Mark-10 ESM303, ESM303H User Manual

Models ESM303 & ESM303H
FORCE TEST STANDS
User’s Guide
Models ESM303 & ESM303H Test Stands User’s Guide
Thank you…
Thank you for purchasing a Mark-10 ESM303 / ESM303H Force Test Stand, designed for producing up to 300 lbF (1.5 kN) of tension and compression force. The ESM303 / ESM303H is an essential component of a force testing system, typically also comprising a force gauge, or load cell with indicator, and grips.
With proper usage, we are confident that you will get many years of great service with this product. Mark-10 test stands are ruggedly built for many years of service in laboratory and industrial environments.
This User’s Guide provides setup, safety, and operation instructions. Dimensions and specifications are
also provided. For additional information or answers to your questions, please do not hesitate to contact us. Our technical support and engineering teams are eager to assist you.
Note: Except where otherwise indicated, references to “ESM303” include models ESM303 and ESM303H.
Before use, each person who is to use the test stand should be fully trained in appropriate operation and safety procedures.
TABLE OF CONTENTS
OVERVIEW .......................................................................................... 3
SETUP AND SAFETY ......................................................................... 5
OPERATION BASICS ....................................................................... 11
TEST FUNCTION SETUP ................................................................. 13
OPERATING MODES ........................................................................ 25
COMMUNICATION ............................................................................ 31
FUNCTION ACTIVATION .................................................................. 33
INSTALLING THE SAFETY SHIELD OPTION ................................. 34
INSTALLING THE TRAVEL INDICATION OPTION ......................... 35
INSTALLING A COLUMN EXTENSION OPTION ............................ 37
MAINTENANCE AND SERVICE ....................................................... 40
TROUBLESHOOTING ....................................................................... 41
SPECIFICATIONS ............................................................................. 42
DIMENSIONS .................................................................................... 42
Models ESM303 & ESM303H Test Stands User’s Guide
1 OVERVIEW
1.1 List of included items
Description
Qty.
ESM303
Qty.
ESM303H
1
Force gauge mounting adapter
1
Indicator / load cell mounting adapter
4
Force gauge / indicator mounting screws, #6-32 x 1/2
4
Indicator mounting screws, #6-32 x 1/2
2
Limit switch thumb screws, #6-32 x 5/8
2
Limit switch thumb screws, #6-32 x 5/8
1
Control panel
1
Control panel
1
Control panel mounting bracket
1
Control panel mounting bracket
- - 1
Adapter, control panel bracket to frame (shipped installed)
1
Screw for mounting control panel bracket to column, 1/4-20 x 1”, with lock washers
2
Screws for mounting control panel bracket to adapter, #8-32 x 1” (shipped installed)
- - 2
Screws for mounting adapter to frame, #10-32 x 1” (shipped installed)
2
Screws for mounting control panel to bracket, 1/4-28 x 1-1/4”, with lock washers
2
Screws for mounting control panel to bracket, 1/4-28 x 1-1/4”, with lock washers
1
Accessories kit:
- #10-32F small hook
- #10-32M medium hook
- #10-32F 2” diameter comp. plate
- #10-32 M/F 2” extension rod
- #10-32 F/F coupler
- Adapter, 5/16-18M to #10-32F
- Set screw, #10-32 x 3/4" (3)
- Hex nut, #10-32 (3)
- Allen wrench set
1
Accessories kit:
- #10-32F small hook
- #10-32M medium hook
- #10-32F 2” diameter comp. plate
- #10-32 M/F 2” extension rod
- #10-32 F/F coupler
- Adapter, 1/2-20M to #10-32F
- Adapter, 1/4-28M to 5/16-18F
- Adapter, 5/16-18M to #10-32F (2)
- Set screw, #10-32 x 3/4" (3)
- Hex nut, #10-32 (3)
- Allen wrench set
1
Resource CD
1
Resource CD
1
Power cord
1
Power cord
Additional optional items are available, such as an adjustable base plate, connection cables, and column extensions. Instructions for installing a column extension and retrofitting the test stand with the travel indication option are provided later on in this user’s guide.
Models ESM303 & ESM303H Test Stands User’s Guide
1.2 Physical features
Note the following physical features. The user’s guide will reference this terminology.
