Marquette Dash 3000, Dash 4000 Service Manual

Dash3000/4000
Patient Monitor
Service Manual
2000966-035 Revision A
Listed below are GE Medical Systems Information Technologies’ trademarks. All other trademarks contained
herein are the property of their respective owners.
900 SC, ACCUSKETCH, AccuVision, APEX, AQUA-KNOT, ARCHIVIST, Autoseq, BABY MAC, C Qwik Connect, CardioServ, CardioSmart, CardioSys, CardioWindow, CASE, CD TELEMETRY, CENTRA, CHART GUARD, CINE 35, CORO, COROLAN, COROMETRICS, Corometrics Sensor Tip, CRG PLUS, DASH, Digistore, Digital DATAQ, E for M, EAGLE, Event-Link, FMS 101B, FMS 111, HELLIGE, IMAGE STORE, INTELLIMOTION, IQA, LASER SXP, MAC, MAC-LAB, MACTRODE, MANAGED USE, MARQUETTE, MARQUETTE MAC, MARQUETTE MEDICAL SYSTEMS, MARQUETTE UNITY NETWORK, MARS, MAX, MEDITEL, MEI, MEI in the circle logo, MEMOPORT, MEMOPORT C, MINISTORE, MINNOWS, Monarch 8000, MULTI-LINK, MULTISCRIPTOR, MUSE, MUSE CV, Neo-Trak, NEUROSCRIPT, OnlineABG, OXYMONITOR, Pres-R-Cuff, PRESSURE-SCRIBE, QMI, QS, Quantitative Medicine, Quantitative Sentinel, RAC RAMS, RSVP, SAM, SEER, SILVERTRACE, SOLAR, SOLARVIEW, Spectra 400, Spectra-Overview, Spectra-Tel, ST GUARD, TRAM, TRAM-NET, TRAM-RAC, TRAMSCOPE, TRIM KNOB, Trimline, UNION STATION, UNITY logo, UNITY NETWORK, Vari-X, Vari-X Cardiomatic, VariCath, VARIDEX, VAS, and Vision Care Filter are trademarks of GE Medical Systems Information Technologies registered in the United States Patent and Trademark Office.
12SL, 15SL, Access, AccuSpeak, ADVANTAGE, BAM, BODYTRODE, Cardiomatic, CardioSpeak, CD TELEMETRY
Cumulus, Event-Link Nimbus, HI-RES, ICMMS, IMAGE VAULT, IMPACT.wf, INTER-LEAD, IQA, LIFEWATCH, Managed Use, MARQUETTE PRISM, MARQUETTE MMS, MRT, MUSE CardioWindow, NST PRO, NAUTILUS, O
®
-LAN, CENTRALSCOPE, Corolation, EDIC, EK-Pro, Event-Link Cirrus, Event-Link
®
RESPONDER, MENTOR, MicroSmart,
SENSOR, Octanet, OMRS, PHi-Res,
2
Premium, Prism, QUIK CONNECT V, QUICK CONNECT, QT Guard, SMART-PAC, SMARTLOOK, Spiral Lok, Sweetheart, UNITY, Universal, Waterfall, and Walkmom are trademarks of GE Medical Systems Information Technologies.
© GE Medical Systems Information Technologies, 2000. All rights reserved.
T-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035 27 November 2000

CONTENTS

1 INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Manual Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Revision History . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Manual Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Intended Audience . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
Responsibility of the Manufacturer . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4
Warnings, Cautions, and Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5
Equipment Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-6
Service Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Service Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-8
Equipment Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-8
2 EQUIPMENT OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-1
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
The Monitoring System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
The Patient Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
Optional RAC 2A Module Housing . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
Optional Wireless LAN System . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
Technical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
Performance Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
ECG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
Invasive Blood Pressure (BP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Noninvasive Blood Pressure (NBP) . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Pulse Oximetry (SPO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-12
Cardiac Output (CO) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Respiration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Temperature (TEMP) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Carbon Dioxide (CO2) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-13
Analog Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Defibrillator Synchronization Pulse . . . . . . . . . . . . . . . . . . . . . . . . . . 2-15
Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Paper Recorder . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
RF Wireless LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
Environmental Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
Physical Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-17
Certification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18
Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-19
Revision A Dash 3000/4000 Patient Monitor i
2000966-035
CONTENTS:
3 INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Back Panel Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Front Panel Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Power Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Ethernet Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-6
Twisted Pair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Concentrator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
Node . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Segment and Branch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Repeater . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Bridge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Twisted Pair Cabling (10BaseT) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Symbol PC Card (Wireless LAN) . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
4 MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Maintenance Schedule . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Manufacturer Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Manufacturer Responsibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Visual Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-4
Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Cleaning Precautions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Cleaning the Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-5
Exterior Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Cleaning the Print Head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-6
Battery Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
Conditioning the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-7
Replacing the Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-8
Electrical Safety Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Recommendations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-9
Wall Receptacle Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Integrity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-11
Ground (Earth) Wire Leakage Current Tests . . . . . . . . . . . . . . . . . . 4-13
Enclosure Leakage Current Test . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-15
Patient (Source) Leakage Current Test . . . . . . . . . . . . . . . . . . . . . .4-17
Patient (Sink) Leakage Current Test
(Mains Voltage on the Applied Part) . . . . . . . . . . . . . . . . . . . . . . .4-19
Test Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-20
Hi-Pot (Dielectric Withstand) Test . . . . . . . . . . . . . . . . . . . . . . . . . . 4-21
DAS Assembly AC Hi-Pot Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-22
Processor/Power Management PCB Hi-Pot Test . . . . . . . . . . . . . .4-23
AC Mains Hi-Pot Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-24
Checkout Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-25
Manufacturer Recommended Test Equipment . . . . . . . . . . . . . . . . . 4-25
Monitor Power-up Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-26
ECG Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-27
ii Dash 3000/4000 Patient Monitor Revision A
2000966-035
CONTENTS:
Respiration Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-30
Temperature Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-31
Cardiac Output Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-32
Invasive Blood Pressure Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-34
Pulse Oximetry Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-36
Noninvasive Blood Pressure Tests . . . . . . . . . . . . . . . . . . . . . . . . . 4-38
Analog Output and Defibrillator Synchronization Tests . . . . . . . . . . 4-40
Battery Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
Graph Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-43
Graph Speed Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-43
Display Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
Speaker Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
Network Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-44
RF LAN Test (option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-44
RAC 2A Module Housing Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-45
PM Form . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-45
Repair Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-46
5 TROUBLESHOOTING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Service Menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Boot Loader Service Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Main Menu Service Mode Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Review Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-8
Error Log Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Battery Alarms and Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-10
Alarm Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-10
Error Message . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-11
Power Source Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Wall Receptacle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-12
Power Cord and Plug . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-13
Data Acquisition Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
ECG Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-14
Lead Fail Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-15
Pace Detect Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-16
Invasive Blood Pressure Functions . . . . . . . . . . . . . . . . . . . . . . . . .5-17
Respiration Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-19
Noninvasive Blood Pressure Functions . . . . . . . . . . . . . . . . . . . . . . 5-21
Wireless LAN Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-22
Service Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Fault/Symptom Analysis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
Acquisition PCB Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-25
Processor PCB Symptoms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-25
Troubleshooting Software Updates - Problems and Solutions . . . . . . . .5-26
Error Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-29
Revision A Dash 3000/4000 Patient Monitor iii
2000966-035
CONTENTS:
6 CONFIGURATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-1
Loading Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Intended Use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Software Loading/Updating Methods . . . . . . . . . . . . . . . . . . . . . . . . .6-3
Software Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Monitor Software Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Maintain Patient Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-5
Problems While Loading Software . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
Record Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-6
Load Software From Diskette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-7
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Connect the PC to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-7
Software Diskettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-8
Update Program Start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-9
Setup Monitor To Accept Download Files . . . . . . . . . . . . . . . . . . . .6-11
Download Files to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-12
Load Software Over The Network . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
About the Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-13
Network Update Diskettes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Copy Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Download Files to the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-15
Complete the Software Download . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Activate Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-17
Configuring a Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-19
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-19
Main Menu Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Set Unit Name . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Set Bed Number . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-21
Patient-Monitor Type . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-22
Set Graph Locations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-23
Admit Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-24
Boot Code Selections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Set Defib Sync Voltage and Pulse Width . . . . . . . . . . . . . . . . . . . . .6-25
Set Line Frequency . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-25
Set CIC and QS Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-25
Set MUSE Protocol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Transcutaneous Pace Blank Length . . . . . . . . . . . . . . . . . . . . . . . . 6-26
Set Country Selection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Wireless LAN . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-27
Completion . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-27
Advanced User Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-28
Set Time and Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-28
Change Software Level . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-29
Change Ethernet Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-30
Review Errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-30
Transferring Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-34
Access the COPY LOGS Menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
Select the Care Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-35
Select the Monitoring Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-35
iv Dash 3000/4000 Patient Monitor Revision A
2000966-035
CONTENTS:
Select the Error Log Date . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
Copy Error Logs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-36
Eject Floppy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-36
7 CALIBRATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-1
Hardware Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
NBP, ECG, BP, and End-tidal CO2 Software Calibration . . . . . . . . . . . . . 7-4
NBP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
ECG or BP Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-7
End-tidal CO2 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-8
8 FIELD REPLACEABLE UNITS AND UPGRADES . . . . . . . . . . . . . . . . . . 8-1
Disassembly Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
Tools Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Before Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
Hardware Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-3
PCB Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-3
After Reassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-5
Handle Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Removing the Handle Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-7
Replacing or Upgrading the Dash 3000 Alarm Light Option . . . . . . .8-8
Display Assembly Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Removing the Display Assembly from the Main Unit . . . . . . . . . . . . . 8-9
Replacing the Backlight Inverter PCB . . . . . . . . . . . . . . . . . . . . . . .8-13
Replacing the Key Pad Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 8-14
Replacing the LCD Color Display . . . . . . . . . . . . . . . . . . . . . . . . . . .8-15
Replacing or Upgrading the Dash 4000 Alarm Light Option . . . . . .8-16
Replacing the Dash 4000 Front Panel PCB . . . . . . . . . . . . . . . . . . .8-17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-17
Main Unit Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-18
DAS and NBP Assemblies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-18
Main and/or Power Supply Assemblies, Speaker or
RF LAN Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-21
Processor/Power Management PCB and Battery Assembly . . . . . .8-22
Power Supply Assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-24
Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-25
RF LAN Upgrade Instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-26
Optional DDW Writer Replacement/Upgrade . . . . . . . . . . . . . . . . . . . . .8-36
Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8-36
Upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-36
Revision A Dash 3000/4000 Patient Monitor v
2000966-035
CONTENTS:
9 ASSEMBLY DRAWINGS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-1
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Theory Of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-4
General Monitor Block Theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
Overall Monitor Block Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-5
User Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
Power Supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Data Acquisition System (DAS) . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-8
Processor/Power Management Subsystem . . . . . . . . . . . . . . . . . . . 9-13
Optional Thermal Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-20
Handle Subassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
Interfaces . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-20
Setup and Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-21
Electrical Diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-22
Dash 3000 Exploded Views, PN 420000-xxx . . . . . . . . . . . . . . . . . . . . . 9-23
Dash 3000 Display Bezel Assembly, PN 419031-001 . . . . . . . . . . . . . .9-28
Dash 3000 Display Bezel Parts List, PN 419031-001 . . . . . . . . . . . . . . 9-29
Dash 3000 Display Bezel Part List , PN 419031-003 . . . . . . . . . . . . . . . 9-31
Dash 4000 Parts List PN 2004323-xxx . . . . . . . . . . . . . . . . . . . . . . . . . . 9-34
Dash 4000 Display Assembly, PN 2004272-001 . . . . . . . . . . . . . . . . . . 9-36
Dash 4000 Display Bezel Parts List, PN 2004272-001 . . . . . . . . . . . . . 9-37
Field Replaceable Units (FRU’s) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Dash 3000 FRUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Dash 4000 FRUs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-38
Port Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-40
Invasive Blood Pressure Cable Connector . . . . . . . . . . . . . . . . . . . . 9-40
Pulse Oximetry (SpO2) Cable Connector . . . . . . . . . . . . . . . . . . . . 9-41
Temperature/CO Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . 9-41
Capnostat III (CO2) Cable Connector . . . . . . . . . . . . . . . . . . . . . . . 9-42
NBP Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
ECG Cable Connector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-43
Input Power Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-44
Network Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-44
Auxiliary Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-45
Defib Sync . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-45
Peripheral Expansion Interface . . . . . . . . . . . . . . . . . . . . . . . . . . . .9-46
vi Dash 3000/4000 Patient Monitor Revision A
2000966-035

1 INTRODUCTION

Revision A Dash 3000/4000 Patient Monitor 1-1
2000966-035
For your notes
INTRODUCTION:
1-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Manual Information

Manual Information

Revision History

Revision Date Comment
A 27 November 2000 Initial release of this manual.

