Mark-10 MR04-05M, MR04-025M User manual

SMART FORCE AND TORQUE SENSORS
User’s Guide
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
1
Thank you…
Thank you for purchasing a Mark-10 Plug & Test® remote sensor, designed for use with Mark-10 models 7i, 5i and 3i force/torque indicators.
With proper usage, we are confident that you will get many years of great service with this product. Mark­10 sensors are ruggedly built for many years of service in laboratory and industrial environments.
This User’s Guide provides setup, safety, and operation
instructions for each individual sensor series. Instructions for using models 5i and 3i indicators are available in their respective user’s guides. For additional information or answers to your questions, please do not hesitate to contact us. Our technical support and engineering teams are eager to assist you.
Before use, each person who is to use Mark-10 sensors and indicators should be fully trained in appropriate operation and safety procedures.
TABLE OF CONTENTS
OVERVIEW ..................................................... 2
SAFETY / PROPER USAGE ..............................3
SETUP ................................................................4
SERIES R01 FORCE SENSORS .......................5
SERIES R02 FORCE SENSORS .......................6
SERIES R03 FORCE SENSORS .......................7
SERIES R04 FORCE SENSORS .......................8
SERIES R05 FORCE SENSORS .......................9
SERIES R06 FORCE SENSORS .................... 10
SERIES R50 TORQUE SENSORS ................. 11
SERIES R51 TORQUE SENSORS ................. 13
SERIES R52 TORQUE SENSORS ................. 15
SERIES R53 TORQUE SENSORS ................. 16
SERIES R55 TORQUE SENSORS ................. 17
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
2
1 OVERVIEW
1.1 General Overview
Plug & Test® sensors accommodate numerous force and torque measurement requirements, from 0.25 to 10,000 lbF (1 N to 50 kN) of force, and from 10 ozFin to 5,000 lbFin (7 Ncm to 550 Nm) of torque. These sensors can be handheld or mounted to a fixture or test stand for more sophisticated testing requirements.
Plug & Test® sensors are used with Mark-10 7i, 5i, and 3i indicators. They may be disconnected from one indicator and connected to another without the need for re-calibration or re-configuration. All such data is saved within a PCB located inside the smart connector.
The model number, serial number, and capacity of the sensor are identified in the rectangular label located on the connector. The model and serial numbers are also identified in the Information screen of the indicator. Refer to the indicator user’s guides for more information.
1.2 Accuracy and Resolution
Indicator accuracy must be combined with sensor accuracy to determine the total accuracy of the system. Since sensors may be used with the 7i, 5i or 3i indicators, the accuracy of the indicator being used must be identified and taken into account, as follows:
Indicator Model
Accuracy
7i / 5i
±0.1% of full scale
3i
±0.2% of full scale
The total system accuracy can be calculated by adding the sensor accuracy and indicator accuracy. Refer to the following examples:
Example 1
Model MR01-100 sensor with Model 5i Indicator
MR01-100
±0.15% of full scale
+
5i
±0.1% of full scale
=
Total
±0.25% of full scale
This translates into a fixed error of up to:
0.25% x 100 lbF = 0.25 lbF Example 2
Model MR50-50Z sensor with Model 3i Indicator
MR50-50Z
±0.35% of full scale
+
3i
±0.2% of full scale
=
Total
±0.55% of full scale
This translates into a fixed error of up to:
0.55% x 50 ozFin = 0.275 ozFin
Because accuracy is defined as a percentage of full scale, the fixed error is possible anywhere on the scale from 0 to the capacity. As such, this value represents an increasingly large error as percentage of reading towards the low end of the scale. It is, therefore, recommended that a sensor is selected with capacity as close as possible to the expected load.
The resolution may be different for some sensors depending on whether a 7i, 5i, or 3i indicator is being
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
3
used. For example, a Series R01 force sensor will display finer resolution when connected to a 5i indicator than when connected to a 3i indicator. Resolution information is shown in the following sections.