1.2.1 ESM303
Base
Base plate
Lower grip
Upper grip
Force gauge
Column
Control panel
Emergency stop
Limit switch (upper)
Test sample
Crosshead
Interface cable
Column cap
Limit switch (lower)
Control panel
mounting bracket
Models ESM303 & ESM303H Test Stands User’s Guide
1.2.2 ESM303H
2 SETUP AND SAFETY
2.1 Mounting
Place the stand on a clean, flat and level work area free from vibration. For the vertically oriented ESM303, It is suggested to mount the stand to a work bench via screws
fastened into the underside of the base. Failure to properly mount the test stand may make it more vulnerable to tipping, especially if a column extension is used, causing a hazardous situation.
2.2 Mounting the control panel
To avoid damage during shipment, the control panel and mounting bracket are packaged separately. Follow these installation instructions:
2.2.1 ESM303
1. Install the mounting bracket to the column utilizing one 1/4-20 screw.
Control panel
mounting bracket
Control panel
Control panel
bracket adapter
Frame cap
Interface cable
Indicator
Crosshead, indicator / load
cell mounting kit
Load cell
Right limit switch
Alternate control panel mounting location
Base plate
Base
Left grip
Test sample
Right grip
Emergency stop
Models ESM303 & ESM303H Test Stands User’s Guide
2. Install the control panel to the mounting bracket utilizing two #8-32 screws.
3. The viewing angle may be adjusted by loosening the lever, positioning the control panel, and re­tightening the lever.
2.2.2 ESM303H
1. Install the control panel to the mounting bracket utilizing two 1/4-28 screws.
2. Install the mounting bracket to the bracket adapter with two #8-32 screws.
3. Install the bracket adapter to the edge of the test frame with two #10-32 screws. This may be done on either the right end or the left end of the frame, where sets of #10-32 holes are provided.
4. The viewing angle may be adjusted by loosening the lever in the rear of the bracket, positioning the control panel, and re-tightening the lever.
2.3 Limit switches
Upper and lower limit switches are supplied to stop travel in both directions. The limit switch thumb screws are shipped banded together, as illustrated below:
Remove the rubber band and instructional tag before use. Refer to later sections for more detailed information about limit switches.
Models ESM303 & ESM303H Test Stands User’s Guide
2.4 Connections and outputs
The following connections and outputs are supplied in the rear of the test stand, as shown in the illustration below:
USB connector
Outputs force only or force and travel data (optional) via USB. Also allows for PC control (optional). Plug one end of the USB cable into this connector, and the other end into a PC’s USB port. Refer to the next sub-section for driver installation information.
Auxiliary limit switch connector
For interfacing an external limit switch, such as an interlock for a machine guard door. A pin diagram is shown in the next sub-section.
Control panel cable connector
Plug the cable into this connector.
Power switch
Refer to the Connecting Power sub-section for important safety information.
Power plug receptacle
Plug the power cord in here. Refer to the Connecting Power sub-section
for important safety information.
2.4.1 Auxiliary limit switch pin diagram
Note that when pins 3 and 4 are not connected, the auxiliary limits are inactive. When the +5V from pin 2 is connected to either pin 3 or pin 4, the respective limit becomes active and the crosshead is prevented from movement in that direction.
2.4.2 USB driver installation
To use this output, install the USB driver provided on the Resource CD, labeled “Mark-10 USB Device”. Installation instructions may also be found on the CD or may be downloaded from www.mark-10.com.
Caution!
Install the USB driver before physically connecting the tester to a PC with the USB cable.
Models ESM303 & ESM303H Test Stands User’s Guide
Further instructions for configuring communication parameters may be found in the Test Function Setup section.
If PC control is used, a full listing of available ASCII commands may be found in the Operating Modes section.
2.5 Installing a force gauge or indicator with load cell
2.5.1 ESM303
Once the test stand is in a stable and secure position, install a force gauge by mating the dowel pin in the
force gauge mounting plate with the blind hole in the rear of the force gauge’s housing. Use the four
thumb screws to secure the gauge. Grips and fixtures may be threaded onto the force gauge and test stand base.
