Manley Laboratories Stereo Limcom, Variable Mu Limiter Owners manual

MANLEY
LABORATORIES, INC.
OWNER'S MANUAL
STEREO "VARIABLE-MU"
LIMITER / COMPRESSOR
MANLEY LABORATORIES, INC.
13880 MAGNOLIA AVE.
CHINO, CA. 91710 TEL: (909) 627-4256 FAX: (909) 628-2482
email: emanley@netcom.com
http://www.manleylabs.com
CONTENTS
SECTION PAGE
INTRODUCTION 3
MAINS CONNECTIONS 4
INSTALLATION 5
FRONT PANEL 6
REAR PANEL 7
EXAMPLE SETTINGS 8
FRONT PANEL ADJUSTMENTS 9
OPERATIONAL NOTES 10
TECHNICAL NOTES 11
MASTERING AND M/S MODELS 12
GAIN REDUCTION CURVES 13
SPECIFICATIONS 14
SERVICE ADJUSTMENTS 15
WARRANTY 16
WARRANTY REGISTRATION 17
BLANK TEMPLATE TO STORE SETTINGS 18
INTRODUCTION
THANK YOU!...
for choosing the Manley Laboratories Stereo Variable-MU Limiter Compressor . The heart of this limiter is the variable gain vacuum tube. In tube-talk MU refers to gain and only a few types of tubes were ever designed for variable-mu operation. This limiting circuitry was designed by David Manley over thirty years ago. Our current version of this unique piece combines that classic sound with the precise cleanliness of modern componentry. The circuit is entirely differential which tends to cancel distortion and hum. It can handle a widely varying range of input signals in a rapid­acting manner without introducing audible coloration common in many limiters. The Attack and Recovery (Release) times have been carefully chosen for todays applications. These properties make this a superb compressor for both individual tracks, mixes and final mastering.
GENERAL NOTES
LOCATION & VENTILATION
The Limiter Compressor Amplifier must be installed in a stable location with ample ventilation. It is recommended, if this unit is rack mounted, that you allow enough clearance on the top and bottom of the unit such that constant movement of air can flow through the ventilation holes. There are 8 tubes inside this unit so it does run quite hot. Good ventilation is definitely encouraged.
WATER & MOISTURE
As with any electrical equipment, this unit should not be used near water or moisture.
SPECIAL NOTES
Tubes may become loose during transit. Straighten and press down each tube before plugging this unit into the mains socket. Furthermore, don't touch the tubes after the unit has been switched on, the tubes become very hot during operation and should only be handled after the power has been turned off and the tubes have cooled.
WARNING!
TO PREVENT THE RISK OF ELECTRIC SHOCK
DO NOT OPEN THE CABINET
REFER SERVICING TO
QUALIFIED PERSONNEL
!
MAINS CONNECTIONS
TROUBLESHOOTING NOTE: Occasionally we have had units sent to us to service that had no problem. The
symtom was clipping only when the unit was in the processing chain. Remove the unit and distortion gone. However in the shop - no problem. The cause was a combination of the somewhat low input impedance of the Limiter/Compressor and the somewhat "wimpy" line driver of the preceding device. The solution - change the order of processing so that a more professional device fed the Limiter / Compressor. The input impedance is over 1.2 Kohm and most devices are supposed to drive 600 ohms
(which is twice as hard) but in this era of ICs and semi-pro gear claiming to be pro well.......
Your unit has been factory set to the correct mains voltage for your country. The voltage setting is marked on the serial badge, located on the rear panel. Check that this complies with your local supply.
Export units for certain markets have a moulded mains plug fitted to comply with local requirements. If your unit does not have a plug fitted the coloured wires should be connected to the appropriate plug terminals in accordance with the following code.
GREEN/YELLOW EARTH terminal BLUE NEUTRAL terminal BROWN LIVE terminal
As the colours of the wires in the mains lead may not correspond with the coloured marking identifying the terminals in your plug proceed as follows;
The wire which is coloured GREEN/YELLOW must be connected to the terminal in the plug which is marked by the letter E or by the safety earth symbol or coloured GREEN or GREEN and YELLOW.
The wire which is coloured BLUE must be connected to the terminal in the plug which is marked by the letter N or coloured BLACK.
