Due to continual product innovation, specifications in this
manual are subject to change.
TRADEMARKSAND COPYRIGHT
Registered trademarks
Trademarks
Copyright
Trademarked names appear throughout this document. Rather
than list the names and entities that own the trademarks or
insert a trademark symbol with each mention of the trademarked
name, the publisher states that it is using the names only for
editorial purposes and to the benefit of the trademark owner with
no intention of improperly using that trademark.
ACCUSKETCH, APEX, AQUA-KNOT, ARCHIVIST, BABY MAC,
CASE, CD TELEMETRY, CENTRA, CRG PLUS, EAGLE, LASER
SXP, MAC, MAC-LAB, MACTRODE, MARQUETTE, MARQUETTE
UNITY NETWORK, MARS, MAX, MEI, MEI in the circle logo,
MINNOWS, MULTI-LINK, MUSE, PRESSURE-SCRIBE, PRES-RCUFF, RAMS, SAM, SEER, SOLAR, ST GUARD, TRAM, TRAMNET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, UNITY NETWORK,
and UNITY NETWORK twist are trademarks of Marquette Medical
Systems registered in the United States Patent and Trademark
Office.
12SL, 15SL, ADVANTAGE, AUTOSEQ, BODYTRODE, CardioServ,
CardioSmart, CardioSys, CD TELEMETRY®-LAN,
CENTRALSCOPE, EDIC, FLEX, HI-RES, IMAGE VAULT, INTERLEAD, LIFEWATCH, MARQUETTE® RESPONDER, MENTOR,
MRT -II, O2SENSOR, OnlineABG, SILVERTRACE, SMART-PAC,
and SMARTLOOK are trademarks of Marquette Medical Systems.
This field service manual has been prepared by the technical
publications staff at Marquette Medical Systems, Inc. It is
intended for use by biomedical electronic technicians or other
qualified service personnel responsible for installation,
maintenance or repair of the Eagle 3000 Patient Monitor
(hereafter referred to as the monitor).
The content of this field service manual is aimed primarily at
biomedical equipment technicians and field service personnel.
The user of this field service manual is expected to have a solid
background in electronics, including strong backgrounds in
analog and digital electronics, as well as microcomputer
technology familiarity.
The field service manual is organized into sections, as follows:
•Section one describes the field service manual, manual
page layout, related documentation, manufacturer
responsibility, notes/cautions/warnings, and abbreviation.
•Section two describes the product, the Marquette Unity
Network, technical specifications, preparation for use,
product part numbers and theory of operation.
•Section three describes the maintenance schedule, visual
inspection, cleaning the monitor, checkout procedures,
leakage current tests and hi-pot tests.
•Section four describes electrostatic discharge, special
components, power source tests, data acquisition tests,
service tips, the service mode menu and a network related
troubleshooting flow chart.
•Section five describes adjustments and non-invasive blood
pressure calibration.
•Section six describes monitor configurations, installing
software, loading software from diskettes, loading software
over the network and setup for use.
•Section seven provides assembly drawings of the monitor.
These include electrical diagrams, schematics, and
exploded views.
EAGLE 3000 PATIENT MONITOR
1 - 2415397-003PAGE REV C
Page 9
PAGE LAYOUT
ABOUT T HIS MANUAL
Section Topic: Each section is
divided into topics. This line
indicates what topic within the
section is covered on this and
possibly subsequent pages.
Left Column: Most pages are
split into two columns. The
left column text indicates topic
sub-titles and summeries of
text found in the right column.
Right Column: The right
column text provides topic
substance and elaborates on
information from text found in
the left column.
Visual Inspection
Inspecting the monitor
Section Title: The top line of
the page always indicates the
section of the manual. Section
topics may also appear next to
the section title.
M
AINTENANCE
The monitor should be carefully inspected prior to each patient
being admitted to the monitoring system. Follow these guidelines
when inspecting the equipment:
• Carefully inspect the monitor for obvious physical damage to
the outer case, display screen and controls. Do not use the
monitor if physical damage is determined. Refer damaged
equipment to qualified service personnel for repair before
using it again on a patient.
• Inspect all external connectors, front and rear, for degraded
pins, prongs and connector housings. Refer damaged
equipment to qualified service personnel for repair before
using it again on a patient.
• Inspect all cable insulation, cable strain-reliefs and cable
connectors for damage, cracks or degradation. Refer
damaged equipment to qualified service personnel for repair
before using it again on a patient.
Section & Page Number: The
number on the left indicates
the section, the number on the
right indicates the page within
the section.
Product Name - Manual Title:
This is found on each page of
the manual.
Page Revision: As changes to
the manual occur, this letter
indicates the current revision
for each page of the manual.
Page Rev A3-3
Eagle 3000 Patient Monitor
415397-003
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 3
Page 10
MANUAL REVISIONS
The following list gives the revision of each release of this
manual. As changes to the manual occur, the following list will
provide a reference to these changes.
Tram X00 Modules/Eagle Monitor Termination Instructions for
BP, CO, and TEMP Cables
•Part number: 403799-016
•Describes how to properly terminate the listed patient cable
connectors.
Eagle 3000 Patient Monitor - Version 3B Update Instructions
•Part number - 415397-060
•Describes methods for updating software in the monitor:
EAGLE 3000 PATIENT MONITOR
1 - 4415397-003PAGE REV C
Page 11
NOTES, CAUTIONSAND W ARNINGS
ABOUT T HIS MANUAL
WHATTHESEINDICATE
Notes, cautions, and warnings all appear in a similar fashion
throughout the manual. These are designed to draw special
attention to particular relevant points of interest.
NOTE
A note conveys special instructions to highlight an
operating procedure, practice, etc. Notes may pre-
cede or follow the applicable text, depending on the
material to be highlighted.
CAUTION
The purpose of a caution is to inform users of this
manual of operating procedures, practices, etc.,
which if not strictly observed, could result in possible
damage to the equipment.
WARNING
A warning provides instructions to users of the
manual that operating procedures, practices, etc., if
not followed, may result in personal injury.
PARTS LISTS
DIMENSIONSPECIFICATIONS
Hardware dimensions in parts lists use either metric or American
standards.
•Metric standards are indicated as items that include the
letter M as a prefix (example: Screw, M 2.0 x 4)
•American standards are indicated as items without a letter
as a prefix (example: Screw, 4-40 x 5/16)
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 5
Page 12
ABOUT T HIS MANUAL
MANUFACTURER
RESPONSIBILITY
EQUIPMENT SYMBOLS
Marquette Medical Systems is responsible for the effects on
safety, reliability, and performance only if:
•assembly operations, extensions, readjustments,
modifications, or repairs are carried out by persons
authorized by Marquette Medical Systems;
•the electrical installation of the relevant room complies with
the requirements of the appropriate regulations; and,
•the patient monitor is used in accordance with the
instructions for use.
Following are descriptions of the symbols used to identify various
safety concerns for patient monitoring equipment.
NOTE
Some symbols may not appear on all equipment.
ATTENTION: Consult accompanying documents
CAUTION: To reduce the risk of electric shock, do NOT remove
cover. Refer servicing to qualified service personnel.
Defibrillator-proof type CF equipment: type CF equipment is
specifically designed for applications where a conductive
connection directly to the heart is established. The paddles
indicate the equipment is defibrillator proof.
Defibrillator-proof type BF equipment: type BF equipment is
suitable for intentional external and internal application to the
patient, excluding direct cardiac application. Type BF equipment
is type B equipment with an F-type isolated (floating) part. The
paddles indicate the equipment is defibrillator proof.
Fuse
Equipotentiality
Alternating current (AC)
EAGLE 3000 PATIENT MONITOR
1 - 6415397-003PAGE REV C
Page 13
EQUIPMENT SYMBOLS (CONT)
ABOUT T HIS MANUAL
EQUIPMENTRATINGSFORPATIENT
APPLIED
PA RT S
PRESS
CF
BF
The following describes the symbols for patient applied parts of
patient monitoring equipment.
Power: I = On; O = Off
Indicates where to press to open the door on the 7160 DDW.
ECG and Respiration
Temperature (TEMP)
Invasive Blood Pressure (BP)
Pulse Oximetry (SPO2)
Non-invasive BP (NBP)
End-tidal CO2 (CO2)
NOTE
The rating of protection against electrical shock
(indicated by symbol for Type B, BF or CF) is achieved
only when used with patient applied parts listed for
use in the Supplies section of the Operator’s Manual.
Do not test the dielectric strength directly at the CO2
connector.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 7
Page 14
HOWTO REACHUS...
ABOUT T HIS MANUAL
The following are telephone numbers and addresses for contacting various Marquette Medical Systems Service and Supplies
Division personnel.
Ordering Supply Items
Ordering Service Parts
Supply items are generally items used during normal operation
of a product. Leadwires, electrode paste, patient cables, and
printer paper are examples of supply items.
•Make telephone inquiries about supply items at:
1-800-558-5102 (U.S. only)
1-407-575-5000 (outside the U.S.)
•Address orders or inquiries to:
Marquette Medical Systems Service and Supplies
P.O. Box 9100
100 Marquette Drive
Jupiter, FL 33468-9100
Attn: Supplies
Service parts are items that are not expended in the normal
operation of the product. They are generally replacements for
defective or malfunctioning items inside the product. Service
parts include PCB assemblies, electronic components, internal
cables and harnesses, software or firmware, and operator and
service manuals. When ordering additional operator manuals,
remember to notate the software version from the start-up
screen.
A part number for the item to be replaced is necessary for ordering a service part. If the part number for the desired item is
unobtainable, the following will be necessary to order the item:
•model and serial number of the equipment,
•part number/name of the assembly where the item is used,
To open a service call with Marquette Medical Systems Service,
contact a Service Dispatcher at:
1-800-558-7044 (U.S. only)
1-407-575-5000 (outside the U.S.)
For any questions about Service Contracts, contact the service
contract operator at:
1-800-552-3248 (U.S. only)
1-407-575-5000 (outside the U.S.)
EAGLE 3000 PATIENT MONITOR
Page 15
H
OWTO REACHUS... (CONT)
ABOUT T HIS MANUAL
Technical Support
For All Hardware
Telemetry
Series 7000/7010
Monitoring Technical Support has a broad base of information
regarding Marquette patient monitoring equipment and provides
assistance with technical questions and concerns.
For technical information regarding Marquette patient monitoring equipment, contact Monitoring Technical Support at:
1-800-558-7822 (U.S. only)
1-407-575-5000 (outside the U.S.)
For technical information regarding Marquette telemetry patient
monitoring equipment, contact Monitoring Technical Support —
Telemetry Products at:
1-800-552-3243 (U.S. only)
1-407-575-5000 (outside the U.S.)
For technical advice concerning Series 7000/7010 patient monitoring equipment, contact Tech Support:
1-800-443-0980 (U.S. only)
1-407-575-5000 (outside the U.S.)
48-Hour Turnaround Repair
Service Address
For Additional
Information
Some Marquette patient monitoring products or components
such as Series 7000 Input Modules, Tram Modules, Series 7700
ECG Telemetry Transmitters, and CD Telemetry Transmitters are
repaired on a 48-hour turnaround basis.
For information regarding 48-hour repair items, contact Monitoring Technical Support — 48-Hour Repair at:
1-800-552-3243 (U.S. only)
1-407-575-5000 (outside the U.S.)
Send all patient monitoring repair items to:
Marquette Medical Systems Service and Supplies
P.O. Box 9100
100 Marquette Drive
Jupiter, FL 33468-9100
Attn: Monitoring Technical Support — Repair
The service telephone operator can direct calls to personnel most
able to assist with information. For other Monitoring Technical
Support information, contact the service telephone operator at:
1-800-558-5120 (U.S. only)
1-407-575-5000 (outside the U.S.)
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 9
Page 16
ABBREVIATIONS
ABOUT T HIS MANUAL
A
AAMI: Association for the
Advancement of Medical
Instrumentation
AC,ac: alternating current
ADC: analog-to-digital
converter
Adj: adjustable
Al: aluminum
Ampl: amplifier
ANSI: American National
Standards Institute, Inc.
ASIC: application specific
integrated circuit
ASYNC COMM: asynchronous
communication
AUI: attachment unit interface
Ave: Avenue
AWG: American Wire Gage
B
B/M: beats per minute
BDGH: binding head
BP: blood pressure
bpm: beats per minute
BT: blood temperature
General monitor block theory .................................... 2-18
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 1
Page 20
PRODUCT DESCRIPTION
EQUIPMENT OVERVIEW
ABOUTTHEMONITOR
Cost-effective design
Standard monitoring
configuration ...
... with options available
“Value-added” features
Network compatibility
The monitor provides the parameters needed in a single,
inexpensive, easy-to-use device. Its compact size and innovative
package allow it to fit in small, tight places making it an ideal
choice for patient monitoring in an operating room, recovery
room, emergency care area as well as outpatient care areas.
Designed to provide high
reliability, long operational
life, minimal downtime,
and low maintenance cost,
the monitor is a costeffective solution to
general purpose patient
monitoring needs. Monitor
configurations include
simultaneous multi-lead
ECG, non-invasive blood
pressure, dual
temperature, pulse
oximetry monitoring.
Optional features that can
be added to all monitors
include an integral twochannel two inch thermal writer, two invasive blood pressures
and end-tidal CO2.
The monitor is equipped with a 9" diagonal high-contrast
electroluminescent (EL) display capable of showing four
waveforms and full digital data for all monitored parameters.
The Trim Knob control and a clinically logical software menu
structure assure ease-of-use and reduced inservice time. Multilead ST-segment monitoring, high quality graphic and tabular
trends, and sophisticated algorithms (which may assist to reduce
false alarms and provide more accurate data) are "value-added"
features. In addition, the monitor is a member of a complete
family of critical care products from the manufacturer, and is
fully compatible with interconnection to the patient monitoring
network.
EAGLE 3000 PATIENT MONITOR
2 - 2415397-003PAGE REV C
Page 21
PRODUCT DESCRIPTION (CONT)
FRONTPANELDESCRIPTION
NOTE
To insure patient safety, use only parts and accessories manufactured or recommended
by the manufacturer. Parts and accessories used must meet the requirements of the
applicable IEC 601 series safety standards, and/or the system configuration must meet
the requirements of the IEC 601-1-1 medical electrical systems standard.
EQUIPMENT OVERVIEW
Patient Input Connectors: Used to
attach patient cables for various
electrodes, sensors and transducers used
in patient signal acquisition
Display:
Monochrome electroluminescent (EL) display panel.
Screen size: 9-inch diagonal
Resolution: 640 x 400 pixels
Trim Knob Control: This is the
control that is used most often to
choose menu items and enter
data.
Rotate the Trim Knob
control to highlight an
item on the display.
Press the Trim Knob
control to select the
highlighted item.
Front Panel Controls: Five backlit
pushbutton operator controls provide
the following functions:
DEFIB SYNC (option): This side panel
jack provides a direct interface between
the monitor and a defibrillator for
synchonization of the two devices during
emergency defibrillation or for
synchronized cardioversion. The signals
available through this connector are:
Outputs —
•Defib sync pulse
•Analog ECG signal
•Analog invasive BP signal
Input —
•Defibrillator triggered marker pulse
RMT ALM (option): This side panel
connector provides interconnection
through the Marquette/Hellige Isolation
Relay to a nurse-call light system.
CO2 Sensor (option): These side panel
connectors provide a direct interface
between the monitor and an End-tidal
CO2 monitoring kit.
NBP Connector: A
pneumatic connector
for attaching a
noninvasive blood
pressure cuff to the
monitor.
Controls power to the
display without disrupting
monitor main power.
Manually starts or stops the
noninvasive blood pressure
function.
Sets zero references for all
invasive blood pressure
functions.
Controls patient alarm
silencing functions.
