Market Forge STM-E Installation Manual

Page 1
STERILMATIC
ANALOG ELECTRIC STERILIZER
INSTALLATION - OPERATION - MAINTENANCE
MODELS
44 Lakeside Avenue, Burlington, Vermont 05401 USA
Telephone: (802) 658-6600 Fax: (802) 864-0183
www.mi.com PN 14-0411 Rev C (12/15)
© 2015 - Market Forge Industries Inc.
Page 2
Your Service Agency’s Address:
Model
Serial number
Sterilizer installed by
Installation checked by
Page 3
IMPORTANT
TABLE OF CONTENTS
WARNING: Improper installa­tion, adjustment, alternation, service or maintenance can cause property damage, in­jury or death. Read the instal­lation, operation and mainte­nance instructions thoroughly before installing or servicing this equipment.
FOR YOUR SAFETY
Do not store or use gasoline or other ammable vapors or liq­uids in the vicinity of this or any other appliance.
The information contained in this manual is important for the prop­er installation, use, and mainte­nance of this sterilizer. Adher­ence to these procedures and instructions will result in satis­factory baking results and long, trouble free service. Please read this manual carefully and retain it for future reference.
INSTALLATION
Introduction .............................................................. 2
Service Connections ..................................................... 3
Installation Instructions ................................................... 4
Optional Cold Water Condenser ........................................... 5
Pan Supports and Bafes ................................................. 6
Optional Stands .......................................................... 8
Optional Recording Thermometer.......................................... 9
OPERATION
Operating Instructions ................................................... 10
Minimum Sterilization Times ............................................. 11
MAINTENANCE
Cleaning ............................................................... 12
Warranty ............................................................... 13
ERRORS: Descriptive, typo­graphic or pictorial errors are subject to correction. Speci­cations are subject to change without notice.
Page 4
Introduction
OPERATING ENVIRONMENTAL CONDITIONS
This unit is designed for commercial use and to be safe at least under the following conditions:
For indoor use only.
For use at altitudes up to 6500 ft (2000 m)
For use at temperatures from 41°F (5°C) to 104°F (40°C).
Maximum relative humidity 80% for temperatures up to 88°F (31°C) decreasing linearly to 50% relative humidity at 104°F (40°C).
Main supply voltage uctuations not to exceed ± 10% of nominal voltage.
Transient overvoltages according to Installation Cat­egories II (in accordance with IEC 664).
Pollution Degree 2 (in accordance with IEC 664).
INTENDED USE - DUTY CYCLE
This unit is intended to be operated intermittently. After a pre-heat cycle, the longest period of use is 60 minutes. The timer allows a maximum of 60 minutes. After each use the unit should be opened for removal and reloading of product. The water level should be checked after each
use and relled when necessary.
SERVICE & TECHNICAL INFORMATION CONTACT
WARNING
This unit should be serviced by qualied ser­vice personnel only.
Your Sterilmatic Sterilizer has been developed to answer the need for a compact, automatic, low-cost steam pres­sure sterilizer. The following instructions cover installa-
tion. Should service be required, it is readily available by
contacting our authorized service agency located nearest to you. The name of your local service company can be
obtained on our website, www.mi.com.
WATER CONDITIONS
When sterilizing culture mediums that generate sul-
de sas or chlorine gas, the inside of the chamber
must be cleaned and rinsed thoroughly without fail.
Market Forge from time to time is asked the question
about using distilled or deionized water for use with our Sterilizer models STM-E and STM-EL. We are always asked why these water choices are not allowed for use with our units and what would be recommended. To ad­dress this situation, we have complied the following as a
means of satisfying these questions:
1. We have found that the use of distilled or deionized water will aggressively attack the pure coat of Alumi­num Alclad, which protects the bottom surface from oxidizing and then eventually pitting (reference: Op­erating and Maintenance Instructions).
2. In addition pitting can also be caused by several other external environmental factors. Few examples are as follows. These conditions have been highlighted in our documentation.
Grains of hardness in the water supply should be as follows (.25 to 2).
