Premium, Prism, QUIK CONNECT V, QUICK CONNECT, QT Guard, SMART-PAC, SMARTLOOK, Spiral
Lok, Sweetheart, UNITY, Universal, Waterfall, and Walkmom are trademarks of GE Medical Systems
Information Technologies.
Each page of this manual has the document part number and revision
letter at the bottom of the page. The revision letter identifies the
document’s update level. The revision history of this document is
summarized below.
Revision History
This manual supplies technical information for service representatives
and technical personnel so they can maintain the equipment to the
assembly level. Use it as a guide for maintenance and electrical repairs
considered field repairable. Where necessary the manual identifies
additional sources of relevant information and technical assistance.
See the operator’s manual for the instructions necessary to operate the
equipment safely in accordance with its function and intended use.
This manual is intended for service representatives and technical
personnel who maintain, troubleshoot, or repair this equipment.
Revision ADash 3000/4000 Patient Monitor1-3
2000966-035
Safety Information
INTRODUCTION: Safety Information
Responsibility of the
Manufacturer
General
GE Medical Systems Information Technologies is responsible for the
effects of safety, reliability, and performance only if:
•Assembly operations, extensions, readjustments, modifications, or
repairs are carried out by persons authorized by GE Medical Systems
Information Technologies.
•The electrical installation of the relevant room complies with the
requirements of the appropriate regulations.
•The equipment is used in accordance with the instructions for use.
This device is intended for use under the direct supervision of a licensed
health care practitioner.
This device is not intended for home use.
Federal law restricts this device to be sold by or on the order of a
physician.
Contact GE Medical Systems Information Technologies for information
before connecting any devices to the equipment that are not
recommended in this manual.
Parts and accessories used must meet the requirements of the applicable
IEC 601 series safety standards, and/or the system configuration must
meet the requirements of the IEC 60601-1-1 medical electrical systems
standard.
Periodically, and whenever the integrity of the device is in doubt, test all
functions.
The use of ACCESSORY equipment not complying with the equivalent
safety requirements of this equipment may lead to a reduced level of
safety of the resulting system. Consideration relating to the choice shall
include:
•use of the accessory in the PATIENT VICINITY; and
•evidence that the safety certification of the ACCESSORY has been
performed in accordance to the appropriate IEC 60601-1 and/or IEC
60601-1-1 harmonized national standard.
If the installation of the equipment, in the USA, will use 240V rather
than 120V, the source must be a center-tapped, 240V, single-phase
circuit.
1-4Dash 3000/4000 Patient MonitorRevision A
2000966-035
INTRODUCTION: Safety Information
Warnings, Cautions,
and Notes
The terms danger, warning, and caution are used throughout this
manual to point out hazards and to designate a degree or level or
seriousness. Familiarize yourself with their definitions and significance.
Hazard is defined as a source of potential injury to a person.
DANGER indicates an imminent hazard which, if not avoided, will
result in death or serious injury.
WARNING indicates a potential hazard or unsafe practice which, if not
avoided, could result in death or serious injury.
CAUTION indicates a potential hazard or unsafe practice which, if not
avoided, could result in minor personal injury or product/property
damage.
NOTE provides application tips or other useful information to assure
that you get the most from your equipment.
Revision ADash 3000/4000 Patient Monitor1-5
2000966-035
INTRODUCTION: Safety Information
Equipment Symbols
Some of the following symbols appear on the equipment.
ATTENTION: Consult accompanying documents before using the
equipment.
In Europe, this symbol means dangerous or high voltage. In the United
States, this symbol represents the caution notice below:
To reduce the risk of electric shock, do NOT remove cover (or back).
Refer servicing to qualified personnel.
Defibrillator-proof type CF equipment; type CF equipment is
specifically designed for applications where a conductive connection
directly to the heart is established. The paddles indicate the equipment
is defibrillator proof.
Defibrillator-proof type BF equipment; type BF equipment is suitable
for intentional external and internal application to the patient,
excluding direct cardiac application. Type BF equipment is type B
equipment with an F-type isolated (floating) part. The paddles indicate
the equipment is defibrillator proof.
Type B equipment; type B equipment is suitable for intentional
external and internal application to the patient, excluding direct
cardiac application.
Equipotentiality
Alternating current (AC)
Power;
Fuse
Battery
Indicates the Ethernet connection for the monitor.
I = ON; O= OFF
1-6Dash 3000/4000 Patient MonitorRevision A
2000966-035
INTRODUCTION: Safety Information
PRESS
814A
815A
816A
817A
818A
Press to open.