Manual Purpose

Intended Audience

Each page of this manual has the document part number and revision letter at the bottom of the page. The revision letter identifies the
document’s update level. The revision history of this document is summarized below.
Revision History
This manual supplies technical information for service representatives and technical personnel so they can maintain the equipment to the assembly level. Use it as a guide for maintenance and electrical repairs considered field repairable. Where necessary the manual identifies additional sources of relevant information and technical assistance.
See the operator’s manual for the instructions necessary to operate the equipment safely in accordance with its function and intended use.
This manual is intended for service representatives and technical personnel who maintain, troubleshoot, or repair this equipment.
Revision A Dash 3000/4000 Patient Monitor 1-3
2000966-035

Safety Information

INTRODUCTION: Safety Information

Responsibility of the Manufacturer

General

GE Medical Systems Information Technologies is responsible for the effects of safety, reliability, and performance only if:
Assembly operations, extensions, readjustments, modifications, or repairs are carried out by persons authorized by GE Medical Systems Information Technologies.
The electrical installation of the relevant room complies with the requirements of the appropriate regulations.
The equipment is used in accordance with the instructions for use.
This device is intended for use under the direct supervision of a licensed health care practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a physician.
Contact GE Medical Systems Information Technologies for information before connecting any devices to the equipment that are not recommended in this manual.
Parts and accessories used must meet the requirements of the applicable IEC 601 series safety standards, and/or the system configuration must meet the requirements of the IEC 60601-1-1 medical electrical systems standard.
Periodically, and whenever the integrity of the device is in doubt, test all functions.
The use of ACCESSORY equipment not complying with the equivalent safety requirements of this equipment may lead to a reduced level of safety of the resulting system. Consideration relating to the choice shall include:
use of the accessory in the PATIENT VICINITY; and
evidence that the safety certification of the ACCESSORY has been performed in accordance to the appropriate IEC 60601-1 and/or IEC 60601-1-1 harmonized national standard.
If the installation of the equipment, in the USA, will use 240V rather than 120V, the source must be a center-tapped, 240V, single-phase circuit.
1-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Safety Information

Warnings, Cautions, and Notes

The terms danger, warning, and caution are used throughout this manual to point out hazards and to designate a degree or level or seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person.
DANGER indicates an imminent hazard which, if not avoided, will result in death or serious injury.
WARNING indicates a potential hazard or unsafe practice which, if not avoided, could result in death or serious injury.
CAUTION indicates a potential hazard or unsafe practice which, if not avoided, could result in minor personal injury or product/property damage.
NOTE provides application tips or other useful information to assure that you get the most from your equipment.
Revision A Dash 3000/4000 Patient Monitor 1-5
2000966-035
INTRODUCTION: Safety Information

Equipment Symbols

Some of the following symbols appear on the equipment.
ATTENTION: Consult accompanying documents before using the equipment.
In Europe, this symbol means dangerous or high voltage. In the United States, this symbol represents the caution notice below:
To reduce the risk of electric shock, do NOT remove cover (or back). Refer servicing to qualified personnel.
Defibrillator-proof type CF equipment; type CF equipment is specifically designed for applications where a conductive connection directly to the heart is established. The paddles indicate the equipment is defibrillator proof.
Defibrillator-proof type BF equipment; type BF equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application. Type BF equipment is type B equipment with an F-type isolated (floating) part. The paddles indicate the equipment is defibrillator proof.
Type B equipment; type B equipment is suitable for intentional external and internal application to the patient, excluding direct cardiac application.
Equipotentiality
Alternating current (AC)
Power;
Fuse
Battery
Indicates the Ethernet connection for the monitor.
I = ON; O= OFF
1-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INTRODUCTION: Safety Information
PRESS
814A
815A
816A
817A
818A
Press to open.
Power
Graph Go/Stop
NBP Go/Stop
Function
Silence Alarm
Classified by Underwriters Laboratories Inc. with respect to electric shock, fire, mechanical and other specified hazards, only in accordance with UL 2601-1, CAN/CSA C22.2 No. 601.1, IEC 60601-1, and, if required, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 60601­1-1.
Revision A Dash 3000/4000 Patient Monitor 1-7
2000966-035
INTRODUCTION: Service Information

Service Information

Service Requirements

Equipment Identification

Follow the service requirements listed below.
Refer equipment servicing to GE Medical Systems Information Technologies’ authorized service personnel only.
Any unauthorized attempt to repair equipment under warranty voids that warranty.
It is the user’s responsibility to report the need for service to GE Medical Systems Information Technologies or to one of their authorized agents.
Failure on the part of the responsible individual, hospital, or institution using this equipment to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards.
Regular maintenance, irrespective of usage, is essential to ensure that the equipment will always be functional when required.
Every GE Medical Systems Information Technologies device has a unique serial number for identification. A sample of the information found on a serial number label is shown below.
D 0 XX 0005 G XX
Month Manufactured
A = January B = February C = March D = April E = May F = June G = July H = August J = September K = October L = November M = December
Year Manufactured
0 = 2000 1 = 2001 2 = 2002
(and so on)
Product Code
Two-character product descriptor
Product Sequence Number
Manufacturing number (of total units manufactured)
Division
F = Cardiology G = Monitoring
Device Characteristics
One or two letters that further describe the unit, for example: P = prototype not conforming to marketing specification R = refurbished equipment S = special product documented under Specials part numbers U = upgraded unit
1-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035

2 EQUIPMENT OVERVIEW

Revision A Dash 3000/4000 Patient Monitor 2-1
2000966-035
For your notes
EQUIPMENT OVERVI EW:
2-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Components

EQUIPMENT OVERVIEW: Components

The Monitoring System

The Patient Monitor

The Dash 3000/4000 patient monitor can function by itself with a built-in writer, or it can be cabled in with the Unity Network via Ethernet. Optional components are, if using Wireless LAN or cabled to Ethernet, a Centralscope central station and the Clinical Information Center.
This device is designed to monitor a fixed set of parameters including ECG, noninvasive blood pressure, impedance respiration, SpO2, and temperature. Invasive pressure and EtCO2 are optional features. Additional specialized features include cardiac output, cardiac calculations, pulmonary calculations, dose calculations, PA wedge (PA wedge is only available with the invasive pressure option), and SAM module interface.
001A
Dash 3000 Monitor
Revision A Dash 3000/4000 Patient Monitor 2-3
2000966-035
Dash 4000 Monitor
051A
EQUIPMENT OVERVIEW: Components

Right Side View

Left Side View

All of the patient cable connectors are located on the right side of the monitor. A Trim Knob control provides single control operation of virtually all monitor functions.
Patient Cable Connectors
002A
On the left of the monitor, you can find the built-in writer and the battery compartment.
Optional Built-in Writer— The built-in, 4 channel writer is located in the center of the left side of the monitor .
Battery Compartment— The battery packs are located in this compartment.
003A
2-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Components

Back Vie w

Network Connector—A cable can be connected to this port for monitors used in patient monitoring network configurations.
Equipotential Terminal—For
measurements in or near the heart we recommend connecting the monitor to the potential equalization system. U se the green and yellow potential equalization cable and connect it to this pin.
AC Power Connector
On the back of the monitor you will find all connectors for equipment and network.
Line V oltage Selector—This s elector is factory set to match the line voltage and frequency rating for your country.
Audible Alarm Enunciator—The internal speaker provides sound for audible alarms. For better sound quality do not block speaker.
Aux Port—Used for RAC 2A and software updates.
Defib Sync Connector—Provides ECG analog output signals to
004A
Peripheral Expansion Port
user-supplied equipment. A 5­volt, 2-millisecond artificial pacer spike is added to the analog output when PACE is on and detection occurs.

Optional Alarm Indicator

An optional alarm indicator can be built into the handle of the Dash 3000 monitor or into the display bezel of the Dash 4000 monitor. When activated, the LED indicator flashes red for CRISIS and WARNING patient status alarms and yellow for all other alarms.
Alarm Indicator
Dash 3000 Monitor
Revision A Dash 3000/4000 Patient Monitor 2-5
2000966-035
Dash 4000 Monitor
052A
EQUIPMENT OVERVIEW: Components

Optional RAC 2A Module Housing

The RAC 2A module housing currently supports the SAM module.
006A
An integral power supply is used to run the RAC 2A and support the needed voltages.
2-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Components

Optional Wireless LAN System

Access Points

The flexibility of the GE Medical Systems Information Technologies Unity Network is increased by using the Wireless LAN system. The Wireless LAN system allows the user to roam from one access point to another, maintaining a strong, seamless connection to the Unity network.
The monitor, with its optional built-in Wireless LAN, functionally performs the same as a monitor connected directly to the Unity network. It can be viewed at the central station and by other GE Medical Systems
Information Technologies’ monitors on the network (i.e. Dash 3000/4000, Eagle 4000, and Solar patient monitors). Monitors with Wireless LAN sends and receives patient data via the access points of the Unity network.
NOTE: Wireless patient monitors that are moved from room to room
must have the monitor type configured as Rover or Rover/Combo monitoring.
To integrate the wireless network with the wired network, one or more access points are necessary. An access point connects the wireless monitor to the wired network infrastructure within the building, and acts as a bridge between the wired and wireless networks. The areas covered by each access point overlap to insure continuous coverage.
NOTE: The monitor will only work with a Symbol Access Point. The
monitor will not communicate directly with a Wireless LAN device from Aironet.
050A
Revision A Dash 3000/4000 Patient Monitor 2-7
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications

Technical Specifications

Due to continual product innovation, specifications are subject to change without notice. The following specifications are accurate as of the date of this publication, and pertain to the Dash 3000/4000 Patient Monitor.

Performance Specifications

Display

Size:
Type:
Color: Active-Matrix Liquid Crystal Display (LCD)
Resolution: 640 by 480 pixels
Dash 3000: Dash 4000:
8.4-inch diagonal
10.4-inch diagonal

Controls

Alarms

Number of traces: 6 (maximum) Number of seconds/trace:
Dash 3000: Dash 4000:
Sweep speed:
All waveforms 6.25, 12.5 or 25 mm/sec (with erase bar)
Waveform display options: Individual 6 waveforms, individual 3 waveforms, full,
Information window: Displays non-real-time information without obstructing
Display organization: Prioritized by parameter
Standard: Trim Knob control plus 5 hard keys: Power, NBP Go/
Categories: Patient Status and System Status
4.9 at 25 mm/sec
5.9 at 25 mm/sec
and full grid modes
the display of real-time information
Stop, Function, Silence Alarm, and Graph Go/Stop
Priorities: 4 levels — Crisis, Warning, Advisory, and Message
Notification: Audible and visual Setting: Default and individual Silencing: 1 minute, current alarm only
2-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035
ECG
EQUIPMENT OVERVIEW: Technical Specifications
Pause: 5 minutes (adult); 3 minutes (neonatal); 5, 15 minutes,
permanent (OR mode)
Volume: Default 70%, 70 dB measured at 1 meter
5 Leadwire cable: I, II, III, V, aVR, aVL, and aVF 10 Leadwire cable (12SL option): V2, V3, V4, V5 and V6 Leads analyzed simultaneously: I, II, III, and V (multi-lead mode) Lead fail: Identifies failed lead Alarms: User-selectable upper and lower heart rate limits Input specifications:
Voltage range: Signal width: Heart rate range: Accuracy: Input impedance:
Common mode: Differential:
Common mode rejection:
±0.5 mV to ±5 mV 40 ms to 120 ms (Q to S) 30 to 300 BPM ±1% or ±1 BPM, whichever is greater
>10 M
at 50/60 Hz
>2.5 M
from dc to 60 Hz
90dB minimum at 50 Hz or 60 Hz
Output specifications:
Frequency response: Display:
Diagnostic: Monitoring: Moderate: Maximum:
Paper Recorder:
Diagnostic: Monitoring: Moderate: Maximum:
Linearity deviation: Noise:
ST segment measurement:
Measurement point: Measurement range:
Measurement accuracy:
Pacemaker detection/rejection:
Input voltage range: Input pulse width: Rise time: Over/under shoot: Baseline drift:
0.05 to 40 Hz
0.05 to 40 Hz
0.05 to 25 Hz 5 to 25 Hz
0.05 to 100 Hz
0.05 to 40 Hz
0.05 to 25 Hz 5 to 25 Hz ±3% (maximum) <30 µV (referred to input)
Adjustable from 0 to 120 ms past the J-point (default: 60 ms adult, 30 ms neonatal) –12.0 to +12.0 mm ±10% or 0.5 mm, whichever is greater
±2 mV to ±700 mV
0.1 ms to 2 ms 10 µs to 100 µs 2 mV (max) <0.5 mV/hour with a ±700-mV, 2-ms pacemaker pulse applied
Revision A Dash 3000/4000 Patient Monitor 2-9
2000966-035

Invasive Blood Pressure (BP)

EQUIPMENT OVERVIEW: Technical Specifications
Number of channels: 2 Transducer sites: Arterial (ART), femoral artery (FEM), pulmonary artery
(PA), central venous (CVP), right atrial (RA), left atrial (LA), intracranial (ICP), and special (SP)
In neonatal mode: umbilical artery catheter (UAC) and umbilical venous catheter (UVC)
Transducer requirements:
Excitation voltage: Transducer output:
Input specifications:
Range: Offset:
Output specifications:
Frequency response: Zero balance range: Zero balance accuracy: Zero balance drift: Accuracy:
Alarms:
5.0 Vdc ±0.1% 5 µV/V/mmHg
–25 mmHg to 300 mmHg ±150 mmHg
dc to 50 Hz ±150 mmHg ±1 mmHg ±1 mmHg over 24 hours ±2% or ±1 mmHg, whichever is greater (exclusive of transducer) User-selectable upper and lower limits for systolic, diastolic, and mean pressures
2-10 Dash 3000/4000 Patient Monitor Revision A
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications

Noninvasive Blood Pressure (NBP)

Measurement technique: Oscillometric Displayed parameters: Systolic, diastolic, and mean pressures, pulse rate, time
of last measurement
Measurement modes: Manual, auto, and stat in adult and OR modes; manual
and auto in neonatal mode
NBP pressure range:
Systolic pressure range
Adult: Pediatric: Neonatal:
Diastolic pressure range
Adult: Pediatric: Neonatal:
Mean pressure range
Adult: Pediatric: Neonatal:
30 to 275 mmHg 30 to 235 mmHg 30 to 135 mmHg
10 to 220 mmHg 10 to 220 mmHg 10 to 110 mmHg
20 to 260 mmHg 20 to 260 mmHg 20 to 125 mmHg
Cuff pressure range:
Adult: Pediatric: Neonatal:
Pressure accuracy: Static: Clinical:
Heart rate detection: 30 to 200 beats per minute Total cycle time: 20 to 40 seconds typical (dependent on heart rate and
Automatic cycle times: 0 to 8 hours Auto zero: Zero pressure reference prior to each cuff inflation Tubing length:
Adult: Neonatal:
Automatic cuff deflation: Cycle time exceeding 3 minutes (90 seconds neonatal),
Cuff sizes:
Disposable: Reusable:
0 to 275 mmHg 0 to 235 mmHg 0 to 135 mmHg
±2% or ±3 mmHg, whichever is greater ±5 mmHg average error 8 mmHg standard deviation
motion artifact)
12 feet 8 feet
power off, or cuff pressure exceeds 294 mmHg (±6 mmHg) adult, 147 mmHg (±3 mmHg) neonatal
Large adult, adult, small adult, pediatric, small pediatric, and infant Thigh, large adult, adult, child, and infant
Alarms: User-selectable upper and lower limits for systolic,
diastolic, and mean pressures
Revision A Dash 3000/4000 Patient Monitor 2-11
2000966-035

