Martel Electronics 3001 User Manual

3001
Operator’s Manual
3001 Operator’s Manual
1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.1 Customer Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.2 Standard Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.3 Options and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.4 Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
1.5 Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2
2.1 Front Panel Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
2.2 Primary Input/Output Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
2.3 Primary Input/Output Display and Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
2.4 Isolated Input Display, Controls, and Terminals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
2.5 Rear Panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
2.6 Display Layouts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11
2.7 Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
3. Getting Started . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4. Primary Inputs and Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.1 DC Voltage Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .14
4.2 DC Current Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
4.3 Resistance Temperature Detector (RTD) and Ohms Measure . . . . . . . . . . . . . . . . . . . . .16
4.4 Resistance Temperature Detector (RTD) and Ohms Source . . . . . . . . . . . . . . . . . . . . . . .18
4.5 Resistance Temperature Detector (RTD) with Custom Coefficients . . . . . . . . . . . . . . . . .19
4.6 Standard Platinum Resistance Thermometer (SPRT) Coefficients . . . . . . . . . . . . . . . . . . .20
4.7 Thermocouple (T/C) Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
4.8 Thermocouple (TC) Source . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
4.9 Pressure Measure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24
5. Isolated Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.1 Voltage Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.2 Current Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
5.3 Pressure Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
6. Output Setpoints . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
7. Application Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
7.1 P/I Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28
7.2 I/P Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .29
7.3 V/I Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
7.4 RTD Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .30
7.5 RTD Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
7.6 Thermocouple Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.7 Thermocouple Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .32
7.8 RTD Indicator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .33
7.9 Precision Current Trip . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .34
7.10 I/I Isolator/Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .35
7.11 Precision Temperature Measurement with IBP-2 Probe . . . . . . . . . . . . . . . . . . . . . . . . .36
8. LCD and Remote Interface Setup Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9. Remote Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .37
9.2 Setting up the RS-232 Port for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
9.3 Setting up the IEEE-488 Port for Remote Control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .38
9.4 Changing Between Local and Remote Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .39
9.5 IEEE-488 Interface Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .40
9.6 Using Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .41
9.7 Checking 3001 Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .45
10. Remote Commands . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.2 Command Summary by Function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .50
10.3 Error Code Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .52
10.4 Remote Command Listing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .54
11. Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.1 Cleaning the Calibrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.2 Replacing a Line Fuse . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .77
11.3 Changing the Line Voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .78
12. Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
12.1 General Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .79
12.2 DC Voltage Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
12.3 DC Voltage Specifications, Isolated Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .80
12.4 DC Current Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.5 DC Current Specifications, Isolated Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.6 Resistance Specifications, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.7 Resistance Specifications, Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .81
12.8 Thermocouple Specification, Output and Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
12.9 RTD and Thermistor Specification, Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
12.10 RTD and Thermistor Specification, Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
12.11 Pressure Measurement Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
13. Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .87
1. Introduction
The Martel series 3001 calibrator is an accurate full-featured temperature, pressure and DC calibrator intended for R & D, manufacturing and calibration lab applications. The unit's simple design and ease of operation allow users to quickly familiarize themselves with its operations and features. Time saving functions like the ability to save, recall and automatically cycle through setpoints for each output range, the ability to enter user definable RTD curves, and a complete remote interface are several key features offered by the 3001.
1.1 Customer Service
Corporate Office:
www.martelcorp.com
e-mail: sales@martelcorp.com
Tel: (603) 434-1433 800-821-0023 Fax: (603) 434-1653
Martel Electronics
3 Corporate Park Drive
Derry, NH 03038
1.2 Standard Equipment
• Power cord
• Thermocouple shorting jumper
1.3 Options and Accessories
• IBP-2 PT100 High Accuracy Probe
• 80029 JKTE Thermocouple Kit
• 80036 RSNB Thermocouple Kit
• PTL-1 Low EMF Beryllium Copper Test Leads
The 3001 and above options are available in a complete kit form.
1.4 Unpacking
Upon receipt of the shipment, inspect the container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping material is saved for their examination.
After examining and removing the contents, save the packing material and carton in the event that re-shipment is necessary.
Remove the Packing List and verify that all of the listed equipment has been received. If there are any questions about the shipment, please call Martel Electronics at 1-800-821-
0023.
