Mantracourt Electronics Limited SGA/A & SGA/D User Manual
1
Page 3
Chapter 1 Introduction to SGA/A & SGA/D
Load Cells
Strain Gauges
Torque Transducers
Pressure Transducers
The Strain Gauge Amplifier SGA
The SGA is a Strain Gauge Amplifier, converting a strain gauge input to a Voltage or current output – otherwise
known as a Signal conditioner.
The SGA provides a wide range of signal conditioning for Strain gauges, Load cells, Pressure and Torque transducers.
Figure 1.1 SGA Signal Conditioner
Inputs
POWER
SUPPLY
110/230 V AC
18-24V DC
STRAIN
GAUGE
AMPLIFIER
INPUT
OUTPUT
±10V
±5V
0-5V
0-20mA
4-20mA
Offered in two versions, the SGA/A for 110/230V AC or 18-24V DC operation and the SGA/D which is DC powered
only.
Transducer SENSITIVITY of between 0.1 mV/V and 30 mV/V are possible.
This is achieved by a combination of gain (span) DIP switches and associated fine adjustment by a potentiometer.
Similarly transducer zero OFFSET and SCALE DEAD BAND of up to 79% can be compensated for in the module.
This is achieved again by a combination of zero DIP switches and associated fine adjustment by a potentiometer.
The module has built-in FILTERS to cancel the field effects of vibration, agitation and electrically noisy
environment.
The on-board low pass filter can be switched in and adjusted (from 1Hz to 5kHz) using a series of DIP switches.
A wide range of proportional output options for currents and voltages can be configured by DIP switch settings.
Both the AC and DC versions are based on a common board and are mounted in an IP65 (NEMA 4X) ABS case.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
2
Page 4
Chapter 2 Installing the SGA/A & SGA/D
Pre Installation
See Specification details in Chapter 8 for details of Environmental Approvals.
Carefully remove the SGA/A unit from its packing. Check that the unit is complete and undamaged.
The SGA/A & SGA/D units can operated in any industrial environment providing the following limits are not
exceeded
Operating Temperature -10 ºC to +50 ºC
Humidity 95% non condensing
Storage temperature -10 ºC to +50 ºC
While the unit is sealed to IP65 (NEMA 4X) it is advisable to follow the following installation practice where possible
• Minimise vibration.
• Do not mount next to strong electrical fields (transformers, power cables)
• Ensure easy access to interior of the module
• Install electrical protection device, as the unit is not internally fused.
• Always ensure the lid is properly fitted and all 4 screws tightened.
• Always ensure the cable gland is sealing against the cable to maintain the IP (NEMA) rating.
Figure 2.1 Dimensions
5
16
148mm (5 ”)160mm (6 ”)
7
8
Allow for cable entry
80mm
(3⅛”)
at either end
50mm
15
(1 ”)
16
Mounting Holes
Depth
55mm(2.16”)
The 4 screws for the lid are captive and must be tightened to maintain the seal.
The 4.5mm (0.18”) holes for the mounting screws in the base are directly behind the screws for the lid.
The box must not be drilled as this would invalidate the IP rating
Allow sufficient space at both sides for the cable entry.
The Nylon 66 cable glands are designed for ROUND cables.
The waterproof entry and strain relief will seal to a higher rating than the enclosure.
Cable diameter should be between 4mm (0.16”) and 7mm (0.27”)
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
NOTE:The SGA/A can be powered from AC or DC sources whichever is available.
It is also possible to connect BOTH AC and DC simultaneously for security of power
supply.
Figure 2.2 Power Connection
220-230 V AC
LL
110 - 120 V AC
18 -24 V DC
N
N
Standard mains 2 or 3 core cable PVC sheathed (unshielded) cable will suffice for the power.
NOTE:Connect the appropriate power to the SGA. For AC powering observe the correct
transformer jumper connections as shown in Figure 2.2 above.
(This diagram is also provided inside the lid).
Connections to the SGA/A & SGA/D input/output signal and the power supply are made via 2.5mm² field terminal
connectors.
Cable entry in the cased versions is via glands in the ends of the case.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
4
Page 6
Figure 2.3 Input (Sensor) Connections
NOTE: Strain Excite is the Excitation to the transducer.
Strain Input is the Signal from the transducer.
The Ref 5V/2.5V is generated internally and used for calibration
The cable connecting the sensor to the SGA should be shielded.
This typical cable data is provided for information only.
The cable should have 2 x twin twisted cables. Ideally with each pair individually shielded and with an overall
shield.
Miniature- twin -round Diameter: 4.8 mm
Impedance: 62 Ohms: Capacitance/m: core to core 120 pF & core to
shield 210 pF
If possible segregate the signal cable from Power Cables; allow a 1meter (3 feet) distance from such cables.
Do not run signal cables parallel to power cables. Cross such cables at right angles.
The ground connection conductor should have sufficient cross-sectional area to ensure a low impedance path to
attenuate RF interference.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
5
Page 7
Output Connections
Two analogue outputs are available from the SGA, proportional DC current and DC voltage.
The ranges available are as follows: -
Output Range
DC voltage ±10V
NB: Maximum Load on voltage ranges is 2mA.
±5V 0 - 10V 0 - 5V
DC current 0 - 20mA NB: Maximum impedance 500R.
4 - 20mA
The DC current support both ‘sink’ and ‘source’ modes of operation.
