MarMonix MCA-207 Instruction Manual

AC/DC True-RMSClamp Meter
Products Code: MCA-207
Instruction Manual
WWW.MARMONIX.CO
1. Measuring Limits
2. Safety Information
3. Features
4. Instrument Layout
5. Specifications
6. DC Current Measurement
7. AC Current Measurement
8. DC Voltage Measurement
9. AC Voltage Measurement
10. Resistance Measurement
11. Capacitance Measurement
12. Frequency Measurement
13. Temperature Measurement
14. Diode Test
15. Continuity Testing
16. Data Hold Functions
17. Peak Hold Functions
18. LCD Backlight
19. Auto Power Off
20. Battery Replacement
1. Measuring Limits
DC Amperes: 0.01A to 1000A AC Amperes: 0.01A to 1000A DC Voltage: 0.1mV to 600V AC Voltage: 0.1mV to 600V
Resistance: 0.1Ὠ to 40MὨ Capacitance: 0.001 nFto 40mF Frequency: 0.001 kHz to 4kHz Temperature: -40˚C to 1000˚C
-40˚F to 1832˚F
2. Safety Information
This manual contains information that must be followed for operating the meter safety and maintaining the meter in a safe operating condition. If the meter is not used in a manner specified in this manual, the protection provided by the meter may be impaired. The Model has been designed and complies with IEC 61010-1 and EN 61010-1 safety requirements for Electronic Measuring Apparatus.
WARNING
Read through the operating instructions thoroughly
and understand the instructions before operating the meter.
Keep the manual at hand so as to enable quick
reference when necessary.
Ensure that the use of the meter is in its intended
applications and follow measuring procedures described in this manual.
Follow all safety and operating instructions to ensure
maximum personal safety during the use and operation of the meter.
Failure to follow the above instructions may cause
injury. Instrument damage and/or damage to equipment under test.
The symbol indicated on the meter means that the
user must refer to related parts in the manual for safe operation of the meter. Be sure to carefully read the instructions following each symbol in this manual.
DANGER is reserved for conditions and actions that are likely to cause serious or fatal injury
WARNING is reserved for conditions and actions that are likely to cause serious or fatal injury.
CAUTION is reserved for conditions and actions that are likely to cause minor injury
DANGER
Never use the meter to measure voltages on a circuit
above the maximum allowable input value on any function.
Do not exceed the maximum allowable input of any
measurement range.
Never touched exposed wiring, connections or any live
circuit when attempting to take measurement.
Do not attempt to make measurements in flammable
gaseous areas, fumes, vapor or dust. The use of the instrument in these areas may cause sparking, which can lead to an explosion.
Do not attempt to use the instrument if its surface or
your hand is wet.
Never open the battery compartment cover when
making measurements.
WARNING
Always inspect the meter and test leads for any signs of
damage or abnormality before use. If the meter or its accessories have any structural defects such as broken test leads, cracked cases, exposed metal parts or the display is not reading, do not attempt to make measurements.
Do not turn the function switch whilst test leads are
connected to the meter.
Do not install substitute parts or make modifications to
the meter. Return the meter to your distributor for repair or re-calibration.
Ensure the meter is switch off before opening the
battery cover when replacing the battery.
Never replace a battery if the surface of the meter is
wet or moist.
CAUTION
Before making measurements ensure that the function
selector switch is set on the appropriate range position.
Always make sure that the plug of each test lead is
inserted fully into the appropriate terminal of the meter.
Ensure that the function selector switch is set to the
“OFF” position after use. When the meter will not be in use for a long period of time, place it in storage after removing the battery.
Do not expose the meter to direct sunlight, extreme
temperatures or moisture.
Do not abrasives or solvents on the meter. To clean it,
use a damp cloth and mild detergent only.
Qualified and trained service technicians should only
perform calibration and repair for the meter.
3. FEATURES
True RMS measurement of AC Current ad AC Voltage Large 4000 count LCD display with a bargraph and a
bright white LED backlight.
Wide measurement range from 0.01A AC/DC up to
1000A AC/DC
Measures AC and DC Voltage up to 600 volts Measures resistance from 0.01Ὠ up to 40MὨ. Capacitance measurements up to 40mF Temperature measurements from -40˚C to +1000˚C and
-40˚F to +1832˚F.
