Malvern Hydro 2000uP User Manual

Hydro 2000
µP
MAN0389-1.0 Hydro 2000uP.book Page i Friday, March 16, 2007 4:11 PM
Hydro 2000
User Manual
MAN0389 Issue 1.0 March 2007
µ
P
MAN0389-1.0 Hydro 2000uP.book Page ii Friday, March 16, 2007 4:11 PM
© Malvern Instruments Ltd. 2007
Malvern Instruments makes every effort to ensure that this document is correct. However, due to Malvern Instruments’ policy of continual product development we are unable to guarantee the accuracy of this, or any other document after the date of publication. We therefore disclaim all liability for any changes, errors or omissions after the date of publication. No reproduction or transmission of any part of this publication is allowed without the express written permission of Malvern Instruments Ltd.
Head office:
Malvern Instruments Ltd. Enigma Business Park, Grovewood Road, Malvern, Worcestershire WR14 1XZ United Kingdom.
Tel + [44] (0)1684-892456 Fax + [44] (0)1684-892789
Windows 2000 and XP are registered trademarks of the Microsoft Corporation.
Tygon is a registered trademark of Norton Co.
Perlast is a registered trademark of Precision Polymer Engineering.
Printed in England
MAN0389-1.0 Hydro 2000uP.book Page i Friday, March 16, 2007 4:11 PM
Table of Contents
Part 1 - Operator’s Guide
Introduction to this manual
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
Access to the dispersion unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Assumed information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Where to get help. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Hardware features
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
What the dispersion unit does . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
How the dispersion unit is controlled . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
How the dispersion unit works . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
Hardware features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
Software features
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
Making a measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Manually controlling the dispersion unit . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Writing an SOP for the dispersion unit . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Making a measurement
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
Measurement guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Instrument preparation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
Anaerobic fill . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
Measuring new samples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-10
Changing cell holder position
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-1
Hydro 2000µP Page i
MAN0389-1.0 Hydro 2000uP.book Page ii Friday, March 16, 2007 4:11 PM
Table of Contents Hydro 2000µP
Part 2 - Appendices
Specification
Specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-1
Chemical compatibility
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1
Regulatory statements
CE Declaration of Conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . C-2
Index
Page ii MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Part 1 ­Operator’s Guide
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Introduction to this manual
Introduction
1
This manual describes the operation of the Micro Precision Hydro 2000µP sample dispersion unit, one of the following:
Instrument Model number
Hydro 2000µP (with ultrasonics) AWA2003
Hydro 2000µP (without ultrasonics) AWA2004
This manual is a supplement to the following manuals:
Mastersizer 2000 User Manual (termed “the main User Manual”).
Mastersizer 2000 Essentials Manual (termed “the Essentials Manual”).
This dispersion unit manual focuses on some specific issues of the dispersion unit that are not covered by the above manuals. It aims to:
Describe what the dispersion unit is and explain in simple terms how it works.
Identify the physical features of the dispersion unit.
Describe the software and explain how to use the dispersion unit to make a
measurement on the system.
Warning!
The dispersion unit or the samples to be measured may be hazardous if misused. Users must read the Health and Safety information in the Essentials Manual before operating the system.
We recommend that users who have never operated a Malvern particle analyser read this manual fully before starting the first measurement.
Those who are more familiar with particle size analysers may wish to jump straight to Chapter 4 of the main User Manual which gives details on making measure-
Hydro 2000µP Page 1-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Chapter 1 Introduction to this manual
1
ments. However, the importance of sample preparation before measurement can­not be overstated so we recommend reading the chapter on sample preparation (Chapter 8 of the main User Manual) as a priority.
Access to the dispersion unit
Within this manual, reference is made to the various people who will have access to the dispersion unit. The following is a list of these people and their responsibilities:
Malvern personnel
Malvern personnel (service engineers, representatives, etc.) have full access to the dispersion unit and are authorised to perform all service procedures that may require the removal of the covers.
Supervisor
The supervisor is the person responsible for the management/safety of the disper­sion unit and of its operation. The supervisor is responsible for the training of the operators. The supervisor can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual.
Warning!
The supervisor/operator must never remove the covers of the instrument or dispersion unit. Removal of the covers by unauthorised personnel will invalidate the warranty of the dispersion unit.
Operator
An operator is a person trained in the use of the dispersion unit. The operator can perform all user maintenance routines identified in Chapter 4 of the Essentials Manual.
Warning!
Failure to follow this guideline could result in exposure to hazardous voltages.
Page 1-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 3 Friday, March 16, 2007 4:11 PM
Introduction to this manual Chapter 1
Assumed information
Naming convention
Within this manual:
The Mastersizer 2000 optical bench is referred to as “the optical bench” or “the
instrument”.
The Hydro 2000µP is referred to in full or as “the dispersion unit”.
The combination of the optical bench, one or more dispersion units and the
computer is referred to as “the system”.
