This equipment is protected in accordance with IEC Safety Class 1. It has been
designed and tested according to IEC Publication 348, 'Safety Requirements for Electronic
Measuring Apparatus', and has been supplied in a safe condition. The following pre-
cautions must be observed by the user to ensure safe operation and to retain the equip-
ment in a safe condition.
Defects and abnormal stresses
Whenever it is likely that protection has been impaired, for example as a result of
damage caused by severe conditions of transport or storage, the equipment shall be made
inoperative and be secured against any unintended operation.
Removal of covers
Removal of the covers is likely to expose live parts although reasonable precautions
have been taken in the design of the equipment to shield such parts. The equipment
shall be disconnected from the supply before carrying out any adjustment, replacement or
maintenance and repair during which the equipment shall be opened. If any adjustment,
maintenance or repair under voltage is inevitable it shall only be carried out by a
skilled person who is aware of the hazard involved.
Note that capacitors inside the equipment may still be charged when the equipment
has been disconnected from the supply. Before carrying out any work inside the equip-
ment, capacitors connected to high voltage points should be discharged; to discharge
mains filter capacitors, if fitted, short together the L (live) and N (neutral) pins of
the mains plug.
Mains plug
The mains plug shall only be inserted in a socket outlet provided with a protective
earth contacts The protective action shall not be negated by the use of an extension
lead without protective conductor. Any interruption of the protective conductor inside
or outside the equipment is likely to make the equipment dangerous.
Fuses
Note that there is a supply fuse in both the live and neutral wires of the supply
lead. If only one of these fuses should rupture, certain parts of the equipment could
remain at supply potential.
To provide protection against breakdown of the supply lead, its connectors, and
filter where fitted, an external supply fuse (e.g. fitted in the connecting plug) should
be used in the live lead. The fuse should have a continuous rating not exceeding 6 A.
Make sure that only fuses with the required rated current and of the specified type
are used for replacement. The use of mended fuses and the short-circuiting of fuse
holders shall be avoided.
RADIO FREQUENCY INTERFERENCE
This equipment conforms with the requirements of lEC Directive 76/889 as to
limits of r.f. interference.
(2a)
Page 3
Section 1 GENERAL INFORMATION
Contents
1.1
1.2
1.3
Section 2 OPERATION
2.
,1
2.
,2
2.
.3
2.
.4
2.
5
2.
,6
2.
.7
2.8
2.9
2.10
2.11
Section 3 TECHNICAL DESCRIPTION
3.1 Circuit summary
3.2 Measuring bridge a. c. supply
3.3 Measuring bridge d. c. supply
3.4 Measuring bridge circuits
3.5 Amplifier detector
3.6 Power supplies
3.7 D. C. Choke Adaptor Type TM 6113
Introduction
Data summary
Accessories
Installation
Controls and connectors
Preliminaries
Connecting the component
Resistance measurement
Inductance measurement
Notes on inductlve loss balancing
Using the D. C. Choke Adaptor Type TM
Capacitance measurement
Notes on capacitive loss balancing
Use of external oscillator and detector
6113
9
10
10
12
16
17
18
21
21
23
24
24
24
25
26
26
Section 4 MAINTENANCE
4.1 General
4. 2 Removal from case
4.3 Access to components
4.4 Mains input adjustments
4.5 Working voltages
4.6 Replacement of valves
4. 7 Adjustment of preset components
4.8 Checking the loss balance diai calibration
4. 9 Access and adjustment to balance control assembly
Section 5 REPLACEABLE PARTS
Universal bridge
D. C. choke adaptor
Section 6 CIRCUIT DIAGRAMS
Fig. 6.1 Power supply and oscillator
Fig. 6.2 Bridge
Fig. 6.3 Detector
Fig. 6.4 D. C. choke adaptor
1313A (1)
28
28
28
28
28
31
31
36
36
41
53
47
49
51
53
Page 4
General information
1.1 INTRODUCTION
TF 1313A is a general-purpose impedance
bridge with one tenth per cent measurement
accuracy over a wide range of inductance, capacitance, and resistance values. It offers exceptional discrimination and resettability, wide-range
loss balancing, and facilities for using an external
oscillator and detector.
