OPERATION MANUAL
MPMS-300/500
PLEASE TAKE THE TIME TO READ THESE INSTRUCTIONS BEFORE
STARTING TO USE THE SCALES

CONTENTS
• PRODUCT SPECIFICATION
• MAINTENANCE AND GENERAL CARE
SECTION 1 – BASIC FUNCTIONS
1.1 – SWITCHING ON
1.2 – SWITCHING OFF
1.3 – SETTING ZERO
1.4 – UNIT SWITCHING
SECTION 2 – ADVANCED FUNCTIONS
2.1 – HOLD FUNCTION
2.2 – SLOW FUNCTION
2.3 – BODY MASS INDEX FUNCTION
2.4 – PRINTING FUNCTION
SECTION 3 – CONFORMITY INFORMATION
SECTION 4 – ERROR MESSAGES

PRODUCT SPECIFICATION
MODEL MPMS-300/500
ACCURACY CLASS CLASS III
CAPACITY 300/500kg
INCREMENT 100/200g
BATTERY TYPE 7.2V 1200mah
MAINS ADAPTOR 12V 300ma
UNIT WEIGHT 8kg
MAINTENANCE AND GENERAL CARE
• Repairs and Servicing should only be carried out by authorised service
agents.
• There are no user serviceable parts in the weighing scale itself.
• The mains power adaptor contains 2 fuses, 1 internal that is not
accessible and 1 external user replaceable fuse. The fuse value is
T400mA and no other fuse value should be used.
• When the adaptor is plugged in to the mains supply a Green LED should
illuminate to indicate that the adaptor is functioning correctly. If the LED
does not illuminate then the external fuse should be checked by
competent personnel.
• There are no other user serviceable parts inside the mains adaptor.
• The weight indicator contains an internal rechargeable battery pack; this
should only be replaced by authorised service agents. Under normal
operating conditions the batteries should last for several years.
• We would recommend using alcohol based wipes or similar when cleaning
the scales.
• Please do not use large amounts of water when cleaning the scales as this
will cause damage to the scales electronics, you should also refrain from
using corrosive liquids or high pressure washers.
• Always disconnect the scales from the mains power supply before
cleaning.

OPERATOR INSTRUCTIONS - 10 KEY VERSION
1.1 - SWITCHING ON THE SCALE
SECTION 1 - BASIC FUNCTIONS
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
PRESS THE ON BUTTON FIRMLY
THE SCALE WILL FIRST TEST
ALL OF THE DISPLAY SEGMENTS
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
THE SCALE WILL NOW SHOW IT'S
CURRENT SOFTWARE VERSION No
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
18.8.8.8.8
P 1.10
THE SCALE WILL NOW GO INTO
WEIGHING MODE AND SHOULD
SHOW 0.0kg ON THE DISPLAY
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0

SECTION 1 - BASIC FUNCTIONS
1.3 - SETTING THE SCALE TO ZERO
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
1.8
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0
1.2 - SWITCHING OFF THE SCALE
PRESS THE OFF BUTTON
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
THE SCALE WILL POWER DOWN
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0
IF FOR SOME REASON THE SCALE SHOWS
A READING OTHER THAN 0.0kg THEN IT CAN
BE RESET TO ZERO (10% OF CAPACITY)
PRESS THE ZERO KEY ONCE
THE SCALE WILL RETURN TO 0.0kg

1.4 - UNIT SWITCHING
THE DEFAULT SETTING IS KG
PRESS THE UNIT KEY ONCE
THE SCALE SWITCHES TO STONES
PRESS THE UNIT KEY ONCE
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
16:12.0
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
220.0
THE SCALE SWITCHES TO POUNDS
PRESS THE UNIT KEY ONCE
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
100.0
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
100.0
THE SCALE NOW RETURNS TO KG
NOTE: UNIT SWITCHING IS NOT AL W AYS AVAILABLE, WEIGHING IN STONES AND POUNDS FOR THE PURPOSES OF
MON
ITORING, T REATMENT OR DIAGNOSES OF PATIENTS WAS MADE ILLEGAL ON THE 1ST JANUARY 2003. IF YOUR
SCALE HAS THE GREEN M STICKER T HEN ONLY KG WILL BE AVAILABLE
SECTION 1 - BASIC FUNCTIONS

SECTION 2 - ADVANCED FUNCTIONS
2.1 - HOLD FUNCTION
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0
PRESS THE HOLD BUTTON ONCE
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
Hold
THE PERSON SHOULD NOW GET ON TO
THE SCALES
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
100.0
AFTER A FEW SECONDS THE SCALE WILL
LOCK ON THE PERSONS WEIGHT, THE
PERSON CAN NOW STAND OFF THE SCALE
AND THE WEIGHT WILL REMAIN
PRESS HOLD AGAIN TO UNLOCK THE
WEIGHT AND RETURN TO 0.0kg
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0