2 SAFETY / PROPER USAGE
Read through the following safety instructions thoroughly before using a sensor:
1. Note the sensor’s capacity before use and ensure that the capacity is not exceeded. Producing a load greater than the indicated safe overload value can damage the sensor. An overload can occur whether the sensor’s indicator is powered on or off.
2. In order to extend the life of the sensor, avoid repetitive shock and impact loading.
3. When moving the sensor to another location, never lift from the cable or strain relief. This can cause damage to the sensor. Always lift the sensor housing itself.
4. Always ensure that load is applied axially with respect to the sensor.
5. Ensure that the sensor is kept away from water or any other electrically conductive liquids at all times.
6. The sensor and indicator should be serviced by a trained technician only. AC power must be disconnected and the indicator must be powered off before the housing is opened.
7. Always consider the characteristics of the sample being tested before initiating a test. A risk assessment should be carried out beforehand to ensure that all safety measures have been addressed and implemented.
8. Typical materials able to be tested include many manufactured items, such as springs, electronic components, fasteners, caps, films, mechanical assemblies, and many others. Items that should not be used with the sensor include potentially flammable substances or products, items that can shatter in an unsafe manner, and any other components that can present an exceedingly hazardous situation when acted upon by a force. Always wear eye and face protection when testing, especially in aforementioned hazardous cases. Extra bodily protection should be worn if a destructive failure of a test sample is possible.
9. In aforementioned hazardous situations, it is strongly recommended that a machine guarding system be employed to protect the operator and others in the vicinity from shards or debris.
10. Sensors have threaded holes or chucks, designed for the mounting of grips, fixtures, or attachments. If any such accessories are used, ensure they are mounted firmly to prevent a potential safety risk to the operator and others in the vicinity. If using an accessory from a supplier other than Mark-10, ensure that it is constructed of suitably rugged materials and components. Similar precautions should be taken when mounting the sensor to a test stand, work bench, or other piece of equipment.
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
4
3 SETUP
The Plug & Test® connector must be inserted into the receptacle of the indicator with the side marked
“Plug & Test
®
Technology” facing up (see Fig. 3.1). When fully inserted, the connector will lock into place
with a “click”.
Fig. 3.1
Appropriate orientation of the Plug & Test® connector.
To release the connector, press both buttons on either side of the indicator housing (see Fig. 3.2). Pull the connector completely out of the indicator by holding the curved aluminum section. DO NOT pull on the cable or strain relief.
Fig. 3.2
Press both buttons on either side of the indicator housing to release the Plug & Test® connector.
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
5
4 SERIES R01 FORCE SENSORS
4.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
4.2 Overview
Tension and compression force may be applied to the surfaces with threaded holes. Attachments may be threaded into these holes. These holes also allow for mounting as required.
4.3 Specifications Accuracy: ±0.15% of full scale Safe overload: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 1.9 to 3.0 lb [0.9 to 1.4 kg], depending on model
4.4 Dimensions (in [mm])
4.5 Capacity x Resolution
Model
No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
gF
kgF N kN
lbF
kgF N kN
MR01-50
50 x 0.02
800 x 0.5
25000 x 10
25 x 0.01
250 x 0.1
-
50 x 0.05
25 x 0.02
250 x 0.2
-
MR01-100
100 x 0.05
1600 x 1
50000 x 20
50 x 0.02
500 x 0.2
-
100 x 0.1
50 x 0.05
500 x 0.5
-
MR01-200
200 x 0.1
3200 x 2
-
100 x 0.05
1000 x 0.5
1 x 0.0005
200 x 0.2
100 x 0.1
1000 x 1
-
MR01-500
500 x 0.2
8000 x 5
-
250 x 0.1
2500 x 1
2.5 x 0.001
500 x 0.5
250 x 0.2
2500 x 2
-
MR01-300
300 x 0.2
4800 x 5
-
150 x 0.1
1500 x 1
1.5 x 0.001
300 x 0.5
150 x 0.2
1500 x 2
-
MR01-750
750 x 0.5
12000 x 10
-
375 x 0.2
3750 x 2
3.75 x 0.002
750 x 0.5
375 x 0.2
3750 x 2
-
MR01-1000
1000 x 0.5
16000 x 10
-
500 x 0.2
5000 x 2
5 x 0.002
1000 x 1
500 x 0.5
5000 x 5
-
MR01-1500
1500 x 1
24000 x 20
-
750 x 0.5
7500 x 5
7.5 x 0.005
1500 x 2
750 x 1
7500 x 10
-
MR01-2000
2000 x 1
32000 x 20
-
1000 x 0.5
10000 x 5
10 x 0.005
2000 x 2
1000 x 1
10000 x 10
-
MR01-5000
5000 x 2 - -
2500 x 1
25000 x 10
25 x 0.01
5000 x 5
2500 x 2
-
25 x 0.02
MR01-10000
10000 x 5 - -
5000 x 2
50000 x 25
50 x 0.02
10000 x 10
5000 x 5
-
50 x 0.05
Model No.
