Six keys on the front panel are used for all functions and control of the instrument.
Some have more than one function, depending on the mode of operation. The main
functions are labeled above the keys and the secondary functions are below the keys
in smaller type. In the list below the secondary functions are in parenthesis. For a
detailed description of the secondary functions see Section 3.
POWER (ENTER)Turns power on and off
UNITS ( )Selects units of measurement
ZERO ( )Zeroes any tare value (up to the full capacity of the
instrument) and clears the peak readings
DATAInitiates a data transmission sequence (if equipped with
the communication option)
MODE (ADVANCE) Switches the display between normal and peak modes
of operation
CLEAR (ESCAPE) Clears peak readings from memory
Orientation
In order to accomodate a variety of testing requirements, the orientation of the loading
shaft may be set up in either of the two positions shown below. In order to change the
loading shaft orientation, simply unscrew two screws on the back side of the housing,
separate the two housing halves, rotate one half 180° and reassemble.
Loading shaft
BG
SERIES
ZERO
UNITSPOWER
CLEAR
MODE
DATA
ESCAPE
ADVANCE
BGSERIES
ZEROUNITSPOWER
CLEAR
MODE
DATA
ESCAPE
ADVANCE
Loading shaft
Upright orientation
(as supplied)
Alternate orientation
(for stand mounting, etc.)
4
Mounting
BGSERIES
User's Guide
DOWEL PIN
MOUNTING PLATE
0.125 [3.2]
Section 2
Gauge shown mounted
on Model TSC test stand
POWER
BOTTOM PLATE
0.187 [4.75]Ø
Recommended use of a dowel pin
The gauge is powered by a 7.2-volt NiCd rechargeable battery. Since batteries are
subject to self-discharge, it may be necessary to recharge the unit after a prolonged
period of storage. Plug the accompanying charger into the AC outlet and insert the
charger plug into the receptacle on the gauge. The gauge may be operated for 8-10
hours after approximately 16-18 hours of charging. Do not use chargers other than
supplied or instrument damage may occur.
There are three levels of low battery voltage indication. At the first level, the display
shows a steady "LO BAT" indicating approximately one hour of charge remaining. The
second level is indicated by a flashing "LO BAT". At the third level, the whole display
except the "LO BAT" indicator will flash for three seconds after which time the gauge
will turn itself off. This prevents the instrument from working at voltages too low for
reliable operation.
CONFIGURATION
Section 3
The Series BG gauges have several features with programmable options allowing
many user-specified choices. In order to access the configuration menu, perform the
following:
1.Turn off the gauge
2.Press and hold MODE
3.Turn on the gauge
4.Release MODE
5
Series BG
The software version number will be displayed for a short time followed by "FLtA". The
following secondary functions of keys are used during the configuration process.
ADVANCEStep through menu choices
ENTERSelect a menu choice
ESCAPEQuit any function (no change)
&Increment or decrement displayed values
The following list shows all configuration options. Italics indicate factory settings.
FLtA - Analog filter
FA 02.5 Hz RC filter disabled
FA 12.5 Hz RC filter enabled
FL Td - Digit al filters
FC 1No filtering of current (displayed) readings
FC 2Average of 2 samples for each reading
FC 4Average of 4 samples for each reading
FC 8Average of 8 samples for each reading
FP 1No filtering of peak readings
FP 2Average of 2 samples for each peak reading
FP 4Average of 4 samples for each peak reading
7-1E7 data bits, 1 stop bit, even parity
7-1o7 data bits, 1 stop bit, odd parity
7-2E7 data bits, 2 stop bits, even parity
7-2o7 data bits, 2 stop bits, odd parity
7-2n7 data bits, 2 stop bits, no parity
8-1E8 data bits, 1 stop bit, even parity
8-1o8 data bits, 1 stop bit, odd parity
8-1n8 data bits, 1 stop bit, no parity
8-2n8 data bits, 2 stop bits, no parity
Ft FFull data (numeric + units)
Ft nNumeric data only
out - Outputs selection (other than RS-232)
SP dSet point outputs disabled
SP ESet point outputs enabled
6
User's Guide
bcd dMitutoyo BCD output disabled
bcd EMitutoyo BCD output enabled
nPOLMitutoyo readings without polarity (absolute value)
PO LMitutoyo readings with polarity; positive for compres-
Et dExternal trigger disabled
Et EExternal trigger enabled in edge mode
Et LExternal trigger enabled in level mode
EtHLData capture during high to low transition of trigger signal
EtLHData capture during low to high transition of trigger signal
sion, negative for tension
Aout - Automatic output (RS-232)
noAutomatic output disabled
1Every sample transmitted
2Every 2nd sample transmitted
4Every 4th sample transmitted
8Every 8th sample transmitted
16Every 16th sample transmitted
32Every 32nd sample transmitted
64Every 64th sample transmitted
128Every 128th sample transmitted
AoFF - Automatic shutoff settings
noDisabled
11-minute automatic shutoff
55-minute automatic shutoff
1010-minute automatic shutoff
2020-minute automatic shutoff
3030-minute automatic shutoff
init - Initial (default) settings
LBPounds as default units
KGKilograms as default units
NNewtons as default units
AAverage mode at turn on
EtE/EtLExternal trigger mode at turn on
TCReal time display at turn on
PEAK TPeak tension display at turn on
PEAK C Peak compression display at turn on
A - Average mode settings
A EAverage mode enabled
A dAverage mode disabled
dE LInitial delay prompt
A tAverage time prompt
trFTrigger force value prompt
7
Series BG
CAL - Calibration. See Section 10.
