Martek Mark 25 Instruction Manual

Martek Instruments, Inc.
Mark 25
Dissolved Oxygen Analyzer
Instruction Manual
WARRANTY POLICY
MARTEK will provide free service at the factory, including parts, labor and transportation back to the customer, for any malfunction of its products which are returned transportation charges prepaid.
Customers desiring to return a product to MARTEK for repair should contact the Service Department by telephone at (800) 628-8834 or fax (919) 790-2375 or via our website at www.martekinstruments.com to obtain return authorization. The information required at this time will be the complete model number and serial number of the product and a brief description of the problem.
All shipments to MARTEK must be freight prepaid and addressed as follows:
MARTEK INSTRUMENTS, INC. 2609 Discovery Drive Raleigh, NC 27616 Attn: Repair Department
A complete and detailed statement of the reason for return must accompany the unit. If possible, include a copy of sample reading or a printout.
Returned units must be packed as well as they were when first shipped. If possible, use the original packing. Do not return detachable cords or manuals with the unit.
MARTEK reserves the right to void this warranty if the product has been subjected to misuse, neglect, accident, improper installation or application, and for consumable items such as batteries, membranes, or solutions.
This warranty is expressly in lieu of all other obligations or liabilities on the part of MARTEK. MARTEK neither assumes nor authorizes any other person or organization to assume on behalf of Martek any other liability in connection with the sales or use of MARTEK instrumentation.
1
MARK 25 TABLE OF CONTENTS
1. IntroductionPage
1.1 Description ……………………………………………………………………………….3
1.2 Specifications ……………………………………………………………………………4
2. Installation
2.1 Panel Cutout for Monitor ..………………………………………………………………5
2.2 Sensor & Flow Chamber Dimensions ..………………………………………………. 5
2.3 Electrical connections to the monitor ……………………………………………….…6
3. Operation
3.1 Keyboard & display ……………………………………………………………………….7
3.2 The CALIB Key...…….….…………………………………………………………………8
3.3 Calibrating Temperature.…………………………………………………………………8
3.4 Calibrating Dissolved Oxygen ..………………………………………………………….9
3.5 Setting Alarms…………….. …….………………………………………………………..9
3.6 Setting Recorder Outputs….……………………………………………………………10
3.7 Setting the Password..…….…………………………………………………….……….11
3.8 Setting the ID Number…………………….……………………………………………..11
3.9 Changing the Serial Output……………………………………………………….……..11
3.10 Setting the Auto-Ranging Function….…………………………………………….….12
3.11 Troubleshooting……………………….…………………………….…………………..13
2
1. Introduction
1.1 Description
The Mark 25 Dissolved Oxygen Analyzer is a simple, low cost, 4 channel, microprocessor-based monitoring system designed for ultrapure water applications. The system consists of a readout module that can be used with up to four galvanic-style dissolved oxygen sensors; each sensor being placed into a separate flow chamber.
A large 4 digit LED display indicates either temperature, parts-per-million (ppm) or parts-per-billion (ppb) dissolved oxygen. An 8-key membrane keypad allows the operator to quickly setup and calibrate the analyzer.
The Mark 25 measures dissolved oxygen in four ranges; 0-20 ppb, 0-200 ppb, 0-2 ppm, and 0-20 ppm. Automatic calibration within the Mark 25 provides a fairly reliable measurement. However, comparison to a referee system remains the most accurate method of calibration available. (Please refer to Martek’s Mark 21 Portable Dissolved Oxygen Analyzer for further information).
Four alarm set points with relay contacts are standard in addition to four analog recorder outputs (0-20 mA) and a serial RS232 digital output for hookup to external recording equipment. For maximum flexibility, alarm set points and analog outputs
can be assigned to any range on any parameter on any channel.
Housed in a ¼ DIN corrosion-resistant plastic case, the readout module can be panel mounted with a minimum of tools. The analyzer accepts power inputs from 97 to 240 VAC 50/60 Hz or 24 Vdc and comes with an ON/OFF switch located on the rear panel. Alarm and recorder wires are easily installed in removable terminal strips.
Sensors connect via a DB-9 connector located on the rear panel of the Mark 25 and the sensors mount easily to any flat surface. In addition, Martek offers adapter plates specifically designed to cover the space occupied by other brands of conductivity analyzers.
Sensors come standard with 6 feet (2 meters) of cable.
