Mar Cor Purification MILLENIUM HX Maintenance Manual

3030226 Rev. D ?
Mar Cor Purification
MILLENIUM HX® REVERSE OSMOSIS UNIT
Operation and Maintenance Manual
Trademarks
MILLENIUM HX® is a registered trademark of Mar Cor Purification
MINNCARE® HD is a registered trademark of Medivators Inc.
Viton® is a registered trademark of E.I. Du Pont De Nemours And Company
Liqui-Clean and Biosan™ are registered trademarks of GE Osmonics, Inc.
Memstor is a trademark of King Lee Technologies
This page intentionally left blank.
3030226 Rev. D A ?
TABLE OF CONTENTS
CHAPTER 1: GENERAL INFORMATION .................................................................... 1-1
1.1 The Manual ................................................................................................ 1-3
1.2 General Notes, Cautions and Warnings ........................................................ 1-3
1.3 Applications ............................................................................................... 1-7
1.4 Intended Use .............................................................................................. 1-7
1.5 Contraindications ....................................................................................... 1-7
1.6 Environmental Considerations ..................................................................... 1-8
1.6.1 Power ............................................................................................................ 1-8
1.6.2 Water ............................................................................................................ 1-8
1.6.3 Drain ............................................................................................................. 1-8
1.6.4 Water Purity ................................................................................................... 1-9
1.6.5 Temperature .................................................................................................. 1-9
CHAPTER 2: INSTALLATION AND INITIAL SYSTEM START-UP ................................... 2-1
2.1 Installation Considerations .......................................................................... 2-3
2.2 Installation Procedures ............................................................................... 2-3
2.3 Product Water AAMI Analysis Procedure ................................................... 2-13
CHAPTER 3: SYSTEM OPERATION .......................................................................... 3-1
3.1 Daily Start-Up ............................................................................................. 3-3
3.2 Shut-Down Procedure ................................................................................. 3-6
3.3 Disinfection Indications ............................................................................... 3-7
3.3.1 Heat Disinfection Procedure (Single Heat) ......................................................... 3-8
3.4 System Storage ......................................................................................... 3-11
3.4.1 Storage Heat Procedure ................................................................................. 3-12
CHAPTER 4: SCREEN OPERATION AND SPECIFICATIONS .......................................... 4-1
4.1 Theory of Operation ................................................................................... 4-3
3030226 Rev. D B ?
4.2 Fluid Components ....................................................................................... 4-3
4.2.1 Solenoid Valves............................................................................................... 4-4
4.2.2 Check Valves .................................................................................................. 4-4
4.2.3 High-Pressure Pump (HPP) ............................................................................... 4-5
4.2.4 Level Switches (L1, L2 and L3) ........................................................................... 4-5
4.2.5 RO Membrane Assembly (ROMA) ..................................................................... 4-5
4.2.6 Feed Port ....................................................................................................... 4-5
4.2.7 Product Port ................................................................................................... 4-5
4.2.8 Waste Port ..................................................................................................... 4-5
4.2.9 Product Return Port ........................................................................................ 4-5
4.2.10 RO Feed Water Conductivity Sensor (WP1) ...................................................... 4-6
4.2.11 Product Water Conductivity Sensor (WP2) ....................................................... 4-6
4.2.12 RO Feed Water Pressure Sensor (PT1) ............................................................. 4-6
4.2.13 Pump Outlet Pressure Sensor (PT2) ................................................................. 4-6
4.2.14 Product Water Pressure Sensor (PT3) .............................................................. 4-6
4.2.15 Pressure Regulator (PR) ................................................................................. 4-6
4.3 System Operation Controls .......................................................................... 4-7
4.4 Alarm Screens .......................................................................................... 4-20
4.5 Optional Equipment .................................................................................. 4-22
4.5.1 Pretreatment Cart ......................................................................................... 4-22
4.5.2 Carbon Filter Tanks ....................................................................................... 4-22
4.5.3 Carbon Block ................................................................................................ 4-22
4.5.4 Softeners ..................................................................................................... 4-23
4.5.5 Dealkalizers .................................................................................................. 4-23
4.6 Specifications ........................................................................................... 4-24
4.6.1 RO Feed Water Requirements ........................................................................ 4-24
4.6.2 Flow Rates ................................................................................................... 4-25
4.6.3 Electrical Requirements ................................................................................. 4-25
4.6.4 Dimensions .................................................................................................. 4-25
4.6.5 Membrane Performance Characteristics .......................................................... 4-25
4.6.6 Environmental Requirement .......................................................................... 4-26
4.6.7 Approvals ..................................................................................................... 4-26