Manually starts or stops
graphs to selected writers.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 3
Page 22
PRODUCT DESCRIPTION (CONT)
REARPANELDESCRIPTION
NOTE
The use of ACCESSORY equipment not complying with the equivalent safety requirements
of this equipment may lead to a reduced level of safety of the resulting system.
Consideration relating to the choice shall include: use of the accessory in the PATIENT
VICINITY; and evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized
national standard.
EQUIPMENT OVERVIEW
Main Power Switch:
When in the on (I)
position, ac power is
applied to the
monitor. When in the
off (0) position, ac
power is removed
from the monitor.
RMT ALARM (remote
alarm) Connector:
Interconnection to the
Marquette/Hellige
Isolation Relay for a
nurse-call light
system.
*Serial Number Label:
(not shown) Describes
the type of equipment,
date and sequence of
product manufacture
for each monitor.
Main Power: A power
cord connected
between this port and
a wall receptacle is
used to provide ac
main power to the
monitor.
* A sample of the information found on the
Serial Number Label is shown below.
F 7 XX 0109 G
ETHERNET Connector:
An IEEE 802.3 AUI cable
and transceiver can be
connected to this port for
monitors used in patient
monitoring network
configurations.
ASYNC COMM
Connector: This
port can be used for
interconnection to a
remote control,
direct digital writer
or other device.
Month
YearProduct
code
A thru M = January thru December
Letter I is not used.
Product
sequence
number
Division identifier
F = Cardiology
G = Monitoring
EAGLE 3000 PATIENT MONITOR
2 - 4415397-003PAGE REV C
Page 23
PRODUCT DESCRIPTION (CONT)
EQUIPMENT OVERVIEW
ABOUTTHEREMOTEALARM
CONNECTOR
The remote alarm (RMT ALRM) connector is for use with a
Marquette/Hellige Isolation Relay (pn 303 555 77) which
provides a relay closure during the following alarms:
•CRISIS Patient Status Alarms, and
•WARNING System Status Alarms.
The signals provided at the RMT ALRM connector activate and
deactivate the Marquette/Hellige Isolation Relay (shown below).
When the monitor is powered up initially or rebooted, the relay
remains turned off until the monitor is finished with its power up
or reboot sequence. Once the monitor finishes powering up or
rebooting, the relay is then energized.
When a CRISIS Patient Status Alarm or a WARNING System
Status Alarm is detected, the relay is turned off by a signal from
the monitor. When the alarm has been cleared, the relay is
turned on once again.
The relay is turned off when AC power is removed from the
monitor.
• Marquette/Hellige Isolation Relay •
NOTE:
Refer to the Installation Instructions that come with
the Marquette/Hellige Isolation Relay for proper
connection to your Nurse Call system.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 5
Page 24
FOR Y OUR NOTES
EQUIPMENT OVERVIEW
EAGLE 3000 PATIENT MONITOR
2 - 6415397-003PAGE REV C
Page 25
MONITOR APPLICATIONS
EQUIPMENT OVERVIEW
STAND-ALONEMONITOR
APPLICATION
PATIENTMONITORINGSYSTEM
APPLICATION
HOSPITAL-WIDENETWORK
APPLICATION
The Marquette Unity Network (hereafter referred to as the
network) provides a method for standardized communication
with various Marquette medical system devices. This versatile
monitor can operate both as a fully functional stand-alone device
and as a component on the network, depending upon the
application.
When connected to the network, the monitor provides access to
other devices for many purposes. Marquette patient monitoring
equipment such as Centralscope central station monitor; Series
7200/7260 direct digital writer; CDT-LAN patient telemetry
system; ADU-LAN; and, Solar or other Eagle patient monitors are
examples of devices that can be used in conjunction with the
monitor when connected to the network.
There are various types of information management and data
base systems devices which may also be integrated with the
monitor via connection to the network. Marquette medical
systems equipment such as MUSE cardiology management
system; MARS UNITY workstation; MARS 24 clinical review
station; MRT II automated vital sign and arrhythmia data
collection system; MAC-Lab cardiac catheterization system; QMI
patient data management system; and, MUSE HIS interface are
examples of systems and data bases which can be integrated
with the monitor on the network.
Patient monitoring system
application
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 7
Page 26
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS
Display
Size:9-inch diagonal
Type:Monochrome, electroluminescent, flat panel
Resolution:640 by 400 pixels
Number of waveform traces:4
Length of trace:5.7 seconds
Sweep speed:22.9 mm/s (with erase bar), all waveforms except CO2
Waveform display options:Full or individual
Information window:Displays all non-real-time information without obstructing the
display of real-time information
Display organization:Prioritized by parameter
Processing
Main processor:MC68EN360 (32-bit microprocessor), 23.4936 MHz clock
frequency
Data acquisition processor:MC68332 (32-bit microprocessor), 15.7248 MHz clock frequency
Program memory:2MB flash EEPROM
Data memory:2MB RAM
Alarms
Classification:4 levels (Crisis, Warning, Advisory, Message)
Notification:Audible and visual
Setting:Default and individual
Silencing:1 minute, current alarm only
Pause
Adult mode:5 minutes
Neonatal ICU mode:3 minutes
OR mode5 minute, 15 minute, or permanent
Volume:Default 70%, 70 dB measured at 1 meter
User Interface:
Trim Knob control:Provides access to all menu-based operations
Hard keys
Silence Alarm:Controls alarm silencing
Graph Go/Stop:Start and stop manual graph operation
NBP Go/Stop:Start and stop non-invasive blood pressure measurement
Zero All:Zero BP transducers
Display On/Off:Blanks the display and disables alarms, for use as screen saver
EAGLE 3000 PATIENT MONITOR
2 - 8415397-003PAGE REV C
Page 27
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
ECG
Standard leads available:I, II, III, V, aVR, aVL, and aVF
Leads analyzed simultaneously: I, II, III, and V (multi-lead mode)
Lead fail:Identifies failed electrodes
Lead fail sensing current
Active electrodes:<25 nA each
Reference electrode:<100 nA
Waveform display aspect ratio:0.34 s/mV
Input specifications
Voltage range:±0.5 mV to ±5 mV
Signal width:40 ms to 120 ms (Q to S)
Input impedance
Common mode:>10 MΩ at 50/60 Hz
Differential:>2.5 MΩ from dc to 60 Hz
Output specifications
Display frequency response
Adult ICU mode:0.05 to 40 Hz
Neonatal Mode:0.5 to 40 Hz
OR Mode:0.05 to 25 Hz
Paper Recorder frequency response
Standard Mode:0.05 to 100 Hz
Neonatal:0.5 to 40 Hz
OR Mode:0.05 to 25 Hz
Common mode rejection:90 dB minimum at 60 Hz
Gain:1000 ±3%
Linearity deviation:±3% (maximum)
Noise:<30 µV (referred to input)
Heart Rate
Heart rate range:30 to 300 beats per minute
Heart rate averaging:8 beats
Display update interval:2 seconds
Response time:<6 seconds (per AAMI EC13)
Limit alarm display:<10 seconds after alarm condition exceeded
ST Segment analysis
Measurement description:ST segment deviation measured and displayed for all acquired
leads, and averaged for anterior, lateral and inferior lead groups
ST display:Three 30-minute trends, or ECG complexes for leads I, II, and V,
and a summation trend
Measurement point:Measures at 60 ms following the J point
Measurement range
Adult:-12.0 mm to + 12.0 mm
Neonatal:-10.0 mm to + 10.0 mm
Display resolution:0.1 mm
ST measurement averaging: 16 beats
Display update interval:2 seconds
Pacemaker detection/rejection
Input voltage range:±2 mV to ±700 mV
Input pulse width:0.1 ms to 2 ms
Rise time:10 µs to 100 µs
Over/under shoot:2 mV (maximum)
Baseline drift:<0.5 V with a ±700-mV, 2-ms pacemaker pulse applied
Alarms:Selectable upper and lower heart rate limits, arrhythmia
detection; lead-failure
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 9
Page 28
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
Invasive Blood Pressure (optional)
Number of channels:2
Transducer sites:Arterial, pulmonary arterial, central venous, left atrial,
intracranial, and special
Transducer requirements
Excitation voltage:±2.5 V dc ±0.1%
Transducer output:50 µV/V/cm Hg
Pressure measurement specifications
Range:-25 mmHg to 300 mmHg
Accuracy:±2% or ±1 mmHg, whichever is greater (independent of
transducer)
Drift:< ±0.1%/°C, and < ±0.1% over and 24-hour period
Zero balance range:±150 mmHg
Zero balance accuracy:±1 mmHg
Zero balance drift:±1 mmHg over 24 hours
Frequency response:dc to 50 Hz (-3 dB)
Input impedance
Common mode:100 KΩ (minimum) at 50/60 Hz
Differential:100 KΩ (minimum) from dc to 60 Hz
Common mode rejection:60 dB (minimum) at 60 Hz
Noise:5 mV peak to peak (maximum) from dc to 30 Hz
Averaging:4 seconds
Display update interval:2 seconds
Alarms:Selectable upper and lower limits for systolic, diastolic, and mean
pressures; sensor failure
Non-invasive Blood Pressure
Measurement technique:Oscillometric
Displayed parameters:Systolic, diastolic, and mean pressures, pulse rate, time of last
measurement
Measurement modes:Manual, auto, and stat
Systolic pressure range:30 to 275 mmHg
Diastolic pressure range:15 to 250 mmHg
Heart rate range:30 to 300 beats per minute
Total cycle time:20 to 40 seconds typical (dependent on heart rate and motion
Overpressure safety valve:Activates when cuff pressure exceeds:
Adult mode:300 mmHg (+30/-0 mmHg) cuff pressure
Neonatal mode:150 mmHg (+15/-0 mmHg) cuff pressure
Maximum pressure leakage:4 mmHg per minute
Automatic cycle times:0 to 24 hours
Auto zero:Zero pressure reference prior to each cuff inflation
Tubing length:12 feet adult, 8 feet neonatal
Automatic cuff deflation:Occurs when power is off or the following limits are exceeded:
Adult mode:300 mmHg (± 3 mmHg) cuff pressure or 3 minutes cycle time
Neonatal mode:150 mmHg (± 2 mmHg) cuff pressure or 90 seconds cycle time
Cuff sizes:
Disposable:Large adult, adult, small adult, pediatric, small pediatric, and
infant
Reusable:Thigh, large adult, adult, child, and infant
Alarms:Selectable upper and lower limits for systolic, diastolic, and mean
Range:0 to 200 breaths per minute (for variation of 1.0 to 10Ω)
Accuracy:± 1 breath per minute
Base impedance:100 to 1000Ω
Detection sensitivity:0.4 to 10 Ω variation (for 0 to 120 breaths per minute)
Excitation current:250 µA
Averaging:8 breaths
at 52.6 kHz
RMS
Display update interval:2 seconds
Waveform display bandwidth:0.1 to 1.8 Hz (-3 dB)
Lead fail:Indicated when base impedance exceeds 1750 ± 250Ω
Apnea detection:Indicated when impedance variation is less than selected
sensitivity or 0.2Ω, whichever is greater
Alarms:Selectable upper and lower respiration rate limits, and user
selectable apnea limit
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 11
Page 30
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
End-tidal C02 (optional)
Monitoring functions:Inspired and expired CO2 measurements, respiration rate
Adult reusable:<5 cc
Adult disposable:<5 cc
Neonatal:<0.5 cc
Sampling:<0.2 cc
CO2 measurement specifications
Range:0 to 100 mmHg
Accuracy
Mainstream:±2 mmHg or 5%, whichever is greater
Sidestream mode:±3 mmHg or 8%, whichever is greater
Display update interval:2 seconds
CO2 averaging:Selectable from single breath, 10 seconds, or 20 seconds
Resolution:1 mmHg
Response time (for 5% step size)
Mainstream adult:<60ms (10 to 90%)
Mainstream neonate:<50 ms (10 to 90%)
Sidestream (with filter):<600ms (10% to 90%) at 180 cc/min
Interference
N2O gas:±2 mmHg or 5%, whichever is greater, with N2O compensation
enabled
O2 gas:±2 mmHg or 5%, whichever is greater, with O2 compensation
enabled
Barometric pressure:±2 mmHg (maximum) from 500 to 800
Water vapor:±0.5 mmHg or 1.5% (maximum), whichever is greater
Anesthetic agent:±0.5 mmHg (maximum), for concentrations of no more than 5% of
halogenated agents
Warm-up time:Less than 15 seconds to initial CO2 indication, full specification
within 60 seconds
Calibration
Factory setting:Factory calibration settings stored in nonvolatile memory within
the sensor. 15 second adapter calibration when switching airway
types
Verification:Zero and reference performance check with on-cable verifier
CO2 Waveform sweep speed: Selectable one-fourth, one-half, and full speed
Respiration rate specifications
Range: (for 5% step size)
Mainstream mode:0 to 120 breaths per minute
Sidestream mode:0 to 50 breaths per minute
Accuracy:±1 breath per minute
Resolution:±1 breath per minute
Sidestream pump specifications
Flow rate:190 ±10 cc/min
Capnostat III sensor specifications
Operating temperature:10 to 40°C
Storage temperature:-30 to 65°C (-22 to 149°F)
Humidity:5 to 95%, relative humidity, non-condensing
Barometric pressure:500 to 800 mmHg
Shock resistance:Able to withstand 6 ft. drop to tile floor
Moisture resistance:Splash resistant sealed transducer
Cleaning and sterilization
Sensor:Transducer, cable, and verifier may be wiped with cold chemical
disinfectant: no steam sterilization, no ETO gas permitted, do not
immerse in fluid
Reusable airway
adapters:Disinfect with buffered glutaraldehyde, ETO gas, isopropyl
alcohol, household bleach; also steam sterilization and
pasteurizable
Alarms:Selectable upper and lower limits for CO2 and respiration rate
400 dots/in at 50 mm/sec
Vertical resolution:200 dots/in
Number of waveform
channels:2
Paper width:50 mm (1.97 in)
Paper speed:1, 5, 10, 12.5, 25, 50 mm/sec
Paper speed accuracy:±5% at 1 and 5 mm/sec; ±2% at 10 mm/sec or faster
Frequency response:Determined by acquisition system
Direct Digital Writer (optional)
Method:Thermal, fixed head
Horizontal resolution:480 dots/in at 25 mm/sec;
Vertical Resolution:200 dots/in
Number of Waveform
Channels:4
Paper width:
7100:108 mm (4.3 in)
7160:50 mm (1.97 in)
Paper length:
7100:46 m (150 ft)
7160:25 m (95 ft)
Paper speed:1, 5, 10, 12.5, 25, 50 mm/sec
Paper speed accuracy:±5% at 1 and 5 mm/sec; ±2% at 10 mm/sec or faster
Frequency response:determined by acquisition system
Analog Output (optional)
ECG
Gain:1 V/mV ±10%
DC offset:±100 mV (maximum)
Noise:5 mV peak to peak (maximum) 0 to 300 Hz
Frequency response:0.05 Hz to 100 Hz –0/+7 Hz
Blood pressure
Gain:10 mV/mmHg ±2%
DC offset:±20 mV (maximum)
Noise:5 mV peak to peak (maximum) 0 to 300 Hz
Frequency response:dc to 50 Hz -0/+2 Hz
Defibrillation Synchronization Pulse (optional)
Marker out
Time delay (R wave peak
to leading edge of pulse):35 ms (maximum)
Amplitude
+5 V selection:3.5V (min) at 1 mA sourcing: 0.5V (max) at 5 mA sinking
+12 V selection:11.0V (min) at 1 mA sourcing: 0.75V (max) at 5 mA sinking
Pulse width:Selectable 10 ms ± 10% or 100 ms ± 10% in service menu
Output impedance:50Ω nominal
Current limit:15 mA nominal, both sourcing and sinking
Marker in
Input threshold:V
Input hysteresis650 mV typical
=±2.5V (min), VIL =±1.5V (max)
IH
Maximum input voltage:±30 V (with respect to ground on pin 2)
Input impedance:10kΩ (min) for -25V <V
Pulse width:1.0 ms (min), V
≥ 2.5V
IN
<25V
IN
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 13
Page 32
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
Environmental Specifications
Power Requirements:90–132 VAC 50/60 Hz
190–264 VAC 50/60 Hz
Power consumption:90–132 VAC 2A
CISPR Publication 11Class B Radiated, Class B Conducted
ClassificationsThe Eagle 3000 Patient Monitor is classified, according to IEC 601-1.