FILTERS
For maximum flexibility in noise suppression and peak capturing ability of the instrument, there are two types of filters available to the user: analog and digital.
The analog filter is a simple RC network with a cutoff frequency of 2.5 Hz and attenuation of 20 dB/decade. It can be either turned on or off.
The digital filter utilizes the moving average technique in which consecutive readings
are "pushed" through a buffer and the displayed reading is the average of the buffer
contents. By varying the length of the buffer, a variable smoothing effect can be
achieved. The BG is equipped with a buffer which can hold up to eight readings. The
number of readings to be averaged can be set to 1,2,4 or 8. The selection of 1 will
disable the filter since the average of a single value is the value itself.
The analog and the digital filters should be disabled or set to their minimum
acceptable values for highest peak capture speed.
EXTERNAL TRIGGER MODE
This mode of operation is useful for measuring electrical contact activation force as
well as synchronization of multiple instruments for a "snap-shot" view of applied
TRIG
GND
USER - SUPPLIED
SWITCH UNDER TEST
3
5
GAUGE
+5V
47K
forces. When this mode is enabled through
the configuration menu (see Section 3),
the MODE key will sequence through an
additional state which is indicated by the
flashing "C" or "T" indicator.
When in this mode, the instrument stops
updating the display when the trigger signal is applied. It is possible to capture the
reading with a normally open contact (high
to low transition of the trigger signal) or a
normally closed contact (low to high transition).
Section 4
Section 5
The display will show the captured reading until ZERO or CLEAR is pressed if the
"edge" mode is set. The "level" mode provides for the display to hold the reading only
until the trigger signal returns to its original state.
Please refer to the diagram for connection details and to Section 3 for configuration
information.
8
User's Guide
SET POINTS
Section 6
This feature is useful for tolerance checking (GO/NO GO) or alarm indication in process
control applications. Two limits, high and low, are specified and stored in the nonvolatile memory of the instrument and all readings are compared to these limits. The
results of the comparisons are indicated
through the three open-collector outputs provided on the 9-pin connector, thus providing
"under", "in range" and "over" signaling.
These outputs can be connected to indica-
+
tors, buzzers or relays as required for the
_
application.
2N3904
GND
4OUT
8
9
5
40 ma MAX
SAME
AS ABOVE
LOAD
24V MAX
After the Set Point mode is enabled through
GAUGE
USER - SUPPLIED
the configuration menu (see Section 3),
pressing the MODE key will sequence
through an additional step indicated by "SP" on the display. To enter or change the
values of the set points press ENTER. The high set point is displayed. Use and
keys to increase or decrease the value and MODE for changing between tension and
compression. When the desired value is displayed, press ENTER and repeat the above
steps for the low set point. After entering both values "donE" will appear on the display.
Press ENTER to store the changes or ESCAPE to quit. In either case "SP" will appear on
the display and the ENTER key may be used for re-entering the set point change mode
or the MODE key for proceeding with the normal operation of the gauge.
AVERAGE MODE
Section 7
This mode is used for obtaining an average force reading over a specified period of
time. Applications include measurement of peel force, muscle strength, frictional force
and any other tests requiring time-averaged readings. There are three user-program-
mable parameters associated with this
FORCE
PEAK
AVG
AVE
TRIGGER
FORCE
mode: trigger force, initial delay and
average time. The programming of
these parameters and the enabling of
the Average mode are done during the
gauge setup. Please refer to the "Configuration" section for more information.
DELAY
is reached and is indicated by a flashing "A". The conclusion of the test is indicated by
an alternating "A" and the calculated value. The readings obtained during the initial
delay will not be part of the average, but the peak value is stored for later recall. A new
test may be started by pressing CLEAR or ZERO.
AVERAGE TIME
Press MODE until "A" is displayed and
TIME
then CLEAR or ZERO to begin testing.
