Mark-10 TT01, MTT01-100, MTT01-50 User Manual

Page 1
Series TT01
DIGITAL CAP TORQUE TESTERS
User’ s Guide
Page 2
Series TT01 Digital Cap Torque Testers User’s Guide
Thank you…
Thank you for purchasing a Mark-10 Series TT01 digital cap torque tester, designed to measure bottle cap application and removal torque.
With proper usage, we are confident that you will get many years of great service with this product. Mark-10 instruments are ruggedly built for both laboratory and industrial environments.
This User’s Guide provides setup, safety, and operation instructions. Dimensions and specifications are also provided. For additional information or answers to your questions, please do not hesitate to contact us. Our technical support and engineering teams are eager to assist you.
Before use, each person who is to use a Series TT01 digital cap torque tester should be fully trained in appropriate operation and safety procedures.
TABLE OF CONTENTS
OVERVIEW .........................................................2
POWER ...............................................................3
SETUP ................................................................4
HOME SCREEN AND CONTROLS ...................5
OPERATING MODES .........................................7
CHANGING THE UNITS .....................................7
DIGITAL FILTERS ..............................................7
SET POINTS .......................................................8
DATA MEMORY AND STATISTICS ..................9
COMMUNICATIONS ........................................10
CALIBRATION .................................................17
PASSWORDS ...................................................20
OTHER SETTINGS ...........................................22
SPECIFICATIONS ............................................24
Page 3
Series TT01 Digital Cap Torque Testers User’s Guide
1 OVERVIEW
1.1 List of included items Qty. Part No. Description
1 MTT01-12 /
Series TT01 digital cap torque tester MTT01-50 / MTT01-100
1 CT004 Sample gripping posts, 4 1 08-1026 Battery (inside the instrument)
1 - Certificate of calibration 1 09-1165 USB cable 1 - Resource CD (USB driver, user’s guides, MESURTM Lite software,
MESUR
TM
gauge DEMO software, User’s Guide) 1 CT001 Flat jaws set, 2 (optional) 1 CT002 Carrying case (optional) 1 CT003 Adjustable jaws set, 2 (optional)
1.2 Safety / Proper Usage
Caution!
Note the torque tester’s capacity before use and ensure that the capacity is not exceeded. Producing a torque greater than 150% of the tester’s capacity can damage the internal sensor. An overload can occur whether the tester is powered on or off.
The tester is designed primarily for the testing of threaded bottle caps, although other items can be tested as well. Items that should not be used with the tester include potentially flammable substances or products, items that can shatter in an unsafe manner, and any other components that can present an exceedingly hazardous situation when acted upon by a load.
The following safety checks and procedures should be performed before and during o peration:
1. Never operate the tester if there is any visible damage to the AC adapter or the tester itself.
2. Ensure that the tester is kept away from water or any other electrically conductive liquids at all times.
3. The tester should be serviced by a trained technician only. AC power must be disconnected and the tester must be powered off before the housing is opened.
4. Always consider the characteristics of the sample being tested before initiating a test. A risk assessment should be carried out beforehand to ensure that all safety measures have been addressed and implemented.
5. Wear eye and face protection when testing, especially when testing brittle samples that have the potential to shatter under force. Be aware of the dangers posed by potential energy that can accumulate in the sample during testing. Extra bodily protection should be worn if a destructive failure of a test sample is possible.
6. In certain applications, such as the testing of brittle samples that can shatter, or other applications that could lead to a hazardous situation, it is strongly recommended that a machine guarding system be employed to protect the operator and others in the vicinity from shards or debris.
7. When the tester is not in use, ensure that the power is turned off.
Page 4
Series TT01 Digital Cap Torque Testers User’s Guide
2 POWER
The TT01 is powered either by an 8.4V NiMH rechargeable battery or by an AC adapter. Since these batteries are subject to self discharge, it may be necessary to recharge the unit after a prolonged period of storage. Plug the accompanying charger into the AC outlet and insert the charger plug into the receptacle on the tester (refer to the illustration below). The battery will fully charge in approximately 8 hours.
USB connector
Serial connector
Power input jack
Caution!
Do not use chargers or batteries other than supplied or instrument damage may occur.
If the AC adapter is plugged in, an icon appears in the lower left corner of the display, as follows: If the AC adapter is not plugged in, battery power drainage is denoted in a five-step process:
1. When battery life is greater than 75%, the following indicator is present:
2. When battery life is between 50% and 75%, the following indicator is present:
3. When battery life is between 25% and 50%, the following indicator is present:
4. When battery life is less than 25%, the following indicator is present:
5. When battery life drops to approximately 2%, the indicator from step 4 will be flashing.
Several minutes after (timing depends on usage and whether the backli ght is turned on or off), a message will appear, “BATTERY VOLTAGE TOO LOW. POWERING OFF”. A 4-tone audio indicator will sound and the tester will power off.