Type of protection against
electrical shock:
Degree of protection against
electrical shock:
Degree of protection against
harmful ingress of water:
Degree of safety of application
in the presence of a flammable
anesthetic mixture with air or
with oxygen or nitrous oxide:
Method(s) of sterilization or
disinfection recommended by the
manufacturer:
Class 1 Equipment
ECG, Respiration, and Invasive Blood Pressure are type CF
equipment. Non-Invasive blood pressure, SpO2, and CO2 are type BF
equipment.
Ordinary Equipment (enclosed equipment without protection against
ingress of water)
Equipment not suitable for use in the presence of a flammable
anesthetic mixture with air or with oxygen or nitrous oxide.
Not applicable
Mode of operation:
2 - 14415397-003PAGE REV C
Continuous operation
EAGLE 3000 PATIENT MONITOR
Page 33
PREPARATION FOR USE
EQUIPMENT OVERVIEW
POWERREQUIREMENTS
EQUIPMENTGROUND
REQUIREMENTS
At least one grounded duplex wall receptacle should be provided
for each monitor. The wall receptacle should be hospital grade
and installed in a suitable junction box. Power should be
provided by a power line dedicated solely to equipment requiring
emergency power.
WARNING
Loss of power to the monitor results in the loss of all
monitoring functions.
The ground pin of the wall receptacles and all exposed metal
parts (beds, radiators, water pipes, etc.) in the patient area
should be connected together and tied to the nearest
equipotential ground point through a bonded grounding system,
or with a 10-AWG stranded copper grounding cable. This
equipotential ground point should be as close to earth ground as
possible. Use only three-prong, polarized, hospital-grade wall
receptacles to accept the three-wire, polarized plug on the power
cord of the monitor.
If a bonded grounding unit is not available, interconnect the
ground pins of all wall receptacles in the patient and monitor
areas with 10-AWG (or larger) stranded copper cables. This
copper cable must connect to the central grounding point. Do
not jumper from ground pin to ground pin, then to the central
grounding point. The ground cabling must not carry current,
such as a grounded neutral, since the current flow will produce
differences in potential along the ground. These potential
differences are the main source for shock hazards to the users
and patients.
Do not rely on conduit as a ground conductor. Plastic (PVC)
pipes or fittings used as conduit break up the ground path,
which can present potential shock hazards. The electrical
ground system must be connected to actual earth ground. If this
is not possible, then a good reference ground such as a metal
cold water pipe or an electrically conductive building component
should be used. It is more important that all grounded objects in
the patient area are at the same potential than at true earth
potential.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 15
Page 34
PREPARATION FOR USE (CONT)
EQUIPMENT OVERVIEW
MONITORVENTILATION
REQUIREMENTS
The monitor is capable of producing as much as 170 BTu per
hour of heat load. This is equivalent to approximately 50 watts
of energy.
WARNING
Failure to properly ventilate the monitor may cause
equipment failure or improper monitoring conditions which may endanger the patient being monitored.
CAUTION
Do not locate the monitor in an enclosed area that
may restrict the heat dissipated by it. Any restriction
in air flow causes a rise in internal temperature
which may result in equipment failure.
CAUTION
The monitor must be located no closer than 4 inches
(10 cm) from any partition or wall. The monitor
should be approximately 12 inches (30 cm) from any
overhead partition or the ceiling.
MOUNTINGRECOMMENDATIONS
SOFTWARESETUP
Tram Critical Care Monitoring System Reference Guide:
•pn. 403799-010
•Manufacturer recommended methods of mounting the
monitor to various locations.
Section 6: Configuration
•Information regarding connection of the monitor to
peripherals
EAGLE 3000 PATIENT MONITOR
2 - 16415397-003PAGE REV C
Page 35
ORDERING INFORMATION
y
EQUIPMENT OVERVIEW
PARTNUMBERSAND
DESCRIPTIONS
EGL3107=A
How the order numbers are
broken down
Below is a breakdown of the product part number used for
ordering the monitor.
0: No options
1: Defib Sync/Analog Out
2: End-tidal CO2 (Defib Sync/Analog Out included/required)
3: STAR writer (Defib Sync/Analog Out included/required)
4: End-tidal CO2 / STAR writer (Defib Sync/Analog Out included/required)
5: Dual invasive BP
6: Dual invasive BP / Defib Sync/Analog Out
7: Dual invasive BP / End-tidal CO2 (Defib Sync/Analog Out included/required)
8: Dual invasive BP / STAR writer (Defib Sync/Analog Out included/required)
9: Dual invasive BP / End-tidal CO2 / STAR writer (Defib Sync/Analog Out included/required)
Eagle 3000 Patient Monitor product line
All models include: ECG, Non-invasive BP, pulse oximetr
and dual temp
Using the sample product order number provided (EGL3107=A),
the following information regarding the monitor configuration
can be determined:
•The “1” in the numeric portion of the order number (3107)
indicates that the monitor includes respiration along with
all of the standard vital sign monitoring functions included
with all monitors: ECG, non-invasive blood pressure, pulse
oximetry and dual temperature.
•The “0” in the numeric portion of the order number (3107)
indicates the monitor is configured with Level 7015
operating software enabled. Level 7015 operating software
includes only basic monitoring functions. Level 7020
operating software includes lethal ECG arrhythmia
detection (V-Fib and V-Tach) along with basic monitor
operating functions.
•The “7” in the numeric portion of the order number (3107)
indicates the monitor also is configured with the following
optional vital sign monitoring functions:
»two invasive blood pressure ports, and
»end-tidal CO2 monitoring which requires the use of an
option interface board. The option interface board
also provides defibrillator synchronization and analog
output jacks.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 17
Page 36
THEORY OF OPERATION
OVERALLMONITORBLOCK
DIAGRAM
EQUIPMENT OVERVIEW
AC Power & Switch
Debug
ECG/Resp
SpO2
Temp 1/Temp 2
BP1
BP2
NBP
EtCO2
J1
J13
W15
Acquisition
J16
J17
J18
J19
Subsystem
Power
Supply
Board
Data
System
(DAS)
Board
NBP
EtCO2
Board
J3J2
J8J8
Processor
J5
J11
Main
Board
J3
J4
J7
J9
J10
J12
J25
J6J6
EL Display
Speaker
5 Hard Keys
Ethernet
RS-422
Debug
Trim Knob
Interface
Option
Board
J21
J22
J23
J14
Defib Sync/Analog Output
Remote Alarm
STAR
Recorder
GENERALMONITORBLOCK
THEORY
The theory of operation for the monitor, as covered in this part of
the section, is intended to provide an overall block level overview
of the monitor for service technicians. A general understanding
of the theory of operation is required to effectively install,
maintain or repair the monitor.
Detailed circuit theory
More detailed theory of operation can be obtained by attending
manufacturer formal technical training classes. Regularly
scheduled technical training classes are held throughout the
year at the manufacturer training facility located in Jupiter,
Florida. If warranted, technical training classes may be
scheduled at customer sites or other locations in the field as
well.
EAGLE 3000 PATIENT MONITOR
2 - 18415397-003PAGE REV C
Page 37
THEORY OF OPERATION (CONT)
EQUIPMENT OVERVIEW
About the power supply board
About the DAS board
The power supply is an off-line forward converter topology with a
two input range voltage rectifier/doubler scheme. The converter
uses current mode control for best overall performance and fault
tolerance. A two-transistor power switch approach was used to
provide maximum ruggedness against input voltage transients
and low conducted/radiated EMI.
Two individual output voltages are provided with complete
overload/fault protection. The supply has six unique functional
sections:
1.Input voltage rectifier/doubler section
2.Power forward converter/magnetics section
3.PWM controller section
4.Fault management section
5.+12V overvoltage protection section
6.Post regulator section
The data acquisition system (DAS) board, located in the monitor,
is responsible for the acquisition of all vital-sign patient data.
Analog sensor/electrode input signals are amplified and
conditioned by hybrid assemblies, then converted to digital data.
The digital patient data is transferred across an isolation barrier
via high-speed opto-couplers to the processor pcb for analysis
and display.
The DAS consists of an isolated and non-isolated section which
are separated by a barrier that is capable of withstanding up to
6000 Vdc with respect to earth ground. Isolation is
accomplished by using a coupled inductor power supply and
opto-isolation for signals crossing the barrier.
About the main processor board
About the interface option board
About the EtCO2 option board
The processor pcb provides signal processing, system control,
user interface, and communications functions for the monitor.
It receives and processes digitized patient data from the isolated
DAS board, text and waveform information for the display,
interfaces with the operator via the front panel switches and
Trim Knob, and communicates with other products on the
network using a built-in Ethernet interface. Additional
capabilities include an asynchronous communications port for
devices like a DDW or remote control.
The development of the interface option board allows the cost of
the main processing board to be reduced by incurring the cost of
additional circuitry and connectors not required in many
monitoring applications. This board provides the electrical
hardware required to interface and control optional monitoring
features. The options available include analog output,
defibrillator synchronization, remote nurse alarm control, an
integrated thermal recorder and an end-tidal CO2 subsystem.
The EtCO2 subsystem connects electrically and mechanically to
the interface option board. An asynchronous communications
port is used to communicate to the main processor board.
High potential tests ................................................... 3-28
AC hi-pot tests .......................................................... 3-30
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 1
Page 40
MAINTENANCE SCHEDULE
MAINTENANCE
MANUFACTURER
RECOMMENDATION
To make sure the monitor remains in proper operational and
functional order, a good maintenance schedule must be adhered
to. The manufacturer's recommendations in this regard are as
follows:
•Inspection: Operators should perform this prior to
admitting each patient to the monitor. Service personnel
should perform this prior to servicing the monitor.
•General Cleaning: Operators should perform this prior to
admitting each patient to the monitor. Service personnel
should perform this after servicing the monitor.
•Checkout Procedures: These should be performed by
qualified service personnel upon receipt of the equipment,
every 12 months thereafter, and each time the monitor is
serviced.
•Leakage Current Tests: These should be performed by
qualified service personnel upon receipt of the equipment,
every 12 months thereafter, and each time the monitor is
serviced.
•Hi-Pot Tests: High-potential tests should be performed by
qualified service personnel whenever any component of the
isolated data acquisition system (DAS) is removed, repaired
or replaced in the monitor.
MANUFACTURERRESPONSIBILITY
NOTE
The Hi-Pot Tests provide a means of checking the
patient isolation circuitry such that a patient receiving defibrillation, while attached or admitted to the
monitor, will receive the full energy of each shock and
that the monitor will not absorb the energy, when
delivered.
•Non-invasive Blood Pressure (NBP) Calibration: NBP
calibration should be performed by qualified service
personnel upon receipt of the equipment and once each
year, thereafter. Refer to Section 5: Calibration, for this
information.
Failure on the part of all responsible individuals, hospitals or
institutions, employing the use of this monitor, to implement the
recommended maintenance schedule may cause equipment
failure and potential operator and patient health hazards. The
manufacturer does not in any manner, unless an Equipment
Maintenance Agreement exists, assume the responsibility for
performing the recommended maintenance schedule. The sole
responsibility rests with all individuals, hospitals, or institutions
utilizing the monitor.
EAGLE 3000 PATIENT MONITOR
3 - 2415397-003PAGE REV C
Page 41
VISUAL INSPECTION
MAINTENANCE
INSPECTINGTHEMONITOR
The monitor should be carefully inspected prior to each patient
being admitted to the monitoring system. Follow these
guidelines when inspecting the equipment:
•Carefully inspect the monitor for obvious physical damage
to the outer case, display screen and controls. Do not use
the monitor if physical damage is determined. Refer
damaged equipment to qualified service personnel for repair
before using it again on a patient.
•Inspect all external connectors, front and rear, for degraded
pins, prongs and connector housings. Refer damaged
equipment to qualified service personnel for repair before
using it again on a patient.
•Inspect all cable insulation, cable strain-reliefs and cable
connectors for damage, cracks or degradation. Refer
damaged equipment to qualified service personnel for repair
before using it again on a patient.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 3
Page 42
CLEANING T HE MONITOR
MAINTENANCE
CLEANINGTHEDISPLAY
CLEANINGTHEEXTERNAL
SURFACES
To clean the display on the monitor, use a soft, clean, lint-free
cloth dampened with a glass cleaner similar to Windex, or a 1:1
mixture of isopropyl alcohol and water.
WARNING
Do not spray glass cleaner or general cleaning solutions directly onto the display. Do not use hospital
disinfectants, like Cidex, on the display.
Clean the external surfaces of the monitor before each time a
patient is admitted to the system. The exterior surfaces may be
cleaned with a lint-free cloth dampened with one of these
approved solutions:
•ammonia (diluted),
•Cidex,
•mild soap (dissolved), or
•sodium hypochlorite bleach (diluted).
MANUFACTURER
RECOMMENDATION
The manufacturer recommends the following guidelines to avoid
damaging the monitor:
•Dilute all cleaning solutions according to respective
manufacturer recommendations.
•Use a clean, dry, lint-free cloth to wipe off excess cleaning
solution after each application.
•Do not pour water or cleaning solutions directly onto the
monitor. Do not allow fluids to run into crevices,
connectors or cooling vents on the monitor.
•Never use these cleaning agents:
»abrasive cleaners or solvents of any kind,
»alcohol-based cleaning agents,
»wax containing a cleaning substance,
»acetone, or
»betadine.
CAUTION
Follow these cleaning instructions exactly. Failure
to follow the instructions may melt, distort, or dull
the finish of the case, blur lettering on the labels, or
cause equipment failures.
EAGLE 3000 PATIENT MONITOR
3 - 4415397-003PAGE REV C
Page 43
CHECKOUT PROCEDURES
MAINTENANCE
ABOUTTHECHECKOUT
PROCEDURES
MANUFACTURERRECOMMENDED
TEST
EQUIPMENT
The following pages contain the checkout procedures for the
monitor. The purpose of the checkout procedures is to provide
service personnel with a method which can be used to verify
operational and functional performance of the monitor. Failure
to attain any of the listed results indicates a potential
malfunction of the monitor.
Perform the checkout procedures upon receipt of the monitor,
every 12 months thereafter, and each time a circuit board is
removed or replaced.
The checkout procedures are based on the assumption that the
monitor being tested is used with known good cables and test
equipment. It also requires that the user be somewhat familiar
with the operation of all test equipment required for the checkout
procedures. For more information concerning the operation of
these components, refer to the respective operator manual.
The following table lists the manufacturer's recommended test
equipment, adaptors, and cables necessary to successfully
complete the checkout procedures. The checkout procedures
were written for the test equipment in the following table. If test
equipment other than the manufacturer's recommendation is
used, it may be necessary to slightly modify some test steps.
Individually Replaceable, 5-Leadwire, AHA411200-0011
BP Adapter700095-0012
Temperature Adaptor402015-0041
TEMP-to-Simulator Cable67700311
Digital manometerSensym PDM200M1
NBP Cuff9461-3011
NBP Tubing414873-0011
Manometer Tubing401582-0012 ft
Coupling46100-0021
Coupling400787-0011
3-Way Tee4745-1011
SpO2 Simulator408610-0011
SpO2 Simulator Cable, Nellcor700232-0041
CO2 SimulatorNovametrix TB12651
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 5
Page 44
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
ECG TESTS
1.Set up the patient simulator as follows:
•Heart rate - 80 bpm,
•Heart rate amplitude - 1.0 mV,
•5-lead ECG patient cable properly attached.
