Martel Electronics BetaGauge PIR PRO User Manual

BetaGauge PIR/PIR PRO
Reference Class
High Accuracy Digital
Pressure Test Gauge
User Reference Manual
This
document
includes information
CSA/ATEX certified
versions of the BetaGauge PI.
1. Introduction
The BetaGauge PI Reference Class (PIR) combines the high accuracy of digital electronics with the convenience and ease of use of an analog test gauge. Accurate to ± 0.04% of reading, the BetaGauge PIR can be used as a calibration reference, or in any application where high accuracy pressure measurement is required.
Many user configurable functions have been designed into the BetaGauge PIR including sampling rate, TARE, damping, auto shut off, and min-max. Once the gauge is configured, settings can be locked and password protected to prevent unauthorized changes to configuration.
1.1 Customer Service
Corporate Office:
www.martelcalibrators.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Drive
Derry, NH 03038
1.2 Standard Equipment
Check to see that your BetaGauge PIR has arrived intact. Batteries are factory installed unless you have purchased the optional 24V powered version, in which case batteries are not supplied or installed. Save the packing materials at least until you have verified that there is no concealed damage.
1.3 Safety information
A Warning identifies conditions and actions that pose hazard(s) to the user; a Caution identifies conditions and actions that may damage the Calibrator or the equipment under test.
Symbols Used
The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
1
Symbol Description
Power OFF
Power ON
Earth ground
Risk of Danger. Important information. Refer to manual.
Battery
Hazardous Voltage
Conforms to ATEX requirements
Conforms to relevant European Union directives.
Wheeled bin, conforms to EC directive 2002/96/EC
Hazard Location Information/Approvals
Ex Hazardous Areas
An Ex-hazardous area as used in this manual refers to an area made hazardous by the potential presence of flammable or explosive vapors. These areas are also referred to as hazardous locations, see NFPA 70 Article 500.
®
LR110460
Class I, Div. 2, Groups A-D
WARNING
Only gauges powered by batteries are approved for use in hazardous areas. 24V versions are not approved for hazardous use.
2
II 3 G EEx nA IIB T6 KEMA 06ATEX0014 X Ta=–10°C... +55°C
Certified by CSA as conforming to relevant Canadian and USA standards
Special Conditions for Safe Use:
Misuse
Should the BetaGauge PIR be exposed to overpressure or sudden physical shock (i.e. being dropped) it should be examined for any damage that may cause a safety concern. If in doubt please return the unit for evaluation to Martel Electronics. Please refer to the Customer Service Section for contact information.
WARNING
The Beta Gauge PIR is not intended for use with flammable substances and is intended for installation only in locations providing adequate protection against the entry of solid foreign objects or water capable of impairing safety.
Caution
To avoid possible damage to calibrator or to equipment
under test:
• If the message changes to “OL” the range limit is exceeded and the pressure source must immediately be removed from the BetaGauge PIR to prevent damage to the pressure transducer inside.
• Maximum torque allowed is 13,5 Nm = 10 ftlbs. NEVER exceed the torque allowed.
2. BetaGauge PI Reference Class Display and Controls
3
Pressure display
Engineering Units
0 - 100% bar graph
Selects an action in configuration. Also turns the back-light on/off
Turns Power on/off
Enters set-up and
configuration menus
Zeros the display,
Scrolls forward through
menu displays
Recalls min and max
readings.
Scrolls backwards
through menu displays
3. Operation
Power: The standard BetaGauge PIR is supplied with 3 AA batteries installed. If you purchased the optional 24Volt powered version, batteries are not installed. Connect a 24V power supply to the terminal block on the rear of the gauge, noting proper polarity. Do not install batteries when external power will be used.
Push the power button momentarily to turn the unit on. Push it again to turn it off.
Set-up and configuration:
Push the CONFIG button to access the user-settable functions on the gauge. Each time the CONFIG button is pressed; the display advances to the next function. Once a function has been set, press ENTER to exit the configuration menu, or CONFIG to continue with further configuration. In order, the configuration menu and operation is as follows:
1. Engineering Units set. The unit is shipped configured to
display PSI. By pressing the and (ZERO and MAX/MIN) keys you can scroll forward and backwards through the 18 standard engineering units plus a one custom unit/scale. When the desired unit is displayed, press ENTER or CONFIG. Pressure will now be displayed in the chosen engineering units. See the Specifications section of this manual for a list of available engineering units. See the Supervisory Mode section for details on setting up custom units.