The tester can be configured to automatically power off following a period of inactivity. Refer to the Other Settings section for details.
If battery replacement is necessary, the battery may be accessed by removing the cover attached to the underside of the base.
Page 5
Series TT01 Digital Cap Torque Testers User’s Guide
3 SETUP
Place the sample between the posts or jaws of the tester, and tighten, using the knob. Always ensure that the sample is gripped firmly, however, excessive tightening of thin-walled closures may result in sample deformation, possibly affecting test results. When the sample is secured, gradually exert torque on the bottle cap by hand.
3.1 Sample gripping methods
Three different sample gripping methods are available, as shown below:
Posts (included) Flat jaws (optional) Adjustable jaws (optional)
3.2 Proper alignment
The center of the bottle cap should be positioned axially with respect to the center of the loading table. Side loading or off-center loading may produce erroneous readings, and can damage the instrument.
3.3 Mounting to a bench The tester can be mounted to a bench using the four threaded holes on the underside of the base.
Page 6
Series TT01 Digital Cap Torque Testers User’s Guide
4 HOME SCREEN AND CONTROLS
4.1 Home Screen
10
No. Name Description 1 Measurement
2 Peaks 3 Primary reading 4 Units
5 Load bar
6 Mode
7 Number of stored 8 Battery / AC
9
8
direction indicator
data points adapter indicator
7
– indicates clockwise direction – indicates counter-clockwise direction
These indicators are used throughout the display and menu. The maximum measured clockwise and counter-clockwise readings. These readings are reset by pressing ZERO or by powering the tester off and on. The current displayed load reading. See Operating Modes section for details. The current measurement unit. Abbreviations are as follows: ozFin – Ounce-inch lbFin – Pound-inch lbFft – Pound-foot Ncm – Newton-centimeter Nm – Newton-meter Analog indicator to identify when an overload condition is imminent. The bar increases either to the right or left from the midpoint of the graph. Increasing to the right indicates clockwise load, increasing to the left indicates counter­clockwise load. If set points are enabled, triangular markers are displayed for visual convenience. This indicator reflects the actual load, which may not correspond to the primary reading (depends on operating mode). The ZERO key does not reset the load bar. See Operating Modes section for details. The current measurement mode. Abbreviations are as follows: RT – Real Time PCW – Peak Clockwise PCCW – Peak Counter-clockwise See Operating Modes section for details about each of these modes The number of stored data points in memory, up to 1000. Displayed only if Memory Storage is enabled for the DATA key. Either the AC adapter icon or battery power icon will be shown, depending on power conditions. Refer to the Power section for details.
1
2
3
4
6
5
Page 7
Series TT01 Digital Cap Torque Testers User’s Guide
9 High / low limit
indicators
Correspond to the programmed set points. Indicator definitions are as follows:
– the primary reading is greater than the upper load limit – the primary reading is between the load limits
– the primary reading is less than the lower load limit
10 Set points
The programmed load limit values. Typically used for pass/fail type testing. One, two, or no indicators may be present, depending on the configuration shown in the Set Points menu item.
4.2 Controls
Primary
Label Primary Function
Powers the tester on and off. Press briefly to power on, press and hold to power off. Active only when the
Secondary
Label Secondary Function
ENTER
Various uses, as described in the following sections.
home screen is displayed.
ZERO MENU
Zeroes the primary reading and peaks. Accesses the main menu.
(UP)
ESCAPE
Navigates up through the menu and sub-menus. Reverts one step backwards through the menu hierarchy.
MODE DATA
Toggles between measurement modes. Stores a value to memory and/or transmits the current reading to an external device, depending on configuration.
(DOWN) DIRECTION
Navigates down through the menu and sub-menus. Toggles between clockwise and counter-clockwise directions while configuring set points and other menu functions.
Note: Measurement units are configured through the menu. Refer to the Changing The Units section for details.
4.3 Menu navigation basics
Most of the tester’s various functions and parameters are configured through the main menu. To access the menu press MENU. Use the UP and DOWN keys to scroll through the items. The cu rrent selection is denoted with clear text over a dark background. Press ENTER to select a menu item, then use UP and DOWN again to scroll through the sub-menus. Press ENTER again to select the sub-menu item.
For parameters that may be either selected or deselected, press ENTER to toggle between selecting and deselecting. An asterisk (*) to the left of the parameter label is used to indicate when the parameter has been selected.