A pH imbalance in the water supply can greatly affect the life to the aluminum cylinder. The pH range that would be recommended is between
7.0-8.5.
The lack of a positive electrical ground can cause an electrolytic reaction that will accelerate pitting.
Another contribution to accelerate pitting is the type of cleaning solutions used or the abrasive scrubbing pads. If a low pH is present with the detergents being used or an abrasive pad, the protective Alclad coating will be removed during the cleaning process.
Spillage of media being sterilized can also con­tribute to the accelerated pitting if it is corrosive.
INSTALLATION
CHLORINE LEVEL ≤ 1 PPM.
IMPORTANT
Market Forge will not be responsible for dam­age resulting from the use of hard or corro­sive water, from failure to drain the unit daily, or from inadequate cleaning procedures.
2
Page 5
Service Connections
ELECTRICAL SPECIFICATIONS
Domestic
Models
STM-E
STM-EL
Export
Models
STM-EX
STM-ELX
kW Hz
9.3 60 45A - 26A -
12.4 60 - 52A - 30A
kW Hz
10.4 50 48A - 16A -
12.4 50 - 52A - 18A
1 Phase
208V 240V 208V 240V
1 Phase
220V 240V 220/380V 240/415V
Unit must be grounded. Main supply voltage fluctuations are not to exceel ± 10% nominal supply voltage.
3 Phase
3 Phase
WATER SUPPLY
Good quality water feed is the responsibility of the owner. Water qual­ity must be within the following general guidelines.
TDS: 40-125 ppm Hardness: 35-100 ppm Chlorine: <0.2 ppm Silica: <13 ppm Chlorides: <25 ppm Chloramine: <0.2 ppm pH: 7.0 - 8.5
The best defense against poor water quality is a water treatment system designed to meet your water quality conditions.
Appliance to be installed with backflow protection according to fed­eral, state or local codes.
SERVICE CONNECTIONS
A Drain - 1/2” (13mm) N.P.T. of 3/8” (10mm) OD copper
(see note 1)
B Steam Exhaust Connection - 5/8” (16mm) (see note 2)
C Electrical Connection - (*see electrical specifications
table)
D Power Supply
NOTES:
1. An air break must be provided if a unit drain line is run.
2. Vent exhaust to atmosphere. B1 is actual connection, but must exit casing at B.
IMPORTANT: Exhaust line must be vented to the outside to elimi­nate the exhaust steam and the accompanying noise from enter­ing the room. Use 1/2” (13mm) copper tubing or suitable alternate. The overall height and length of the line should not rise more then 4’ (1.2 meters) above the unit and exceed 15’ (4.5 meters) with a minimum of bends. The line should slope downward after leaving the sterilizer in order to ensure condensate drainage.
IMPORTANT: Failure to comply with this outline will affect the sterilization process.
When an exhaust condenser is supplied; the following services must be provided: 1/2” (13mm) cold water: 1” (25mm) waste: 115V electrical line.
DIMENSIONS ARE IN INCHES [MM]
18.69 [475]
3.5
[89]
15.5
[394]
D
9.34 [237]
C
A
CONDUIT
B
B1
SEE NOTE 2
1.5 [38]
4.75 [121]
25 [635]
30 [762]
4.63 [117]
2 [57]
23 [584]
1 [25]
6 [152]
Customer provided Drain Catch Pan
Approx. 4-6 quarts (3.8-5.7 liters)
B
Figure 1
SEE NOTE 2
B1
A
4 [102] ADJ. LEGS
DOOR IN
OPEN POSITION
21 [533] CHAMBER
9 [230]
9 [228]
CLEARANCE
Locate electric box as close to
C
sterilizer as possible and per
local electrical codes
12 [305]
11 [280]
CLEARANCE
CONDUIT
3 [76]
D
3
INSTALLATION
Page 6
Installation Instructions
INSTALLATION
Set sterilizer on counter, using the 6” (152mm) legs pro­vided or assemble the optional stainless steel stand with under-shelf.
If your Sterilmatic includes a water-cooled exhaust con­denser, we recommend the use of the Sterilmatic stand, part number 95-6060.
