Due to continual product innovation, specifications in this
manual are subject to change.
TRADEMARKSAND COPYRIGHT
Registered trademarks
Trademarks
Copyright
Trademarked names appear throughout this document. Rather
than list the names and entities that own the trademarks or
insert a trademark symbol with each mention of the trademarked
name, the publisher states that it is using the names only for
editorial purposes and to the benefit of the trademark owner with
no intention of improperly using that trademark.
ACCUSKETCH, APEX, AQUA-KNOT, ARCHIVIST, BABY MAC,
CASE, CD TELEMETRY, CENTRA, CRG PLUS, EAGLE, LASER
SXP, MAC, MAC-LAB, MACTRODE, MARQUETTE, MARQUETTE
UNITY NETWORK, MARS, MAX, MEI, MEI in the circle logo,
MINNOWS, MULTI-LINK, MUSE, PRESSURE-SCRIBE, PRES-RCUFF, RAMS, SAM, SEER, SOLAR, ST GUARD, TRAM, TRAMNET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, UNITY NETWORK,
and UNITY NETWORK twist are trademarks of Marquette Medical
Systems registered in the United States Patent and Trademark
Office.
12SL, 15SL, ADVANTAGE, AUTOSEQ, BODYTRODE, CardioServ,
CardioSmart, CardioSys, CD TELEMETRY®-LAN,
CENTRALSCOPE, EDIC, FLEX, HI-RES, IMAGE VAULT, INTERLEAD, LIFEWATCH, MARQUETTE® RESPONDER, MENTOR,
MRT -II, O2SENSOR, OnlineABG, SILVERTRACE, SMART-PAC,
and SMARTLOOK are trademarks of Marquette Medical Systems.
This field service manual has been prepared by the technical
publications staff at Marquette Medical Systems, Inc. It is
intended for use by biomedical electronic technicians or other
qualified service personnel responsible for installation,
maintenance or repair of the Eagle 3000 Patient Monitor
(hereafter referred to as the monitor).
The content of this field service manual is aimed primarily at
biomedical equipment technicians and field service personnel.
The user of this field service manual is expected to have a solid
background in electronics, including strong backgrounds in
analog and digital electronics, as well as microcomputer
technology familiarity.
The field service manual is organized into sections, as follows:
•Section one describes the field service manual, manual
page layout, related documentation, manufacturer
responsibility, notes/cautions/warnings, and abbreviation.
•Section two describes the product, the Marquette Unity
Network, technical specifications, preparation for use,
product part numbers and theory of operation.
•Section three describes the maintenance schedule, visual
inspection, cleaning the monitor, checkout procedures,
leakage current tests and hi-pot tests.
•Section four describes electrostatic discharge, special
components, power source tests, data acquisition tests,
service tips, the service mode menu and a network related
troubleshooting flow chart.
•Section five describes adjustments and non-invasive blood
pressure calibration.
•Section six describes monitor configurations, installing
software, loading software from diskettes, loading software
over the network and setup for use.
•Section seven provides assembly drawings of the monitor.
These include electrical diagrams, schematics, and
exploded views.
EAGLE 3000 PATIENT MONITOR
1 - 2415397-003PAGE REV C
PAGE LAYOUT
ABOUT T HIS MANUAL
Section Topic: Each section is
divided into topics. This line
indicates what topic within the
section is covered on this and
possibly subsequent pages.
Left Column: Most pages are
split into two columns. The
left column text indicates topic
sub-titles and summeries of
text found in the right column.
Right Column: The right
column text provides topic
substance and elaborates on
information from text found in
the left column.
Visual Inspection
Inspecting the monitor
Section Title: The top line of
the page always indicates the
section of the manual. Section
topics may also appear next to
the section title.
M
AINTENANCE
The monitor should be carefully inspected prior to each patient
being admitted to the monitoring system. Follow these guidelines
when inspecting the equipment:
• Carefully inspect the monitor for obvious physical damage to
the outer case, display screen and controls. Do not use the
monitor if physical damage is determined. Refer damaged
equipment to qualified service personnel for repair before
using it again on a patient.
• Inspect all external connectors, front and rear, for degraded
pins, prongs and connector housings. Refer damaged
equipment to qualified service personnel for repair before
using it again on a patient.
• Inspect all cable insulation, cable strain-reliefs and cable
connectors for damage, cracks or degradation. Refer
damaged equipment to qualified service personnel for repair
before using it again on a patient.
