out - Outputs selection (other than RS-232)
SP d Set point outputs disabled
SP E Set point outputs enabled
bcd d Mitutoyo BCD output disabled
bcd E Mitutoyo BCD output enabled
nPOL Mitutoyo readings without polarity (absolute value)
POL Mitutoyo readings with polarity; positive for compression, negative for
tension
Et d External trigger disabled
Et E External trigger enabled in edge mode
Et L External trigger enabled in level mode
EtHL Data capture during high to low transition of trigger signal
EtLH Data capture during low to high transition of trigger signal
Aout - Automatic output (RS-232)
no Automatic output disabled
1 Every sample transmitted
2 Every 2nd sample transmitted
4 Every 4th sample transmitted
8 Every 8th sample transmitted
16 Every 16th sample transmitted
32 Every 32nd sample transmitted
64 Every 64th sample transmitted
128 Every 128th sample transmitted
AoFF - Automatic shutoff settings
no Disabled
1 1-minute automatic shutoff
5 5-minute automatic shutoff
10 10-minute automatic shutoff
20 20-minute automatic shutoff
30 30-minute automatic shutoff
init - Initial (default) settings
LB Pounds as default units
KG Kilograms as default units
N Newtons as default units
A Average mode at turn on
EtE/EtL External trigger mode at turn on
TC Real time display at turn on
PEAK T Peak tension display at turn on
PEAK C Peak compression display at turn on
A - Average mode settings
A E Average mode enabled
A d Average mode disabled
dEL Initial delay prompt
At Average time prompt
trF Trigger force value prompt
CAL - Calibration. See Section 10.
Series CG
5
OUTPUTS
RS-232
The data transmission can be initiated by pressing the DATA key or by an external
device by sending ASCII "?" to the gauge. The gauge will respond by sending the
current reading in either full or numeric format, depending on the configuration settings
(see Section 3). Polarity sign indicates tensile (-) or compressive (+) forces. The
transmitted string has the following format:
[POLARITY (SPACE OR -)][DA T A][SP ACE][UNITS (IF ENABLED)][CRLF]
The display will flash "Err" and no data will be transmitted if DATA is pressed during
the average computation while in the Average mode or during the input scanning in the
External trigger mode.
Mitutoyo BCD
This output is useful for connection to data collectors, printers, multiplexers or any
other device capable of accepting Mitutoyo BCD data. The transmission is initiated by
the
DATA key or by the receiving device (see Section 3 for settings).
Analog
This output can be used for chart recorders, oscilloscopes, data acquisition systems,
or any other compatible devices with analog inputs. The output produces ±1 volt at full
scale of the instrument. The polarity of the signal is positive for compression and
negative for tension.
1 RS-232 receive Input
2 RS-232 transmit Output
3 Mitutoyo request Input
External trigger "
Input bit 3 "
4 Mitutoyo clock Output
"Within" set point output "
Output bit 2 "
5 Signal ground --6 +Analog signal Output
7 +12V DC Output
8 Mitutoyo ready Output
"Under" set point output "
Output bit 1 "
9 Mitutoyo data Output
"Over" set point output "
Output bit 0 "
Section 9
I/O connector pin diagram
7
DB-9P
6
1298345
User's Guide
10