Thank you for purchasing a Mark-10 Series 2 digital force gauge, designed
for tension and compression force testing applications from 2 to 500 lbF (10
to 2,500 N) full scale. The Series 2 is an essential component of a force
testing system, typically also comprising a test stand and grips.
With proper usage, we are confident that you will get many years of great
service with this product. Mark-10 force gauges are ruggedly built for many
years of service in laboratory and industrial environments.
This User’s Guide provides setup, safety, and operation instructions.
Dimensions and specifications are also provided. For additional information
or answers to your questions, please do not hesitate to contact us. Our
technical support and engineering teams are eager to assist you.
Before use, each person who is to use the Series 2 force gauge should
be fully trained in appropriate operation and safety procedures.
Certificate of calibration with data
Medium hook, #10-32M (requires G1039)
Coupling, #10-32F/F
Flat head, #10-32F
Cone, #10-32F
Chisel point, #10-32F
V-groove, #10-32F
Extension rod, 5”, #10-32F
Large hook, 5/16-18M (requires G1037)
Coupling, 5/16-18F/F
Flat head, 5/16-18F
Cone, 5/16-18F
Chisel point, 5/16-18F
V-groove, 5/16-18F
Extension rod, 5”, 5/16-18F
Caution!
Note the force gauge’s capacity before use and ensure that the capaci ty is not exceeded.
Producing a force greater than 200% of the gauge’s capacity can damage the internal load cell. An
overload can occur whether the gauge is powered on or off.
Typical materials suitable for testing include many manufactured items, such as springs, electronic
components, fasteners, caps, films, mechanical assemblies, and many others. Items that should not be
used with the gauge include potentially flammable substances or products, items that can shatter in an
unsafe manner, and any other components that can present an exceedingly hazardous situation when
acted upon by a force.
The following safety checks and procedures should be performed before and during o peration:
1. Never operate the gauge if there is any visible damage to the AC adapter or the gauge itself.
2. Ensure that the gauge is kept away from water or any other electrically conductive liquids at all
times.
3. The gauge should be serviced by a trained technician only. AC power must be disconne cted and
the gauge must be powered off before the housing is opened.
4. Always consider the characteristics of the sample being tested before initiating a test. A risk
assessment should be carried out beforehand to ensure that all safety measures have been
addressed and implemented.
5. Wear eye and face protection when testing, especially when testing brittle samples that have the
potential to shatter under force. Be aware of the dangers posed by potential energy that can
accumulate in the sample during testing. Extra bodily protection should be worn if a destructive
failure of a test sample is possible.
6. In certain applications, such as the testing of brittle samples that can shatter, or other applications
that could lead to a hazardous situation, it is strongly recommended that a machine guarding
system be employed to protect the operator and others in the vicinity from shards or debris.
7. When the gauge is not in use, ensure that the power is turned off.
2 POWER
Caution!
Do not use AC adapters other than supplied or instrument damage may occur.
The gauge is powered either by an 8.4V rechargeable battery or by an AC adapter (input jack is located
in the left side of the housing).
If the AC adapter is plugged in, an icon appears in the lower left corner of the display, as follows:
If the AC adapter is not plugged in, battery power drainage is denoted in a five-step process:
1. When battery life is greater than 75%, the following indicator is present:
2. When battery life is between 50% and 75%, the following indicator is present:
3. When battery life is between 25% and 50%, the following indicator is present:
4. When battery life is less than 25%, the following indicator is present:
5. When battery life drops to approximately 2%, the indicator from step 4 will be flashing.
Several minutes after (timing depends on usage and whether the backli ght is turned on or
off), a message appears, “BATTERY VOLTAGE TOO LOW. POWERING OFF”. An audio
tone will sound and the gauge will power off.
The gauge can be configured to automatically power off following a period of inactivity. Refer to the Other Settings section for details.
Battery life is approximately 24 hours with the backlight off, or 7 hours with the backlight on. To change
the battery, loosen the two captive screws on the back side of the housing and separate the two housing
halves. A foam strip separates the battery from the PCB, as shown in the illustration below:
Ensure that the foam is positioned above the battery when reassembling the housing. Exercise care whe n
reassembling the two halves of the housing, ensuring that internal wires do not interfere.
3 MECHANICAL SETUP
3.1 Loading shaft orientation
In order to accommodate a variety of testing requirements, the orientation of the loading shaft may be
oriented in either of the two positions shown below. To change the loading shaft orientation, loosen the
two captive screws on the back side of the housing, separate the two housing halves, rotate one half 180
degrees, and reassemble. Exercise care when reassembling the two halves of the housing, ensuring that
internal wires do not interfere.
Although the gauge may be used by hand, proper mounting is important if attached to a fixture or test
stand. The round steel insert with a hole in the back of the housing is provided to withstand the load
during a test. A mating dowel pin should be used (see illustration below). Mounting plates on Mark-10 test
stands include a dowel pin and clearance holes for the four threaded holes located near the corners of
the housing. An additional two holes are supplied for metric screws. These holes are designed to
accommodate screws in order to hold the gauge in place (Mark-10 test stands include a set of thumb
screws for gauge mounting). The screws must not
dowel pin properly can result in a hazardous situation.
be used for load bearing purposes. Failure to use a
3.3 Mounting attachments to the gauge
The force gauge’s threaded loading shaft is designed to accommodate common grips and attachments
with female mounting holes. To mount a grip, gently thread it onto the shaft. Ensure that the grip or fixture
is positioned to ensure axial load with respect to the loading shaft of the force gauge. When using a grip,
ensure that it secures the sample in such a way that it is prevented from slipping out during a test,
preventing a potential safety risk to the operator and others in the vicinity. If using a grip or fixture from a
supplier other than Mark-10, ensure that it is constructed of suitably rugged materials and components.
Do not use jam nuts or tools to tighten grips or attachments onto the shaft. Finger-tighten only