Power
Graph Go/Stop
NBP Go/Stop
Function
Silence Alarm
Classified by Underwriters Laboratories Inc. with respect to electric
shock, fire, mechanical and other specified hazards, only in accordance
with UL 2601-1, CAN/CSA C22.2 No. 601.1, IEC 60601-1, and, if
required, IEC 60601-2-27, IEC 60601-2-30, IEC 60601-2-34, IEC 606011-1.
Revision ADash 3000/4000 Patient Monitor1-7
2000966-035
INTRODUCTION: Service Information
Service Information
Service
Requirements
Equipment
Identification
Follow the service requirements listed below.
•Refer equipment servicing to GE Medical Systems Information
Technologies’ authorized service personnel only.
•Any unauthorized attempt to repair equipment under warranty voids
that warranty.
•It is the user’s responsibility to report the need for service to GE
Medical Systems Information Technologies or to one of their
authorized agents.
•Failure on the part of the responsible individual, hospital, or
institution using this equipment to implement a satisfactory
maintenance schedule may cause undue equipment failure and
possible health hazards.
•Regular maintenance, irrespective of usage, is essential to ensure
that the equipment will always be functional when required.
Every GE Medical Systems Information Technologies device has a unique
serial number for identification. A sample of the information found on a
serial number label is shown below.
D 0 XX 0005 G XX
Month
Manufactured
A = January
B = February
C = March
D = April
E = May
F = June
G = July
H = August
J = September
K = October
L = November
M = December
Year
Manufactured
0 = 2000
1 = 2001
2 = 2002
(and so on)
Product Code
Two-character
product
descriptor
Product
Sequence
Number
Manufacturing
number (of total
units
manufactured)
Division
F = Cardiology
G = Monitoring
Device Characteristics
One or two letters that
further describe the unit,
for example:
P = prototype not
conforming to marketing
specification
R = refurbished equipment
S = special product
documented under Specials
part numbers
U = upgraded unit
1-8Dash 3000/4000 Patient MonitorRevision A
2000966-035
2EQUIPMENT OVERVIEW
Revision ADash 3000/4000 Patient Monitor2-1
2000966-035
For your notes
EQUIPMENT OVERVI EW:
2-2Dash 3000/4000 Patient MonitorRevision A
2000966-035
Components
EQUIPMENT OVERVIEW: Components
The Monitoring
System
The Patient Monitor
The Dash 3000/4000 patient monitor can function by itself with a built-in
writer, or it can be cabled in with the Unity Network via Ethernet.
Optional components are, if using Wireless LAN or cabled to Ethernet, a
Centralscope central station and the Clinical Information Center.
This device is designed to monitor a fixed set of parameters including
ECG, noninvasive blood pressure, impedance respiration, SpO2, and
temperature. Invasive pressure and EtCO2 are optional features.
Additional specialized features include cardiac output, cardiac
calculations, pulmonary calculations, dose calculations, PA wedge (PA
wedge is only available with the invasive pressure option), and SAM
module interface.
001A
Dash 3000 Monitor
Revision ADash 3000/4000 Patient Monitor2-3
2000966-035
Dash 4000 Monitor
051A
EQUIPMENT OVERVIEW: Components
Right Side View
Left Side View
All of the patient cable connectors are located on the right side of the
monitor. A Trim Knob control provides single control operation of
virtually all monitor functions.
Patient Cable
Connectors
002A
On the left of the monitor, you can find the built-in writer and the
battery compartment.
Optional Built-in Writer—
The built-in, 4 channel
writer is located in the
center of the left side of the
monitor .
Battery Compartment—
The battery packs are
located in this
compartment.
003A
2-4Dash 3000/4000 Patient MonitorRevision A
2000966-035
EQUIPMENT OVERVIEW: Components
Back Vie w
Network Connector—A cable
can be connected to this port
for monitors used in patient
monitoring network
configurations.
Equipotential Terminal—For
measurements in or near the heart
we recommend connecting the
monitor to the potential
equalization system. U se the
green and yellow potential
equalization cable and connect it
to this pin.
AC Power Connector
On the back of the monitor you will find all connectors for equipment and
network.
Line V oltage Selector—This s elector
is factory set to match the line
voltage and frequency rating for
your country.
Audible Alarm Enunciator—The
internal speaker provides sound
for audible alarms. For better
sound quality do not block
speaker.
Aux Port—Used for
RAC 2A and software
updates.
Defib Sync Connector—Provides
ECG analog output signals to
004A
Peripheral Expansion Port
user-supplied equipment. A 5volt, 2-millisecond artificial pacer
spike is added to the analog
output when PACE is on and
detection occurs.
Optional Alarm Indicator
An optional alarm indicator can be built into the handle of the Dash 3000
monitor or into the display bezel of the Dash 4000 monitor. When
activated, the LED indicator flashes red for CRISIS and WARNING
patient status alarms and yellow for all other alarms.