Check to see if your basic calibrator package is complete. It should include:
• 3001 Calibrator
• Instruction Manual
1
• AC Line Cord
• Thermocouple Shorting Jumper
• NIST Certificate
1.5 Safety Information
Symbols Used
The following table lists the International Electrical Symbols. Some or all of these symbols may be used on the instrument or in this manual.
Symbol Description
AC (Alternating Current)
AC-DC
Battery
CE Complies with European Union Directives
DC
Double Insulated
Electric Shock
Fuse
PE Ground
Hot Surface (Burn Hazard)
Read the User’s Manual (Important Information)
Off
On
Canadian Standards Association
2
The following definitions apply to the terms "Warning" and "Caution".
• " Warning" identifies conditions and actions that may pose hazards to the user.
• " Caution" identifies conditions and actions that may damage the instrument being used.
Use the calibrator only as specified in this manual, otherwise personal injury and/or damage to the calibrator may occur.
Warning
To avoid possible electric shock or personal injury:
• Do not apply more than the rated voltage between the terminals, or between any
terminal and chassis ground. See specifications for supported ranges.
• Follow all equipment safety procedures.
• Always use the power cord and connector suitable for the voltage and outlet of the
location in which you are working.
• Do not use the calibrator if it is damaged. Before you use the calibrator, inspect the
case. Look for cracks or missing plastic. Pay particular attention to the insulation surrounding the connectors.
• Inspect the test leads for damaged insulation or exposed metal. Check test lead
continuity. Replace damaged test leads before you use the calibrator.
• Never operate the calibrator with the cover removed or the case open, and never
remove the cover or open the case without first disconnecting the power source and all test leads.
• Select the proper function and range for your measurement.
• Never touch the probe to a voltage source when the test leads are plugged into the
current terminals.
• When using the probes, keep your fingers away from the probe contacts. Keep your
fingers behind the finger guards on the probes.
• Connect the common test lead before you connect the live test lead. When you
disconnect test leads, disconnect the live test lead first.
• Do not use the calibrator if it operates abnormally. Protection may be impaired. When
in doubt, have the calibrator serviced.
• Do not operate the calibrator around explosive gas, vapor, or dust.
• When using a pressure module, make sure the process pressure line is shut off and
depressurized before you connect it, or disconnect, it from the pressure module.
• Disconnect test leads before changing to another measure or source function.
• Have the calibrator serviced only by qualified personnel, and use only specified
replacement parts.
• Use only the replacement fuse(s) specified in this manual.
• To avoid a violent release of pressure in a pressurized system, shut off the valve and
slowly bleed off the pressure before you attach the pressure module to the pressure line.
3
Caution
To avoid possible damage to the calibrator or to the equipment under test:
• Use the proper terminals, function, and range for your measurement or sourcing application.
• To avoid mechanically damaging the pressure module, never apply more than 10 ft-lb of torque between the pressure module fittings, or between the fittings and the body of the module.
• To avoid damaging the pressure module from overpressure, never apply pressure above the rated maximum printed on the module.
• To avoid damaging the pressure module from corrosion, use it only with specified materials. Refer to the pressure module documentation for material compatibility.
Make sure to remove all test connections before powering up the 3001. Failure to do so could lead to damage of the 3001 or the unit under test.
2. Calibrator Description
2.1 Front Panel Overview
Figure 1 shows the overall layout of the front panel. Each of the three major divisions is described in detail in the following sections.
Figure 1 - Front Panel
Item Name Description
1 Primary input/output terminals See section 2.2 for details
2 Primary input/output display and controls See section 2.3 for details
3 Isolated input display, controls, and terminals See section 2.4 for details
4
2.2 Primary Input/Output Terminals
Figure 2 describes the primary input/output terminals in detail. The display and controls for these terminals are described in the next section.
Figure 2 - Primary Input/Output Terminals
Item Name Description
1 VOLTS DC voltage output terminals. See notes 1 and 2 below.
2 mA DC current output terminals. See notes 1 and 2 below.
3 RTD/O OUTPUT Two wire RTD and Ohms output terminals. See notes 1
and 2 below.
4 TC INPUT/OUTPUT Thermocouple input and output terminals. These termi-
nals accept a miniature polarized thermocouple plug with flat in-line blades spaced 7.9 mm (0.312 in.) center to cen­ter.
5 4W RTD/O INPUT Four wire RTD and Ohms input terminals. See notes 1
and 3 below.
