Martel IVC-222HPII Reference Manual

IVC-222HPII
Precision Voltage/Current
Calibrator
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Table of Contents
1.1 Customer Service . . . . . . . . . . . . . . . .2
1.2 Unpacking . . . . . . . . . . . . . . . . . . . . .2
2. Operating Procedure . . . . . . . . . . . . . . .3
3. Operating Considerations and
Precautions . . . . . . . . . . . . . . . . . . . . . . .4
4. Accuracy . . . . . . . . . . . . . . . . . . . . . . . . .6
5. Maintenance . . . . . . . . . . . . . . . . . . . . . .7
6. Rechargeable Battery Option . . . . . . .7
7. Specifications . . . . . . . . . . . . . . . . . . . .8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
1
IVC-222HPII
WARNING
Before attempting to interface the
IVC-222HPII with any other device, carefully
read the following instructions.
1. Introduction
The Martel Model IVC-222HPII is designed to be a simple, low cost voltage/current calibrator for field and laboratory purposes. It combines the latest digital and analog circuitry to achieve its rated specifications in a small, reliable pack­age.
The unit is capable of sourcing at ranges of 200mV, 2V, 20V, and 24mA. Each range is dis­played on the LCD readout. Upon power-up, the Model IVC-222HPII will return to the range selected before the unit was turned off.
This unit has the capability to store and recall up to two values in the setpoints provided by the SP1 and SP2 keys.
The keys and the s t keys help achieve needed values by selecting and scroll­ing digits.
The IVC-222HPII has an RS-232 jack which can be accessed via a custom interface cable avail­able from Martel.
s
s
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2
1.2 Unpacking
1. Upon receipt of shipment, inspect the
container and equipment for any signs of damage. Take particular note of any evidence of rough handling in transit. Immediately report any damage to the shipping agent.
NOTE: The carrier will not honor any claims unless all shipping material is saved for their examination. After examining and removing contents, save packing material and carton in the event re-shipment is necessary.
2. Remove the Packing List and verify that all
equipment has been received. If there are any questions about the shipment, please call Martel Electronics at 800-821-0023.
Set-up
Become familiar with the designations and polarities of the four output jacks located on the top of the calibrator. Refer to the back panel label for polarity information.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
2. Operating Procedure
1. After power-up, select the desired output
range. The 200mV, 2V, 20V, and 24mA ranges can be selected by pressing the RANGE key until the desired range is shown on the LCD readout.
2. Observe the proper polarity when
connecting the output jacks. Red is (+) and Black is (-).
3. Adjust output using the control keys.
When pressed, the keys select a digit on the LCD readout to be scrolled. When these keys are pressed and held down, a blinking cursor appears and covers up one of the digits indicating which digit is selected. After a second more, the cursor starts to move in the direction selected.
The s t keys scroll the digit up or down, respectively. As the outputs are scrolled, the integral LCD readout displays the voltage or current present at the output jacks.
4. The SP1 or SP2 keys will store or recall a
value when pressed. To store a value in SP1 or SP2, scroll to the
desired output. Press and hold down the SP1 or SP2 keys until the word “STORED” flashes onto the LCD readout.
To recall a value in SP1 or SP2: momentarily press the SP1 or SP2 keys, and the value needed from that particular setpoint should appear on the LCD display.
5. The voltage output jacks are disabled while
in the current output mode. Likewise, the current output jacks are disabled when in
3
s
s
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
the voltage output mode. Therefore, only one type of output, either voltage or current can be simulated at one time.
6. The purpose of the RS-232 jack on the
IVC-222HPII is to allow data from the IVC­222HPII to be transferred to a data collection device such as a PC or a printer. When connecting the IVC-222HPII to a computer, the following specifications will apply:
a. Baud Rate 2400 b. 8 data bits with 1 stop bit c. No Parity
Other commands can be sent to the IVC­222HPII from the computer to do all of the functions such as range change, output adjust, setpoint control, etc.