The process of averaging begins as
soon as the programmed trigger force
9
Series BG
GAUGE CONTROL LANGUAGE
The instrument can be controlled by an external device through the RS-232 channel.
The following is a list of supported commands and their interpretations. All commands
must be terminated with a Carriage Return character (hex 0D) or with a Carriage
Return/Line Feed combination (hex 0D+0A). The gauge responses are always terminated with a Carriage Return/Line Feed.
AEnable Average mode
ADDisable Average mode
AMSelect Average mode (if enabled)
ATnAverage time. n=0.1-300.0 seconds
DELnInitial delay. n=0.1-300.0 seconds
TRFnTrigger force. n=value (+ for compression, - for tension)
S PEnable Set Point mode
S P DDisable Set Point mode
SPHnHigh set point. n=value (+ for compression, - for tension)
SPLnLow set point. n=value (+ for compression, - for tension)
AOFFnAuto-shutoff. n=0,1,5,10,20,30 minutes. 0=always on
AOUTnAuto-transmit every nth reading. n=0,1,2,4,8,16,32,64,128. 0=disabled
LBSwitch units to pounds
KGSwitch units to kilograms
NSwitch units to Newtons
GSwitch units grams
ETEnable External trigger mode
ET DDisable External trigger mode
ETEEdge triggered External trigger mode
ETLLevel triggered External trigger mode
HLReading captured on a high to low transition
LHReading captured on a low to high transition
Section 8
CURCurrent mode (real time display)
P TPeak Tension mode
PCPeak Compression mode
CLRClear peaks, start a new average, or external trigger test
ZZero display and perform the CLR function
?Request the displayed reading
?CRequest the current reading
?PTRequest the peak tension reading
?PCRequest the peak compression reading
?ETRequest the reading obtained during the External trigger mode
?ARequest the average reading obtained during the Average mode
FLTCnDigital filter for current (displayed readings). n=1,2,4,8
FLTPnDigital filter for peak readings. n=1,2,4,8
FLTAnAnalog filter (2.5 Hz). n=1,2. 1=on, 0=off
10
User's Guide
FULLRS-232 transmission with units
NUMRS-232 transmission without units (only numeric values)
MITEnable Mitutoyo output
MITDDisable Mitutoyo output
POLMitutoyo outputs with polarity. (+ for compression, - for tension)
NPOLMitutoyo outputs without polarity (absolute value)
PMPrint/send data to a Mitutoyo compatible device
SnSet output bit (open collector, pull to ground). n=0,1,2
CnClear output bit. n=0,1,2
RnRead current status of output bit or level of input pin. n=0,1,2,3
SAVESave current settings in nonvolatile memory
CALEnter Calibration mode. See Section 10 for more information
LISTList current settings and status. Here is a typical LIST output:
Any detected errors are reported back by means of the following error codes.
*10Illegal command
*11Not applicable; e.g. SPHn command without enabling the set points
*21Invalid specifier; e.g. AOFF2
*22Value too large
*30Calibration weight too high
*31Calibration weight too low
*50Communication error
*51Command string too long
V3.00;LB;PC;FLTC8;FLTP1;FLTA1;AOUT00;AOFF05;FULL;MIT;POL;B0
All fields are separated by ";". The first field shows the software version, the
last field shows the remaining battery power (B0=full charge, B3=minimum
power). All other fields show the status of settings and features using the
same abbreviations as the commands to set them.
Following is a sample BASIC program illustrating the use of some commands. It switches
the units to kilograms and sets the display to zero. Press any key to get a reading on the
screen. Use "ESC" to exit the program.
10 CLS: OPEN "COM1:9600,N,8,1,RS,CS,DS,CD,LF" AS #1
20 PRINT #1 "KG"
30 PRINT #1 "Z"
40 PRINT "PRESS ANY KEY FOR READING OR <ESC> TO EXIT"
50 KEYPRS$=INKEY$: IF KEYPRS$="" THEN 50
60 IF KEYPRS$=CHR$(27) THEN SYSTEM
70 PRINT #1 "?"
80 LINE INPUT #1,A$
90 PRINT A$
100 GOTO 40
11
Series BG
OUTPUTS
Section 9
RS-232
The data transmission can be initiated by pressing the DATA key or by an external
device by sending ASCII "?" to the gauge. The gauge will respond by sending the
current reading in either full or numeric format, depending on the configuration settings
(see Section 3). Polarity sign indicates tensile (-) or compressive (+) forces. The
transmitted string has the following format:
[POLARITY (SP ACE OR -)][DATA][SPACE][UNITS (IF ENABLED)][CRLF]
The display will flash "Err" and no data will be transmitted if DA TA is pressed during the
average computation while in the Average mode or during the input scanning in the
External trigger mode.
