Marquette Dash 3000, Dash 4000 Service Manual

Dash3000/4000
Patient Monitor
Service Manual
2000966-035 Revision A
Listed below are GE Medical Systems Information Technologies’ trademarks. All other trademarks contained
herein are the property of their respective owners.
900 SC, ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, C Qwik Connect, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROLAN, COROMETRICS, Corometrics Sensor Tip, CRG PLUS, DASH, Digistore, Digital DATAQ, E for M, EAGLE, Event-Link, FMS 101B, FMS 111, HELLIGE, IMAGE STORE, INTELLIMOTION, IQA, LASER SXP, MAC, MAC-LAB, MACTRODE, MANAGED USE, MARQUETTE, MARQUETTE MAC, MARQUETTE MEDICAL SYSTEMS, MARQUETTE UNITY NETWORK, MARS, MAX, MEDITEL, MEI, MEI in the circle logo, MEMOPORT, MEMOPORT C, MINISTORE, MINNOWS, Monarch 8000, MULTI-LINK, MULTISCRIPTOR, MUSE, MUSE CV, Neo-Trak, NEUROSCRIPT, OnlineABG, OXYMONITOR, Pres-R-Cuff, PRESSURE-SCRIBE, QMI, QS, Quantitative Medicine, Quantitative Sentinel, RAC RAMS, RSVP, SAM, SEER, SILVERTRACE, SOLAR, SOLARVIEW, Spectra 400, Spectra-Overview, Spectra-Tel, ST GUARD, TRAM, TRAM-NET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, Trimline, UNION STATION, UNITY logo, UNITY NETWORK, Vari-X, Vari-X Cardiomatic, VariCath, VARIDEX, VAS, and Vision Care Filter are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
12SL, 15SL, Access, AccuSpeak, ADVANTAGE, BAM, BODYTRODE, Cardiomatic, CardioSpeak, CD TELEMETRY
Cumulus, Event-Link Nimbus, HI-RES, ICMMS, IMAGE VAULT, IMPACT.wf, INTER-LEAD, IQA, LIFEWATCH, Managed Use, MARQUETTE PRISM, MARQUETTE MMS, MRT, MUSE CardioWindow, NST PRO, NAUTILUS, O
®
-LAN, CENTRALSCOPE, Corolation, EDIC, EK-Pro, Event-Link Cirrus, Event-Link
®
RESPONDER, MENTOR, MicroSmart,
SENSOR, Octanet, OMRS, PHi-Res,
2
Premium, Prism, QUIK CONNECT V, QUICK CONNECT, QT Guard, SMART-PAC, SMARTLOOK, Spiral Lok, Sweetheart, UNITY, Universal, Waterfall, and Walkmom are trademarks of GE Medical Systems Information Technologies.
© GE Medical Systems Information Technologies, 2000. All rights reserved.
T-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035 27 November 2000