Pulse Oximetry (SPO2)

EQUIPMENT OVERVIEW: Technical Specifications
Parameters monitored: Arterial oxygen saturation (SpO2) and peripheral pulse
rate (PPR) SpO2 range: 50 - 100% PPR range: 30 - 300 beats per minute

Cardiac Output (CO)

Accuracy:
SpO2: PPR:
Alarms: User-selectable upper and lower limits for SpO2 and
Availability: Included in 7020 and 7025 software packages. Not
Input specifications:
Probe type: Catheter manufacturers: Catheter sizes:
Abbott catheter sizes: Arrow catheter sizes:
Baxter catheter sizes: Ohmeda catheter sizes: Other catheter sizes:
Injectate volume:
Actual accuracy depends on probe. Please reference
manufacturer’s specifications.
± 2% (70 - 100% SpO2) ±1 standard deviation
± 3% (50 - 69% SpO2) ±1 standard deviation
± 3 beats per minute
PPR
available in 7015 software package.
In-line or bath probe
Abbott, Arrow, Baxter, Ohmeda, or other
5.5F (75 cm), 7F (85 cm), 7.5F (110 cm), and
8F (110 cm)
5, 6, 7, or 7.5F
5, 6, 7, 7.5 or 8F
5, 7, or 7.5F
Cardiac coefficient entered manually
3, 5, or 10 cc Output specifications:
Parameters displayed: Range:
Cardiac output: Blood temperature: Injectate temperature:
Accuracy:
Cardiac output: Blood temperature: Injectate temperature:
Frequency response:
2-12 Dash 3000/4000 Patient Monitor Revision A
2000966-035
Cardiac output, blood temperature, injectate
temperature, trial number
0.2 - 15 liters per minute
30 - 42°C
0 - 30°C
±5% (liters of blood/min)
±0.2°C
±0.3°C
dc to 15 Hz ±2 Hz
EQUIPMENT OVERVIEW: Technical Specifications

Respiration

Temperature (TEMP)

Measurement technique: Impedance variation detection Range:
Respiration rate: Base impedance: Detection sensitivity:
Accuracy:
Respiration rate ±1 BrPM
Waveform display bandwidth: 0.1 to 1.8 Hz (–3 dB) Alarms: User-selectable upper and lower respiration rate limits,
Number of channels: 2 Input specifications:
Probe type: Temperature range:
Resolution:
0 - 200 breaths per minute
100 - 1000
0.4 to 10
and user-selectable apnea limit
YSI Series 400 or 700 thermistor (determined by input
cable)
0°C to 45°C (32°F to 113°F)
±0.1°C
at 52.6 kHz excitation frequency
variation

Carbon Dioxide (CO2)

Output specifications:
Parameters displayed: Accuracy:
Alarms:
Information displayed: Inspired and expi red car bon diox ide co ncen tra tion s in %,
Measurement technique: Non-dispersive infrared absorption, dual wavelength
Sensor type: Novametrix Medical Systems’ Capnostat III Patient interface: Compatible with Novametrix Medical Systems’
Airway adaptors
Types: Dead space/chamber volumes:
Adult reusable: Adult disposable: Neonatal:
T1, T2
(independent of source)
±0.1°C for YSI series 400 probes;
±0.3°C for YSI series 700 probes
User-selectable upper and lower limits for T1, T2
mmHg or kPa, respiration rate, continuous CO2 wavef orm
ratiometric
Capnogard monitoring product
Adult reusable (standard), adult disposable, neonatal
<5 cc
<5 cc
<0.5 cc
Revision A Dash 3000/4000 Patient Monitor 2-13
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications
CO2 measurement specifications:
Measurement range:
Pi CO2/Fi CO2: Pe CO2/Fe CO2:
RR: Accuracy: Display update interval: CO2 waveform sweep speed: CO2 averaging: CO2 measurement stability: Resolution: Noise:
0 to 100 mmHg/0 to 13% 0 to 100 mmHg/0 to 13% 0 to 120 breaths/min
±5% of reading or ±2 mmHg, whichever is greater 2 sec Selectable 6.25, 12.5, or 25 mm/sec Selectable from single breath, 10 sec, or 20 sec Accuracy maintained over 8 hours 1 mmHg 2% of reading or 0.5 mmHg (maximum), whichever is greater
60 Hz interference:
<0.5 mmHg at 38 mmHg
Step response time:
Adult:
Neonatal
<60 ms (10-90%) <50 ms (10-90%)
Interference:
N2O gas:
±5% of reading or ±2 mmHg (maximum), whichever is greater, with N2O compensation enabled
O2 gas:
±5% of reading or ±2 mmHg (maximum), whichever is greater, with O2 compensation enabled
Barometric pressure:
Water vapor:
±2 mmHg (maximum) from 500 to 800 mmHg ±1.5% of reading or ±0.5 mmHg (maximum), whichever is greater
Anesthetic agent:
±0.5 mmHg (maximum) for concentration of no more than 5% of halogenated agents
Airway adapter variability:
±3% of reading or ±1.5 mmHg (maximum), whichever is greater, with same or different adapter; not applicable after adapter zero
Warm-up time:
Less than 15 seconds to initial CO2 indication, full specification within 120 seconds; waveform immediate upon power up
Calibration:
Factory settings:
Factory calibration settings stored in nonvolatile memory within the sensor; 15 second adaptor calibration when switching airway types
Verification:
Zero and span performance check with on-cable verifier
Respiration rate specifications:
Range (for 5% step size):
Accuracy:
Resolution:
0-120 breaths per minute ±1 breath per minute ±1 breath per minute
Barometric pressure sensor specifications:
Range: Accuracy:
425 to 817 mmHg (56 to 109 kPa) ±25 mmHg
Alarms: User-selectable upper and lower limits for CO2 and RR.
2-14 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Analog Output

Defibrillator Synchronization Pulse

EQUIPMENT OVERVIEW: Technical Specifications
ECG:
Gain: DC offset: Noise: Frequency response: Time delay:
Blood pressure:
Gain: DC offset: Noise: Frequency response: Time delay:
1 V/mV ±10% ±100 mV (max) <5 mVp-p (0-300 Hz)
0.05 Hz to 100 Hz +7/–0 Hz 40 ms monitoring filter, 35 ms diagnostic filter
10 mV/mmHg ±2% ±20 mV (max) <5 mVp-p (0-300 Hz) dc to 50 Hz +2/–0 Hz 40 Hz filter, 37 ms
Marker out:
Time delay: Amplitude (selectable in Service
menu):
+5 V selection:
+12 V selection: Pulse width: Output impedance:
Current limit:
Marker in:
Input threshold: Input hysteresis: Maximum input voltage: Input impedance: Pulse width:
35 ms (maximum), R-wave peak to leading edge of pulse.
3.5 V (min) at 1 mA sourcing; 0.5 V (max) at 5 mA sinking.
11.0 V (min) at 1 mA sourcing; 0.75 V (max) at 5 mA sinking. 10 ms ±10% or 100 ms ±10% (selectable in Service menu). 50
nominal
15 mA nominal, both sourcing and sinking.
V
= ±2.5 V (min); VIL = ±1.5 V (max)
IH
650 mV typical ±30 V (with respect to ground on pin 3) 10 k
(min) for -25 V < VIN < 25 V
1.0 ms (min), VIN > 2.5 V
Revision A Dash 3000/4000 Patient Monitor 2-15
2000966-035

Battery

Paper Recorder

EQUIPMENT OVERVIEW: Technical Specifications
Battery type: Exchangeable Lithium-Ion Number of batteries: 2 Battery weight: 0.36 kg (0.8 lbs) each Voltage: 11.1 V (nominal) Capacity: 3.9 Ah Charge time: Less than 4 hours each Run time: 4 to 5 hrs
Method: Thermal dot array Horizontal resolution: 480 dots/in at 25 mm/sec Vertical resolution: 200 dots/in

RF Wireless LAN

Number of waveform channels: 4 Paper width: 50 mm (1.97 in) Paper length: 30 m (100 ft) Paper speed: 0.1, 0.5, 1, 5, 10, 12.5, 25, and 50 mm/sec (± 2%)
Transmission technique: Frequency hopping spread spectrum Frequency: Country dependent, specific settings received from
access point. Within 2400 to 2500 MHZ range.
Frequency hopping characteristics:
Radio data rate: 1 and 2 Mbps Radio output power: 160 mW (including antenna gain) 1 Mbps range: Open environment: over 850 ft. (260)
2 Mbps range: Open environment: over 425 ft. (130)
Country dependent, specific settings received from access point. IEEE 802.11 compliant
Typical hospital environment: between 150 and 200 ft. (45 to 60 m)
Typical hospital environment: between 100 and 150 ft.
(30 to 45 m) Modulation: Binary GFSK Applicable standards: US: FCC Part 15 Class B
Europe: ETS 300 328 and ETS 300 826
2-16 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Environmental Specifications

EQUIPMENT OVERVIEW: Technical Specifications
NOTE: The system may not meet its performance specifications if
stored or used outside the manufaturer’s specified temperature and humidity range.
Power requirements:
90-132VAC
190-264 VAC Power consumption: 75 watts (fully loaded) Cooling: Convection Heat dissipation: 240 Btu/hr (max) Battery operation time:
General: Battery age will affect operating time. Operating conditions:
Ambient temperature:
While charging batteries:
Capnostat III sensor Relative humidity: Vibration: Altitude:
50/60 Hz 2.0A 50/60 Hz 1.0A
0 to 40°C (32 to 104°F) 0 to 35°C (32 to 95°F) 10° to 40° C (50° to 104°F) 5 to 95% at 40°C MIL-STD 810E, Method 514.4, Category 1
-610 to 4, 570 m (-2,000 to 15,000 ft)

Physical Specifications

Storage conditions (Do not exceed):
Maximum: Minimum: CO2 Sensor: Batteries:
Equipment Type: Portable per IEC 60601-1
Height:
Dash 3000 Dash 4000
Width:
Dash 3000 Dash 4000
Depth:
Dash 3000 Dash 4000
Weight (without batteries)
Dash 3000 Dash 4000
70°C (158°F) at 95% relative humidity –40°C (–40°F) at 15% relative humidity –30 to 65°C (–22 to 149°F) –20 to 60°C (–4 to 140°F)
26 cm (10.25 inches)
27.38 cm (10.78 inches)
28 cm (11.0 inches)
29.26 cm (11.5 inches)
20 cm (8 inches)
24.26 cm (9.55 inches)
5.08 kg (11.2 lbs)
5.53 kg (12.2 lbs)
Revision A Dash 3000/4000 Patient Monitor 2-17
2000966-035

Certification

Safety

EQUIPMENT OVERVIEW: Technical Specifications
UL 2601-1 classified. UL classified for CAN/CSA C22.2 No. 601.1 IEC 60601-1 and EN 60601-1 Certified CE marking for Council Directive 93/42/EEC concerning medical devices Radio and Telecommunication Terminal Equipment Directive

Electromagnetic Compatibility Compliance (EMC)

The Dash 3000/4000 system meets the requirements of EN 60601-1-2
(1993–04) Medical Electrical Equipment, Part 1: General Requirements for Safety, 2. Collateral Standard: Electromagnetic compatibility— Requirements and tests.
Exceptions
SpO2 Parameter — EN 60601-1-2 clause 36.202.1—IMMUNITY: Radiated Immunity:
The level of compliance is 1 volt per meter. If operating under the conditions defined in EMC Standard EN60601-1-2 (Radiated Immunity 3 volts per meter), field strength above 1 volt per meter may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
CO2 Parameter — EN 60601-1-2 clause 36.202.1—IMMUNITY: Radiated Immunity:
‘The level of compliance is 1 volt per meter. If operating under the conditions defined in EMC Standard EN60601-1-2 (Radiated Immunity 3 volts per meter), field strength above 1 volt per meter may cause waveform distortions and erroneous numeric data at various electromagnetic interference (EMI) frequencies.
Recommendations
Review the AAMI EMC Committee technical information report (TIR-18) titled Guidance on electromagnetic compatibility of medical devices for clinical/biomedical engineers - Part 1: Radiated radio-frequency electromagnetic energy. This TIR provides a means to evaluate and manage the EMI environment in the hospital.
The following actions can be taken:
managing (increasing) distance between sources of EMI and susceptible devices.
managing (removing) devices that are highly susceptible to EMI
lower power from internal EMI sources under hospital control (i.e. paging systems)
labeling devices susceptible to EMI
educate staff (nurses and doctors) to be aware of, and to recognize,
2-18 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Warranty

EQUIPMENT OVERVIEW: Technical Specifications
potential EMI related problems
Standard: One year. Other options are available. Contact your
sales representative for more information.
Revision A Dash 3000/4000 Patient Monitor 2-19
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications
2-20 Dash 3000/4000 Patient Monitor Revision A
2000966-035

3 INSTALLATION

Revision A Dash 3000/4000 Patient Monitor 3-1
2000966-035
For your notes
INSTALLATION:
3-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Connections

INSTALLATION: Connections

Back Panel Connections

ETHERNET

On the back of the monitor you will find all connectors for equipment and network.
AC Power Connector
Aux Port
Defib Sync Connector
004A
Peripheral Expansion Port
ETHERNET