4.7 Disposal ................................ ................................ ................................ ... 4-26
4.8 Electromagnetic Interference .................................................................... 4-27
4.9 Service Assistance ..................................................................................... 4-31
4.10 Return Materials Authorization (RMA) Procedure ..................................... 4-31
4.11 Symbols and Abbreviations ..................................................................... 4-32
3030226 Rev. D C ?
CHAPTER 5: ROUTINE MAINTENANCE ................................................................... 5-1
5.1 Routine Maintenance.................................................................................. 5-3
5.1.1 Per Treatment Day Maintenance ...................................................................... 5-4
5.1.2 Weekly Maintenance....................................................................................... 5-4
5.1.3 Monthly Maintenance ..................................................................................... 5-5
5.1.4 Quarterly Maintenance ................................................................................... 5-5
5.1.5 Annual Maintenance ....................................................................................... 5-6
5.1.6 As Required .................................................................................................... 5-6
5.2 Specific Maintenance Procedures and Instructions ....................................... 5-7
5.2.1 Leak Repair .................................................................................................... 5-7
5.2.2 Conductivity/Temperature Sensor Verification ................................................... 5-9
5.2.3 Membrane Replacement ............................................................................... 5-10
5.3 Cleaning Indications .................................................................................. 5-12
5.3.1 Cleaning Procedure ....................................................................................... 5-13
5.4 Chemical Sanitization Procedure ................................................................ 5-19
5.5 Product Water Microbiological Testing ...................................................... 5-25
5.5.1 Product Water Culture Procedure ................................................................... 5-25
5.5.2 Endotoxin Testing Procedure .......................................................................... 5-27
5.6 Chemical Storage Procedure ...................................................................... 5-28
5.7 Auto Run .................................................................................................. 5-30
CHAPTER 6: TROUBLESHOOTING ........................................................................... 6-1
6.1 RO Fails to Run ........................................................................................... 6-3
6.2 RO Feed Pressure Alarm .............................................................................. 6-3
6.3 Product Water Quality Alarm ...................................................................... 6-4
6.4 This Unit Must Stand Upright Alarm ............................................................. 6-4
6.5 Product Water High Temperature ................................................................ 6-5
6.6 RO Feed Water Quality Alarm ...................................................................... 6-5
6.7 Pump Pressure Alarm .................................................................................. 6-6
6.8 Product Pressure High Alarm ....................................................................... 6-6
6.9 Service Required Component Failure Detected ............................................. 6-7
6.10 Product Flow Too Low ............................................................................... 6-8
6.11 Product Flow Too High .............................................................................. 6-9
6.12 Leaks ........................................................................................................ 6-9
3030226 Rev. D D ?
6.13 Low Percent Rejection ............................................................................. 6-10
6.14 TDS/µS Value High .................................................................................. 6-11
CHAPTER 7: DRAWINGS ........................................................................................ 7-1
7.1 Component Identification ........................................................................... 7-3
7.1.1 Front View ..................................................................................................... 7-3
7.1.2 Rear View....................................................................................................... 7-4
7.1.3 Electrical Right Side View ................................................................................. 7-5
7.1.4 Electrical Left Side View ................................................................................... 7-6
7.1.5 Membrane Assembly....................................................................................... 7-7
7.1.6 Top Hydraulic View ......................................................................................... 7-8
7.1.7 Center Hydraulic View ..................................................................................... 7-9
