Martel Electronics BetaGauge 330 Data sheet

Calibration Technology Starts Here
Hand Held Pressure Calibrator with Integral Electric Pressure Pump
The new
etaGauge 330, technicians will have a small, lightweight hand held calibrator that
standard of pressure calibration
Before the BetaGauge 330, users had to manually generate pressures with tedious and hard to use
manual pumps or use large, heavy and unwieldy boxes with limited portability and limited battery life.
at a mere 2.5 pounds. The housing is contoured to easily fit a technician’s hand with good balance top to bottom.
Performance doesn’t stop with the pump. The BetaGauge 330 provides ±0.025% of full scale accuracy on its
internal, isolated pressure sensor. Temperature compensation on its internal sensor ensures accuracy in field
applications. An external pressure module connection supports all 29 ranges of BetaPort-P
B
generates pressures up to 150 PSIG using a high performance integral electric pump.
BetaGauge 330
Martel Electronics Corporation introduces a revolution in pressure
calibration technology for the process industries. With the release of its new
pressure modules for
even greater measurement capability
Continued on back.
General Features
.
Integrated internal electrically operated pneumatic pump: generates pressure to 150 PSI and vacuum to -12 PSI (-1 to 10 Bar, -10 to 100 kPa)
.
Accuracy to 0.025% FS
.
Source/measure 4-20 mA
.
Small, rugged compact design operates on eight (8) standard Alkaline AA batteries
-
Only 8” high x 4” wide x 2-3/8” deep (20 cm x 10 cm x 6 cm)
- Lightweight at just 2.5 pounds (1.2 kg)
.
External pressure module interface supports all 29 BetaPort-P compatible pressure modules (requires optional BPPA-100 module adapter)
.
Pt100 RTD input for temperature measurement, accurate to
0.1°C (0.2°F) (measurement only)
.
24 V loop power to power device under test
.
Store and recall up to 5 user defined setups
.
Large graphic style LCD with backlight can be configured to display up to 3 variables displayed simultaneously (user controlled)
.
Ideal for gas flow calibration (custody transfer) applications
.
Pressure displayed in the user’s choice of 19 engineering units
.
Switch test is automatic using internal sensor or external pressure module
.
IP56 Rated
.
Percent error function
Patent Pending
For more information:
- Call us toll free at 800-821-0023 or 603-434-1433
-
Fax us at 603-434-1653
- Email us at sales@martelcorp.com and ask for a BetaGauge 330 Pressure brochure
- Or go to our website, www.martelcalibrators.com and click on BetaGauge 330
BetaGauge 330
Specifications
anges
R
Available Pressure
PSIG: -12 to 30 PSI /-0.8 to 2 bar
-12 to 150 PSI /-0.8 to 10 bar (electric pump)
-12 to 300 PSI / -0.8 to 20 bar
A 0 to 24 mA, max, load 1000 Ohms
m RTD Pt100, -50 to 150˚C Engineering Units
psi, bar, mbar, kPa, kgcm2, cmH2O@4°C, cmH2O@20°C, mH2O@4°C, mH2O@20°C, inH2O@4°C, inH2O@20°C, inH2O@60°F, mmHg@0°C, inHg@0°C, ftH2O@4°C, ftH2O@20°C, ftH2O@60°F
Instrument Power Supply 24 VDC at 24 mA
Accuracy
Pressure
±0.025% of full scale on internal sensor module dependent for external sensors Add 0.005% FS/˚C outside of 15 to 35˚C
mA ±0.015% of reading ± 2 counts
(simulation and measurement)
Volts ±0.015% of reading ± 2 counts
(measurement only)
Temperature ±0.1°C (0.2°F) (measurement only)
(Pt100 RTD)
General
Integrated internal electrically operated pneumatic pump:
generates -12 PSI to 150 PSI pressure
(-0.8 to 10 Bar, -8 to 100 kPa) Manual pump version -12 to 300 PSI / -0.8 to 20 Bar Environmental
Operating Temperature -10 °C to +50 °C
Storage Temperature -20 °C to +60 °C Power Requirements
Battery
Battery Life 300 pump cycles to 150 PSI minimum
Physical
Dimensions 8” high x 4” wide x 2-3/8” deep
Weight 2.5 pounds (1.2 kg) Connections
Pressure: 1/8” female NPT
Electrical: standard banana jacks
RTD: 4 pin LEMO style
External module: 6 pin LEMO style Included Accessories - NIST traceable calibration certificate,
batteries installed, test leads, test hose fittings, user guide,
carrying case.
Eight (8) standard Alkaline AA batteries
1000 pump cycles to 30 PSI
(20 cm x 10 cm x 6 cm)
(electric pump)
(manual pump)
BETAGAUGE 330 Hand Held Pressure Calibrator with Integral Electric Pressure Pump
A Pt100 RTD input is provided for temperature
easurements accurate to ±0.1 °C (0.2 °F).
m In addition, the BetaGauge 330 measures and simulates 4-20 mA loop current signals. It can measure up to 30 VDC. An internal 24 VDC Instrument Power Supply supplies power to a transmitter under test.
Like its calibrator siblings, the 330 displays up to 3 calibration values at one time from the choice of internal pressure sensor, external pressure sensor, temperature from the optional RTD probe or electrical (mA or VDC). The display is a large back lit graphics style LCD that’s easy to read in any lighting condition.
The new BetaGauge 330 Pressure Calibrator provides a number of convenience features. Switch test can be performed on internal or external pressure input. Percent (%) Error and damping functions are provided. Up to five frequently used setups can be stored and retrieved with one button recall.
The compact, rugged design operates on eight (8) standard AA Alkaline batteries. Due to the power saving designs incorporated in the BetaGauge 330, it can perform approximately 300 calibration cycles to 150 PSI on one set of batteries.
The 330 comes in 3 ranges of 30 PSI, 150 PSI and 300 PSI. The 300 PSI version uses a manually operated handpump.
Like all Martel calibrators, the BetaGauge 330 comes from the factory ready to go to work with batteries installed, NIST traceable calibration certificate, test leads, carrying case and user manual.