Menu commands
Menu commands from the Malvern software are referred to in the form main menu-menu item. As an example, the command Configure-New SOP refers to
selecting the New SOP item in the Configure menu. The same rules apply for sub-menus of sub-menus, so that Tools-Options-Instrument Port refers to the Instrument Port item in the Options sub-menu, which itself is a sub-menu of the Tools menu. Menu commands are always shown in bold text.
Where to get help
Full details on where and how to obtain help can be found in Chapter 1 of the Mastersizer 2000 User Manual.
The Essentials Manual gives information on the following:
Site requirements.
Health and Safety.
Maintenance.
Installation (in case the system has to be moved after its initial installation by
Malvern Instruments personnel).
Hydro 2000µP Page 1-3
MAN0389-1.0 Hydro 2000uP.book Page 4 Friday, March 16, 2007 4:11 PM
Chapter 1 Introduction to this manual
Page 1-4 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Hardware features
Introduction
2
This chapter gives a simple overview of the dispersion unit. All key aspects of the dispersion unit’s operation are explained in very simple terms. This chapter covers:
What the dispersion unit does.
How the dispersion unit is controlled.
How the dispersion unit works.
The hardware features in detail.
Hydro 2000µP Page 2-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
2
What the dispersion unit does
The sole purpose of any sample dispersion unit is to prepare the sample then deliver it to the measurement zone of the optical bench so that it can be measured.
The Hydro 2000µP allows the Mastersizer to be used for particle-in-liquid particle sizing and has been specifically designed so that only small amounts of sample and suspending liquids (dispersants) are required.
The small volume of the Hydro 2000µP is ideal when using solvent dispersants or when samples and dispersants are either expensive or hazardous.
The materials used in the manufacture of the dispersion unit maximise the range of samples which can be measured.
Other features of the dispersion unit include:
Variable speed pump – allows a wide range of particle sizes and densities to
be suspended and circulated.
Ultrasonic system (optional) – allows particle agglomerates to be dispersed
and assists in removing bubbles from the system.
Warning!
When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used.
Temperature sensing – an internal sensor monitors the temperature of the
dispersant. The software can be configured so that measurements are only made between user specified temperature limits. This can be used to ensure that the instrument can only be used when it has reached the correct tempera­ture.
How the dispersion unit is controlled
The dispersion unit is controlled using a single software dialogue with sliders for the pump speed and ultrasonic power (Chapter 3 has details). The dialogue also has pump pause, drain valve and Anaerobic fill buttons.
When controlled through a Standard Operating Procedure (SOP), the software tells the user what value to set for each parameter.
Page 2-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 3 Friday, March 16, 2007 4:11 PM
Hardware features Chapter 2
How the dispersion unit works
The dispersion unit looks like this:
9
10
4
6
2
5
7
1
8
3
ill 5473
The dispersion unit comprises two main units:
The control unit which contains most of the control electronics.
The cell unit which contains the cell, the sample well, pump, ultrasonic probe
(optional) and the drain facilities.
The dispersant is injected into the cell via the dispersant inlet port using a syringe
. Once preliminary measurements (“backgrounds”) are made, the sample to be
measured is added to the well
. The pump agitates the dispersant to ensure that
the sample does not settle out and continually circulates the sample through the cell
and then back to the well.
Within the cell, the dispersant and sample flow between two glass windows windows are precisely mounted within the measurement zone of the optical bench where a laser within the sample scatter the laser light. This scattered laser light is measured by the optical bench and used to calculate the particle size distribution of the sample.
Hydro 2000µP Page 2-3
. The
is passed through the sample/dispersant mixture. The particles
MAN0389-1.0 Hydro 2000uP.book Page 4 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
The optional ultrasonic probe can be used to break up particle agglomerates and help to remove bubbles from the system.
Once the sample has been measured, an internal drain valve is operated and the dis­persant and sample are drained to an external beaker
.
The lid for the well
should always be fitted during a measurement. The lid is
locked in place by rotating the locking lever
All functions of the dispersion unit are controlled using the Malvern software.
Hardware features
This section describes the hardware in detail.
Main system components
The diagram below shows a typical installation for the Hydro 2000µP. Its main components are identified. These components are examined in more detail later in this section.
10
9
.
4
1
3
8
2
7
6
11
5
ill 5407
Power supply
The external DC Power Supply Unit (PSU) supplies all power requirements to the dispersion unit.
Page 2-4 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 5 Friday, March 16, 2007 4:11 PM
Hardware features Chapter 2
Warning!
The PSU used on the Hydro 2000µP is not designed for operation in wet conditions. Do not place it in areas that are prone to spillages.
Warning!
Only use the power supply and cables supplied by Malvern Instruments.
Control unit
The control unit contains most of the control electronics for the dispersion unit. Power and communications cables are connected to the control unit. A control cable is connected from the front of the control unit to the front of the cell unit.
Cell unit
The cell unit contains the cell, well, pump and drain facilities. The cell unit is designed to circulate the sample and dispersant mixture and pass it through the analysing laser beam of the optical bench.