Section & Page Number: The
number on the left indicates
the section, the number on the
right indicates the page within
the section.
Product Name - Manual Title:
This is found on each page of
the manual.
Page Revision: As changes to
the manual occur, this letter
indicates the current revision
for each page of the manual.
Page Rev A3-3
Eagle 3000 Patient Monitor
415397-003
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 3
MANUAL REVISIONS
The following list gives the revision of each release of this
manual. As changes to the manual occur, the following list will
provide a reference to these changes.
Tram X00 Modules/Eagle Monitor Termination Instructions for
BP, CO, and TEMP Cables
•Part number: 403799-016
•Describes how to properly terminate the listed patient cable
connectors.
Eagle 3000 Patient Monitor - Version 3B Update Instructions
•Part number - 415397-060
•Describes methods for updating software in the monitor:
EAGLE 3000 PATIENT MONITOR
1 - 4415397-003PAGE REV C
NOTES, CAUTIONSAND W ARNINGS
ABOUT T HIS MANUAL
WHATTHESEINDICATE
Notes, cautions, and warnings all appear in a similar fashion
throughout the manual. These are designed to draw special
attention to particular relevant points of interest.
NOTE
A note conveys special instructions to highlight an
operating procedure, practice, etc. Notes may pre-
cede or follow the applicable text, depending on the
material to be highlighted.
CAUTION
The purpose of a caution is to inform users of this
manual of operating procedures, practices, etc.,
which if not strictly observed, could result in possible
damage to the equipment.
WARNING
A warning provides instructions to users of the
manual that operating procedures, practices, etc., if
not followed, may result in personal injury.
PARTS LISTS
DIMENSIONSPECIFICATIONS
Hardware dimensions in parts lists use either metric or American
standards.
•Metric standards are indicated as items that include the
letter M as a prefix (example: Screw, M 2.0 x 4)
•American standards are indicated as items without a letter
as a prefix (example: Screw, 4-40 x 5/16)
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 5
ABOUT T HIS MANUAL
MANUFACTURER
RESPONSIBILITY
EQUIPMENT SYMBOLS
Marquette Medical Systems is responsible for the effects on
safety, reliability, and performance only if:
•assembly operations, extensions, readjustments,
modifications, or repairs are carried out by persons
authorized by Marquette Medical Systems;
•the electrical installation of the relevant room complies with
the requirements of the appropriate regulations; and,
•the patient monitor is used in accordance with the
instructions for use.
Following are descriptions of the symbols used to identify various
safety concerns for patient monitoring equipment.
NOTE
Some symbols may not appear on all equipment.
ATTENTION: Consult accompanying documents
CAUTION: To reduce the risk of electric shock, do NOT remove
cover. Refer servicing to qualified service personnel.
Defibrillator-proof type CF equipment: type CF equipment is
specifically designed for applications where a conductive
connection directly to the heart is established. The paddles
indicate the equipment is defibrillator proof.
Defibrillator-proof type BF equipment: type BF equipment is
suitable for intentional external and internal application to the
patient, excluding direct cardiac application. Type BF equipment
is type B equipment with an F-type isolated (floating) part. The
paddles indicate the equipment is defibrillator proof.
Fuse
Equipotentiality
Alternating current (AC)
EAGLE 3000 PATIENT MONITOR
1 - 6415397-003PAGE REV C
EQUIPMENT SYMBOLS (CONT)
ABOUT T HIS MANUAL
EQUIPMENTRATINGSFORPATIENT
APPLIED
PA RT S
PRESS
CF
BF
The following describes the symbols for patient applied parts of
patient monitoring equipment.
Power: I = On; O = Off
Indicates where to press to open the door on the 7160 DDW.
ECG and Respiration
Temperature (TEMP)
Invasive Blood Pressure (BP)
Pulse Oximetry (SPO2)
Non-invasive BP (NBP)
End-tidal CO2 (CO2)
NOTE
The rating of protection against electrical shock
(indicated by symbol for Type B, BF or CF) is achieved
only when used with patient applied parts listed for
use in the Supplies section of the Operator’s Manual.
Do not test the dielectric strength directly at the CO2
connector.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 7
HOWTO REACHUS...
ABOUT T HIS MANUAL
The following are telephone numbers and addresses for contacting various Marquette Medical Systems Service and Supplies
Division personnel.