RAC 2A Housing Connectors

The ETHERNET connector provides an ANSI/IEEE 802.3 10BaseT Ethernet standard interface to the Unity Network.
The RAC 2A module housing connects to the monitor via a standard category 5 patch cable (PN 418335-002) which plugs into the AUX port on the monitor and to the Auto Port on the back of the RAC 2A module housing.
The RAC 2A module housing does not have an Analog Output connector.
Power Switch
AC Power
Auto Port to the
monitor’s Aux Port
Async Comm
007A
Revision A Dash 3000/4000 Patient Monitor 3-3
2000966-035
INSTALLATION: Connections

Defib Sync

AC Pow er

The connector provides ECG analog output signals to user-supplied equipment.
CAUTION
Equipment damage. Connect all peripheral equipment before plugging the power cord into an AC outlet. Otherwise, connectors may be damaged.
Use this connector to apply power to the monitor. The monitor will be powered at all times when using AC power (there is no AC power switch). The monitor is preset at the factory for a specific AC voltage. Before applying power, be sure the power requirements match your power supply. Refer to the label on the back of the unit for the voltage and current requirements.
3-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INSTALLATION: Connections

Front Panel Indicators

AC/ Battery Power Indicators
Battery A and B Charge Status Indicators
009A
Dash 3000 Monitor’s Front Panel

AC Power Indicator

Battery Power Indicator

Power and battery indicators are located on the front panel of the monitor.
Battery A and Battery B Charge S tatus Indicators
Dash 4000 Monitor’s Front Panel
The indicator illuminates green when AC power is applied to the monitor. The indicator is not illuminated when the monitor is not powered.
The indicator illuminates yellow when the monitor is battery powered. The indicator is not illuminated when the monitor is not powered or when AC power is applied.
AC/Battery Power Indicator
053A

Battery Charging/Ready Indicators

Power Up

An icon for each battery pack indicates its charging status. The battery icon illuminates yellow when the respective battery is being charged. If both batteries are present and require charging, then both icons will illuminate even though they will be charged sequentially. The battery icon illuminates green when the respective battery is fully charged.
When the monitor is operating under battery power the battery icons will not be illuminated. The icons are also not illuminated when the respective battery is either not being charged, not installed, or has failed.
NOTE
No specific information is given to distinguish a failed battery pack condition from a condition where the battery is not installed or is not being charged.
After making all connections, plug the power cord into an AC wall outlet.
When all cables are properly connected, press the power button to turn the monitor on. All four front panel indicators will illuminate until the power-up sequence is complete. After approximately 10 seconds you should see a display on the screen.
Revision A Dash 3000/4000 Patient Monitor 3-5
2000966-035
INSTALLATION: Ethernet Communication

Ethernet Communication

Overview

Twisted Pair

Concentrator

Ethernet is a local area network used as the main link of the GE Medical Systems Information Technologies’ Unity network, a comprehensive
information communication system. The Unity network offers the high rate of communication of 10 megabits per second. The Ethernet connector connects to an Ethernet transceiver directly or via a transceiver cable. This local area network links all patient monitors, central stations, and other GE Medical Systems Information Technologies’ equipment throughout the hospital. Depending on the construction of the hospital, thick-net, thin-net, or twisted pair cabling is used.
Twisted pair is the most popular cabling because it is easy to install and flexible to work with. It uses the star topology with a concentrator as the hub of the segment. Each of the network devices is connected directly to the concentrator so longer lengths of cable are required. A maximum of 100 meters or 328 feet is the longest length of twisted pair cable used. The number of devices is limited to the amount of connectors at the concentrator.
The concentrator is simply a transceiver that passes all network data between any two branches in the LAN. Note that the concentrator passes all network data between the two branches, regardless of whether or not one node is sending data to another node on the same branch.
To implement the star topology, each network device is connected to a concentrator. The concentrator functions as a central hub and simply passes all network data between each network device in the star segment. Typically, the concentrator supports 8 to 12 network devices and may be linked to other concentrators to form larger networks.
3-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
INSTALLATION: Ethernet Communication

Node

Segment and Branch

Each network device or node is assigned an address number and requires a transceiver to interface between the network device and the network. For thick-net and thin-net cabling a transceiver and a serial drop cable connects to the main trunk. The serial drop cable is sometimes referred to as an AUI (attachment unit interface) transceiver cable. For twisted pair cabling, the transceiver to connected directly to the network device.
Some Ethernet systems are comprised of smaller, stand-alone Ethernet systems (called branches or segments) that are connected by bridges, concentrators, or repeaters. Many nodes on the Ethernet network may be serviced by one segment or branch. Each segment may support many patient monitors, central stations, and auxiliary devices.
For example, one segment may connect all the patient monitors and central stations in the ICU (Intensive Care Unit) and another may connect the monitoring system in the CCU (Critical Care Unit). Each segment could be a fully-functioning stand-alone system if they were not connected to each other. However, with a bridge or repeater to connect the ICU (one segment) with the CCU (the other segment), information can pass between any of the nodes (patient monitors and central stations) on either branch similar to a patient transfer from one unit to another.
A section is a single length of twisted pair cable with a RJ-45 connector on each end. A section goes from one twisted pair transceiver to the concentrator. A segment is comprised of all the sections of twisted pair cable connected in a star formation to one concentrator.

Repeater

A repeater is used to extend the length of cabling when the distance required exceeds the length of the cable specifications. It is simply a transceiver that passes all network data between any two segments. Note that the repeater passes all network data between the two segments, regardless of whether or not the one node is sending data to another node on the same segment.
Revision A Dash 3000/4000 Patient Monitor 3-7
2000966-035
INSTALLATION: Ethernet Communication

Bridge

Twisted Pair Cabling (10BaseT)

A bridge is more selective than a repeater with the data that it passes between segments. It also acts as a transceiver between two segments, but it only passes signals if a node on one of the segments is attempting to communicate with a node on the other segment. Since the majority of communication on the network occurs within a single segment, the bridge does not pass all of the data from one segment to the other. This lowers the amount of data traffic passing between segments, and makes the network more efficient than a system that is connected with repeaters.
Twisted pair is an IEEE 802.3 local area network that uses flat and small diameter cable containing four pairs of twisted wires to connect devices. Twisted pair operates at the same speed as thin-net and thick­net (10 megabits/second), but the cable distances extended up to 100 meters (328 feet).
A twisted pair transceiver passes data back and forth between the network device and the LAN. It is attached directly to the network device at the at the 15-pin D-type connector. The twisted pair cable is connected from the RJ-45 connector at the transceiver and the RJ-45 connector at the concentrator.
NOTE: Some devices (like Octacomm/Solar 8000M patient monitor)
have 10BaseT standard meaning that the RJ-45 connector is part of the product and the twisted pair transceiver is not required.

Symbol PC Card (Wireless LAN)

The Symbol PC card, installed in the monitor, uses a 2.4 GHz frequency band and a Frequency Hopping spread spectrum (FHSS). The Frequency Hopping spread spectrum meets IEEE 802.11 standards.
Two diversity antennas, installed in the handle of the monitor, radiates the RF energy through the air to a Symbol Access Point. The Symbol Access Point also uses a 2.4 GHz frequency band and the Frequency Hopping spread spectrum.
NOTE: Refer to the Wireless LAN (Symbol Access Point) Installation
and Service Manual for detailed information on the Symbol Access point.
NOTE: The following is required for the monitor to roam from access
point to access point while maintaining Wireless LAN communication with the Unity Network.
Installation of the RF LAN (Wireless LAN) option.
Configuration and verification of the monitor’s RF LAN operation.
Activation of the RF LAN by disconnecting the monitor’s Ethernet cable.
3-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035

4 MAINTENANCE

Revision A Dash 3000/4000 Patient Monitor 4-1
2000966-035
For your notes
MAINTENANCE:
4-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Maintenance Schedule

Maintenance Schedule

Manufacturer Recommendations

To ensure the monitor is always functional when required, qualified service personnel should perform the following regular maintenance.
Visual Inspection: Perform a visual inspection upon receipt of the equipment, every 12 months thereafter, and prior to servicing the unit.
Cleaning: Clean the unit upon receipt of the equipment, every 12 months thereafter, and each time the unit is serviced.
Conditioning the Batteries: Condition the batteries once every two months or as needed.
Calibrating the NBP, BP, ECG, and End-tidal CO2 Software: Calibrate the software upon receipt of the equipment, every 12 months thereafter, and each time the unit is opened for service.
Electrical Safety Tests: Perform safety tests upon receipt of the equipment, every 12 months thereafter, and each time the unit is serviced.
Checkout Procedure: Perform the checkout upon receipt of the equipment, every 12 months thereafter, and each time the unit is serviced.
Clearing the Stored Patient Data Memory: Admit and discharge a test patient every 12 months to clear the monitor’s stored patient data memory.

Manufacturer Responsibility

CAUTION
Failure on the part of all responsible individuals, hospitals or institutions, employing the use of this device, to implement the recommended maintenance schedule may cause equipment failure. The manufacturer does not, in any manner, assume the responsibility for performing the recommended maintenance schedule, unless an Equipment Maintenance Agreement exists. The sole responsibility rests with the individuals, hospitals, or institutions utilizing the device.
Revision A Dash 3000/4000 Patient Monitor 4-3
2000966-035

Visual Inspection

MAINTENANCE: Visual Inspection
The monitor and it’s components should be carefully inspected prior to installation, once every 12 months thereafter and each time the equipment is serviced.
Carefully inspect the equipment for physical damage to the case, the display screen, and the keypad. Do not use the monitor if damage is determined. Refer damaged equipment to qualified service personnel.
Inspect all external connections for loose connectors or frayed cables. Have any damaged connectors or cables replaced by qualified service personnel.
Inspect the display face for marks, scratches, or other damage. Physical damage to a CRT display face may pose an implosion hazard. Have the CRT replaced by qualified service personnel if necessary.
Safety labels and inscription on the device are clearly legible.
4-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Cleaning

MAINTENANCE: Cleaning

Cleaning Precautions

Use one of the following approved solutions:
Cidex solution, or
Sodium hypochlorite bleach (diluted), or
Mild soap (diluted)
Lint-free cloth
Dust Remover (compressed air)
To avoid damage to the equipment surfaces, never use the following cleaning agents:
organic solvents,
ammonia based solutions,
acetone solution,
alcohol based cleaning agents,
Betadine solution,
a wax containing a cleaning substance, or
abrasive cleaning agents.

Cleaning the Display

Exterior Cleaning

To clean the display, follow the recommendations of the display’s manufacturer. In general you will need to use a soft, clean, lint-free cloth dampened with a glass cleaner.
CAUTION
To avoid getting liquid into connector openings, do not spray glass cleaning or general cleaning solutions directly onto the product’s surface.
Clean the exterior surfaces with a clean, lint-free cloth and one of the cleaning solutions listed in the table above.
Wring the excess solution from the cloth. Do not drip any liquid into open vents, switches, plugs, or connectors.
Dry the surfaces with a clean cloth or paper towel.
Revision A Dash 3000/4000 Patient Monitor 4-5
2000966-035
MAINTENANCE: Cleaning

Cleaning the Print Head

Materials Required

Procedure

PRINTHEAD
PAPER DRIVE ROLLER
Heavy usage causes debris to build up on the print head. This build can cause the printed images to appear distorted. It is recommended that this procedure be performed when necessary, depending on usage.
A nonabrasive material/cloth and isopropyl alcohol are all that are necessary to perform this procedure.
This procedure should be performed in the order listed.
1. Disconnect the power cord from the mains source.
2. Open the writer door to expose the print head.
3. Remove paper roll.
4. Locate print head shown in figure at left. A flashlight may help illuminate the print head for closer examination.
5. Wipe print head with alcohol and a nonabrasive material/cotton swab in an side to side motion. Continue wiping until the cloth/swab wipes clean.
6. Wipe paper drive roller clean of any bits of paper and debris with alcohol and a nonabrasive material.
4-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Battery Maintenance

Battery Maintenance

Charging

Conditioning the Batteries

Frequency Guidelines

The battery is charged whenever the monitor is connected to AC power, regardless whether the monitor is turned on or turned off.
A battery conditioning cycle occurs when the following has been completed.
1. The battery is fully charged without interruption.
2. The battery is discharged until the monitor shuts down.
3. The battery is charged until the battery status light turns green in color.
To maintain useful life, use the following guidelines to condition a battery:
Once every two months,
When the run time of the battery becomes noticeably shorter,
When the predicted run times become noticeably inaccurate, or
When the associated battery is requesting a conditioning cycle (i.e., the CHECK BATT STATUS error message is displayed in the ECG waveform area and CONDITION is displayed for BATTERY QUALITY in the Battery Status information window).

Recommendations

Procedure

Conditioning a battery is best done on an external charger (see instructions included with the charger). However, a conditioning cycle can also be run on the monitor.
To condition a battery on the monitor, follow this procedure:
1. Disconnect the monitor from the patient and remove it from service.
2. Insert the battery in need of conditioning in one of the battery slots in the monitor, and leave the other slot EMPTY.
3. Apply AC power to the monitor and allow the battery to charge uninterrupted until the Charging Status indicator on the front panel turns green.
4. Remove AC power and allow the monitor to run from the battery until it shuts off.
5. Apply AC power again to the monitor and allow the battery to charge uninterrupted until the Charging Status indicator on the front panel turns green.
6. This battery is now conditioned and the monitor can be returned to service.
Revision A Dash 3000/4000 Patient Monitor 4-7
2000966-035
MAINTENANCE: Battery Maintenance

Replacing the Batteries

1. Open the battery door. The battery door is on the left side of the monitor, along the bottom.
2. In the middle is a retainer. Turn this away from the battery you are replacing.
3. Remove the faulty batteries.
Retainer
4. Replace with a new battery. The monitor uses two exchangeable lithium-ion batteries. Install the battery with the connection pins facing down and inserted first.
5. Close the battery cover. The retainer needs to be straight up for the door to close.
6. Verify that the monitor operates correctly:
Confirm that the Battery IDs with a battery icon displays in the
lower right corner of the monitor.
Verify that the Battery LEDS illuminate either green or amber.