Two jumpers JP1 & JP2 provide the means of selecting the desired mode.
Figure 2.4 Output Connections
In ‘Sink’ mode the +ve end of the load is connected to the internal +15V supply on the SGA and the -ve end is
connected to the SGA output. The current through the load is ‘sunk’ by the SGA towards ground (0V).
N.B. In this mode neither connection to the output load is electrically common to the load cell.
Select this option by fitting the two jumpers, JP1 and JP2 to the ‘outside’ positions (See Figure 3.2)
In ‘Source’ mode the +ve end of the load is connected to the SGA output and the current is ‘sourced’ by the SGA
output through the load towards ground (0V).
This mode has the advantage that the negative output connection is common to the load cell
‘- Excitation’ terminal.
Select this option by fitting the two jumpers, JP1 and JP2 to the ‘inside’ positions (See Figure 3.2)
See Chapter 3 for switch settings and details of SINK & SOURCE jumpers.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
6
Page 8
Chapter 3 Switch Settings
O
utput Option
Switch Positions
e.g. The switches in Figure 3.1 are depicted as ALL ON.
Figure 3.1 Output Settings–Switch 4
Use switch 4 to select the required output and, if
required, the low pass filter and 5V Excitation.
(See Tables 3.1 and 3.2)
- Full Scale n/a n/a 4mA Note 1 0mA Note 1 0V 0V -10V -5V
N.B. Full scale output on the voltage ranges is achieved with a bi-polar (±) input
Note 1 Negative inputs can be accommodated on the current output ranges by setting the ‘Zero’ switch SW2 to
+50% (Table 3.8) and setting SW1 to twice the required mV/V setting (Table 3.6).
↓↓↓↓↓↓
Table 3.2 Switch 4
Analogue Output and Excitation Voltage Options - SW4
SW4 1 2 3 4 5 6 7 8
±10V
±5V
0-10V
0-5V
0-20mA X X X
4-20mA X X X 1↑1↑1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
Filter out X X X X X
Filter in X X X X X
10V Exc X X X X X
5V Exc X X X X X
Switch settings (0 = Off 1 = On X = Don’t Care)
Important:
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
7
0↓0↓0↓
0↓1↑0↓
0↓1↑1↑
1↑1↑1↑
Low pass filtering is switched into operation by setting SW4/6 ‘ON’↑ and SW4/7 ‘OFF’↓.
Reverse these settings to bypass the filter.
It should be noted that either one of these switches MUST be on but not BOTH
X X
X X
X X
X X
0↓0↓1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
1↑=Filter in 1↑Filter out 1↑=10V Exc 0↓=5V Exc
0↓1↑1↑=10V Exc 0↓=5V Exc
1↑0↓1↑=10V Exc 0↓=5V Exc
1↑=Filter in 1↑Filter out 1↑
1↑=Filter in 1↑Filter out 0↓
Page 9
Example :- 0-10 Volt output with no filter required.
Table 3.3
SW4 1 2 3 4 5 6 7 8
0-10V
0 ↓ 1 ↑ 1 ↑ X X 0 ↓ 1 ↑ X
SW4
tsrrsssrst
Output Filter Settings –Switch 3
The SGA incorporates a second order (-12dB/oct) low pass filter which can be switched in to improve the
performance and output signal quality in electrically noisy environments.
It can also be used to reduce the effects of high frequency fluctuations in the load or applied force to the load cell.
The cut off frequency of the filter is set by the DIP switch SW3 as illustrated in the table below
** Note:
A SECONDARY low pass filter, with a cut off frequency of 800Hz, can be switched into the SGA input by fitting a
link to JP3 (see Figure 3.2)
Important:
Low pass filtering is switched into operation by setting SW4/6 ‘ON’↑ and
SW4/7 ‘OFF’↓. Reverse these settings to bypass the filter.
It should be noted that either one of these switches MUST be on but not BOTH
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
8
Page 10
Example :The Switch Settings for a cut-off frequency of 50 Hz setting is illustrated below.
Note: SW4/6 must be ‘ON’ and SW4/7 must be ‘OFF’.