CONTENTS

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2 EQUIPMENT OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
The Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
The Patient Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Optional RAC 2A Module Housing . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Optional Wireless LAN System . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Invasive Blood Pressure (BP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Noninvasive Blood Pressure (NBP) . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Pulse Oximetry (SPO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Cardiac Output (CO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Temperature (TEMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Carbon Dioxide (CO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Defibrillator Synchronization Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Paper Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
RF Wireless LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19
Revision A Dash 3000/4000 Patient Monitor i
2000966-035
CONTENTS:
3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Back Panel Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Front Panel Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Ethernet Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Twisted Pair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Concentrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Node . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Segment and Branch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Repeater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Twisted Pair Cabling (10BaseT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Symbol PC Card (Wireless LAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
4 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Manufacturer Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Manufacturer Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Cleaning Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Cleaning the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning the Print Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Battery Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Conditioning the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Wall Receptacle Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Wire Leakage Current Tests . . . . . . . . . . . . . . . . . . 4-13
Enclosure Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Patient (Source) Leakage Current Test . . . . . . . . . . . . . . . . . . . . . .4-17
Patient (Sink) Leakage Current Test
(Mains Voltage on the Applied Part) . . . . . . . . . . . . . . . . . . . . . . .4-19
Test Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-20
Hi-Pot (Dielectric Withstand) Test . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
DAS Assembly AC Hi-Pot Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Processor/Power Management PCB Hi-Pot Test . . . . . . . . . . . . . .4-23
AC Mains Hi-Pot Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Checkout Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-25
Manufacturer Recommended Test Equipment . . . . . . . . . . . . . . . . . 4-25
Monitor Power-up Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
ECG Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-27
ii Dash 3000/4000 Patient Monitor Revision A
2000966-035
CONTENTS:
Respiration Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Temperature Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-31
Cardiac Output Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Invasive Blood Pressure Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-34
Pulse Oximetry Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-36
Noninvasive Blood Pressure Tests . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
Analog Output and Defibrillator Synchronization Tests . . . . . . . . . . 4-40
Battery Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
Graph Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
Graph Speed Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
Display Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
Speaker Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
Network Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
RF LAN Test (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
RAC 2A Module Housing Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-45
PM Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Repair Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-46
5 TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Service Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Boot Loader Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Main Menu Service Mode Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Review Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Error Log Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Battery Alarms and Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Power Source Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Wall Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Power Cord and Plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Data Acquisition Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
ECG Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Lead Fail Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Pace Detect Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Invasive Blood Pressure Functions . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Respiration Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Noninvasive Blood Pressure Functions . . . . . . . . . . . . . . . . . . . . . . 5-21
Wireless LAN Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-22
Service Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Fault/Symptom Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Acquisition PCB Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
Processor PCB Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-25
Troubleshooting Software Updates - Problems and Solutions . . . . . . . .5-26
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-29
Revision A Dash 3000/4000 Patient Monitor iii
2000966-035
CONTENTS:
6 CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Loading Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Software Loading/Updating Methods . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Software Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Monitor Software Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Maintain Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Problems While Loading Software . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Record Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Load Software From Diskette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Connect the PC to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Software Diskettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
Update Program Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Setup Monitor To Accept Download Files . . . . . . . . . . . . . . . . . . . .6-11
Download Files to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-12
Load Software Over The Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
Network Update Diskettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Copy Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Download Files to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
Complete the Software Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Activate Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Configuring a Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Main Menu Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Set Unit Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Set Bed Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Patient-Monitor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Set Graph Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
Admit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-24
Boot Code Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Set Defib Sync Voltage and Pulse Width . . . . . . . . . . . . . . . . . . . . .6-25
Set Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-25
Set CIC and QS Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Set MUSE Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Transcutaneous Pace Blank Length . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Set Country Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Wireless LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-27
Advanced User Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-28
Set Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Change Software Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-29
Change Ethernet Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-30
Review Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-30
Transferring Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
Access the COPY LOGS Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
Select the Care Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-35
Select the Monitoring Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
iv Dash 3000/4000 Patient Monitor Revision A
2000966-035
CONTENTS:
Select the Error Log Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
Copy Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
Eject Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
7 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Hardware Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
NBP, ECG, BP, and End-tidal CO2 Software Calibration . . . . . . . . . . . . . 7-4
NBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
ECG or BP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
End-tidal CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
8 FIELD REPLACEABLE UNITS AND UPGRADES . . . . . . . . . . . . . . . . . . 8-1
Disassembly Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Before Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Hardware Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
PCB Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
After Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Handle Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Removing the Handle Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Replacing or Upgrading the Dash 3000 Alarm Light Option . . . . . . .8-8
Display Assembly Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Removing the Display Assembly from the Main Unit . . . . . . . . . . . . . 8-9
Replacing the Backlight Inverter PCB . . . . . . . . . . . . . . . . . . . . . . .8-13
Replacing the Key Pad Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Replacing the LCD Color Display . . . . . . . . . . . . . . . . . . . . . . . . . . .8-15
Replacing or Upgrading the Dash 4000 Alarm Light Option . . . . . .8-16
Replacing the Dash 4000 Front Panel PCB . . . . . . . . . . . . . . . . . . .8-17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-17
Main Unit Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
DAS and NBP Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-18
Main and/or Power Supply Assemblies, Speaker or
RF LAN Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
Processor/Power Management PCB and Battery Assembly . . . . . .8-22
Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-24
Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-25
RF LAN Upgrade Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
Optional DDW Writer Replacement/Upgrade . . . . . . . . . . . . . . . . . . . . .8-36
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-36
Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
Revision A Dash 3000/4000 Patient Monitor v
2000966-035
CONTENTS:
9 ASSEMBLY DRAWINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Theory Of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
General Monitor Block Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Overall Monitor Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Data Acquisition System (DAS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Processor/Power Management Subsystem . . . . . . . . . . . . . . . . . . . 9-13
Optional Thermal Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Handle Subassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
Setup and Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-21
Electrical Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Dash 3000 Exploded Views, PN 420000-xxx . . . . . . . . . . . . . . . . . . . . . 9-23
Dash 3000 Display Bezel Assembly, PN 419031-001 . . . . . . . . . . . . . .9-28
Dash 3000 Display Bezel Parts List, PN 419031-001 . . . . . . . . . . . . . . 9-29
Dash 3000 Display Bezel Part List , PN 419031-003 . . . . . . . . . . . . . . . 9-31
Dash 4000 Parts List PN 2004323-xxx . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34
Dash 4000 Display Assembly, PN 2004272-001 . . . . . . . . . . . . . . . . . . 9-36
Dash 4000 Display Bezel Parts List, PN 2004272-001 . . . . . . . . . . . . . 9-37
Field Replaceable Units (FRU’s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Dash 3000 FRUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Dash 4000 FRUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Port Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-40
Invasive Blood Pressure Cable Connector . . . . . . . . . . . . . . . . . . . . 9-40
Pulse Oximetry (SpO2) Cable Connector . . . . . . . . . . . . . . . . . . . . 9-41
Temperature/CO Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . 9-41
Capnostat III (CO2) Cable Connector . . . . . . . . . . . . . . . . . . . . . . . 9-42
NBP Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
ECG Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
Input Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-44
Network Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-44
Auxiliary Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-45
Defib Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-45
Peripheral Expansion Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-46
vi Dash 3000/4000 Patient Monitor Revision A
2000966-035

