127(Due to continuing product innovation, specifications in this manual are subject to change without
notice.
Listed below are GE Medical Systems Information Technologies trademarks used in this document. All other
trademarks contained herein are the property of their respective owners.
DASH and UNITY NETWORK are trademarks of GE Medical Systems Information Technologies registered
in the United States Patent and Trademark Office.
Unity Network
®
Interface Device (ID) is a trademark of GE Medical Systems Information Technologies.
Each page of this document has the document part number and revision
letter at the bottom of the page. The revision letter changes whenever
the document is updated.
Manual Purpose
This manual supplies technical information for service representatives
and technical personnel so they can maintain the equipment to the
assembly level. Use it as a guide for maintenance and electrical repairs
considered technician repairable. Whe re necess ary the manua l identifie s
additional sources of relevant information and technical assistance.
Introduction: Manual Information
RevisionDateComment
A20 November 2002Initial release.
Intended Audience
Naming Conventions
See the operator’s manual for the instructions necessary to operate the
equipment safely in accordance with its function and intended use.
This manual is intended for service representatives and technical
personnel who maintain, troubleshoot, or repair this equipment.
In this manual, the Unity Network Interface Device (ID) is referred to as
the Unity Network ID connectivity device.
Revision AUnity Network ID Connectivity Device1-3
2009517-002
Definitions
Introduction: Manual Information
Black textIndicates keys on the keyboard, text to be entered, or h ardware items
such as buttons or switches on the equipment.
Italicized textIndicates software terms that identify menu items, buttons, or options
in various windows.
Ctrl+EscIndicates a keyboard operation. A (+) sign between the names of two
keys indicates that you must press and hold the first key while
pressing the second key once.
For example, “Press Ctrl+Esc” means to press and hold down the
Ctrl key while pressing the Esc key.
<Space>Indicates you must press the spacebar. When instructions are given
for typing a precise text string with one or more spaces, the point
where the spacebar must be pressed is indicated as: <Space>. The
purpose of the < > brackets is to ensure you press the spacebar
when required.
EnterIndicates you must press the “Enter” or “Return” key on the
keyboard. Do not type “enter”.
1-4Unity Network ID Connectivity DeviceRevision A
2009517-002
Introduction: Safety Information
Safety Information
Responsibility of the Manuf acturer
GE Medical Systems Information Technologies is responsible for the
effects of safety, reliability, and performance only if:
n
assembly operations, extensions, readjustments, modifications, or
repairs are carrie d out by persons aut horized b y GE Medica l Systems
Information Technologies, Inc.;
n
the electrical installation of the relevant room complies with the
requirements of the appropriate regulations; and
n
the device is used in accordance with the instructions for use.
General
This device is intended for use under the direct supervision of a licensed
health care practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a
physician.
Contact GE Medical Systems Information Technologies before connecting
any devices or versions of devices not recommended in this manual to the
connectivity device.
Parts and accessories used must meet t he requireme nts of t he appli cable
IEC 60601 series sa fety st andard s, and/ or t he sy stem co nfigura tion mus t
meet the requirements of the IEC 60601-1-1 medical electrical systems
standard.
Periodically, and whenever the integrity of the device is in doubt, test all
functions.
The use of ACCESSORY equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice shall
include:
n
use of the accessory in the PATIENT VICINITY; and
n
evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate IEC 60601-1 and/or IEC
60601-1-1 harmonized national standard.
If the installation of the equipment, in the USA, will use 240V rather
than 120V, the source must be a center-tapped, 240V, single-phase
circuit.
Revision AUnity Network ID Connectivity Device1-5
2009517-002
Introduction: Safety Information
001
002
003
4P41
004
Warnings, Cautions, and Notes
The terms danger, warning, and caution are used throughout this
manual to point out hazards and to designate a degree or level of
seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person.
DANGER indicates an imminent hazard which, if not avoided, will
result in death or serious injury.
WARNING indicates a potential hazard or unsafe practice which, if not
avoided, could result in death or serious injury.
CAUTION indicates a potential hazard or unsafe practice which, if not
avoided, could result in minor personal injury or product/property
damage.
NOTE provides application tips or other useful information to assure
that you get the most from your equipment.
Equipment Symbols
The following symbols appear on the equipment.
ATTENTION: Consult accompanying documents before using the
equipment.
Equipotentiality
Power:
l = ON; O = OFF
Medical Equipment
With respect to electric shock, fire and mechanical hazards only in
accordance with UL 2601-1, and CAN/CSA C22.2 NO. 601.1.