7.1.8 Main Manifold .............................................................................................. 7-10
7.1.9 Lower Hydraulic View .................................................................................... 7-11
7.1.10 Tank View .................................................................................................. 7-12
7.2 Flow Schematics ....................................................................................... 7-13
7.2.1 Run Mode .................................................................................................... 7-13
7.2.2 Divert to Drain .............................................................................................. 7-14
7.2.3 Heat/Chemical Purge..................................................................................... 7-15
7.2.4 Heat/Chemical Recirculation .......................................................................... 7-16
7.3 Electrical Diagram ..................................................................................... 7-17
7.4 Maintenance Schedule .............................................................................. 7-18
CHAPTER 8: SPARE PARTS ................................................................ ..................... 8-1
8.1 Spare Parts List ........................................................................................... 8-3
Millenium HX RO System
3030226 Rev. D 1-1 General Information
Millenium HX RO System
CHAPTER 1:
GENERAL INFORMATION
Millenium HX RO System
3030226 Rev. D 1-2 General Information
This page intentionally left blank.
Millenium HX RO System
3030226 Rev. D 1-3 General Information
1.1 The Manual
This manual has been prepared to provide the operator with information and instructions regarding the installation, use, maintenance and troubleshooting of the Mar Cor Purification Millenium HX Reverse Osmosis System.
CAUTION:
When used as a medical device, Federal law restricts this device to sale by or on the order of a physician. Per CFR 801.109 (b)(1).
The manual has been written in narrative form supplemented with schematics and drawings for clarification. The operator can perform most procedures mentioned in this manual. Any exceptions will be clearly identified by a qualifying statement.
1.2 General Notes, Cautions and Warnings
Words in BOLD CAPITAL letters are used to identify labels on the device and key safety or qualifying statements. A list of all symbols and abbreviations is located at the end of this chapter. This safety summary does not contain all of the safety statements in the manual. Other safety statements are included within the manual text and are enhanced and defined as follows:
NOTE:
Statements that provide further clarification.
CAUTION:
Statements identifying conditions or practices that could result in equipment or other property damage.
WARNING:
Statements identifying conditions or practices that could result in personal injury or loss of life.
WARNING:
Read this manual prior to operating or servicing this device. This manual must be read and understood and no deviations from the procedures shall be made. Keep this and other associated manuals for future reference and for new operators or qualified service personnel. A note sheet is provided at the end of each chapter for operators to make notations that may be valuable to other users.
Millenium HX RO System
3030226 Rev. D 1-4 General Information
WARNING:
Only chemicals listed in the manual may be used with the Millenium HX RO. Other chemicals may not be compatible with the materials of construction. When using chemical cleaners or sanitizers, ensure the system is properly rinsed prior to use for dialysis.
WARNING:
The selection of water treatment equipment for dialysis is the responsibility of the user and the product water should be tested periodically in accordance with AAMI and CSA standards.
WARNING:
This device will reach temperatures up to 180°F (82°C). Only appropriate heat resistant materials can be used for fluid pathways that are disinfected with hot water.
WARNING:
After installation and subsequent use, if any device in the water treatment system is changed or replaced, the user should conduct appropriate tests to ensure that the revised system meets the initial design criteria.
WARNING:
The product water quality is proportional to the RO feed water quality. The product water may not meet quality limits if RO feed water quality deteriorates significantly. The user is responsible for monitoring the RO feed water quality.
WARNING:
Carbon cannot be regenerated and must be replaced with new beds when exhausted or contaminated.
WARNING:
A suitable carbon filter should always be present as pretreatment for the RO in order to prevent chlorine/chloramine damage to the thin film RO membrane and harm to the patient. Mar Cor Purification strongly recommends the use of two carbon devices, used in a series configuration.
WARNING:
This RO can only operate in the vertical position. This machine has been equipped with a sensor that will prevent operation of the unit while lying horizontal. Damage to equipment may result if operated in the horizontal position.
Millenium HX RO System
3030226 Rev. D 1-5 General Information
WARNING:
Some settings are user adjustable. Entering the SETTINGS screen and/or adjusting any settings should only be done by properly trained personnel and according to facility policies and procedures.