The wire which is coloured BROWN must be connected to the terminal in the plug which is marked by the letter L or coloured RED.
DO NOT CONNECT/SWITCH ON THE MAINS SUPPLY UNTIL ALL OTHER CONNECTIONS HAVE BEEN MADE.
INSTALLATION
5
Please refer to the REAR PANEL section for a full layout of the rear panel of your Stereo limiter compressor.
1. Set the voltage change-over switch on the back of the unit to the voltage you are using.
2. Be certain that the unit has a 2 Amp SLO-BLO (120 Volt operation) or 1 Amp SLO-BLO (for 240/220 Volt use) fuse.
3. Connect the input signal(s) to the 3 pin female XLR's on the rear panel. The input XLR's are TRANSFORMER COUPLED balanced wired as follows:
PIN 1: GROUND PIN 2: (+) POSITIVE GOING PHASE PIN 3: ( - ) NEGATIVE GOING PHASE
4. Connect a 3 pin XLR cable to the male output XLR on the rear panel. The output XLR is balanced and the pinout is wired as follows:
PIN 1: GROUND PIN 2: (+) POSITIVE GOING PHASE PIN 3: ( - ) NEGATIVE GOING PHASE
The output impedance is 600 Ohms nominal.
5. Connect a standard IEC mains cable to the IEC mains socket on the rear panel.
6. Check that the power is switched off on the unit.
7. Connect the IEC mains cable to a 50/60 Hz AC source of the proper selected voltage.
8. Power up the unit and allow it enough time to stabilize before using (approx 15 to 30 minutes).
METER
ADJUST
ATTACK
OUTPUT
THRESHOLD
RECOVERY
COMPRESS
LIMIT
BALANCE
BYPASS
IN
FRONT PANEL
A IN / BYPASS Switched in bypass mode, (down position) all effects of the limiting circuitry are
bypassed and will not affect the audio signal. In the bypass mode , the audio signal passes from the input directly to the output.
B. COMP. / LIMIT Selects compression ( 1.5:1) or limiting (4:1) function. Compression is soft -knee,
Limit is a sharper knee. At greater than 12dB of limiting the ratio increases (up to 20:1 maximum)
C. RECOVERY Recovery times can be selected between
a. VERY SLOW 8 seconds b. SLOW 4 seconds c. MEDIUM .6 seconds d. FAST .4 second e. VERY FAST .2 second
D. THRESHOLD Continuously variable gain reduction threshold control.
Determines the necessary amplitude for compression or limiting to take effect. Most extreme effect is at the MIN (fully counter clockwise) position. The lower the threshold, the more that gets limited.
E. OUTPUT ATTENUATE Continuously variable attenuation of output signal leaving the amplifier circuitry.
Full attenuation occurs at the full counterclockwise position. This control is active only in IN mode.
F. ATTACK Continuously variable sensitivity of transient detection. Determines the necessary
length of transient to initiate gain reduction. Fast =25mS, Med =50mS, Slow =70mS. Fully counter clockwise at the slowest setting will prevent most percussive signals from causing limiting or compression.
G. DUAL INPUT Continuously variable attenuation of input signal entering the amplifier limiting
circuitry Full attenuation occurs at the fully counterclockwise position. This control is active only in IN mode.
I. LINK When switched to "SEP" left side is independent from the right side. When
switched to "LINK" both channels are reduced in gain by the same number of dB. This insures that the stereo image is unchanged. The METERs remain independent to help set THRESHOLDs.
I. POWER ON/OFF Power is supplied to the unit when switched UP. Meters will illuminate
H. GR METER Measures in decibels amount of reduction
6
8
10
12
MANLEY
LABORATORIES INC.
REDUCTION dB
IN
BYPASS BALANCE
4
2
ADJUST METER
0
MANLEYSTEREO "VARIABLE-MU" LIMITER COMPRESSOR
LIMIT
COMPRESS
RECOVERY
SLOW FAST
THRESHOLD
MIN MAX
OUTPUT
ATTACK
SLOW FAST
DUAL INPUT
SEP LINK
POWER
FASTSLOW
MAXMINFASTSLOW
A H B C D E F G H I E F C D B H A
10
12
MANLEY
LABORATORIES INC.
REDUCTION dB
6
8
4
2
0
Loading...
+ 12 hidden pages