A B C D E
F
MR01-50
2.40
[61.0]
2.00
[50.8]
0.46
[11.7]
0.90
[22.9]
1/4-28
UNF
MR01-XXXXX:
20 ft [6 m]
MR01-XXXX-1: 10 in [254 mm]
MR01-100
MR01-200
MR01-300
0.71
[18.0]
1.10
[27.9]
1/2-20
UNF
MR01-500
MR01-750
MR01-1000
MR01-1500
0.96
[24.4]
1.40
[35.6]
MR01-2000
MR01-5000
3.90
[99.1]
3.00
[76.2]
3/4-16
UNF
MR01-10000
1.21 [30.7]
1.63 [41.4]
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
6
5 SERIES R02 FORCE SENSORS
5.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
5.2 Overview
Compression force may be applied to the button in the center of the top surface of the sensor (visible in the picture above). DO NOT apply load to the cover on the underside of the sensor. Threaded holes are supplied to permit mounting to various surfaces.
5.3 Specifications
Accuracy: ±0.5% of full scale Safe overload: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: MR02-100 - MR02-2000: 0.3 lb [0.1 kg]
MR02-5000 - MR02-10000: 0.5 lb [0.2 kg]
5.4 Dimensions (in[mm])
Model
ØA
ØB C D
ØE F ØG
MR02-100 – MR02-2000
1.25
[31.8]
0.32 [8.1]
0.05 [1.3]
0.39 [9.9]
1.0
[25.4]
-
0.83
[21.1]
MR02-5000 – MR02-10000
1.50
[38.1]
0.43
[10.9]
0.07 [1.8]
0.62
[15.8]
1.25
[31.8]
0.25 [6.4]
1.08
[27.4]
5.5 Capacity x Resolution
Model
No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
gF
kgF N kN
lbF
kgF N kN
MR02-100
100 x 0.05
1600 x 1
50000 x 20
50 x 0.02
500 x 0.2
-
100 x 0.1
50 x 0.05
500 x 0.5
-
MR02-200
200 x 0.1
3200 x 2
-
100 x 0.05
1000 x 0.5
1 x 0.0005
200 x 0.2
100 x 0.1
1000 x 1
-
MR02-500
500 x 0.2
8000 x 5
-
250 x 0.1
2500 x 1
2.5 x 0.001
500 x 0.5
250 x 0.2
2500 x 2
-
MR02-1000
1000 x 0.5
16000 x 10
-
500 x 0.2
5000 x 2
5 x 0.002
1000 x 1
500 x 0.5
5000 x 5
-
MR02-2000
2000 x 1
32000 x 20
-
1000 x 0.5
10000 x 5
10 x 0.005
2000 x 2
1000 x 1
10000 x 10
-
MR02-5000
5000 x 2 - -
2500 x 1
25000 x 10
25 x 0.01
5000 x 5
2500 x 2
-
25 x 0.02
MR02-10000
10000 x 5 - -
5000 x 2
50000 x 25
50 x 0.02
10000 x 10
5000 x 5
-
50 x 0.05
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
7
6 SERIES R03 FORCE SENSORS
6.1 Unpacking and Assembly
Carefully remove the sensor from the box. Extra care should be taken for very low capacity models. No assembly is required.