WARNING:
Do not attempt to tamper with or perform any unauthorized modifications to the device including programming changes. Any unauthorized work on the device could cause damage to the equipment or harm to personnel if any safety features are circumvented.
WARNING:
While a water treatment system may produce water of sufficient quality to meet the requirements of AAMI and CSA standards, distribution of the water may degrade its quality to the point where it no longer meets the requirements of the above mentioned standards if the distribution system is not maintained appropriately.
WARNING:
DO NOT REMOVE COVERS OR PANELS: To avoid electrical shock hazard, do not remove covers or panels when power is supplied to the device. Do not operate the device when covers or panels are removed.
WARNING:
Connect this device to a proper ground connection in accordance with the National Electrical Code. DO NOT under any circumstances remove the ground wire or ground prong from any power plug. DO NOT use an extension cord with this equipment. Keep the surface under the RO free of water.
WARNING:
During operation of the equipment, position the RO so there is access to the power switch located on the back of the machine.
WARNING:
Do not, under any circumstances, remove any Caution, Warning or any other descriptive labels from the devices until the conditions warranting the label are eliminated.
WARNING:
To avoid fire or explosion, do not operate this device in an explosive environment or near flammable products.
Millenium HX RO System
3030226 Rev. D 1-6 General Information
WARNING:
Do not remove the back panel of the RO during operation. System is under pressure during operation. Only remove panel when RO is turned OFF.
WARNING:
To avoid physical and/or equipment harm, move the RO by using both handles at all times.
CAUTION:
Do not cover or enclose the RO during operation to ensure the unit properly dissipates heat.
CAUTION:
The input water quality to the RO must meet the National Primary Drinking Water Standards of the United States Environmental Protection Agency (EPA) or Canadian Federal and Provincial guidelines governing drinking water in Canada. If the input water does not meet these standards, the RO may not produce water that meets AAMI and CSA standards.
NOTE:
Use proper wiring and connection methods to satisfy hospital electrical codes.
NOTE:
Use proper water and drain connections to prevent contamination of the RO and danger to the patient. The Product/Return Ports are male quick disconnect to prevent mix up of connection ports.
NOTE:
The Millenium HX RO’s hot water disinfection process is effective in reducing at
least 99.9% of an introduced bacteria load of Pseudomonas aeruginosa under normal operation conditions.
NOTE:
Clean the HMI display screen as necessary with an alcohol based fluid and a soft towel as required to maintain the screen’s cleanliness.
NOTE:
This device does not supply hot water to disinfect hemodialysis machines.
Millenium HX RO System
3030226 Rev. D 1-7 General Information
1.3 Applications
The Millenium HX Reverse Osmosis (RO) System is designed to purify water by forcing water through a semi-permeable RO membrane. Water purified by reverse osmosis has had approximately 95% of the dissolved inorganic contaminants and approximately 99% of all microorganisms, pyrogens, particulates and organics with a molecular weight greater than 300 removed. The quality of the purified water, referred to as product water or permeate, depends on the quality of the RO feed water and normally is equal to or exceeds standards set forth by AAMI and CSA for water treatment equipment hemodialysis applications. Refer to Technote 103 for AAMI and CSA Product Water Quality Standards.
1.4 Intended Use
The Mar Cor Purification Millenium HX Portable Reverse Osmosis Water Purification System is intended to be used as dialysis accessory to produce water through reverse osmosis for use with hemodialysis equipment.
The Millenium HX can be connected to hemodialysis equipment used in hospitals, clinics and in home environments, in conjunction with the appropriate pre and post treatment units, as a part of a water treatment system designed to meet current AAMI and Federal (US.) Standards.
1.5Contraindications
WARNING:
This system is not designed, sold, or intended for use in producing water for injection.
Mar Cor Purification systems are not intended to be used outside of the device specifications and limitations, as outlined in this manual and other related material. If this equipment is used in a manner not specified by the operating instructions, the protection provided by the equipment may be impaired.
Millenium HX RO System
3030226 Rev. D 1-8 General Information
1.6 Environmental Considerations
Prior to the installation of the Millenium HX RO, it will be necessary to provide utilities and create an environment suitable for the trouble free operation of the RO and its accessories.