2.2 - SLOW FUNCTION
SECTION 2 - ADVANCED FUNCTIONS
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
0.0
THIS FUNCTION IS SIMILAR TO THE HOLD FUNCTION BUT INSTEAD OF HOLDING THE WEIGHT
THE WEIGHT DISPLAY IS SLOWED DOWN CONSIDERABLY
AT 0.0kg PRESS THE SLOW BUTTON
PRESS THE SLOW BUTTON A SECOND
TIME TO DISABLE THE FUNCTION
THIS FUNCTION CAN BE VERY USEFUL WHEN WEIGHING DISABLED PATIENTS, CHILDREN OR BABIES

SECTION 2 - ADVANCED FUNCTIONS
2.3 - BMI (BODY MASS INDEX) FUNCTION
FOR THIS EXAMPLE WE WILL ASSUME WE HAVE A PERSON WEIGHING 100kg AND 180.0cm TALL
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
100.0
PLACE THE PERSON ON THE SCALE
PRESS THE BMI BUTTON ONCE
THE SCALE WILL NOW GO INTO HEIGHT
ENTRY MODE AND THE FIRST DIGIT WILL
FLASH - THE PERSON BEING WEIGHED
CAN STEP OFF THE SCALE AT THIS POINT
THE ARROW KEYS WILL ADJUST THE
FLASHING FIGURE UP OR DOWN AND THE
BMI KEY WILL MOVE TO THE NEXT DIGIT
USING THE UP/DOWN ARROWS AND THE BMI
BUTTON SET THE HEIGHT IN CM TO 180.0
AS SHOWN ON THE LEFT
WHEN FINISHED PRESS THE UNIT BUTTON
AND THE PERSONS BMI FIGURE WILL BE
SHOWN ON THE DISPLAY AS BELOW
PRESS THE BMI BUTTON ONCE TO RETURN
TO NORMAL WEIGHING MODE
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
POWER
BMI HOLD
PRETARE
NET WEIGHT
kg
st
lb
ON
OFF
ZERO
BMI
HOLD
UNITS
PRINT
SLOW
160.0
180.0
38.0

SECTION 2 - ADVANCED FUNCTIONS
2.4 PRINT FUNCTION
AN OPTIONAL EXTERNAL PRINTER IS AVAILABLE FOR ALL MODELS, WHEN FITTED THE
PATIENTS WEIGHT, HEIGHT, AND BMI RESULT CAN BE PRINTED ON TO A PAPER ROLL.
ONCE THE PERSON HAS BEEN WEIGHED AND THERE BMI CALCULATED, SIMPLY PRESS
THE PRINT KEY TO PRODUCE THE FOLLOWING TICKET
Gross Wei ght:
Tare Weight :
Net Weigh t:
Patient Height:
Patient BMI
100.0 kg
0.0 kg
100.0 kg
180.0 cm
38.0

ROHS COMPLIANCE
EU Directive 2002/95/EC restricts the use of the 6 substances below in the
manufacture of specified types of electrical equipment.
• The product does not contain any of the restricted substances in
concentrations and applications banned by the directive;
• and for components, the product is capable of being worked on at higher
temperatures required by lead-free soldering
The restricted substances and maximum allowed concentrations in the
homogenous material are, by weight:
Substance Concentration
Lead 0.1%
Mercury 0.1%
PBB (Polybrominated Biphenyls) 0.1%
PBDE (Polybrominated Diphenyl Ethers) 0.1%
Hexavalant Chromium 0.1%
Cadmium 0.01%
PRODUCT CONFORMITY
Harmonised Standards to Which Conformity is Declared –
EN60601-1-2 / EN60950 / EN55022 / EN50082-1
89/336/EEC – Electromagnetic Compatibility Directive
73/23/EEC – Low Voltage Directive
93/42/EEC – Medical Devices Directive
90/384/EEC – Non Automatic Weighing Instrument Directive
0120

ERROR MESSAGES
Lo
The internal battery is low on
charge, charge the scale
errl
Indicates that the signal from
the load cell/s is too low *
1
err
Indicates that the signal from
the load cell/s is too high *
1
errp
Indicates that there is a fault
with the scales software *
1
Oooooo
At top of LCD Display
Indicates that the signal from
the load cell/s is above the
Zero Range *
2
At bottom of LCD Display
oooooo
Indicates that the signal from
the load cell/s is below the
Zero Range *
2
*1 – These errors are normally caused by a serious fault on the
scales such as a faulty load cell or wiring. You should contact our
Service Department.
*2 – These errors are usually less serious and can be rectified by re-
calibration, however they can indicate an underlying problem with the
scales. Zero Range is set to 10% of Scale Capacity E.g. Capacity
200.0kg = Zero Range +/-20.0kg