1 INTRODUCTION

Revision A Dash 3000/4000 Patient Monitor 1-1
2000966-035
For your notes
INTRODUCTION:
1-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Manual Information

Manual Information

Revision History

Revision Date Comment
A 27 November 2000 Initial release of this manual.

Manual Purpose

Intended Audience

Each page of this manual has the document part number and revision letter at the bottom of the page. The revision letter identifies the
document’s update level. The revision history of this document is summarized below.
Revision History
This manual supplies technical information for service representatives and technical personnel so they can maintain the equipment to the assembly level. Use it as a guide for maintenance and electrical repairs considered field repairable. Where necessary the manual identifies additional sources of relevant information and technical assistance.
See the operator’s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use.
This manual is intended for service representatives and technical personnel who maintain, troubleshoot, or repair this equipment.
Revision A Dash 3000/4000 Patient Monitor 1-3
2000966-035

Safety Information

INTRODUCTION: Safety Information

Responsibility of the Manufacturer

General

GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE Medical Systems Information Technologies.
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
This device is intended for use under the direct supervision of a licensed health care practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a physician.
Contact GE Medical Systems Information Technologies for information before connecting any devices to the equipment that are not recommended in this manual.
Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1-1 medical electrical systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all functions.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
use of the accessory in the PATIENT VICINITY; and
evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.
If the installation of the equipment, in the USA, will use 240V rather than 120V, the source must be a center-tapped, 240V, single-phase circuit.
1-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Safety Information