Disposal

4-8 Dash 3000/4000 Patient Monitor Revision A
When the battery no longer holds a charge, it should be replaced. The batteries are recycleable. Remove the old battery from the monitor and follow your local recycling guidelines.
WARNING
Explosion Hazard. DO NOT incinerate the battery or store at high temperatures.
2000966-035
MAINTENANCE: Electrical Safety Tests

Electrical Safety Tests

General

Recommendations

Electrical safety tests provide a method of determining if potential electrical health hazards to the patient or operator of the device exist.
These instructions are intended for every component in the system. If the Tram-rac housing does not have its own power supply, it should remain connected to the monitor throughout the safety tests.
Record the date and results on the “Maintenance/Repair Log” included at the end of this chapter.
GE Medical Systems Information Technologies recommends that you perform all safety tests presented in this chapter:
upon receipt of the device (monitor and its associated equipment),
every twelve months thereafter, and
each time the main enclosure is disassembled or a circuit board is removed, tested, repaired, or replaced.
CAUTION
Failure to implement a satisfactory maintenance schedule may cause undue equipment failure and possible health hazards. Unless you have an Equipment Maintenance Contract, GE Medical Systems Information Technologies does not in any manner assume the responsibility for performing the recommended maintenance procedures. The sole responsibility rests with the individual or institution using the equipment. GE Medical Systems Information Technologies service personnel may, at their discretion, follow the procedures provided in this manual as a guide during visits to the equipment site.

Test Conditions

Revision A Dash 3000/4000 Patient Monitor 4-9
Electrical safety tests may be performed under normal ambient conditions of temperature, humidity, and pressure.
2000966-035
MAINTENANCE: Electrical Safety Tests

Test Equipment

The manufacturer recommended test equipment required to perform electrical safety tests is listed below. Equivalent equipment may be substituted as necessary.
Required Tools/Special Equipment
Item Part Number
Leakage Current Tester
120 V (or equivalent) 240 V (or equivalent)
Multimeter
DALE 600 DALE 600E
4-10 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests

Wall Receptacle Test

Ground (Earth) Integrity

Before starting the tests, the wall receptacle from which the monitoring device will get electrical power must be checked. This test checks the condition of the wall receptacle to ensure correct results from leakage tests.
For international wall receptacles, refer to the internal standards agencies of that particular country. Use a digital multimeter to ensure the wall receptacle is wired properly.
If other than normal polarity and ground is indicated, corrective action must be taken before proceeding. The results of the following tests will be meaningless unless a properly wired wall receptacle is used.
Listed below are two methods for checking the ground (earth) integrity,
“Ground Continuity Test” and “Impedance of Protective Earth Connection.” These tests determine whether the device's exposed metal and power inlet's earth (ground) connection has a power ground fault
condition.

Ground Continuity Test

Perform the test method below that is required by your Country/Local governing safety organization.
Completion of this test is checked by the following steps:
1. Disconnect the DUT (device under test) from the wall receptacle.
2. Connect the negative(-) lead of the ohm meter to the protective earth terminal (ground pin in power in-let connector) or the protective earth pin in the MAINS PLUG (ground pin in power cord). Refer to the US 120Vac power cord figure on the left.
3. Set the Ohm meter to the milliohm (mΩ) range.
4. Connect the positive (+) lead of the Ohm meter to all exposed metal surfaces on the DUT. If the metal surfaces are anodized or painted scrape off a small area in a inconspicuous area for the probe to make contact with the metal.
5. Resistance should read to pass:
0.1 ohm or less without power cord
0.2 ohms or less with power cord
Revision A Dash 3000/4000 Patient Monitor 4-11
2000966-035
MAINTENANCE: Electrical Safety Tests

Impedance of Protective Earth Connection

This test unlike a ground continuity test will also stress the ground system by using special ground bond testers i.e. Kikusui (model 872 or
TOS 6100) or Associated Research model HYAMP® Jr. Model 3030D.
This test normally is only required as a manufacturing production test to receive safety agency compliance (i.e. IEC601-1).
Some country agency's do require this test after field equipment repairs (i.e. Germany's DIN VDE 0751 standards).
Consult your country/local safety agency if in question.
Compliance is checked by the following steps:
1. A current not less than 10A and not exceeding 25A from a current source with a frequency of 50 or 60 Hz with a no-load voltage not exceeding 6 V is passed for at least 5 s through the protective earth terminal or the protective earth pin in the mains plug and each accessible metal part which could become live in case of failure in basic insulation.
2. The voltage drop between the parts described is measured and the impedance determined from the current and voltage drop. It shall not exceed the values indicated.
for equipment without a power supply cord the impedance between the protective earth terminal and any accessible metal part which is protectively earthed shall not exceed 0.1 ohms
For equipment with a power supply cord the impedance between the protective earth pin in the mains plug and any accessible metal part which is protectively earthed shall not exceed 0.2 ohms.
When taking this measurement move the unit's power cord around, no fluctuations in resistance should be observed.
4-12 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests

Ground (Earth) Wire Leakage Current Tests

Perform this test to measure current leakage through the ground (earth) wire of the equipment during normal operation.
1. Set the leakage tester switches as follows:
Selector knob - 1,
GND switch - OPEN,
Polarity switch - NORM,
•Power switch - OFF.
2. Connect the DMM to the METER jacks on the leakage tester. Set the DMM to measure AC millivolts.
3. Connect the power cord of the device under test to the power receptacle on the rear of the leakage tester.
NOTE: The device under test is to be tested at its normal operating
voltage.
4. Set the leakage tester power switch to ON.
5. Set the power switch of the device under test to ON.
6. Read the current leakage indicated on DMM. If the reading is greater than the appropriate specification below, the device under test fails and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under
test is powered from 100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is
powered from a centered-tapped 200-240 V/50-60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is
powered from a non-center-tapped, 200-240 V/50-60 Hz, single­phase circuit
NOTE: Center-tapped and non-center-tapped circuits produce different
leakage currents and the UL and IEC limits are different.
Revision A Dash 3000/4000 Patient Monitor 4-13
2000966-035
MAINTENANCE: Electrical Safety Tests
7. Set the polarity switch on the leakage tester to RVS (reverse).
8. Read the current leakage indicated on DMM. If the reading is greater than the appropriate specification below, the device under test fails and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under
test is powered from 100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is
powered from a centered-tapped 200-240 V/50-60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is
powered from a non-center-tapped, 200-240 V/50-60 Hz, single­phase circuit
NOTE: Center-tapped and non-center-tapped circuits produce different
leakage currents and the UL and IEC limits are different.
9. Set the leakage tester power switch to OFF.
NOTES:The MD (measuring device) is the circuitry defined by the
appropriate standard for measuring leakage current.
The measuring devices, defined by various standard organizations (IEC, UL, etc.), produce almost identical test measurement results.
4-14 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests

Enclosure Leakage Current Test

Perform this test to measure current leakage through exposed conductive surfaces on the device under test during normal operation.
1. Set the leakage tester switches as follows:
Selector knob - 2,
GND switch - OPEN, and
Polarity switch - NORM.
2. Connect a meter lead between the CHAS connector on the rear of the leakage tester and an unpainted, non-anodized chassis ground on the unit under test.
3. Set the leakage tester power switch to ON.
4. Read the current leakage indicated on DMM. If the reading is greater than the appropriate specification below, the device under test fails and should be repaired and tested again.
300 microamperes (0.3 volts on the DMM), and the device under
test is powered from 100-120 V/50-60 Hz
300 µA (0.3 volts on the DMM), and the device under test is
powered from a centered-tapped 200-240 V/50-60 Hz, single phase circuit
500 µA (0.5 volts on the DMM), and the device under test is
powered from a non-center-tapped, 200-240 V/50-60 Hz, single­phase circuit
NOTE: Center-tapped and non-center-tapped circuits produce
different leakage currents and the UL and IEC limits are different.
Revision A Dash 3000/4000 Patient Monitor 4-15
2000966-035
MAINTENANCE: Electrical Safety Tests
5. Set the polarity switch to RVS and observe the same meter readings as in the previous step.
6. Set the GND switch on the leakage tester to CLOSED.
7. Read the current leakage indicated on DMM. If the reading is greater than the appropriate specification below, and the device under test is powered from 100-240 V/50-60 Hz, the device under test fails and should be repaired and tested again.
100 microamperes (0.1 volts on the DMM), and the device under
test is powered from 100-240 V/50-60 Hz
8. Set the polarity switch to RVS and observe the same meter readings as in the previous step.
9. Set the leakage tester power switch to OFF and remove the meter lead connected in step 2.
4-16 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Patient (Source) Leakage Current Test

MAINTENANCE: Electrical Safety Tests

Test Equipment

Equipment required to perform these tests is listed below. Equivalent equipment may be substituted as necessary.
Name Manufacturer Par t Numbe r
ECG Test Body GE Medical Systems
Information Technologies
Temp/CO Test Body GE Medical Systems
Information Technologies
This procedure only applies to Class I (grounded/earthed) equipment, and measures the leakage current from the ECG and TEMP/CO connectors of the device to ground.
1. Set leakage tester switches as follows:
Selector knob – 3,
GND switch – GND OPEN,
Polarity switch – NORM,
Power switch – OFF.
2. Connect an ECG test body to the ECG connector of the DUT.
3. Connect a short length of cable between the ECG test body installed in the last step and the jacks on the top of the leakage tester.
MT-3387
MT-3644
4. Set the leakage tester power switch to ON.
5. Set the rear panel power switch of the device to ON.
6. Read the leakage current indicated on the DMM.
If the reading is greater than 50 µA (0.05 volts on the DMM), the device under test fails this test and should be repaired and tested again.
NOTE: The AAMI and IEC single fault condition (ground open) is 50
µA, whereas the normal condition (ground closed) is less.
7. Change the leakage tester polarity switch to the RVS position.
8. Read the leakage current indicated on the DMM.
If the reading is greater than 50 µA (0.05 volts on the DMM), the device under test fails this test and should be repaired and tested again.
NOTE: The AAMI and IEC single fault condition (ground open) is 50
µA, whereas the normal condition (ground closed) is less.
Revision A Dash 3000/4000 Patient Monitor 4-17
2000966-035
MAINTENANCE: Electrical Safety Tests
9. Change the GND switch to the CLOSED position.
GND
10. Read the leakage current indicated on the DMM.
11. If the reading is greater than 10 µA (0.01 volts on the DMM), the
device under test fails this test and should be repaired and tested again.
12. Change the leakage current switch to the RVS position.
13. Read the leakage current indicated on the DMM.
14. If the reading is greater than 10 µA (0.01 volts on the DMM), the device under test fails this test and should be repaired and tested again.
15. Set the power switch of the leakage tester to OFF.
16. Repeat all previous steps for the TEMP/CO connector using the appropriate test body.
4-18 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests
Patient (Sink) Leakage Current Test
(Mains Voltage on the Applied Part)
This procedure only applies to Class I (grounded/earthed) equipment, and measures the leakage current from a mains voltage source into the ECG and TEMP/CO connectors.
1. Set the leakage tester switches as follows:
Selector knob – 5,
GND switch – CLOSED,
Polarity switch – NORM.
2. Disconnect the test cable from the leakage tester PATIENT JACKS (TOP) and reconnect it to the PATN JACK connector on the front panel of the leakage tester.
WARNING
The following step will cause high voltage (120 VAC to 240 VAC) to appear at the PATN JACK on the leakage tester. Do not touch the PATN JACK posts or ECG lead clips during this test as an electrical shock will occur.
3. Set power switch on the leakage tester to ON.
4. Read leakage current indicated on DMM.
If the reading is greater than the appropriate specification below, the device under test fails this test and should be repaired and tested again.
10 µA, (0.01 volts on the DMM) at 120 VAC without the patient
cable.
20 µA (0.02 volts on the DMM) at 240 VAC without the patient
cable.
NOTE: The 10 and 20 µA limit are based on internal design
standards.
50 µA (0.05 volts on the DMM) at 120-240 VAC with the patient
cable.
NOTE: The 50 µA limit is common to all standards. AAMI ES-1
standard requires using the patient cable.
5. Change the leakage tester polarity switch to the RVS position.
Revision A Dash 3000/4000 Patient Monitor 4-19
2000966-035
MAINTENANCE: Electrical Safety Tests
6. Read the leakage current indicated on the DMM.
If the reading is greater than the appropriate specification below, the device under test fails this test and should be repaired and tested again.
10 µA (0.01 volts on the DMM) at 120 VAC without the patient
cable.
20 µA (0.02 volts on the DMM) at 240 VAC without the patient
cable.
NOTE: The 10 and 20 µA limits are based on internal design
standards.
50 µA (0.05 volts on the DMM) at 120-240 VAC with the patient
cable.
NOTE: The 50 µA limit is common to all standards. AAMI ES-1
standard requires using the patient cable.

