Return For Calibration/Repair................................54
3
INTRODUCTION:
GENERAL DESCRIPTON
The Model 320 pressure calibrator is a compact,
versatile, functionally styled calibrator. The unit is
menu driven, so technicians can immediately begin
using it to its fullest capabilities without extensive
training.
The Model 320 faceplate consists of three functional
areas:
At the top is a two line, alphanumeric liquid crystal
display (LCD) for simltaneous indication of pressure
and milliamps/volts readings and operator prompting.
The center section is a membrane keypad for operator
entry.
The lower left section contains recessed input jacks
for connecting the supplied test leads to the device
being calibrated.
The pressure connectors are at the lower end of the
case and are 1/8” FNPT.
The Loop Power Supply (LPS) is an option that
provides 24 VDC to power the device under test if
4
desired. The LPS is stored in a pouch under the
Model 320 calibrator for easy access.
BATTERY CHARGING
Units are shipped with the internal batteries in an
uncharged state. Before using the Model 320 or the
LPS as a portable unit, it is necessary to charge the
batteries for a minimum of 14 to 16 hours. The
calibrator battery charger is a wall plug-in type (see
below) that connects to a jack on the right-hand side
of the instrument. Battery operation time is 50 hours,
typically for the Model 320 and 6 hours nominal at 20
milliamps for the LPS.
5
OPERATION:
The Model 320 calibrator is user friendly and menu
driven. A first time user can normally start operating
the unit with few or no instructions. The following
information is intended to provide a detailed
understanding of the unit.
POWER ON SEQUENCE
The calibrator is turned on or off by pressing the
On/Off key. Each time it is turned on, a series of
internal tests and calibrations are performed. The
start-up procedures provide a pass/fail test of certain
internal circuitry. If the tests determine the unit has an
internal failure, a fault message is displayed and the
unit will not operate. Refer to the Service Information
on page 54 for information on how to contact the
factory.
The start-up calibrations provide for dynamic
recalibration of the internal calibrator support
circuitry, by applying the primary reference voltages
to points in the circuitry and recording the readings as
references for further measurements. The start-up
6
auto calibration does not in any way affect the primary
calibration done at the factory or at a field metrology
lab using available software and suitable standards.
During the time the start-up tests and calibrations are
being performed, certain informational displays are
presented to show the status of the calibrator. The
first message displayed will be:
Beta Gauge
01/04/94
This indicates the date in MM/DD/YY format of the
last factory calibration. The example indicates the
Model 320 was calibrated on Jan. 4, 1994. If the unit
has been recalibraed in the field, the date will be
displayed in a MM-DD-YY format. See page 26 for a
description of the calibration procedure.
The next display will be:
property of
YOUR NAME HERE
7
This gives the name of the owner of the calibrator.
The preceding display shows the default setting as the
calibrator is shipped. The name can be changed
through the operator setups. See page 24 for how to
change the owner’s name.
The next display will be:
Auto Calibration
in progress
This display will remain until the internal tests and
calibrations are finished. If the auto calibration
message fails to clear, the unit must be returned for
service. See page 54 for details. The next display will
be the functions that had been selected when the
calibrator was last used:
inH2O mA
0.0 0.000
This display indicates the calibrator is operating
properly and is ready to go to work.
8
BATTERY CHECK
The Bat Ck key may be pressed at any time after
completion of the calibrator startup routine to diplay
the current batttery voltage for approximately 1
second before the normal display returns. The current
input funciton is not disturbed by making a battery
check measurement.
The battery voltage of the Model 320 calibrator should
be between 5.3 and 4.6 volts. When the voltage falls
below 4.8 volts, a warning message is displayed
approximately every minute. When the voltage falls
below 4.6 volts for 1 minute, the calibrator will turn
itself off to prevent damage to the batteries due to
possible reversal.
The batteries of the LPS are checked by plugging the
cable provided with the LPS into the battery charger
jack located on the side of the LPS. The other end of
the cable should be plugged into the Model 320 V
input. When voltage is selected on the keypad, the
LPS battery voltage will be displayed on the Model
320.
9
The LPS battery voltage should be between 10.6 and
9.2 volts, indicating a full charge. When the voltage
falls below 8.8 volts, the LPS will turn itself off to
prevent damage to the batteries due to possible cell
reversal.
SELECT KEY
The Select Key is used for two functions. First, while
in Setup, pressing Select will switch the dispay from
two sets of menu screens. Second, Select is used in
the field calibration described on page 26.
ZERO KEY
When the Model 320 is first turned on, a small zero
offset may appear. Pressing the zero key wil eliminate
it. A warm up of up to four minutes may be required
to achieve nominal accuracy. For absolute pressure sensors only, the following procedure may be used.