Alarm Indicator
Dash 3000 Monitor
Revision ADash 3000/4000 Patient Monitor2-5
2000966-035
Dash 4000 Monitor
052A
EQUIPMENT OVERVIEW: Components
Optional RAC 2A
Module Housing
The RAC 2A module housing currently supports the SAM module.
006A
An integral power supply is used to run the RAC 2A and support the
needed voltages.
2-6Dash 3000/4000 Patient MonitorRevision A
2000966-035
EQUIPMENT OVERVIEW: Components
Optional Wireless
LAN System
Access Points
The flexibility of the GE Medical Systems Information Technologies
Unity Network is increased by using the Wireless LAN system. The
Wireless LAN system allows the user to roam from one access point to
another, maintaining a strong, seamless connection to the Unity
network.
The monitor, with its optional built-in Wireless LAN, functionally
performs the same as a monitor connected directly to the Unity network.
It can be viewed at the central station and by other GE Medical Systems
Information Technologies’ monitors on the network (i.e. Dash 3000/4000,
Eagle 4000, and Solar patient monitors). Monitors with Wireless LAN
sends and receives patient data via the access points of the Unity
network.
NOTE: Wireless patient monitors that are moved from room to room
must have the monitor type configured as Rover or Rover/Combo
monitoring.
To integrate the wireless network with the wired network, one or more
access points are necessary. An access point connects the wireless
monitor to the wired network infrastructure within the building, and
acts as a bridge between the wired and wireless networks. The areas
covered by each access point overlap to insure continuous coverage.
NOTE: The monitor will only work with a Symbol Access Point. The
monitor will not communicate directly with a Wireless LAN
device from Aironet.
050A
Revision ADash 3000/4000 Patient Monitor2-7
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications
Technical Specifications
Due to continual product innovation, specifications are subject to change
without notice. The following specifications are accurate as of the date of
this publication, and pertain to the Dash 3000/4000 Patient Monitor.
Performance
Specifications
Display
Size:
Type:
Color:Active-Matrix Liquid Crystal Display (LCD)
Resolution:640 by 480 pixels
Dash 3000:
Dash 4000:
8.4-inch diagonal
10.4-inch diagonal
Controls
Alarms
Number of traces:6 (maximum)
Number of seconds/trace:
Dash 3000:
Dash 4000:
Sweep speed:
All waveforms6.25, 12.5 or 25 mm/sec (with erase bar)
5 Leadwire cable:I, II, III, V, aVR, aVL, and aVF
10 Leadwire cable (12SL option):V2, V3, V4, V5 and V6
Leads analyzed simultaneously:I, II, III, and V (multi-lead mode)
Lead fail:Identifies failed lead
Alarms:User-selectable upper and lower heart rate limits
Input specifications:
Voltage range:
Signal width:
Heart rate range:
Accuracy:
Input impedance:
Common mode:
Differential:
Common mode rejection:
±0.5 mV to ±5 mV
40 ms to 120 ms (Q to S)
30 to 300 BPM
±1% or ±1 BPM, whichever is greater
0.05 to 25 Hz
5 to 25 Hz
±3% (maximum)
<30 µV (referred to input)
Adjustable from 0 to 120 ms past the J-point
(default: 60 ms adult, 30 ms neonatal)
–12.0 to +12.0 mm
±10% or 0.5 mm, whichever is greater
±2 mV to ±700 mV
0.1 ms to 2 ms
10 µs to 100 µs
2 mV (max)
<0.5 mV/hour with a ±700-mV, 2-ms pacemaker pulse
applied
Revision ADash 3000/4000 Patient Monitor2-9
2000966-035
Invasive Blood
Pressure (BP)
EQUIPMENT OVERVIEW: Technical Specifications
Number of channels:2
Transducer sites:Arterial (ART), femoral artery (FEM), pulmonary artery
(PA), central venous (CVP), right atrial (RA), left atrial
(LA), intracranial (ICP), and special (SP)
In neonatal mode: umbilical artery catheter (UAC) and
umbilical venous catheter (UVC)
Transducer requirements:
Excitation voltage:
Transducer output:
Input specifications:
Range:
Offset:
Output specifications:
Frequency response:
Zero balance range:
Zero balance accuracy:
Zero balance drift:
Accuracy:
Alarms:
5.0 Vdc ±0.1%
5 µV/V/mmHg