3
1.2 Specifications
ParameterRangeAccuracyResolution
Dissolved Oxygen0-20.00 ppb+0.1 ppb 0.01 ppb
0-200.0 ppb+1.0 ppb 0.1 ppb 0-2.000 ppm+0.01 ppm 0.001 ppm 0-20.00 ppm+0.1 ppm 0.01 ppm
Temperature0-50.0o C+0.1o C 0.01o C
Power: 97-240 VAC, 50/60 Hz, 24Vdc, fuse protected Recorder Output: Four, scaleable 0-20 mA DC analog & RS232C serial ASCII digital output Alarm Output: Four solid-state relays, 5 amp, 250 V Weight: 2 lbs (0.9 kg) including one sensor with 6 feet (2 meters) Construction: Readout Module Case: ABS plastic, NEMA 12 (IP65)
Sensor: Type 316 Stainless Steel, Delrin, CPVC, Platinum & lead, Teflon
Cabling: 8-conductor, 24 gauge, PVC jacket, shielded, 0.25 inch (0.63 cm) Dimensions: ¼ DIN case, 6.51 x 3.78 x 3.78 (165.5 x 96 x 96 mm) LWH
Installation: Panel mount (mounting hardware included). Recommend mounting flow chamber
and sensor vertically.
Flow rate: 50-250 ml/min
4
2. Installation
2.1 Panel Cutout for Monitor2.2 Flow Chamber Mounting Bracket
3.60” 2.20”
.266” dia.
3.60” 2.00”
1.25”
Panel Cut-out
2.3 Sensor & Flow Chamber Dimensions
2.25”
4.25” 3.50”
5
2.4 Electrical connections to the monitor
Alarm OutputsRecorder Outputs
RS232 Digital Output
RS232 1 2 3 4
ON/OFF Switch
AC Cord
Sensors
To set up the Mark 25 for operation, simply plug the sensor connector in, then plug the AC power cord into a properly grounded outlet. If recorder, digital or alarm outputs are to be used, make certain the Mark 25 is unplugged before connecting the cable to the appropriate recorder or computer. Follow the wiring diagram below for hooking up the Mark 25 alarm and analog recorder outputs:
ALARM RECORDER
Channel 1 NO/NC Channel 1 4-20 mA
COMMON
Channel 2 NO/NC Channel 2 4-20 mA
COMMON
Channel 3 NO/NC Channel 3 4-20 mA
COMMON
Channel 4 NO/NC Channel 4 4-20 mA
COMMON
6
3. Operation
o
3.1 Keyboard and display
The Mark 25 front panel consists of a 4 digit red LED display, 3 parameter and channel LED’s, an alarm LED, and an 8­key keypad. The display is used to present all parameter measurements, operator prompts and system errors.
o
C
PPM ALARM
PPB
DISSOLVED OXYGEN ANALYZER
READ Range CALIB
C
PPM MED
PPB HIGH RUN
LOW
ZERO
SPAN
www.martekinstruments.com
The keys are arranged in two rows, four keys in each row.
READ key - When the <READ> key is pressed, the Mark 25 will display rd. If one of the bottom keys is selected, the parameter high-lighted in red will be displayed and its corresponding LED lit.
RANGE key - Range changes can be made manually for dissolved oxygen. To display and/or change the range, press the <RANGE> key, then select one of the following keys:
LOW = rG 1: 0-20 ppb MED = rG 2: 0-200 ppb HIGH = rG 3: 0-2 ppm DOWN ARROW = rG 4: 0-20 ppm UP ARROW = rG A: Auto-ranging
NOTE: Manually changing the range for dissolved oxygen changes the analog recorder output scaling. See Section 3.6 for details.
The top row of keys consist of three color-outlined function keys and are used to perform instrument operations such as selecting which parameter to view, what range the parameter is to be viewed in, parameter calibration and system set up.
The bottom row of keys are multi­colored, multi-legend keys and are used in conjunction with the function keys. The same colored legends correspond to the same colored function key.
The arrow keys are used to view channels, select functions during setup and calibration, and increment or decrement a displayed value.
Viewing a different channel is easily done by pressing the <UP> or <DOWN> arrow key. The parameter LED’s to the left of the Mark 25 display also indicate which channel is being viewed. Channel ID’s are as follows:
Red LED - Channel 1 Green LED - Channel 2 Red/Green LED - Channel 3 Flashing Red LED - Channel 4
7
3.2 The CALIB key
All setup and calibration functions are conducted via the <CALIB> key. The <CALIB> allows you to set the following:
Calibrate temperature and dissolved oxygen
Assign and set alarm points
Assign and set recorder outputs
Set a password
Set up an analyzer ID and serial
output
The Mark 25 can be calibrated automatically or manually for temperature.