Press the Zero Key. The calibrator will request the
user to enter the calibration access code (found on the
factory calibration data sheet.) After entering the code
and pressing the enter key, the calibrator will prompt
the user to enter the new zero (actual barometric
10
pressure). A barometric standard of at least .025%
accuracy should be used to determine atmospheric
pressure. When entering this value, proper decimal
place entry must be observed.
The “Zero” key is inactive during user calibration.
Only numeric entry (9) is active in this mode.
MIN/MAX KEY
The Model 320 calibrator constantly monitors the
displayed readings and stores the minimum and
maximum values internally.
Pressing the Min/Max key will cause the minimum
and maximum values to be displayed. The maximum
values will be displayed on the top line and the
minimum values will be displayed on the bottom line.
Pressing the Min/Max key a second time will return
the Model 320 to a normal display mode and will reset
the minimum and maximum values to the current
reading.
The minimum and maximum values are also reset
whenever a new function or engineering unit is
selected.
11
ENG UNIT KEY
The Eng Unit key allows the user to select any of the 8
standard or 1 user definable engineering units to be
activated for the pressure display. Each time the Eng
Unit key is pressed, the next engineering unit is
selected. Engineering units which exceed PSI
resolution, will not be valid. Example, 50,000 counts,
when converted from the 5000.0 PSIG converted to
mmHg equals 258,575 mmHg. This exceeds 50,000
counts and will not be an active unit of measurement.
If a user defined engineering unit has been entered, it
is displayed as the 9th entry. The user defined entry
can always be identified by a “u” in the first position,
followed by the name.
The user defined engineering uits are configured from
the operator setup functions (UsrEng).
DAMP KEY
The Damp key alternately enables or disables
damping of the displayed pressure measurements.
When damping is active, the Model 320 displays a
running average of the last 16 measurements rather
12
than the instantaneous values. This feature is useful
when measuring pressures that are pulsating or
otherwise unstable.
The Model 320 makes 2-3 pressure measurements per
second. When the Damp mode is active, the Model
320 updates the display each time a measurement is
made. The displayed is the average of the last 16
measurements.
PORT SELECT KEY
The Port Select key is used to select either or both of
the pressure ports to be displayed on dual sensor 320
calibrator models.
If the Model 320 is equipped with dual pressure
sensors, pressing the Port Select key will alternately
select the left port, the right port, or both ports.
If both ports are selected, the Model 320 will display
both pressures simultaneously but milliamps or volts
cannot be displayed.
If one port is selected, the Model 320 will display one
pressure and either milliamps or volts.
13
MA/V KEY
The mA/V key will alternate the electrical
measurement being displayed between milliamps and
volts.
% KEY
Pressing the % key will cause the Model 320
calibrator to switch from displaying in the selected
engineering units to displaying the readings as a
percentage. The Model 320 percent function is a
powerful feature that greatly simplifies the calibration
of I/P, E/P, P/I, P/E AND P/P transmitters. When the
percent function is active, the input to the instrument
being calibrated is displayed as a percent of full scale,
and the output from the instrument being calibrated is
displayed as an error percent.
To utilize the percent function, the operator should
first utilize the setup function (% func) to enter the
desired zero and full scale values required for the
calibrator. This is described on page 21.
14
Once the proper zero and full scale values have been
entered, the operator presses the % key to activate the
percent function. The zero and span adjustments to
the transmitter can now be quickly adjusted by
applying aproximately 0% and 100% inputs to the
transmitter as shown on the display, and adjusting the
output for 0% error as shown on the display.
After the zero and span adjustments have been made,
the linearity can be quickly checked by applying
various intermediate inputs and noting the error
directly in percent. If there is no error, 0% will be
displayed for any pressure input.
The application note on page 45 gives examples for
using the percent function.
Pressing the % key a second time will return the
display to the selected engineering units. The percent
mode is also terminated if a new function or
engineering unit is selected.
TRIP RESET KEY
The Trip Rst key is used in conjunction with the trip
15
detect mode that is available through the setup menus.
When trip detect is active, the Model 320 calibrator
will monitor the electrical measurement for a
significant change of state. When a change of state is
detected, the last pressure measurement is frozen on
the display, and the message “tripped” is displayed
over the electrical measurement.
When the Trip Rst key is pressed, the Model 320 will
resume displaying the instantaneous pressure and
monitoring for a change of state at the electrical
terminals.
The message “ready” will be displayed over the
electrical measurement.
A change of state is signified by a 10% change in
either the volts or milliamps reading. When trip detect
is active, voltage measurements will be made on the
30 V scale and milliamps measurements will be made
on the 50 mA scale. The Model 320 makes 2-3
electrical measurements per second. To guarantee that
the change of state will be detected, the change must
exist for a minimum of ½ second.
For a hook-up diagram, refer to the application note
on page 43.
16
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.