6.2 Overview
Tension and compression force may be applied to the threaded holes in the load cell shaft and opposite flat surface. Attachments may be threaded into these holes. Finger-tighten only. These holes also allow for mounting as required.
6.3 Specifications
Accuracy: ±0.15% of full scale Safe overload: MR03-025 – MR03-2: 200% of full scale
MR03-5 – MR03-100: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 0.4 lb [0.2 kg]
6.4 Dimensions (in[mm])
6.5 Capacity x Resolution
Model
No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
gF
kgF N mN
lbF
gF
kgF
N
MR03-025
0.25 x 0.0001
4 x 0.002
100 x 0.05
-
1 x 0.0005
1000 x 0.5
0.25 x 0.0002
100 x 0.1
-
1 x 0.001
MR03-05
0.5 x 0.0002
8 x 0.005
250 x 0.1
-
2.5 x 0.001
2500 x 1
0.5 x 0.0005
250 x 0.2
-
2.5 x 0.002
MR03-2
2 x 0.001
32 x 0.02
1000 x 0.5
1 x 0.0005
10 x 0.005
-
2 x 0.002
-
1 x 0.001
10 x 0.01
MR03-5
5 x 0.002
80 x 0.05
2500 x 1
2.5 x 0.001
25 x 0.01
-
5 x 0.005
-
2.5 x 0.002
25 x 0.02
MR03-10
10 x 0.005
160 x 0.1
5000 x 2
5 x 0.002
50 x 0.02
-
10 x 0.01
-
5 x 0.005
50 x 0.05
MR03-20
20 x 0.01
320 x 0.2
10000 x 5
10 x 0.005
100 x 0.05
-
20 x 0.02
-
10 x 0.01
100 x 0.1
MR03-50
50 x 0.02
800 x 0.5
25000 x 10
25 x 0.01
250 x 0.1
-
50 x 0.05
-
25 x 0.02
250 x 0.2
MR03-100
100 x 0.05
1600 x 1
50000 x 20
50 x 0.02
500 x 0.2
-
100 x 0.1
-
50 x 0.05
500 x 0.5
Model No.
A
MR03-XXX
24 in [610 mm], coiled, when retracted 10 ft [3 m], coiled, when stretched
MR03-XXX-1
10 in [254 mm], straight
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
8
7 SERIES R04 FORCE SENSORS
7.1 Unpacking and Assembly
Carefully remove the sensor from the box. Extra care should be taken for very low capacity models. No assembly is required.
7.2 Overview
Tension and compression force may be applied to the surfaces with threaded holes. Attachments may be threaded into these holes.
Finger-tighten only. These holes also allow for mounting as required.
7.3 Specifications
Accuracy: ±0.2% of full scale Safe overload: 200% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 0.015 lb [0.007 kg]
7.4 Dimensions (in[mm])
7.5 Capacity x Resolution
Model
No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
gF
kgF N mN
lbF
gF
kgF
N
MR04-025
0.25 x 0.0001
4 x 0.002
100 x 0.05
-
1 x 0.0005
1000 x 0.5
0.25 x 0.0002
100 x 0.1
-
1 x 0.001
MR04-05
0.5 x 0.0002
8 x 0.005
250 x 0.1
-
2.5 x 0.001
2500 x 1
0.5 x 0.0005
250 x 0.2
-
2.5 x 0.002
MR04-2
2 x 0.001
32 x 0.02
1000 x 0.5
1 x 0.0005
10 x 0.005
-
2 x 0.002
-
1 x 0.001
10 x 0.01
MR04-5
5 x 0.002
80 x 0.05
2500 x 1
2.5 x 0.001
25 x 0.01
-
5 x 0.005
-
2.5 x 0.002
25 x 0.02
MR04-10
10 x 0.005
160 x 0.1
5000 x 2
5 x 0.002
50 x 0.02
-
10 x 0.01
-
5 x 0.005
50 x 0.05
MR04-20
20 x 0.01
320 x 0.2
10000 x 5
10 x 0.005
100 x 0.05
-
20 x 0.02
-
10 x 0.01
100 x 0.1
MR04-50
50 x 0.02
800 x 0.5
25000 x 10
25 x 0.01
250 x 0.1
-
50 x 0.05
-
25 x 0.02
250 x 0.2
MR04 -100
100 x 0.05
1600 x 1
50000 x 20
50 x 0.02
500 x 0.2
-
100 x 0.1
-
50 x 0.05
500 x 0.5
OK
NOT OK
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
9
8 SERIES R05 FORCE SENSORS
8.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
8.2 Overview
Designed for pull and push testing. Firmly grip the handles when producing tension or compression force. Force may be applied to the surface with threaded hole. Attachments may be threaded into
these holes. Finger-tighten only.