For parameters requiring the input of a numerical value, use the UP and DOWN keys to increment or decrement the value. Press and hold either key to auto-increment at a gradually increasing rate. When the desired value has been reached, press ENTER to save the change and revert back to the sub-menu item, or press ESCAPE to revert back to the sub-menu item without saving. Press ESCAPE to revert one step back in the menu hierarchy until back into normal operating mode.
Refer to the following sections for details about setting up particular functions an d parameters.
Page 8
Series TT01 Digital Cap Torque Testers User’s Guide
5 OPERATING MODES
Caution!
In any operating mode, if the capacity of the instrument has been exceeded by more than 110%, the display will show “OVER” to indicate an overload. A continuous audible tone will be sounded (if beeps are enabled) until the MENU key has been pressed or the load has been reduced to a safe level.
Three operating modes are possible with the TT01 torque tester. To cycle between the modes, press
MODE while in the home screen.
5.1 Real time (RT)
The primary reading corresponds to the live measured reading.
5.2 Peak Clockwise (PCW)
The primary reading corresponds to the peak clockwise reading observed. If the actual load decreases from the peak value, the peak will still be retained in the primary reading area of the display. Pressing
ZERO will reset the value.
5.3 Peak Counter-clockwise (PCCW)
Same as above, but for counter-clockwise readings.
6 CHANGING THE UNITS
The TT01 can display five different measurement units. To change the unit, select Units from the menu. The display will list the available units, as follows:
UNITS
ozFin * lbFin lbFft Ncm Nm
The tester will always power on with the unit selected in this sub-menu.
7 DIGITAL FILTERS
Digital filters are provided to help smooth out the readings in situations where there is mechanical interference in the work area or test sample. These filters utilize the moving average technique in which consecutive readings are pushed through a buffer and the displayed reading i s the average of the buffer contents. By varying the length of the buffer, a variable smoothing effect can be achieved. The selection of 1 will disable the filter since the average of a single value is the value itself.
To access digital filter settings, select Filters from the menu. The display will appear as follows:
Page 9
Series TT01 Digital Cap Torque Testers User’s Guide
DIGITAL FILTERS
(1 = Fastest)
Current Reading 8 Displayed Reading 1024
Two filters are available:
Current Reading – Applies to the peak capture rate of the instrument. Displayed Reading – Applies to the primary reading on the display.
Available settings: 1,2,4,8,16,32,64,128,256,512,1024. It is recommended to keep the current reading filter at its lowest value for best performance, and the displayed reading filter at its highest value for best stability.
8 SET POINTS
8.1 General Information
Set points are useful for tolerance checking (pass/fail) and triggering an external device in process control applications. Two limits, high and low, are programmed in the tester, and the primary reading is compared to these limits. The results of the comparisons are indicated via on-screen indicators as well as through the three outputs provided on the 15-pin connector, thus providing “under”, “in range”, and “over” signaling. These outputs can be connected to indicators, buzzers, or relays as required for the application. On-screen indicators are described in the next sub-section.
8.2 Configuration
To configure set points, select Set Points from the menu. The screen will appear as follows:
SET POINTS Upper Disabled * Upper Enabled
5.00 Lower Disabled * Lower Enabled
3.50
Either one, two, or none of the set points may be enabled. To toggle between the clockwise and counter­clockwise directions, press the DIRECTION key.
If two set points have been enabled, they are displayed in the upper left corner of the display. If only one set point has been enabled, the word “OFF” will appear in place of the value. If no set points have been enabled, the upper left corner of the display will be blank.
Page 10
Series TT01 Digital Cap Torque Testers User’s Guide
8.3 Set Point Indicators
When set points are enabled, the following indicators are shown to the left of the primary reading:
– the displayed value is greater than the upper load
limit (NO GO HIGH)
– the displayed value is between the limits (GO)
– the displayed value is less than the lower load
limit (NO GO LOW)
Note: Set point indicators reference the displayed reading, not necessarily the current live load.
9 DATA MEMORY AND STATISTICS
Series TT01 testers have storage capacity of 1,000 data points. Readings may be stored, viewed, and output to an external device. Individual, or all, data points may be deleted. Statistics are calculated for the data presently in memory.
Individual data points may be saved to memory by pressing the DATA key, or automatically via the Break Detection function, whichever is enabled. Refer to the Communications section for details. Once data storage has been enabled, the data record number 0000 will appear below the primary reading in the home screen. The record number will increment each time DATA is pressed or the automatic data storage function is activated. When memory is full the message “MEMORY FULL” will be flashed at the bottom of the display and a double audio tone will be sounded.
