08-1026 Battery Installation Instructions for Legacy Gauges
The 8.4V NiMH 08-1026 battery is designed for use with new Mark-10 gauges. They can also be used with legacy gauges, however, some modifi cations are required. Failure to perform this modifi cation will not cause damage to the gauge, however, the period of time per battery charge will be signifi cantly reduced. Applicable legacy gauge series include Series BG, BGI, EG, and CG. Refer to the instructions below:
2
3
4
5
www.GlobalTestSupply.com
1
Separate the two halves of the gauge housing by loos­ening the captive screws. Then pull the two halves apart.
The separated halves ap­pear as above. The PCB of interest is the auxiliary PCB, shown on the right half, adjacent to the battery. Remove the battery before beginning work on the modi­ cations.
Identify the R4 resistor (brown color), with the thermistor (silver color) sol­dered to it. The connection between the two compo­nents is highlighted with a yellow arrow.
Using a pair of wire cutters, cut the connection between the components.
Twist the thermistor away from the resistor, or cut it out completely. Note the absence of the connection, identifi ed by the yellow ar- row. Then, install the new battery and reassemble the two halves of the housing.