6 Pressure module input connector.
Note 1: These terminal binding posts are made of a special copper alloy to reduce thermal
EMF's. They support the use of either discreet wires or standard banana plugs, and the HI/LO pairs are spaced for standard dual banana plugs.
Note 2: Caution. Do not exceed a maximum of 100 volts to chassis ground.
Note 3: Caution. Do not exceed a maximum of 20 volts to chassis ground.
5
2.3 Primary Input/Output Display and Controls
Figure 3 describes the primary input/output display and controls in detail.
Figure 3 - Primary Input/Output Display and Controls
Item Name Description
1 Display A 2 line, 16 character, display providing all visual user
feedback for the primary output and input operations.See section 2.6 for layout details, and section 2.7 for possible error messages.
2 Numeric and secondary Output value data entry keys.Secondary function selection
function keys per the text printed above the numeric key. Press the
key followed by the numeric key to select the func-
tion.
OUTPUT Change RTD/Ohms or Thermocouple to output mode.
INPUT Change RTD/Ohms or Thermocouple to input mode.
ZERO Zero the input for Pressure, Thermocouple millivolts, or
RTD ohms
SETUP Adjust the LCD Contrast, LCD Backlight, and Remote
Interface Configuration as described in section 8.
CJC Select internal or external cold junction compensation for
Thermocouple temperature measurements. When external compensation is selected, XCJC is displayed at the start of the second line.
°C/°F Select Centigrade or Fahrenheit units for RTD and
Thermocouple temperature measurements.
6
SET Set a new value for a preset output setpoint as described
in section 6.
RECALL Recall a preset output setpoint as described in section 6.
AUTOSET Initiate automatic stepping of preset output setpoints as
described in section 6.
RNG LOCK Select Auto-range or Range Lock for voltage output.
LOCAL Press to regain local control of the 3001 after the remote
command REMOTE has been received; in this case all keys except this one are ignored. When the remote com­mand LOCKOUT has been received, all keys are ignored including this one and the remote command LOCAL must be received to regain local control.
EXP Press during entry of a RTD custom curve coefficient to
begin entering the exponent.
3 Function keys
Select DC voltage or current output mode, and toggle between them.
Select Thermocouple or RTD/Ohms input/output mode, and toggle between them.
Select Pressure input mode.
In Thermocouple mode, cycle through the thermocouple types, including millivolts.In RTD/Ohms mode, cycle through the RTD types, including ohmsIn Pressure mode, cycle through the pressure units.
For all output modes, except Thermocouple, toggle between Standby and Operate modes.In Standby mode, any change to the output value in the display is not driven to the terminals until the Operate mode is selected.In Operate mode, each change to the output value in the dis­play is driven to the terminals immediately, except for DC voltages greater then 30V when the mode reverts to Standby automatically for safety reasons.
Changes the calibrator output or parameter to the numeric value typed into the keypad.
7
Clears a partial keypad entry and reverts the calibrator out­put or parameter to its last known value.
Prepares for selection of a secondary function via the numeric keypad according to the text above each key. The display changes to SHIFT ENABLED until a numeric
key is pressed. To cancel the selection press again.
4 Cursor controls Press or to position the cursor under the digit in
an output value that is to be incremented or decrement-
ed.Press to increment the digit in the output value where the cursor is positioned.
Press to decrement the digit in the output value where the cursor is positioned.
The and keys are also used to adjust LCD Contrast level, LCD Backlight level, and Remote Interface Configuration selections as described in section 8.
2.4 Isolated Input Display, Controls, and Terminals
Figure 4 describes the isolated input display, controls, and terminals in detail.
Figure 4 - Isolated Input Display, Controls, and Terminals
8
Item Name Description
1 Display A 2 line, 16 character, display providing all visual user
feedback for the isolated input operations. See section 2.6 for layout details, and section 2.7 for possible error mes­sages.
2 Function keys
Select DC voltage and current input mode. Subsequent presses of this key cycle through the ranges: 10V, 100V, and 50mA.
When using 50mA mode to test a 2 wire loop powered transmitter that is disconnected from its wiring, press this key to activate an internal 24V power supply in series with the current measuring circuit. Press the key again to deactivate the 24V supply.
When using 50mA mode to test a HART configuration device, press this key to activate an internal 250 ohm resis­tor in series. Press the key again to deactivate the resis­tor.Note that activating this resistor drops the maximum load driving capability from 1000 ohms at 20mA to 750 ohms at 20mA.