First, level unit in place, then adjust rear legs to pitch the unit forward 1/4” (6mm) to insure positive drainage of the cylinder.
ELECTRICAL
Connect to proper electrical supply box and disconnect switch as shown on one of the following schematic dia­grams - 208 or 240 volts, single or three phase. Connec­tion is made from the rear of the unit, through the conduit to the terminal box located at the front of the unit.
Electric supply connection for STM-E & STM-EL
Connect to proper electrical supply as indicated on name­plate on top of unit. The power supply cord is brought in from the rear of the unit, through the conduit and the con­nection is made at the terminal box located at the front of
the unit.
Electric supply connection for STM-EX & STM-ELX (export models)
Connect to proper electrical supply as indicated on name­plate on top of unit. Connection is made from the rear of the unit, through the conduit to the terminal box located at the front of the unit. All control circuits are 220 volts.
In order to accomplish this, a current-carrying grounded neutral must be provided.
Thus, a three phase system must be 4-wires. Most elec-
trical codes require, and we recommend, that a separate
switch be located within sight of the sterilizer.
OUTSIDE VENTING
Connect 1/2” (13mm) nominal tubing exhaust to outside vent connection located on top of unit, within the control housing.
IMPORTANT: Exhaust line must be vented to the out­side to eliminate the exhausted steam and the accom­panying noise from entering the room.
Use 1/2” (13mm) copper tubing or suitable alternate. The overall height and length of the line should not rise more then 4 feet (1.2 meters) above the unit and exceed 15 feet (4.5 meters) with a minimum of bends. The line should slope downward after leaving the sterilizer in order to in­sure condensate drainage.
WATER-COOLED EXHAUST CONDENSER
If outside venting is not possible, an optional water-cooled condenser is available for connection to an open drain. If
required order part no. 95-0436 kit.
RECORDING THERMOMETER
If a recording thermometer is provided, refer to installation guide provided with recorder.
TRAY SUPPORTS
Install side tray supports. Tray supports are attached by means of key-hole clearance slots which are slipped over studs located on the sides of the Sterilmatic chamber.
BAFFLE INSTALLATION
To insure maximum drying of packs, a bafe is supplied with your Sterilmatic. Place perforated splash bafe in bottom of the sterilizing chamber. Install small bafe with
no perforation at the rear of the upper tray support chan­nel.
NOTE: The perforated bafe is not to be used as a shelf
to place media or other items. It is intended to eliminate splashing.
OPERATION CHECK
To check for proper operation of unit:
3. Close drain valve by turning handle clockwise.
WARNING
DO NOT OPEN DRAIN VALVE WHILE UNIT IS OPERATING. PREMATURE OPENING MAY RESULT IN SCALDING OF OPERATOR.
4. Fill chamber with 4 to 6 quarts (3.7 to 5.6 liters) of ordinary tap water. DO NOT USE DISTILLED OR DE­IONIZED WATER.
5. Close chamber door.
6. Set exhaust selector to INSTRUMENTS AND PACKS (fast exhaust) or LIQUIDS (slow exhaust).
7. Set timer to 15 minutes. Cycle will go to completion automatically.
NOTE: Cycle timer will not start until sterilizing
temperature is obtained.
INSTALLATION
4
Page 7
Optional Cold Water Condenser
Item Part No. Description Qty
1 95-2119 Steam condensing unit 1
2 95-2219 Thermostat Box Assy. 1
3 95-0086 Exhaust line 1
5 15-7057 Copper tubing 3/8 OD 22.25”
6 10-1775
7 10-2500 Lockwasher, 1/4 2
8 10-2308 Hex Nut, 1/4-20 2
9 10-1812
10 10-2505 Lockwasher, 10 2
11 10-2340 Hex Nut, 10-32 2
12 95-4009 Front Template (7” Lg)
13 95-4010 Back Template (11” Lg) 1
Rd. Hd. Mach. Screw, 1/4-20
Rd. Hd. Mach. Screw, 10­32
2
2
1
SERVICE CONNECTIONS REQUIRED
A 1/2” IPS Cold Water Connection
B 1” IPS Drain Connection (See Note 3)
C 115V Elec. Connection 7/8 Ø knockout (cond. unit)