NOTE: A custom RS-232 cable from Martel
Electronics will be needed. The part number is C232SJ and can be ordered directly from Martel.
When connected to a computer, the IVC­222HPII will begin to send data as soon as the calibrator is turned on.
3. Operating Considerations
and Precautions
1. While operating in the voltage mode, the
calibrator’s output circuitry can sink or source up to 1mA max. Loads greater than 1mA may cause the calibrator to drop out of its accuracy specification. (See Figure 1)
When there is an excessive load, the word “OVERLOAD” will appear intermittently on the LCD readout. Long term short circuit
4
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
conditions will not harm the calibrator but will cause the battery to decay. Maximum short circuit is 25 mA.
Figure 1
IVC-222HPII operating as a Precision Voltage
Source
2. In the current mode, the maximum load
driving capacity of the IVC-222HPII is 1000 ohms. If connections to the mA jacks are open or the load is greater than the compliance voltage while trying to source a current, the word “OPEN LOOP” will appear intermittently on the LCD readout. (See Figure 2)
Figure 2
IVC-222HPII operating as a Current Source
3. Current loop devices that derive their power
supply from an external power supply can be calibrated as shown in Figure 3.
5
(+)
Volts
Volts
Max. Load
+
+
+
mA
+
mA
1 mA
O to 24mA
up to
1000 ohms
Unit
Under
Test
Unit
Under
Test
(–)
(+)
(–)
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
Figure 3
IVC-222HPII indicating current using an
external power supply
4. This instrument is equipped with an
automatic low battery indicator. When the LCD “BAT” indicator appears, the battery should be replaced immediately so that the calibrator’s output accuracy is maintained.
4. Accuracy
The IVC-222HPII is checked against an NIST traceable reference before shipment to verify that each range falls within the .015%, ±2 counts of accuracy specification. Long term accuracy, beyond one year, should remain within .030%, ±2 counts of FSR. All of these ratings are based on a 18°C to 28°C ambient temperature.
From 18°C to 0°C and from 28°C to 50°C, there is an accuracy change of .005% of full scale per °C. Therefore, allowances must be made for error caused by wide temperature variations.
6
(+)
Volts
Power
Supply
+
+
mA
4 to 20mA
(+)
Unit
Under
Test
30 Volts
Max.
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
5. Maintenance
Generally, with normal usage, the calibrator should hold its rated specifications for at least 12 months. Beyond this, it should be returned for re-calibration annually. If after the stated warranty period, the device falls out of calibration, or needs repair, it can be returned
6. Rechargeable Battery
Option
When continual service is required, a rechargeable “NI-Cad” 9V battery and wall mount AC adapter/charger may be purchased. The AC adapter may be left on continually to keep the charge up on the battery while the instrument is in use or can be used to recharge a low battery over a 12-14 hour period. The AC adapter/charger can also be used solely as an AC power supply without battery installed.
When using the “Ni-Cad” battery, it is advisable to occasionally allow the instrument to operate until the “BAT” indicator appears then recharge. This procedure will cycle the battery and prevent any “memory” effects that are associated with this type of battery.
7
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com
7. Specifications
Voltage Output mA Output Output Range: 200mV 2V 20V 24mA Resolution: 0.01mV 0.1mV 1mV 1mA Accuracy: 18°C to 28°C:
All voltage ranges:
±0.015% of Reading ±2 count
Current range 0 to 20mA:
±0.015% of Reading ±2 count
Current range above 20mA:
±0.030% of Reading ±2 count
Temperature 0 to 18°C Stability: 28 to 50°C
±50 ppm/C degrees
Load Capability: ±1mA for rated 1000 ohms
accuracy
Capacitive Load: No limitations N/A Output Protection: Current limited Internal Resettable fuse Power: 9V Alkaline battery or optional NiCd Case Dimensions: 1.43" x 3.15" x 5.7" Weight: 12 oz.
Specifications subject to change without notice
8
Find Quality Products Online at: sales@GlobalTestSupply.com
www.GlobalTestSupply.com