Select Pressure input mode. Subsequent presses of this key cycle through the pressure units.Pressure input mode uses the pressure module connector on the primary input/output side. Both sides may be selected to pressure mode simultaneously and can be set to display the same pressure measurement in different units if desired.
3 Input Terminals Common input terminals for DC voltage and current. See
notes 1 and 2 below.
Note 1: These terminal binding posts are made of a special copper alloy to reduce thermal
EMF's. They support the use of either discreet wires or standard banana plugs, and the HI/LO pairs are spaced for standard dual banana plugs.
Note 2: Caution. Do not exceed a maximum of 100 volts to chassis ground.
9
2.5 Rear Panel
Figure 5 describes the rear panel layout.
Figure 5 - Rear Panel
Item Description
1 RS-232 9 pin connector for remote control of the 3001 via any computer's serial interface.
2 GPIB IEEE 488.2 connector for remote control of the 3001 via a GPIB bus.
3 Service port for updating the 3001 firmware.
4 Chassis ground terminal internally connected to the ground prong of the AC power inlet.
Warning To avoid shock hazard, connect the factory supplied 3 conductor power
cord to a properly grounded power outlet. Do not use a 2 conductor adapter or extension cord as this will break the protective ground. Use the chassis ground terminal for a protective ground wire if there is any question about the grounding through the 3 conductor power cord.
5 Standard IEC AC power inlet for 120/240 VAC.
6 Main power on/off switch.
7 Power line voltage selector and fuse compartment. See section 11 for instructions on
changing the line voltage selector and changing fuses.
Warning To prevent electrical shock, only remove the line voltage selector and
fuse holder when the power cord is removed.
10
2.6 Display Layouts
a) Primary Voltage and Current Display
Figure 6 - Primary Voltage and Current Display Layout
Item Description
1 Operating mode:
AUTO: Auto-range LOCK: Range lock rem remote operation SP# Automatic stepping of preset setpoints
2 Present range and output mode
3 Output state:
Stby Standby, terminals inactive Opr Operating, terminals are active with output per the displayed value
4 Output value
5 Units
b) Primary Thermocouple and RTD Display
Figure 7 - Primary Thermocouple and RTD Display Layout
Item Description
1 Output mode selection: RTD, TC, or rem for remote operation
2 Input or output selection
3 RTD or thermocouple type selection
11
4 Output state for RTD outputs:
Stby Standby, terminals inactive
Opr Operating, terminals are active with output per the displayed value Blank for RTD inputs Cold junction selection for thermocouple inputs and outputs:
XCJC External cold junction compensation; the 3001 automatic cold junction
compensation is turned off, i.e. 0 mV is always 0°C
Blank Internal cold junction compensation; the 3001 automatically measures
the ambient temperature at the thermocouple terminals and compen­sates the measurement, i.e. 0 mV is ambient temperature
5 Input or output value
6 Units
c) Primary and Isolated Pressure Display
Figure 8 - Primary and Isolated Pressure Display Layout
Item Description
1 UnitsOn the primary display, rem appears to the left during remote operation
2 Input value
d) Isolated Voltage and Current Display
Figure 9 - Isolated Voltage and Current Display Layout
12
Item Description
1 Selected range and input mode:·
• 10V RANGE or 100V RANGE for DC voltage·
• 50mA RANGE for basic DC current·
• 24mA LPWR for loop powered 2 wire transmitter·
• 24mA LPWR HART for loop powered 2 wire transmitter with 250 ohm resistor in circuit
2 Input value
3 Units
2.7 Error Messages
The following error messages may appear on either display.
Table 1 - Error Messages
Message Description
OVER RANGE The value entered on the numeric keypad exceeds the range of the output
mode selected.
OVER LOAD For DC voltage output mode, the current required to generate the output
exceeds the 3001 specifications.For DC current mode, the resistance of the circuit exceeds the 3001 specifications.
OL For input modes, the measured value exceeds the upper limit of the selected
input mode range.For output modes, when the range is locked, the present automatically recalled preset setpoint exceeds the upper limit of the locked range. The output is set to zero for the duration of this setpoint.
-OL For input modes, the measured value exceeds the lower limit of the selected input mode range.
13
3. Getting Started
After unpacking the 3001 and becoming familiar with the layout and general operation of the unit as described in the previous section, it is ready to set up for operation.