The single direct-reading dial used for L,
C, and R measurements has coarse aiid fine concentric controls; the coarse one moves in 1%
steps and the fine gives continuously variable
interpolation between steps. The variable and
non-linear sensitivity of the detector facilitates
easy measurement of completely unknown comp-
onents. An external audio oscillator can be
plugged in where L and C measurements are required at frequencies other than the internal 1 and
10kc/s; the detector output is available externally
to allow an oscilloscope or headphones to be used
for balance indication.
This bridge can be used for precision evaluation of resistance, capacitance and inductance,
for measuring the dissipation factor of capacitors
and Q of inductors, and for quickly identifying
unknown components. Measurements on high-loss
components are made easier by the low D and Q
ranges of the loss balance control and the relative
independence between the adjustment of the main
and loss balance controls. Its high discrimination
and resettability make it particularly suitable for
comparative measurements such as checking the
difference between an unknown and a standard
component. Measurements on inductors carrying
d. c. can be made by using Adaptor TypeTM6113,
which is available as an optional accessory, while
an external d. c. supply can be connected direct
for polarizing capacitors.
Fig. 1.1 0-1% Universal bridge TF 1313A
1313A (la)
Page 5
1.2 DATA SUMMARY
Resistance measurement
General information
Range
Accuracy :
Residual resistance
Inductance measurement
Rangé :
Accuracy :
3 mΩ to 110 MΩ. in eight ranges with maxima of 11Ω, 110Ω,
1.1 kΩ, 11 kΩ, 110 kΩ, 1.1 MΩ, 11 MΩ, 110 MΩ.
Basic measurement error :
±0.1% of reading, or ±0. 015% of rangé maximum, whichever
11 Ω and 11 MΩ ranges basic errors, and additional ±0.1% of
reading.
11 MΩ X 10 rangé - basic errors, and additional^±0.15% of
reading.
Less than 0.003 Ω.
0.1 µH to 110 H in seven ranges, with maxima of 110 μH,
1.1 mH, 11 mH, 110 mH, 1.1 H, 11 H, 110 H.
Basic measurement error at 1 kc/s :
±0.1% of reading, or ±0. 015% of rangé maximum whichever is
greater.
Residual inductance :
Capacitance measurement
Range :
Accuracy :
(when D is not greater
than 0.031)
Basic measurement error at 10 kc/s :
±0. 2% of reading, or ±0. 025% of range maximum, whichever is
greater.
Range errors :
1.1 mH to 11 H ranges inclusive - basic errors only.
110 ΜH and 110 H ranges basic errors, and additional ±0.1%
of reading.
Additional errors at Low Q :
± (0.1 xf/Q)%, where f is in kc/s.
Typically 0. 05 ΜH.
0.1 pF to 110 µF in seven ranges, with maxima of 110 pF, 1100 pF,
0.011 µF, 0.11 ΜF, 1.1 ΜF, 11 ΜF, 110 ΜF.
Basic measurement error at 1 kc/s :
±0.1% of reading, or ±0. 015% of range maximum, whichever
is greater. (When D is greater than 0. 031 additional error
is
typically
±0.3D2%).
1313A (lc)
Page 6
General information
Accuracy :
(when D is not greater
than 0.031)
- continued
Residual capacitance
Temperature range
Temperature coefficient
Q and D measurement
Range :
Basic measurement error at 10 kc/s :
±0. 2% of reading, or ±0. 025% of range maximum, whichever
is greater. (When D is greater than 0.031 and less than 0. 31,
additional
error is typically
±6D2%.)
Range errors ;
1100 pF to 11 ΜF ranges inclusive basic errors only.
110 pF and 110 µF ranges basic errors, and additional ±0.1%
of reading.
Less than 0. 05 pF.
18 C to 28 C for the stated accuracies.
Additional error of ±0.01% per degree C, for temperatures between
10°C and 18°C, and 28°C and 35°C.
Low Q range :
Normal Q range
Normal D range
High D range :
1 kc/s
0 to 3
0.5 to 31
0.0005 to 0.031
0,005 to 3
10 kc/s
0 to 30
5 to 310
0.005 to 0.031 (limited)
Not required
Accuracy :
Bridge sources and detector
Internal sources :
External oscillators :
Additional L and C errors Typically :
Normal Q : ±5% of reading, ±0. 5% of full scale.
Normal D : ±5% of reading.
Low Q and High D : ±10% of reading, ±3% of full scale.