Designed to the international safety standard IEC61010
CAT III 600V/ CAT II 1000V, Pollution degree 2.
Auto power off after approximately 20 minutes to
conserve battery life.
Continuity Buzzer and Diode Test Frequency measurement up to 4kHz. Peak Hold to record the minimum and maximum
readings for current and voltage.
Data Hold switch used to freeze readings on display.
4. INSTRUMENT LAYOUT
5. SPECIFICATIONS
Clamp Size 30mm Opening Approx Diode Test Test Current of 0.3mA typical; Open circuit voltage 1.5V DC Continuity check Threshold<35Ὠ; Test current<1mA
Low battery indication “ ” is displayed Overrange Indication “OL” is displayed Measurements Rate 2 per second, nominal Input Impedance 10MὨ (VDC and VAC) Display 4000 count LCD AC Current 50/60Hz (AAC) AC Voltage Bandwidth 50/60 (VAC) Operating Temperature -10˚C to 50˚C (14˚F to 122˚F) Storage Temperature -30˚C to 60˚C (-22˚F to 140˚F) Relative Humidity UP to 85% Over Voltage Category III 600V Battery One DC 9V IEC6F22 1604 Battery Auto Power Off Approx 20 minutes Dimensions 229 x 80 x 49mm Weight 303g
5.SPECIFICATIONS
Range
Measuring Range
Resolution
Accuracy(% of Readings)
40A
0~40.00A
0.01A
±(2.8%+10digits)
400A
0~400.0A
0.1A
±(2.8%+8digits)
1000A
0~1000A
1A
±(3.0%+8digits)
Range
Measuring Range
Resolution
Accuracy(% of Readings)
40A
0~40.00A
0.01A
±(2.8%+10digits)
400A
0~400.0A
0.1A
±(2.8%+8digits)
1000A
0~1000A
1A
±(3.0%+8digits)
Range
Measuring Range
Resolution
Accuracy(% of Readings)
Range
Measuring Range
Resolution
Accuracy(% of Readings)
DC Current
AC Current
DC Voltage
400mV 0~400.0mV 0.1mV ±(0.8% + 2digits)
4V 0~4.00V 0.001V
±(1.5% + 2digits) 40V 0~40.00V 0.01V
400V 0~400V 0.1V
600V 0~600V 1V ±(2.0% + 2digits)
AC Voltage
400mV 0~400.0mV 0.1mV ±(0.8% + 2digits)
4V 0~4.00V 0.001V
±(1.5% + 2digits) 40V 0~40.00V 0.01V
400V 0~400V 0.1V
600V 0~600V 1V ±(2.0% + 2digits)
Note: No Autoranging on 400mV AC Range
Resistance
Range
Measuring Range
Resolution
Accuracy(% of Readings)
Range
Measuring Range
Resolution
Accuracy(% of Readings)
Range
Measuring Range
Resolution
Accuracy(% of Readings)
400Ὠ 0~400.0 Ὠ 0. Ὠ ±(1.0% + 4 digits)
4k Ὠ 0~4.000k Ὠ 1 Ὠ
±(1.5% + 2 digits) 40k Ὠ 0~40.00k Ὠ 10 Ὠ
400kV 0~400.0k Ὠ 100 Ὠ
4M Ὠ 0~4.000M Ὠ 1k Ὠ ±(2.5% + 5digits)
40M Ὠ 0~40.00M Ὠ 10k Ὠ ±(3.5% + 10 digits)
Capacitance
4nF 0~4.000nF 0.001nF ±(5.0% + 30 digits)
40nF 0~40.00nF 0.01nF ±(5.0% + 20 digits)
400nF 0~400.0nF 0.1nF
±(3.0% + 5 digits) 4μF 0~4.000μF 0.001μF
40μF 0~40.00μF 0.01μF
400μF 0~400.0μF 0.1μF ±(4.0% + 10 digits)
4mF 0~4.000mF 0.001mF ±(4.5% + 10 digits)
40mF 0~40.00mF 0.01mF ±(5.0% + 10 digits)
Frequency
4kHz 0~4.000kHz
0.00kHz
±(1.5% + 2 digits)
Sensitivity: 5Vrms min
Temperature
Accuracy(% of Readings)
Range Measuring Range Resolution
˚C -40˚C~1000˚C 1˚C ±(2.5˚% + 3˚C) ˚F -40.0˚F~1832˚F 1˚F ±(2.5% + 5˚F)
6. DC Current Measurement
WARNING: Ensure that the test leads are
disconnected from the meter before making current measurements.