NOTE: Automatic, electronic calibration is a less accurate method of calibration. For best results, the Mark 25 should be calibrated to ASTM or NIST- traceable standards.
To calibrate temperature, press the <CALIB> key. The display will flash oC.
To select temperature, press the <UP ARROW> key. The flashing will stop. Next, press the <ZERO> key. The Mark 25 will display three calibration bars and automatically zero the temperature before returning to the monitoring mode.
Once the sensors have been connected to the Mark 25, the power cord can be plugged into any outlet supplying 97-240 V AC 50/60 Hz power. When the Mark 25 is switched on, it will immediately start monitoring temperature.
Pressing the <CALIB> key will cause the Mark 25 to display a flashing oC.
If you continue pressing the <DOWN ARROW> key the display will flash the following prompts: dO, ALA, rEC, PAS, uSr. After uSr, the Mark 25 returns to monitoring temperature.
Here are two basic rules to remember when using the <CALIB> key:
1. Pressing the <UP ARROW> key will select any flashing prompt.
2. Pressing the <RUN> key will return the Mark 25 to the monitoring mode.
3.3 Calibrating Temperature
To span temperature, press the <CALIB> key, then the <UP ARROW> key when the temperature prompt is displayed. Next, select the <SPAN> key. The Mark 25 will display Aut which is short for Automatic.
To automatically calibrate temperature, press the <UP ARROW> key. After the calibration bars appear, the Mark 25 will briefly display a span temperature of
50.0 before returning to monitor mode. The Mark 25 has now been calibrated
for temperature.
To manually calibrate temperature, simply press the <DOWN ARROW> key when Aut is displayed. The Mark 25 will momentarily display three lines, then a number indicating the input voltage will appear in the display. Once this number has stabilized, press the <CALIB> key and select the appropriate span value using the up or down arrow keys (the display defaults to 25 deg. C). When complete, press the <CALIB> key to
8
enter the data into the Mark 25’s memory.
3.4 Calibrating Dissolved Oxygen
Calibrating dissolved oxygen follows the same procedure as the temperature calibration. However, there are several differences.
Press the <CALIB> key, then the <DOWN ARROW> key until dO is flashing. Follow the same procedure for zero calibration for dissolved oxygen that was used for temperature.
To span dissolved oxygen, press the <CALIB> key and select the D.O. prompt. Next, press the <UP ARROW> key when the Aut prompt is displayed. The Mark 25 will display the calibration bars then briefly display 20.00 ppm before returning to the monitor mode.
NOTE: If the auto-ranging function is OFF while the Mark 25 is monitoring dissolved oxygen and the measurement exceeds the current selected range, the Mark 25 will display Err. If an error message occurs, you must select a higher range.
will display three bars. The bars indicate no alarm set point. Use the <UP> or <DOWN ARROW> keys to select temperature (t), or dissolved oxygen (do). Press the <CALIB> key to select the desired parameter.
The Mark 25 will then display Lo. To select a low set point, press the <CALIB> key. If a high alarm set point is desired, use the <UP> or <DOWN ARROW> key to change the display to hi then press the <CALIB> key to select.
The Mark 25 will display .000 (dissolved oxygen is selected) though the decimal point may vary depending on the range previously selected.
Use the <UP> or <DOWN ARROW> key to change the value to the desired alarm set point. When the desired set point is reached, press the <CALIB> key to set and activate alarm.
NOTE: For convenience, pressing the <RANGE> key will advance the digit left of the decimal point.
To disable alarm, return the set point value to .000.
3.5 Setting Alarms
The Mark 25 has four alarm set points that can be assigned to any parameter on any channel.
Press the <CALIB> until the display indicates the alarm prompt, ALA. Next, press the <UP ARROW> key. The display will prompt with AL 1 and all parameter LED’s will be lit. The <UP ARROW> key can be used to change the alarm set point to 2, 3, or 4.
To assign a particular alarm set point, press the <CALIB> key. The Mark 25
Once an alarm set point has been assigned to a particular parameter on a channel, it cannot be assigned to another parameter or channel until it has been disabled.
When the Mark 25 reaches an alarm state, the Alarm LED will light and the alarm relay activate. The factory-set position for the alarm relays is NORMALLY OPEN. To change to a NORMALLY CLOSED position, contact Martek Instruments.
NOTE: The Alarm LED will blink if there are active alarms that are not being
9
viewed and will remain steady if the channel viewed is in an alarm state.