2.5.2 ESM303H
1. Once the test stand is in a stable and secure position, install an indicator via four thumb screws.
2. Install a load cell to the crosshead using the supplied hardware, and insert the Plug & TestTM connector into the indicator, as shown in the image below:
2.6 Installing an indicator / load cell mounting kit (optional for ESM303)
The optional AC1062 mounting kit is available to mount a Series R01 or Series R03 load cell (force sensor) and an indicator onto the crosshead, instead of a force gauge. To install, follow these instructions:
1. Remove the force gauge mounting plate by loosening the two 5/16-18 flat head screws, as shown below:
2. Install the indicator bracket onto the side of the crosshead, utilizing two #10-32 flat head screws, as shown below:
Models ESM303 & ESM303H Test Stands User’s Guide
3. Install the L-bracket with the same two flat head screws, as shown below:
4. Mount the load cell to the L-bracket utilizing the included hardware, as shown below.
5. Mount the indicator utilizing the four provided thumb screws, and insert the Plug & TestTM connector into the indicator. Orient the housing of the indicator such that the connector is pointed downwards. Refer to the indicator’s user’s guide for details.
Models ESM303 & ESM303H Test Stands User’s Guide
2.7 Installing the 09-1214 / 09-1252 interface cable
If using test stand functions requiring communication between the force gauge / indicator and test stand, such as data output, overload protection, preload, etc., connect the interface cable between the instrument’s connector and the crosshead’s connector. This cable may be used with a Series 4, 5, or 7 instrument. Refer to the image below:
or
09-1214 (ESM303)
09-1252 (ESM303H)
Note: The 09-1214 and 09-1252 cables power the instrument. When connected, the instrument automatically powers on when the test stand is powered on. The instrument’s AC adapter is not necessary under these conditions.
2.8 Safety / proper usage
Typical materials able to be tested by the ESM303 include many raw materials and manufactured items, such as springs, metals, plastics, electronic components, mechanical assemblies, packaging materials, and many others. Potentially flammable substances or products should not be used with the ESM303. For samples which can shatter or otherwise present a physical danger to the operator, a safety shield is required, such as model no. ESM303-003. Refer to the Safety Shield section for installation instructions.
Ensure that the grip or fixture is positioned to ensure axial load with respect to the load axis of the load cell or force gauge. When using a grip, ensure that it secures the sample in such a way that it is prevented from slipping out during a test, preventing a potential safety risk to the operator and others in the vicinity. If using a grip or fixture from a supplier other than Mark-10, ensure that it is constructed of suitably rugged materials and components.
A safety shield is recommended
2.9 Connecting power
Plug one end of the power cord into its receptacle at the rear of the stand and the other end into a wall outlet with local earth ground (3-prong connector).
Before turning on power, the following safety checks and procedures should be performed:
1. Never operate the test stand if there is any visible damage to the power cord or the test stand itself. The ESM303 is powered by 110V/220V. Any contact with this high voltage can cause serious injury or even death.
2. Ensure that the test stand is kept away from water or any electrically conductive liquids at all times.
3. Make sure the electrical outlet powering the test stand has local earth ground (3-prong connector).
Models ESM303 & ESM303H Test Stands User’s Guide
4. The test stand should be serviced by a trained technician only. Power must be disconnected before the column covers are removed.
After the above safety checks and procedures have been performed, the test stand may be powered on and is ready for operation.
3 OPERATION BASICS
3.1 Operational safety
The following safety checks and procedures should be performed before and during operation:
1. Always consider the characteristics of the sample being tested before initiating a test. A risk assessment should be carried out beforehand to ensure that all safety measures have been addressed and implemented.
2. Wear eye and face protection when testing, especially when testing brittle samples that have the potential to shatter under force. Be aware of the dangers posed by potential energy that can accumulate in the sample during testing. Extra bodily protection should be worn if a destructive failure of a test sample is possible.
3. Keep away from moving parts of the test stand. Loose articles of clothing should not be worn. Long hair should be covered to avoid a hazardous situation. A Crush Hazard warning label is located on the base of the test stand. It appears as follows:
Definition: Keep any body parts and clothing clear of the area between the base of the test stand and the moving crosshead.
4. In those applications where samples can shatter, or other applications that could lead to a hazardous situation, use of a machine guard is strongly recommended.
5. When the test stand is not in use, ensure that the power is turned off to prevent accidental engagement of any of the controls.
Models ESM303 & ESM303H Test Stands User’s Guide
3.2 Controls
Label
Function
1
SOFT KEYS
Functions are determined by the corresponding text on the display.
2
UP / RIGHT
Commences movement in the up (ESM303) or right (ESM303H) direction.