Warnings, Cautions, and Notes

The terms danger, warning, and caution are used throughout this manual to point out hazards and to designate a degree or level or seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person.
DANGER indicates an imminent hazard which, if not avoided, will result in death or serious injury.
WARNING indicates a potential hazard or unsafe practice which, if not avoided, could result in death or serious injury.
CAUTION indicates a potential hazard or unsafe practice which, if not avoided, could result in minor personal injury or product/property damage.
NOTE provides application tips or other useful information to assure that you get the most from your equipment.
Revision A Dash 3000/4000 Patient Monitor 1-5
2000966-035
INTRODUCTION: Safety Information

Equipment Symbols

Some of the following symbols appear on the equipment.
ATTENTION: Consult accompanying documents before using the equipment.
In Europe, this symbol means dangerous or high voltage. In the United States, this symbol represents the caution notice below:
To reduce the risk of electric shock, do NOT remove cover (or back). Refer servicing to qualified personnel.
Defibrillator-proof type CF equipment; type CF equipment is specifically designed for applications where a conductive connection directly to the heart is established. The paddles indicate the equipment is defibrillator proof.
Defibrillator-proof type BF equipment; type BF equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application. Type BF equipment is type B equipment with an F-type isolated (floating) part. The paddles indicate the equipment is defibrillator proof.
Type B equipment; type B equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application.
Equipotentiality
Alternating current (AC)
Power;
Fuse
Battery
Indicates the Ethernet connection for the monitor.
I = ON; O= OFF
1-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Safety Information
PRESS
814A
815A
816A
817A
818A
Press to open.
Power
Graph Go/Stop
NBP Go/Stop
Function
Silence Alarm
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No. 601.1, IEC 60601-1, and, if required, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601­1-1.
Revision A Dash 3000/4000 Patient Monitor 1-7
2000966-035
INTRODUCTION: Service Information

Service Information

Service Requirements

Equipment Identification

Follow the service requirements listed below.
Refer equipment servicing to GE Medical Systems Information Technologies’ authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that warranty.
It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.
Every GE Medical Systems Information Technologies device has a unique serial number for identification. A sample of the information found on a serial number label is shown below.
D 0 XX 0005 G XX
Month Manufactured
A = January B = February C = March D = April E = May F = June G = July H = August J = September K = October L = November M = December
Year Manufactured
0 = 2000 1 = 2001 2 = 2002
(and so on)
Product Code
Two-character product descriptor
Product Sequence Number
Manufacturing number (of total units manufactured)
Division
F = Cardiology G = Monitoring
Device Characteristics
One or two letters that further describe the unit, for example: P = prototype not conforming to marketing specification R = refurbished equipment S = special product documented under Specials part numbers U = upgraded unit
1-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035

2 EQUIPMENT OVERVIEW

Revision A Dash 3000/4000 Patient Monitor 2-1
2000966-035
For your notes
EQUIPMENT OVERVI EW:
2-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Components

EQUIPMENT OVERVIEW: Components

The Monitoring System

The Patient Monitor

The Dash 3000/4000 patient monitor can function by itself with a built-in writer, or it can be cabled in with the Unity Network via Ethernet. Optional components are, if using Wireless LAN or cabled to Ethernet, a Centralscope central station and the Clinical Information Center.
This device is designed to monitor a fixed set of parameters including ECG, noninvasive blood pressure, impedance respiration, SpO2, and temperature. Invasive pressure and EtCO2 are optional features. Additional specialized features include cardiac output, cardiac calculations, pulmonary calculations, dose calculations, PA wedge (PA wedge is only available with the invasive pressure option), and SAM module interface.
001A
Dash 3000 Monitor
Revision A Dash 3000/4000 Patient Monitor 2-3
2000966-035
Dash 4000 Monitor
051A
EQUIPMENT OVERVIEW: Components