3.6 Setting Recorder Outputs
The Mark 25 comes standard with four 4-20 mA analog recorder outputs and one RS232 digital output. Like the alarm set points, the four analog recorder outputs can be assigned to any range of any parameter on any channel. Press the <CALIB> until the display indicates the recorder prompt, rEC. Next, press the <UP ARROW> key. The display will prompt with ou 1 and all parameter LED’s will be lit. The <UP ARROW> key can be used to change the recorder output to 2, 3, or 4.
To assign a particular analog output, press the <CALIB> key. The Mark 25 will display three bars. The bars indicate no analog output assigned. Use the <UP> or <DOWN ARROW> keys to select temperature (t) or dissolved oxygen (do).
NOTE: A decimal point will be present for do which indicates which one of the four ranges the channel is currently using. The range can be changed by pressing the <RANGE> key.
When the desired parameter is displayed, the standard 4-20 mA output range can be changed to any value desired. In addition, a “balance” is provided to “fine-tune” each end of the recorder range.
To change the lower end of the recorder range, press the <ZERO> key. The Mark 25 will display 4.00. Use the <UP> or <DOWN ARROW> key to change the number to the desired value. When the desired value is reached, press the <CALIB> key to set the value in memory.
At this point, the Mark 25 will display
000. This is the “balance”. Use the <UP> or <DOWN ARROW> key to change the value in order to make the lower end of the recorder range agree with the external recorder. When the desired balance is reached, press the <CALIB> key to set the value in memory.
The same procedure is used to change and “balance” the upper range of the recorder output. To exit the menu, simply press the <RUN> key.
To disable any assigned analog recorder output, select the particular output and press the <CALIB> key. Once the parameter is displayed, use the <UP> or <DOWN ARROW> keys until three bars are displayed. Press the <CALIB> key to disable the output.
Once a recorder output has been assigned to a particular parameter on a channel, it cannot be assigned to another parameter or channel until it has been disabled.
NOTE: When the auto-ranging function is OFF, changing a dissolved oxygen range with an assigned analog output automatically changes the scaling of the recorder output to the new range. When the auto-ranging function is ON, the assigned analog output will remain in the range it was assigned to.
When the RS232 digital output is used, calibration is not necessary. However, the receiving computer must be set up to accommodate the Mark 25’s digital output configuration:
Baud Rate - 9600 Parity - Even Word length - 7 Stop Bit - 1
10
A communication program such as PROCOMM or Microsoft HYPERLINK can be used to write data from the Mark 25 to a computer file which can then be imported into a spreadsheet program such as EXCEL or LOTUS NOTES. The data will be displayed in the following format:
ID, Chan, Temp, Dissolved Oxygen.
3.7 Setting the Password
The Mark 25 can be assigned a numeric password to prevent unauthorized personnel from changing any of the settings.
Press the <CALIB> until the display indicates the password prompt, PAS. Next, press the <UP ARROW> key. The display will prompt with 1999 and all parameter LED’s will be lit and flashing. Use the <UP> or <DOWN ARROW> keys to change the displayed value to any number other than 0. When the desired value has been reached, press the <CALIB> key to set and activate the password.
Once the password is enabled, the display will default to 1999 whenever the password prompt is selected. To gain access to any other function, you must use the arrow keys to scroll to the correct value password then press the <CALIB> key.
To disable the password, use the <UP ARROW> key to move the value to 000 then press the <CALIB> key. The password function will then be disabled.
3.8 Setting the ID Number
The Mark 25 can be assigned a three digit numeric ID number that will be displayed on the digital output.
Press the <CALIB> until the display indicates the user prompt, uSr. Next, press the <UP ARROW> key. The display will prompt with id and all parameter LED’s will be lit and flashing. When the <CALIB> is pressed, the Mark 25 will display 000. To assign an ID number, use the <UP> or <DOWN ARROW> keys to change the displayed value to any number other than 000. When the desired value has been reached, press the <CALIB> key to set the ID number.
To disable the ID number, use the <UP ARROW> key to move the value to 000 then press the <CALIB> key. The ID function will then be disabled.
3.9 Changing the Serial Output
The Mark 25 provides a standard RS232 digital output. This output can be configured in one of two ways:
1. Polling - The Mark 25 will provide the digital output only when polled by the host computer.
2. Periodic - The Mark 25 will automatically provide a digital output at a 30 second interval.
Press the <CALIB> until the display indicates the user prompt, uSr. Next, press the <UP ARROW> key. The display will prompt with id and all parameter LED’s will be lit and flashing. Use the <UP> or <DOWN ARROW> key to advance to the serial prompt, SEr. When the <CALIB> is pressed, the Mark 25 will display PoL. To change the output to periodic interval, press either
11
the <UP> or <DOWN ARROW> key until the Mark 25 displays PEr.