1. Set the function switch to 1000A or 400A or 40A
range position, and make sure that the current under test does not exceed the upper limit of the measuring range you have selected. The meter automatically defaults to DC Current.
2. Press the DC ZERO button once, the sign will
appear on the LCD indicating Range is zero.
3. Press the trigger to open up the transformer jaws
and clamp around the single conductor under test.
4. Read the display.
Note:
During current measurements keep the
transformer jaws fully closed, otherwise this will affect the accuracy of the measurement.
When large currents the transformers jaws
may buzz. This is not a fault and does not affect the accuracy of the reading.
7. AC True RMS Current Measurement
WARNING: Ensure that the test leads
are disconnected from the meter
before making current measurements.
1. Set the function switch to 1000A~ or 400A~ or
40A~range position, and make sure that the current under test does not exceed the upper limit of the measuring range you have selected.
2. Press the MODE button to select the AC Current range.
The meter automatically defaults to Dc Current.
3. Press the trigger to open up the transformer jaws and
clamp around the single conductor under test.
4. Read the display.
Note:
During current measurements keep the transformer
jaws fully closed, otherwise this will affect the accuracy of the measurement.
When large currents the transformer jaws may buzz.
This is not a fault and does not affect the accuracy of the reading.
The meter is set to default to “Auto Range” mode. Pressing the PEAK hold button will allow the user to
record the peak maximum and minimum readings on the AC volt range. See page 17 for further instructions for “PEAK” Hold function.
8. DC Voltage Measurement
1. Set the function switch to range position. The meter automatically defaults to DC Volts.
2. Insert the test lead into the “V Ὠ TEMP HZ” input terminal and the black lead to the “COM” terminal.
3. Connect the other end of the test leads to the circuit under test.
4. Read the display. If a “-“ sign is displayed, the red leads is the negative potential.
9. AC true RMS voltage measurement
1. Set the function switch range position.
2. Press the MODE button to select the AC Volt range. The meter automatically defaults to DC Volts.
3. Insert the red test lead into the “V Ὠ TEMP HZ” input terminal and the black lead to the “COM” terminal.
4. Connect the other end of the test leads to the circuit under test. Take the reading on the display. Note:
The meter is set to default to “Auto Range” mode. Pressing the PEAK hold button will allow the user to
record the peak maximum and minimum readings on the AC volt range. See page 7 for further instructions for “PEAK” Hold function.
10. Resistance Measurement
1. Set the function switch to range position. The meter automatically defaults to resistance range.
2. Insert the test lead into the “V Ὠ TEMP HZ” input terminal and the black lead to the “COM” terminal.
3. Connect the other end of the test leads to the circuit or component under test. Take the reading on the display. Note:
The meter is set to default to “Auto Range” mode. WARNING before attempting to make a
resistance measurement, ensure there is no voltage present on the circuit under test.
11. Capacitance Measurement
1. Set the function switch to CAP range position.
2. Insert the test lead into the “V Ὠ TEMP HZ” input terminal and the black lead to the “COM” terminal.
3. Connect the other end of the test leads to the circuit or component under test. Take the readings on the display. Note:
In capacitance range the meter is Auto Ranging only. CAUTION to avoid damage to the meter or the
equipment under test, remove all power from the circuit and discharge all capacitors before measuring capacitance.
Large value capacitors should be discharged through
an appropriate resistance load. Use the DC Voltage function to confirm that the capacitors are discharged.
12. Frequency Measurement
1. Set the unction switch to range position. The
meter automatically defaults to Frequency range.
2. Press the MODE button for 3 seconds to select the
Frequency range. The meter automatically defaults to DC Voltage.
3. Insert the test lead into the “V Ὠ TEMP HZ”input
terminal and the black lead to the “COM” terminal.
4. Connect the other end of the test leads to the circuit
or component under test. Take the readings on the display. Note:
In the Frequency Range the meter is Auto Ranging
only.
