1. Introduction
The BetaProbe TI/TI+ is designed to be a high precision digital
thermometer that can be used in place of mercury liquid in glass (LIG)
thermometers. It can also be used as a reference standard to check the
accuracy of other types of digital or analog thermometers. Because of its
intrinsically safe certification it can be used in many of the same
hazardous locations as analog thermometers.
The probe houses a fast responding thin film (TI) or wire wound (TI+)
sensor that allows very accurate readings to be taken quickly and with
minimal insertion depth.
A rotatable probe mount allows for both horizontal and vertical
orientation of the display module.
A unique feature of the BetaProbe TI/TI+ is the trend indication arrows
which the user can configure to show when the readings are considered
to be stable enough to record the reading.
Additional features such as min/max storage, selectable sample rate,
auto-shutoff, backlight and damping allow the user to tailor the TI/TI+ for
their specific application.
Battery life is up to 300 hours. With the auto-shutoff feature the typical
user may only have to change the batteries a few times per year.
1.1 Customer Service
Corporate Office:
www.martelcalibrators.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Drive
Derry, NH 03038
1.2 Standard Equipment
Check that your BetaProbe TI/TI+ has arrived intact. Batteries are
factory installed. Save the packing materials at least until you have
verified that there is no concealed damage.
1.3 Safety information
A Warning identifies conditions and actions that pose hazard(s) to the
user; a Caution identifies conditions and actions that may damage the
Calibrator or the equipment under test.
1
Symbols Used
The following table lists the International Electrical Symbols. Some or all
of these symbols may be used on the instrument or in this manual.
Symbol Description
Power OFF
Power ON
Earth ground
This unit was designed to meet the above requirements using:
EN 60079-0:2006 ELECTRICAL APPARATUS FOR
EXPLOSIVE GAS ATMOSPHERES -- PART
0: GENERAL REQUIREMENTS (IEC
60079-0:2004 (MOD))
EN 60079-11:2007
EXPLOSIVE ATMOSPHERES -- PART 11:
EQUIPMENT PROTECTION BY INTRINSIC
SAFETY "I" (IEC 60079-11:2006 (EQV) +
CORRIGENDUM DEC. 2006 (EQV))
Risk of Danger. Important information. Refer to manual.
Battery
Hazardous Voltage
Conforms to ATEX requirements
Conforms to relevant European Union directives.
Wheeled bin, conforms to EC directive 2002/96/EC
Hazard Location Information/Approvals
Ex Hazardous Areas
An Ex-hazardous area as used in this manual refers to an area made
hazardous by the potential presence of flammable or explosive vapors.
These areas are also referred to as hazardous locations.
Ex ib IIB T4 Gb (–10°C ≤ Ta ≤+50°C)
ITS10ATEX27114X
Ex ib IIB T4 Gb
II 2 G IECEx ITS10.0049X 0344
Manufactured by Martel Electronics, Inc.,
3 Corporate Park Dr.
Derry, NH, USA
Special Conditions for Safe Use:
Misuse
Should the BetaProbe TI/TI+ be overheated or exposed to sudden
physical shock (i.e. being dropped) it should be examined for any
damage that may cause a safety concern. If in doubt please return the
unit for evaluation to Martel Electronics. Please refer to the Customer
Service Section for contact information.
WARNING
Check for proper operation of the instrument before taking it into a
hazardous area. Should the instrument not operate as expected do
not take the instrument into a hazardous area and return it to the
manufacturer for inspection.
WARNING
The BetaProbe TI/TI+ is intended for installation only in locations
providing adequate protection against the entry of solid foreign objects
or water capable of impairing safety.
Caution
To avoid possible damage to calibrator or to equipment
under test:
If the message changes to “OL” the range limit is exceeded and the
BetaProbe TI/TI+ must immediately be removed from the heat source to
prevent damage.
2 3