8.3 Specifications
Accuracy: ±0.15% of full scale Safe overload: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 2 lb [0.9 kg]
8.4 Dimensions (in[mm])
8.5 Capacity x Resolution
Model
No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
kgF N kN
lbF
kgF
N
MR05-500
500 x 0.2
8000 x 5
250 x 0.1
2500 x 1
2.5 x 0.001
500 x 0.5
250 x 0.2
2500 x 2
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
10
9 SERIES R06 FORCE SENSORS
9.1 Overview
The MR06-200 is designed for field pull testing of crimped wire terminals. Carefully remove the sensor from the box. No assembly is required.
9.2 Instructions
Rotate the upper grip to an appropriately sized slot. Depress the cam shaft button, located in the tensioning mechanism, and insert the end of the wire sample in the hole located in the cam shaft. Release the cam shaft button. Wrap the wire at least one full clockwise turn around the cam shaft. Then insert the terminal lead in the slot on the upper grip.
Begin to compress and release the handles slowly. Tension force values will be displayed on the indicator. After the test is complete, remove any debris from the upper grip assembly, and depress the button on the cam shaft to remove the wire.
9.3 Specifications
Accuracy: ±0.5% of full scale Safe overload: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 2.5 lb [1.2 kg]
9.4 Dimensions (in[mm])
9.5 Capacity x Resolution
Model No.
With Model 7i / 5i Indicator
With Model 3i Indicator
lbF
ozF
kgF
N
kN
lbF
kgF
N
MR06-200
200 x 0.1
3200 x 2
100 x 0.05
1000 x 0.5
1 x 0.0005
200 x 0.2
100 x 0.1
1000 x 1
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
11
10 SERIES R50 TORQUE SENSORS
10.1 Unpacking and Assembly
Carefully remove the sensor from the box. For models MR50-10Z, MR50-20Z, and MR50-50Z, remove the protective tubing inserted around the chuck. Save it for future transportation needs. No assembly is required.
10.2 Overview
Designed for clockwise and counter-clockwise torque testing. The sensor may be handheld or mounted to a test stand, fixture, or other equipment. Bits or fixtures may be placed in the chuck, although extra care should be taken when handling low capacity models.
10.3 Specifications
Accuracy: ±0.35% of full scale Safe overload: MR50-10Z - MR50-50Z: 300% of full scale
MR50-12 - MR50-100: 150% of full scale
Chuck opening range: MR50-10Z - MR50-50Z: 0.062 - 0.375 in [1.6 - 9.5 mm]
MR50-12 - MR50-100: 0.078 - 0.5 in [2.0 - 12.7 mm] Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: From 1.4 lb [0.6 kg]
10.4 Calibration Calibration should be performed in a vertical orientation, especially for sensors with capacities of 50 ozFin [35 Ncm] or less. Horizontal orientation subjects the sensor to side loads resulting from the weight of the chuck and attachments. Such side loads can be significant enough to skew the readings out of tolerance.