1-6Unity Network ID Connectivity DeviceRevision A
2009517-002
Service Information
Service Requirements
n
n
n
n
n
Introduction: Service Information
Refer equipment servicing to GE Medical Systems Information
Technologies authorized service personnel only.
Any unauthorized attempt to repair equipment under warr anty voids
that warranty.
It is the user’s responsibility to report the need for service to GE
Medical Systems Information Technologies or to one of it s au thor ized
agents.
Failure on the part of the responsible individual, hospital, or
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and
possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure
that the equipment will always be functional when required.
Equipment Identification
Every GE Medical Systems Information Technologies device has a
unique serial number for identification. The serial number appears on
the product label of each unit.
D 0 XX 0005 GXX
Month
Manufactured
A = January
B = February
C = March
D = April
E = May
F = June
G = July
H = August
J = September
K = October
L = November
M = December
Year
Manufactured
0 = 2000
1 = 2001
2 = 2002
(and so on)
Product Code
Two-character
product descriptor
Product Sequence
Number
Manufacturing
number (of total units
manufactured)
Division
F = Cardiology
G = Monitoring
N = Freiburg
Hellige
Device Characteristics
One or 2 letters that further
describe the unit, for example:
P = prototype not conforming to
marketing specification
R = refurbished equipment
S = special product
documented under Specials
part numbers
U = upgraded unit
Revision AUnity Network ID Connectivity Device1-7
2009517-002
For your notes
Introduction: Service Information
1-8Unity Network ID Connectivity DeviceRevision A
2009517-002
2Equipment Overview
Revision AUnity Network ID Connectivity Device2-1
2009517-002
For your notes
2-2Unity Network ID Connectivity DeviceRevision A
2009517-002
Equipment Overview: System Components
System Components
System Operation
The Unity Network ID connectivity device acquires digital data from up
to eight peripheral devices, processes this data, and transmits the
formatted data to the monitoring network. The data can then be
displayed on a clinical information system, central station, and/or GE
Medical Systems Information Technologies patient monit or.
127(
Contact your GE Medical Systems Information Technologies
representative for information about the monitors and software
versions that are capable of receiving and displaying information
from a Unity Network ID connectivity device.
Peripheral devices include anesthesia machines, continuous cardiac
output monitors, gas analyzers, IV pumps, patient monitors, pulse
oximeters, transcutaneous monitors, urometers, and ventilators.
Stand-Alone Mode
The Unity Network ID connectivity device operates in either a standalone mode or a peripheral mode of operation.
In the stand-alone mode, the Unity Network ID connectivity device
provides data via the Unity Net work to one or more Unity Network
devices capable of viewing remote patient data.
Unity Network ID
Interface Device
Central
Station
Unity Monitoring Network
Peripheral
Adapter
Bedside
Devices
Patient
Monitor
Clinical
Information
System
To Other
Unity Network
Devices
Network
Gateway
To Clinical
Information
System
Network
PATIENT’S ROOM
005
The diagram is an example of patient data flow from peripheral devices
to a clinical information system:
Revision AUnity Network ID Connectivity Device2-3
2009517-002
Peripheral Mode
Equipment Overview: System Components
n
Each peripheral device requires a unique interface adapter to obtain
the patient data.
n
Each adapter is connected by cabling to one of the eight peripheral
device connectors on the Unity Network ID connectivity device.
n
The monitoring network delivers the patient data to a network
gateway that segregates the monitoring network from the clinical
information system. The gateway separates the real-time network
and the transaction-oriented network to assure reliable delivery of
real-time data.
n
The gateway sends the patient data to a clinical information system
or other parts of the hospital’s enterprise-wide network.
In the peripheral mode, the Unity Network ID connectivity device
provides its collected data to only one Unity Network device, such as a
GE patient monitor. While in this mode, other viewing devices on the
Unity Network cannot view data collected by the Unity Network ID
connectivity device. This mode of operation allows for data collected by
non-GE Medical Systems Information Technologies equipment to be
merged with data collected locally by a GE patient monitor.