Additional D or 1/Q error at 1 kc/s and below :
less than ±0. 0005 with correction supplied (on top of instrument), or
less than ±0. 0015 without correction.
(Above 1 kc/s multiply by f kc/s.)
1 kc/s and 10 kc/s oscillators for L and C measurement,
accuracy ±2. 5%; output level, depending on loading, up to 750 mV.
D. C. supply for R measurement; less than, 100 mW component
loading.
Frequency range: 20 c/s to 30 kc/s.
Input level required : 3 to 20 V depending on frequency.
(An external tuned detector is also necessary to achieve the quoted
measurement accuracies.)
Frequency
20 c/s
100 c/s
20 kc/s
30 kc/s
% error
±.05
±.03
±0.2
±0.35
1313A (lb)
Page 7
Test terminal voltage at balance:
General information
CAPACITOR BIAS :
Up to 350 V d. c. may be applied for polarizing electrolytic
capacitors.
POWER SUPPLY
A. C. mains :
100 to 130 V and 200 to 250 V, 45 to 180, 275 to 300, 366 to 400 c/s.
25 VA.
DIMENSIONS & WEIGHT
1.3 ACCESSORIES
Supplied
Three telephone plugs, type P40, for external
oscillator and detector and bias jacks.
Available
D. C. Choke Adaptor, type TM 6113 ; enables d. c.
currents up to 200 mA from an external supply
to be passed through inductors under test at
1 kc/s in the range 100 mH to 100 H; fitted with
test leads for attaching to bridge terminals.
Errors introduced by the adaptor do not
generally exceed 3% and may be eliminated by
simple substitution methods.
Page 8
Operation
2.1 INSTALLATION
The 0.1% Universal Bridge is normally dispatched with its valves in position and with its
mains input circuit adjusted ready for immediate
use on 240 V within the basic supply frequency
range 45 to 400 c/s. Note that harmonics of
supply frequencies of 200, 250 and 333 c/s may
give rise to interference which will obscure the
balance point.
If required, the instrument may be adjusted
for operation from other supplies wlthin the ran-
ges 100 to 130 and 200 to 250 V. To check, or
alter the tappings on the mains transformer, refer
to Section 4.4.
2.2 CONTROLS AND CONNECTIONS
(T) TEST TERMINALS. Connect the component
between the Hl and DET terminals.
(2) COMPONENT PLATFORM and ABRIDGED
OPERATING INSTRUCTIONS. Supports
components, isolating them from chassis.
If approximate value unknown, set BALANCE
controls to half scale, and then select range
giving lowest meter reading.
BALANCE METER.
minimum reading.
SENSITIVITY control. Adjust initially to
give half scale meter reading. As balance
is approached increase sensitivity to give
required discrimination.
COARSE BALANCE control. Set to half
scale while searching for correct range,
and then adjust for minimum meter reading.
FINE BALANCE control. . Adjust for
minimum meter reading once minimum
for COARSE BALANCE control has been
foimd. Note that when the instrument
has been idle for some time the protective
oil film may, before dispersion, cause
erratic balance, A number of brisk
rotations will restore the contact surface.
Adjust for appropriate mea-
Adjust controls for
EXT D-Q terminals. Linked for measure-
ments at 1 kc/s to 10 kc/s; connect to external potentiometer, to give continuous
coverage when using an external oscillator
at frequencies below 1 kc/s.
BIAS jack. Connect external d. c. supply
of up to 350 V for polarizing capacitors.
SUPPLY switch.
PILOT LAMP.
2.3 PRELIMINARIES
Having checked that the instrument is adjusted
for the supply voltage it will use, proceed as follows :-
1. If necessary, mechanically zero the meter.
2. Connect the mains lead to the power supply.
3. Set the SUPPLY switch to ON; the pilot
light should now glow.
LOSS BALANCE control. Adjust to obtain
final balance, (minimum meter reading) in
conjunction with FINE BALANCE control.
FINE D-Q control. Adjust for final balance
only when LOSS BALANCE control is too
coarse.
1 kc/s - 10 kc/s switch. Select 1 kc/s
except for low value inductors. LOSS BALANCE dial readings multiplied by 10 when
this switch is at 10 kc/s. Inopera.tive for
resistance measurements.
D-Q switch. Select range appropriate to
component under test. Inoperative for resistance measurements.
EXT OSC. Connect an external oscillator
with frequencies from 20 c/s to 35 kc/s.