Sample well
The well in the cell unit is where sample to be measured is added. The sample well and connecting pipes have a capacity of less than 20ml.
Dispersant input/syringe
The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. Always use a syringe with a maximum capacity of 20 to 25ml.
Drain beaker
Once the measurement is complete (or during a flush routine to clean the system) an internal drain valve can be operated. The contents of the cell will drain through the drain port into the drain beaker.
If the end of the drain pipe is under the surface of the waste liquid, the back pres­sure created may prevent the cell from draining correctly. Typically ensure that the end of the tube is approximately 3cm below the rim of the beaker (slide the beaker holder until the correct length is achieved).
Hydro 2000µP Page 2-5
MAN0389-1.0 Hydro 2000uP.book Page 6 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
Warning!
If using hazardous dispersants or samples, do not allow the beaker to become over-full. If dispersants or samples are noxious, the instrument should ideally be used in a fume cupboard.
ill 5643
Drain beaker holder
The drain beaker may hold hazardous samples and dispersants. The drain beaker holder secures the drain beaker to prevent it from being accidentally knocked over when the cell unit is removed from the optical bench.
We recommend that the beaker holder is always used.
Cell unit holder
The cell unit holder is a “parking” area for the cell when it is not in use. This is par­ticularly useful when more than one dispersion unit is in use. It allows the cell to be removed and stored without the need to disconnect it from the dispersion unit. When the cell is “parked” for a short period make sure that the cell is full of disper­sant. If it is to be parked for longer periods, the unit should be drained and the cell windows dried. The Essentials manual describes these procedures.
Well lid
The well lid is used to reduce the escape of hazardous fumes when measuring using hazardous samples and dispersants. It also reduces the risk of accidental splashing from the well.
Always use the well lid when making measurements or moving the cell unit. Always lock the lid into place with the lid locking lever
.
Lid locking lever
The lid locking lever is used to lock the well lid in place when measurements are taking place.
Page 2-6 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 7 Friday, March 16, 2007 4:11 PM
Hardware features Chapter 2
Dispersant valve
The dispersant valve is used to prevent sample or dispersant leaking when the dis­persant input syringe is removed.
Two types of valve are available:
Aqueous valve – this is disposable and can be used with non aggressive dis-
persants such as water. The valve is easily identifiable as it is made of transpar­ent, purple plastic.
Non-aqueous valve – suitable for use with a large range of chemically aggres-
sive dispersants. The non-aqueous valve is white and has a tap on its side.
Replacement valves are available from a Malvern representative.
Controller unit
This section examines the controller unit in more detail. As stated above, the con­troller unit contains most of the control electronics for the dispersion unit.
Rear panel
The rear panel of the controller unit is shown below and its main features identi­fied.
DRAIN OPEN/CLOSED
CAUTION
Flash tests can
permanently damage
this equipment
WARNING :
To prevent electric shock do not remove screws. No user s erviceabl e parts ins ide. Refer servicing to Malvern trained service personnel.
DC IN 24V
2.9A
4
1
DC power input
Power input from the external power supply.
Hydro 2000µP Page 2-7
ACCESSORY COMMS
IN OUT
2
3
ill 5409
MAN0389-1.0 Hydro 2000uP.book Page 8 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
Warning!
The PSU is not designed for operation in wet conditions. Do not place it in areas prone to spillages.
Warning!
Only use the PSU and cables supplied by Malvern Instruments.
Accessory comms “in” connector
The communications cable from the optical bench or from another dispersion unit connects here.
Accessory comms “out” connector
If more than one dispersion unit is connected to the system and this is the first dis­persion unit in line, a communication cable will be connected from this connector to the Accessory comms - in connector on the second dispersion unit.
Caution!
The termination plug supplied with the optical bench must be fitted here if this is the only dispersion unit connected or it is the last dispersion unit in the line.
Manual drain
The manual drain button can be pressed to drain the well. Users may wish to do this if, for example, they need to drain the dispersion unit without having to start the control software. Pressing the button once (for about half a second) will open the drain valve. Pressing the button a second time will close the drain valve. If the drain valve is left open for more than 15 minutes it will automatically close.
Page 2-8 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 9 Friday, March 16, 2007 4:11 PM
Hardware features Chapter 2
Front panel
The front panel of the controller unit is shown below and its main features identi­fied.
1
2
CONTROL ULTRASOUND
3
4
HYDRO
ill 5410
Status indicator
The status indicator illuminates in one of three colours when the dispersion unit is powered up:
Green if the dispersion unit is functioning correctly and the cell unit has been
loaded into the optical bench (i.e. the dispersion unit is “active”).
Amber if the dispersion unit is functioning correctly but the cell has not been
loaded into the optical bench (i.e. the dispersion unit is at “standby” and is parked in the cell unit holder).
Red if the software has detected a error. If an error message does not appear on
the screen, selecting Configure-Accessories displays the accessory control dialogue which should show the type of error.
