V- Negative power supply input.
LP- Negative supply for LED backlighting.
TST Connect to V+ to displayallsegments exceptdecimal points and annunciators.
It should not be operated for more than afewseconds as the DC voltage applied to the LCD may “bur n” the display.This connection is normally at 5V below V+ and is the ground for the digital section ofthemeter.
V+ Auxilliary V+ for connections to decimal points DP1, DP2 or DP3. DP1 199.9 Connect to V+ to display required decimal point. DP2 19.99 Connect to V+ to display required decimal point. DP3 1.999 Connect to V+ to display required decimal point.
SAFETY WARNING
Tocomply with the Low VoltageDirective (LVD93/68/EEC), input voltages to themodule’s connectors must not exceed500VRMS. If voltages to be measured exceed 500VRMS, then fit scaling resistors externally to the module. If the mains power supply is connected to the module, then all the module’s connectors must be considered to be floating at mains potential. Consequently due caution must be exercised when connecting to the module’s connector functions under these conditions. The user must ensure that the incorporationof the DPM into the user’s equipment conforms tothe relevant sections of BS EN 61010 (Safety Requirements for Electrical Equipmentfor Measuring, Control andLaboratory Use).
APPLICATIONS
WARNING : Do not apply in excess of 500V a.c. to the meter.
Voltage
Current
Voltage
Current
Voltage
Current
V+
V-
V+
V-
V+
V-
V+
V-
V+
V-
V+
V-
+5V
-5V
+5V
-5V
+5V
-5V
110/240V
PSU 303
0V
110/240V
PSU 303
0V
110/240V
PSU 303
0V
Phase 1
Phase 2
Phase 3
Neutral
Load
Load
Load
ACV
COM
ACI
COM
ACV
COM
ACI
COM
ACV
COM
ACI
COM
3-Phase Current and Voltage
Measurement
Phase
Neutral
Phase
Neutral
Phase
LP+
V+
V-
LP-
Neutral
Powering the LED backlighting on a DPM 970
from the PSU 303's +5V and COM outputs.
Voltage
Load
ACV
COM
ACI
COM
Current
V+
V-
V+
V-
Single Phase Current and Voltage
Measurement
Load
ACI
CT
COM
V+
V-
Using a current transformer (CT) to
measure currents in excess of 10A a.c.
+5V
COM
-5V
+5V
-5V
+5V
-5V
110/240V
PSU 303
0V
110/240V
PSU 303
0V
110/240V
PSU 303
0V
Specifications liable to change without prior warning DPM 970 Issue 5 August/2002 M.C. Applies to DPM 970/4
Loading...