Ordering Supply Items
Ordering Service Parts
Supply items are generally items used during normal operation
of a product. Leadwires, electrode paste, patient cables, and
printer paper are examples of supply items.
•Make telephone inquiries about supply items at:
1-800-558-5102 (U.S. only)
1-407-575-5000 (outside the U.S.)
•Address orders or inquiries to:
Marquette Medical Systems Service and Supplies
P.O. Box 9100
100 Marquette Drive
Jupiter, FL 33468-9100
Attn: Supplies
Service parts are items that are not expended in the normal
operation of the product. They are generally replacements for
defective or malfunctioning items inside the product. Service
parts include PCB assemblies, electronic components, internal
cables and harnesses, software or firmware, and operator and
service manuals. When ordering additional operator manuals,
remember to notate the software version from the start-up
screen.
A part number for the item to be replaced is necessary for ordering a service part. If the part number for the desired item is
unobtainable, the following will be necessary to order the item:
•model and serial number of the equipment,
•part number/name of the assembly where the item is used,
To open a service call with Marquette Medical Systems Service,
contact a Service Dispatcher at:
1-800-558-7044 (U.S. only)
1-407-575-5000 (outside the U.S.)
For any questions about Service Contracts, contact the service
contract operator at:
1-800-552-3248 (U.S. only)
1-407-575-5000 (outside the U.S.)
EAGLE 3000 PATIENT MONITOR
H
OWTO REACHUS... (CONT)
ABOUT T HIS MANUAL
Technical Support
For All Hardware
Telemetry
Series 7000/7010
Monitoring Technical Support has a broad base of information
regarding Marquette patient monitoring equipment and provides
assistance with technical questions and concerns.
For technical information regarding Marquette patient monitoring equipment, contact Monitoring Technical Support at:
1-800-558-7822 (U.S. only)
1-407-575-5000 (outside the U.S.)
For technical information regarding Marquette telemetry patient
monitoring equipment, contact Monitoring Technical Support —
Telemetry Products at:
1-800-552-3243 (U.S. only)
1-407-575-5000 (outside the U.S.)
For technical advice concerning Series 7000/7010 patient monitoring equipment, contact Tech Support:
1-800-443-0980 (U.S. only)
1-407-575-5000 (outside the U.S.)
48-Hour Turnaround Repair
Service Address
For Additional
Information
Some Marquette patient monitoring products or components
such as Series 7000 Input Modules, Tram Modules, Series 7700
ECG Telemetry Transmitters, and CD Telemetry Transmitters are
repaired on a 48-hour turnaround basis.
For information regarding 48-hour repair items, contact Monitoring Technical Support — 48-Hour Repair at:
1-800-552-3243 (U.S. only)
1-407-575-5000 (outside the U.S.)
Send all patient monitoring repair items to:
Marquette Medical Systems Service and Supplies
P.O. Box 9100
100 Marquette Drive
Jupiter, FL 33468-9100
Attn: Monitoring Technical Support — Repair
The service telephone operator can direct calls to personnel most
able to assist with information. For other Monitoring Technical
Support information, contact the service telephone operator at:
1-800-558-5120 (U.S. only)
1-407-575-5000 (outside the U.S.)
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0031 - 9
ABBREVIATIONS
ABOUT T HIS MANUAL
A
AAMI: Association for the
Advancement of Medical
Instrumentation
AC,ac: alternating current
ADC: analog-to-digital
converter
Adj: adjustable
Al: aluminum
Ampl: amplifier
ANSI: American National
Standards Institute, Inc.
ASIC: application specific
integrated circuit
ASYNC COMM: asynchronous
communication
AUI: attachment unit interface
Ave: Avenue
AWG: American Wire Gage
B
B/M: beats per minute
BDGH: binding head
BP: blood pressure
bpm: beats per minute
BT: blood temperature
General monitor block theory .................................... 2-18
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 1
PRODUCT DESCRIPTION
EQUIPMENT OVERVIEW
ABOUTTHEMONITOR
Cost-effective design
Standard monitoring
configuration ...
... with options available
“Value-added” features
Network compatibility
The monitor provides the parameters needed in a single,
inexpensive, easy-to-use device. Its compact size and innovative
package allow it to fit in small, tight places making it an ideal
choice for patient monitoring in an operating room, recovery
room, emergency care area as well as outpatient care areas.