To view, edit, and output stored readings and statistics, select Memory from the menu. The screen appears as follows:
MEMORY
View Data
View Statistics Output Data Output Statistics Output Data & Stats Clear All Data
9.1 View Data
All the saved data points may be viewed. The record number is displayed, along with the corresponding value and presently set unit of measurement. Any readings may be deleted individually. To do so, scroll to the desired reading and press DELETE. The letter “D” will appear to the left of the record number, indicating that the gauge is in Delete mode, as follows:
0001 2.458 lbFin 0002 2.224 lbFin 0003 2.446 lbFin 0004 1.890 lbFin D 0005 2.098 lbFin 0006 1.998 lbFin 0007 2.042 lbFin
Page 11
Series TT01 Digital Cap Torque Testers User’s Guide
Press ENTER to delete the value. To exit Delete mode, press DELETE again. Any number of readings may be individually deleted, however, all readings may also be cleared simultaneously. Refer to the Clear
All Data sub-section for details.
9.2 Statistics
Statistical calculations are performed for the saved values. Calculations include number of readings, minimum, maximum, mean, and standard deviation.
9.3 Output Data
Press ENTER to output data to an external device. The display will show, “SENDING DATA…”, then “DATA SENT”. If there was a problem with communication, the display will show, “DATA NOT SENT”. Saved data can be downloaded by some Mark-10 data collection programs. Refer to their respective user’s guides for details.
9.4 Output Statistics
Press ENTER to output statistics to an external device. The display will show, “SENDING STATS…”, then “STATS SENT”. If there was a problem with communication, the display will show, “STATS NOT SENT”.
9.5 Output Data & Stats Press ENTER to output data and statistics to an external device. The display will show, “SENDING DATA”, then “SENDING STATS…”, then “DATA SENT”, then “STATS SENT”. If there was a problem with communication, the display will show, “DATA NOT SENT” and/or “STATS NOT SENT”.
9.6 Clear All Data Press ENTER to clear all data from the memory. A prompt will be shown, “CLEAR ALL DATA?”. Select Yes to clear all the data, or No to return to the sub-menu.
For output of data and/or statistics, RS-232 or USB output must be enabled. Data formatting is <CR><LF> following each value. Units can be either included or excluded. Output of data via the Mitutoyo output is possible, however, output of statistics is not. Refer to the Communications section for details.
Note: Data is not retained while the gauge is powered off.
10 COMMUNICATIONS
Communication with the TT01 tester is achieved through the micro USB or 15-pin seri al ports located in the rear of the housing, as shown in the illustration in the Power section. Communication is possible only when the tester is in the main operating screen (i.e. not in a menu or configuration area).
10.1 Installing the USB driver It is recommended that the USB driver be installed before physically connecting the tester to the
PC with a USB cable.
1. Insert the Resource CD supplied with the tester into the CD/DVD drive in the computer. Then, navigate in Windows Explorer or My Computer to one of the following folders on the CD:
Windows 2000 through Vista - “Win_2K_XP_S2K3_Vista” Windows 7 - “Windows_7”
2. Execute the installer application “Mark-10USBInstaller.exe” by double-clicking it. When the program launches, one of the following windows will appear, depending on the operating system:
10
Page 12
Series TT01 Digital Cap Torque Testers User’s Guide
or
Click “Install”.
3. The next screen appears as follows:
Click “Continue Anyway”.
4. After installation completes the following screen may appear in non-Windows 7 operating systems.
Restart the computer before connecting a Mark-10 USB device.
5. After Windows has restarted, plug in the device. The following will occur: Windows 7 Operating Systems When the Mark-10 USB device has been plugged into a USB
port, the driver will automatically be found. When the driver installation is complete, a message will appear as follows: “The MARK-10 USB DEVICE driver is now installed and ready to use”.
Non-Windows 7 Operating Systems When the Mark-10 USB device has been plugged into a USB port, the following screen appears:
11
Page 13
Series TT01 Digital Cap Torque Testers User’s Guide
Select “No, not this time”, then click “Next”.
6. The next screen appears as follows:
Select “Install the software automatically (Recommended)”, then click “Next”.
7. The next screen appears as follows:
Click “Continue Anyway”.
12
Page 14
Series TT01 Digital Cap Torque Testers User’s Guide
8. The next, and final, screen appears as follows:
Click “Finish”. The Mark-10 USB device is now installed and ready to use. The COM port number assigned by Windows may be identified in Device Manager, or in the communication application being used, such as MESURgauge or HyperTerminal.