Test Completion

GND
7. Set the power switch on the leakage tester to OFF.
8. Repeat all previous steps for the TEMP/CO connector using the appropriate test body.
Disconnect all test equipment from the device. Disconnect the device power cord plug from the leakage tester power receptacle. Disconnect the leakage tester from the wall receptacle.
4-20 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests

Hi-Pot (Dielectric Withstand) Test

Recommendations

The high potential (Hi-Pot) tests provide a method of checking patient isolation circuits and protect patients connected to the device under test from potential electrical health hazards. These tests are recommended for direct patient-connected medical devices to check the integrity of the patient isolation circuitry after any isolated component in the device has been repaired.
The manufacturer recommends that hi-pot tests be performed whenever a circuit board in the patient-isolated DAS assembly of the device under test is removed, repaired, or replaced.
WARNING
Failure to perform hi-pot tests may cause undue equipment failure and possible health hazards. The manufacturer does not in any manner, unless an Equipment Maintenance Agreement exists, assume the responsibility for performing these recommended hi-pot tests. The sole responsibility rests with the individuals, hospitals or institutions utilizing this equipment. Manufacturer service representatives may, at their discretion, use this procedure as a helpful guide during visits to the equipment site.
Test Conditions Test Equipment

Preparation

These tests may be performed under normal ambient conditions of temperature, humidity, and pressure.
Equipment required to perform these tests is listed below. Equivalent equipment may be substituted as necessary.
Name Manufacturer Part Number
AC/DC Hi-Pot Generator
ECG Test Body GE Medical Systems
Information Technologies
Aux/Ethernet Test Body GE Medical Systems
Information Technologies
Temp/CO Test Body GE Medical Systems
Information Technologies
Power Cord Hi-Pot Body GE Medical Systems
Information Technologies
Follow these steps in the same order in which they are listed.
1. Set up the AC/DC Hi-Pot Generator in the following manner:
Power switch – ON,
VOLTAGE RANGE selector – MEDIUM (10 kVA),
RAISE VOLTAGE selector – 0 volts,
OUTPUT & CURRENT selector – 2 mA range, and
Allow the tester to warm up for 15 minutes before continuing
with this test.
Hipotronics AD125
MT-3387
MT-5265
MT-3644
MT-4542
2. Connect the ground pin on the power cord connector of the device under test to the ground of the AC/DC Hi-Pot Generator.
Revision A Dash 3000/4000 Patient Monitor 4-21
2000966-035
MAINTENANCE: Electrical Safety Tests

DAS Assembly AC Hi-Pot Test

Perform the AC hi-pot tests for both the ECG and Temp/CO side panel connectors of the device under test. This only needs to be performed when the DAS assembly is opened.
CAUTION
Never attempt to perform this test on any of the other front panel connectors of the device under test. Damage to the device under test may occur if this test is performed on any of the other front panel connectors.
1. Attach the black lead from the Hi-Pot generator to the ground prong of the power cord.
2. Install an ECG dead body plug (pn MT-3387) and the Temp/CO dead body plug (pn MT-3644) on the side panel. Connect the red high voltage lead from the Hi-Pot generator to the exposed lead of the ECG dead body plug. Jumper the exposed lead of the Temp/CO dead body plug to the black high voltage lead. Set the current limit to 1 mA on the Hi-Pot generator.
NOTE: During this test, watch the analog meter to ensure the
current level never exceeds 1 mA. If it does, the unit has failed the test and must be repaired then tested again.
WARNING
The following steps cause high voltage (4000 V AC) to appear at the test body.
3. Set the voltage switch to AC and the scale to 10KV. Turn ON the Hi­Pot generator and bring up the voltage to 4000 VAC RMS for a period of 60 seconds. The breakdown warning lamp or buzzer must not activate. Turn OFF the Hi-Pot generator.
4. Connect the red high voltage lead from the Hi-Pot generator to the exposed lead of the Temp/CO dead body plug. Jumper the exposed lead of the ECG dead body plug to the black high voltage lead. Repeat step three.
5. If the device under test fails, repairs must be made and the unit must be tested again.
6. This completes the AC hi-pot test for the DAS assembly.
4-22 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Electrical Safety Tests

Processor/Power Management PCB Hi-Pot Test

This test is only required when you repair the isolated Ethernet or Aux circuitry. This test pertains to the Ethernet and Aux ports on the rear of the unit.
CAUTION
Never attempt to perform this test on any of the other rear panel connectors of the monitor. Damage to the monitor may occur if this test is performed on any of the other rear panel connectors.
1. Attach the black lead from the Hi-Pot generator to the ground prong of the power cord.
2. Install the Dash AUX/Ethernet Hi-Pot test body (pn MT-5265) to the AUX and Ethernet connectors on the rear of the monitor. Connect the high voltage lead from the Hi-Pot generator to the exposed lead of the Ethernet port test body. Jumper the exposed lead of the AUX connector test body to the black high voltage lead.
NOTE: During this test, watch the analog meter to ensure the
current level never exceeds 1 mA. If it does, the unit has failed the test and must be repaired and tested again.
WARNING
The following step can cause high voltage (1500 VAC) to appear at the test body.
3. Set the voltage switch to AC and the scale to 10KV. Turn ON the Hi­Pot generator and bring up the voltage to 1500 VAC RMS for a period of 60 seconds. The breakdown warning lamp or buzzer must not activate. Turn OFF the Hi-Pot generator.
4. Connect the high voltage lead from the Hi-Pot generator to the exposed lead of the AUX test body. Jumper the exposed lead of the Ethernet lead dead body to the black high voltage lead. Repeat step three.
5. If the unit under test fails, repairs must be made and the unit must be tested again.
6. This completes the processor/power management PCB Hi-Pot test.
Revision A Dash 3000/4000 Patient Monitor 4-23
2000966-035
MAINTENANCE: Electrical Safety Tests

AC Mains Hi-Pot Test

Perform the following steps to hi-pot the AC mains dielectric relative to ground. This applies only to actual board repair!
1. Set up the AC/DC Hi-Pot Generator in the following manner:
VOLTAGE RANGE selector - MEDIUM (10 kVA),
RAISE VOLTAGE selector - 0 volts,
OUTPUT & CURRENT selector - 5 mA range,
Power switch - ON, and
allow the tester to warm up for 15 minutes before continuing
with this test.
2. Connect the ground pin on the power cord connector of the monitor to the BLACK ground of the AC/DC Hi-Pot Generator.
3. Connect the LINE and NEUTRAL to the RED lead of the hi-pot test.
WARNING
To avoid electric shock by accidently shorting line voltage to ground, make and use a receptacle adapter that connects the LINE voltage and NEUTRAL together separate from any ground potential.
4. Slowly turn the RAISE VOLTAGE selector to 1500 volts.
5. Wait for 60 seconds. If the breakdown warning lamp illuminates or the buzzer activates before the time expires, then the unit has failed the test and should be repaired then tested again.
6. Slowly turn the RAISE VOLTAGE selector to 0 volts.
7. Set the HIGH VOLTAGE switch to OFF. The high voltage indicator should turn off.
8. If the unit under test fails, repairs must be made and the unit must be tested again.
This completes the AC mains hi-pot test.
4-24 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures

Checkout Procedures

These checkout procedures provide service personnel with a method to verify operational and functional performance of the monitor. Failure to attain any of the listed results indicates a potential malfunction of the monitor.
Perform the checkout procedures when you receive the monitor, every twelve months thereafter, and each time you service the unit.
The checkout procedures are based on the assumption that the tested monitor has known good cables and test equipment. It also requires that the user be familiar with the operation of all test equipment required for the checkout procedures. For more information concerning the operation of these components, refer to the respective operator manual(s).

Manufacturer Recommended Test Equipment

The following table lists GE Medical Systems Information Technologies’ recommended test equipment, adaptors, and cables you need to successfully complete the checkout procedures. The checkout procedures are written for the test equipment in the following table. If you use test equipment other than those GE Medical Systems Information Technologies recommends, you may need to slightly modify some test steps.
Description Part Number Qty
Multifunction Micro-simulator MARQII 1 Cardiac Output Simulator II 900028-001 1 Patient cable, 5-leadwire, AHA 403061-001 1 Leadwire Set, 5-Leadwire, AHA 403066-005 1 BP Adapter 700095-001 2 Temperature Adaptor 402015-004 1 TEMP-to-Simulator Cable 6770031 1 CO Adaptor 900028-001 1 SpO
Simulator 408610-001 1
2
SpO
Simulator Cable, Nellcor 700232-004 1
2
Revision A Dash 3000/4000 Patient Monitor 4-25
2000966-035
MAINTENANCE: Checkout Procedures

Monitor Power-up Tests

1. Remove the batteries and unplug the monitor from AC power to turn it off.
2. Restore the batteries to the monitor and plug the monitor into AC power to turn it on.
3. Verify all four front panel indicators illuminate on power up.
4. Verify the AC indicator stays illuminated.
NOTE: If the AC LED stays on, but the screen is blank, the monitor
is likely in “standby mode” (battery charging). Press the POWER button to enter the normal mode.
If the AC indicator is on, continue with the tests.
If either of the CHARGING STATUS indicators is yellow, wait
for the battery(ies) to fully charge and the indicators to illuminate green. The batteries may take up to four hours to charge.
If the battery “fuel gauge” displays the word “ERROR,” the
battery may be asleep. See “Error Message” on page 5-11.
5. Verify the optional alarm indicator lights both red and amber on power up.
6. Verify an audio “Beep” sounds at the end of Boot up.
7. Test all of the front panel keys and the Trim Knob control. Verify that an audio “Beep” sounds after each key press.
8. Check battery power for both batteries.
Pull the AC plug and open the battery door. Verify one LED in
the battery compartment is on (batteries must have more than 10% charge).
Pull that battery out and verify the other LED lights, thus
indicating the unit is powered by the other battery.
Reinstall battery and plug in monitor.
4-26 Dash 3000/4000 Patient Monitor Revision A
2000966-035

ECG Tests

0
MAINTENANCE: Checkout Procedures
15B

5 Lead ECG Test

Perform this test if the monitor acquires 5 lead ECG data.
1. Set up the patient simulator as follows.
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
5-lead ECG patient cable properly attached.
2x gain for MARQI or MARQII simulator.
2. Attach the ECG patient cable and ECG leadwire set to the ECG/ RESP connector on the monitor and the leadwire connectors on the top of the patient simulator.
3. Admit the patient simulator to the monitor.
4. Observe the following:
ECG lead II is displayed and is noise-free,
Heart rate of 80 ±1 bpm is displayed,
With QRS tones enabled, an audible tone sounds with each R-
Wave (QRS complex).
5. Verify all six ECG leads are available to view and are noise-free.
6. Select DETECT PACE and set to PACE 2.
7. Select the VP2 pacemaker pulse on the simulator.
8. Observe the following while you view ECG leads I, II, III, aVL, aVF, and V5:
a “P” appears above the PVC count indicating pacemaker pulse
detection is enabled, and
the heart rate still reads 80 ±1 bpm.
Revision A Dash 3000/4000 Patient Monitor 4-27
2000966-035
MAINTENANCE: Checkout Procedures
9. Disable pacemaker pulse detection on the monitor and return the simulator to these conditions:
Heart rate – 80 bpm,
Heart rate amplitude – 1.0 mV,
5-lead ECG patient cable properly attached.
10. Select ECG lead II to view in the top trace position on the monitor display.
11. Disconnect the RA leadwire from the patient simulator.
12. Observe that:
a RA FAIL message appears on the display, and
lead III automatically displays in place of lead II in the top trace
position.
13. Reconnect the RA leadwire to the patient simulator.
14. Inject a 1-millivolt calibration signal using the patient simulator and start a manual graph.
15. Observe that the calibration pulse properly displays and graphs.

12SL and ACI-TIPI ECG Test

016A
16. This completes the 5 Lead ECG test. Continue to the next steps of these checkout procedures.
Perform this test if your monitor uses the 12SL ACI-TIPI ECG option.
1. Set up the patient simulator as follows:.
Heart rate – 80 bpm.
Heart rate amplitude – 1.0 mV.
12SL ECG patient cable (5-leads with V leads) properly attached.
2x gain for MARQI or MARQII simulator.
2. Select ECG from the monitor menu. Then, select 12 Lead ECG Analysis.
3. Verify that the monitor is displaying 10 noise-free leads.
4. Select 12LD ECG Now. Wait for the monitor to acquire and analyze the data.
5. Select Transmit-Print.
6. Verify the 12SL ECG prints at the graph location assigned in the
monitor’s Graph Setup -->12SL Graph Location.
If there is no graph location is assigned, an error message
appears on the bottom of the monitor’s display.
4-28 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
7. Verify the ECG is transmitted to the MUSE Cardiovascular Information System. Verify the ECG prints out correctly as defined by the MUSE system.
If no MUSE system is connected , an error message appears on
the bottom of the monitor’s display.
8. Delete this test 12SL ECG from the MUSE system’s edit list.
Revision A Dash 3000/4000 Patient Monitor 4-29
2000966-035
MAINTENANCE: Checkout Procedures

Respiration Tests

1. With the ECG patient cable still connected to the ECG/RESP connector of the monitor, set up the patient simulator as follows:
Respiration (RESP) baseline impedance – 750Ω,
RESP R – 0.5Ω,
RESP lead select – I & II,
RESP rate (respirations per minute) – 30.
2. Set up the monitor as follows:
RESP waveform – on,
RESP waveform lead select – lead II (RESP waveform derived
from ECG lead II).
3. Observe the following:
RESP parameter window appears on the monitor with a reading
of 30 ±2 (respirations per minute),
RESP waveform appears distortion-free on the monitor.
4. Change the RESP waveform lead select of the monitor to lead I (RESP waveform derived from ECG lead I).
5. Observe the following:
RESP parameter window appears on the monitor with a reading
of 30 ±2 (respirations per minute),
RESP waveform appears distortion-free on the monitor.
6. Disconnect the ECG patient cable from the ECG/RESP connector of the monitor. Proceed to the next steps in these checkout procedures.
4-30 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures

Temperature Tests

1. Set up the patient simulator for a temperature output of 37°C.
2. Attach the temperature adaptor cable to the TEMP/CO connector of the monitor.
3. Set the switch on the temperature adaptor to the 400 position.
4. Attach the temperature simulator cable from the SERIES 400 TEMPERATURE OUTPUT connector of the patient simulator to the T1 connector of the temperature adaptor.
5. Verify a TEMP parameter window appears on the monitor display with a T1 reading of 37.0° ±0.4° C.
6. Move the temperature simulator cable from the T1 connector of the temperature adaptor to the T2 connector of the temperature adaptor.
7. Verify a T2 reading of 37.0° ±0.4° C in the TEMP parameter window on the monitor display.
8. Remove the temperature adaptor and temperature simulator cable from the monitor and patient simulator.
Revision A Dash 3000/4000 Patient Monitor 4-31
2000966-035
MAINTENANCE: Checkout Procedures