Power on/off
This is the main power switch for the dispersion unit.
Control cable
The control cable carries all power and communication signals between the control unit and the cell unit (with the exception of ultrasonics).
Hydro 2000µP Page 2-9
MAN0389-1.0 Hydro 2000uP.book Page 10 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
Ultrasonic control cable (optional)
If the ultrasonic option is fitted, this cable will carry power and control signals for the ultrasonic probe between the control unit and the cell unit.
Warning!
The Ultrasonic control cable carries high voltages. Under normal operat­ing procedures this is perfectly safe.
Inspect the cable at regular intervals for any sign of damage. Never operate the equipment if the cable is damaged.
Cell unit
The cell unit is shown below with its main features identified.
10
8
2
9
1
5
4
3
6
7
11
Well cone
The well cone is where the sample is added to the cell unit. Typically, dispersant is added by filling the well to within 1 to 2mm of the top of the well. Perform an
Page 2-10 MAN 0389
ill 5495
MAN0389-1.0 Hydro 2000uP.book Page 11 Friday, March 16, 2007 4:11 PM
Hardware features Chapter 2
anaerobic fill (to remove bubbles) and then remove dispersant directly from the well using a pipette so that the well is about a quarter full.
Overflow
If the well is overfilled any excess is directed to the drain port via the overflow.
Control cable
The control cable carries all power and communication signals between the control unit and the cell unit (with the exception of ultrasonics).
Ultrasonic control cable (optional)
If the ultrasonic option is fitted this cable carries control signals for the ultrasonic transducer between the control unit and the cell unit.
Dispersant inlet port
The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. To avoid introducing bubbles into the system, use a slow, steady pressure when injecting the dispersant; see Chapter 2 for details.
Warning!
Always consult the Materials Safety Data Sheet for any handling and safety information for all dispersants used. Only use syringes with a maximum capacity of 20-25ml.
The dispersant for suspending the sample and flushing the system is injected into this port using a syringe. To avoid introducing bubbles into the system, use a slow, steady pressure when injecting the dispersant; see Chapter 2 for details.
Fill the well until dispersant is visible within the well cone.
Dispersant valve
The dispersant valve must be used on the dispersant inlet to prevent sample/disper­sant from leaking when the dispersant syringe is removed.
Two types of valve are available, as described earlier. Replacement valves are availa­ble from a Malvern representative.
Drain port
When the drain valve is operated, the contents of the cell will flow out through this port. A tube will guide the contents to a waste beaker. Always ensure that the waste beaker has enough capacity before draining the cell.
Well lid
The well lid should always be in place when making measurements (except when adding sample). The lid reduces the amount of vapour released when using sol­vents as a dispersant. It also stops dispersant from welling up and stops air being sucked in.
Hydro 2000µP Page 2-11
MAN0389-1.0 Hydro 2000uP.book Page 12 Friday, March 16, 2007 4:11 PM
Chapter 2 Hardware features
Additionally, when measurements are performed using solvents, users may occa­sionally find that the background levels are very high. This is due to temperature gradients within the solvent. The normal cure is to wait for several minutes for thermal equilibration, or even to warm the solvent beforehand. In many cases, a much simpler and quicker solution is to ensure that the lid is in place.
Lid locking lever
The lid locking lever is used to lock the well lid into place during measurements.
Cell windows
The cell windows allow the analyser beam of the optical bench to pass through the cell and hence the particle field. The cell windows can be removed to allow clean­ing or replacement.
Caution!
The cell windows are delicate and should be treated with caution; refer to the Essentials Manual for details on cleaning the windows.
Window ring lock screws
The windows can be removed for cleaning. The window ring lock screws can be loosened to allow the window ring and window to be removed; refer to the Essen- tials Manual for details.
Page 2-12 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Software features
Introduction
3
This chapter describes all the physical features of the dispersion unit. It covers:
Making a measurement – the basics of making a measurement using the dis-
persion unit.
Manually controlling the dispersion unit.
Writing a Standard Operating Procedure (SOP) for the dispersion unit.
Software features – the features of the Malvern software that are specific to the
dispersion unit.
Hydro 2000µP Page 3-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Chapter 3 Software features
3
Making a measurement
The dispersion unit can either be controlled manually or automatically by running a Standard Operating Procedure (SOP). Making a measurement is fully docu­mented in Chapter 4 of the main User manual.
The system automatically detects which dispersion unit and cell is connected. If more than one dispersion unit is connected, the system detects all dispersion units connected, but only the dispersion unit that has its cell installed on the optical bench will be “active”.
Manually controlling the dispersion unit
To control the dispersion unit manually either select Configure-Accessory or, when making a manual measurement, click the Accessory button in the Meas- urement Display window. This accessory control dialogue appears:
7
1
8
2
4
3
6
This dialogue is specific to the Hydro 2000µP. It has separate Control and Errors tabs. Use it to alter the pump speed and (optionally) the ultrasonic power by mov­ing the relevant slider bar.