WARNING:
Ordinary pipe dope must be avoided since it may leach objectionable and potentially dangerous impurities into the water.
CAUTION:
It is necessary to keep the RO from freezing to prevent serious damage to the RO.
CAUTION:
The use of mobile telephones or communication equipment in the vicinity of the Millenium HX could adversely influence the performance of the machine. See specification in Section 4.8.
CAUTION:
Millenium HX needs special precautions regarding EMC and needs to be installed and put into service according to the EMC information provided in the Operators and Service Manuals.
1.6.1 Power
The RO operates on 115/230 VAC single-phase power. Histories of power failure, power surges, and low line voltages should be noted and reported to the manufacturer or their agent as they may create adverse conditions for the operation of equipment.
1.6.2 Water
Reverse osmosis systems require a reliable water source (rapid pressure fluctuation of greater than 10 PSI should be avoided to prevent damage). The RO needs a minimum of 20 PSI pressure at a minimum flow of 1.5 gallons per minute. It is important that water is flowing at the design flow rate when testing water pressure rather than reading static pressure. Inlet water must contain less than 0.1 ppm free chlorine.
1.6.3 Drain
A drain outlet is required for the RO. The drain must have a minimum capacity of 2 gallons per RO per minute of continuous flow for as long as and as often as the RO is operating.
Millenium HX RO System
3030226 Rev. D 1-9 General Information
1.6.4 Water Purity
Plumbing materials can significantly contribute to the contamination of the water. Metallic pipes (with the exception of 316 stainless steel) must be eliminated once the purification process has begun. Schedule 40 or 80 PVC pipes, polypropylene, PVDF, and other FDA recommended materials are suitable for most grades of purified water. Care must also be exercised in the choice of a thread sealant.
1.6.5 Temperature
The RO is normally connected to a cold water supply. The performance specifications of the RO are based on 77°F (25°C) RO feed water temperature. The Millenium HX contains a VFD motor that will increase the pump speed when operating on cold water helping to increasing the amount of product water produced.
Millenium HX RO System
3030226 Rev. D 1-10 General Information
NOTES:
Millenium HX RO System
3030226 Rev. D 2-1 Installation And
Initial System Start-Up
Millenium HX RO System
CHAPTER 2:
INSTALLATION AND
INITIAL SYSTEM START-UP
Millenium HX RO System
3030226 Rev. D 2-2 Installation And
Initial System Start-Up
This page intentionally left blank.
Millenium HX RO System
3030226 Rev. D 2-3 Installation And
Initial System Start-Up
2.1 Installation Considerations
1. Select a location for the RO near the required utilities and dialysis equipment.
2. The RO requires 115 or 230 VAC for operation depending on model.
3. The RO requires nominally 1.5 GPM water at 20 PSI. Warmer RO feed water
(greater than 77°F (25°C)) will produce more permeate (product water), and will consume more water. Inadequate RO feed water flow and pressure may cause the RO to shut down due to low pressure. The water may also require some pretreatment to ensure maximum performance and suitability. Mar Cor Purification recommends that RO feed water be chlorine/chloramine free and solids above 5 microns in size to be filtered out.
4. A free flowing drain with a minimum capacity of 2 GPM is required for the RO.
5. Back flow prevention may be required in accordance with local codes.
6. Special connections may be needed to connect to the facility water supply.
Connections vary with different facilities.
2.2 Installation Procedures
CAUTION:
If a carbon tank is present on the Millenium HX RO, the carbon tank must be completely rinsed of all carbon fines and ash before connecting to the RO. Otherwise, irreparable damage to the TF membrane and RO internals will result (refer to Carbon Tank O&M Manual).
WARNING:
The Millenium HX RO is stored with Storage Solution from the factory. Follow all Material Safety Data Sheet (MSDS) guidelines that are included with the RO.
CAUTION:
Supply water with a pH above 8.5 and containing chloramines will adversely affect the operation of the RO membrane.
CAUTION:
Any pre-treatment device supplying water to the RO must be ready for use prior to operating the RO.