Right Side View

Left Side View

All of the patient cable connectors are located on the right side of the monitor. A Trim Knob control provides single control operation of virtually all monitor functions.
Patient Cable Connectors
002A
On the left of the monitor, you can find the built-in writer and the battery compartment.
Optional Built-in Writer— The built-in, 4 channel writer is located in the center of the left side of the monitor .
Battery Compartment— The battery packs are located in this compartment.
003A
2-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Components

Back Vie w

Network Connector—A cable can be connected to this port for monitors used in patient monitoring network configurations.
Equipotential Terminal—For
measurements in or near the heart we recommend connecting the monitor to the potential equalization system. U se the green and yellow potential equalization cable and connect it to this pin.
AC Power Connector
On the back of the monitor you will find all connectors for equipment and network.
Line V oltage Selector—This s elector is factory set to match the line voltage and frequency rating for your country.
Audible Alarm Enunciator—The internal speaker provides sound for audible alarms. For better sound quality do not block speaker.
Aux Port—Used for RAC 2A and software updates.
Defib Sync Connector—Provides ECG analog output signals to
004A
Peripheral Expansion Port
user-supplied equipment. A 5­volt, 2-millisecond artificial pacer spike is added to the analog output when PACE is on and detection occurs.

Optional Alarm Indicator

An optional alarm indicator can be built into the handle of the Dash 3000 monitor or into the display bezel of the Dash 4000 monitor. When activated, the LED indicator flashes red for CRISIS and WARNING patient status alarms and yellow for all other alarms.
Alarm Indicator
Dash 3000 Monitor
Revision A Dash 3000/4000 Patient Monitor 2-5
2000966-035
Dash 4000 Monitor
052A
EQUIPMENT OVERVIEW: Components

Optional RAC 2A Module Housing

The RAC 2A module housing currently supports the SAM module.
006A
An integral power supply is used to run the RAC 2A and support the needed voltages.
2-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Components

Optional Wireless LAN System

Access Points

The flexibility of the GE Medical Systems Information Technologies Unity Network is increased by using the Wireless LAN system. The Wireless LAN system allows the user to roam from one access point to another, maintaining a strong, seamless connection to the Unity network.
The monitor, with its optional built-in Wireless LAN, functionally performs the same as a monitor connected directly to the Unity network. It can be viewed at the central station and by other GE Medical Systems
Information Technologies’ monitors on the network (i.e. Dash 3000/4000, Eagle 4000, and Solar patient monitors). Monitors with Wireless LAN sends and receives patient data via the access points of the Unity network.
NOTE: Wireless patient monitors that are moved from room to room
must have the monitor type configured as Rover or Rover/Combo monitoring.
To integrate the wireless network with the wired network, one or more access points are necessary. An access point connects the wireless monitor to the wired network infrastructure within the building, and acts as a bridge between the wired and wireless networks. The areas covered by each access point overlap to insure continuous coverage.
NOTE: The monitor will only work with a Symbol Access Point. The
monitor will not communicate directly with a Wireless LAN device from Aironet.
050A
Revision A Dash 3000/4000 Patient Monitor 2-7
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications

Technical Specifications

Due to continual product innovation, specifications are subject to change without notice. The following specifications are accurate as of the date of this publication, and pertain to the Dash 3000/4000 Patient Monitor.

Performance Specifications

Display

Size:
Type:
Color: Active-Matrix Liquid Crystal Display (LCD)
Resolution: 640 by 480 pixels
Dash 3000: Dash 4000:
8.4-inch diagonal
10.4-inch diagonal