Table 3.5
SW3 1 2 3 4 5 6 7 8
50Hz 1
SW3
trrrrrrrrt
1
1
1
1
1
1
1
↑
↑
↑
↑
↑
↑
↑
↑
Output Current Mode and Input Filter Settings – Jumpers JP1, JP2 & JP3
Figure 3.2
Refer to Figure 2.4 for details of wiring connections to J1.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
9
Page 11
Span (Gain) Setting Switch SW1
28.30 mV/V
23.10 mV/V
15.60 mV/V
29.30 mV/V
24.60 mV/V
17.80 mV/V
30.30 mV/V
25.90 mV/V
19.70 mV/V
10.50 mV/V
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
1
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
2
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
3
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
4
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
5
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
6
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
7
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
8
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
x
49
45
41
37
33
29
25
21
17
13
13.20 mV/V
Ranges 1 to 60 – from 0.06 mV/V to 30.30 mV/V
↑ = ON (1) ↓ = OFF (0). SW1/8 switches on the shunt cal function – see Table 3.8
Table 3.6 – SW1
1
0.06 mV/V
1
2
3
5
0.28 mV/V
1
2
3
9
0.44 mV/V
1
2
3
0.61 mV/V
1
2
3
0.75 mV/V
1
2
3
1.41 mV/V
1
2
3
2.07 mV/V
1
2
3
2.91 mV/V
1
2
3
3.46 mV/V
1
2
3
4.00 mV/V
1
2
3
4.63 mV/V
1
2
3
5.54 mV/V
1
2
3
1
2
3
53
21.50 mV/V
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
4
5
6
2
0.11 mV/V
7
8
x
6
0.31 mV/V
7
8
x
10
0.50 mV/V
7
8
x
14
0.65 mV/V
7
8
x
18
0.80 mV/V
7
8
x
22
1.49 mV/V
7
8
x
26
2.35 mV/V
7
8
x
30
2.95 mV/V
7
8
x
34
3.72 mV/V
7
8
x
38
4.05 mV/V
7
8
x
42
4.89 mV/V
7
8
x
46
5.72 mV/V
7
8
x
50
7
8
x
54
3
0.17 mV/V
7
0.34 mV/V
11
0.55 mV/V
15
0.70 mV/V
19
0.91 mV/V
23
1.78 mV/V
27
2.49 mV/V
31
3.19 mV/V
35
3.73 mV/V
39
4.26 mV/V
43
5.12 mV/V
47
7.50 mV/V
51
55
4
0.23 mV/V
8
0.39 mV/V
12
0.60 mV/V
16
0.75 mV/V
20
1.20 mV/V
24
1.99 mV/V
28
2.63 mV/V
32
3.35 mV/V
36
4.00 mV/V
40
4.36 mV/V
44
5.34 mV/V
48
52
56
1 2 3 4 5 6 7 8
x
57
27.10 mV/V
58
59
60
1 2 3 4 5 6 7 8
x
Please Note:
When using 5V Excitation (SW4 switch 8 = OFF), divide the transducer's mV/V output by two and set SW1 to the
nearest setting shown in table 3.6 above
e.g. for 2.5mV/V with 5V excitation choose the 1.2mV/V setting
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
10
Page 12
Example :-
A strain gauge has a sensitivity of 2.809 mV /V - Select Switch Setting number 28 from Table 3.6 and fine tune with
potentiometer PI
Table 3.7
SW1 1 2 3 4 5 6 7 8
2.63 mV/V 0
1
1
0
0
0
↓
↑
↑
↓
↓
↓
0
X
↓
SW1
Refer to Chapter 4 for calibration details.
tsrrssssst
Shunt Calibration Switch SW1/8
SW1/8 connects a 120k 50ppm surface mount resistor across the ‘+Excitation’ and ‘+ Input’ terminals of the SGA.
This shunts one arm of the connected load cell to produce a known change in output which can be used for
calibration or checking the integrity of the load cell and associated wiring.
Table 3.8
SW1 1 2 3 4 5 6 7 8
Shunt Cal ON x x x x x x x 1
↑
Shunt Cal OFF x x x x x x x 0
↓
The 120k resistor can be taken out of circuit and replaced by a user defined leaded component by carefully cutting
the fine link as shown in Figure 3.3. Use the right hand pad and either of the left hand pads to fit the new
component.
The surface mount resistor can be reinstated by re-connecting the two pads either side of the cut link.
Figure 3.3
Zero (Offset) Setting Switch SW2
This offset can be used to compensate for the transducer zero error, to tare the scale dead load or to shift the
output.
These settings allow the user to calibrate a zero offset. The range allows for up to 79% of the span.
Potentiometer P2 provides fine adjustment.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
11
Page 13
Table 3.9
SW2 1 2 3 4 5 6 7 8
% + ve Offset - ve Offset 40% 20% 10% 5% 2% 1%
Example :-
An installation has a tare of 15 kg with a 200kg strain gauge which gives an output of 6.37mV/V at 10V excitation.
The tare equates to 7.5% (15/200). Set the switches to nearest % (5 + 2) and fine trim with Potentiometer P2.
The tare must be subtracted therefore the ‘- ve Offset’ switch SW2/2 should be ‘ON’.
The calibrated zero mV reading would be 4.78 mV i.e. 7.5% of 63.7mV
Table 3.10
SW2 1 2 3 4 5 6 7 8
7.5% 0
SW2
tsrsssrrst
Note
1
0
0
0
1
1
0
↓
↑
↓
↓
↓
↑
↑
↓
SW2 /1 & 2 should never be 'ON ' together. Either one or other should be 'ON ' if an
offset is required; otherwise both switches should be 'OFF '.
Switch settings 3 to 8 are ADDITIVE. The offset value of each switch is added to give
a total offset of 78%.
Fine adjustment is provided by potentiometer P2.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
12
Page 14
Chapter 4 Calibration
The SGA/A & SGA/D provides the excitation supply and signal conditioning to cater for a wide range of strain
gauges, load cells, pressure transducers or torque transducers.
Output
Select the analogue output range as detailed in Chapter 3, Figure 3.1,Table 3.1 & 3.2 by means of SW4.
Zero Offset
Select the offset as detailed in Chapter 3, Table 3.9 by means of SW2.
Having selected the polarity and the offset nearest to that required with the switches use the fine potentiometer P2
to achieve the final setting.
Sensitivity
Select the sensitivity as detailed in Chapter 3, Table 3.6 by means of SW1.
Switches 1-4 of SW1 provide fine setting of the SGA sensitivity while switches 5-7 give coarse control.