D Electrical Connection
Figure 2
Figure 3
5
INSTALLATION
Page 8
Pan Supports and Bafes
1. Locate the mounting studs on the inside of the cham­ber. There are two rack mounting studs on each side.
Mounting Studs
Figure 4
2. Taking one pan support and positioning rack so that the pan stop is facing the rear of the unit and the wires are facing toward the center of the unit. The pan stop is a piece of sheet metal welded to the rack with a 65° bend.
4. After installing one pan support rack correctly, you
can install the upper bafe. Position the bafe so that
the 45° bend is facing up towards the front of the unit
(see “Figure 6”). Slide the mounting tab onto the at bend on the pan stop bracket. The bafe should now
stay in place by itself, but in a tilted state (see “Figure 7”).
Baffle
Figure 6
3. Begin to hang the pan support by placing the rear key-way slot onto the rear mounting stud and slide the rack until the slot sits on the mounting stud. When this is done correctly the front mounting stud will be in position to place the front key-way slot. Slide the rack down into its correct position.
Keyway Slot
Figure 5
Pan Stop
Figure 7
INSTALLATION
6
Page 9
5. Position the second pan support rack into the cavity
and slide the other mounting tab onto the rack at
bend while the pan support rack is not on the mount­ing studs. Hang the pan support by placing the rear key-way slot onto the rear mounting stud and slide the rack until the slot sits on the mounting stud. When this is done correctly the front mounting stud will be in position to place the front key-way slot. Slide the rack down into its correct position.
6. Place the Perforated Water Bafe so that it sits on the
bottom of the inside of the sterilizer chamber.
Pan Supports and Bafes
Perforated Water Baffle
Figure 8
CAUTION
Do not cover the holes in the Perforated Water
Bafe by using it as a shelf. This will result in a disrupted ow of steam.
7
INSTALLATION
Page 10
Optional Stands
STERILMATIC OPEN STAND:
Market Forge Sterilmatic Stand can be supplemented with an Optional Stand for utility use where maximum compactness is desired.
The sturdy, stainless steel unit is equipped with adjust­able leg extensions which allow the unit to be installed
and leveled over existing contours in the oor.
The open design lends itself to maximum sanitary condi­tions because of the ease with which periodic cleaning can be done.
Though simple in design and appearance, the sterilmatic stand is the ideal arrangement for mounting in that it al­lows secondary air to circulate.
STERILMATIC OPEN STAND WITH CONDENSER
Market Forge can provide the open stand with an optional steam condenser system for use where steam exhaustion into the room is undesirable.
The condenser is automatically controlled by the ther­mostat. The normal factory thermostat setting is 130°F (54°C). The open under-shelf of the stand gives added utility providing a handy tabouret for utensils and access for drainage of water from the sterilizing chamber.
Part No. 50 Hz Description
10-4653 10-4653 Thermostat
10-4035 10-7074 3/8” Solenoid
10-5731 10-5731 1/2” Water Stop Valve
95-2106 95-2106 Water Injection Assy.
95-1680 95-1680 Shelf
OPERATION
8
Page 11
Optional Recording Thermometer
INSTALLATION FOR OPTIONAL RECORDING THER­MOMETER
The optional recording thermometer may be surface or panel mounted whichever is preferred.
1. Remove the top cover of the sterilizer. Remove the
pipe plug located in the T-tting that also accommo­dates the safety relief valve and which is connected
directly to the sterilizing chamber. (See gure to the
right)
2. Insert the RTD temperature sensor of the tempera-
ture recorder through the rectangular cutout in the top cover then into the pipe from which the plug was removed.
3. Make sure RTD temperature sensor is protruding
down into the sterilizer cavity but not too much so as
to interfere with the sheet metal bafe plate. Tighten the compression tting. Replace top cover onto the
sterilizer.
4. For mounting the recorder on a wall, refer to the re-
corder manual supplied with the recorder.