Controls

Alarms

Number of traces: 6 (maximum) Number of seconds/trace:
Dash 3000: Dash 4000:
Sweep speed:
All waveforms 6.25, 12.5 or 25 mm/sec (with erase bar)
Waveform display options: Individual 6 waveforms, individual 3 waveforms, full,
Information window: Displays non-real-time information without obstructing
Display organization: Prioritized by parameter
Standard: Trim Knob control plus 5 hard keys: Power, NBP Go/
Categories: Patient Status and System Status
4.9 at 25 mm/sec
5.9 at 25 mm/sec
and full grid modes
the display of real-time information
Stop, Function, Silence Alarm, and Graph Go/Stop
Priorities: 4 levels — Crisis, Warning, Advisory, and Message
Notification: Audible and visual Setting: Default and individual Silencing: 1 minute, current alarm only
2-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035
ECG
EQUIPMENT OVERVIEW: Technical Specifications
Pause: 5 minutes (adult); 3 minutes (neonatal); 5, 15 minutes,
permanent (OR mode)
Volume: Default 70%, 70 dB measured at 1 meter
5 Leadwire cable: I, II, III, V, aVR, aVL, and aVF 10 Leadwire cable (12SL option): V2, V3, V4, V5 and V6 Leads analyzed simultaneously: I, II, III, and V (multi-lead mode) Lead fail: Identifies failed lead Alarms: User-selectable upper and lower heart rate limits Input specifications:
Voltage range: Signal width: Heart rate range: Accuracy: Input impedance:
Common mode: Differential:
Common mode rejection:
±0.5 mV to ±5 mV 40 ms to 120 ms (Q to S) 30 to 300 BPM ±1% or ±1 BPM, whichever is greater
>10 M
at 50/60 Hz
>2.5 M
from dc to 60 Hz
90dB minimum at 50 Hz or 60 Hz
Output specifications:
Frequency response: Display:
Diagnostic: Monitoring: Moderate: Maximum:
Paper Recorder:
Diagnostic: Monitoring: Moderate: Maximum:
Linearity deviation: Noise:
ST segment measurement:
Measurement point: Measurement range:
Measurement accuracy:
Pacemaker detection/rejection:
Input voltage range: Input pulse width: Rise time: Over/under shoot: Baseline drift:
0.05 to 40 Hz
0.05 to 40 Hz
0.05 to 25 Hz 5 to 25 Hz
0.05 to 100 Hz
0.05 to 40 Hz
0.05 to 25 Hz 5 to 25 Hz ±3% (maximum) <30 µV (referred to input)
Adjustable from 0 to 120 ms past the J-point (default: 60 ms adult, 30 ms neonatal) –12.0 to +12.0 mm ±10% or 0.5 mm, whichever is greater
±2 mV to ±700 mV
0.1 ms to 2 ms 10 µs to 100 µs 2 mV (max) <0.5 mV/hour with a ±700-mV, 2-ms pacemaker pulse applied
Revision A Dash 3000/4000 Patient Monitor 2-9
2000966-035

Invasive Blood Pressure (BP)

EQUIPMENT OVERVIEW: Technical Specifications
Number of channels: 2 Transducer sites: Arterial (ART), femoral artery (FEM), pulmonary artery
(PA), central venous (CVP), right atrial (RA), left atrial (LA), intracranial (ICP), and special (SP)
In neonatal mode: umbilical artery catheter (UAC) and umbilical venous catheter (UVC)
Transducer requirements:
Excitation voltage: Transducer output:
Input specifications:
Range: Offset:
Output specifications:
Frequency response: Zero balance range: Zero balance accuracy: Zero balance drift: Accuracy:
Alarms:
5.0 Vdc ±0.1% 5 µV/V/mmHg
–25 mmHg to 300 mmHg ±150 mmHg
dc to 50 Hz ±150 mmHg ±1 mmHg ±1 mmHg over 24 hours ±2% or ±1 mmHg, whichever is greater (exclusive of transducer) User-selectable upper and lower limits for systolic, diastolic, and mean pressures
2-10 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications

Noninvasive Blood Pressure (NBP)

