Thank you for purchasing a Mark-10 Series ES Force Measurement Test Stand. We are confident that you
will get many years of great service from this product.
Mark-10 test stands are ruggedly built for many years of service in laboratory and industrial environments.
This User’s Guide provides unpacking, setup, and operator instructions, along with dimensions and
specifications. For additional information or answers to your questions, our technical support and
engineering teams are eager to help you.
Thank you again for your purchase and happy testing!
1. Carefully unpack the stand and inspect for any damage. Inspect the contents to ensure that you
have received a test stand complete with all accessories – see List of Included Items.
2. Place the stand on a firm, flat and level working surface free from vibration to ensure accurate
readings.
3. Save original packaging materials in case future transportation is required.
4 OPERATION
Mark-10 force gauges with capacities up to 500 lbF [2,500 N] can be
mounted directly to the stand without adapters. Mount the gauge using
the thumb screws provided.
The loading mechanism can be repositioned along the column by
loosening the wing nut on the left side of the housing, repositioning the
mechanism as desired, and tightening the wing nut, as illustrated in Fig.
1.
To initiate a compression test, push down on the lever. Note that the
lever is spring loaded. To prevent damage to the ES05, keep your hand
on the lever until it returns to its original position.
Install the necessary attachments and make sure they are secure, safe
to operate, and set up properly for your sample. Then zero out the force
gauge and begin operation.
To maintain smooth functioning of the stand, avoid overloads.
If the optional base plate has been ordered, install it by tightening the
four screws provided, as illustrated in Fig. 2.
Lever mechanism can be repositioned
along the column.
Optional base plate can be installed
with four screws.
Measures up to 1" of travel, with 0.001" resolution.
2.2 ES002 – digital indicator kit
Measures up to 1" / 25 mm of travel, with 0.0005" / 0.01 mm resolution. Includes
zero and in/mm selector buttons.
2.3 AC1060 – Base plate with threaded holes
Optional mounting plate contains a matrix of threaded holes for grip and fixture
mounting.
3 UNPACKING AND SETTING UP
1. Carefully unpack the stand and inspect for any damage. Inspect the contents to ensure that you
have received a test stand complete with all accessories – see List of Included Items.
2. ES10: Install the loading lever into the hub.
ES20: Install the handwheel at the end of the lead screw, then tighten the set screw.
3. Place the stand on a firm, flat and level working surface free from vibration to ensure accurate
readings. If preferred, the test stand can be secured to a work bench with two screws or bolts
through the two clearance holes in the base – see Operation section for further details.
Place the stand on a clean, flat, and level surface. It is
recommended that the stand be mounted to the surface with
two 5/16 screws through the clearance hole in the base.
Install a force gauge onto the gauge plate with four thumb
screws. Mark
auges with capacities up to 500 lbF
/ 2,500 N mount directly to the stand without adapters.
The loading lever on the ES10 can be adjusted to allow for
ease of operation. Reposition the lever using one of the
three tapped holes located in the hub
If the optional dial indicator kit (ES001) or digital indicator kit
(ES002) is purchased separately, it can be installed using
the hardware provided.
Install the necessary attachments and make sure they are
secure, safe to operate, and set up properly for your sample.
Then zero out the force gauge and begin operation.
To maintain smooth functioning of the stand, avoid
overloads.
6" [150 mm] travel, 0.0005" [0.01 mm] resolution, SPC output for automated data
collection. Requires travel stops kit. To output data to a PC via USB, use
communication adapter part no. MU100 and 09-1066 cable.
3 UNPACKING AND SETTING UP
1. Carefully unpack the stand and inspect for any damage. Inspect the contents to ensure that you
have received a test stand complete with all accessories – see List of Included Items.
2. Place the stand on a firm, flat and level working surface free from vibration to ensure accurate
readings.
3. Save original packaging materials in case future transportation is required.
4 OPERATION
Place the stand on a clean, flat, and level surface. It is recommended that the stand be mounted to the
surface with two 5/16 screws through the clearance hole in the base.
Install a force gauge onto the gauge plate with four thumb screws. Mark-10 force gauges with capacities
up to 500 lbF / 2,500 N mount directly to the stand without adapters.
Install the necessary attachments and make sure they are secure, safe to operate, and set up properly for
your sample. Then zero out the force gauge and begin operation.
The base is removable so that the stand may be used as a component of a larger testing system. Proper
care should be taken during installation to avoid injury.
To maintain smooth functioning of the stand, avoid overloads.