To return to the monitor mode, press the <RUN> key.
3.10 Sensor Design & Maintenance
The dissolved oxygen sensor is a galvanic-style sensor using a platinum cathode, lead anode, potassium iodide electrolyte and a gas-permeable Teflon membrane.
5. Screw the Membrane Ring onto the Outer Body and gently tighten.
6. Using the dropper from the electrolyte bottle, carefully fill the four small flow-through holes located around the platinum tip of the Spindle.
7. Fill the Outer Body with electrolyte just above the interior ledge of the Outer Body.
The dissolved oxygen sensor consists of five principle parts:
Spindle
Outer body
Filler Valve
Membrane Ring
Teflon Membrane
Refer to the photos at the back of this manual for additional information.
The sensor is normally shipped dry from the factory without electrolyte or a Teflon membrane installed. The following procedure should be used when changing the membrane or replacing the electrolyte.
1. Turn the Filler Valve clockwise until it stops.
2. Unscrew the black Outer Body counter-clockwise and remove it from the gray Spindle. Empty out any electrolyte still in the Outer Body.
3. Unscrew the stainless steel Membrane ring from the Outer Body and remove the old membrane.
8. With the Filler Valve facing up at a 45o angle, screw the Outer Body back onto the Spindle and gently tighten (some electrolyte should flow out of the Filler Valve).
9. Gently screw out the Filler Valve until it stops (two or three threads should be visible).
10. Follow the normal calibration procedure.
If the dissolved oxygen sensor is not going to be immediately used, fill the red vinyl wetting cap with water and place over the Membrane Ring squeezing as much air from the cap as possible.
If the sensor is not to be used within two weeks, empty the electrolyte out of the sensor, dry all of the parts and store the probe in a cool place.
Normally, very little maintenance is required for the dissolved oxygen sensor and it can operate for weeks, even months, without the need to change the electrolyte or membrane, especially if the sample stream has little or no dissolved oxygen present.
4. Place a new membrane in the bottom of the Membrane Ring and place several drops of electrolyte onto the Membrane.
However, over time or in a sample stream with a high dissolved oxygen content, lead oxide will form on the lead
12
anode portion of the Spindle. A possible indication of lead oxide build-up is lower than normal readings due to less surface area on the lead anode. The lead strip will appear dull and possibly have yellow deposits near the edges.
mounted on-line, stainless steel fittings with ¼ inch stainless steel tubing is recommended. If the installation is a portable application, nylon or heavy-wall Tygon tubing may be used.
To clean the lead anode, simply take a toothbrush or plastic scouring pad (i.e. Scotch-Brite) and gently scrub the lead anode with distilled water until it shines. You should also check the platinum cathode tip and the four flow-through holes for dirt or debris. If any exists, use a fine wire to clean the holes.
NOTE: Do NOT scrub the platinum cathode. However, you can polish the
platinum tip using a polishing cloth. Lastly, make sure the o-rings used in
the Outer Body, Filler Valve, and Flow Chamber are intact and in good condition.
When placing the dissolved oxygen sensor into the stainless steel Flow Chamber, follow the steps outlined below:
1. Gently remove the red vinyl wetting cap from the end of the dissolved oxygen sensor.
2. Carefully insert the dissolved oxygen sensor into the flow chamber until it rests flush with the top of the chamber. Any twisting of the probe should always be clockwise. Place the locking nut over the dissolved oxygen sensor cable and secure it firmly onto the flow chamber.
3.11 Troubleshooting
The Martek Mark 25 represents over thirty years of experience in water quality measurements and, as such, has been designed to provide simple, reliable operation for many years.
To obtain optimum performance from the Mark 25, care should be taken to ensure the sensor has adequate flow and is free of obstructions. In addition, it is important to maintain correct calibration on a regular basis.
Comparison of the Mark 25 to a referee system is the preferred method of validation, however, calibration using the Mark 25’s internal circuitry is also acceptable.
In the event a problem develops with the Mark 25 that cannot be resolved through calibration or maintenance, contact Martek Instruments directly via telephone, fax, or e-mail located at our website, ww.martekinstruments.com.
3. To hook up sample flow to the Flow Chamber, place the inlet flow on the left fitting (with the mounting bracket facing away from you) and the outlet flow on the right fitting. If the Flow Chamber is to be permanently
13
14
Loading...