Measurement technique: Oscillometric Displayed parameters: Systolic, diastolic, and mean pressures, pulse rate, time
of last measurement
Measurement modes: Manual, auto, and stat in adult and OR modes; manual
and auto in neonatal mode
NBP pressure range:
Systolic pressure range
Adult: Pediatric: Neonatal:
Diastolic pressure range
Adult: Pediatric: Neonatal:
Mean pressure range
Adult: Pediatric: Neonatal:
30 to 275 mmHg 30 to 235 mmHg 30 to 135 mmHg
10 to 220 mmHg 10 to 220 mmHg 10 to 110 mmHg
20 to 260 mmHg 20 to 260 mmHg 20 to 125 mmHg
Cuff pressure range:
Adult: Pediatric: Neonatal:
Pressure accuracy: Static: Clinical:
Heart rate detection: 30 to 200 beats per minute Total cycle time: 20 to 40 seconds typical (dependent on heart rate and
Automatic cycle times: 0 to 8 hours Auto zero: Zero pressure reference prior to each cuff inflation Tubing length:
Adult: Neonatal:
Automatic cuff deflation: Cycle time exceeding 3 minutes (90 seconds neonatal),
Cuff sizes:
Disposable: Reusable:
0 to 275 mmHg 0 to 235 mmHg 0 to 135 mmHg
±2% or ±3 mmHg, whichever is greater ±5 mmHg average error 8 mmHg standard deviation
motion artifact)
12 feet 8 feet
power off, or cuff pressure exceeds 294 mmHg (±6 mmHg) adult, 147 mmHg (±3 mmHg) neonatal
Large adult, adult, small adult, pediatric, small pediatric, and infant Thigh, large adult, adult, child, and infant
Alarms: User-selectable upper and lower limits for systolic,
diastolic, and mean pressures
Revision A Dash 3000/4000 Patient Monitor 2-11
2000966-035

Pulse Oximetry (SPO2)

EQUIPMENT OVERVIEW: Technical Specifications
Parameters monitored: Arterial oxygen saturation (SpO2) and peripheral pulse
rate (PPR) SpO2 range: 50 - 100% PPR range: 30 - 300 beats per minute

Cardiac Output (CO)

Accuracy:
SpO2: PPR:
Alarms: User-selectable upper and lower limits for SpO2 and
Availability: Included in 7020 and 7025 software packages. Not
Input specifications:
Probe type: Catheter manufacturers: Catheter sizes:
Abbott catheter sizes: Arrow catheter sizes:
Baxter catheter sizes: Ohmeda catheter sizes: Other catheter sizes:
Injectate volume:
Actual accuracy depends on probe. Please reference
manufacturer’s specifications.
± 2% (70 - 100% SpO2) ±1 standard deviation
± 3% (50 - 69% SpO2) ±1 standard deviation
± 3 beats per minute
PPR
available in 7015 software package.
In-line or bath probe
Abbott, Arrow, Baxter, Ohmeda, or other
5.5F (75 cm), 7F (85 cm), 7.5F (110 cm), and
8F (110 cm)
5, 6, 7, or 7.5F
5, 6, 7, 7.5 or 8F
5, 7, or 7.5F
Cardiac coefficient entered manually
3, 5, or 10 cc Output specifications:
Parameters displayed: Range:
Cardiac output: Blood temperature: Injectate temperature:
Accuracy:
Cardiac output: Blood temperature: Injectate temperature:
Frequency response:
2-12 Dash 3000/4000 Patient Monitor Revision A
2000966-035
Cardiac output, blood temperature, injectate
temperature, trial number
0.2 - 15 liters per minute
30 - 42°C
0 - 30°C
±5% (liters of blood/min)
±0.2°C
±0.3°C
dc to 15 Hz ±2 Hz
EQUIPMENT OVERVIEW: Technical Specifications

Respiration

Temperature (TEMP)