Designed to provide high
reliability, long operational
life, minimal downtime,
and low maintenance cost,
the monitor is a costeffective solution to
general purpose patient
monitoring needs. Monitor
configurations include
simultaneous multi-lead
ECG, non-invasive blood
pressure, dual
temperature, pulse
oximetry monitoring.
Optional features that can
be added to all monitors
include an integral twochannel two inch thermal writer, two invasive blood pressures
and end-tidal CO2.
The monitor is equipped with a 9" diagonal high-contrast
electroluminescent (EL) display capable of showing four
waveforms and full digital data for all monitored parameters.
The Trim Knob control and a clinically logical software menu
structure assure ease-of-use and reduced inservice time. Multilead ST-segment monitoring, high quality graphic and tabular
trends, and sophisticated algorithms (which may assist to reduce
false alarms and provide more accurate data) are "value-added"
features. In addition, the monitor is a member of a complete
family of critical care products from the manufacturer, and is
fully compatible with interconnection to the patient monitoring
network.
EAGLE 3000 PATIENT MONITOR
2 - 2415397-003PAGE REV C
PRODUCT DESCRIPTION (CONT)
FRONTPANELDESCRIPTION
NOTE
To insure patient safety, use only parts and accessories manufactured or recommended
by the manufacturer. Parts and accessories used must meet the requirements of the
applicable IEC 601 series safety standards, and/or the system configuration must meet
the requirements of the IEC 601-1-1 medical electrical systems standard.
EQUIPMENT OVERVIEW
Patient Input Connectors: Used to
attach patient cables for various
electrodes, sensors and transducers used
in patient signal acquisition
Display:
Monochrome electroluminescent (EL) display panel.
Screen size: 9-inch diagonal
Resolution: 640 x 400 pixels
Trim Knob Control: This is the
control that is used most often to
choose menu items and enter
data.
Rotate the Trim Knob
control to highlight an
item on the display.
Press the Trim Knob
control to select the
highlighted item.
Front Panel Controls: Five backlit
pushbutton operator controls provide
the following functions:
DEFIB SYNC (option): This side panel
jack provides a direct interface between
the monitor and a defibrillator for
synchonization of the two devices during
emergency defibrillation or for
synchronized cardioversion. The signals
available through this connector are:
Outputs —
•Defib sync pulse
•Analog ECG signal
•Analog invasive BP signal
Input —
•Defibrillator triggered marker pulse
RMT ALM (option): This side panel
connector provides interconnection
through the Marquette/Hellige Isolation
Relay to a nurse-call light system.
CO2 Sensor (option): These side panel
connectors provide a direct interface
between the monitor and an End-tidal
CO2 monitoring kit.
NBP Connector: A
pneumatic connector
for attaching a
noninvasive blood
pressure cuff to the
monitor.
Controls power to the
display without disrupting
monitor main power.
Manually starts or stops the
noninvasive blood pressure
function.
Sets zero references for all
invasive blood pressure
functions.
Controls patient alarm
silencing functions.
Manually starts or stops
graphs to selected writers.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 3
PRODUCT DESCRIPTION (CONT)
REARPANELDESCRIPTION
NOTE
The use of ACCESSORY equipment not complying with the equivalent safety requirements
of this equipment may lead to a reduced level of safety of the resulting system.
Consideration relating to the choice shall include: use of the accessory in the PATIENT
VICINITY; and evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate IEC 601-1 and/or IEC 601-1-1 harmonized
national standard.
EQUIPMENT OVERVIEW
Main Power Switch:
When in the on (I)
position, ac power is
applied to the
monitor. When in the
off (0) position, ac
power is removed
from the monitor.
RMT ALARM (remote
alarm) Connector:
Interconnection to the
Marquette/Hellige
Isolation Relay for a
nurse-call light
system.
*Serial Number Label:
(not shown) Describes
the type of equipment,
date and sequence of
product manufacture
for each monitor.
Main Power: A power
cord connected
between this port and
a wall receptacle is
used to provide ac
main power to the
monitor.
* A sample of the information found on the
Serial Number Label is shown below.
F 7 XX 0109 G
ETHERNET Connector:
An IEEE 802.3 AUI cable
and transceiver can be
connected to this port for
monitors used in patient
monitoring network
configurations.
ASYNC COMM
Connector: This
port can be used for
interconnection to a
remote control,
direct digital writer
or other device.