EXT DET socket. Connect an external
detector to monitor output, when using an
external oscillator.
4. Before using the instrument, allow a short
warm-up period. Two or three minutes is
normally sufficient for most purposes.
If you are using this Bridge for the first
time, it will be helpful at this stage to get to know
the recommended technique of reading the main
dial, before making a measurement.
The coarse control moves the pointer of the
outer dial in steps equal to one dial division; the
fine control for the inner dial is continuously
variable, and the range 0 to 100 of the inner scale
is equivalent to one division of the outer scale.
The coarse control is a switch with light indexing
for ease of operation; no attempt should be made
to obtain a setting between the marks.
The value of a component is most easily
found by first noting the reading of the outer scale
followed by the reading of the inner scale, and then
putting in the decimal point by referring to the figure displayed in the nearest scale window. Typical
examples are shown in Fig. 2.2.
1313A (la)
Page 10
operation
2.4 CONNECTING THE COMPONENT
Connect the component between the HI and
DET terminals, using short leads where possible.
Small components can often be directly supported
by the test terminals; those which can't be sup
ported should be laid on the insulated component
platform. Note the following precautions :
(a) Impedance between the HI side of the comp
onent and earth, shunts the 'balance' arm
of the bridge, and may result in an inaccurate
reading.
(b) Stray pickup on the lead to the DET terminal
is applied to the input of the bridge detector
and may cause difficulty in obtaining a clear
balance indication. Using a screened lead,
with the screen connected to the E terminal,
helps to avoid this effect.
(c) The presence of an earth on either side of
the component will shortcircuit the bridge
and make measurement impossible.
(d) The bridge energizing voltages appearing at
the test terminals (see Data Summary) are
too small to cause damage to the component
or shock to the operator.
(e) When measuring screened inductors, the
screen may be left disconnected, or connected
to the E terminal.
2.5 RESISTANCE MEASUREMENT
Resistance can be measured at d. c. only.
The 1 kc/s10 kc/s switch and the loss balance
controls are inoperative.
Set the LCR switch to R (or Rx10 if the anti
cipated value is above 11 MΩ).
Fig. 2.2 L: Reading 3,782 mH
C:
Reading
R: Reading 83,22 M Ω
0.006652ΜF
If the resistance value is completely unknown,
set the RANGE switch to its highest setting.
Set the COARSE and FINE BALANCE pointers
to mid scale.
Adjust the SENSITIVITY control to give a
meter deflection of no more than half scale.
1313A (1)
Page 11
Adjust the RANGE switch to give the lowest
meter reading. If the resistance value is
approximately known select the appropriate
range.
Balance the bridge by adjusting first the
COARSE and finally the FINE BALANCE
Controls, to give a minimum meter def
lection, progressively advancing the
SENSITIVITY control as required.
Read the resistance value from the BAL
ANCE scales as described in Section 2.3.
measuring Low wattage resistors. The general
Circuit used for resistance measurement is shown
in Fig. 2. 3. The bridge source resistance is
30 Ω and the first two range resistors are 1 Ω and
10 Ω. Using this Circuit to calculate the power
dissipation in resistors under test it will be found
that the maximum is 100 mW for resistors between
1050 n.
If a resistor of less than 100 mW rating is
to be measured the power dissipation can be red
uced by adding a resistor of known value in series
with the test resistor. The value of such a res
Note : It may be observed that many composition
resistors show a continuous drift in value.
Measuring low-wattage resistors
On the lower ranges, the effect of the 4 V
source e. m. f. may have to be considered when
1313A (1)
SERIES RESISTOR
RANGE RESISTOR
TEST RESISTOR
Fig. 2.3 Circuit used in measuring low-wattage resistors
II
Page 12
Operation
istor should be no larger than is sufficient to reduce the dissipation to an acceptable level otherwise
measurement discrimination will suffer. Alternatively, at the expense of accuracy, selecting a
higher range than usual will have the desired
effect.
Detector zero
The detector circuit used for resistance
measurements is extremely sensitive and is sus
ceptible to thermal e. m. f. 's especially on the
10 Ω, 10 MΩ and 100 MΩ ranges. Such an e. m. f.
will produce a residual meter deflection and an
internal preset control is provided to enable this
to be offset. If the control requires resetting
allow at least 20 minutes from switching on the
Bridge for internal thermal stabilization and then
remove the top plate as described in Section 4. 3.