This arrangement allows the SGA to cover a wide range of strain gauge sensitivities without sacrificing stability and
ease of set up.
Locate the required sensitivity in the table and set switches 1-7 of SW1 accordingly.
Potentiometer P1 provides fine trimming and range overlap to enable the SGA to be calibrated precisely to any
given value within its ranges.
Note1If therangeisrepeatedinthetablee.g.4mV/V(4.0,4.05and4.0mV/V) choose the
setting which has the greatest number of switches 1-4 set to ‘off’ i.e.SW1=[1000][000].Thiswillenablefinertrimmingtothefinal valueusingpotentiometerPI.
The sensitivity settings shown in Table 3.6 assume that the load cell is fully loaded. The sensitivity settings can be
used to maximise the output when the full range of the load cell is not being used. Here are a couple of examples.
Example1A 2.5mV/V load cell provides 10V for an l00Ib load. However it is never loaded above
50lb
The sensitivity setting can be set to 1.25 mV/V.
Table 3.6 /20 (1.20mV/V SW1 = [1101][000]
Example 2 When a reduced output is required from a fully loaded transducer, use a less
sensitive switch setting.
For an 8 volt output from a fully loaded 2.5mV/V load cell use the 3.19mV/V setting
i.e. (10/8x2.5=3.125mV/V)
Table 3.6 /31 (1.20mV/V SW1 =[0010][000]
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
13
Page 15
The SGA/A & SGA/D can be calibrated with the transducer connected, provided that two calibration points can be
implemented. e.g. by applying known weights or forces. If this is not possible, a stable mV source or load cell
simulator can be used provided that the precise sensitivity (mV/V) and full range output (kg) of the transducer is
known. In this case the 'Ref (5V/2.5V)' output should be connected to 'Strain Input-' and the mV source applied
between ‘Strain Input+’ and ‘Strain Input-’.
Actual calibration is carried out in the following way:-
1. Set the correct switch settings on SW1 as described above using the transducer's calibration sheet supplied by the
manufacturer. This is normally specified as sensitivity or full range output and should be in mV/V
2. Apply the known low calibration conditions (weight, force or mV/V: this may be zero if required), and note the
analogue output, having ensured that the SW1 settings are correct for the transducer sensitivity as step 1 above.
3. Apply the known high calibration conditions (for optimum accuracy this should be at least 75% of full load) and
note the analogue output.
4. Use the fine trim control, P1, to obtain the required change in Volts or mA, between the two calibration points
(steps 2 and 3). e.g. If the required output at the low calibration point is 0V and the required output at the high
calibration point is 7.5V, adjust P1 in step 4 to produce a change of 7.5V between the calibration points.
Initially, the low calibration point may not produce 0V at the output. If this is the case, note the reading, e.g.
0.5V, apply the high calibration conditions and trim P1 for the required change in output. i.e. Trim the output
for 0.5 + 7.5 = 8V.
5. Use the fine ‘Zero’ control, P2 in conjunction with the coarse switches SW2/3-8 and polarity switches SW2/1 and
2 to set the output to the required absolute values. Each switch within SW2 offsets the output by a particular
percentage of full scale as shown in Table 3.9
N.B. It may be necessary to repeat these steps until the required output is achieved.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
14
Page 16
Figure 4.1 Calibration Connections using Millivolt Source
J2
Strain Input
(Strain Gauge Output)
Strain Excite ( 10V/5V)
-
Excitation
+
Shield (0V)
Ref (5V/2.5V)
-
+
-
mV Source
1. The ‘Ref (5V/2.5V)’ should be connected to ‘Strain Input-’ and the mV source applied between ‘Strain Input+’ &
‘Strain Input-’
2. Set the correct switch settings on SW1 as described above using the transducer's calibration sheet supplied by the
manufacturer. This is normally specified as sensitivity or full range output and should be in mV/V
3. Ensure the Zero and Span switch settings are correct, as detailed in Chapter 3, Tables 3.6 & 3.9
4. Apply the known low calibration conditions and fine adjust P2.
5. Apply the known high calibration conditions and fine adjust P1
6. Repeat steps 3 and 4 until the required output is achieved.
+
Hint If the required output at the low calibration point is 0V and the required output at
the high calibration point is 7.5V, adjust P1 in step 5 to produce a change of 7.5V
between the calibration points. Initially, the low calibration point may not produce 0V
at the output. If this is the case, note the reading, e.g. 0.5V, apply the high
calibration conditions and trim P1 for the required change in output, i.e. Trim the
output for 0.5 + 7.5 = 8V.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
15
Page 17
Chapter 5 Troubleshooting
1. No output
a) Check power supply is present (LED is on).
b) Check the output connections are correct.
c) Check terminations (ensure insulation is not trapped in terminal, cable break etc.)
d) Check the sensor is connected (typically reading 350 Ohm across Strain Excite + and – and also
Strain Input + and – of J2) with the power off.
e) Check the Excitation voltage (J2) is at 10V DC
1.a For voltage output
a) Check V out+ and V out- terminals are wired
b) Check the load is connected and is not open or short circuited
c) Check SW4 settings are correct for Voltage Output see Chapter 3, Table 3.2
d) Check Span and Zero settings (SW1 and SW2)
1.b For current output
a) Check Isink+ and Isink- terminals are used for 'Sink' current output
b) Check Isource+ and Isource- terminals are used for 'Source' current output.