The following steps should be followed to set up the 3001 for operation:
a) Before attaching the power cord to the rear connector, check that the line voltage
selector is set appropriately for your location. The 3001 is shipped from the factory with the line voltage set for the country of purchase. To verify the line voltage setting, check the indicator on the power line voltage selector and fuse compartment cover; see figure 5 in section 2.5 for the location.
Confirm that the setting is correct according to the following guidelines:
Line Voltage (50/60Hz) Selector Position
90 to 135 VAC 120 VAC position
220 to 250 VAC 240 VAC position
If the setting is not correct, follow the instructions in section 11.3 to change it.
b) Once the voltage selection has been made, making sure that the power switch is off,
connect the AC power cord to the 3001; see figure 5 in section 2.5 for the location.
c) Turn on the 3001 using the rear panel mounted rocker switch. The 3001 should
power up within a few seconds, briefly displaying the model number and firmware version in the primary display before reverting to the normal input/output display.
NOTE: If a proper power up display does not occur within 30 seconds, turn the power off, wait a few seconds, and repower the unit. If the problem persists, report the problem to Martel immediately.
Warm up time is twice the time since last warmed up, to a maximum of 30 minutes. For good stability it is best to leave the 3001 on all the time.
4. Primary Inputs and Outputs
4.1 DC Voltage Output
The 3001 can source DC voltages from 0 V to 100 V, using the following four ranges for maximum accuracy: .1 V, 1 V, 10 V, and 100 V.
a) Disconnect any test leads from external devices.
b) Press the key to select DC voltage and current mode, if not already selected.
If DC current mode is displayed, press the key again to cycle to DC voltage mode.
c) Connect the unit under test to the voltage output terminals of the 3001 as shown in
figure 10.
14
Figure 10 - DC Voltage Output Connection
d) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple solution when small changes to an output value are required, or if specific decades need to be incremented or decremented.
e) When DC voltage mode is first selected, the 3001 is placed in the standby (Stby)
mode which puts the positive (+) output jack into a high impedance state
(>100k ohm) for safety. To place the output into the active state, press the key which toggles between the standby and operate modes.
The standby mode is also activated in the following situations:
• If a fault occurs during operation, such as an overload or short circuit condition.
• As a safety feature for all new outputs over 30 VDC. Refer to the product specification
section of this manual for maximum drive currents.
Warning. Scrolling the output when the output value is already over 30V will not place
the 3001 in standby mode for each new value.
Warning. Automatic setpoints over 30V will not place the 3001 in standby mode for
each new value.
f) The 3001 can be locked to a specific voltage range by entering a value in that range
and then selecting the secondary RNG LOCK function by pressing the and
keys.
4.2 DC Current Output
The 3001 can source DC current from 0 mA to 100 mA.
a) Disconnect any test leads from external devices.
b) Press the key to select DC voltage and current mode, if not already selected.
If DC voltage mode is displayed, press the key again to cycle to DC current mode.
15
c) Connect the unit under test to the current output terminals of the 3001 as shown in
figure 11.
Figure 11 - DC Current Output Connection
d) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the
or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades need to be incremented or decremented.
e) When DC current mode is first selected, the 3001 is placed in the standby (Stby)
mode which puts the positive (+) output jack into a high impedance state
(>100k ohm) for safety. To place the output into the active state, press the key which toggles between the standby and operate modes.
The standby mode is also activated in the following situations:
• No connection made to the output terminals.
• The voltage compliance for a given output current is exceeded. The 3001 has a typical voltage compliance of 10V so that 4-20mA application loads of up to 500 ohms can be driven. At maximum current of 100mA, the maximum load is 100 ohms.
4.3 Resistance Temperature Detector (RTD) and Ohms Measure
The 3001 can measure all common RTD types, 5 custom RTD curves, and a custom SPRT in °F or °C, plus basic resistance from 0 to 4000 ohms.
The following common RTD types are supported:
Pt 385 100, 200, 500, 1000
Pt 3926 100
Pt 3916 (JIS) 100
Ni120 120
Cu 427 (Minco) 10
YSI 400
16
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/mode, if not already
selected. If thermocouple mode is displayed, press the key again to cycle to RTD/ mode.
c) If output mode is displayed, select input mode by pressing the and
keys.
d) Press the key to select the desired RTD type, the custom curve, the SPRT, or
the desired ohms range. The set up and use of custom RTD coefficients is described in section 4.5. The set up and use of SPRT coefficients is described in section 4.6.
e) Connect the unit under test to the 4 wire RTD/input terminals of the 3001 as shown
in figure 12.
