•Excellent accuracy traceable to National
Standards
•Power levels from -70 dBm
(100 pW) to +44 dBm (25 W)
•Internal power reference
•Audible Pass/Fail limits alarm
•Rugged case
•Four digit display
The 6970 RF Power Meter provides precision
microwave power measurements in a hand
portable battery-powered package. A wide
range of user features, including duty cycle, relative power measurements and limit checking
make the 6970 Power Meter highly versatile.
The same wide range of power sensors used
with the IFR 6960 Power Meters and the 6200B
series Microwave Test Set are used by the 6970.
Field and Bench Use
The 6970 is designed for both field and bench use. For the installation and maintenance engineer the instrument replaces large and
expensive meters with no loss of measurement performance.
Rechargeable batteries mean that there is no need for an AC supply;
this greatly eases measurements at remote sites such as an antenna
tower.
Additional benefits of the 6970 are lower cost and portability. An
optional soft carrying pouch with both belt loop and shoulder strap
enable the 6970 to be easily carried in field environments
For bench operation the instrument takes up little bench space and
it can be powered from the AC adapter.
Battery operation is ideal for field operation
High Accuracy
The 6900 series power sensors all have an excellent return loss specification to minimise mismatch errors in power measurements.
Correction for power sensor frequency response and high power
non-linearity is also provided by entry of calibration factor and linearity factor values into the 6970.
For the very latest specifications visit www.aeroflex.com
Page 2
All the power sensors are individually calibrated and supplied with a
calibration data chart.
The auto ZERO function reduces offset errors on each of the 6970
ranges. This increases measurement accuracy at low power levels
and improves the measurement sensitivity.
Built-In Power Reference
The built-in power reference provides an accurate 0 dBm (1 mW) 50
MHz calibration signal. Power sensors can be calibrated against this
reference for improved measurement accuracy and traceability.
The power reference is automatically switched on during a CAL procedure or it can be enabled at any time to provide a traceable 0 dBm
signal for verification.
Operation
The instrument front panel consists of a keyboard and a four digit
LCD with annunciators. Power levels are clearly displayed on a digital readout with an analog bargraph to facilitate peaking and nulling.
Power levels may either be displayed in logarithmic units such as dBm
or dBV or in Watts.
To save battery life, the unit powers down six minutes after the last
key press. 6970 will operate for 7 hours from a full battery charge.
When the instrument powers down the last operator set-up is stored
and this will be the state the 6970 assumes when it is next switched
on.
Upper and lower power limits can be set. An audio and visual alarm
alerts the user of a measurement that is out of limits. This is useful
to set a power to a pre-determined level as well as for monitor-ing
purposes.
The 6970 is fully auto-ranging but manual range hold is also possible.
Manual range selection is used when measuring modulated or unstable signals. The display indicates either AUTO or MAN as appropriate and will also provide information of over/under-range conditions.
Wide Range of Power Sensors
The 6970 uses the same rugged IFR power sensors that have been
developed for the 6960 Power Meter and 6200B series Microwave Test
Set. A wide range of 17 sensors is available covering the frequencies
from 30 kHz to 46 GHz. Power level measured depends on sensor
type, sensors cover from -70 dBm (0.1 nW) to +44 dBm (25 W).
To cater for the different frequency ranges connector types include N
type, PC 3.5 mm, 2.92 mm and N type (75 W).
SSPPEECCIIFFIICCAATTIIOONN
Frequency Range
30 kHz to 46 GHz
Depending on sensor used
Power Range
-70 dBm (0.1 nW) to +44 dBm (25 W)
Depending on sensor used
Power Sensors Supported
6910 series (-30 dBm to +20 dBm)
6920 series (-70 dBm to -20 dBm)
6930 series (-15 dBm to +35 dBm)
6930 series opt 2 (-5 dBm to +44 dBm)
Instrumentation Accuracy Including Carryover
±1%range 0, 1, 2
±5%range 3
Power Accuracy
After calibration using 0 dBm, 50 MHz power reference: ±0.2 dB
Measuring a 50 MHz signal in the centre of the power sensor dynamic
range, from a source with a return loss better than 14 dB
DISPLAY
Resolution
4 digits for positive readings
3 digits for negative readings
Units
dBm, nW, mW, mW, W, mV, V, dBV
Annunciators
Analog bargraph, Battery low indicator, Indication of auto-ranging or
manual range, Upper and lower limits
CORRECTION
Linearity Factor
Ability to enter in range 0.01 to 15 with 0.01 resolution
Defaults to standard setting
Calibration Factor
Ability to enter in range 0.01% to 200%
Resolution:0.01 to 99.99%
0.1 100% and above
Auto-Calibration
Ability to calibrate against a 0 dBm (1 mW),
50 MHz power reference
Auto-Zero
Removes DC offset from gain stages and power sensor
Produces audible tones when the measured power is above or below
programmed limits
Batteries
3 rechargeable NiCd AA size
Operating time
>7 hours from full charge when new
Operating Time
>7 hours from full charge when new
Time to Recharge
<14 hour
Power Consumption
250 mW
DC Input Requirement
9 V to 21 V, 120 mA via a 2.1 mm power connector
GENERAL
Operating Temperature Range
0 to +55°C
Storage Temperature Range
-40 to +55°C
Storage Humidity Range
Up to 85% RH at +40°C
Storage Altitude Range
Up to 4600 m (15000 ft)
DIMENSIONS AND WEIGHT
HeightWidthDepthWeight
50 mm88 mm190 mm550 g
1.9 in3.5 in7.5 in
ELECTROMAGNETIC COMPATIBILITY
Conforms with the protection requirements of the EEC Council Directive
89/336/EEC. Conforms with the limits specified in the following standards:
IEC/EN61326-1 : 1997, RF Emission Class B, Immunity Table 1,
Performance Criteria B
SAFETY
Conforms with the requirements of EEC Council Directive 73/23/EEC
(as amended) and the product safety standard IEC / EN 61010-1 :
2001 + C1 : 2002 + C2 : 2003 for class 3 portable equipment, for
use in a Pollution Degree 2 environment. The instrument is designed to
be operated from an Installation Category 1 supply.