10.2 Serial / USB
To set up RS-232 and USB communication, select Serial/USB Settings from the menu. The screen appears as follows:
SERIAL/USB SETTINGS RS-232 Selected * USB Selected + Baud Rate + Data Format
Select either RS-232 or USB output. Configure the baud rate and data format as required for the application. Default values are as follows:
Baud Rate: 9,600
Data Format: Numeric + units Auto Output: Disabled
Other communication settings are permanently set to the following:
Data Bits: 8 Stop Bits: 1 Parity: None
Individual data points may be transmitted by pressing DATA. Continuous data may also be requested from a PC or other external device. Automatic output of individual data points is also possible. Refer to the Automatic Output and Zeroing sub-section for details.
13
Page 15
Series TT01 Digital Cap Torque Testers User’s Guide
%
10.3 Command Set Series TT01 testers will respond to the following ASCII commands via USB or RS-232:
? Request the displayed reading MEM Transmit all stored readings STA Transmit statistics
All commands must be terminated with a Carriage Return character or with a Carriage Return/Line Feed combination. The tester’s responses are always terminated with a Carriage Return/Lin e Feed. Any detected errors are reported back by means of error code *10 (illegal command).
10.4 Mitutoyo BCD settings This output is useful for connection to data collectors, printers, multiplexers, or any other device capable of accepting Mitutoyo BCD data. Individual data points may be transmitted by pressing DATA or by requesting it from the Mitutoyo communication device (if available). Automatic output of individual data points is also possible. Refer to the Break Detection sub-section for details. To enable Mitutoyo output, select the desired format – either with polarity or without polarity. The screen appears as follows:
MITUTOYO BCD
* Disabled Ena w/o Polarity Ena w/Polarity
10.5 Break Detection Three functions can be triggered automatically upon bottle cap removal (break detection), defined as a 50% drop in load from the peak. These functions automate the process of transmitting and saving multiple individual test data, as well as zeroing the display. If beeps are enabled, an audible tone will sound when the automatic output, automatic storage, and automatic zero functions have occurred. The display will appear as follows:
BREAK DETECTION + Auto Output Auto Storage Auto Zero + Break Settings Trig. Threshold 10
Function Description
Auto Output
Automatically transmits the peak torque reading upon break detection. Press ENTER to access the Auto Output Settings sub-menu, where the output type can be selected. Refer to Section 10.5.1 for details.
Auto Storage Auto Zero Break Settings
Automatically stores the peak to memory upon break detection. Automatically zeroes the display following data transmission and/or storage. Press ENTER to access the Break Detection Settings sub-menu to configure the trigger for auto output and storage, and the delay to zero the display. Refer to Section 10.5.2 for details.
14
Page 16
Series TT01 Digital Cap Torque Testers User’s Guide
10.5.1 Auto Output Settings
Select the output type. One, both, or no output types may be enabled. The display will appear as follows:
AUTO OUTPUT
SETTINGS
RS-232/USB Output Mitutoyo Output
10.5.2 Break Detection Settings
Set up the trigger for auto output and storage, and the delay to zero the display The display will appear as follows:
BREAK DETECTION
SETTINGS
Trig. Threshold 10 % Auto Zero Delay 5 sec.
Trig. Threshold
Auto Zero Delay
Sets the percentage of full scale at which the break detection feature becomes active. This threshold is provided to ignore torque peaks that can occur during sample loading and unloading. Available settings: 5-50%, in 5% increments. Sets the time delay before the torque value is cleared from the screen. Available settings: 1-60 sec., in 5 sec. increments.
10.6 Analog Output This output can be used for chart recorders, oscilloscopes, data acquisition systems, or any other compatible devices with analog inputs. The output produces ±1 volt at full scale of the instrument. The polarity of the signal is positive for clockwise and negative for counter-clockwise.
10.7 DATA Key Settings
To configure the functions of the DATA key, select DATA Key from the menu. The display will appear as follows:
DATA KEY
* RS232/USB Output Mitutoyo Output Memory Storage
RS232/USB Output Mitutoyo Output Memory Storage
Outputs data via the serial and USB ports Outputs data via Mitutoyo (Digimatic) through the serial port Stores a reading to memory (refer to the Memory section for details)
Any combination of the above functions may be selected.