Month
YearProduct
code
A thru M = January thru December
Letter I is not used.
Product
sequence
number
Division identifier
F = Cardiology
G = Monitoring
EAGLE 3000 PATIENT MONITOR
2 - 4415397-003PAGE REV C
PRODUCT DESCRIPTION (CONT)
EQUIPMENT OVERVIEW
ABOUTTHEREMOTEALARM
CONNECTOR
The remote alarm (RMT ALRM) connector is for use with a
Marquette/Hellige Isolation Relay (pn 303 555 77) which
provides a relay closure during the following alarms:
•CRISIS Patient Status Alarms, and
•WARNING System Status Alarms.
The signals provided at the RMT ALRM connector activate and
deactivate the Marquette/Hellige Isolation Relay (shown below).
When the monitor is powered up initially or rebooted, the relay
remains turned off until the monitor is finished with its power up
or reboot sequence. Once the monitor finishes powering up or
rebooting, the relay is then energized.
When a CRISIS Patient Status Alarm or a WARNING System
Status Alarm is detected, the relay is turned off by a signal from
the monitor. When the alarm has been cleared, the relay is
turned on once again.
The relay is turned off when AC power is removed from the
monitor.
• Marquette/Hellige Isolation Relay •
NOTE:
Refer to the Installation Instructions that come with
the Marquette/Hellige Isolation Relay for proper
connection to your Nurse Call system.
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 5
FOR Y OUR NOTES
EQUIPMENT OVERVIEW
EAGLE 3000 PATIENT MONITOR
2 - 6415397-003PAGE REV C
MONITOR APPLICATIONS
EQUIPMENT OVERVIEW
STAND-ALONEMONITOR
APPLICATION
PATIENTMONITORINGSYSTEM
APPLICATION
HOSPITAL-WIDENETWORK
APPLICATION
The Marquette Unity Network (hereafter referred to as the
network) provides a method for standardized communication
with various Marquette medical system devices. This versatile
monitor can operate both as a fully functional stand-alone device
and as a component on the network, depending upon the
application.
When connected to the network, the monitor provides access to
other devices for many purposes. Marquette patient monitoring
equipment such as Centralscope central station monitor; Series
7200/7260 direct digital writer; CDT-LAN patient telemetry
system; ADU-LAN; and, Solar or other Eagle patient monitors are
examples of devices that can be used in conjunction with the
monitor when connected to the network.
There are various types of information management and data
base systems devices which may also be integrated with the
monitor via connection to the network. Marquette medical
systems equipment such as MUSE cardiology management
system; MARS UNITY workstation; MARS 24 clinical review
station; MRT II automated vital sign and arrhythmia data
collection system; MAC-Lab cardiac catheterization system; QMI
patient data management system; and, MUSE HIS interface are
examples of systems and data bases which can be integrated
with the monitor on the network.
Patient monitoring system
application
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 7
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS
Display
Size:9-inch diagonal
Type:Monochrome, electroluminescent, flat panel
Resolution:640 by 400 pixels
Number of waveform traces:4
Length of trace:5.7 seconds
Sweep speed:22.9 mm/s (with erase bar), all waveforms except CO2
Waveform display options:Full or individual
Information window:Displays all non-real-time information without obstructing the
display of real-time information
Display organization:Prioritized by parameter
Processing
Main processor:MC68EN360 (32-bit microprocessor), 23.4936 MHz clock
frequency
Data acquisition processor:MC68332 (32-bit microprocessor), 15.7248 MHz clock frequency
Program memory:2MB flash EEPROM
Data memory:2MB RAM
Alarms
Classification:4 levels (Crisis, Warning, Advisory, Message)
Notification:Audible and visual
Setting:Default and individual
Silencing:1 minute, current alarm only
Pause
Adult mode:5 minutes
Neonatal ICU mode:3 minutes
OR mode5 minute, 15 minute, or permanent