Peripheral Mode Operation Using Dash Pro Patient Monitor
PATIENTíSROOM
DashPro
PatientMonitor
UnityNetworkID
InterfaceDevice
Adapter
UnityMonitoringNetwork
Peripheral
Adapter
Bedside
Devices
UnityNetwork
ToOther
Devices
006
2-4Unity Network ID Connectivity DeviceRevision A
2009517-002
Equipment Overview: System Components
Peripheral Mode Operation Using Solar 8000M/9500 Monitor
PATIENTíSROOM
Solar8000M/9500
Monitor
M-PortNetwork
Unity Network ID Components
UnityNetworkID
InterfaceDevice
UnityMonitoringNetwork
Peripheral
Adapter
Bedside
Devices
ToOther
UnityNetwork
Devices
034
Status LEDs
RS 232
Service
Connector
Power Switch
Power Connector
The Unity Network ID connectivity device components include:
I
0
Peripheral
Device
Connectors
Ethernet
Connector
u
RS 232 Service Connector—A dedicated serial port to
communicate with a personal computer or terminal for
configuration and programming.
u
Status LEDs— Eight bi-color light-emitting-diodes that indicate
proper system function, peripheral device compatibility
problems, and communication errors or delays.
u
Peripheral Device Connectors—Communication ports for eight
peripher al de v ic e s.
u
Ethernet Connector—A single communication port for the
monitoring network.
u
Power Connector—A receptacle for the AC power cord.
u
Power Switch—Power on/off switch with indicator light.
007
Revision AUnity Network ID Connectivity Device2-5
2009517-002
Interface Adapters
Equipment Overview: System Components
&$87,21
Use of the wrong interface a d apter may cause improper
operation of the supported peripheral device. Verify that
the interface adapter on th e peripheral device is the
correct adapter and is operational before the device is
used on a patient.
Interface
Adapters
008
Peripheral devices not manufactured by GE Medical Systems
Information Technologies require a special interface adapter for
connection to the Unity Network ID connectivity device. Dash Pro 3000/
4000 monitors require a special interconnect cable with integral interface
adapter. Interface adapters are available with different connector pin
configurations and are specifically programmed to allow communication
between the specific peripheral device and the Unity Network ID
connectivity device.
127(
Solar 8000M/9000 monitors do not require the use of interface
adapters for connection to the Unity Network ID connectivity de vice.
These monitors utilize a direct Ethernet connection. Refer to the
Installation chapter of this manual for further information.
127(
If your peripheral device software is updated, please complete and
fax the software upgrade notification page found in this manual.
2-6Unity Network ID Connectivity DeviceRevision A
2009517-002
Supported Devices
This section provides the li st of supported peripheral devices compatible
with the Unity Network ID connectivity device. The devi ces are
organized by type, and part numbers for the corresponding interface
adapter are provided.
127(
Equipment Overview: Supported Devices
Due to continuous product innovation, this list may no longer be
comprehensive. If necessary, call your sales representative for a
current list of supported peripheral devices.
Refer to the host monitor operator documentation for further
information about supported devices and monitored parameters.
Revision AUnity Network ID Connectivity Device2-7
2009517-002
Anesthesia Machines
Equipment Overview: Supported Devices
Supported Anesthesia Machines
Supported Device
North American Dräger
Medical Narkomed 2B
Anesthesia Workstation
North American Dräger
Medical Narkomed 2C
Anesthesia Workstation
North American Dräger
Medical Narkomed 2C
(with CO2+Gas)
Anesthesia Workstation
North American Dräger
Medical Narkomed 3
Anesthesia Workstation
North American Dräger
Medical Narkomed 4
Anesthesia Workstation
2-12Unity Network ID Connectivity DeviceRevision A
2009517-002
Equipment Overview: Supported Devices
Supported Ventilators
Nellcor Puritan Bennett
Infant Star Neonatal
Ventilator
Nellcor Puritan Bennett
Infant Star 500 Neonatal
Ventilator
Nellcor Puritan Bennett
Infant Star 950 Neonatal
Ventilator
Nellcor Puritan Bennett
Adult Star, Adult Star 1500,
and Adult Star 2000
Ventilators
Siemens Medical Servo
Ventilator (SV) 300
Siemens Medical Servo
Ventilator (SV) 900 C/D/E
Bear Medical Bear 1000
Ventilator