5
ill 7776
Page 3-2 MAN 0389
i
MAN0389-1.0 Hydro 2000uP.book Page 3 Friday, March 16, 2007 4:11 PM
Software features Chapter 3
It features the following:
Pump Speed control
The speed of the pump (which controls the speed at which the sample is circu­lated) is altered by moving the slider. Use the up and down arrows under the slider for fine adjustment of the pump speed.
Ultrasound control (optional)
Warning!
When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used.
The power of the ultrasonic probe is altered by moving the slider. Use the up and down arrows under the slider for fine adjustment of the ultrasonic power.
Dispersant temperature
The temperature of the dispersant within the cell is continually monitored and dis­played here. Temperature monitoring can be used in conjunction with the Sam- pler Settings dialogue to limit the range of temperatures at which measurements can be made.
Pause Pump button
Pressing the Pause Pump button stops the pump motor. Pressing the button a second time restarts the pump. The indicator to the right of the button shows red if the pump is paused, and green when the pump is running.
Note
If having difficulty with bubbles in the system, they can often be removed by turning the pump on and off in quick succession using Pause Pump.
Drain Valve button
Pressing the Drain Valve button will open the drain valve, allowing the contents of the cell to be emptied into the waste beaker. The indicator to the right of the button will show red if the drain is open and green when the drain is closed.
Anaerobic fill button
Due to the small volume of the Hydro 2000µP and associated problems of remov­ing bubbles from the system, a special fill sequence called Anaerobic fill can be used. Chapter 4 gives details.
Hydro 2000µP Page 3-3
MAN0389-1.0 Hydro 2000uP.book Page 4 Friday, March 16, 2007 4:11 PM
Chapter 3 Software features
Error indicators
The following indicators on the Errors tab illuminate if an error condition has been detected.
Connection box disconnected – shows red if the control cable from the
control unit to the cell unit is not connected.
Dispersant Temperature Probe – shows red if the temperature is outside
the range 10°C - 35°C or there is an internal error with the temperature sensor. Contact a Malvern representative if this error condition persists.
Drain Valve – illuminates if the drain valve has failed to drive (i.e. it has stuck
open or closed). This may be caused by leaving a sample in the cell for long periods - this may have set and bonded the drain valve. Never leave the sample in the cell for long periods - always flush the cell with clean dispersant. Contact a Malvern representative if this error indicator illuminates.
Pump – illuminates if there is a problem with the pump controller electronics
Pump over temperature – illuminates if the pump is over its operating tem-
perature range. The pump will stop until its temperature returns to the operat­ing range. The pump will typically overheat if the sample dispersant is too viscous or the sample material is too large, causing jamming.
Pump speed control – illuminates if the difference between the speed
demanded and the actual speed exceeds 500rpm for over 10 seconds.
Ultrasound – illuminates if there is an error with the ultrasound control elec-
tronics.
Ultrasound displacement – illuminates if the current supplied to drive the
ultrasound probe is too high.
Ultrasound timeout – illuminates if the ultrasound is left on for over 15
minutes.
Status indicators
These indicators give the current status of the pump and ultrasonic probe. If this indicator shows green then the pump or ultrasonic probe are currently in use. If the Ultrasonics indicator is greyed out then the unit does not have the ultrasonic option installed.
Page 3-4 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 5 Friday, March 16, 2007 4:11 PM
Software features Chapter 3
Writing an SOP for the dispersion unit
A Standard Operating Procedure (SOP) can be configured to control all settings for the dispersion unit automatically. When an SOP is run the software (depending on how the SOP is set up) automatically runs a measurement, giving instructions to the operator such as when to add dispersant, sample, etc. The SOP even reports if the wrong cell unit is fitted.
The SOP wizard is described in full in the main User Manual. Here we just describe those dialogues which are specific to the Hydro 2000µP. These are the Sampler Selection and Sampler Settings dialogues.
Sampler Selection dialogue
The Sampler Selection dialogue is shown below:
It features:
Sample handling unit list
Select the dispersion unit to be used from the drop down list
. All dispersion
units available will be listed, even those that are not attached to the system. This allows an SOP to be written remotely from the system.
All systems the SOP is distributed to must have the dispersion unit connected.
1
2
ill 7774
The system allows up to two of each dispersion unit to be connected to the optical bench. In this situation, the dispersion units will be identified as unit A and B - in
Hydro 2000µP Page 3-5
MAN0389-1.0 Hydro 2000uP.book Page 6 Friday, March 16, 2007 4:11 PM
Chapter 3 Software features
the case of this dispersion unit there would be an entry for Hydro 2000µP (A) and Hydro 2000µP (B). If using just one dispersion unit, select the “A” option.
Sample handling unit options
This dispersion unit is supplied in two versions, with ultrasonics and without ultra­sonics. Use this pull down list to select the model in use.