Millenium HX RO System
3030226 Rev. D 2-4 Installation And
Initial System Start-Up
NOTE:
The Product Water and the Product Return connections on the RO are a male quick disconnect fitting. All other quick disconnect fittings on the cabinet are female disconnects to prevent cross connection. In addition, the RO fittings are color coded. Green is RO feed water, red is RO waste water and the RO product and return connectors are white.
1. Unpack the RO and inspect the contents for the following;
Millenium HX RO
Manual (this document)
Hose Kit
Y-Connector with adapter hose and sample port
Power Cord
Chemical Bottle
AAMI analysis kit
2. Inspect the RO for damage. When removing the RO from the packaging, you
should use 2 people to lift out of the box.
NOTE:
If using pre-treatment equipment, refer to the specific O&M manual for proper placement and hose connections.
3. Connect the RO feed inlet hose connector to the water supply, and the QD end
to the “Feed” connection on the RO. This is a green quick disconnect fitting on the RO.
4. Connect the waste water line to the red Waste” QD fitting on the back of the
RO and route to drain in accordance with the local plumbing code.
5. Assemble the product line. Take the two ¼” silicone lines from the hose kit.
Place hose clamps over the tubing and install the Y connection. Secure the hose clamps to secure the line on the Y connection. On the product side of the line, cut the silicone tubing a few inches back from the Y connector. Place the hose clamps on the tubing ends. Install the sample port and secure the hose clamps. Connect the product line between the Product and the Return connector on the back of the RO. The Y connector will be connected to the mating item on the dialysis machine. Do not connect the Y connector to the hemodialysis machine until the completion of this installation procedure.
Millenium HX RO System
3030226 Rev. D 2-5 Installation And
Initial System Start-Up
6. Plug the power cord into a 115 VAC (230 VAC) grounded outlet and reset the
GFCI if necessary (60 Hz RO only). Do not use an extension cord or an ungrounded adapter.
7. Turn the water supply to the RO ON.
8. Turn ON the power module switch located on the rear of the RO. The touch
screen will illuminate and read PLEASE WAIT as the RO starts up. Once the startup is complete, navigate to the MAIN Screen by pressing EXIT if necessary.
9. Verify the time and date on the RO. If they need to be adjusted, touch in the
area of the time and date and follow the on screen instructions to adjust. The time does not automatically adjust for day light savings.
NOTE:
The RO is ready for initial startup.
NOTE:
The RO SETTINGS have been set at the factory. It may be necessary to adjust the Product Quality Alarm set point and the Membrane Performance Monitor Setting to allow the RO to operate without interruption.
Millenium HX RO System
3030226 Rev. D 2-6 Installation And
Initial System Start-Up
10. Press the RUN button.
11. The RO will start in a flush and divert mode where the product water is diverted
to the internal tank. The machine will operate in this flush mode for a user specific time (adjustable from 30-300 seconds) indicated on the run screen. The RO will not come out of divert mode until the flush is complete and the TDS/µS is below the alarm set point. The symbol in the lower right hand side of the display indicates the machine is in divert mode. If water quality does not improve as the RO is operating, the RO may shutdown because of product water quality. If this occurs, clear the alarm screen and press RUN again.
12. After the RO startup flush is complete and the product water is below the alarm
set point, the divert valve will open and start supplying product water through the product water line. The symbol in the lower right hand side of the display indicates the machine is supplying water to the product water line.
13. Let the RO run for 15 minutes or until the percent rejection and TDS/µS values
have stabilized.
Millenium HX RO System
3030226 Rev. D 2-7 Installation And
Initial System Start-Up
WARNING:
Ensure that TDS/µS and membrane performance (% rejection) are normal for the regions water quality. Consult your facility physician. An appropriate AAMI analysis and bacteria/Endotoxin test meeting AAMI or CSA requirements of the product water must be obtained and interpreted before using the RO unit for hemodialysis treatment, refer to Section 2.3 Product Water AAMI Analysis Procedure” and Section 5.5.1 Product Water Culture Procedure”.