Cardiac Output Tests

1. Connect the cardiac output (CO) cable adaptor to the TEMP/CO connector of the monitor.
2. Connect a simulator cable between the CO cable adaptor and the CO simulator.
3. Set the CO simulator to output blood temperature (BT) readings, as found in the following table:
Simulator BT Setting Monitor BT Reading Range
30.3°C 30.1 – 30.5
35.1°C 34.9 – 35.3
36.0°C 35.8 – 36.2
37.0°C 36.8 – 37.2
41.7°C 41.5 – 41.9
4-32 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
4. Verify a CO parameter window appears on the monitor display with correct BT readings as shown in the table above.
5. Set the CO simulator to output injectate temperature (IT) readings, as found in the following table:
Simulator IT Setting Monitor IT Reading Range
0.0°C –0.3 – +0.3
8.0°C 7.7 – 8.3
15.0°C 14.7 – 15.3
24.0°C 23.7 – 24.3
29.6°C 29.3 – 29.9
6. Verify correct IT readings appear on the monitor display, as shown in the table above.
7. Disconnect the CO cable adaptor from the TEMP/CO connector of the monitor. This completes the CO tests.
Revision A Dash 3000/4000 Patient Monitor 4-33
2000966-035
MAINTENANCE: Checkout Procedures

Invasive Blood Pressure Tests

BP1 Connector (AR1) Tests

The invasive blood pressure (BP) tests provide a method of verification for both BP connectors (BP1 and BP2) of a monitor equipped with this optional function. Follow these steps:
1. Set up the patient simulator as follows:
Blood pressure (BP) polarity – POS,
BP output – 0 mmHg.
1. Connect the BP simulator cable from the BLOOD PRESSURE 1 ­120/80 connector of the patient simulator to the BP1 (left-most BP) connector of the monitor.
2. Verify the AR1 parameter window, waveform label, corresponding graticules, and waveform appear on the monitor display, along with a BP waveform requiring zero reference.
3. Press the FUNCTION key on the front panel of the monitor to zero­reference the AR1 BP waveform.
4. Change the patient simulator BP output to 200 mmHg.
5. Observe a reading of 200/200 (200) ± 4 mmHg in the AR1 parameter window on the monitor display.
6. Change the patient simulator BP output to WAVE (simulated BP waveform).
7. Set the AR1 BP waveform gain on the monitor to auto.
4-34 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
8. Observe a distortion-free AR1 BP waveform and a reading of approximately 120/80 (93) in the AR1 parameter window on the monitor display.
9. Disconnect the BP simulator cable from the BP1 connector of the monitor. Continue to the next step for the BP2 test.
10. Again, set up the patient simulator as follows:
BP polarity – POS,
BP output – 0 mmHg.

BP2 Connector (PA2) Tests

1. Connect the BP simulator cable to the BP2 (right-most BP) connector of the monitor.
2. Verify a PA2 parameter window, waveform label and corresponding graticules appear on the monitor display, along with a PA2 BP waveform requiring zero reference.
3. Press the FUNCTION key on the front panel of the monitor to zero reference the PA2 BP waveform.
4. Change the patient simulator BP output to 200 mmHg.
5. Observe a reading of 200/200 (200) ± 4 mmHg in the PA2 parameter window on the monitor display.
6. Change the patient simulator BP output to WAVE (simulated BP waveform).
7. Set the PA2 BP waveform gain on the monitor to auto.
8. Observe a distortion-free PA2 BP waveform and a reading of approximately 120/80 (93) in the PA2 parameter window on the monitor display.
9. Remove the BP simulator cable from the BP2 connector of the monitor. This completes the BP tests.
Revision A Dash 3000/4000 Patient Monitor 4-35
2000966-035
MAINTENANCE: Checkout Procedures

Pulse Oximetry Tests

1. Set the pulse oximetry (SpO2) simulator power switch to the off position.
2. Connect the Nellcor-style SpO connector of the monitor and the SpO
simulator cable between the SpO2
2
simulator.
2
3. Set up the SpO
simulator as follows:
2
SPO2 – 99% (using the white NELLCOR values),
PULSE RATE – 100 B/M (beats per minute),
MODE – NELLCOR,
Power switch – on.
4. Verify a SPO
parameter window and waveform label appear on the
2
monitor display.
5. Verify the following appear on the monitor display:
Sinusoidal SpO
•SPO
% parameter reading of 97-102 (%),
2
waveform,
2
PPR parameter reading of 97-103 (beats per minute).
6. Verify accuracy of the SPO values shown on the SpO the SpO
simulator settings from the following table:
2
% values (these are the white NELLCOR
2
simulator) on the monitor display using
2
SpO2 Simulator Setting Displayed SPO2% Value
99% 97 – 102
85.5% 83 – 88
68.4% 66 – 71
4-36 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
7. Verify accuracy of the PPR values on the monitor display using the SpO
simulator pulse rates from the following table.
2
Simulator PULSE RATE
Displayed PPR Value
70 B/M 68 – 72 100 B/M 97 – 103 160 B/M 156 – 164
8. Press the INTERFERENCE TEST button on the SpO
simulator for
2
30 seconds.
9. Verify the displayed SPO
% value remains 97–102%, or an
2
interference detection message is displayed and XX is displayed in the SpO
parameter window in place of an SPO2% value.
2
10. Set the SpO2 simulator power switch to the off position.
11. Disconnect the Nellcor-style SpO SpO
connector. This completes the SpO2 tests.
2
simulator cable from the monitor
2
Revision A Dash 3000/4000 Patient Monitor 4-37
2000966-035
MAINTENANCE: Checkout Procedures

Noninvasive Blood Pressure Tests

1. Attach the digital manometer, noninvasive blood pressure (NBP) cuff, tees and tubing, as shown in the illustration below, to the NBP connector of the monitor.
2. Set the digital manometer power switch to the on position.
3. Set the digital manometer range switch to 1000 mmHg.
To perform the noninvasive blood pressure (NBP) tests, version 1A software is assumed to be installed in the monitor.
Using the Trim Knob control, access the SERVICE MODE menu starting from the MAIN menu.
1. Select MORE MENUS -> MONITOR SETUP -> SERVICE MODE ->
2. Enter password using the Trim Knob control to select the day and month from monitor screen with leading zeros. (e.g. July 4 = 0407).
4-38 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
3. Select CALIBRATE-> CALIBRATE NBP-> CHECK CAL OFF-> START->.
The text on the menu item changes from CHECK CAL OFF to CHECK CAL IN PROGRESS.
Verify the readings in the NBP parameter window on the monitor
display and readings on the digital manometer are equal (± 1 mmHg) for at least one full minute. If the readings are not equal for at least one full minute, the NBP circuit requires calibration.
3
4. Select CHECK CAL IN PROGRESS-> STOP->. The pneumatic control circuit of the monitor vents air pressure in
the pneumatic circuit of the monitor to atmosphere and causes the NBP cuff to deflate.
4
5. Remove the NBP test setup apparatus from the monitor. The NBP tests are complete.
Revision A Dash 3000/4000 Patient Monitor 4-39
2000966-035
MAINTENANCE: Checkout Procedures

Analog Output and Defibrillator Synchronization Tests

DEFIB Sync Connector: ECG

1. Use the figure at the left as a reference for connecting the oscilloscope to the DEFIB SYNC connector, located on the back panel of the monitor, for performing these tests.
2. Test the ECG, Arterial BP, and Marker Out signals from the DEFIB SYNC connector. They should closely resemble the waveforms in the figures below.
Signal Pin: — 7
Ground Pin: — 3
Probe Type: — x10
Time/Division: — 0.2S
Volts/Division: — 0.5V

DEFIB Sync Connector: Arterial BP

023A
Signal Pin: — 6
Ground Pin: — 5
Probe Type: — x10
Time/Division: — 0.2S
Volts/Division: — 0.2V
024A
4-40 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures
There are two Marker Out traces shown below. The upper Marker Out figure references the frequency aspects of the signal. The lower Marker Out figure references the pulse width aspects of the signal.
NOTE: The Marker Out amplitude and the pulse width are configured
in the boot menu as described in the configuration chapter. The following two graphs indicate an amplitude of 5V and a pulse width of 10ms.

DEFIB Sync Connector: Marker Out (Frequency)

DEFIB Sync Connector: Marker Out (Pulse Width)

Signal Pin: — 1
Ground Pin: — 8
Probe Type: — x10
Time/Division: — 0.2S
Volts/Division: — 1V
025A
Signal Pin: — 1
Ground Pin: — 8
Probe Type: — x10
Time/Division: — 5mS
Volts/Division: — 1V
026A
Revision A Dash 3000/4000 Patient Monitor 4-41
2000966-035
MAINTENANCE: Checkout Procedures

Verify Markers

3. Attach a jumper wire between pin-1 (Marker Out) and pin-2 (Marker In) of the DEFIB SYNC connector located on the back of the monitor. Verify negative spikes in each of the QRS Complex (ECG waveform) R-Waves on the monitor display, similar to those shown in the illustration below.
4. Remove the jumper wire installed in the previous step, from the DEFIB SYNC connector. This completes the defibrillator synchronization tests.
4-42 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures

Battery Tests

1. Disconnect the power cord plug from the wall receptacle.
2. Verify the BATTERY front panel indicator illuminates. This
indicates operation from the monitor’s battery power.
3. Setup the patient simulator as follows:
ECG heart rate – 80 bpm,
ECG amplitude – 1.0 mV,
5-lead patient cable attached.
4. Observe the following:
ECG Lead II is displayed and is noise-free,
Heart rate of 80 ±1 bpm is displayed,
•With QRS VOLUME enabled, an audible tone sounds with each
R-Wave.
5. Verify all six ECG leads are selectable for display on the monitor.
6. Connect the power cord plug to the wall receptacle.
7. Verify the AC front panel indicator illuminates. This indicates the monitor is operating from wall receptacle (AC) power.
8. Verify the CHARGING STATUS front panel indicator illuminates for a few minutes.
An amber glow indicates the monitor battery is charging.
A green glow indicates the monitor batteries are fully charged.

Graph Test

Graph Speed Test

Using the Trim Knob control, access the SERVICE MODE menu starting from the MAIN menu.
1. Select MORE MENUS -> MONITOR SETUP -> SERVICE MODE ->
2. Enter password using the Trim Knob control to select the day and month from monitor screen with leading zeros. (e.g. July 4 = 0407).
3. Select GRAPH TEST PATTERN-> START->.
4. Verify the following:
•Fonts.
•Shading.
•Triangle Pattern.
•No missing dots.
Select GRAPH TEST PATTERN-> STOP->
5.
Using the Trim Knob control, access the GRAPH SETUP menu starting from the MAIN menu.
1. Select MORE MENUS -> MONITOR SETUP -> GRAPH SETUP ->
2. Select SPEED:25 (default).
3. Verify that all eight speeds work.
.
Revision A Dash 3000/4000 Patient Monitor 4-43
2000966-035
MAINTENANCE: Checkout Procedures

Display Test

Speaker Test

Network Test

1. Hold the NBP GO/STOP and the FUNCTION keys and press the Trim Knob control at the same time.
2. Release the Trim Knob control immediately.
3. Continue holding the NBP GO/STOP and the FUNCTION keys.
4. Select “Video Test Screens.”
5. Test all screens:
White Screen.
Red Screen.
Blue Screen.
Green Screen.
Vertical Bars.
1. Change the alarm volume of the monitor to 100%.
2. Verify the speaker volume of the monitor changes accordingly.
3. Return the volume of the monitor to the level it was previously set to, before you changed it for this test.
1. Verify that the monitor is connected to the Unity-MC (Mission Critical) network.

RF LAN Test (option)

2. Select VIEW OTHER PATIENTS.
3. Select SELECT ANOTHER CARE UNIT.
4. Verify that you can see at least one care unit.
5. Select a care unit.
6. Select SELECT A BED TO VIEW.
7. Select a bed.
8. Verify that the patient window appears on the monitor’s split-screen.
1. If the monitor has the Wireless LAN, disconnect the Ethernet cable and verify Wireless LAN communication still exist between beds,
2. Reconnect the Ethernet Cable.
3. Return to the MAIN MENU.
4-44 Dash 3000/4000 Patient Monitor Revision A
2000966-035
MAINTENANCE: Checkout Procedures

RAC 2A Module Housing Test

Electrical Safety

Operation

Because the RAC 2A module housing has a separate power supply, perform electrical safety tests separate from the monitor.
Refer to the Electrical Safety Tests section of this chapter and complete the following test.
1. Wall Receptacle Test
2. Ground (earth) Continuity Test,
3. Ground (earth) Wire Leakage Tests, and
4. Enclosure Leakage Current Test.
To test the RAC 2A module housing for proper operation with your patient monitor complete the following steps:
1. Attach the communications interface cable to the Auto port on the RAC 2A housing and the AUX port of the monitor.
2. Install a SAM module into the RAC 2A module housing.
3. Apply power to the patient monitor and the RAC 2A module housing.
4. Operate the module and check for proper output displayed on the monitor.