2.Attach the ECG patient cable and ECG leadwire set to the
ECG/RESP connector on the monitor and the leadwire
connectors on the top of the patient simulator.
3.Admit the patient simulator to the monitor.
4.Observe the following:
•ECG lead II is displayed and is noise-free,
•Heart rate of 80 ±1 bpm is displayed,
•With QRS tones enabled, an audible tone sounds with
each R-Wave (QRS complex).
5.Verify all seven ECG leads are available for viewing and are
noise-free.
6.Select DETECT PACE and set to NORMAL.
7.Select the VP2 pacemaker pulse on the simulator.
8.Observe the following while viewing ECG leads II, III, aVR,
aVF, and V:
•a P appears above the PVC count indicating
pacemaker pulse detection is enabled, and
•the heart rate still reads 80 ±1 bpm.
9.Disable pacemaker pulse detection on the monitor and
return the simulator to these conditions:
•Heart rate - 80 bpm,
•Heart rate amplitude - 1.0 mV,
•5-lead ECG patient cable properly attached.
10.Select ECG lead II for viewing in the top trace position on
the monitor display.
11.Disconnect the RA leadwire from the patient simulator.
12.Observe the following:
•a RA FAIL message appears on the display, and
•lead III automatically displays in place of lead II in the
top trace position.
13.Reconnect the RA leadwire to the patient simulator.
14.Inject a 1-millivolt calibration signal using the patient
simulator and start a manual graph.
15.Observe that the calibration pulse is properly displayed and
graphed.
16.This completes the ECG tests. Continue to the next steps
of these checkout procedures.
EAGLE 3000 PATIENT MONITOR
3 - 6415397-003PAGE REV C
Page 45
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
RESPIRATIONTESTS (OPTIONAL)
Respiration tests completion
1.With the ECG patient cable still connected to the ECG/
RESP connector of the monitor, set up the patient simulator
as follows:
•Respiration (RESP) baseline impedance - 750Ω,
•RESP R - 0.5Ω,
•RESP lead select - I & II,
•RESP rate (respirations per minute) - 30.
2.Set up the monitor as follows:
•RESP waveform - on,
•RESP waveform lead select - lead II (RESP waveform
derived from ECG lead II).
3.Observe the following:
•RESP parameter window appears on the monitor with
a reading of 30 ±2 (respirations per minute),
•RESP waveform appears distortion-free on the
monitor.
4.Change the RESP waveform lead select of the monitor to
lead I (RESP waveform derived from ECG lead I).
5.Observe the following:
•RESP parameter window appears on the monitor with
a reading of 30 ±2 (respirations per minute),
•RESP waveform appears distortion-free on the
monitor.
6.Disconnect the ECG patient cable from the ECG/RESP
connector of the monitor. Proceed to the next steps in
these checkout procedures.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 7
Page 46
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
TEMPERATURETESTS
Temperature tests completion
1.Set up the patient simulator for a temperature output of
37°C.
2.Attach the temperature adaptor cable to the TEMP
connector of the monitor.
3.Set the switch on the temperature adaptor to the 400
position.
4.Attach the temperature simulator cable from the SERIES
400 TEMPERATURE OUTPUT connector of the patient
simulator to the T1 connector of the temperature adaptor.
5.Verify a TEMP parameter window appears on the monitor
display with a T1 reading of 37.0°±0.4° C.
6.Move the temperature simulator cable from the T1
connector of the temperature adaptor to the T2 connector of
the temperature adaptor.
7.Verify a T2 reading of 37.0°±0.4° C in the TEMP parameter
window on the monitor display.
8.Remove the temperature adaptor and temperature
simulator cable from the monitor and patient simulator.
EAGLE 3000 PATIENT MONITOR
3 - 8415397-003PAGE REV C
Page 47
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
INVASIVEBLOODPRESSURE
(OPTIONAL) TESTS
BP1 connector (AR1) tests
BP1 test completion
BP2 connector (PA2) tests
Invasive blood pressure tests
completion
The invasive blood pressure (BP) tests provide a method of
verification for both BP connectors (BP1 and BP2) of a monitor
equipped with this optional function. Follow these steps:
1.Set up the patient simulator as follows:
•Blood pressure (BP) polarity - POS,
•BP output - 0 mmHg.
2.Connect the BP simulator cable from the BLOOD
PRESSURE 1 - 120/80 connector of the patient simulator
to the BP1 (left-most BP) connector of the monitor.
3.Verify the AR1 parameter window, waveform label,
corresponding graticules, and waveform appear on the
monitor display, along with a BP waveform requiring zero
reference.
4.Press the ZERO ALL push-button on the front panel of the
monitor to zero-reference the AR1 BP waveform.
5.Change the patient simulator BP output to 200 mmHg.
6.Observe a reading of 200/200 (200) ± 4 mmHg in the AR1
parameter window on the monitor display.
7.Change the patient simulator BP output to WAVE
(simulated BP waveform).
8.Set the AR1 BP waveform gain on the monitor to auto.
9.Observe a distortion-free AR1 BP waveform and a reading of
approximately 120/80 (93) in the AR1 parameter window
on the monitor display.
10.Disconnect the BP simulator cable from the BP1 connector
of the monitor. Continue to the next step for the BP2 test.
11.Again, set up the patient simulator as follows:
•BP polarity - POS,
•BP output - 0 mmHg.
12.Connect the BP simulator cable to the BP2 (right-most BP)
connector of the monitor.
13.Verify a PA2 parameter window, waveform label and
corresponding graticules appear on the monitor display,
along with a PA2 BP waveform requiring zero reference.
14.Press the ZERO ALL push-button on the front panel of the
monitor to zero reference the PA2 BP waveform.
15.Change the patient simulator BP output to 200 mmHg.
16.Observe a reading of 200/200 (200) ± 4 mmHg in the PA2
parameter window on the monitor display.
17.Change the patient simulator BP output to WAVE
(simulated BP waveform).
18.Set the PA2 BP waveform gain on the monitor to auto.
19.Observe a distortion-free PA2 BP waveform and a reading of
approximately 120/80 (93) in the PA2 parameter window on
the monitor display.
20.Remove the BP simulator cable from the BP2 connector of
the monitor. This completes the BP tests.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 9
Page 48
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
PULSEOXIMETRYTESTS
1.Set the pulse oximetry (SpO2) simulator power switch to the
off position.
2.Connect the Nellcor-style SpO2 simulator cable between the
SpO2 connector of the monitor and the SpO2 simulator.
3.Set up the SpO2 simulator as follows:
•SPO2 - 99% (using the white NELLCOR values),
•PULSE RATE - 100 B/M (beats per minute),
•MODE - NELLCOR,
•Power switch - on.
4.Verify a SPO2 parameter window, waveform label and
corresponding graticules appear on the monitor display.
5.Verify the following appear on the monitor display:
•Sinusoidal SpO2 waveform,
•SPO2% parameter reading of 97-102 (%),
•PPR parameter reading of 97-103 (beats per minute).
6.Verify accuracy of the SPO2% values (these are the white
NELLCOR values shown on the SpO2 simulator) on the
monitor display using the SpO2 simulator settings from the
following table:
Pulse oximetry tests completion
SpO2 Simulator SettingDisplayed SPO2% Value
99%97 – 102
85.5%83 – 88
68.4%66 – 71
7.Verify accuracy of the PPR values on the monitor display
using the SpO2 simulator pulse rates from the following
table:
Simulator PULSE RATEDisplayed PPR Value
70 B/M68 – 72
100 B/M97 – 103
160 B/M156 – 164
8.Press the INTERFERENCE TEST button on the SpO
simulator for 30 seconds.
9.Verify the displayed SPO2% value remains 97-102%, or an
interference detection message is displayed and XX is
displayed in the SpO2 parameter window in place of an
SPO2% value.
10.Set the SpO2 simulator power switch to the off position.
11.Disconnect the Nellcor-style SpO2 simulator cable from the
monitor SpO2 connector. This completes the SpO2 tests.
2
EAGLE 3000 PATIENT MONITOR
3 - 10415397-003PAGE REV C
Page 49
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
NONINVASIVEBLOODPRESSURE
TESTS
: PRE-TESTSETUP
1.Attach the digital manometer, noninvasive blood pressure
(NBP) cuff, tees and tubing, as shown below, to the NBP
connector of the monitor.
NBP tubing,
pn 414873-001
6-inch diameter PVC
pipe (or 1-pound
coffee can):
Wrap the NBP cuff
around this for tests.
NBP cuff,
pn 9461-301:
Any size NBP cuff
will be sufficient.
Coupling,
pn 46100-002:
Connects manometer
tubing to NBP tubing.
3-way tee,
pn 4745-101:
Connects manometer
tubing, NBP tubing
and NBP cuff tubing.
Digital manometer,
Sensym PDM200M:
Note: A calibrated
mercury manometer
may be substituted.
Coupling,
Manometer tubing,
pn 401582-001:
Approximately two
feet in length.
2.Set the digital manometer power switch to the on position.
3.Set the digital manometer range switch to 1000␣ mmHg.
pn 400787-001:
Connects manometer
tubing to NBP cuff
tubing.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 11
Page 50
0106
SERVICE
MODE
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
NONINVASIVEBLOODPRESSURE
TESTS
Enter the service menus of the
monitor
To perform the noninvasive blood pressure (NBP) tests, current
software is assumed to be installed in the monitor.
1.From the main menu of the monitor, rotate the Trim Knob
control to highlight MONITOR SETUP and press the Trim
Knob control to select it.
ALARM
CONTROL
PATIENT
DATA
MONITOR
SETUP
PATIENT:
DISCHARGED
2.Rotate the Trim Knob control to highlight SERVICE MODE,
and press the Trim Knob control to select it.
MAIN
MENU
WAVEFORMS
ON/OFF
UNIT ALARMS:
OFF
DISPLAY:
INDIVIDUAL
BRIGHTNESS:
100%
LEARN THE
MONITOR
PARAMETERS
ON/OFF
SOFTWARE
REVISION
GRAPH
SETUP
SOFTWARE
COMPATIBILITY
MONITOR
DEFAULTS
SERVICE
MODE
3.A service menu password window will appear on the
monitor display. A password is required to prevent nonservice personnel from accessing the service menus. The
password is four numbers that represent the date that
currently resides in a memory circuit within the monitor
(please note that this may or may not be the correct date).
In the password, the first two numbers, starting from the
left, represent the day and the second two numbers
represent the month of whatever date that currently resides
in the memory circuits of the monitor. For example, the
seventh day of the third month (June 1st) would be
represented in the password as 0106 (ddmm). Note the
date that is currently on the monitor display and follow
these steps to enter the password;
•Rotate the Trim Knob control to highlight the
password number that you would like to change.
•To change the highlighted number. press the Trim
Knob control.
•Rotate the Trim Knob control until the correct number
is displayed in the selected field.
•To enter the number, press the Trim Knob control.
•Repeat these steps until all password numbers are
correctly displayed.
•Once you have entered the correct password numbers,
rotate the Trim Knob control to highlight SERVICE
MODE in the enter password window.
•Press the Trim Knob control one more time to enter
the password and access the service menus of the
monitor.
3 - 12415397-003PAGE REV C
EAGLE 3000 PATIENT MONITOR
Page 51
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
NBP calibration menu of the
monitor
The service menus should appear on the monitor display. These
next steps guide you through the service menus associated with
checking NBP calibration. If desired test results are not
obtained, NBP calibration will be necessary.
4.Rotate the Trim Knob control to highlight CALIBRATE and
press the Trim Knob control to select it. Next, rotate the
Trim Knob control to highlight CALIBRATE NBP and press
the Trim Knob control to select it.
MAIN
MENU
SOFTWARE
LEVEL
MAIN
MENU
PREVIOUS
MENU
REVIEW
ERRORS
PATIENT-MONITOR TYPE:
ADULT-ICU
CALIBRATE
NBP
CO2
SERVICE
ADMIT MENU:
STANDARD
SET UNIT
NAME
CALIBRATE
SET BED
NUMBER
SET INTERNET
ADDRESS
STAR TEST
PATTERN
TIME AND
DATE
Start the NBP calibration test
5.Rotate the Trim Knob control to highlight CHECK CAL OFF
and press the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
OFF
6.Rotate the Trim Knob control to highlight START and press
the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
OFF
> START
STOP
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 13
Page 52
CHECKOUT PROCEDURES (CONT)
Verify NBP calibration7.The text on the menu item will change from CHECK CAL
MAINTENANCE
OFF to CHECK CAL IN PROGRESS. Verify the readings in
the NBP parameter window on the monitor dispaly and
readings on the digital manometer are equal (± 1 mmHg) for
at least one full minute. If the readings are not equal for at
least one full minute, the NBP circuit requires calibration.
24-JAN-1996 09:27
II
V
OPENS POPUP TO START/STOP A CALIBRATION CHECK
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
IN PROGRESS
DISCHARGED
LEADS FAIL
ICU-BED5
X / X
CUFF 250
X
150
50
140
40
mmHg
ADT
8.Rotate the Trim Knob control to highlight CHECK CAL IN
PROGRESS and press the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
IN PROGRESS
E
C
G
N
B
P
EAGLE 3000 PATIENT MONITOR
3 - 14415397-003PAGE REV C
Page 53
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
Stop the NBP calibration test
9.Rotate the Trim Knob control to highlight STOP and press
the Trim Knob control to select it. The pneumatic control
circuit of the monitor will vent air pressure in the
pneumatic circuit of the monitor to atmosphere, causing
the NBP cuff to deflate.
24-JAN-1996 09:29
II
V
OPENS POPUP TO START/STOP A CALIBRATION CHECK
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
IN PROGRESS
>
START
STOP
DISCHARGED
LEADS FAIL
ICU-BED5
X / X
CUFF 236
X
150
140
mmHg
E
50
C
G
N
40
B
P
ADT
Noninvasive blood pressure tests
completion
10.Remove the NBP test setup apparatus from the monitor.
The NBP tests are complete.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 15
Page 54
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
END-TIDAL CO2 TESTS
1. Return to the main menu of the monitor.
2. With the Capnostat sensor attached to the front panel
connector of the monitor, put the sensor on the zero
reference (→ 0 ←) cell.
3. Use the Trim Knob control to select the CO2 parameter
menu. Rotate the Trim Knob control to highlight
CAL SENSOR TO ZERO CELL, and press the Trim Knob to
select it. Select READY and press the Trim Knob.
A CALIBRATING message will appear in the CO
parameter box.
4. After zero calibration is complete, put the sensor on the
reference (REF) cell.
5. Verify the reading in the CO
mmHg.
parameter box displays 38 ±2
2
2
EAGLE 3000 PATIENT MONITOR
3 - 16415397-003PAGE REV C
Page 55
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
DEFIBRILLATOR
SYNCHRONIZATION
DEFIB. SYNC.
ECG - 1
Ground - 2
Marker Out - 3
DEFIB SYNC connector:
DEFIB SYNC connector:
TESTS
6 - Arterial BP
5 - Analog Ground
4 - Marker In
Arterial BP
ECG
1.Use the figure at the left as a reference for connecting the
oscilloscope to the DEFIB SYNC connector, located on the
front panel of the monitor, for performing these tests.
2.Test the ECG, Arterial BP and Marker Out signals from the
DEFIB SYNC connector. They should closely resemble the
waveforms in the figures below. Note that there are two
Marker Out traces shown below. The upper Marker Out
figure references the frequency aspects of the signal. The
lower Marker Out figure references the pulse width aspects
of the signal.