Volume:Default 70%, 70 dB measured at 1 meter
User Interface:
Trim Knob control:Provides access to all menu-based operations
Hard keys
Silence Alarm:Controls alarm silencing
Graph Go/Stop:Start and stop manual graph operation
NBP Go/Stop:Start and stop non-invasive blood pressure measurement
Zero All:Zero BP transducers
Display On/Off:Blanks the display and disables alarms, for use as screen saver
EAGLE 3000 PATIENT MONITOR
2 - 8415397-003PAGE REV C
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
ECG
Standard leads available:I, II, III, V, aVR, aVL, and aVF
Leads analyzed simultaneously: I, II, III, and V (multi-lead mode)
Lead fail:Identifies failed electrodes
Lead fail sensing current
Active electrodes:<25 nA each
Reference electrode:<100 nA
Waveform display aspect ratio:0.34 s/mV
Input specifications
Voltage range:±0.5 mV to ±5 mV
Signal width:40 ms to 120 ms (Q to S)
Input impedance
Common mode:>10 MΩ at 50/60 Hz
Differential:>2.5 MΩ from dc to 60 Hz
Output specifications
Display frequency response
Adult ICU mode:0.05 to 40 Hz
Neonatal Mode:0.5 to 40 Hz
OR Mode:0.05 to 25 Hz
Paper Recorder frequency response
Standard Mode:0.05 to 100 Hz
Neonatal:0.5 to 40 Hz
OR Mode:0.05 to 25 Hz
Common mode rejection:90 dB minimum at 60 Hz
Gain:1000 ±3%
Linearity deviation:±3% (maximum)
Noise:<30 µV (referred to input)
Heart Rate
Heart rate range:30 to 300 beats per minute
Heart rate averaging:8 beats
Display update interval:2 seconds
Response time:<6 seconds (per AAMI EC13)
Limit alarm display:<10 seconds after alarm condition exceeded
ST Segment analysis
Measurement description:ST segment deviation measured and displayed for all acquired
leads, and averaged for anterior, lateral and inferior lead groups
ST display:Three 30-minute trends, or ECG complexes for leads I, II, and V,
and a summation trend
Measurement point:Measures at 60 ms following the J point
Measurement range
Adult:-12.0 mm to + 12.0 mm
Neonatal:-10.0 mm to + 10.0 mm
Display resolution:0.1 mm
ST measurement averaging: 16 beats
Display update interval:2 seconds
Pacemaker detection/rejection
Input voltage range:±2 mV to ±700 mV
Input pulse width:0.1 ms to 2 ms
Rise time:10 µs to 100 µs
Over/under shoot:2 mV (maximum)
Baseline drift:<0.5 V with a ±700-mV, 2-ms pacemaker pulse applied
Alarms:Selectable upper and lower heart rate limits, arrhythmia
detection; lead-failure
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 9
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
Invasive Blood Pressure (optional)
Number of channels:2
Transducer sites:Arterial, pulmonary arterial, central venous, left atrial,
intracranial, and special
Transducer requirements
Excitation voltage:±2.5 V dc ±0.1%
Transducer output:50 µV/V/cm Hg
Pressure measurement specifications
Range:-25 mmHg to 300 mmHg
Accuracy:±2% or ±1 mmHg, whichever is greater (independent of
transducer)
Drift:< ±0.1%/°C, and < ±0.1% over and 24-hour period
Zero balance range:±150 mmHg
Zero balance accuracy:±1 mmHg
Zero balance drift:±1 mmHg over 24 hours
Frequency response:dc to 50 Hz (-3 dB)
Input impedance
Common mode:100 KΩ (minimum) at 50/60 Hz
Differential:100 KΩ (minimum) from dc to 60 Hz
Common mode rejection:60 dB (minimum) at 60 Hz
Noise:5 mV peak to peak (maximum) from dc to 30 Hz
Averaging:4 seconds
Display update interval:2 seconds
Alarms:Selectable upper and lower limits for systolic, diastolic, and mean
pressures; sensor failure
Non-invasive Blood Pressure
Measurement technique:Oscillometric
Displayed parameters:Systolic, diastolic, and mean pressures, pulse rate, time of last
measurement
Measurement modes:Manual, auto, and stat
Systolic pressure range:30 to 275 mmHg
Diastolic pressure range:15 to 250 mmHg
Heart rate range:30 to 300 beats per minute
Total cycle time:20 to 40 seconds typical (dependent on heart rate and motion
Overpressure safety valve:Activates when cuff pressure exceeds:
Adult mode:300 mmHg (+30/-0 mmHg) cuff pressure