Hamilton Medical Veolar
Ventilator
46—VNTV1A420915-008
49 and 107—VNTV1A
107—VNTV1A
216 and 218—VNTV1A420915-009
COM-PROM V2.0
COM-PROM V2.01
1.0 and 2.0 of Servo
VNT Flow,
VNTV1A420915-011
VNT Pres
—VNTV1A420915-002
Computer Module 990
9.7 and A3.3VNT Flow,
VNTV1A420915-005
VNT Pres
E V31E.4 N31D.2 R030.0VNT Flow,
VNTV1A420915-007
VNT Pres
Hamilton Medical Amadeus
Ventilator
Hamilton Medical Galileo
Ventilator
Drager Medical Babylog
8000/8000SC Infant Care
Ventilator
Dräger Medical Evita
Intensive Care Ventilator
Dräger Medical Evita 2
Intensive Care Ventilator
Dräger Medical Evita 2
Dura Intensive Care
Ventilator
Dräger Medical Evita 4
Intensive Care Ventilator
Datex-Ohmeda Aestiva
3000 Anethesia Ventilator
Datex-Ohmeda 7800/7810
Anethesia Ventilators
A33X.0 N33A.6 R33A.0/
A33X.0 N33A.6 N01S.1
GMP01.21b GCP01.202
GPT01.00
Device v3.02, v4.02, v4.03,
v4.04, and v5.00
VNT Flow,
VNT Pres
VNT Flow,
VNT Pres
VNT Flow,
VNT Pres
VNTV1A
VNTV1A420915-060
VNTV1A420915-017
All with Medibus v3.00
Dräger Medibus Version
VNT PresVNTV1A420915-040
3.0 and Device Version 1.0
Dräger Medibus Version
3.0 and Device Version 1.0
Dräger Medibus Version
4.0 and Device Version 1.0
Dräger Medibus Version
4.0 and Device Version 1.0
VNT Pres,
CO2 Exp
VNT Pres,
CO2 Exp
VNT Pres,
CO2 Exp
VNT, CO2V1A420915-041
VNT, CO2V1A420915-042
VNT, CO2V1A420915-043
1.0—VNTV1A420915-050
1500-9001-000
—VNTV1A420915-019
CATV00557
Revision AUnity Network ID Connectivity Device2-13
2009517-002
Equipment Overview: Supported Devices
Supported Ventilators
Datex-Ohmeda 7900
Anethesia Ventilator
Bird VIP Infant-Pediatric,
6400ST, and 8400ST
Ventilators
2.8—VNTV1A420915-049
AnyV1A420915-020
2-14Unity Network ID Connectivity DeviceRevision A
2009517-002
Equipment Overview: Technical Specifications
Technical Specifications
Due to continuous product innovation, specifications for the Unity
Network ID connectivity device are subject to change without notice. The
following specifications are accurate as of the date of this publication.
Performance Specifications
Main ProcessorMotorola MC68EN360 25 MHz
NetworkType: Ethernet 10 BaseT
IndicatorsPower: Green LED on AC switch
Status LEDs: Eight bicolor LEDs
n
Solid green indicates OK
n
Any other condition (color or state) indicates caution
Device InputsNumber: Eight devices
EIA Standard: RS-232
Baud Rates: 22 standard baud rates 50-115.2k
28.8, 38.4k, 57.6k, 115.2k
Data Bits: 5 to 8
Stop Bits: 1 or 2
Parity: None, odd or even
Connector: RJ-45
Device ID Facility: Interface adapter provides identification and connector adaptation for “plug and
play”
Environmental Specifications
Power RequirementsInput voltage: 90 -264 V ac; 50/60 Hz; Single Phase
Operating ConditionsTemperature: 10°C to 40°C (50°F to 104 °F)
Humidity: 15% to 95% (noncondensing)
Storage ConditionsTemperature: –40°C to 70°C (40°F to 158 °F)
Humidity: 15% to 95% (noncondensing)
Physical Specifications
Height211 mm (8.3 in)
Width213 mm (8.4 in)
Depth66 mm (2.6 in)
Weight1.1 kg (2.4 lbs)
Revision AUnity Network ID Connectivity Device2-15
2009517-002
Equipment Overview: Technical Specifications
4P41
004
Certification
Safety StandardsMedical Equipment
With respect to electric shock, fire and mechanical hazards only in accordance
with UL 2601-1, and CAN/CSA C22.2 NO. 601.1.
EC 60601-1 certified.
CE marking for the 93/42 EEC Medical Device Directive.
Separation DeviceThe Unity Network ID connectivity device qualifies as a “Separation Device” as defined by the IEC
60601-1-1 Medical Electrical System Standard.
Classification
Type of protection against
electrical shock
Degree of protection against
electrical shock
Degree of protection against
harmful ingress of water
Degree of safety of application in
the presence of a flammable
anesthetic mixture with air or with
oxygen or nitrous oxide
Method(s) of sterilization or
disinfection recommended by the
manufacturer
Mode of operationContinuous operation
Class I
Not Applicable
Ordinary Equipment (enclosed equipment without protection against ingress of water)
Equipment not suitable for use in the presence of a flammable anesthetic mixture with air or with
oxygen or nitrous oxide.
Not Applicable
2-16Unity Network ID Connectivity DeviceRevision A
2009517-002
Loading...
+ 70 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.