Figure 12 - RTD/Input Connection
f) Press the and keys to toggle the RTD display between °F and °C.
g) For best accuracy, it is advisable to zero the RTD resistance circuit(s) daily, or if the
3001 is being used outside of the ambient temperature range of 18 to 28 °C. The maximum offset from unit calibration that can be zeroed out is ±0.1 ohm for the high range and ±0.01 ohm for the low range.
To zero a RTD resistance circuit:
• Select the RTD measure function as described above, and press the key until the low or high ohms range is selected.
• Short the RTD/terminals with the test leads normally used for RTD/ measurements.
• Allow at least 3 minutes for the test leads and terminals to stabilize to the same temperature.
• Press the and keys to zero the ohms range.
17
4.4 Resistance Temperature Detector (RTD) and Ohms Source
The 3001 can source all common RTD types and 5 custom RTD curves in °F or °C, plus basic resistance from 5 to 4000 ohms.
The following common RTD types are supported:
Pt 385 100, 200, 500, 1000
Pt 3926 100
Pt 3916 (JIS) 100
Ni120 120
Cu 427 (Minco) 10
YSI 400
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/mode, if not already
selected. If thermocouple mode is displayed, press the key again to cycle to RTD/ mode.
c) If input mode is displayed, select output mode by pressing the and keys.
d) Press the key to select the desired RTD curve or ohms range. The set up
and use of custom RTD coefficients is described in section 4.5.
e) Connect the unit under test to the RTD/output terminals of the 3001 as shown in
figure 13.
Figure 13 - RTD/Output Connection
f) Press the and keys to toggle the RTD display between °F and °C.
g) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the
or cursor key to select a digit to modify, followed by the
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades need to be incremented or decremented.
18
h) When RTD/mode is first selected, the 3001 is placed in the standby (Stby) mode
which puts the positive (+) output jack into a high impedance state (>100k ohm) for
safety. To place the output into the active state, press the key which toggles between the standby and operate modes.
4.5 Resistance Temperature Detector (RTD) with Custom Coefficients
The 3001 has the capability to store coefficients for up to 5 custom RTD curves.
To enter the coefficients for a custom RTD curve:
a) Select RTD measure or source mode as described in the preceding sections.
b) Press the key until the USR_DEF type is selected.
c) Press the key to display the curve selection prompt "RTD CUSTOM (1-
5)".
d) Press the numeric key corresponding to the custom RTD curve to be entered.
e) At the "SET(1)/RECALL(2)" prompt, press to select custom RTD curve
data entry.
f) At the "ENTER MIN TEMP" prompt, enter the minimum temperature limit for the
custom RTD curve, and press the key.
g) At the "ENTER MAX TEMP" prompt, enter the maximum temperature limit for
the custom RTD curve, and press the key.
h) At the "ENTER R0" prompt, enter the nominal resistance value (R0) for the
custom RTD curve, and press the key.
i) At the "ENTER COEFF A" prompt, enter the first (A) coefficient for the custom
RTD curve, and press the key. To enter a coefficient that includes an
exponent, enter the mantissa, press the and keys to select the EXP
function, enter the exponent, and press the key.
j) When prompted, enter the second (B) and third (C) coefficients in the same
manner.
k) To abort the curve entry without saving any changes, press the key.
To use a custom RTD curve:
a) Select RTD measure or source mode as described in the preceding sections.
b) Press the key until the USR_DEF type is selected.
c) Press the key to display the curve selection prompt "RTD CUSTOM (1-
5)".
19
d) Press the numeric key corresponding to the custom RTD curve to be used.
e) At the "SET(1)/RECALL(2)" prompt, press to recall the custom RTD curve
coefficients.
f) To use a different custom RTD curve, press the key twice to display the
USR_DEF selection prompt.
The USR_DEF function of the 3001 uses the Calendar-Van Dusen equation for sourcing and measuring custom RTD's. The C coefficient is only used for the subrange -260 to 0 degrees Celsius. Only the A and B coefficients are needed for the subrange 0 to 630 degrees. The R0 value is the resistance of the probe at 0 degrees Celsius.