–25 mmHg to 300 mmHg
±150 mmHg
dc to 50 Hz
±150 mmHg
±1 mmHg
±1 mmHg over 24 hours
±2% or ±1 mmHg, whichever is greater (exclusive of
transducer)
User-selectable upper and lower limits for systolic,
diastolic, and mean pressures
2-10Dash 3000/4000 Patient MonitorRevision A
2000966-035
EQUIPMENT OVERVIEW: Technical Specifications
Noninvasive Blood
Pressure (NBP)
Measurement technique:Oscillometric
Displayed parameters:Systolic, diastolic, and mean pressures, pulse rate, time
of last measurement
Measurement modes:Manual, auto, and stat in adult and OR modes; manual
and auto in neonatal mode
NBP pressure range:
Systolic pressure range
Adult:
Pediatric:
Neonatal:
Diastolic pressure range
Adult:
Pediatric:
Neonatal:
Mean pressure range
Adult:
Pediatric:
Neonatal:
30 to 275 mmHg
30 to 235 mmHg
30 to 135 mmHg
10 to 220 mmHg
10 to 220 mmHg
10 to 110 mmHg
20 to 260 mmHg
20 to 260 mmHg
20 to 125 mmHg
Cuff pressure range:
Adult:
Pediatric:
Neonatal:
Pressure accuracy:
Static:
Clinical:
Heart rate detection:30 to 200 beats per minute
Total cycle time:20 to 40 seconds typical (dependent on heart rate and
Automatic cycle times:0 to 8 hours
Auto zero:Zero pressure reference prior to each cuff inflation
Tubing length:
Adult:
Neonatal:
Automatic cuff deflation:Cycle time exceeding 3 minutes (90 seconds neonatal),
Cuff sizes:
Disposable:
Reusable:
0 to 275 mmHg
0 to 235 mmHg
0 to 135 mmHg
±2% or ±3 mmHg, whichever is greater
±5 mmHg average error
8 mmHg standard deviation
motion artifact)
12 feet
8 feet
power off, or cuff pressure exceeds 294 mmHg (±6
mmHg) adult, 147 mmHg (±3 mmHg) neonatal
Large adult, adult, small adult, pediatric, small pediatric,
and infant
Thigh, large adult, adult, child, and infant
Alarms:User-selectable upper and lower limits for systolic,
diastolic, and mean pressures
Revision ADash 3000/4000 Patient Monitor2-11
2000966-035
Pulse Oximetry
(SPO2)
EQUIPMENT OVERVIEW: Technical Specifications
Parameters monitored:Arterial oxygen saturation (SpO2) and peripheral pulse
RR:
Accuracy:
Display update interval:
CO2 waveform sweep speed:
CO2 averaging:
CO2 measurement stability:
Resolution:
Noise:
0 to 100 mmHg/0 to 13%
0 to 100 mmHg/0 to 13%
0 to 120 breaths/min
±5% of reading or ±2 mmHg, whichever is greater
2 sec
Selectable 6.25, 12.5, or 25 mm/sec
Selectable from single breath, 10 sec, or 20 sec
Accuracy maintained over 8 hours
1 mmHg
2% of reading or 0.5 mmHg (maximum), whichever is
greater
60 Hz interference:
<0.5 mmHg at 38 mmHg
Step response time:
Adult:
Neonatal
<60 ms (10-90%)
<50 ms (10-90%)
Interference:
N2O gas:
±5% of reading or ±2 mmHg (maximum), whichever is
greater, with N2O compensation enabled
O2 gas:
±5% of reading or ±2 mmHg (maximum), whichever is
greater, with O2 compensation enabled
Barometric pressure:
Water vapor:
±2 mmHg (maximum) from 500 to 800 mmHg
±1.5% of reading or ±0.5 mmHg (maximum),
whichever is greater
Anesthetic agent:
±0.5 mmHg (maximum) for concentration of no more
than 5% of halogenated agents
Airway adapter variability:
±3% of reading or ±1.5 mmHg (maximum), whichever
is greater, with same or different adapter; not applicable
after adapter zero
Warm-up time:
Less than 15 seconds to initial CO2 indication, full
specification within 120 seconds; waveform immediate
upon power up
Calibration:
Factory settings:
Factory calibration settings stored in nonvolatile
memory within the sensor; 15 second adaptor
calibration when switching airway types
Verification:
Zero and span performance check with on-cable verifier
Respiration rate specifications:
Range (for 5% step size):
Accuracy:
Resolution:
0-120 breaths per minute
±1 breath per minute
±1 breath per minute
Barometric pressure sensor
specifications:
Range:
Accuracy:
425 to 817 mmHg (56 to 109 kPa)
±25 mmHg
Alarms:User-selectable upper and lower limits for CO2 and RR.
2-14Dash 3000/4000 Patient MonitorRevision A
2000966-035
Loading...
+ 220 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.