c) Check the load is connected and is not open circuit
d) Check load does not exceed 500 Ohms.
e) In 'Sink' mode check 15 V is present at +ve terminal of load.
f) In 'Source' mode check the -ve terminal of the load is connected to ground.
g) In 'Sink' mode check the load is isolated from the load cell (sensor) excitation.
h) In 'Source' mode check the -ve output is common to the -ve Excitation.
i) Check output SW 4 settings are correct for current see Chapter 3, Table 3.2
j) Check Span and Zero settings (SW1 and SW2) see Chapter 3, Table 3.6 & 3.9
2. Low Output
This is when an output is present but not of sufficient magnitude to meet the required value.
b) Check power supply is within specified limits (i.e. is not low)
c) Check the sensor is connected (typically reading 350 Ohm across Strain Excite + and – and also
Strain Input + and – of J2) with the power off.
d) Check the Excitation voltage (J2) is at 10V DC
e) Check the calibration. Incorrect setting of the calibration Span switches are the most common cause of low
output - particularly when associated with ± Voltage outputs. Refer to the calibration instructions in Chapter 4.
Refer to tutorial on the calibration set-up.
f) Check the Zero (offset) is correct for the sensor. This too is a common reason for low outputs.
3. High output
This is when an output is present but higher (in span or zero) than required.
b) Check the sensor is connected (typically reading 350 Ohm across Strain Excite + and – and also
Strain Input + and – of J2) with the power off.
c) Check the Excitation voltage (J2) is at 10V DC
d) Check the Zero (offset) is correct for the sensor. This is a common reason for high outputs where the offset is
either omitted or incorrect for the sensor. Refer to the calibration instructions in Chapter 4
e) Refer to tutorial on the calibration set-up
f) Check the calibration. Incorrect setting of the calibration span switches is the most common cause of high output
- particularly when associated with ± Voltage outputs.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
16
Page 18
4. Unstable Output
This is when the output is unstable or varies. The cause could be (a) poor installation or (b) a noisy environment.
Poor Installation -This is when an output is present but higher or lower (in span or zero) than expected:
a) Check the installation for problems and repair where necessary
b) Poor termination
c) High resistance on cable leads
d) Low insulation impedance
e) Proximity to High Voltage Equipment – Transformers, Contactors, Motors etc.
Noisy Environment-
a) Check if the source can be found and remove noise
b)
Check the cable shielding and ensure it is correctly installed and terminated
5.Calibration
This section assumes that the unit is providing an output that is not stuck at top or bottom of the scale.
(See paragraphs 1 to 4 if this is the case)
Ensure you have the calibration set-up correctly installed i.e.mV source and output as required.
Ensure you are connected to the correct sensor and not to another adjacent unit.
Ensure you have the correct calibration data from the sensor manufacturer. This must include a certified table with
offset, zero and linearity.
Ensure the temperature and other environmental parameters are within specification and where necessary taken
into account when calibrating should such parameters have an effect on the calibration.
6. Fine Span (Gain) and Zero (Offset ) Adjustment Problems
If the adjustment cannot reach the maximum output desired then, check the tare is not too high.
If the potentiometer does not alter the output the unit must be repaired – remove from service.
It is always wise to check a known good SGA against the problem installation before rejecting the suspect SGA.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
17
Page 19
Chapter 6 Product Care
A worn out component, excessive use in harsh environments, an overly zealous operator; regrettably some
circumstances necessitate repair.
At Mantracourt Electronics Ltd we can't guarantee that a product will never require repairing. We can, however,
promise a repair service of exceptional quality, one which is governed by a rigorous procedure.
Detailed below is our pledge to you: a defined set of ground rules and procedures to which we will adhere. All we
ask in return is that you assist us with our procedure, such that we can maintain our promise to you. Please note
that warranty repairs may not be available on overdue accounts, and that a strict interpretation of our conditions of
trading invalidates warranty claims where late payment has occurred.
Please refer to ‘Customer Repair Service Procedure’ document – contact your supplier for a copy.
In the unlikely event you have problems with the SGA module we would advise that you take the following
precautions:-
• The unit is installed as instructed.
• Recommended spares are kept in stock. We can assist.
• Sufficient expertise available for first line maintenance.
• Routine maintenance checks are performed – annually is recommended.
• The necessary documentation for the product is available to the maintenance personnel.
We recommend you keep on file – as a minimum
• This Manual
• The settings of the switches and links on the SGA card
• The calibration figures for the attached sensors
• The instrument loop to which the output is connected
• A record of the ‘normal’ output – if applicable
• A maintenance record of the SGA
• A contact phone number from the supplier for assistance
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
18
Page 20
Chapter 7 Glossary
AWG American Wire Gauge.
Background Noise The total noise floor from all sources of interference in a measurement
Bipolar The ability of a signal conditioner to display either positive or negative
Bridge Resistance The resistance measured across the excitation terminals of a strain
Calibration The process of adjusting an instrument or compiling a deviation chart so
CMR
(Common-Mode
Rejection)
Common Mode Rejection
Ratio
Deadband / hysteresis (Hysteresis) In a digital controller, there may be one switching point at
Drift A change of a reading or a set point value over long periods due to
Dual Power supply The SGA/A can have a Dual Power Supply. An AC supply can be connected
Excitation The external application of electrical voltage applied to a transducer for
Fine Adjustment The Zero and Span calibration have a Fine Adjustment to give accuracy to
Full Bridge A Wheatstone bridge configuration utilizing four active elements or strain
Full Range Output The algebraic difference between the minimum output and maximum
Gain Gain is otherwise identified as SPAN. It relates to the proportional output
Ground 1)The electrical neutral line having the same potential as the surrounding
Input Impedance The resistance measured across the excitation terminals of a transducer.