If the user tries to run an SOP that has been defined for an dispersion unit with ultrasonics on an dispersion unit without that option, a message is displayed, telling them to connect the correct dispersion unit.
Sampler Settings dialogue
The Sampler Settings dialogue is shown below:
1
2
3
4
ill 7775
Its features are:
Pump speed
The pump speed is altered by moving the slider or entering the required speed in the RPM box to the right of the slider.
Selecting the Advanced button displays the Vari-flow options. Some materials such as metal flakes have been found to form structures when they are circulated through the cell at normal pump speeds. These structures give rise to a secondary diffraction pattern that is incorrectly interpreted as a small peak at 1000 microns.
Vari-flow allows these materials to be dispersed in the well in the normal way and then the pump is briefly turned off or reduced before measurement of the sample. When the pump is turned off, the structures in the sample are randomised and the
Page 3-6 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 7 Friday, March 16, 2007 4:11 PM
Software features Chapter 3
measurement can be made without the occurrence of the secondary diffraction pat­tern.
When the flake material is suspended in a volatile solvent, such as white spirit, vari­flow is even more useful since it assists the rapid thermal stabilisation of the optics. When the sample well is filled with fresh solvent, very high backgrounds can often be observed. These arise from thermal gradients in the solvent causing steering of the laser beam onto the inner rings of the internal detector. As the temperature equilibrates, the background will fall to normal levels (80-100) and the measure­ment can proceed. Pausing the pump after the solvent has had a chance to thor­oughly disperse through the system will accelerate the fall of the background to normal levels.
Ultrasonics
Warning!
When using solvents, especially those with low ignition points, minimise the use of ultrasonics. There is a remote possibility that localised heating caused by the ultrasonic action may cause an explosive reaction. Check the Material Safety Data Sheets for all substances used.
The ultrasonics radio buttons enable the ultrasonic transducer. The options are:
None – no ultrasound.
Continuous (from sample addition) – ultrasound is activated after sample
is added and will run continuously.
Pre-measurement – ultrasound is activated for a set time prior to measure-
ment. For pre-measurement mode, type in the required duration in seconds in the Duration box.
Continuous (from start) – ultrasound is active throughout the measure-
ment. Ultrasound will start after the electrical background is complete. The stabilising period will add a delay between the electrical background and the optical alignment. During this delay ultrasound will be active to allow bubbles to be driven from the dispersant.
Either click and drag the slider bar to set the level, or type the exact level required in the box to the right of the slider bar.
Determine the optimum ultrasound level and duration using manual control, ensuring that the lowest setting giving satisfactory dispersion is obtained. Record the settings and transfer these to the SOP in the Sampler Settings dialogue using the slider controls.
Hydro 2000µP Page 3-7
MAN0389-1.0 Hydro 2000uP.book Page 8 Friday, March 16, 2007 4:11 PM
Chapter 3 Software features
Pre-Measurement Delay
A pre-measurement delay can be added before the measurement to allow thermal equilibration to occur after ultrasound has been applied. If this is not done the anal­ysis may show the presence of large particles which are not in the sample. This is caused by the refractive index differences observed in the sample cell when the dis­persant is not equilibrated. Choose a delay sufficiently long enough for adequate thermal equilibration to occur.
To set the delay, highlight either the minutes or seconds in the entry box, then use the selection arrows to alter the time; alternatively the time can be typed in. A delay of between 0 and 59.59 minutes can be chosen.
Temperature Monitoring
It is possible to monitor the temperature of the sample and dispersant and to set the software to only make measurements between specified limits.
Temperature monitoring is enabled by selecting the Allow measurements only…. check box.
The options are:
Lower/upper limit – this section specifies the temperature range within
which the sample/dispersant must be for the measurement to start. This limit must be in the range of 10° to 35°C.
Stabilising temperature – the time for which the temperature must be stable
before the measurement proceeds.
Time-out period – the measurement will be abandoned if the temperature is
not stabilised within this time period.
Page 3-8 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Making a measurement
Introduction
4
This chapter describes how to make a measurement using the Hydro 2000µP. It covers:
Measurement guidelines.
Instrument preparation.
Anaerobic fill.
Measuring new samples.
Hydro 2000µP Page 4-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
4
Measurement guidelines
The Hydro 2000µP uses very low volumes (approximately 20ml) of dispersant and sample. When using such small quantities, take the following precautions when making a measurement:
Minimise the amount of sample lost during sample preparation, or through
sample adhering to the sides of the well.
Cleanliness of the system is critical: very small amounts of cross-contamination
can bias the results.
Manage the pump speed to prevent the introduction of bubbles, while also
ensuring that the sample is kept in suspension. This is true for all wet sample dispersion units but particularly important with the small dispersant volumes of the Hydro 2000µP.
The basic procedure
Add dispersant to the well by injecting the requested dispersant through the inlet port. Inject the dispersant slowly until the level of the dispersant is within 1 to 2mm of the internal rim of the well.