13. Temperature Measurement
1. Set the functions switch to TEMP range position. The meter automatically defaults to ˚C range.
2. Insert the meter white temperature adaptor into the “V Ὠ TEMP HZ” input terminal and the “COM” terminal. Ensure that the “-“ marking on the adaptor is inserted into the “COM” terminal and the “+” marking on the adaptor is inserted into the “V Ὠ TEMP HZ” input terminal.
3. Connect any K-Type probe into the meter adaptor and measure the temperature of the apparatus or area required.
4. Read the temperature directly on the display.
5. To change the measuring unit from ˚C to ˚F, press the MODE button to select the ˚F unit. Note:
The meter is set to default to “Auto Range” mode.
14. Diode Test
1. Set the function switch to range position.
2. Press the MODE button to select the Diode Test range. The meter automatically defaults to resistance.
3. Insert the red test lead into the “V Ὠ TEMP HZ” input terminal and the black lead to the “COM” terminal.
4. Connect the other end of the test leads to the components under test. Read the display. Note:
Use the diode test to check diodes, Transistors, silicon
controlled rectifiers (SCR’s) and other semiconductor devices.
The test sends a current through a semiconductor
junction, then measures the junction’s voltage drop.
Normal forward voltage drop (forward bias) for a good
silicon diode is between 0.4V to 0.9V. a reading higher than that indicates a leaky (defective) diode. A zero reading indicates a shorted diode.
Reverse the test leads connection (reverse bias) across the diode. The display shows “OL” if the diode is good. Any other readings indicate the diode is shorted or resistive (defective).
15. Continuity Testing
1. Set the function switch to range position.
2. Press the MODE button to select the Continuity Test range. The meter automatically defaults to Resistance.
3. Insert the red test lead into the “V Ὠ TEMP HZ”input terminal and the black lead to the “COM” terminal.
4. Short the tip of the test leads and makes sure the display reads “0” and the buzzer beeps.
5. Connect the tip of the test leads to the circuit or component under test. The display reads the resistance and the buzzer beeps when the reading is not more than about 35Ὠ. Note:
Using resistance and continuity function in a live circuit
will produce false results and may damage the instrument.
In many cases the suspicious components must be
disconnected from the circuit under test to obtain accurate results.
WARNING before attempting to make a test, ensure
there is no voltage present on the circuit.
16. DATAHOLD
This is a function used to freeze the reading on the display, ideal for later viewing.
1. Press the HOLD button once. When the HOLD is
activated, the meter beeps, freezes the reading, and displays the “HOLD” indicator on the LCD.
2. To deactivate the “HOLD” function, press HOLD
button once, the meter will beep and the meter will start reading new measurements.
17. PEAK HOLD
This is a function used to record the maximum and minimum readings on the display for the voltage and current ranges.
1. Press the PEAK button once. This will record the
PEAK maximum, the meter beeps, and displays the “P MAX” indicator on the LCD.
2. Press the PEEK button again. This will record the
“PEAK” minimum, the meter beeps, and displays the “P MIN” indicator on the LCD.
3. To deactivate the “PEEK” function, press button for 3
seconds, the meter will beep and the meter will start reading new measurements.
18. LCD Backlight
The backlight is ideally used in dark or dimly lit areas.
1. Press the button for 3 seconds, and the bright
white light will come on.
2. To turn off the backlight press the button for 3
seconds. Note: the use of the backlight will reduce the battery life considerably.
19. Auto Power Off
This feature automatically turns off the meter after approximately 20 minutes from the last measurement taken. To turn on the meter, press any button or move the rotary switch to any position. Ensure the test leads are disconnected from any circuit to avoid injury or meter damage.
20. Battery Replacement
When the sign appears on the LCD, this indicates the battery should be replaced. Use the following procedure to replace the standard 9V battery (IEC 6F
22) battery.
1. Disconnect the test leads from any live source and
remove the leads from the input terminals.
2. Rotate the function switch to the OFF position.
3. The battery cover is secured to the bottom of the
case by a screw. Using a screwdriver remove the screw from the battery cover and remove the battery cover from the meter.
4. Remove the old battery and replace it with a new
IEC 6F 22 9V battery.
5. Replace the battery cover and reinstall the screw.
Loading...