Checkout Procedures Completion

PM Form

This completes all tests associated with the checkout procedures.
1. Discharge the test patient admitted during the “ECG Tests” on
page 4-27.
2. Set all test equipment power switches to the off position.
3. Unplug the monitor from AC power.
4. Remove all test equipment from the monitor.
Due to continuing product innovation and because specifications in this manual are subject to change without notice, a PM form is not included with this manual. For the latest PM form regarding this product, contact GE Medical Systems Information Technologies Service.
If repairs/adjustments were made or any parts replaced, describe this in the area provided on the PM form.
Also include comments regarding any unusual environmental conditions that may affect the operation or reliability of the equipment in the area provided on the PM form.
On the following pages a repair log is included for your convenience to record the repair history of this product.
Revision A Dash 3000/4000 Patient Monitor 4-45
2000966-035

Repair Log

Unit Serial Number:
Institution Name:
Date Maintenance/Repair Technician
MAINTENANCE: Checkout Procedures
4-46 Dash 3000/4000 Patient Monitor Revision A
2000966-035

5 TROUBLESHOOTING

Revision A Dash 3000/4000 Patient Monitor 5-1
2000966-035
For your notes
TROUBLESHOOTING:
5-2 Dash 3000/4000 Patient Monitor Revision A
2000966-035

Service Menus

TROUBLESHOOTING: Service Menus
There are two distinct service menus for the monitor. The SERVICE MODE menu is found in the monitor’s Main Menu and is used for various
functions like calibration, video tests, and downloading monitor interface software. The Boot Loader SERVCIE MENU is found in the Boot Code and is used when downloading the Boot Code and main processor code.
Both service menus are generally used by qualified field engineers and factory service personnel to troubleshoot, repair, or download new software to the patient monitor.
WARNING
The Boot Loader SERVICE MENU and the SERVICE MODE menu is intended for qualified personnel only. It
is possible to lose patient data, damage the operating software for this monitor, and even affect the Unity Network. Do not ‘experiment’ with any commands found in the service menus.

Boot Loader Service Menu

Use the Boot Loader service menu when downloading new Boot Code or Main Code software to the patient monitor or when the patient monitor exhibits a serious failure. Activate the Boot Loader program as follows:
1. Hold down NBP Go/Stop and Function on the front panel.
2. Press and release the Trim Knob control.
3. Keep holding NBP Go/Stop and Function until the Boot Loader information appears on the display.
Following is a list of options in the boot code service menu;
CHANGE INTERNET ADDRESS—This menu selection allows changes to the Ethernet address, gateway address, and internet mask.
WARNING
Duplication of an Internet address on a network causes data loss and possible Unity Network problems. If you change the factory assigned Internet address, you must record all other Internet addresses used on your network to avoid duplication.
SHOW INSTALL OPTIONS—This menu list the options installed on the monitor.
SET CONFIGURATION—This menu contains options for configuring the monitor. Refer to Boot Code Selections in the Configuration chapter of this manual.
SERIAL DOWNLOAD MAIN—This option is used when downloading
Revision A Dash 3000/4000 Patient Monitor 5-3
2000966-035
TROUBLESHOOTING: Service Menus
software from a laptop PC.
SERIAL DOWNLOAD BOOT—This option is used when downloading software from a laptop PC.
SERIAL DOWNLOAD DAS MAIN—This option is used when downloading software from a laptop PC.
SERIAL DOWNLOAD DAS BOOT—This option is used when downloading software from a laptop PC.
SERIAL DOWNLOAD WRITER MAIN—This option is used when downloading software from a laptop PC.
SERIAL DOWNLOAD WRITER BOOT—This option is used when downloading software from a laptop PC.
VIDEO TEST SCREENS—Various color screens for testing the display.
BATTERY SIMULATION—This option is for engineering use only.
WAKE UP BATTERY—This option is used when the battery is dead. See “Error Message” on page 5-11.

Main Menu Service Mode Menu

OPTIONS MENU—A unique password is required for each option. Contact your sales/service representative to obtain a password. You must provide your product serial number and Ethernet address. (The Ethernet address is displayed in the Boot Code banner information.)
The SERVICE MODE menu option items provide the user access to several general and technical built-in software functions of the monitor. Only persons responsible for configuring and maintaining the monitor should access the service mode menu option items.
WARNING
The Service Mode menu is intended for use only by qualified service technicians. Experimentation with service mode menu option items can be detrimental to the monitor. Lost patient data, damaged operating system software for the monitor, even network related problems are but a few examples of problems that can be induced as the result of tampering with service mode menu option items.
5-4 Dash 3000/4000 Patient Monitor Revision A
2000966-035
TROUBLESHOOTING: Service Menus
Access the Service Mode

About Service Mode Menu Option Items

Access the SERVICE MODE menu starting from the MAIN menu.
1. Select MORE MENUS -> MONITOR SETUP -> SERVICE MODE ->
2. Enter password using the Trim Knob control to select the day and month from monitor screen with leading zeros. (e.g. July 4 = 0407).
The Service Mode menu is used for initial setup and configuration as well as for troubleshooting. ALWAYS exercise caution when using any of these password-protected functions.
The service technician can use the Service Mode menu to:
relay software information to design engineers;
calibrate and troubleshoot NBP functions of the monitor;
set admit menu options, software feature levels and operating
mode of the monitor;
configure the monitor unit name, bed number and Internet
address for use on the network; and
enter or change the time and date on the monitor.

Service Mode Menu Option Item s

Do not use any of these options unless specifically instructed to do so.
WARNING
Some of the service mode menu option items are to be used only by qualified service technicians and others are for general use. Because of this, unnecessary tampering with service mode menu option items for experimentation purposes is not recommended by GE Medical Systems Information Technologies and may cause a malfunction of the monitor.
Following is a list of options in the main code service menu;
REVIEW ERRORS—This menu selection is for advanced troubleshooting by GE Medical Systems Information Technologies’ engineers. Error log data can be transferred over the network to a central station and then loaded onto a diskette for review. (Review Errors is discusses in greater detail later in this chapter.)
CALIBRATE—For checkout or calibration of the noninvasive blood pressure, ECG analog output, BP analog output, CO
service menu functions of the monitor.
service, and SAM
2
BATTERY SERVICE—This is a complete collection of battery data for troubleshooting the batteries.
Revision A Dash 3000/4000 Patient Monitor 5-5
2000966-035
TROUBLESHOOTING: Service Menus
PATIENT-MONITOR TYPE—Select the type of monitor desired, i.e adult, neonatal or operating room. Refer to Chapter 6, Configuration, for detailed procedures.
WARNING
Changing the patient-monitor type will default the admit function to STANDARD configuration. Different alarms and parameters are activated for each selection.
NOTE: The keypad/remote control is DIDCA programmed for specific
monitor types. The error message WARNING: REMOTE MISMATCHED WITH MONITORING MODE displays if the monitor and keypad/remote control do not match.
MENU SETUP—This menu selection provides the following sub-menus: (Refer to Chapter 6, Configuration, for detailed procedures.)
ADMIT MENU: STANDARD This menu selection allows you to determine the function of the patient monitor. The four variables include stationary or ambulatory (telemetry) patient monitoring with a monitor that always stays in one room (STANDARD) or a monitor that moves from room to room (ROVER).
SOFTWARE LEVEL This menu selection displays the software feature level this monitor is using. It allows setting the level to a lower setting than the software feature level setting in Boot Code.
MONITOR DEFAULTS PASSWD This menu selection allows you to set the monitor so that a password is REQUIRED or NOT REQUIRED for entry into the MONITOR DEFAULTS menu section. If selected, the password will be the same as the SERVICE MODE MENU password.
MONITOR SETTINGS—This menu selection provides the following sub­menus: (Refer to Chapter 6, Configuration, for detailed procedures.)
SET UNIT NAME This menu selection allows changes to the care unit name. After initial setup, this name should not be changed or communication to the central station will be corrupted. Note that the care unit name must be registered exactly the same in the central station and the patient monitor.
SET BED NUMBER This menu selection allows changes to the bed number. After initial setup, this number should not be changed or communication to the central station will be corrupted. Note that the bed number must be registered exactly the same in the central station and the patient monitor.
SET INTERNET ADDRESS This menu selection allows changes to the internet (IP) address.
5-6 Dash 3000/4000 Patient Monitor Revision A
2000966-035
TROUBLESHOOTING: Service Menus
WARNING
Duplication of an internet (IP) address on a network causes lost data. If you change the factory assigned internet address, you must first record all other internet addresses used on your network to avoid duplication.
An incorrect internet address may also prevent the monitor from viewing other monitors on the network even though the unit names match. Whether or not this can occur depends on the network topology at the installed site.
GRAPH TEST PATTERN—This menu selection allows you to run a graph test pattern. The choices are START and STOP.
TIME AND DATE—This menu selection allows changes to the time and date and may affect the time and date for the entire monitoring network. (Refer to Chapter 6, Configuration, for detailed procedures.)
WARNING
Loss of patient history. This menu should rarely be used because patient histories will be lost.
Revision A Dash 3000/4000 Patient Monitor 5-7
2000966-035
TROUBLESHOOTING: Service Menus

Review Errors

About the Monitor Error Log

Downloading the Error Log

Accessing the Review Errors Menu Option Item

The REVIEW ERRORS menu is an advanced troubleshooting tool used by GE Medical Systems Information Technologies’ engineering
personnel. Some of the information recorded in the monitor error log can be useful for field service troubleshooting.
This section provides an introduction to error log usage and meaning. Because the information contained in the error log is engineering­oriented, the intent of the manual is to simply provide a general understanding of this monitor function.
This section includes a method for downloading error log data over the network to a central station. Once downloaded to a central station, you can load the error log data onto floppy diskettes or review it on the central station.
To access the error log and learn more about the REVIEW ERRORS menu option item, follow these steps:
1. Rotate and press the Trim Knob control to select REVIEW ERRORS from the Service Mode Menu.
2. The review errors menu option items include four possible selections; one each for viewing output or input errors along with one each for clearing output or input errors. Rotate and press the Trim Knob control to scroll to and select VIEW OUTPUT ERRORS from the Review Errors Menu.
3. The RUN TIME ERROR LOG pop-up window appears on the left side of the monitor display. One time-dated output software error appears in the pop-up window at a time.
Use the Trim Knob control to scroll through each logged error and peruse all of the parameters associated with each output software error. Rotate the Trim Knob control to move the cursor (>) to a position for viewing the NEXT or PREVIOUS error as well as the position that allows the user to QUIT viewing output errors.
Selecting QUIT closes the RUN TIME ERROR LOG pop-up window and returns to the Review Errors Menu.
4. The VIEW INPUT ERRORS menu causes a RUN TIME ERROR LOG pop-up window to appear on the monitor display. The pop-up window now displays input software errors and provides basically the same information as the VIEW OUTPUT ERRORS pop-up window provided. The appearance of both pop-up windows are similar, the difference being errors that are logged as input versus output to/from the monitor.
5. To clear out the stored run time error logs, use the Trim Knob control to select the CLEAR OUTPUT ERRORS or CLEAR INPUT ERRORS menu, respectively.
Immediately after you clear one of the error logs, a message appears on the upper right side of the display. The message verifies the actuation of the Trim Knob control for this function.
5-8 Dash 3000/4000 Patient Monitor Revision A
2000966-035
TROUBLESHOOTING: Service Menus

Error Log Information

This part of the section describes in greater detail what information the error log contains and what can be learned from error logs.
An error log in the monitor can hold up to 50 events. As an event occurs, error information is stored in the log. Subsequent events are stored sequentially as they occur. When the 50-event limit is reached, subsequent errors replacing the oldest error(s) in the log.
A sample of the monitor error log pop-up window appears as follows:
When using the error log to troubleshoot a problem with the monitor, the following parameters from the pop-up window that are of greatest interest are:
PROCESS NAME—The task that was operating when the event or problem occurred,
ERROR CODE—A software code for the type of event or problem that occurred,
SEVERITY—Indicates the level of impact of the event or problem on the system,
DATE—The date the event or problem occurred,
TIME—The time the event or problem occurred, and

Error Logs

Severity of the Error

ERROR NUMBER—A sequential number used to identify each event or problem.
INPUT ERROR—Additional information used to determine the cause of the error.
Error logs contain more than just operating system errors. Many events that occur that might have an impact upon the system are entered into the log. These logs may be requested by Tech Support on occasion to aid in troubleshooting the monitor. The logs are developed to aid engineering for internal diagnostics of the monitor. Contact Tech Support if you need clarification of any of the error logs.
Severity is a measure of how the event/error affected the system. There are three levels of severity. The following is a list of these levels accompanied by a brief description of each:
CONTINUE—The event or error was logged, the task may or may not have completed, but the system was able to continue operating. Most error log entries have this severity level.
FATAL—The event or error was logged, the task did not complete, and the system was unable to continue operating as recovery was not possible. This level of severity in an event or error is always followed by an automatic warm start.
FORCED RESTART—The operating system restarted normally after a known condition, such as an Internet address change, patient discharge, etc.
Revision A Dash 3000/4000 Patient Monitor 5-9
2000966-035
TROUBLESHOOTING: Battery Alarms and Messages

Battery Alarms and Messages

Alarm Conditions

Messages Displayed in the ECG Waveform Area
BATTERY LOW System WARNING
POWERING DOWN System WARNING
CHECK BATTERY STATUS System MESSAGE
BATTERY ERROR System WARNING
Battery alarms occur when the following conditions occur:
Low Battery,
Empty Battery,
Battery Failures, or
Charger Failures.
Battery Alarm Cause
Critical Low Battery battery of run time remaining (10 minutes if one battery, 20 minutes if two batteries).
Empty Battery remaining.
Battery Failure while using or charging the battery.
Battery Failure occurred while using or charging the battery.
Only 10 minutes per
There is no battery run time
A minor failure has occurred
A serious failure has
CHECK BATT STATUS NOTE: INTERNAL CHARGER
FAILED, CALL SERVICE also
appears in the Battery Status information window.
Messages Displayed in the Battery Status Information Window
INTERNAL CHARGER FAILED, CALL SERVICE
NOTE: CHECK BATT STATUS also appears in the ECG waveform area.
CONDITION None
System MESSAGE
Battery Alarm Cause
System MESSAGE
Charger Failure have failed.
Charger Failure have failed.
Condition conditioning cycle.
Charger communications
Charger communications
The battery is requesting a
5-10 Dash 3000/4000 Patient Monitor Revision A
2000966-035
Loading...