All 5 of the custom RTD curves are set to PT385 at the factory, as shown in Table 2.
Table 2 - Default Custom RTD Coefficients
Curve Subrange R0 Coefficient A Coefficient B Coefficient C
1 0 to 630 100 3.9083e10-3 -5.775e10-7 0
2 -260 to 0 100 3.9083e10-3 -5.775e10-7 -4.183e10-12
3 0 to 630 100 3.9083e10-3 -5.775e10-7 0
4 -260 to 0 100 3.9083e10-3 -5.775e10-7 -4.183e10-12
5 0 to 630 100 3.9083e10-3 -5.775e10-7 0
Table 3 shows the coefficients for RTD types PT391 and PT392. The C coefficient is only used for temperatures below 0 degrees Celsius.
Table 3 - Other Common RTD Coefficients
RTD Type R0 Coefficient A Coefficient B Coefficient C
PT392 100 3.9848e10-3 -5.87e10-7 -4.0e10-12
PT391 100 3.9692e10-3 -5.8495e10-7 -4.2325e10-12
4.6 Standard Platinum Resistance Thermometer (SPRT) Coefficients
The SPRT function of the 3001 uses ITS-90 standard coefficients as a basis for measuring a SPRT. The five custom coefficients are entered as deviations from the standard coefficients, and as such, all of them are set to zero at the factory.
The coefficients A- and B- represent the A
4
and B4coefficient, obtained when the SPRT is
calibrated at the triple points of argon, mercury and water. This covers the 83.8058K to
273.16K subrange. Coefficients A, B and C can represent different coefficients based on which subranges of the SPRT has been calibrated. For example, if the 273.15K to
933.473K subrange was used, A, B and C would represent A
7
, B7and C7whereas if the
273.15K to 692.67K subrange was used, A and B would represent A
8
and B8and C=0.
To enter the deviation coefficients for a custom SPRT:
a) Select RTD measure mode as described in the preceding section.
20
b) Press the key until the SPRT type is selected.
c) Press the key to display the action prompt "SET(1)/RECALL(2)".
d) Press to select custom SPRT data entry.
e) At the "ENTER MIN TEMP" prompt, enter the minimum temperature limit for the
custom SPRT, and press the key.
f) At the "ENTER MAX TEMP" prompt, enter the maximum temperature limit for
the custom SPRT, and press the key.
g) At the "ENTER RTPW" prompt, enter the nominal resistance value (RTPW) for
the custom SPRT, and press the key.
h) At the "ENTER COEFF A" prompt, enter the first (A) deviation coefficient for the
custom SPRT, and press the key. To enter a coefficient that includes an
exponent, enter the mantissa, press the and keys to select the EXP
function, enter the exponent, and press the key.
i) When prompted, enter the second (B), third (C), fourth (A-), and fifth (B-)
deviation coefficients in the same manner.
j) To abort the SPRT entry without saving any changes, press the key.
To use a custom SPRT:
a) Select RTD measure mode as described in the preceding section.
b) Press the key until the SPRT type is selected.
c) Press the key to display the action prompt "SET(1)/RECALL(2)".
d) Press to recall the custom SPRT curve coefficients.
e) To use a different custom SPRT, press the key twice to display the SPRT
selection prompt.
4.7 Thermocouple (T/C) Measure
The 3001 can measure all common thermocouple types in °F or °C, plus basic millivolts from -10.0 to 75.0 mV.
The following common thermocouple types are supported:
B, C, E, J, K, L, N, R, S, T, U, XK, BP
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/O mode, if not already
21
selected. If RTD/O mode is displayed, press the key again to cycle to thermocouple mode.
c) If output mode is displayed, select input mode by pressing the and
keys.
d) Press the key to select the desired thermocouple type or the millivolt range.
e) Connect the unit under test to the thermocouple terminals of the 3001 using a
standard T/C miniplug as shown in figure 14. One pin is wider than the other; do not attempt to force the plug in the wrong polarization. The T/C wire used for the connection must match the thermocouple type selected for proper cold junction compensation. If the 3001 and the T/C miniplug are at different temperatures, wait at least 3 minutes for the miniplug and terminals to stabilize to the same temperature.