The illustration below depicts a recommended vertical setup:
For further calibration instructions, refer to the indicator’s user’s guide.
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
12
10.5 Dimensions (in[mm])
10.6 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator
With Model 3i indicator
ozFin
lbFin
lbFft
gFcm
kgFmm
Nmm
Ncm
Nm
ozFin
lbFin
kgFmm
Ncm
MR50-10Z
10 x 0.005 - - 700 x 0.5
7 x 0.005
70 x 0.05
7 x 0.005 - 10 x 0.01 - 7 x 0.005
7 x 0.005
MR50-20Z
20 x 0.01 - - 1400 x 1 14 x 0.01
140 x 0.1
14 x 0.01 - 20 x 0.02 - 14 x 0.01
14 x 0.01
MR50-50Z
50 x 0.02 - - 3600 x 2 36 x 0.02
350 x 0.2
35 x 0.02 - 50 x 0.05 - 36 x 0.05
35 x 0.05
MR50-12 - 12 x 0.005
1 x 0.0005 - 140 x 0.1 - 135 x 0.1
1.35 x 0.001 - 12 x 0.01
140 x 0.1
135 x 0.1
MR50-50 - 50 x 0.02
4 x 0.002 - 580 x 0.5 - 570 x 0.5
5.7 x 0.005 - 50 x 0.05
580 x 0.5
570 x 0.5
MR50-100 - 100 x 0.05
8 x 0.005 - 1150 x 0.5 - 1150 x 0.5
11.5 x 0.005 - 100 x 0.1
1150 x 1
1150 x 1
Model No.
A
MR50-10Z - MR50-50Z
4.82 [122.4]
MR50-12 - MR50-100
5.19 [131.8]
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
13
11 SERIES R51 TORQUE SENSORS
11.1 Unpacking and Assembly
Carefully remove the sensor from the box. Series R51 sensors are available with three interchangeable chuck attachments and bit holder. To attach the chuck or bit holder, align the pin located on the end of the sensor body with the hole located on the attachment (see Fig. 10.1). Thread the ring onto the sensor body to keep the attachment in place.
11.2 Overview
Designed for clockwise and counter-clockwise torque testing. The sensor may be handheld or mounted to a test stand, fixture, or other equipment.
Bits or fixtures may be placed in the chuck, although extra care should be taken when handling low capacity models.
11.3 Specifications
Accuracy: ±0.2% of full scale Safe overload:
MR50-10Z - MR50-50Z: 300% of full scale
MR50-12 - MR50-100: 150% of full scale
Chuck opening range:
MR50-10Z - MR50-50Z: 0.062 - 0.375 in [1.6 - 9.5
mm]
MR50-12 - MR50-100: 0.078 - 0.5 in [2.0 - 12.7 mm]
Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing)
Weight: From 1.4 lb [0.6 kg]
11.4 Calibration Calibration should be performed in a vertical orientation, especially for sensors with capacities of 50 ozFin [35 Ncm] or less. Horizontal orientation subjects the sensor to side loads resulting from the weight of the chuck and attachments. Such side loads can be significant enough to skew the readings out of tolerance.
The illustration below depicts a recommended vertical setup:
Fig. 10.1
The attachment and main sensor body mate with a pin and threaded ring.
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
14
For further calibration instructions, refer to the indicator’s user’s guide.