Linearity The closeness of a calibration curve to a specified straight line. Linearity
Load The electrical demand of a process expressed as power (watts), current
Load Impedance The impedance presented to the output terminals of a transducer by the
Load cell The load cell is one of a series of Strain Gauge sensors that the SGA input
Low Pass Filter
system, independent of the presence of a data signal. (See noise)
readings.
gauge.
that it’s reading can be correlated to the actual value being measured.
The ability of an instrument to eliminate the effect of AC or DC noise
between signal and ground. Normally expressed in dB at dc to 60 Hz. One
type of CMR is specified between SIG LO and PWR GND. In differential
meters, a second type of CMR is specified between SIG LO and ANA GND
(METER GND).
The ability of an instrument to reject interference from a common
voltage at its input terminals with relation to ground. Usually expressed
in db (decibels).
which the signal increases and another switching point at which the signal
decreases. The difference between the two switching points is hysterisis.
several factors including change in ambient temperature, time, and line
voltage.
along with a DC supply for additional security.
normal operation.
the calibration. These are potentiometers P1 and P2 for span and zero
respectively.
gauges.
output.
to the sensor input. Calibration of the SGA is determined by setting the
Gain (Span) and Offset (Zero).
The amount of amplification used in an electrical circuit.
ground. 2) The negative side of power supply. 3) Reference point for an
electrical system.
is expressed as the maximum deviation of any calibration point on a
specified straight line during any one calibration cycle.
(amps) or resistance (ohms).
associated external circuitry.
is designed to accept. (Torque Sensor, Pressure & temperature
transducers).
The SGA Module has a low pass filter to remove unwanted signals on the
output. This can be set to suit the installation, from DC to 5kHz.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
19
Page 21
Millivolt One thousandth of a volt, 10-3 volts symbol mV.
NEMA 4/ UL Type 4 A standard from the National Electrical Manufacturers Association, which
defines enclosures, intended for indoor or outdoor use primarily to
provide a degree of protection against windblown dust and rain, splashing
water, and hose-directed water.
Noise An unwanted electrical interference on the signal wires.
Null A condition, such as balance, which results in a minimum absolute value
of output.
Offset Offset is otherwise identified as Zero. It relates to the proportional
output to the sensor input. Calibration of the SGA is determined by
setting the Offset (Zero) and Gain (Span).
Potentiometer Two potentiometers (variable resistors) are used in the SGA for fine
calibration.
Pressure Transducer The Pressure Transducer is one of a series of Strain Gauge sensors that
the SGA input is designed to accept. (Torque Sensor, Load Cell and
Temperature transducers).
Proportional Outputs
The Voltage or Current outputs are calibrated to be directly proportional
to the input from the sensor. The output is, within the sensor limits,
taken as linear and no linearity compensation is required within the SGA.
Resolution The input corresponding to a one-unit change in the least significant digit
of the data acquisition /display equipment (Good resolution is not
necessarily equal to good accuracy.)
Sensing Element That part of the transducer, which reacts directly in response to the
input.
Sensitivity The minimum change in input signal to which an instrument can respond.
This is the relationship between the change in strain gauge output to the
level or magnitude of the SGA output
Signal Conditioner A circuit module that offsets attenuates, amplifies, linearizes and/or
filters the signal for input to an A/D converter. A typical output signal
conditioning is 4 to 20 mA.
The SGA is essentially a Signal Conditioner –more specifically known as a
Strain Gauge Amplifier - in that it conditions (alters) the input signal from
a load cell to an electrical output
Single card assembly The SGA has only the one Printed Circuit Board assembly on which all the
components are mounted. The assembly is then mounted inside an
environmentally rugged enclosure.
Span Span is otherwise identified as GAIN. It relates to the proportional output
to the sensor input. Calibration of the SGA is determined by setting the
Span (Gain) and Zero (Offset).
Span Adjustment The ability to adjust the gain of a process or strain meter so that a
specified display span in engineering units corresponds to a specified
signal span. For instance, a display span of 200°F may correspond to the
16 mA span of a 4-20 mA transmitter signal.
Stability The quality of an instrument or sensor to maintain a consistent output
when a constant input is applied.
Strain Gauge The strain gauge is a resistance bridge device where the bridge value
alters linearly and proportionally to the force exerted on it – be it
temperature, pressure, torque or load. The SGA is designed to convert
this change to a proportional electrical signal.
Torque Transducer The Torque Transducer is one of a series of STRAIN GAUGE sensors that
the SGA input is designed to accept. (Torque Sensor, Load Cell and
Temperature transducers).
Zero Zero is otherwise identified as Offset. It relates to the proportional
output to the sensor input. Calibration of the SGA is determined by
setting the Span (Gain) and Zero (Offset).
