Martel Electronics 303, 306 Data sheet

The Martel Electronics 303 and 306 Temperature
303 and 306
Temperat ure Meters
Meters
measurement of temperature. Both instruments provide ±0.2% of reading or better accuracy and 0.1°C/°F resolution over their measurement ranges. Both feature input protection to 60 VDC or 24 Vrms. REL, HOLD, and MIN/MAX functions simplify use. Both provide automatic power off to preserve and extend battery life. The 306 provides data logging for up to 16,000 records.
provide a variety of options for the accurate
Features:
Dual inputs
-200 to +1370 °C measurement range Dual display (303), triple display (306)
Auto power off Time function (306) RS232 interface 16,000 record data logging (306)
Martel Electronics, P.O. Box 770, Londonderry, NH 03053
Tel: 800-821-0023, E-mail: sales@martelcorp.com, W eb: www .mar telcorp.com
303/ThermoLink® Software
The Martel 303 Temperature Meter can be supplied with ThermoLink software and an RS-232 cable which con­nects the 303 to an unused COMM port on a PC or laptop. With the 303 connected to the PC and the ThermoLink software running, realtime contr ol and analysis of its data can be achieved. With an extended length RS-232 cable, the 303 Temperature Meter and PC can be used as a remote temperature measurement system.
On start-up, the ThermoLink software automatically scans the available COMM ports on the PC or laptop, and establishes communication with the 303 Temperature Meter . Once communications is established, temperature data is continually downloaded from the 303. The data sampling rate can be adjusted in a number of ways.
Clicking on the Graphical section of the main screen opens a graphic display of the data being downloaded in realtime from the 303. At the top of the display is a text field where the time interval between data samples can be adjusted. A number of the parameters of the graph can be modified to better display the range of data being displayed. Controls are pr ovided for starting, stopping, and pausing the recording of data to the graph.
Clicking on the Control Panel section of the main screen opens an image of the 303, which provides interactive, realtime control of all the 303’s functions.
Clicking on the Auto Arrange section of the main screen opens the Control Panel, Realtime Graph, and Tabular windows, arranged in a non-overlapping display.
Clicking on the Help section of the main screen accesses an indexed HTML document that provides detailed information on menus and controls within the ThermoLink software.
Clicking on the Tabular section of the main screen opens a tabular display of the data being downloaded in realtime from the 303.
Model 303
The 303 is a dual-input, dual display, temperature meter which accepts both Type-J and -K thermocouples. It features a measurement range of -200 to +760 °C (Type-J) and -200 to +1370 °C (Type-K). REL, HOLD, and MAX/AVG/MIN functions, and temperature unit selection are available at the push of a button. Measurement of T1, T2, or T1-T2 is selected by push button as well. An RS232 interface is available for downloading measurement data to a PC for analysis.
Loading...
+ 2 hidden pages