The control is located behind the Balance
switch assembly and is labelled RV11 SET ZERO.
Set the RANGE switch and BALANCE controls to
10 MΩ, short circuit the DET and E terminals
with copper wire and adjust RV11 for minimum
meter reading with maximum SENSITIVITY.
Residual and connection resistance
where this is considered to be significant the connections should be measured separately and allowed
for in the result.
Further extraneous errors can be avoided
by not handling the resistor under test and making
sure that it is isolated from earth or any external
supplies.
2.6 INDUCTANCE MEASUREMENT
When measuring inductance, the loss balance controls are used to balance out the resistive
component of the inductor. The loss balance dial
is directreading at 1 kc/s only; readings on either
scale must be multiplied by 10 when measuring at
10 kc/s further details are given in Section 2. 7.
The main balance dial is direct reading at all frequencies.
Most audio and powerfrequency indiietors
may be measured at Ikc/s using the 030 Q range.
For small lowQ inductors such as r. f. coils it is
better to use 10 kc/s and the 03 Q range. Large
inductors with a low selfresonant frequency, or
small inductors with a saturable core need special
precautions and these are described later in the
section.
When measuring resistors near the lower
limit of the range, the residual resistance (R0)
should be deducted from the value obtained. Also
note that the component connections can have a
significant resistance.
An approximate value of R0 is given in the
Data Simimary, but the actual value for a partic-
ular bridge may be found as follows :
Short circuit the HI and DET terminals
with a thick copper strap (of 'zero' resistance). Set the LCR switch to R, and the
RANGE switch to its lowest setting.
Balance the bridge with the FINE BALANCE
control to as near zero on the meter as
possible. Read the residual resistance
from the FINE BALANCE scale; the '02'
calibration is equivalent to 0.002 Ω.
Typical component connections have a
resistance of about 1 mΩ per inch (0.4 mΩ per cm).
This amounts of 0.1% per inch at 1 n. Therefore
Set the LCR switch to L.
Set the DQ switch to Q = 030.
Set the 1 kc/s 10 kc/s switch to 1 kc/s.
Set the LOSS BALANCE controlto about
10.
Set the FINE Q and the FINE and COARSE
BALANCE controls to about midway.
Set the SENSITIVITY control to give a
meter deflection of no more than half
scale.
12
1313A (1)
Page 13
Operation
Set the LCR switch to L.
Set the D-Q switch to Q = 0-3.
Set the 1 kc/s-10 kc/s switch to 10 kc/s.
Tum the LOSS BALANCE control fully
clockwise.
Set the FINE Q control mid-way.
Set the RANGE switch to its highest setting.
Set the COARSE BALANCE control to 0.
First measure the approximate inductance
as follows : -
Set the SENSITIVITY contrbl to give a
meter deflection of no more than half scale.
Using the FINE BALANCE control only
search for a null. If the best null is below
05 reduce the RANGE by one step. Once
an apparent null has been obtained on
scale adjust the LOSS BALANCE and FINE
BALANCE controls suecessively to obtain
a minimum meter reading.
If the inductor Q is extremely low, the
balance indication will be rather flat.
Read the inductance value from the FINE
BALANCE scale noting that its range is
1/10th of that indicated in the lower left
hand window. This will only be an approx-
imate value, although the sharper the balance indication the more accurate the value
obtained.
To measure the coil accurately, proceed
as follows :-
' 8 ) Select the RANGE suitable for the approx-
imate value obtained above, usually two
(T) lower. Set the COARSE BALANCE pointer
to the value and re-balance the bridge with
(4) slight adjustment of the LOSS BALANCE
control to obtain the lowest meter reading.
14
1313A (1)
Page 14
Operation
Offset the COARSE BALANCE control a
step at a time (usually counter clockwise)
re-balancing with the LOSS BALANCE
control and advancing the SENSITIVITY
control as necessary. Use the FINE Q
control if the loss adjustment becomes
critical.
It is important to progress to the lowest
possible reading, ultimately using the
FINE BALANCE and FINE Q controls to
obtain a zero reading - or at least within
one or two divisions - on the meter scale,
with the SENSITIVITY control fully clock-
wise.