11.5 Dimensions (in[mm])
11.6 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator
With Model 3i indicator
ozFin
lbFin
lbFft
gFcm
kgFmm
Nmm
Ncm
Nm
ozFin
lbFin
kgFmm
Ncm
MR51-10Z
10 x 0.005 - - 700 x 0.5
7 x 0.005
70 x 0.05
7 x 0.005 - 10 x 0.01 - 7 x 0.005
7 x 0.005
MR51-20Z
20 x 0.01 - - 1400 x 1 14 x 0.01
140 x 0.1
14 x 0.01 - 20 x 0.02 - 14 x 0.01
14 x 0.01
MR51-50Z
50 x 0.02 - - 3600 x 2 36 x 0.02
350 x 0.2
35 x 0.02 - 50 x 0.05 - 36 x 0.05
35 x 0.05
MR51-12 - 12 x 0.005
1 x 0.0005 - 140 x 0.1 - 135 x 0.1
1.35 x 0.001 - 12 x 0.01
140 x 0.1
135 x 0.1
MR51-50 - 50 x 0.02
4 x 0.002 - 580 x 0.5 - 570 x 0.5
5.7 x 0.005 - 50 x 0.05
580 x 0.5
570 x 0.5
MR51-100 - 100 x 0.05
8 x 0.005 - 1150 x 0.5 - 1150 x 0.5
11.5 x 0.005 - 100 x 0.1
1150 x 1
1150 x 1
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
15
12 SERIES R52 TORQUE SENSORS
12.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
12.2 Overview
Designed for clockwise and counter-clockwise torque testing. The sensor contains threaded holes on each side to permit mounting to a larger testing system.
12.3 Specifications
Accuracy: ±0.35% of full scale Safe overload: MR52-10Z – MR52-50Z: 300% of full scale
MR52-12 – MR52-100: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: 1.2 lb [0.5 kg]
12.4 Dimensions (in[mm])
12.5 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator
With Model 3i indicator
ozFin
lbFin
lbFft
gFcm
kgFmm
Nmm
Ncm
Nm
ozFin
lbFin
kgFmm
Ncm
MR52-10Z
10 x 0.005 - - 700 x 0.5
7 x 0.005
70 x 0.05
7 x 0.005 - 10 x 0.01 - 7 x 0.005
7 x 0.005
MR52-20Z
20 x 0.01 - - 1400 x 1
14 x 0.01
140 x 0.1
14 x 0.01 - 20 x 0.02 - 14 x 0.01
14 x 0.01
MR52-50Z
50 x 0.02 - - 3600 x 2
36 x 0.02
350 x 0.2
35 x 0.02 - 50 x 0.05 - 36 x 0.05
35 x 0.05
MR52-12 - 12 x 0.005
1 x 0.0005 - 140 x 0.1 - 135 x 0.1
1.35 x 0.001 - 12 x 0.01
140 x 0.1
135 x 0.1
MR52-50 - 50 x 0.02
4 x 0.002 - 580 x 0.5 - 570 x 0.5
5.7 x 0.005 - 50 x 0.05
580 x 0.5
570 x 0.5
MR52-100 - 100 x 0.05
8 x 0.005 - 1150 x 0.5 - 1150 x 0.5
11.5 x 0.005 - 100 x 0.1
1150 x 1
1150 x 1
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
16
13 SERIES R53 TORQUE SENSORS
13.1 Unpacking and Assembly
Carefully remove the sensor from the box. Install the four posts in the desired positions on the sliders. If alternative gripping fixtures were purchased, install them using the supplied hardware.
13.2 Overview
Designed for clockwise and counter-clockwise bottle cap torque testing.
Turning the knob moves the sliders along the length of the screw to engage to disengage the sample. The posts or alternative gripping fixtures secure the sample while the cap is twisted by hand. The sensor contains threaded holes on the underside of the base for bench mounting.