15
Page 17
Series TT01 Digital Cap Torque Testers User’s Guide
10.8 I/O Connector Pin Diagram (female)
Pin No. Description Input / Output
1 Signal Ground --­2 Counter-clockwise Overload Output 3 RS-232 Receive Input 4 RS-232 Transmit Output 5 +12V DC Output 6 Analog Output Output 7 Clockwise Overload Output 8 Mitutoyo Clock
Output
Output Bit 2
9 Mitutoyo Data
Output
Output Bit 0
10 Mitutoyo Request
Input
Input Bit 3 11 “Under” Set Point Output 12 “Over” Set Point Output 13 “Within” Set Point Output 14 --- --­15 Mitutoyo Ready
Output
Output Bit 1
16
Page 18
Series TT01 Digital Cap Torque Testers User’s Guide
11 CALIBRATION
11.1 Initial Physical Setup
The TT01 should be mounted to a fixture rugged enough to withstand a load equal to the full capacity of the instrument. Suitable certified calibration equipment is required, and caution should be taken while handling such equipment. A calibration kit is available from Mark-10.
11.2 Calibration Procedure
1. Select Calibration from the menu. The display will appear as follows: CALIBRATION
ENTER # CAL POINTS
(1 TO 10)
CLOCKWISE:
5
COUNTER-CLOCKWISE:
5
The tester can be calibrated at up to 10 points in each direction. Enter the number of calibration points for each direction. At least one point must be selected for each direction.
Note: To achieve the accuracy specification of ±0.3% of full scale, it is recommended to calibrate the TT01 at 5 or more even increments in both the clockwise and counter-clockwise directions. For example, Model MTT01-50 (with capacity of 50 lbFin) should be calibrated at 10, 20, 30, 40, and 50 lbFin loads in each direction.
2. To escape the Calibration menu at any time, press ESCAPE. The display will appear as follows: CALIBRATION
NOT COMPLETE
CANCEL EXIT W/O SAVING
Selecting “CANCEL” will revert back to the Calibration setup. Selecting “EXIT W/O SAVING” will return to the menu without saving changes.
3. After the number of calibration points has been entered, press ENTER. The display will appear as
follows:
CALIBRATION
OFFSET
Place tester vertical THEN PRESS ZERO
4. Place the TT01 in a fixture free from vibration, then press ZERO. The tester will calculate offsets,
and the display will appear as follows:
17
Page 19
Series TT01 Digital Cap Torque Testers User’s Guide
CALIBRATION
OFFSET
Please wait…
CALIBRATION
CALIBRATION
OFFSET
Sen.Offset Adj.Passed Ana.Offset Adj.Passed
Sen.Offset Adj.Failed Ana.Offset Adj.Failed
If failed:
5. The following screen appears after the offsets have been calculated: CALIBRATION
CLOCKWISE
Attach necessary weight fixtures.
THEN PRESS ENTER
Attach weight fixtures (cable attachment, cable, hook, etc), as required. Do not yet attach any weights or apply any calibration loads. Then press ENTER.
6. The display will appear as follows: CALIBRATION
CLOCKWISE
Optionally exercise load cell a few times.
THEN PRESS ENTER
Optionally exercise the internal sensor several times (at full scale, if possible), then press ENTER.
7. The display will appear as follows: CALIBRATION
CLOCKWISE
GAIN ADJUST
APPLY FULL SCALE LOAD
50.000 LBFIN +/-20%
THEN PRESS ENTER
OFFSET
Apply torque equal to the full scale of the instrument, then press ENTER.
18
Page 20
Series TT01 Digital Cap Torque Testers User’s Guide
8. After displaying “PLEASE WAIT…” the display will appear as follows: CALIBRATION
CLOCKWISE ENSURE NO LOAD THEN PRESS ZERO
Remove the torque applied in Step 7, leave the fixtures in place, then press ZERO.
9. The display will appear as follows: CALIBRATION
CLOCKWISE
APPLY LOAD
1 OF 5
ENTER LOAD:
10.000 LBFIN
THEN PRESS ENTER
Use the UP and DOWN keys to adjust the torque value as required. The torque values default to even increments, as indicated by the previously entered number of data points described in Step
1. Then press ENTER. Repeat the above step for the number of data points selected.
10. After all the clockwise calibration points have been completed, the display will appear as follows:
CALIBRATION
CLOCKWISE COMPLETE
REVERSE DIRECTION
FOR CCW Attach necessary weight fixtures.
THEN PRESS ENTER
Press ENTER.
11. At the completion of the counter-clockwise calibration, the display will appear as follows: CALIBRATION
COMPLETE
SAVE & EXIT EXIT W/O SAVING
To save the calibration information, select “SAVE & EXIT”. To exit without saving the data select “EXIT W/O SAVING”.
12. Any errors are reported by the following screens:
19
Page 21
Series TT01 Digital Cap Torque Testers User’s Guide
CALIBRATION
Units must be kgFmm.
PLEASE TRY AGAIN
PRESS ENTER
Displayed at the start of calibration if a disallowed unit is selected.
LOAD NOT STABLE
PLEASE TRY AGAIN
Ensure that the load is not swinging, oscillating, or vibrating in any manner. Then try again.