Mitutoyo BCD
This output is useful for connection to data collectors, printers, multiplexers or any
other device capable of accepting Mitutoyo BCD data. The transmission is initiated by
the DATA key or by the receiving device (see Section 3 for settings).
Analog
This output can be used for chart recorders, oscilloscopes, data acquisition systems,
or any other compatible devices with analog inputs. The output produces ±1 volt at full
scale of the instrument. The polarity of the signal is positive for compression and
negative for tension.
Mount the gauge firmly with the loading shaft pointing downward. Go into the configuration mode as described in the previous section and select the calibration sub-menu
by pressing ENTER three times when the display shows 'CAL'. After the display shows
'null' press ZERO, while insuring that there is no weight on the loading shaft other than
the weight of the required attachments (hooks, etc.). The next displayed prompt is
'SPAn' at which time apply the exact weight equal to the full capacity of the gaugein pounds and press ENTER. A successful calibration procedure is indicated by 'donE'
on the display. Press ENTER to save the new calibration data and to return to normal
operation. In some cases the display will show 'nnnn' or 'uuuu' to indicate excessive or
insufficient calibration weight. This can be caused by incorrect weights, tare weight of
over 10% of the full capacity of the gauge or an overloaded sensor. The calibration
procedure may be aborted without changing the previous calibration information at
any time by pressing ESCAPE.
Section 10
SPECIFICATIONS Section 11
General
Accuracy:±0.2% of full scale ±1 count
Tare capacity:110% of capacity. Display shows "----" at 110%
Overload capacity:150% of capacity. Display shows "----" at 110%
Sampling rate:65 samples per second
Display update:2.5-10 times per second in normal mode, depending on
Display:4-1/2-character LCD 0.3" [7.6 mm] high
Load cell deflection:Maximum 0.010" [0.25 mm] at full scale
Outputs:
RS-232:Baud rates between 300 and 9600
Mitutoyo:Standard Mitutoyo SPC BCD output
Analog:±1 VDC ±0.25% FS
Connector:9-pin D-type male
Power:7.2 NiCd battery or included AC adapter/charger
Battery life:8-10 hours per charge
Weight:0.9 lbs [0.4 kg]
filter settings. 65 times per second in peak mode
Capacity x graduation
BG0120.12 x 0.00005 lbF50 x 0.02 gF0.5 x 0.0002 N
BG0250.25 x 0.0001 lbF100 x 0.05 gF1 x 0.0005 N
BG050.5 x 0.0002 lbF250 x 0.1 gF2 x 0.001 N
BG22 x 0.001 lbF1 x 0.0005 kgF10 x 0.005 N
BG55 x 0.002 lbF2 x 0.001 kgF25 x 0.01 N
BG1010 x 0.005 lbF5 x 0.002 kgF50 x 0.02 N
BG2020 x 0.01 lbF10 x 0.005 kgF100 x 0.05 N
BG5050 x 0.02 lbF25 x 0.01 kgF250 x 0.1 N
BG100100 x 0.05 lbF50 x 0.02 kgF500 x 0.2 N
BG200200 x 0.1 lbF100 x 0.05 kgF1000 x 0.5 N
BG500500 x 0.2 lbF250 x 0.1 kgF2500 x 1 N
13
Series BG
Dimensions
in [mm]
MODELMODEL
BG025-BG200
BG012-BG200
BG500
B
POWER UNITS
DATA
2.53 [64.3]
MODE
ADVANCE
BGSERIES
ZERO
CLEAR
ESCAPE
0.200 [5.8]
0.312 [7.9]
ØA
Ø A
BCTHREAD
0.19 [4.8]
0.44 [11.2]
0.420 [10.7]
C
2.250 [57.2]
#6-32 X4
1.50 [38.1]
0.35 [8.9]
0.31 [7.9]
Ø 0.188 [4.8]
2.130 [54.0]
#10-32
5/16-18
THREAD
0.88 [22.4]
5.90 [149.9]
4.500 [114.3]
WARRANTY Section 12
Mark-10 Corporation expressly warrants to its buyer for three (3) years from the date
of delivery that the goods sold are free from defects in workmanship and materials.
Mark-10 Corporation will, at its option, repair or replace or refund the purchased price
of goods found to be defective. This remedy shall be the buyer's sole and exclusive
remedy. Any modification, abuse, exposure to corrosive environment or use other than
intended will void this warranty. This warranty is in lieu of all other warranties, including implied warranties of merchantability and fitness for an intended purpose. In no
event shall Mark-10 Corporation be liable for any incidental and consequential damages
in connection with goods sold or any part thereof.
14
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.