Signal Pin:1
Ground Pin:5
Probe Type:x10
Time/Division: 0.2S
Volts/Division: 0.5V
Signal Pin:6
Ground Pin:5
Probe Type:x10
Time/Division: 0.2S
Volts/Division: 0.2V
DEFIB SYNC connector:
Marker Out (frequency)
Signal Pin:3
Ground Pin:2
Probe Type:x10
Time/Division: 0.2S
Volts/Division: 1V
DEFIB SYNC connector:
Marker Out (pulse width)
Signal Pin:3
Ground Pin:2
Probe Type:x10
Time/Division: 5mS
Volts/Division: 1V
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 17
Page 56
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
Verify defib sync markers
Defibrillator synchronization tests
completion
3.Attach a jumper wire between pin-3 (Marker Out) and pin-4
(Marker In) of the DEFIB SYNC connector located on the
front of the monitor. Verify negative spikes in each of the
QRS Complex (ECG waveform) R-Waves on the monitor
display, similar to those shown in the illustration below.
Observe normal
R-waves before
the jumper is
installed.
Observe the negative spikes in the R-waves
while the jumper is installed. The spikes are
small, and they can be difficult to see at times.
An "X" is also placed on these ECG waveforms.
XXXX
4.Remove the jumper wire installed in the previous step from
the DEFIB SYNC connector. This completes the
defibrillator synchronization tests.
EAGLE 3000 PATIENT MONITOR
3 - 18415397-003PAGE REV C
Page 57
CHECKOUT PROCEDURES (CONT)
MAINTENANCE
SPEAKERTESTS
Checkout procedure tests
completion
1.Change the alarm volume of the monitor to 100%.
2.Verify the speaker volume of the monitor changes
accordingly.
3.Return the volume of the monitor to the level it was
previously set to, before you changed it for this test.
This completes all tests associated with the checkout procedures.
Disconnect the monitor from all test equipment in the following
manner:
1.Set all test equipment power switches to the off position.
2.Set the monitor rear panel power switch to the off (0)
position.
3.Remove all test equipment from the monitor.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 19
Page 58
ELECTRICAL SAFETY T ESTS
MAINTENANCE
CURRENTLEAKAGETESTS
Manufacturer recommendation
Leakage current tests provide a method of determining if
potential electrical health hazards to the patient exist in the
monitor. These tests generally are required by the National Fire
Protection Agency (NFPA) as a part of National Electrical Code
(NEC) guidelines for medical device electrical safety.
It is recommended that these tests be performed upon receipt of
the equipment, once per year thereafter, and each time the main
enclosure is disassembled or a circuit board is removed, tested,
repaired, or replaced.
WARNING
Failure to perform leakage current tests may cause
undue equipment failure and potential health hazards to patients connected to the monitor. The
manufacturer does not in any manner, unless an
Equipment Maintenance Agreement exists, assume
the responsibility for performing the leakage current
tests. The sole responsibility rests with the individual or institution using the equipment. Manufacturer service representatives may, at their discretion, use this procedure as a helpful guide during
visits to the equipment site.
Test conditions
Test equipment
Leakage current tests may be performed under normal ambient
conditions of temperature, humidity, and pressure.
The Manufacturer recommended test equipment required to
perform leakage current tests is listed below. Equivalent
equipment may be substituted as necessary.
NameManufacturerPart Number
Digital MultimeterFluke8060A
Leakage Tester - 115V/60HzMEIMT-1216-01
ECG test bodyMEIMT -3387
EAGLE 3000 PATIENT MONITOR
3 - 20415397-003PAGE REV C
Page 59
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
WALLRECEPTACLETESTS
If other than normal polarity and
ground is indicated
Before starting the tests, the wall receptacle from which the
monitor will get electrical power must be checked. This test
checks the condition of the wall receptacle to ensure correct
results from leakage tests.
Connect the leakage tester to the wall receptacle. Observe the 0,
K, and R lamps with the GND switch in the down position. For
safe conditions, the lamps should reflect normal polarity and
ground as shown below.
0KRCondition
OnOnOffNormal polarity and ground
OffOnOnReverse polarity
OffOnOffNo ground
OnOffOffNo neutral
OffOffOnNo neutral/reverse polarity
OffOffOffNo power
If other than normal polarity and ground is indicated, corrective
action must be taken before proceeding to the following steps.
The results of the following steps will be meaningless unless a
properly wired wall receptacle is used.
SURFACECONTINUITYTESTS
Ground
Pin
Power cord plug (120 Vac)
The surface continuity test provides a method of checking the
integrity of the monitor relative to proper internal and external
electrical ground. This test determines whether the monitor has
a power ground fault.
1.Disconnect the monitor (unit under test) from any wall
receptacle.
2.Connect the negative lead of a digital multimeter (DMM) to
the ground pin of the unit under test’s power cord plug.
The figure to the left shows the location of the ground pin
on a 120 Vac power cord plug used in the United States. If
your monitor uses a different voltage, or you live in a
different country, your outlet will look different.
3.Set the DMM to the milliohms (mΩ) range.
4.Connect the positive lead of the DMM to any exposed metal
surface on the unit under test.
5.Read the resistance displayed on the DMM. If the
resistance is higher than 100 mΩ, the unit under test fails
this test and should be repaired and tested again.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 21
Page 60
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
GROUNDWIRETOGROUNDTESTS
Electrical diagram:
ground wire to ground tests
Perform this test to measure leakage current through the ground
wire of the monitor during normal operation.
1.Set the leakage tester switches as follows:
•Selector knob - 1,
•GND switch - OPEN,
•Polarity switch - NORM,
•Power switch - OFF.
2.Connect the DMM to the METER jacks on the leakage
tester. Set the DMM to measure AC millivolts.
3.Connect the power cord of the monitor to the power
receptacle on the rear of the leakage tester.
4.Set the leakage tester power switch to ON.
5.Set the rear panel power switch of the monitor to ON.
6.Read leakage current indicated on DMM. If the reading is
greater than:
•300 microamperes (µA, read as 0.3 volts on the
DMM), and the monitor is operating at 120 V/60 Hz
(U.S.); or
•500 µA (0.5 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
7.Set the polarity switch on the leakage tester to RVS
(reverse).
8.Read the leakage current indicated on the DMM. If the
reading is greater than:
•300 µA, (0.3 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•500 µA (0.5 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
9.Set the leakage tester power switch to OFF.
Leakage Tester
HIGH
LOW
GND
POWER CORD
V*
1K
0.15µF
10
NORM
RVS
GND
POWER CORD
UNIT
UNDER
TEST
*Meter reading: 1 mV = 1 µA (leakage current)
EAGLE 3000 PATIENT MONITOR
3 - 22415397-003PAGE REV C
Page 61
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
CHASSISTO GROUNDTESTS
Perform this test to measure leakage current through exposed
conductive surfaces on the monitor during normal operation.
1.Set the leakage tester switches as follows:
•Selector knob - 2,
•GND switch - OPEN,
•Polarity switch - NORM.
2.Connect a meter lead between the CHAS connector on the
rear of the leakage tester and an unpainted, non-anodized
chassis ground on the unit under test.
3.Set the leakage tester power switch to ON.
4.Read the leakage current indicated on the DMM. If the
reading is greater than:
•300 µA, (0.3 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•500 µA (0.5 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
5.Set the polarity switch to RVS and observe the same meter
readings as in the previous step.
6.Set the GND switch on the leakage tester to CLOSED.
7.Read the leakage current indicated on the DMM. If the
reading is greater than:
•300 µA, (0.3 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•500 µA (0.5 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
8.Set the polarity switch to RVS and observe the same meter
readings as in the previous step.
9.Set the leakage tester power switch to OFF and remove the
meter lead connected in step 2.
Electrical diagram:
Leakage Tester
chassis to ground tests
HIGH
LOW
GND
POWER CORD
1K
V*
*Meter reading: 1 mV = 1 µA (leakage current)
0.15µF
10
NORM
RVS
GND
POWER CORD
Probe to exposed chassis
UNIT
UNDER
TEST
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 23
Page 62
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
PATIENTSOURCETESTS
This test checks leakage current from the ECG/RESP connector
of the monitor relative to ground.
1.Set leakage tester switches as follows:
•Selector knob - 3,
•GND switch - GND OPEN,
•Polarity switch - NORM,
•Power switch - OFF.
2.Connect an ECG test body to the ECG/RESP connector of
the monitor.
3.Connect a short length of cable between the ECG test body
installed in the last step and the jacks on the top of the
leakage tester.
4.Set the leakage tester power switch to ON.
5.Set the rear panel power switch of the monitor to ON.
6.Read the leakage current indicated on the DMM.
If the reading is greater than 10 µA (10 mV on the DMM),
the unit under test fails this test and should be repaired
and tested again.
7.Change the leakage tester polarity switch to the RVS
position.
8.Read the leakage current indicated on the DMM.
•10 µA, (0.01 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•50 µA (0.05 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
9.Change the GND switch to the CLOSED position.
Electrical diagram:
Leakage Tester
patient source tests
HIGH
LOW
GND
POWER CORD
V*
*Meter reading: 1 mV = 1 µA (leakage current)
1K
0.15µF
10
NORM
RVS
GND
PATIENT JACKS
(TOP)
POWER CORD
TEST BODY
UNIT
UNDER
TEST
EAGLE 3000 PATIENT MONITOR
3 - 24415397-003PAGE REV C
Page 63
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
Patient source tests (Cont)
10.Read the leakage current indicated on the DMM.
•10 µA, (0.01 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•50 µA (0.05 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
11.Change the leakage tester polarity switch to the RVS
position.
12.Read the leakage current indicated on the DMM.
•10 µA, (0.01 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•50 µA (0.05 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
13.Set the power switch of the leakage tester to OFF.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 25
Page 64
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
PATIENTSINKTESTS
This tests ECG connector leakage current from a 115 or 220 V
source into the ECG/RESP connector of the monitor.
1.Set the leakage tester switches as follows:
•Selector knob - 5,
•GND switch - CLOSED,
•Polarity switch - NORM.
2.Disconnect the test cable from the leakage tester PATIENT
JACKS (TOP) and reconnect it to the PATN JACK connector
on the front panel of the leakage tester.
WARNING
The following step will cause high voltage (120 Vac to
240 Vac) to appear at the PATN JACK on the leakage
tester. Do not touch the PATN JACK posts or ECG
lead clips during this test as an electrical shock will
occur.
3.Set power switch on the leakage tester to ON.
4.Read leakage current indicated on DMM.
If the reading is greater than:
•10 µA, (0.01 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•50 µA (0.05 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
ac
Electrical diagram:
Leakage Tester
patient sink tests
HIGH
LOW
GND
POWER CORD
1K
V*
*Meter reading: 1 mV = 1 µA (leakage current)
120K
0.15µF
10
NORM
RVS
GND
PATN JACK
(FRONT)
POWER CORD
TEST BODY
(Keep test body
cable length as
short as possible.)
UNIT
UNDER
TEST
EAGLE 3000 PATIENT MONITOR
3 - 26415397-003PAGE REV C
Page 65
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
Patient sink tests (Cont)
Patient sink tests completion
TESTCOMPLETION
5.Change the leakage tester polarity switch to the RVS
position.
6.Read the leakage current indicated on the DMM.
If the reading is greater than:
•10 µA, (0.01 volts on the DMM), and the monitor is
operating at 120 V/60 Hz (U.S.); or
•50 µA (0.05 volts on the DMM), and the monitor is
operating at 220-240 V/50-60 Hz (non-U.S.);
the unit under test fails this test and should be repaired
and tested again.
7.Set the power switch on the leakage tester to OFF.
Disconnect all test equipment from the monitor. Disconnect the
monitor power cord plug from the leakage tester power receptacle. Disconnect the leakage tester from the wall receptacle.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 27
Page 66
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
HI-POT (DIELECTRIC
WITHSTAND) TESTS
Manufacturer recommendation
The high potential (Hi-Pot) tests provide a method of checking
patient isolation circuits and protect patients connected to the
monitor from potential electrical health hazards. These tests are
recommended for direct patient-connected medical devices to
check the integrity of the patient isolation circuitry after any
isolated component in the device has been repaired.
The manufacturer recommends that hi-pot tests be performed
whenever a circuit board in the patient-isolated portion of the
monitor is removed, repaired, or replaced. Examples of patientisolated components include, but are not limited to, the front
panel patient cable connectors, the isolated power supply, or any
patient data acquisition assemblies.
WARNING
Failure to perform hi-pot tests may cause undue
equipment failure and possible health hazards. The
manufacturer does not in any manner, unless an
Equipment Maintenance Agreement exists, assume
the responsibility for performing these recommended
hi-pot tests. The sole responsibility rests with the
individuals, hospitals or institutions utilizing this
equipment. Manufacturer service representatives
may, at their discretion, use this procedure as a
helpful guide during visits to the equipment site.
Test conditions
Test equipment
Pretest preparation
These tests may be performed under normal ambient conditions
of temperature, humidity, and pressure.
Equipment required to perform these tests is listed below.
Equivalent equipment may be substituted as necessary.
NameManufacturerPart Number
AC/DC Hi-Pot GeneratorHipotronicsAD125
ECG Test BodyMEIMT-3387
Follow these steps in the same order in which they are listed.
1.Set up the AC/DC Hi-Pot Generator in the following
manner:
•Power switch - ON,
•VOLTAGE RANGE selector - MEDIUM (10 kVA),
•RAISE VOLTAGE selector - 0 volts,
•OUTPUT & CURRENT selector - 1 mA range, and
•Allow the tester to warm up for 15 minutes before
continuing with this test.
2.Connect the ground pin on the power cord connector of the
monitor to the ground of the AC/DC Hi-Pot Generator.
EAGLE 3000 PATIENT MONITOR
3 - 28415397-003PAGE REV C
Page 67
ELECTRICAL SAFETY T ESTS (CONT)
MAINTENANCE
High Potential Tests (Cont)
Perform the AC hi-pot tests only on the ECG/RESP front panel
connector of the monitor.
CAUTION
Never attempt to perform this test on any of the other
front panel connectors of the monitor. Damage to the
monitor may occur if this test is performed on any of
the other front panel connectors.
1.Install the ECG test body in the ECG/RESP front panel
connector of the monitor.
2.Connect one end of a high voltage lead to the exposed lead
of the test body.
3.Connect the other end of the high voltage lead to the AC
OUT connector of the AC/DC Hi-Pot Generator.
WARNING
The following step will cause high voltage (4000 Vac)
to appear at the test body.
4.Set the HIGH VOLTAGE switch to ON. The high voltage
indicator should illuminate with this action.
NOTE
During this test, watch the analog meter to ensure
the current level never exceeds 1␣ mA. If it does, the
unit has failed the test and must be repaired then
tested again.
5.Slowly turn the RAISE VOLTAGE selector to 4000 volts.
6.Wait for 60 seconds. If the breakdown warning lamp
illuminates or the buzzer activates before the time expires,
then the unit has failed the test and should be repaired
then tested again.
7.Slowly turn the RAISE VOLTAGE selector to 0 volts.
8.Set the HIGH VOLTAGE switch to OFF. The high voltage
indicator should turn off.
9.If the unit under test fails, repairs must be made and the
unit must be tested again.
10.This completes the AC hi-pot test.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0033 - 29
Page 68
FORYOURNOTES
MAINTENANCE
EAGLE 3000 PATIENT MONITOR
3 - 30415397-003PAGE REV C
Page 69
4
TROUBLESHOOTING
Power Source Tests ................................................................ 4-2
DAS board symptoms ............................................... 4-20
Main processor board symptoms............................... 4-20
Power supply board symptoms.................................. 4-20
Isolating Problems on a Network .......................................... 4-21
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 1
Page 70
POWER SOURCE T ESTS
TROUBLESHOOTING
WALLRECEPTACLE
Neutral
Line
Ground
Voltage tests
Ground-to-neutral resistance test
Use this procedure to confirm AC power from the wall receptacle
which the monitor is plugged into.
Use a digital multimeter (DMM) to verify the wall receptacle is
wired correctly. This is accomplished by performing a:
•voltage measurement between all three connections of the
wall receptacle;
•ground-to-neutral loop resistance measurement.
A standard wall receptacle consists of three connections: line,
neutral and ground. The figure at left indicates the location of
each on a 120 Vac wall receptacle commonly used in the United
States. The location and shape of pins may be different on wall
receptacles used in countries other than the United States.