5. Electrical connections: The 115/120VAC unit comes
with a 6’ power cord and three prong plug. This can be plugged into any 120V outlet. The 220/230VAC unit comes with a 6’ power cord and three prong plug. However, due to the many varieties of plugs and out­lets you may need to replace the plug with the ap­propriate plug.
The recording thermometer is a 24-hour clock and should be run continuously.
Charts should be changed every 24 hours.
TO REMOVE THE CHART
1. Unscrew and remove the chart ‘hub’ knob (attached to the chain).
Remove Plug
Figure 10
2. Grasp the edge of the chart and lift straight off.
TO INSTALL A NEW CHART
1. Slide edge of the chart under the two raised tabs lo­cated under the stylus.
2. Locate the center hole of the chart over the center, threaded rod. Rotate chart to line up the current time of day directly underneath the stylus print head.
3. Screw the chart ‘hub’ knob back in place, hand tight only.
Figure 9
9
INSTALLATION
Page 12
Operating Instructions
IMPORTANT
Make sure the drain valve is closed. Fill bot­tom of the sterilizer chamber with approxi­mately six quarts of water or just below ledge at bottom of door opening. (If water supply is known to be hard or corrosive, a source of treated water should be used.) DO NOT USE DISTILLED OR DEIONIZED WATER.
1. LOAD STERILIZER: Use proper sterilizer loading procedures when placing materials in sterilizer cham­ber. All solid containers or instruments must be placed so that water or air will not be trapped in them.
2. CLOSE DOOR: Grasp handle, and holding it in verti­cal position, pull door down until bottom of door rests in the bottom of door opening. Then rotate handle for­ward, engaging the lower curved portion under the horizontal rod in the casting at the bottom of the door opening. Push handle all the way down and back until door is locked securely in position.
3. SET EXHAUST SELECTOR: Located at center of the control housing mounted on top of the unit, to correct position. Unit is now ready to start. All items, other than solutions, may be sterilized with selector at
“Instruments”. Solutions require a low exhaust. Place selector at “Liquids”.
4. DETERMINE CORRECT STERILIZATION TIMES:
(Referring to next page for minimum required times in
the sterilization guide and times table.)
NOTE: In no case should the timer be set to less
than 15 minutes. Sterilization will not be ac­complished in less than 15 minutes exposure time.
5. TURN TIMER: Located at upper right front of steril­izer. Select desired length of sterilizing period. This turns power supply on and starts the cycle after pres­sure-temperature combination has been reached. Amber pilot light indicates that the timer is running.
6. When the sterilizer chamber reaches the selected temperature, the timed exposure cycle will begin. When the exposure cycle is completed, the electric supply will be opened automatically. When the cham­ber pressure gauge located at the top of the control housing reads “0”, the door may be opened. (Release handle and let go to avoid possible contact with re­maining steam.) When opening the door, allow a few seconds for steam to escape from chamber before opening completely.
7. To assist in drying racks, release door handle after pressure has been attained at start of cycle. Pres­sure in chamber will keep door closed. The use of a wire basket will provide better drying for dressings. At end of sterilizing cycle, release door handle and open slightly. Do not lift door to open position. This will allow steam and moisture to escape. Allow door to remain in this position for 15 to 20 minutes before removing load. Small packs can be dried successfully with this procedure. We do not recommend the ster­ilization of large packs, such as linens. Be sure con-
densate bafes are in position in the chamber.
8. Remove load and check water level for next opera­tion.
STERILIZATION GUIDE
PACKS (Linens, gloves, etc.): Use wire basket
to facilitate drying. Be sure condensate bafes are
in place. Place packs on edge and arrange load in chamber, so that only minimal resistant to passage of steam through the load will exist. NOTE: Place gloves in upper two-thirds of chamber.
JARS, CANISTERS (etc.): Place containers on side to allow for displacement of air and complete contact of steam to surfaces. Drying is also facilitated.
PETRI DISHES, PIPETTES, DESICCATORS (etc.): Should be inverted.
UTENSILS, TREATMENT TRAYS: Placed on edges to facilitate drying.