2. Set Auto Off. The auto-shut off can be set in 1 minute
increments from 1 to 30 minutes or “off” (continuous operation). The unit is shipped set for 30 minutes. Use the and keys to set the desired interval. The “off” setting is at the low end of the choices, below 1 minute.
3. Display battery voltage. Actual voltage and a percent of life
bargraph indicate battery condition
4. Display actual temperature. The BetaGauge PIR is
temperature compensated, this displays the temperature measured by the internal sensor. The value can be set to degrees F or degrees C using the arrow keys.
5. Set damping. Choices are “on” and “off” set with the and
keys. Turning damping on will smooth readings from pulsating pressure sources.
4
6. Set sample rate: This determines how often pressure is
sampled and the display is updated. Choices are 0.5, 1, 3, and 10 samples/second. Note that 10/sec provides the fastest response time.
7. Set TARE. This allows you to set a constant offset value,
which is then subtracted from the measured pressure. For example if a TARE is set at 30 PSI, and the measured pressure is 37 PSI, the displayed value will be 7 PSI. A pressure of 27 PSI would be displayed as –3 PSI.
The tare value is set manually with the ▲ and ▼ keys, and is based on the engineering units and resolution selected for display. TARE value can be set to the maximum range of the gauge.
The bar graph will always display the actual pressure based on the full range of the gauge regardless of the tare setting. This is done for safety to insure that even with a “0” reading that pressure is being applied to the gauge.
8. Function Lock: Access to each of the settable parameters
above can be turned “off” once set, to prevent unauthorized changes to configuration. This is accomplished through a password protected “supervisory mode”. Press ENTER to access the supervisory mode, or CONFIG to return to normal operation.
4. Supervisory Mode
Press ENTER when “FUnC LOCK” is displayed, 0PWRD will be displayed on the gauge. The password to enter supervisory mode is 101, set using the ▲ and ▼ keys. Holding a key continuously will cause the display to advance more quickly for faster setting. The password is factory set and cannot be changed
1. Your BetaGauge PIR is shipped from the factory with all setting access “unlocked” or available to be changed.
2. In supervisory mode each of the parameters can be locked or unlocked using the ▲ and ▼ keys. Select LOC (lock) for those parameters you do not want to be accessible, and UnLOC (unlock) for those can be accessed.
3. In order, the functions that can be unlocked, locked or accessed are:
• Zero function (enable/disable)
• Set pressure units (enable/disable)
5
• Auto shutdown adjustment (enable/disable)
• Damping settings (enable/disable)
• Sample rate setting (enable/disable)
• Tare setting (enable/disable)
• Custom engineering units (set scale factor)
4. Use the CONFIG key to scroll through the above choices, and the and keys to lock and unlock features. Press CONFIG to continue scrolling through the parameters, pressing ENTER at any point saves your settings and returns the gauge to normal operation.
When a function is “locked, it cannot be accessed or changed from its current state. To change a locked function, enter the supervisory mode, and unlock the function. Once it is changed, you may enter supervisory mode to lock access again.
5. Setting a custom engineering unit or scale: The last menu choice in supervisory mode is SET FACTR. This allows you to set a multiplier factor from 0.001 to 100, creating a custom scale. The set factor will be multiplied by the PSI measured, the result will be displayed.
For example: 40 PSI is the equivalent of 1000 lbs of product in a tank. You want to display the product weight , using a 100 PSI gauge. By setting a factor of 25, a 40 PSI pressure would display as 1000 (40 x 25). The engineering unit displayed on the BetaGauge PIR will be “Cust”.
5. Normal Operation
Turning the backlight on and off: Press the ENTER button.
Zeroing the display: Press and hold the ZERO button.
MAX/MIN: The BetaGauge PIR stores minimum and maximum
pressure values in memory. Pressing the MAX/MIN button once will display the minimum pressure from memory. Pressing the MAX/MIN button again will display the maximum pressure from memory. After about 2 seconds, the gauge returns to normal (live display) operation. To clear the MAX/MIN memory registers, press and hold the MAX/MIN button for 2 or more seconds until “CLr” is displayed.
The analog bar graph at the bottom of the display indicates the applied pressure level relative to the full range of the gauge. Keep in mind that if a TARE value has been programmed into the
6
Loading...
+ 16 hidden pages