Because the stirring and the circulation of the sample is carried out in the chamber below the well cone, keep the level of dispersant in the well low. If sample is seen adhering to the sides of the well cone, use a pipette to withdraw some of the disper­sant from the well to rinse the sides of the well. Once the sample has been added, it is important to replace the well lid and lock it in place using the locking lever.
Minimising sample loss
Sample loss by adhesion
To prevent sample adhering to the side of the sample well during sample addition, it is typical to have the dispersant at a level a quarter of the way up the well. This allows the sample to be taken into the well cavity by the dispersant.
A typical procedure is:
1. Fill the well to 1-2mm below the top of the well.
2. Run an anaerobic fill sequence (press the Anaerobic fill button on the acces-
sory control dialogue).
3. Once complete, use a pipette to remove dispersant so that the well is about a
quarter full.
Page 4-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 3 Friday, March 16, 2007 4:11 PM
Making a measurement Chapter 4
Sample loss during sample preparation
Choose a sample preparation method that ensures that all particles are transferred to the well. A recommended method is to prepare a slurry in a small container and add this to the well using a pipette. Add as a maximum 2ml of sample volume, but 1ml is preferred.
Material will usually be seen collecting in the bottom of the container.
To take a representative sample of the material:
1. Make the slurry more dilute than is normal for larger volume sample disper-
sion units. This will ensure that adding slurry does not lead to an excessive concentration.
2. Tilt the container at an angle.
3. Agitate the dispersant by continually filling and discharging the pipette to
ensure the material is suspended in the dispersant (rather than sinking to the bottom).
4. Once dispersed, finally fill the pipette with as much of the dispersant as possi-
ble and transfer it to the well drop by drop, pausing between drops to allow the sample to fully disperse i.e. monitor the obscuration value until it stabilises.
5. When adding the sample, ensure that it is added directly to the dispersant
rather than trickling it down the sides of the well walls. This prevents sample loss through adhesion.
Cleanliness
Cleanliness is essential; since only 10-50mg of sample is added, it only takes a small amount of cross-contamination from a previous measurement to bias results signif­icantly. Consider the following points when making measurements:
When cleaning, rinse the system three or four times. It doesn’t use much dis-
persant, and gives better performance for the next sample.
The first rinse of a clean is sometimes best done by injecting 5-10ml bursts of
dispersant into the well while the pump is running and the drain valve open. This tends to rinse larger particles through the system.
When changing dispersants, blow each dispersant out of the fill tube with a
syringe full of air (with the drain open) to reduce cross contamination.
When changing over from an oily organic solvent (vegetable oil, etc.) to water,
use a biological drain cleaner (containing food enzymes) to clean out the last traces of oil.
Hydro 2000µP Page 4-3
MAN0389-1.0 Hydro 2000uP.book Page 4 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
Bubbles and pump speeds
Control the pump speed to prevent the introduction of bubbles while ensuring that the sample is kept in suspension. This is true for all wet dispersion units but partic­ularly important with the small dispersant volumes of the Hydro 2000µP. The Mastersizer sees bubbles as particles and measures these. Bubbles will vary in size but are typically in the region of 100 microns. In many cases bubbles can be clearly seen as a separate peak when the measurement data is analysed.
Consider the following points when making measurements:
When filling the system with dispersant, turn off the pump to prevent it intro-
ducing bubbles.
Fill the system with dispersant slowly to avoid trapping bubbles.
When the system is first filled, bring the pump speed up slowly to clear any
remaining bubbles, without breaking them into smaller ones. Ideally use the Anaerobic fill sequence (described later) to remove air and bubbles from the system.
Take care with the pump speed. The system is designed, and balanced, for
water. Other dispersants may tend to well up in the well or pump chamber as the pump speed increases. Full speed (5000 rpm) is too much for most disper­sants, so do not use 100% pump by default. See the table in the Anaerobic fill section later for examples of settings.
Choose a pump speed high enough to keep the particles in suspension. Using
too high a pump speed will tend to draw air into the system.
Always fit the well lid during measurements.
Finally, on the last wash before filling the system to make a measurement, we
recommend that the pump is set to off whilst draining. This is because the action of running the pump whilst draining can cause the formation of small air bubbles in drops of dispersant that may remain in the system after draining. If the system is then filled immediately in order to perform a measurement, these air bubbles are circulated and sized in addition to the required sample.
Page 4-4 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 5 Friday, March 16, 2007 4:11 PM
Making a measurement Chapter 4
Instrument preparation
Before starting a measurement, ensure that the correct cell unit is installed in the cell area and then power up the system. The stages are outlined below.
Powering up
Power up the system by:
Switching on the dispersion unit(s).
Switching on the optical bench.
Switching on the computer and starting the software by double-clicking on the
Mastersizer 2000 icon.
Filling the cell with clean dispersant
Before a measurement fill the cell with clean, degassed dispersant (when the meas­urement is run the software asks if this has been done). Take care not to let air (either dissolved gasses or bubbles) into the system. The following procedure is recommended to fill the cell unit.