Measuring iron-cored inductors
To avoid errors in measurement, inductors
having a low self-resonant frequency should not be
tested at 10 kc/s. If their self-resonant frequency
is lower than about 5 kc/s, then test results evén
at 1 kc/s will be in error - inductance will appear
high, and magnification low. It will be found that
1 henry is the practical upper limit for measurements at 10 kc/s.
These errors will increase rapidly as the
self-resonant frequency of the inductor approaches
the test frequency, and it is impossible to measure
an inductor which resonates below 1 kc/s, since
it will exhibit capacitive reactance. This effective
capacitance may be measured on the bridge, but
it will appear to have a poor dissipation factor.
The inherent capacitance of the bridge at the test
terminals is low - approximately 0. 05 pF - and
will contribute little to lowering the self-resonant
frequency of the component.
Shunt capacitance however, due to the pos-
itioning of the components and arrangement of the
test leads, may lead to appreciable errors in
measurement. Such errors, which will increase
with frequency, can be minimized by careful positionlng of the component, using short well-spaced
leads and,, whenever possible, supporting it on the
insulated component platform.
Core saturation
If the inductor is of such a type that its
core is easily saturated, it is advisable to use
10 kc/s and to switch to the highest range that
still covers the component value (for example by
measuring a 10 mH inductor on the 10- to 100 mH
rangé instead of the 0- to 10 mH range) thereby
reducing the current through the inductor windings.
Note : Iron- or ferrite-cored inductors are part-
icularly liable to pick up stray radiation
which, applied to the bridge via the DET terminál,
will tend to mask the balance point indication. To
check for this effect, insert an unconnected tele-
phoné plug into the EXT OSC jack in order to switch
off the internal oscillator. Any reading on the
méter is now due to stray pick-up, and the inductor
should be turned or repositioned to minimize the
effect.
ResiduaI inductance
When measuring inductance near the lower
limit of the range, the residual inductance (Lo)
should be deducted from the value obtained.
An approximate value of Lo is given in the
Data Summary, but the actual value for a particular
bridge may be found as follows :-
Short-circuit the Hl and DET terminals
with a thick copper strap.
Set the LCR switch to L, and the RANGE
switch to its lowest setting.
Switch to 10 kc/s; Q = 0-3; set the coarse
BALANCE control to zero, and the fine
control to '10 '.
Balance the bridge in the normal manner
to obtain a balance as near zero, on the
meter, as possible.
Read the residual inductance from the
FINE BALANCE scale; the '10' calibration
is equivalent to 0.1 µH.
Screening cans on h. f. coils
These are usually best connected to the E
terminál but if already connected to the inductor
then jóin to the DET terminál. 10 kc/s test frequency should be used as the shorted turn effect of
an aluminlum or copper can rapidly reduces at
lower frequencies.
1313A (1)
15
Page 15
Page 16
Page 17
Page 18
Page 19
Page 20
Page 21
Page 22
Page 23
Page 24
Page 25
Page 26
Page 27
Page 28
Page 29
Page 30
Page 31
Page 32
Page 33
Page 34
Page 35
Page 36
Page 37
Page 38
Page 39
Page 40
Page 41
Page 42
Page 43
Fig. 6.1 Power supply and oscillator
47 1313 A (Id)
Page 44
Page 45
Page 46
D.C. CHOKE ADAPTOR TYPE TM 6113
CAPACITORS
KNOBS
CI Cer 2000 pF 20% 500 V 26364210
C2 Cer 2000 pF 20% 500 V 26364210
C3 Cer 2000 pF 20% 500 V 26364210
C4 Cer 2000 pF 20% 500 V 26364210
C5 Cer 2000 pF 20% 500 V 26364210
06 Cer 2000 pF 20% 500 V 26364210
C7 Cer 2000 pF 20% 500 V 26364210
C8 0.047µF(S.I.C.)*
C9 0.047µF(S.I.C.)*
CIO 1533 pF (S.I. C.)
Cll Elec 1 µF 275 V 26452101
C12 Var 5005500 pF 26857329
C13 Elec
68ΜF
+20%50%315V 26415358
Knob for C12 TB 28666
Knob for SA TO 17848/4
INDUCTORS
LI 360 mH coil assembly TB 31309
L2 360 mH coil assembly TB 31309
SWITCH
SA Rotary
•Adjust this capacitance value to archive
resonance at 1kc/s swith associated inductor
TC 4428/545
1313A (lb)
Fig. 6.4 D.C. choke adaptor
53
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.