INSTRUMENT SETS: Place instruments set in trays
having mesh or perforated bottoms. Place trays at
on shelves.
COMBINING FABRICS & HARD GOODS: Place hard goods on lowest shelves.
PLASTIC UTENSILS: DO NOT stack or nest plastic items.
LIQUIDS: Sterilize medium liquids separately from other supplies or materials. Set exhaust selector to
proper position (liquids).
SMALL ITEMS: Sterilize small items in baskets, or trays.
OPERATION
10
Page 13
TIME (Minutes) ARTICLES
15 Glassware, empty, inverted.
Instruments, metal in covered or open tray, padded or unpadded.
Needles, unwrapped.
Pipettes, blood diluting, serological, volumetric, etc
Tubing glass (6mm), (10mm) inverted
20 Flasked solutions 75-250 ml.
Instruments, metal combined with other materials in covered and/or padded tray.
Instruments wrapped in double thickness muslin.
Rubber gloves, catheters, drains, tubing, etc. Unwrapped or wrapped in muslin or paper.
30 Brushes in dispensers, in cans of individually wrapped.
Dressings, wrapped in paper or muslin, small packs only.
Flasked solutions 500-1000 ml.
Syringes, unassembled, individually packaged in muslin or paper.
Minimum Sterilization Times
Needles, luer, individually packaged in glass tubes or paper.
45 Flasked solutions 1500-2000 ml.
11
OPERATION
Page 14
Cleaning
DAILY CLEANING PROCEDURE (AT THE END OF EACH DAY):
When sterilizing culture mediums that generate sul-
de gas or chlorine gas, the inside of the chamber
must be cleaned and rinsed throughly without fail.
1. Remove bottom splash bafe.
IMPORTANT
STERILIZING CHAMBER MUST BE CLEANED AND DRAINED DAILY USING THE FOLLOW­ING PROCEDURE. WASH WETTED PORTION OF THE CYLINDER THOROUGHLY BY ADDING A MILD DETERGENT TO WATER IN CYLIN­DER.
2. If a soft cloth or brush is used with the detergent and
does not completely remove the surface lm, a nylon
soap pad should be used. After washing thoroughly rinse with clean water. Dry cylinder* and leave door open overnight.
* The Sterilmatic cylinder is constructed of corrosion re­sistant Alclad aluminum alloy. The protective properties of this material afforded to the interior portion of the cylinder which is exposed to water may be destroyed by allowing
a lm to form. Such a lm can be caused by salts or other
contaminants in the water. Corrosion may also occur if water is not drained daily.
MAINTENANCE
12
Page 15
Warranty
STERILIZER (AUTOCLAVE) WARRANTY
MODELS: STM-E, STM-EL, STM-EX* AND STM-ELX*
We warrant to the original purchaser that the sterilizers manufactured by Market Forge Industries, Inc. will be free from de­fects in material and factory workmanship if properly installed and operated under normal conditions. Within one year from date of original installation, or within 15 months from date of shipment from factory, whichever is sooner, we will repair or replace that part of any such machine that becomes defective at no cost to the customer.
This warranty is effective for One (1) Year Parts and 90 Days Labor, Travel and Mileage.
This warranty does not apply to damage resulting from use of hard or corrosive water, from failure to drain and dry cylinder daily or from inadequate cleaning procedures. Nor does it cover any part or assembly, which has been subjected to ac­cident, alteration, or is from a machine where the serial number has been removed or altered. Normal service adjustments are not covered by this warranty.
Any defect during the warranty period shall be brought to the attention of a factory authorized service agency or the dealer
from whom the equipment was purchased. He will be authorized to furnish or arrange for repairs or replacements within
the terms of the warranty.
NOTE: This warranty only applies to the USA and Canada. Elsewhere, warranty covers parts only for one year as de-
scribed above.
* Export Model.
MARKET FORGE INDUSTRIES
44 Lakeside Avenue
Burlington, Vermont 05401 USA
Telephone: (802) 658-6600
Fax: (802) 864-0183
www.mi.com
13
MAINTENANCE
Loading...