CALIBRATION
CLOCKWISE
LOAD TOO LOW
PLEASE TRY AGAIN
The calibration load does not match the set value.
CALIBRATION
COUNTER-CLOCKWISE
LOAD TOO CLOSE
TO PREVIOUS
PLEASE TRY AGAIN
The entered calibration point is too close to the previous point.
12 PASSWORDS
Two separate passwords may be set to control access to the Calibration section and to the menu and other keys. To access the passwords setup screen, select Passwords from the menu. The display will appear as follows:
PASSWORDS Calibration Menu Key Mode Key Zero Key Data Key
20
Page 22
Series TT01 Digital Cap Torque Testers User’s Guide
12.1 Calibration Password
Select Calibration from the sub-menu. The display will appear as follows:
CALIBRATION PASSWORD
* Disabled Enabled Set Password (0000 – 9999)
5000
To set the password, select Enabled, then Set Password. Use the UP and DOWN keys to increment and decrement the value, from 0 to 9999. When the desired value has been selected, press ENTER, then
ESC to exit the sub-menu.
12.2 Menu Key Password
If enabled, a password must be provided every time the MENU key is selected. Select Menu Key from the sub-menu. Follow the same procedure as described in the previous sub-section.
12.3 Locking Out Other Keys
Other keys may be locked out individually. Select any combination of keys (MODE, ZERO, DATA) by pressing ENTER in the Passwords sub-menu. Pressing a locked key will prompt the message “KEY PROTECTED” and then revert to the previous screen.
12.4 Password Prompts
If passwords have been enabled, the following will be displayed when pressing the MENU key or accessing the Calibration section:
ENTER PASSWORD
(0000 – 9999)
5000
Use the UP and DOWN keys to select the correct password, then press ENTER to continue. If the incorrect password has been entered, the display will appear as follows:
INCORRECT PASSWORD
Reset password
Request code:
XXXX
Press ENTER or ESC
To re-enter the password, press ESC to exit to the home screen. Then, access the desired function and enter the password again when prompted.
21
Page 23
Series TT01 Digital Cap Torque Testers User’s Guide
If the password has been misplaced, it can be reset. Press ENTER to generate a request code. The
request code must be supplied to Mark-10 or a distributor, who will then provide a corresponding activation code. Enter the activation code to disable the password.
13 OTHER SETTINGS
13.1 Automatic Shutoff
The tester may be configured to automatically power off following a period of inactivity while on battery power. Inactivity is defined as the absence of any key presses or load changes of 100 counts or less. To access these settings, select Automatic Shutoff from the menu. The display will appear as follows:
AUTOMATIC SHUTOFF
* Disabled Enabled Set Minutes 5
Select Disabled to disable automatic shutoff. Select Enabled to enable it. The length of time of inactivity is programmed in minutes via the Set Minutes parameter. Available settings: 5-30, in 5 minute increments.
Note: If the AC adapter is plugged in, the tester will ignore the Automatic Shutoff setting and remain powered on until the POWER key is pressed.
13.2 Backlight
Several initial settings are available upon powering on the tester. To access these settings, select Backlight from the menu. The display will appear as follows:
BACKLIGHT Off On * Auto Set Minutes 1
Select Off for the backlight to be off upon powering on the tester. Select On for the backlight to be on upon powering on the tester. Select Auto for the backlight to be on upon powering on the tester, but will shut off after a period of
inactivity (as defined in the Automatic Shutoff sub-section). The backlight will turn on again when activity resumes. The length of time of inactivity is programmed in minutes via the Set Minutes parameter. Available settings: 1-10, in 1 minute increments.
Note: If the AC adapter is plugged in, the tester will ignore these settings and keep the backlight on. Selecting the On or Off setting in the Backlight menu will manually turn the backlight on or off.
22
Page 24
Series TT01 Digital Cap Torque Testers User’s Guide
13.3 LCD Contrast
The contrast of the display may be adjusted. Select LCD Contrast from the menu. The screen will appear as follows:
LCD CONTRAST
Set Contrast
10
Press ENTER to modify the contrast. Select a value from 0 to 25, with 25 producing the most contrast.
13.4 Beeps
Audible tones can be enabled for all key presses and alerts, such as overload, set point value reached, etc. The Set Point alert can be configured to be either a momentary tone or a continuous tone (until the load is restored to a value between the set points). To configure the functions for which audibl e tones will apply, select Beeps from the menu. The screen will appear as follows:
BEEPS
Keys * Alerts Set Points * Momentary Continuous
13.5 Initial Mode The default measuring mode at power-on may be configured. To access this parameter, select Initial Mode from the menu. The screen will display the available modes. An example is as follows:
INITIAL MODE
* Real Time Peak Clockwise Peak Counter-clockwise
The default value is Real Time.