Perform the following tests:
1.Use a DMM to measure the voltage between the three
connections.
•Select the AC voltage scale on the DMM.
•Measure the voltage from line to neutral, line to
ground, and neutral to ground and make sure these
are correct. With a correctly wired wall receptacle
used in the United States, the following readings
should be obtained:
Line to neutral: 120 V
Line to ground: 120 V
Neutral to ground: < 3 V
Readings other than these indicate improper wiring.
Have the wall receptacle checked by an electrician.
2.Use a DMM to measure the ground-to-neutral loop
resistance.
ac
ac
ac
CAUTION
Do not check the ground-neutral loop resistance
unless the wall receptacle is correctly wired.
•Select the milliohms (mΩ) scale on the DMM.
•Measure resistance across the power cord ground and
neutral.
•Measure from the ground lug on the rear power
connector to any exposed metal of the monitor. The
resistance between the ground and neutral
connections, after the ohmmeter is nulled, must be
less than 100 mΩ. If not, have the wall receptacle
checked by an electrician.
EAGLE 3000 PATIENT MONITOR
4 - 2415397-003PAGE REV C
Page 71
POWER SOURCE T ESTS (CONT)
TROUBLESHOOTING
POWERCORDANDPLUG
MAINPOWERANDDISPLAY
POWER
CONTROL
Verify the power cord being used with the monitor is good. The
following are a couple of things to check for in this regard:
•Failure of the power cord strain relief is very common.
Often times users of the equipment will pull on the power
cord itself, rather than the power cord plug, to unplug the
monitor from a wall receptacle. If in doubt, test for
continuity through each conductor of the power cord
connector and plug.
•Verify line, neutral, and ground conductors are properly
connected to the power cord plug and are not shortcircuited. Rewire and tighten these, or replace the power
cord, as necessary.
Turn the rear panel main power switch of the monitor to the on
(1) position. During normal operation, the main power switch is
typically left in the on position. The DISPLAY ON/OFF front
panel control on the monitor is used for turning the display on or
off, depending on whether a patient is admitted to the monitor or
not.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 3
Page 72
DATA ACQUISITION T ESTS
TROUBLESHOOTING
ECG FUNCTIONS
1.Connect the Marquette Multifunction Microsimulator, pn
MARQ1, and appropriate patient cables, to the ECG
connector of the monitor. Turn the monitor and the patient
simulator on.
2.Set the monitor to display leads I, II, III, and V
simultaneously:
•From the main menu, select MONITOR SETUP.
•Make sure the DISPLAY menu item shows
INDIVIDUAL. If it shows FULL, change it to
INDIVIDUAL.
•Select WAVEFORMS ON/OFF from the menu.
•Set the displayed waveforms for the following ECG
leads:
ECG 1: LEAD II
WAVEFORM 2: LEAD V
WAVEFORM 3: LEAD I
WAVEFORM 4: LEAD III
3.Set the patient simulator to output calibration (cal) pulses
at 1.0 mV.
4.Measure the cal pulse () amplitude. These should be:
•Lead I: 0.5 mV
•Lead II: 1 mV
•Lead III: 0.5 mV
•Lead V: –0.5 mV
5.It may be necessary to run a graph to accurately measure
the cal pulses. Perform these steps to graph all four
waveforms.
•From the main menu, select GRAPH & ALARMS.
•Select GRAPH CONTROL from the menu.
•Set the graphed waveforms for the following ECG
leads:
ECG 1: LEAD II
WAVEFORM 2: LEAD V
•Press the GRAPH GO/STOP front panel control on the
monitor to start and stop a manual graph.
•Verify the printed graph shows proper cal pulses.
EAGLE 3000 PATIENT MONITOR
4 - 4415397-003PAGE REV C
Page 73
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
ECG functions (Cont)
ECG WAVEFORMSAREDISPLAYED
INCORRECTLY
ECG waveforms are not displayed
at all
6.Change the patient simulator output from cal pulses to an
80-bpm ECG waveform.
•The displayed ECG waveforms should be similar to
those shown in the figure below.
•If this is the case, the ECG functions of the
acquisition pcb, as well as communication between
the acquisition and processor pcb's, are functioning
as designed.
II
V
I
III
If the calibration pulses were not correct, test the patient
simulator using a working monitor. If the patient simulator is
functioning as designed, calibration of the acquisition pcb may
be necessary. Refer to Section 5: Calibration information in this
regard.
1.If displayed ECG waveforms contain a significant amount of
noise (see figure at left), check the ECG patient cables.
2.Test the patient simulator and ECG patient cables on a
working monitor to verify the ECG signal.
3.If the ECG signal, patient simulator and ECG patient cables
are good, the acquisition pcb is suspect and may need to be
replaced.
4.Test the ECG patient cables on a working monitor.
5.Test the patient simulator on a working monitor.
6.Swap the acquisition pcb into a working monitor. If the
symptoms follow the pcb into the working monitor, replace
the acquisition pcb.
7.If none of these first three steps provide any results, swap
the processor pcb and/or power supply pcb into a working
monitor.
Lead fail functions
PAGE REV C415397-0034 - 5
Perform the following steps to test lead fail detection function:
1.With the monitor displaying leads I, II, III, and V from the
patient simulator, remove the RA leadwire from the patient
simulator.
2.The monitor should display a RA FAIL message. Lead fail
detection is functioning properly if this is the case. Lead
fail detection is not functioning, if this is not the case. The
acquisition pcb is suspect. Swap the pcb with a working
monitor to verify the malfunction.
3.Reattach the RA leadwire to the patient simulator.
EAGLE 3000 PATIENT MONITOR
Page 74
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
PACEDETECTFUNCTIONS
1.With the monitor displaying leads I, II, III, and V, set the
patient simulator to output a VP1 (ventricular pacemaker
simulation #1) waveform.
2.Enable the pacemaker detection function of the monitor:
•select ECG from the display main menu,
•select DETECT PACE and set to PACE 1.
3.Verify the heart rate remains at approximately 80 bpm.
4.Select the VP2 output (ventricular pacemaker simulation
#2) on the patient simulator. The heart rate number may
disappear from the display for a few seconds and return to
the screen shortly thereafter. Verify the heart rate is at
approximately 80 bpm. Verify the pacemaker spikes
display at the same amplitude.
5.Disable the pacemaker detection function of the monitor.
Verify the displayed pacemaker spikes have a different
amplitude than in the previous step.
6.Select the AVS output (A/V sequential pacemaker
simulation) on the patient simulator. Again, verify the
displayed pacemaker spikes are at different amplitudes.
7.Enable the pacemaker detection function of the monitor
once again.
8.Verify a stable heart rate display of approximately 80 bpm.
Verify the pacemaker spikes are again at the same
amplitude.
9.Disable the pacemaker detection function of the monitor.
Pace detect functions are not
working properly
If the pacemaker detection test results are not correct, as
described above:
•Verify the patient simulator is functioning correctly by
testing it on a working monitor,
•The acquisition pcb is suspect. Swap a working
acquisition pcb into the monitor and perform these
test to verify correct operation.
EAGLE 3000 PATIENT MONITOR
4 - 6415397-003PAGE REV C
Page 75
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
INVASIVEBLOODPRESSURE
FUNCTIONS
Setup BP1
Setup BP2
Zero-reference both BP's
Generate dynamic BP waveforms
Verify dynamic BP results
The invasive blood pressure (BP) test procedure requires the use
of the following patient simulator: Marquette Multifunction
Microsimulator, pn. MARQ1. If use of a different patient
simulator is necessary, adjust the procedure steps/readings
accordingly.
1.Connect the BLOOD PRESSURE 1 output of the patient
simulator to the P1 patient connector on the front panel of
the monitor.
2.Connect the BLOOD PRESSURE 2 output of the patient
simulator to the P2 patient connector on the front panel of
the monitor.
3.Properly zero-reference each BP input:
•Set the patient simulator BP output to 0 mmHg
•Press the ZERO ALL front panel control on the
monitor.
4.Set the patient simulator BP output to WAVE.
5.Setup the BP scales on the monitor for auto gain:
•Select AR1 from the main menu of the monitor
•Select ART SCALES from the AR1 menu
•Select AUTO gain from the ART SCALES menu
•Return to the main menu of the monitor and setup
auto gain for the PA2 waveform as you did for AR1.
Once the BP waveforms are setup as described above verify
the following:
•Both the AR1 and PA2 BP waveforms are noise-free,
as shown in the figure at the left.
•BP displayed parameters are within tolerance as
indicated in the following list:
BP Parameter:AR1PA2
Systolic (mmHg)116 – 12428 – 32
Diastolic (mmHg)78 – 829 – 11
NOTE
These tests are designed for use with a MEI Multifunction Microsimulator, pn. MARQ1. Accuracy
specifications of the patient simulator in combination with the monitor (±2% or 1 mmHg, whichever is
greater) is how the parameter values listed above
were derived. Use of any other manufacturer patient
simulator and associated specifications will potentially change these test results.
Generate static BP waveforms
Verify static BP results
PAGE REV C415397-0034 - 7
6.Set the patient simulator BP output to 200 mmHg, static
pressure.
7.Verify the BP channels are working correctly if systolic,
diastolic, and mean pressure values for both AR1 and PA2
are displaying parameter readings between 194 and 206
mmHg.
EAGLE 3000 PATIENT MONITOR
Page 76
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
BP WAVEFORMSDONOTAPPEAR
CORRECTLY
BP waveforms do not appear on
ONTHEDISPLAY
the display at all
If the BP waveforms displayed on the monitor appear noisy or
distorted (example shown on the left), test the Patient simulator
and simulator test cables and on a working monitor to determine
the source of the problem.
1.If the static pressure test results were inaccurate, test the
Patient simulator and simulator test cables and on a
working monitor to determine the source of the problem.
2.If the patient simulator and associated test cables are
determined to be functioning correctly, the acquisition pcb
is suspect. Swap the acquisition pcb into a working
monitor to determine if replacement is necessary.
3.If the AR1 or PA2 parameter labels, readings and associated
waveforms do not display on the monitor, verify the patient
simulator and associated test cables on a working monitor.
4.Inspect the BP front panel connectors on the monitor for
bent or broken pins.
5.Perform continuity tests between the front panel connectors
of the monitor, front panel flex circuit assembly located
behind the front panel connectors and connection to the
acquisition pcb.
6.If the patient simulator and associated test cables are
determined to be functioning correctly and the continuity
tests yield no malfunction, the acquisition pcb is suspect.
Swap the acquisition pcb into a working monitor to
determine if replacement is necessary.
EAGLE 3000 PATIENT MONITOR
4 - 8415397-003PAGE REV C
Page 77
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
RESPIRATIONFUNCTIONS
(OPTIONAL)
Inspiration Markers
Expiration Markers
1.Connect the Marquette Multifunction Microsimulator, pn.
MARQ1, and appropriate patient cables to the ECG/RESP
front panel connector on the monitor.
2.Adjust the patient simulator to output a respiration
waveform using the following settings:
•Rate BPM - 30
•Baseline Impedance Ohms - 750,
•∆R Ohms - 2.0.
3.Enable the respiration function of the monitor:
•Select MONITOR SETUP from the main menu display
on the monitor,
•Select PARAMETERS ON/OFF from the monitor setup
menu.
Next, turn and push the Trim Knob to:
•scroll to and select RR in the parameters on/off pop-
up window.
•toggle and select ON in the RR line of the parameters
on/off pop-up window.
Verify the following:
•Respiration rate is displayed and accurate.
•Respiration waveform is displayed and noise-free.
•Markers appear in the displayed respiration waveform
(refer to figure at left). These indicate the points at
which the monitor senses inspiration and expiration
for determination of the respiration rate.
No respiration waveform or rate
appear on the display
PAGE REV C415397-0034 - 9
If the respiration waveform or rate does not appear on the
monitor display, perform the following steps to isolate the
problem:
•Vary the baseline impedance on the patient simulator
•Vary the ∆R on the patient simulator.
•Test the patient simulator and appropriate patient
cables on a working monitor to determine the source
of the problem.
•If none of the previous recommendations corrects the
problem, the acquisition pcb is suspect. Swap the
pcb into a working monitor to determine the source of
the problem and replace as necessary.
EAGLE 3000 PATIENT MONITOR
Page 78
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
Markers do not appear on the
respiration waveform; respiration
rate is inaccurate
Respiration functions work
properly when using a patient
simulator but not on an actual
patient
If the markers on the respiration waveform do not appear on the
display or the respiration rate count is inaccurate, try changing
the respiration sensitivity level on the monitor. To do this, use
the Trim Knob on the monitor to:
•Scroll to and select RR (respiration parameter) from
the monitor main menu,
•Scroll to and select SENSITIVITY from the respiration
parameter menu, and
•Scroll to and select a different sensitivity percentage
(%) from the sensitivity menu
NOTE
Usually, a lower respiration sensitivity % level rectifies this problem.
Refer to the Operator’s Manual for detailed information regarding
patient preparation relative to respiration monitoring functions.
Achieving optimum results for respiration waveforms and
accurate respiration rate detection by the monitor requires
proper preparation for ECG electrode placement on the patient.
An example of a noisy respiration waveform, usually due to bad
patient preparation, is shown at the left.
NOTE
With patients that exhibit excessively high baseline
chest impedance, proper respiration monitoring will
be extremely difficult, if not impossible.
EAGLE 3000 PATIENT MONITOR
4 - 10415397-003PAGE REV C
Page 79
DATA ACQUISITION T ESTS (CONT)
TROUBLESHOOTING
NON-INVASIVEBLOODPRESSURE
FUNCTIONS
Perform the non-invasive blood pressure (NBP) Checkout
Procedure found in Section 3: Maintenance. This procedure will
determine whether or not the NBP functions of the monitor are
working as designed or whether the monitor requires NBP
calibration.
If, after performing the prescribed checkout procedure, it is
determined that there are potential problems that NBP
calibration does not cure, try the following:
1.If calibration is unsuccessful and cannot be properly
performed, there could be leaks in the pneumatic circuit
plumbing. The following steps will assist you in
determining this:
•The NBP cuff and tubing is the easiest area to inspect
for leaks and is also the most likely area for failure in
this regard. Closely inspect these items for cracks or
leaks. Test the NBP cuff and tubing on a working
monitor to determine the source of the problem.
•If the NBP cuff and tubing are determined to be good
after testing them on a working monitor, the leaks are
probably internal to the monitor. Disassemble the
monitor and check inspect all internal tubing and
connections in the pneumatic circuit plumbing.
2.If no leaks are found after performing the previous step, the
NBP pump assembly is suspect. Swap the NBP pump
assembly with one from a working monitor and/or replace
as necessary.
NBP alarms occur continuously
Cannot get NBP readings from a
patient in under 3 minutes
NBP displayed readings are
inaccurate
If the monitor is not configured properly, a variety of NBP
problems may occur. To determine monitor configuration, rotate
then push the Trim Knob to:
•Scroll to and select MONITOR SETUP from the main
menu of the monitor,
•Scroll to and select SERVICE MODE from the monitor
setup menu of the monitor and enter the two-digit
numeric day and month shown in the upper-left
corner of the monitor display,
•Scroll to and select PATIENT-MONITOR TYPE from
the service mode menu of the monitor.
Verify the configured monitor type matches the environment in
which the monitor is being used. If it is set to a neonatal ICU
when the monitor is used for the adult ICU application or vice
versa, problems listed to the left may occur.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 11
Page 80
SERVICE MODE MENU
TROUBLESHOOTING
ABOUTTHESERVICEMODEMENU
Service mode menu option items
The SERVICE MODE menu option items provide the user access
to several general and technical built-in software functions of the
monitor. Only persons responsible for configuring and
maintaining the monitor should access the service mode menu
option items.