3
STOP
Stops crosshead movement.
4
DOWN / LEFT
Commences movement in the down (ESM303) or left (ESM303H) direction.
5
FollowMeTM
Enables force-activated crosshead positioning. If this option is not installed, the message “NOT INSTALLED” flashes.
6
ZERO TRAVEL
Zeroes the travel display. If this option is not installed, the message “NOT INSTALLED” flashes.
7
EMERGENCY STOP
Press this button at any time to stop test stand motion. Rotate clockwise to release.
3.3 Modes overview
The ESM303 has three functional modes:
1. OPERATING MODE This is the operating mode in which testing sequences can be started and stopped.
2. TEST FUNCTION SETUP In this mode, test functions are configured, such as rate of speed, number of cycles, password editing, and other functions.
3. FUNCTION ACTIVATION
Many test stand functions, such as cycling, auto return, distance measurement, and others are ordered individually as options. The Function Activation menu provides a means of identifying which functions were purchased. Through this menu, it is also possible to enable any functions not originally purchased via an activation code. All functions are temporarily enabled for the first 160 hours of operation, as described below:
2
3 4 1
5
6
7
Models ESM303 & ESM303H Test Stands User’s Guide
DEMO MODE
The ESM303 is shipped in Demo Mode, during which time all available functions are temporarily activated for 160 hours of operation. At power-up, a counter displays the number of hours remaining, as follows:
Press STOP to continue. At the end of this period, any functions not originally purchased will be deactivated,
and will no longer be accessible from the Test Function Setup menu. Demo Mode can be suspended at any time by pressing and holding STOP while turning on power to the test stand. This mode can be re-enabled in the same manner, and will be active for the remaining time period.
Refer to the Function Activation section for instructions for field activation.
R E M A I N I N G D E M O T I M E : 1 6 0 H O U R S
4 TEST FUNCTION SETUP
This section provides configuration instructions for each function.
Function
Standard / Optional
Travel indication
Optional
Speed - same setting applies to both directions
Standard
Independent up and down speeds
Optional
High speed range extension
Optional
Low speed range extension
Optional
Auto return
Optional
Cycling with dwell time
Optional
Programmable travel limits
Optional
Overload protection
Optional
Preload
Optional
Loadholding
Optional
Break detection
Optional
PC control
Optional
FollowMeTM
Optional
Profiles
Optional
Communication settings
Standard
Speed / distance unit of measurement
Standard
Directional button configuration
Standard
Password protection
Standard
All ESM303 test stands are supplied in Demo Mode, as explained in the Overview section. After Demo Mode expires, only installed functions will be displayed in Test Function Setup. To access Test Function Setup menu, press menu from the Operating Mode home screen, which appears as follows:
S P E E D : 0 . 0 0 m e n u m i n m a x S E
T
Models ESM303 & ESM303H Test Stands User’s Guide
After pressing menu, the initial Test Function Setup screen appears as follows:
Label
Description
ESC
Exits Test Function Setup, reverts to Operating Mode
<
Scrolls to the previous function
>
Scrolls to the next function
ENTR
Selects the function, allowing it to be modified
When the functions have been configured and are ready to be saved, press ESC to exit Test Function Setup.
Note:
Changes can be made to an unlimited number of settings before saving.
If the PROFILES function is installed, the changes may be saved to the desired profile name. The screen appears as follows:
Label
Description
ESC
Exits Test Function Setup, reverts to Operating Mode
+
Scrolls to the next saved profile
EDIT
Press to edit the profile name. The first character will be flashing. Press + to change the character and –> to advance to the next character.
SAVE
Saves the settings under the specified profile name
Refer to the Profiles sub-section for further details on selecting and deleting profiles. If PROFILES is not installed, the screen appears as follows:
Make the appropriate selection.
4.1 Speed
If the Independent Speeds option has not been installed, both directional speeds will be the same, and is programmed via the SPEED function. If the Independent Speeds option is installed, UP SPEED / RIGHT SP and DN SPEED / LEFT SP functions will be present, and may be set individually.
Default setting: 10 in/min / 250 mm/min Available settings: 0.02 – 45 in/min / 0.5 – 1,100 mm/min
S P E E D : 2 0 . 0 0 E S C < − − > E N T R S A V E A S : E X A M P L E 1 E S C + E D I T S A V E S A V E C H A N G E S ? N O Y E