BetaGauge Pressure Modules
Description:
Martel Electronics offers 29 standard pressure
modules, covering gauge, vacuum, absolute,
compound, and differential measurements. All
modules are directly compatible with the BetaGauge
II. With the Model BPPA-100 BETA Port Pressure
Adapter, all modules are fully compatible with
the Martel MC-1200 and MC-1000 Multi-Function
Calibrators, the BetaGauge 330, 321, 311 and 301
Pressure Calibrators, the DMC-1400 Documenting
Multi-Function Calibrator, and the Martel Electronics
M2001 and 3001 Laboratory/Bench Standards.
Pressure ranges may be displayed in any of 13 userselectable units. The choice of pressure unit may be
restricted by limitations on resolution of the
instrument display of the particular calibrator the
module is used with.
Specifications subject to change without notice
High Accuracy Digital Modules
Features:
• 29 standard ranges
• Gauge, vacuum, absolute, compound, and dif-
ferential measurements
• Accuracy specifi ed over 15 °C to 35 °C range
• Isolated and non-isolated measurements, range
dependant
Description:
Martel Electronics offers 29 standard pressure
modules, covering gauge, vacuum, absolute,
compound, and differential measurements. All
modules are directly compatible with the BetaGauge
II, as is a DC measurement module.
With the Model BPPA-100 BETA Port Pressure
Adapter, all modules (with the exception of the
DC measurement model) are fully compatible with
the Martel MC-1200 and MC-1000 Multi-Function
Calibrators, the BetaGauge 321, 311 and 301 Pressure
Calibrators, the DMC-1400 Documenting Multi-
Function Calibrator, and the Martel Electronics M2001
and 3001 Laboratory/Bench Standards.
Pressure ranges may be displayed in any of 13 user-
selectable units. Water density correction factors
of 4 °C, 20 °C, or 60 °F can be selected for either
water column unit. The choice of pressure unit may
be restricted by limitations on resolution of the
instrument display of the particular calibrator the
module is used with.
For optimum mechanical
strength, external pressure
connection is made by a
FNPT 316SS connector
welded to a stainless steel
metal plate.
1
Model BPPA-100 BETA Port
Pressure Adapter
/8”
BetaPort-P Pressure Module Specifications (all % of Full Scale)
Parameter/Range Positive1 Vacuum
Accuracy Accuracy Over-Pressure Notes
Isolated Gauge (PSIG)2
0 to 15 (0 to 1 Bar) ±0.025% 300 %
0 to 30 (0 to 2 Bar) ±0.025% 300 %
0 to 500 (0 to 35 Bar) ±0.025% 200 %
0 to 1000 (0 to 70 Bar) ±0.025% 200 %
0 to 1500 (0 to 100 Bar) ±0.035% 200 %
0 to 3000 (0 to 200 Bar) ±0.05% 200 %
0 to 5000 (0 to 340 Bar) ±0.05% 200 %
0 to 10000 (0 to 700 Bar) ±0.1% 120 %
Isolated Absolute (PSIA)
0 to 15 (0 to 1 Bar) ±0.04% 300 %
0 to 30 (0 to 2 Bar) ±0.025% 300 %
0 to 50 (0 to 3.5 Bar) ±0.03% 300 %
0 to 100 (0 to 7 Bar) ±0.025% 300 %
0 to 300 (0 to 20 Bar) ±0.025% 200 %
Non Isolated Compound (PSIG)
-0.4 to 0.4 (-20 to 20mBar) ±0.1% ±0.15% 400 %
-1 to 1 (-70 to 70 mBar) ±0.05% ±0.1% 400 %
-5 to 5 (-350 to 350 mBar) ±0.075% ±0.1% 400 %
-7.2 to 7.2 (-500 to 500 mBar) ±0.07% ±0.1% 300 %
-10 to 10 (-700 to 700 mBar) ±0.03% ±0.05% 300 %
-15 to 15 (-1 to 1 Bar) ±0.04% ±0.04% 300 %
-15 to 30 (-1 to 2 Bar) ±0.025% ±0.025% 300 %
Isolated Compound (PSIG)
-12 to 50 (-0.8 to 3.5 Bar) ±0.03% ±0.03% 300 %
-12 to 100 (-0.8 to 7 Bar) ±0.025% ±0.025% 300 %
-12 to 150 (-0.8 to 10 Bar) ±0.03% ±0.03% 200 %
-12 to 300 (-0.8 to 20 Bar) ±0.025% ±0.025% 200 %
Differential (PSID)
0 to 5 (0 to 350 mBar) ±0.075% 400 % 3
0 to 30 (0 to 2 Bar) ±0.025% 300 % 3
0 to 50 (0 to 3.5 Bar) ±0.03% 300 % 3
2
2
2
2
Notes:
1. Accuracy is percent of full scale range, over the 15 °C to 35 °C temperature range. Includes the pressure/temperature hysteresis in psi.
The accuracy statement shown in the specification table is the base accuracy from 15 °C to 35 °C. Outside this temperature range, add
an additional ±0.0015% of FS per °C. (For the 0.3 and 1psi ranges add an additional +
deviation of a particular BetaPort-P Pressure Module, use the following formula: Deviation = ±%FS , ±T/P H, ±tempco Where ±T/P H =
thermal/pressure hysteresis in psi where applicable, And ±tempco = ±0.0015% FS/°C when the temperature is outside the 15-35 °C
temperature range
2. The Gauge, Vacuum, and Compound type range measurements are relative to atmospheric pressure. The Absolute type is a
measurement made relative to absolute zero (perfect vacuum). The Differential type is a measurement made relative to the pressure
applied to the low-pressure port of the module.
3. The maximum static pressure is 200 psig (14 bar).
3 Corporate Park Drive, Unit 1, Derry, NH 03038 USA
Tel: +1 (603) 434-1433 • Fax: +1 (603) 434-1653 • E-mail: sales@martelcorp.com • Internet: www.martelcalibrators.com
p/n 0219410, rev. D, 11/09
0.005% of FS per °C) To calculate the allowed