Figure 14 - Thermocouple Input Connection
f) Press the and keys to toggle the thermocouple display between °F and
°C.
g) Press the and keys to toggle the cold junction compensation between
the internal temperature sensor and an external reference.
h) For best accuracy, it is advisable to zero the T/C millivolt circuit daily, or if the 3001 is
being used outside of the ambient temperature range of 18 to 28 °C. The maximum offset from unit calibration that can be zeroed out is ±1 mV.
To zero the T/C millivolt circuit:
• Select the thermocouple measure function as described above, and press the
key until the millivolt range is selected.
• Insert the supplied thermocouple shorting jumper into the thermocouple terminals.
• Allow at least 3 minutes for the jumper and terminals to stabilize to the same temperature.
• Press the and keys to zero the T/C millivolt circuit.
22
4.8 Thermocouple (TC) Source
The 3001 can source all common thermocouple types in °F or °C, plus basic millivolts from -10.0 to 75.0 mV.
The following common thermocouple types are supported:
B, C, E, J, K, L, N, R, S, T, U, XK, BP
a) Disconnect any test leads from external devices.
b) Press the key to select thermocouple and RTD/mode, if not already
selected. If RTD/mode is displayed, press the key again to cycle to thermocouple mode.
c) If input mode is displayed, select output mode by pressing the and
keys.
d) Press the key to select the desired thermocouple type or the millivolt range.
e) Connect the unit under test to the thermocouple terminals of the 3001 using a
standard T/C miniplug as shown in figure 15. One pin is wider than the other; do not attempt to force the plug in the wrong polarization. The T/C wire used for the connection must match the thermocouple type selected for proper cold junction compensation. If the 3001 and the T/C miniplug are at different temperatures, wait at least 3 minutes for the miniplug and terminals to stabilize to the same temperature.
Figure 15 - Thermocouple Output Connection
f) Press the and keys to toggle the thermocouple display between °F and
°C.
g) Press the and keys to toggle the cold junction compensation between
the internal temperature sensor and an external reference.
h) Use the numeric keypad to enter the desired output value and press the key.
Alternatively, use the
or cursor key to select a digit to modify, followed by the
23
or cursor key to ramp the digit up or down. This method offers a simple
solution when small changes to an output value are required, or if specific decades need to be incremented or decremented.
i) For best accuracy, it is advisable to zero the T/C millivolt circuit daily, or if the 3001 is
being used outside of the ambient temperature range of 18 to 28 °C. This procedure is described in section 4.7 on thermocouple measurements.
4.9 Pressure Measure
The 3001 can support the following types of pressure modules:
• BETA Calibrators Corporation BETA Port-P Modules
• Fluke Corporation Model 700 Series Modules
• Mensor Corporation Model 6100 Modules
The BETA modules require the use of the BETA BPPA-100 Adapter and offer the best performance vs. cost. The adapter allows the user to "Hot Swap" a pressure module for quick range changes. For applications that require very high accuracy the Mensor 6100 Series, while expensive, will yield the best accuracy. Pressure modules from Fluke will plug directly into the 3001. It may be helpful to discuss your pressure needs with Martel Technical Support before you purchase modules.
a) Connect the pressure module to the 3001 as shown in figure 16.
Figure 16 - Pressure Module Connection
b) Press the key. The 3001 automatically senses which pressure module is
attached and sets its range accordingly.
c) Press the key to select the desired pressure units for display.
d) Before attaching the module to the pressure source, zero the module as described in
the instruction sheet that came with the module. Procedures vary, but all end with
pressing the and keys.
e) Attach the module to the pressure source according to the instruction sheet that
came with the module, taking care to follow all safety precautions when dealing with high pressures.
24
5. Isolated Inputs
5.1 Voltage Input
The 3001 can measure DC voltages from 0 V to 100 V, using the following two ranges for maximum accuracy: 10 V, and 100 V.
a) Disconnect any test leads from external devices.
b) Press the key to select isolated DC voltage and current input mode, if not
already selected. If the desired DC voltage mode is not displayed, press the key again to cycle to the desired DC voltage mode.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 17.
Figure 17 - Isolated DC Voltage Input Connection
5.2 Current Input
The 3001 can measure DC current from 0 mA to 50 mA.
a) Disconnect any test leads from external devices.
b) Press the key to select isolated DC voltage and current input mode, if not
already selected. If the DC current mode is not displayed, press the key again to cycle to it.
c) Connect the unit under test to the isolated voltage/current input terminals of the 3001
as shown in figure 18.
25
Loading...
+ 65 hidden pages