Neonatal mode:150 mmHg (+15/-0 mmHg) cuff pressure
Maximum pressure leakage:4 mmHg per minute
Automatic cycle times:0 to 24 hours
Auto zero:Zero pressure reference prior to each cuff inflation
Tubing length:12 feet adult, 8 feet neonatal
Automatic cuff deflation:Occurs when power is off or the following limits are exceeded:
Adult mode:300 mmHg (± 3 mmHg) cuff pressure or 3 minutes cycle time
Neonatal mode:150 mmHg (± 2 mmHg) cuff pressure or 90 seconds cycle time
Cuff sizes:
Disposable:Large adult, adult, small adult, pediatric, small pediatric, and
infant
Reusable:Thigh, large adult, adult, child, and infant
Alarms:Selectable upper and lower limits for systolic, diastolic, and mean
Range:0 to 200 breaths per minute (for variation of 1.0 to 10Ω)
Accuracy:± 1 breath per minute
Base impedance:100 to 1000Ω
Detection sensitivity:0.4 to 10 Ω variation (for 0 to 120 breaths per minute)
Excitation current:250 µA
Averaging:8 breaths
at 52.6 kHz
RMS
Display update interval:2 seconds
Waveform display bandwidth:0.1 to 1.8 Hz (-3 dB)
Lead fail:Indicated when base impedance exceeds 1750 ± 250Ω
Apnea detection:Indicated when impedance variation is less than selected
sensitivity or 0.2Ω, whichever is greater
Alarms:Selectable upper and lower respiration rate limits, and user
selectable apnea limit
EAGLE 3000 PATIENT MONITOR
PAGE REV C415397-0032 - 11
EQUIPMENT OVERVIEW
PERFORMANCE SPECIFICATIONS (CONT)
End-tidal C02 (optional)
Monitoring functions:Inspired and expired CO2 measurements, respiration rate
Adult reusable:<5 cc
Adult disposable:<5 cc
Neonatal:<0.5 cc
Sampling:<0.2 cc
CO2 measurement specifications
Range:0 to 100 mmHg
Accuracy
Mainstream:±2 mmHg or 5%, whichever is greater
Sidestream mode:±3 mmHg or 8%, whichever is greater
Display update interval:2 seconds
CO2 averaging:Selectable from single breath, 10 seconds, or 20 seconds
Resolution:1 mmHg
Response time (for 5% step size)
Mainstream adult:<60ms (10 to 90%)
Mainstream neonate:<50 ms (10 to 90%)
Sidestream (with filter):<600ms (10% to 90%) at 180 cc/min
Interference
N2O gas:±2 mmHg or 5%, whichever is greater, with N2O compensation
enabled
O2 gas:±2 mmHg or 5%, whichever is greater, with O2 compensation
enabled
Barometric pressure:±2 mmHg (maximum) from 500 to 800
Water vapor:±0.5 mmHg or 1.5% (maximum), whichever is greater
Anesthetic agent:±0.5 mmHg (maximum), for concentrations of no more than 5% of
halogenated agents
Warm-up time:Less than 15 seconds to initial CO2 indication, full specification
within 60 seconds
Calibration
Factory setting:Factory calibration settings stored in nonvolatile memory within
the sensor. 15 second adapter calibration when switching airway
types
Verification:Zero and reference performance check with on-cable verifier
CO2 Waveform sweep speed: Selectable one-fourth, one-half, and full speed
Respiration rate specifications
Range: (for 5% step size)
Mainstream mode:0 to 120 breaths per minute
Sidestream mode:0 to 50 breaths per minute
Accuracy:±1 breath per minute
Resolution:±1 breath per minute
Sidestream pump specifications
Flow rate:190 ±10 cc/min
Capnostat III sensor specifications
Operating temperature:10 to 40°C
Storage temperature:-30 to 65°C (-22 to 149°F)
Humidity:5 to 95%, relative humidity, non-condensing
Barometric pressure:500 to 800 mmHg
Shock resistance:Able to withstand 6 ft. drop to tile floor
Moisture resistance:Splash resistant sealed transducer
Cleaning and sterilization
Sensor:Transducer, cable, and verifier may be wiped with cold chemical
disinfectant: no steam sterilization, no ETO gas permitted, do not
immerse in fluid
Reusable airway
adapters:Disinfect with buffered glutaraldehyde, ETO gas, isopropyl
alcohol, household bleach; also steam sterilization and
pasteurizable
Alarms:Selectable upper and lower limits for CO2 and respiration rate
EAGLE 3000 PATIENT MONITOR
2 - 12415397-003PAGE REV C
Loading...
+ 120 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.