14. Once the RO has achieved a steady state, note the percent rejection, and
product TDS/µS (Run Screen), and the Combined Pump Feed Water Condition TDS/µS (V-1 screen). You will need these values when setting the operating parameters in the SETTINGS screen.
15. After a minimum of 30 minutes of run time, perform a Single Heat disinfection
of the RO (steps 17 through 25).
16. From the RUN screen, press and hold the STOP button for 2 seconds. The RO
will continue to operate for 60 seconds in a flush mode. When in flush, the RO RUN button is not visible until the process is complete.
NOTE:
When in stop flush, the message STOP FLUSH ACTIVE is visible on the screen indicating that the RO is flushing the membrane with low TDS/µS water.
17. Press and hold the blue banner at the top of the screen that states PRESS AND
HOLD FOR CLEANING MODES to enter the Chemical or Heating Operations selection screen.
Millenium HX RO System
3030226 Rev. D 2-8 Installation And
Initial System Start-Up
18. Review the instructions on the screen and press and hold the HEAT PROCESS
SELECT button.
WARNING:
During heat disinfection, the RO heats the internal water to approximately 176F (80C). To avoid injury, do not touch the RO connections during the heat cycle.
19. Press the ACTIVATE FAST COOL cooling method. Default cooling method is Slow
Cool. Review the instruction on the display. Verify all instructions have been completed. Once verified, press and hold the SINGLE HEAT button.
20. The machine starts to purge, filling the internal tank to the proper level with
product quality water.
21. Once the purge process is complete, the RO’s internal water is circulated and
heated to a target temperature of 176F (80C).
Millenium HX RO System
3030226 Rev. D 2-9 Installation And
Initial System Start-Up
22. Once the target temperature is achieved, the system will maintain the
temperature for 30 minutes by automatically cycling the heater modules ON/OFF.
23. After the system is held at the target temperature for 30 minutes, the RO enters
cool down. With ACTIVATE FAST COOL selected, the inlet solenoid cycles to
introduce small amounts of RO feed water to slowly reduce the RO’s
temperature.
24. After the RO reaches 120°F (49°C), the RO will enter a POST HEAT RINSE. This
will further cool and rinse the RO for a minimum of 5 additional minutes.
Millenium HX RO System
3030226 Rev. D 2-10 Installation And
Initial System Start-Up
25. After the POST HEAT RINSE is complete, the RO is ready for normal use. Press
the EXIT button to return to the CHEMICAL OR HEATING OPERATIONS screen.
26. Press EXIT to return to the Main Screen.
NOTE:
Entering the SETTINGS screen and adjusting any setting should only be done by properly trained personnel and according to facility policies and procedures.
27. From the Main Screen, press SETTINGS.
Millenium HX RO System
3030226 Rev. D 2-11 Installation And
Initial System Start-Up
28. Enter a value of 999 and press to access the alarm set point screens.
29. Press the PRODUCT QUALITY ALARM. Once in the set point screen, set the
alarm set point per facility protocol or if no protocol exists, set at two times the steady state operating TDS (µS). Once complete, press .
30. Press MEMBRANE PERFORMANCE and input the operational % Rejection noted
in step 14. Once complete, press .
Millenium HX RO System
3030226 Rev. D 2-12 Installation And
Initial System Start-Up
31. Press the Feed Water Condition Alarm. Set the RO feed water quality alarm 50
TDS/100 µS above the steady state Combined Pump Feed Water Condition noted in step 14. If the Combined Pump Feed Water Condition is greater than or within 50 TDS/100 µS of the alarm limits of 1000 TDS/2000 µS, disable the alarm to prevent nuisance alarms. To disable the alarm, set the TDS setting to
1001 or conductivity setting to 2001 µS. Once complete, press .
32. Press RETURN TO MAIN.
33. The RO product may be used in conjunction with hemodialysis treatment
equipment only after obtaining test results, which validate compliance with accepted standards.
AAMI or CSA Product Water Panel
Microbiological Culture (colony count) results
Endotoxin (pyrogen) level testing
34. The Millenium HX RO is now ready for use.
Loading...
+ 128 hidden pages