Measurement technique: Impedance variation detection Range:
Respiration rate: Base impedance: Detection sensitivity:
Accuracy:
Respiration rate ±1 BrPM
Waveform display bandwidth: 0.1 to 1.8 Hz (–3 dB) Alarms: User-selectable upper and lower respiration rate limits,
Number of channels: 2 Input specifications:
Probe type: Temperature range:
Resolution:
0 - 200 breaths per minute
100 - 1000
0.4 to 10
and user-selectable apnea limit
YSI Series 400 or 700 thermistor (determined by input
cable)
0°C to 45°C (32°F to 113°F)
±0.1°C
at 52.6 kHz excitation frequency
variation

Carbon Dioxide (CO2)

Output specifications:
Parameters displayed: Accuracy:
Alarms:
Information displayed: Inspired and expi red car bon diox ide co ncen tra tion s in %,
Measurement technique: Non-dispersive infrared absorption, dual wavelength
Sensor type: Novametrix Medical Systems’ Capnostat III Patient interface: Compatible with Novametrix Medical Systems’
Airway adaptors
Types: Dead space/chamber volumes:
Adult reusable: Adult disposable: Neonatal:
T1, T2
(independent of source)
±0.1°C for YSI series 400 probes;
±0.3°C for YSI series 700 probes
User-selectable upper and lower limits for T1, T2
mmHg or kPa, respiration rate, continuous CO2 wavef orm
ratiometric
Capnogard monitoring product
Adult reusable (standard), adult disposable, neonatal
<5 cc
<5 cc
<0.5 cc
Revision A Dash 3000/4000 Patient Monitor 2-13
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications
CO2 measurement specifications:
Measurement range:
Pi CO2/Fi CO2: Pe CO2/Fe CO2:
RR: Accuracy: Display update interval: CO2 waveform sweep speed: CO2 averaging: CO2 measurement stability: Resolution: Noise:
0 to 100 mmHg/0 to 13% 0 to 100 mmHg/0 to 13% 0 to 120 breaths/min
±5% of reading or ±2 mmHg, whichever is greater 2 sec Selectable 6.25, 12.5, or 25 mm/sec Selectable from single breath, 10 sec, or 20 sec Accuracy maintained over 8 hours 1 mmHg 2% of reading or 0.5 mmHg (maximum), whichever is greater
60 Hz interference:
<0.5 mmHg at 38 mmHg
Step response time:
Adult:
Neonatal
<60 ms (10-90%) <50 ms (10-90%)
Interference:
N2O gas:
±5% of reading or ±2 mmHg (maximum), whichever is greater, with N2O compensation enabled
O2 gas:
±5% of reading or ±2 mmHg (maximum), whichever is greater, with O2 compensation enabled
Barometric pressure:
Water vapor:
±2 mmHg (maximum) from 500 to 800 mmHg ±1.5% of reading or ±0.5 mmHg (maximum), whichever is greater
Anesthetic agent:
±0.5 mmHg (maximum) for concentration of no more than 5% of halogenated agents
Airway adapter variability:
±3% of reading or ±1.5 mmHg (maximum), whichever is greater, with same or different adapter; not applicable after adapter zero
Warm-up time:
Less than 15 seconds to initial CO2 indication, full specification within 120 seconds; waveform immediate upon power up
Calibration:
Factory settings:
Factory calibration settings stored in nonvolatile memory within the sensor; 15 second adaptor calibration when switching airway types
Verification:
Zero and span performance check with on-cable verifier
Respiration rate specifications:
Range (for 5% step size):
Accuracy:
Resolution:
0-120 breaths per minute ±1 breath per minute ±1 breath per minute
Barometric pressure sensor specifications:
Range: Accuracy:
425 to 817 mmHg (56 to 109 kPa) ±25 mmHg
Alarms: User-selectable upper and lower limits for CO2 and RR.
2-14 Dash 3000/4000 Patient Monitor Revision A
2000966-035
Loading...