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
20
Page 22
Zero Adjustment The ability to adjust the display of a process or strain meter so that zero
on the display corresponds to a non-zero signal, such as 4 mA, 10 mA, or
1 V dc.
Zero Offset The difference expressed in degrees between true zero and an indication
given by a measuring instrument. See Zero Suppression
Zero Suppression The span of the SGA can be offset from zero (zero suppressed) such that
neither limit of the span will be zero. For example, an SGA which
measures a load of a 100kg span from 400kg to 500kg° is said to have
400kG zero suppression.
AC Alternating Current
DC Direct Current
Hz Hertz (Frequency)
IP66 UK Environmental Specification
kHz kilohertz (Frequency)
mA milliamps
mm millimetres
NEMA 4X US Environmental Specification
SC Signal Conditioner
SGA Strain Gauge Amplifier
V Volts
mV millivolts
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
21
Page 23
Chapter 8 Specifications for SGA/A & SGA/D Load Cell Amplifiers
Parameter Min Typical Max Units
Power supply (SGA/A):- (110/230Vac) 50 - 60Hz - 110/230 - V AC
Power supply dc: - 18 - 24 V DC (See note1)
Power supply current dc: - (depends on loading) 50 90 200 mA
Bridge excitation (10V range) 9.75 10 10.25 V (See note 2)
Bridge excitation (5V range) 4.85 5 5.15 V (See note 2)
Connections:
Field screw terminals - 2.5mm² rising clamp.
Enclosure:
ABS case 160 x 80 x 55 sealed to IP65 fitted with 3 off cable glands.
Controls:
Gain pot
Offset pot
Coarse gain switches
Coarse offset switches
Filter cut-off switches
Output mode switch
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
22
Page 24
CE Approvals
European EMC Directive
Low Voltage Directive
2004/108/EC
BS EN 61326-1:2006
BS EN 61326-2-3:2006
2006/95/EC
BS EN 61010-1:2001
Rated for Basic Insulation
Normal Condition
Pollution Degree 2
Permanently Connected
Insulation Category lll
Warranty
All SGA products from Mantracourt Electronics Ltd., ('Mantracourt') are warranted against defective material and
workmanship for a period of (3) three years from the date of dispatch.
If the 'Mantracourt' product you purchase appears to have a defect in material or workmanship or fails during normal
use within the period, please contact your Distributor, who will assist you in resolving the problem. If it is necessary
to return the product to 'Mantracourt' please include a note stating name, company, address, phone number and a
detailed description of the problem. Also, please indicate if it is a warranty repair.
The sender is responsible for shipping charges, freight insurance and proper packaging to prevent breakage in
transit.
'Mantracourt' warranty does not apply to defects resulting from action of the buyer such as mishandling, improper
interfacing, operation outside of design limits, improper repair or unauthorised modification.
No other warranties are expressed or implied. 'Mantracourt' specifically disclaims any implied warranties of
merchantability or fitness for a specific purpose. The remedies outlined above are the buyer’s only remedies.
'Mantracourt' will not be liable for direct, indirect, special, incidental or consequential damages whether based on
the contract, tort or other legal theory.
Any corrective maintenance required after the warranty period should be performed by 'Mantracourt' approved
personnel only.
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
23
Page 25
Figure 8.1 Connection Details
S
G
A
/
A
&
S
G
A
/
D
C
o
n
n
e
c
t
i
o
n
d
e
t
a
i
l
s
R
e
f
:
N
o
m
i
n
a
l
l
y
a
t
5
V
(
2
.
5
V
i
f
u
s
i
n
g
5
V
e
x
c
i
t
a
t
i
o
n
)
C
o
n
n
e
c
t
t
o
‘
S
t
r
a
i
n
I
n
p
u
t
+
’
&
‘
-
’
t
o
c
h
e
c
k
z
e
r
o
.
I
f
u
s
i
n
g
a
m
V
s
o
u
r
c
e
f
o
r
c
a
l
i
b
r
a
t
i
o
n
c
o
n
n
e
c
t
‘
R
e
f
’
t
o
‘
S
t
r
a
i
n
I
n
p
u
t
-
’
+
E
x
c
i
t
a
t
i
o
n
-
E
x
c
i
t
a
t
i
o
n
+
O
u
t
p
u
t
-
O
u
t
p
u
t
S
t
r
a
i
n
g
a
u
g
e
V
o
l
t
a
g
e
o
u
t
p
u
t
(
2
m
A
m
a
x
)
+
+
o
d
e
c
u
r
r
e
n
t
o
u
t
p
u
t
m
s
m
a
x
)
N
.
B
.
n
o
t
t
o
P
S
U
n
e
g
a
t
i
v
e
g
a
u
g
e
s
c
r
e
e
n
m
o
d
e
c
u
r
r
e
n
t
o
u
t
p
u
t
m
s
m
a
x
)
N
.
B
.
c
o
m
m
o
n
n
e
g
a
t
i
v
e
&
s
t
r
a
i
n
g
a
u
g
e
-
Strain Inpu
t
+ Strain Inpu
t
{
DC power
+
DC powe
r
-
urc
e
-/
V out
-
V out
+
I sin
k
+
urce +/I sink
-
‘Sink’ m
(500 Oh
common
or strain
‘Source’
(500 Oh
to PSU
screen
+
Jp1
Jp2
I so
I so
P1
Jp3
Ref (2.5V/5V)
Screen (0V)
+
(5V/10V)
-
Strain Excite
Screen
J1
Sw8
Shunt cal.