WARNING
The Service Mode menu is intended for use only by
qualified service technicians. Experimentation with
service mode menu option items can be detrimental
to the monitor. Lost patient data, damaged operating
system software for the monitor, even network related problems are but a few examples of problems
that can be induced as the result of tampering with
service mode menu option items.
Access to the service mode menu option items is necessary for
the following service-related functions of the monitor:
•REVIEW ERRORS - For troubleshooting difficult equipment
problems or network problems on a software engineering
level,
•ADMIT MENU - For setup or configuration of the monitor to
admit a patient with one of the following network
configuration features enabled:
»Standard, or
»Rover.
•CALIBRATE - For checkout or calibration of the noninvasive blood pressure and CO2 functions of the monitor,
•STAR TEST PATTERN - Generates a print test for an
optional built-in thermal array printer,
•SOFTWARE LEVEL - If enabled for 7020 feature, the menu
allows you to select 7020 or 7015.
•PATIENT - MONITOR TYPE - For setup or configuration of
one of three monitor operating modes. The three modes of
operation for the monitor are:
»Adult ICU,
»Neonatal ICU, or
»Operating Room.
•SET UNIT NAME - For setup or configuration of the monitor
care unit name,
•SET BED NUMBER - For setup or configuration of the
monitor bed number or bed name,
•SET INTERNET ADDRESS - For setup or configuration of
the monitor Internet address for the network,
•TIME AND DATE - For entering or changing the monitor
time and date.
EAGLE 3000 PATIENT MONITOR
4 - 12415397-003PAGE REV C
Page 81
SERVICE MODE MENU (CONT)
TROUBLESHOOTING
Access to the service mode
menu
Select monitor setup from the
main menu
Select service mode from the
monitor setup menu
Enter the service menu password
0106
SERVICE
MODE
Begin setup by entering into the service mode menu of the
monitor. Follow these steps:
1.Make sure all cables are properly connected to the monitor.
2.Apply AC power to the monitor.
•Plug the power cord into a working AC power wall
receptacle and turn the monitor rear panel main
power switch to the on (1) position,
•Press the DISPLAY ON/OFF front panel control on the
monitor. The display should be on.
3.Use the Trim Knob control to scroll to MONITOR SETUP in
the monitor main menu and press the Trim Knob control to
select it.
ALARM
CONTROL
PATIENT
DATA
MONITOR
SETUP
PATIENT:
DISCHARGED
4.Use the Trim Knob control to scroll to SERVICE MODE in
the monitor setup menu and press the Trim Knob control
to select it.
MAIN
MENU
WAVEFORMS
ON / OFF
UNIT ALARMS:
OFF
DISPLAY:
INDIVIDUAL
BRIGHTNESS:
100%
TIME AND
DATE
LEARN THE
MONITOR
PARAMETERS
ON / OFF
SOFTWARE
REVISION
GRAPH
SETUP
SOFTWARE
COMPATIBILITY
MONITOR
DEFAULTS
SERVICE
MODE
5.A service menu password window will appear on the monitor display, as shown in the figure at the left. A password
is required to prevent non-service personnel from accessing
the service menus. The password is four numbers that
represent the date that currently resides in a memory
circuit within the monitor (please note that this may or may
not be the correct date). In the password, the first two
numbers, starting from the left, represent the day and the
second two numbers represent the month of whatever date
that currently resides in the memory circuits of the monitor. For example, the seventh day of the third month (June
1st) would be represented in the password as 0106 (ddmm).
Note the date that is currently on the monitor display and
follow these steps to enter the password;
•Rotate the Trim Knob control to highlight the pass-
word number that you would like to change.
•To change the highlighted number, press the Trim
Knob control.
•Rotate the Trim Knob control until the correct number
is displayed in the selected field.
•To enter the number, press the Trim Knob control.
•Repeat these steps until all password numbers are
correctly displayed.
•Once you have entered the correct password numbers,
rotate the Trim Knob control to highlight SERVICE
MODE in the enter password window.
•Press the Trim Knob control one more time to enter
the password and access the service menus of the
monitor.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 13
Page 82
SERVICE MODE MENU (CONT)
TROUBLESHOOTING
ABOUTSERVICEMODEMENU
OPTION
ITEMS
Service mode menu option items are used for many purposes in
the monitor. The majority of the functions of these menu option
items are for initial setup and configuration. Some of the
functions are for troubleshooting as well. Caution should always
be exercised when using any of these password-protected
functions.
Service mode menu option items are used by service technicians
to: relay software information to design engineers; calibrate and
troubleshoot NBP functions of the monitor; set admit menu
options, software feature levels and operating mode of the
monitor; configure the monitor unit name, bed number and
Internet address for use on the network; and enter or change the
time and date on the monitor. None of these options should be
used unless specifically instructed to do so.
WARNING
Some of the service mode menu option items are to
be used only by qualified service technicians and
others are for general use. Because of this, unnecessary tampering with service mode menu option
items for experimentation purposes is not recommended by the manufacturer and may cause a
malfunction of the monitor.
REVIEWERRORS
About the monitor error log
Downloading the error log
Accessing the review errors menu
option item
The REVIEW ERRORS menu option item is mostly used as an
advanced troubleshooting technique by manufacturer
engineering personnel. Some of the information recorded in the
monitor error log can be useful for field service troubleshooting.
Details included in this part of the section will provide an
introduction to error log usage and meaning. Because the
information contained in the error log is engineering-oriented,
the intent of the manual is to simply provide a general
understanding of this monitor function.
A method for downloading error log data over the network to a
central station is included in this part of the section. Once
downloaded to a central station, the error log data can be loaded
onto floppy diskettes, or reviewed on the central station.
To access the error log and learn more about the REVIEW
ERRORS menu option item, follow these steps:
1.Rotate and press the Trim Knob control to scroll to and
select REVIEW ERRORS from the service mode menu
option items.
MAIN
MENU
SOFTWARE
LEVEL
PATIENT - MON ITOR TYPE:
REVIEW
ERRORS
ADULT - ICU
STANDARD
SET UNIT
NAME
CALIBRATEADMIT MENU:
SET BED
NUMBER
SET INTERNET
ADDRESS
STAR TEST
PATTERN
TIME AND
DATE
EAGLE 3000 PATIENT MONITOR
4 - 14415397-003PAGE REV C
Page 83
MAIN
MENU
PREVIOUS
MENU
Status Register :
Program Counter :
User Stack Pointer :
Super Stack Pointer :
Heap Pointer :
Process Name :
Error Code :
Severity :
Date :
Time :
Error Number :
RUN TIME ERROR LOG
VIEW OUTPUT
ERRORS
>
NEXT
PREVIOUS
QUIT
2000
10017A
981794
94FFAE
910F78
system
warmstart:0
Continue
JUL 19 1995
13: 34: 52
1
415230-002 2A 11JAN96
SERVICE MODE MENU (CONT)
TROUBLESHOOTING
Viewing output errors
Run time error log pop-up
window
2.The review errors menu option items include four possible
selections; one each for viewing output or input errors
along with one each for clearing output or input errors.
Rotate and press the Trim Knob control to scroll to and
select VIEW OUTPUT ERRORS from the review errors menu
option items.
MAIN
MENU
PREVIOUS
MENU
VIEW OUTPUT
ERRORS
CLEAR OUTPUT
ERRORS
VIEW INPUT
ERRORS
CLEAR INPUT
ERRORS
3.The RUN TIME ERROR LOG pop-up window appears on the
left side of the monitor display. One time-dated output
software error appears in the pop-up window at a time.
Use the Trim Knob to navigate
PAGE REV C415397-0034 - 15
through the error log
The Trim Knob control can be used to scroll through each
logged error, perusing all of the parameters associated with
each output software error. Rotate the Trim Knob control
to move the cursor ( > ) to a position for viewing the NEXT
or PREVIOUS error as well as the position that allows the
user to QUIT viewing output errors.
Selecting QUIT closes the run time error log pop-up
window and returns to the review errors menu option
items.
EAGLE 3000 PATIENT MONITOR
Page 84
SERVICE MODE MENU (CONT)
MAIN
MENU
PREVIOUS
MENU
Status Register :
Program Counter :
User Stack Pointer :
Super Stack Pointer :
Heap Pointer :
Process Name :
Error Code :
Severity :
Date :
Time :
Error Number :
Parameter :
RUN TIME ERROR LOG
VIEW INPUT
ERRORS
>
NEXT
PREVIOUS
QUIT
0
22234
0
0
20EFB6C
dasmsg_in
2001
Continue
JUN 25 1995
09: 37: 06
5
4F7
VIEW OUTPUT
ERRORS
CLEAR OUTPUT
ERRORS
415230-002 2A 11JAN96
TROUBLESHOOTING
View input errors
4.The VIEW INPUT ERRORS menu option item, when selected
using the Trim Knob, causes a RUN TIME ERROR LOG
pop-up window to appear on the monitor display. The popup window now displays input software errors and provides
basically the same information as the VIEW OUTPUT
ERRORS pop-up window provided. The appearance of both
pop-up windows are similar, the difference being errors that
are logged as input versus output to/from the monitor.
4 - 16415397-003PAGE REV C
Clearing the error log
5.To clear out the stored run time error logs, use the Trim
Knob to scroll to and select the CLEAR OUTPUT ERRORS
or CLEAR INPUT ERRORS menu option item, respectively.
OUTPUT SOFTWARE ERRORS HAVE BEEN CLEARED
MAIN
MENU
PREVIOUS
MENU
VIEW OUTPUT
ERRORS
CLEAR OUTPUT
ERRORS
VIEW INPUT
ERRORS
CLEAR INPUT
ERRORS
Immediately following the assertion of the Trim Knob to
clear one of the error logs, a message appears directly
above the menu option items, on the right side of the
display. The message verifies the actuation of the Trim
Knob for this function.
EAGLE 3000 PATIENT MONITOR
Page 85
SERVICE MODE MENU (CONT)
TROUBLESHOOTING
MOREABOUTREVIEWERRORS
How much data actually is in the
error log
Using information in the error log
This part of the section describes in greater detail what
information the error log contains and what can be learned from
error logs.
An error log in the monitor is constructed as a circular file (not
referring to a wastepaper basket). This circular file can hold up
to 50 events. As an event occurs, error information is stored in
the log. Subsequent events are stored sequentially as they occur.
When the 50-event limit is reached, the next error (the 51st
error) is written over the first event that was logged, erasing that
event and replacing it with the latest event. The 52nd event is
written over the second event, and so on. If errors occur
infrequently the error log could span a period of weeks and
months, maybe even years. For example; if a problem with the
network begins, repeating frequently, the error log might consist
only of errors from the last few hours. In any case the error log
will contain the most recent 50 errors that were detected and
recorded.
A sample of the monitor error log pop-up window appears as
follows:
RUN TIME ERROR LOG
Status Register :
Program Counter :
User Stack Pointer :
Super Stack Pointer :
Heap Pointer :
Process Name :
Error Code :
Severity :
Date :
Time :
Error Number :
2000
10017A
981794
94FFAE
910F78
system
warmstart:0
Continue
JUL 19 1995
13: 34: 52
1
415230-002 2A 11JAN96
Error log categories of greatest
interest for troubleshooting
purposes
When using the error log to troubleshoot a problem with the
monitor, the following parameters from the pop-up window that
are of greatest interest are:
•Process Name: The task that was operating when the event
or problem occurred,
•Error Code:A software code for the type of event or
problem that occurred,
•Severity:Indicates the level of impact of the event or
problem on the system,
•Date:The date the event or problem occurred,
•Time:The time the event or problem occurred,
and
•Error number: A sequential number (0-50) used to identify
each event or problem.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 17
Page 86
SERVICE MODE MENU (CONT)
TROUBLESHOOTING
ERRORLOGS
Error code descriptions
Something to remember about the error log is that it contains
more than just operating system errors. Many events that occur
that might have an impact upon the system are entered into the
log. The 700-series of error codes include many such events.
Some of the event/error codes you might find useful are
described in the following table:
Error CodeDescription
1A00-1AFFNetwork errors were detected.
703Diagnostics test were completed.
70BInternet address was changed. The network
address for the monitor was changed. Any network address changes should only be done by
qualified service personnel.
70ETime was changed from this monitor.Helps
determine how the system-wide time may have
been altered.
70FDate was changed from this monitor. Helps
determine how the system-wide time may have
been altered.
Severity of the error
Severity is a measure of how the event/error affected the system.
There are three levels of severity. The following is a list of these
levels accompanied by a brief description of each:
•Continue: The event or error was logged, the task may or
may not have completed, but the system was able to
continue operating. Most error log entries will have this
severity level.
•Fatal: The event or error was logged, the task did not
complete, and the system was unable to continue operating
as recovery was not possible. This level of severity in an
event or error is always followed by an automatic warm
start.
•Forced Restart; The operating system restarted normally
after a known condition, such as an Internet address
change, patient discharge, etc.
EAGLE 3000 PATIENT MONITOR
4 - 18415397-003PAGE REV C
Page 87
SERVICE T IPS
TROUBLESHOOTING
FAULT/SYMPTOMANALYSIS
The power LED on the front panel
flashes every 2 seconds
This information is provided for the benefit of service technicians
responsible for the maintenance and repair of the monitor. The
symptoms covered in this part of the Troubleshooting section
represent only a select number of faults that you may encounter
and by no means are intended to cover every possible failure that
may occur.
A systematic approach to the diagnosis of problems as well as a
general understanding of the architecture, both hardware and
software, of the monitor are essential to ensure successful
troubleshooting of this device. The manufacturer recommends
formal service training before repairs are attempted on the
monitor. The Service Tips listed below combined with formal
training should provide the service technician with skills
necessary to service and repair a monitor, in the event of a
malfunction.
Reason: The +5 Vdc supply voltage is in an over-current
condition.
•Replace the main processor board. This is the only board
that uses the +5 Vdc supply voltage. The power supply pcb
assembly is working properly by the fact that it is restarting
after detection of the over-current condition, thus causing
the front panel LED to flash at a regular interval.
•Check the power supply board. This board can be benchtested as per the Calibration section of this manual.
Video problem - the patient
waveforms are displayed correctly
but the alphanumerics are
displayed improperly or are not
displayed at all or vice-versa
Video problem - there are bars/
strips of pixels missing on the
display in rows/columns. Or only
one row/column of pixels on the
display is missing or never turned
on. The remaining portion of the
display functions properly.
Reason: The graphics processing (video) circuitry on the
processor pcb has problems attempting to "clock-out" the text
information data that is stored in video memory circuits (VRAM).
•Replace the main processor board. This is the only area of
the monitor where text (alphanumerics) information and
graphics (patient waveforms) information is processed
separately.
Reason: The active matrix display has a defective row or column
driver. If the entire driver has failed, the display will have a
whole missing strip of display area. If just a part of the driver
has failed, the display will have only a single missing line of
display area.
•Replace the display assembly. This is an EL
(electroluminescent) monochrome display assembly. This
type of display has specific drivers for rows and columns
integrated into the display assembly and, therefore, cannot
be repaired.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 19
Page 88
SERVICE T IPS (CONT)
TROUBLESHOOTING
DAS BOARDSYMPTOMS
MAINPROCESSORBOARD
SYMPTOMS
POWERSUPPLYBOARDSYMPTOMS
+12 Vdc (+12 MAIN) supply
applications
Symptoms relative to patient signal acquisition such as missing
parameter text and waveform(s) may be associated with
acquisition pcb assembly failure. It is important that you are
able to distinguish the difference between the general format of
the display, which is generated by the processor pcb assembly,
versus the patient signals and data that is associated with these
patient signals, a function of data acquisition, which is generated
by the acquisition pcb assembly.
Symptoms with network communications, asynchronous
communications, NBP control, analog output, audio/sound
generation, and communications as well as other display-related
problems all may be associated with processor pcb assembly
failure. All of these are functions controlled by microcontroller or
graphics processing circuitry located on the processor pcb
assembly.