13.3 Specifications
Accuracy: ±0.5% of full scale Safe overload: MR53-10Z - MR53-50Z: 300% of full scale / MR53-12 - MR53-100: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: MR53-10Z - MR53-50Z: 2.2 lb [1.0 kg] / MR53-12 - MR53-100: 5.4 lb [2.5 kg]
13.4 Dimensions (in[mm])
MR53-10Z - MR53-50Z: MR53-12 - MR53-100:
13.5 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator
With Model 3i indicator
ozFin
lbFin
lbFft
gFcm
kgFmm
Nmm
Ncm
Nm
ozFin
lbFin
kgFmm
Ncm
MR53-10Z
10 x 0.01 - - 700 x 0.5
7 x 0.005
70 x 0.05
7 x 0.005 - 10 x 0.01 - 7 x 0.005
7 x 0.005
MR53-20Z
20 x 0.02 - - 1400 x 1
14 x 0.01
140 x 0.1
14 x 0.01 - 20 x 0.02 - 14 x 0.01
14 x 0.01
MR53-50Z
50 x 0.05 - - 3600 x 5
36 x 0.05
350 x 0.5
35 x 0.05 - 50 x 0.05 - 36 x 0.05
35 x 0.05
MR53-12 - 12 x 0.01
1 x 0.001 - 140 x 0.1 - 135 x 0.1
1.35 x 0.001 - 12 x 0.01
140 x 0.1
135 x 0.1
MR53-50 - 50 x 0.05
4 x 0.005 - 580 x 0.5 - 570 x 0.5
5.7 x 0.005 - 50 x 0.05
580 x 0.5
570 x 0.5
MR53-100 - 100 x 0.1
8 x 0.01 - 1150 x 1 - 1150 x 1
11.5 x 0.01 - 100 x 0.1
1150 x 1
1150 x 1
www.GlobalTestSupply.com
Plug & Test® Smart Force & Torque Sensors User’s Guide
17
14 SERIES R55 TORQUE SENSORS
14.1 Unpacking and Assembly
Carefully remove the sensor from the box. No assembly is required.
14.2 Overview
Designed for clockwise and counter-clockwise torque testing. Square drives on both ends permit inline use with a torque wrench, or for use in OEM and other applications.
14.3 Specifications
Accuracy: ±0.6% of full scale Safe overload: 150% of full scale Operating temperature: 40ºF – 100ºF [5ºC – 38ºC] Operating humidity: 96% max. (non-condensing) Weight: MR55-20 – MR55-1000: 1.3 lb (0.6 kg)
MR55-5000: 1.9 lb (0.9 kg)
14.4 Dimensions (in[mm])
14.5 Capacity x Resolution
Model
No.
With Model 7i / 5i indicator
With Model 3i indicator
lbFin
lbFft
kgFmm
Ncm
Nm
lbFin
kgFmm
Ncm
MR55-20
20 x 0.02
1.5 x 0.002
230 x 0.2
220 x 0.2
2 x 0.002
20 x 0.02
230 x 0.2
220 x 0.2
MR55-50
50 x 0.05
4 x 0.005
580 x 0.5
570 x 0.5
5.7 x 0.005
50 x 0.05
580 x 0.5
570 x 0.5
MR55-100
100 x 0.1
8 x 0.01
1150 x 1
1150 x 1
11.5 x 0.01
100 x 0.1
1150 x 1
1150 x 1
MR55-200
200 x 0.2
16 x 0.02
2300 x 2
2200 x 2
22 x 0.02
200 x 0.2
2300 x 2
2200 x 2
MR55-400
400 x 0.5
32 x 0.05
4600 x 5
4500 x 5
45 x 0.05
400 x 0.5
4600 x 5
4500 x 5
MR55-1000
1000 x 1
80 x 0.1
11500 x 10
11000 x 10
110 x 0.1
1000 x 1
11500 x 10
11000 x 10
MR55-5000
5000 x 5
400 x 0.5
55 x 0.05 kgFm
55000 x 50
550 x 0.5
5000 x 5
55 x 0.05 kgFm
550 x 0.5 Nm
Model No.
Drive A B C D
ØE
F
MR55-20
1/4"
0.30 [7.5]
1.73
[44.0]
0.32 [8.0]
0.39
[10.0]
1.77
[45.0]
2.32
[59.0]
MR55-50
MR55-100
MR55-200
3/8"
0.41
[10.5]
1.73
[44.0]
0.43
[11.0]
MR55-400
MR55-1000
1/2"
0.59
[15.0]
0.63
[16.0]
MR55-5000
3/4"
0.89
[22.5]
2.11
[53.5]
0.94
[24.0]
0.77
[19.5]
2.01
[51.0]
2.56
[65.0]
www.GlobalTestSupply.com