On
5V Exc
8
Sw4
Sw4
1 2 3 4 5 6 7 8
1 2 3 4 5 6 7
Analogue Output
Analogue Output
7
7
6
6
Analogue Output
0 1 0 x x1=No Filter
1 2 3 4 5
0 0 0 x x1=No Filter
0 1 0 x x1=No Filter
0 1 1 x x1=No Filter
Sw4
1 1 1 x x1=No Filter
Sw41 2 3 4 5
+/- 10V 0 0 0 x x1=No Filter
+/- 5V
0-10V0 1 1 x x1=No Filter
0-5V1 1 1 x x1=No Filter
8
Sw3
7
5 6
4
Filter
3
1 2
1
0
1
0
0
1
0
1
Span (Gain) mV/V
x x x x x
x x x 0 01=No Filter
x x x 1 11=No Filter
x x x x x
x x x x x
Sw1
1 2 3 4 567 567 567 567
Filter in x x x x x
0-20mA x x x 0 01=No Filter
4-20mA x x x 1 11=No Filter
J3
L
N
Off
8
Shunt cal. resistor
Sw1
Span (Gain)
1 2 3 4 5 6 7
001 000 111 110
0 1 1 1 0.06 0.31 0.75 4.0
1 0 1 1 0.11 0.61 1.41 7.5
0 0 1 1 0.17 0.91 1.99 10.5
1 1 0 1 0.23 1.20 2.49 13.2
0 1 0 1 0.28 1.49 2.95 15.6
0 = Off
1 = On
Filter
Sw3
8
Sw2
7
5 6
4
3
Zero (Offset)
Zero
P2
1 0 0 1 0.34 1.78 3.35 17.8
0 0 0 1 0.39 2.07 3.72 19.7
1 1 1 0 0.44 2.35 4.05 21.5
0 1 1 0 0.50 2.63 4.36 23.1
40 20 10 5 2 1
-
1 2 3 4 5 6 7 8
1 2
+
%
Sw2
1 0 1 0 0.55 2.91 4.63 24.6
0 0 1 0 0.60 3.19 4.89 25.9
1 1 0 0 0.65 3.46 5.12 27.1
0 1 0 0 0.70 3.73 5.34 28.3
1 0 0 0 0.75 4.00 5.54 29.3
0 0 0 0 0.80 4.26 5.72 30.3
220-230V
SGA/A
L
Mantracourt Electronics Limited SGA/A & SGA/D User Manual
Sw3 1 2 3 4 5 6 7 8
1Hz 0 0 0 0 0 0 1 1
5Hz 1 0 0 0 0 1 1 1
10Hz 1 1 0 0 1 1 1 1
50Hz 1 1 1 1 1 1 1 1
100Hz 0 0 0 0 0 0 0 0
500Hz 1 0 0 0 0 1 0 0
1kHz 1 1 0 0 1 1 0 0
Sw3 1 2 3 4 5 6 7 8
1Hz 0 0 0 0 0 0 1 1
5Hz 1 0 0 0 0 1 1 1
10Hz 1 1 0 0 1 1 1 1
L
110-120V
N
J3
N
J3
50Hz 1 1 1 1 1 1 1 1
5kHz 1 1 1 1 1 1 0 0
100Hz 0 0 0 0 0 0 0 0
500Hz 1 0 0 0 0 1 0 0
1kHz 1 1 0 0 1 1 0 0
5kHz 1 1 1 1 1 1 0 0
24
Page 26
Other Mantracourt Products
REGIS TERED FI RM
www.mantracourt.co.uk
Signal Conditioning
SGA
STRAIN GAUGE AMPLIFIER
Connect up to 4 strain gauges
Proportional mA and/or Voltage output
Simple DIL switch configuration
Set Sensitivity and Low pass filter and output
Simple - Reliable - Rugged
LCA
IN-LINE INTELLIGENT
STRAIN GAUGE AMPLIFIER
2 Set Points
4 to 20 mA AND 0 to 10 V (isolated) outputs
RS 232/485 Communications port
On-Board easy to use Programmer
Auto tare – Auto calibrate – and much more
…..
Fieldbus Connectivity In Two Excellent Packages
DSC The Digital Strain Card DCell The ‘in-cell’ Digital Strain Puck
ICA
IN-CELL
STRAIN GAUGE AMPLIFIER
Proportional mA or Voltage output
Single strain gauge applications
small in size – Big on Specification
ADW15
72 mm DIN Module – Display & Controller
10 mm LED Display (Configurable)
Sensitivity from 0.5 mV to 200 mV/V
10 V @150 mA Excitation
Isolated I/O100mSec sample rate
Set Point Relays
4 to 20 mA Output
Programmable via keypad
Mount this package adjacent to the strain
gauge
Plug-in-and-go-sensor High accuracy
Integrate the electronics with the load cell,
remove the cost, space and bother of additional
electronics and have a direct output provided in
REAL ENGINEERING UNITS.
Accuracy (1 part in half a million)
C
In the interests of continued product development, Mantracourt Electronics Limited reserves the right to alter product specifications without prior notice.
DESIGNED & MANUFACTURED IN THE UK
Code No. 517-150 Issue 2.8 01.09.10
Mount this package directly into the strain gauge
pocket
A quantum leap in the quality of measurement.
Temperature compensated
Unwanted Signal noise filter
Sensor specific calibration
Elimination induced noise on signals
ISO 9001