POWER REFERENCE
Frequency
50 MHz ±0.05 MHz
Power level
0 dBm (1 mW)
Uncertainty
±0.7% traceable to National Standards
Accuracy
±1.2% worst case for one year
Output Connector
N (female), 50
mm power sensors.
Ω
. Adapters are supplied with 75 Ω, 3.5 mm and 2.92
For the very latest specifications visit www.aeroflex.com
When ordering please quote the full ordering number information.
Ordering Numbers
Versions
6970RF Power Meter with built-in power reference 6970 RF
Power Meter must be used with one of the AC Adapters
listed below. One Adapter must therefore be ordered with
each instrument
54441/016AC Adapter - UK Style
54441/017AC Adapter - European Style
54441/018AC Adapter - USA Style
Supplied with
43138/6631.5 m Power Sensor Cable
46882/182Operating Manual
46882/183Summary Card
23421/6412.1 mm Power Connector Plug
Accessories
54112/159Carrying Pouch
54311/171DC Supply Lead
46882/207Service Manual
54417/002Waveguide 22 to 2.92 mm Transformer
POWER SENSORS - STANDARD
56910/90010 MHz to 20 GHz (-30 dBm to +20 dBm) Type N.
56911/90010 MHz to 20 GHz (-30 dBm to +20 dBm) APC 7.
56912/90030 kHz to 4.2 GHz (-30 dBm to +20 dBm) Type N.
56913/90010 MHz to 26.5 GHz (-30 dBm to +20 dBm) MPC
3.5.
56914/00110 MHz to 40 GHz (-30 dBm to +20 dBm)
2.92 mm.
56914/00210 MHz to 40 GHz (-30 dBm to +20 dBm) 2.92 mm
plus waveguide 22 coax transition and calibration
table.
56914/00310 MHz to 46 GHz (-30 dBm to +20 dBm)
2.92 mm.
56919/900
POWER SENSORS - LOW POWER
56920/90010 MHz to 20 GHz (-70 dBm to -20 dBm) Type N.
56923/90010 MHz to 26.5 GHz (-65 dBm to -20 dBm) MPC
75 Ω 30 kHz to 3 GHz (-30 dBm to +20 dBm) Type N
3.5
56924/00110 MHz to 40 GHz (-65 dBm to -20 dBm) 2.92 mm.
56924/00210 MHz to 40 GHz (-65 dBm to -20 dBm) 2.92 mm
plus waveguide 22 coax transition and calibration
table
56924/00310 MHz to 46 GHz. (-65 dBm to -20 dBm)
2.92 mm.
Power Sensors - High power
56930/90010 MHz to 18 GHz (-15 dBm to +35 dBm) Type N.
56932/90030 kHz to 4.2 GHz (-15 dBm to +35 dBm) Type N.
56934/00110 MHz to 40 GHz (-15 dBm to +30 dBm) 2.92 mm
56934/00210 MHz to 40 GHz (-15 dBm to +30 dBm) 2.92 mm
plus waveguide 22 coax transition and calibration
table
56934/00310 MHz to 46 GHz (-15 dBm to +30 dBm) 2.92 mm
56930/00210 MHz to 18 GHz (-5 dBm to +44 dBm) Type N
56932/00230 kHz to 4.2 GHz (-5 dBm to +44 dBm) Type N
Part No. 46891/066, Issue 7, 02/04
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.