The power supply pcb assembly provides power that is used
throughout the monitor. All of the supply voltages are
distributed to the processor pcb assembly for various
application. Below is a list of the supply voltages and where and
how these voltages are applied. Problems in any of the following
areas may be associated with power supply pcb assembly failure.
Following is a list of functions for each of the supply voltages
generated on the power supply pcb:
•DAS board - main power source
•Ethernet transceiver - power source
•Display assembly - power source
•Defib marker out - power source for defib sync jack
•Audio amplifier - power source (speaker)
•NBP compressor (pump assembly) and solenoid valves power source
•Main memory - FLASH memory programming power source
+5 Vdc (+5) supply applications
±12 Vdc (±12 ANALOG) supply
applications
4 - 20415397-003PAGE REV C
•Main processor board - logic power source
•Display assembly - logic power source
•Analog ECG/blood pressure - signal generation for defib
sync jack
EAGLE 3000 PATIENT MONITOR
Page 89
TROUBLESHOOTING
ISOLATING PROBLEMSONA NETWORK
Start
BS: Bedside Monitor
CS: Central Station Monitor
Is more
than one CS having
a problem?
Y
NOTE- If there is only one
CS on the network,
A
Is there
any display at all at
the CS?
'A' may pertain only if a
network problem exists.
Y
Is there a
NO COMM
message at the
CS?
Y
Under LISTWHAT
no BS's should be
available. Only CS
is.
Attempt to view
BS's on another
CS.
N
than one BS display
Refer to the
troubleshooting
N
section in the CS
service manual.
Under LISTWHAT,
the BS's on the
N
network should be
present. Possible
incorrect UNIT.
NOTE- You should
be able to view BS's
on another CS.
Is more
missing at the
CS?
End
N
Y
Is there a
NO PARAMETERS
message at the
CS?
Y
Verfiy the BS is
operating. Also
verfiy BS is under
AC power and not in
battery fail.
End
Connect AUI cable
to the ETHERNET
port at the BS. The
BS should appear
on the network.
N
N
Is there a
NO COMM
message at the
CS?
Y
Under LISTWHAT,
check to see if the
BS in question is on
the network.
BS should not be
present. There is a
problem at the BS.
Go to the BS and
check the AUI cable.
Is the AUI
cable connected at
the BS?
N
Under LISTWHAT,
check to see if the
BS in question is on
the network.
BS should be
present. The UNIT
name at the BS
should be incorrect.
At the BS, enter the
correct UNIT name.
The BS should
appear on the
network.
Y
Possible problem at
CS. Defective LAN
pcb, AUI cable or
MAU (Transceiver).
Replace MAU.
If BS's ID
themselves, it's a
defective MAU.
If BS's still don't ID
themselves.
Replace AUI.
NOTE-If the MAU or AUI cable do not solve
the problem, the CS LAN Controller pcb
may be defective. Check status LEDs and
event log. Reboot system and check for
errors.
End
B
End
End
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0034 - 21
Page 90
TROUBLESHOOTING
ISOLATING PROBLEMSONA NETWORK (CONT)
A
BS: Bedside Monitor
CS: Central Station Monitor
Is there a
NO COMM
message at the
CS?
Y
Under LISTWHAT,
there should not be
any BS's listed.
Only the CS itself
should be listed..
Verify with
ocilliscope proper
network signal
voltage levels
( 0 Vdc to -2Vdc)
Do
voltage
levels swing from
0Vdc to -4Vdc?
Y
Excessive collisions
across the network.
Possible hardware
failure. BS's/CS's
'talking' too much.
Are
voltage levels OK?
N
Possible open in
network cable. NO
COMM message
appears in all BS
windows at CS.
Remove all AUI
cables from each
MAU. Remove one
50Ω terminator from
the network cable.
Using an ohmmeter,
measure from shield
to center conductor.
High Ω indicates an
open in the cable.
Under LISTWHAT,
the BS's should be
N
listed.
Probable incorrect
UNIT name at either
the BS or the CS.
End
N
NOTE-A device 'talking' too much on the
network may cause dropout at the CS.
A possible ground-loop or an excessive
ground potential difference may exist.
Normal ground potential difference is
1Vac or less.
Under LOG directory
at the CS, list and
Y
read the contents
from the current
and/or previous day.
Inspect SCOPE ER
-RORS under REVIEW ERRORS at
BS. If errors are ex
-cessive, reboot BS.
Unsuccessful boot
indicates BS failure.
End
NO COMM
message at CS
indicates possible
short in the network
cable.
Remove all AUI
cables from each
MAU. Remove one
50Ω terminator from
the network cable.
Using an ohmmeter,
measure from shield
to center conductor.
0 ohms indicates a
short in the cable.
For Thin-net:
Reinstall the 50Ω
terminator. Break
the network at mid-
point. Measure
resistance each
way.
For Thick-net:
A time domain
reflectometer (TDR)
is necessary to
scope the cable.
Isolate shorted
section of network
cable. Replace
defective section or
defective connector.
NOTE-You can check
the network cable for a
short using an o'scope.
With the system still
connected, scope one
end of the cable
monitoring the network
activity. If the data
signal swings above
and below the dc
baseline, there is a
short in the cable. The
data signal should
swing from 0 to -2Vdc
since it is ECL logic.
For Thin-net:
Reinstall the 50Ω
terminator. Break
the network at mid-
point. Measure
resistance each
way.
For Thick-net:
A time domain
reflectometer (TDR)
is necessary to
scope the cable.
Isolate opened
section of network
cable. Replace
defective section or
defective connector.
End
End
EAGLE 3000 PATIENT MONITOR
4 - 22415397-003PAGE REV C
Page 91
TROUBLESHOOTING
ISOLATING PROBLEMSONA NETWORK (CONT)
B
Possible defective
MAU or AUI cable.
Also check for a bad
connection on the
MAU.
Possible incorrect
internet address.
On ETHERNET
connector at BS,
check if 12Vdc is
present at pin 13.
Is 12Vdc present?
Possible defective
N
BS: Bedside Monitor
CS: Central Station Monitor
processor pcb
Y
Possible defective
MAU or AUI cable.
NOTE-There is a possibility that the MAU is defective. On a busy
network, it may be difficult to determine whether the MAU is detecting carriers properly. The MAU may be continually detecting
carriers and will not allow the BS to transmit.
This section summarizes the calibration for the monitor.
Noninvasive blood pressure (NBP) is the only function that
requires software calibration. The manufacturer recommends
performing the NBP software calibration upon receipt of the
monitor initially, and once each year thereafter. The NBP software calibration should also be performed whenever the monitor
is opened for service purposes. This will ensure the pneumatic
circuit plumbing has not developed any air leaks as a result of
disassembly.
Leakage current tests, checkout procedures and hi-pot tests are
recommended by the manufacturer if a circuit board has been
repaired or replaced in the monitor. These can be found in
Section 3: Maintenance.
EAGLE 3000 PATIENT MONITOR
5 - 2415397-003PAGE REV C
Page 95
NON-INVASIVE BLOOD PRESSURE
CALIBRATION
ABOUTTHEPROCEDURE
MANUFACTURER
RECOMMENDATION
TESTEQUIPMENT
The overall accuracy of non-invasive blood pressure (NBP)
readings by the monitor depend on the following:
•the zero pressure reading, and
•the voltage span of the NBP sensor in the monitor.
This procedure provides a method of verifying these items are
accurate and also checks the NBP pneumatic circuit plumbing
for leaks.
The manufacturer recommends performing this procedure upon
initially receiving the monitor, before it is used on a patient, and
once each year thereafter. Also, perform the procedure each time
the monitor is opened for service or repair, simply to verify the
NBP pneumatic circuit plumbing did not develop inadvertent air
leaks.
The following items are required to successfully complete the
NBP calibration procedure:
•Manometer (Sensym PDM200M or mercury manometer),
•NBP tube, pn. 414873-001,
•NBP cuff, pn. 9461-301 (any size will work),
•Something to wrap the NBP cuff around (PVC pipe or coffee
can),
•The table below lists items for connecting the NBP tube
between the manometer and NBP cuff:
When the NBP cuff is used in this procedure, it must
be tightly wrapped around a rigid cylinder or pipe.
Do not put the NBP cuff around a human arm during
the calibration procedures due to the potential for
injury.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0035 - 3
Page 96
NON-INVASIVE BLOOD PRESSURE (CONT)
CALIBRATION
CALIBRATIONPROCEDURE
Enter the proper password to
access the service menu
0106
SERVICE
MODE
1.Remove all cables except for the power cord from the
monitor.
2.Apply power to the monitor.
•Plug the power cord into a working AC power wall
receptacle and turn the monitor rear panel main
power switch to the on (1) position,
•Press the DISPLAY ON/OFF front panel control on the
monitor. The display should be on.
3.Use the Trim Knob control to scroll to MONITOR SETUP in
the monitor main menu and press the Trim Knob control to
select it.
ALARM
CONTROL
PATIENT
DATA
MONITOR
SETUP
PATIENT:
DISCHARGED
4.Use the Trim Knob control to scroll to SERVICE MODE in
the monitor setup menu and press the Trim Knob control
to select it.
MAIN
MENU
WAVEFORMS
ON/OFF
UNIT ALARMS:
OFF
DISPLAY:
INDIVIDUAL
BRIGHTNESS:
100%
LEARN THE
MONITOR
PARAMETERS
ON/OFF
SOFTWARE
REVISION
GRAPH
SETUP
SOFTWARE
COMPATIBILITY
MONITOR
DEFAULTS
SERVICE
MODE
5.A service menu password window will appear on the
monitor display, as shown in the figure at the left. A
password is required to prevent non-service personnel from
accessing the service menus. The password is four
numbers that represent the date that currently resides in a
memory circuit within the monitor (please note that this
may or may not be the correct date). In the password, the
first two numbers, starting from the left, represent the day
and the second two numbers represent the month of
whatever date that currently resides in the memory circuits
of the monitor. For example, the seventh day of the third
month (June 1st) would be represented in the password as
0106 (ddmm). Note the date that is currently on the
monitor display and follow these steps to enter the
password;
•Rotate the Trim Knob control to highlight the
password number that you would like to change.
•To change the highlighted number, press the Trim
Knob control.
•Rotate the Trim Knob control until the correct number
is displayed in the selected field.
•To enter the number, press the Trim Knob control.
•Repeat these steps until all password numbers are
correctly displayed.
•Once you have entered the correct password numbers,
rotate the Trim Knob control to highlight SERVICE
MODE in the enter password window.
•Press the Trim Knob control one more time to enter
the password and access the service menus of the
monitor.
EAGLE 3000 PATIENT MONITOR
5 - 4415397-003PAGE REV C
Page 97
NON-INVASIVE BLOOD PRESSURE (CONT)
CALIBRATION
Service menus
Enter the NBP calibration menus
The service menus should appear on the monitor display. These
next steps guide you through the service menus associated with
checking NBP calibration. If desired test results are not
obtained, NBP calibration will be necessary.
6.Rotate the Trim Knob control to highlight CALIBRATE and
press the Trim Knob control to select it. Next, rotate the
Trim Knob control to highlight CALIBRATE NBP and press
the Trim Knob control to select it.
MAIN
MENU
SOFTWARE
LEVEL
MAIN
MENU
PREVIOUS
MENU
REVIEW
ERRORS
PATIENT-MONITOR TYPE:
ADULT-ICU
CALIBRATE
NBP
CO2
SERVICE
ADMIT MENU:
STANDARD
SET UNIT
NAME
CALIBRATE
SET BED
NUMBER
SET INTERNET
ADDRESS
STAR TEST
PATTERN
TIME AND
DATE
7.Rotate the Trim Knob control to highlight CAL ZERO OFF,
and then press the Trim Knob control to select it.
NBP zero calibration
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
OFF
8.Rotate the Trim Knob control to highlight START, and then
press the Trim Knob control to select it. The CAL ZERO
menu item will show that it’s IN PROGRESS, and when it’s
done it will show that it’s OFF again.
MAIN
MENU
PREVIOUS
MENU
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL ZERO
IN PROGRSS
CAL GAIN
OFF
> START
STOP
CHECK CAL
OFF
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
OFF
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0035 - 5
Page 98
NON-INVASIVE BLOOD PRESSURE (CONT)
CALIBRATION
NBP calibration setup
9.Connect a cuff and manometer to the monitor as shown
below.
NBP tubing,
pn 414873-001
6-inch diameter PVC
pipe (or 1-pound
coffee can):
Wrap the NBP cuff
around this for tests.
NBP cuff,
pn 9461-301:
Any size NBP cuff
will be sufficient.
Coupling,
pn 46100-002:
Connects manometer
tubing to NBP tubing.
3-way tee,
pn 4745-101:
Connects manometer
tubing, NBP tubing
and NBP cuff tubing.
Digital manometer,
Sensym PDM200M:
Note: A calibrated
mercury manometer
may be substituted.
Set up the manometer10.Turn the manometer on and adjust the range switch to the
Manometer tubing,
pn 401582-001:
Approximately two
feet in length.
1000␣ mmHg setting.
Coupling,
pn 400787-001:
Connects manometer
tubing to NBP cuff
tubing.
EAGLE 3000 PATIENT MONITOR
5 - 6415397-003PAGE REV C
Page 99
NON-INVASIVE BLOOD PRESSURE (CONT)
CALIBRATION
Start the gain calibration test
11.Rotate the Trim Knob control to highlight CAL GAIN OFF,
and then press the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL ZERO
OFF
CAL GAIN
OFF
CHECK CAL
OFF
12.Rotate the Trim Knob control to highlight CAL GAIN OFF,
and then press the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL GAIN
OFF
ENTER CAL
PRESSURE
13.Rotate the Trim Knob control to highlight START, and then
press the Trim Knob control to select it. The second line of
text on the CAL GAIN menu item changes from HOLDING to
INFLATING. Then, the monitor starts pumping up the
pressure bulb or cuff—the audible whirring sound of the
NBP pump motors will occur and an increase in displayed
pressures on both the monitor and the manometer will be
observed.
MAIN
MENU
PREVIOUS
MENU
MAIN
MENU
PREVIOUS
MENU
MAIN
MENU
PREVIOUS
MENU
CAL GAIN
OFF
CAL GAIN
HOLDING
ENTER CAL
PRESSURE
CAL GAIN
INFLATING
ENTER CAL
PRESSURE
> START
STOP
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0035 - 7
Page 100
NON-INVASIVE BLOOD PRESSURE (CONT)
CALIBRATION
Verify the pneumatic circuit
plumbing does not have air leaks
Perform the software calibration
of the monitor
14.The pump shuts off at about 250 mmHg, and the pressure
drops slowly to about 240 mmHg before stabilizing. The
second line of text on the CAL GAIN menu item changes
from INFLATING back to HOLDING. If the pressure
continues to drop at a rate of 1 mmHg or more for every five
seconds, there is a leak in the NBP plumbing. If there is a
leak in the NBP plumbing, repair it and restart this
calibration procedure.
MAIN
MENU
PREVIOUS
MENU
MAIN
MENU
PREVIOUS
MENU
CAL GAIN
INFLATING
ENTER CAL
PRESSURE
CAL GAIN
HOLDING
ENTER CAL
PRESSURE
15.Rotate the Trim Knob control to highlight ENTER CAL
PRESSURE and press the Trim Knob control to select it.
MAIN
MENU
PREVIOUS
MENU
CAL GAIN
HOLDING
ENTER CAL
PRESSURE
16.An ENTER CAL PRESSURE pop-up window will appear.
Use the Trim Knob control to select a pressure value that is
1 mmHg lower than the current manometer reading.
MAIN
MENU
PREVIOUS
MENU
ENTER CAL
PRESSURE
247
17.When the manometer falls to exactly the value that you
selected in the pop-up window, press the Trim Knob control
to enter the value.
EAGLE 3000 PATIENT MONITOR
5 - 8415397-003PAGE REV C
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.