13.6 Information / Welcome Screen
The following screen is displayed at power-up and can be accessed at any time by selecting Information from the menu:
Digital Torque Tester
Series TT01 Model No: MTT01-50 Serial No: 1234567 Version: 1.0 (c) Mark-10 Corp.
23
Page 25
Series TT01 Digital Cap Torque Testers User’s Guide
14 SPECIFICATIONS
14.1 General
Accuracy: Sampling rate: Power:
Battery life: Measurement units:
Outputs:
Configurable settings: Safe overload: Weight:
Included accessories:
Warranty:
14.2 Dimensions
IN [MM]
7.8 [199.3]
±0.3% of full scale 7,000 Hz AC or rechargeable battery. Low battery indicator appears when battery level is low, and
tester powers off automatically when power reaches critical stage.
Backlight on: Backlight off:
up to 7 hours of continuous use
up to 24 hours of continuous use
ozFin, lbFin, lbFft, Nm, Ncm
USB / RS-232: Mitutoyo (Digimatic): Analog: General purpose: Set points:
Configurable up to 115,200 baud.
Serial BCD suitable for all Mitutoyo SPC-compatible devices.
±1 VCD, ±0.25% of full scale at capacity.
Three open drain outputs, one input.
Three open drain lines.
Digital filters, outputs, set points, automatic shutoff, default units, default mode, passwords, key tones, audio alarms, backlight, calibration
150% of full scale (display shows “OVER” at 110% and above)
8.4 lb [3.8 kg] Set of four posts, AC adapter, battery, USB cable, resource CD (USB driver, MESUR
TM
Lite software, MESURTMgauge DEMO software, and user’s guide), NIST-traceable certificate of calibration
3 years (see individual statement for further details)
WITH OPTIONAL
FLAT JAWS:
5.9 [149.9] MAX
1.3 [33.0] MIN
4.4 [111.8] MAX 0 MIN
WITH OPTIONAL
ADJUSTABLE JAWS:
Ø4.6 [116.8] MAX Ø0 MIN
10.7 [270.5]
Ø7.5 [189.4] MAX Ø0.5 [12.7] MIN
4.8 [121.8]
1.3 [33.0]
Ø6.0 [151.9] MAX Ø1.3 [33.0] MIN
Ø4.4 [111.8] MAX Ø0.25 [6.35] MIN
24
Page 26
Series TT01 Digital Cap Torque Testers User’s Guide
14.3 Factory Settings Parameter Setting
Set points Upper Disabled (defaults to 80% of full scale, clockwise, when enabled) Lower Disabled (defaults to 40% of full scale, clockwise, when enabled) Filters Current 8 Displayed 1024 DATA Key Functions RS-232/USB Output Enabled Mitutoyo Output Disabled Memory Storage Enabled Backlight Auto Minutes 1 Serial/USB RS-232 Output Selected Disabled USB Output Selected Enabled Baud Rate 9,600 Data Format Numeric + units Mitutoyo BCD Output Disabled Break Detection Automatic Output RS-232/USB Output Disabled Mitutoyo Output Disabled Automatic Storage Disabled Automatic Zero Disabled Break Detection Settings Trigger Threshold 10% Auto Zero Delay 5 sec. Automatic Shutoff Enabled Minutes 5 Beeps Keys Enabled Alerts Enabled Set Points Momentary LCD Contrast 10 Initial Mode Real Time Units lbFin Passwords All passwords disabled
14.4 Capacity x Resolution
Model No. ozFin lbFin lbFft Ncm Nm MTT01-12 MTT01-50
MTT01-100
192 x 0.1 12 x 0.005 1 x 0.0005 135 x 0.1 1.35 x 0.001 320 x 0.2 50 x 0.02 4 x 0.002 570 x 0.5 5.7 x 0.005 800 x 0.5 100 x 0.05 8 x 0.005 1150 x 0.5 11.5 x 0.005
25
Page 27
Series TT01 Digital Cap Torque Testers User’s Guide
NOTES:
26
Page 28
Series TT01 Digital Cap Torque Testers User’s Guide
Mark-10 Corporation has been an innovator in the force and torque measurement fields
since 1979. We strive to achieve 100% customer satisfaction through excellence in product
design, manufacturing and customer support. In addition to our standard line of products we
can provide modifications and custom designs for OEM applications. Our engineering team
is eager to satisfy any special requirements. Please contact us for further information or
suggestions for improvement.