Degassing the dispersant
Gases dissolved in the dispersant can form bubbles in the cell. To prevent this from happening, we recommend that the dispersant is degassed prior to adding it to the cell unit. The following procedure is suggested if water is used as a dispersant.
Warning!
This procedure may not be suitable for all dispersants, especially disper­sants with low boiling/ignition points.
To de g a s w a t er :
1. Choose a vessel that has a lid and fill the vessel with water. Do not fill the vessel
completely at this point.
2. Loosely fit the lid, but do not tighten it. This ensures that pressure does not
build up in the vessel, while preventing dust getting in.
3. Place the open vessel in an ultrasonic bath that can be heated.
4. Set the ultrasonic power to 60W and the water bath to 70°C.
5. Degas for 20 minutes.
Hydro 2000µP Page 4-5
MAN0389-1.0 Hydro 2000uP.book Page 6 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
6. Once the water has been degassed, a number of bubbles may have formed on
the inside surface of the vessel. To remove these, fit the lid to the vessel and very carefully turn the vessel on its side (see the diagram below). Slowly rotate the bottle through one complete revolution - the bubbles will be removed as they meet the air/fluid interface.
Vessel
Heater
Ultrasonic bath
ill 5638
Filling the syringe
Take care to avoid air entering the syringe when filling it with dispersant.
Page 4-6 MAN 0389
Bubbles
ill 5639
Warning!
Avoid using a syringe with a capacity of more than 20ml. The additional capacity may cause the cell to overflow, causing hazardous conditions within the cell unit.
MAN0389-1.0 Hydro 2000uP.book Page 7 Friday, March 16, 2007 4:11 PM
Making a measurement Chapter 4
To fill the syringe:
1. Insert the syringe into the vessel of dispersant.
2. Draw back the syringe very slowly. Pulling back too quickly will cause “boil-
ing” on the surface of the syringe plunger (see the diagram below) or air may be dragged back through the syringe seal.
Syringe
Syringe plunger
“Boiling”
Syringe body
ill 5640
3. Hold the syringe with the tip pointing upwards, and push the plunger in to dis-
place any air (as shown below). It may be necessary to tap the body of the syringe to displace any air bubbles.
i
Hydro 2000µP Page 4-7
ill 5641
Note
Avoid having to refill the syringe half way through filling the cell, as dis­connecting and reconnecting the syringe may introduce air into the cell. When filling the standard 20ml syringe, fully draw back the plunger to take up enough dispersant to fill the cell (its approximate capacity is 18ml).
MAN0389-1.0 Hydro 2000uP.book Page 8 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
Filling the cell
Finally, fill the cell with the clean dispersant.
To fill the cell:
1. Connect the syringe to the valve attached to the front of the cell unit.
2. Hold the syringe with the plunger facing upwards (as shown below) so that any
bubbles in the syringe rise to the top and are not injected into the cell.
Valve
Pipe to cell unit
3. Filling the cell now depends on which of the two types of dispersant valve
being used:
For the non-aqueous valve (suitable for use with a large range of chemi-
cally aggressive dispersants) open the tap on the side of the valve. Slowly press the plunger of the syringe (typically over about a 20 second period). Fill the cell until the sample well is almost full (to within 1 or 2mm of the internal well rim. Once full, close the tap on the side of the valve. Note that there is no need to remove the syringe from the valve.
The aqueous valve (used with non aggressive dispersants such as water)
does not have a tap to control the flow through the valve but acts as a non­return valve. To fill the cell, slowly press the plunger of the syringe (typi­cally over about 20 seconds). Fill the cell until the sample well is almost full (to within 1 or 2mm of the internal well rim).
Note
For the valve to function correctly as a non-return valve, the syringe must
i
be removed from the valve once the cell is full.
ill 5642
Page 4-8 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 9 Friday, March 16, 2007 4:11 PM
Making a measurement Chapter 4
Anaerobic fill
Due to the small volume of the Hydro 2000µP and associated problems of remov­ing bubbles from the system, a special fill sequence called Anaerobic fill can be used. During an anaerobic fill the pump speed is increased in steps to a target speed. A short pause is added between steps to allow bubbles to disperse.
To use anaerobic fill click the Anaerobic fill button on the accessory control dia­logue’s Control tab. The Anaerobic fill dialogue will appear. This contains a field for setting the final pump speed. This is the speed the pump will run at after the fill process has finished. During a manual measurement the last value used will be recalled. During an SOP this will be the pump speed specified in the SOP.
For best results set the pump speed field 500rpm above the final speed required. Once the Anaerobic fill sequence is complete the final pump speed can be reset. Example values are given below.
Material Pump speed (SOP) Anaerobic fill
Glass Beads in:
Water
Oils
Alcohols
Latex in water 2000rpm 2500rpm
2500rpm
3000rpm
2500rpm
3000rpm
3500rpm
3000rpm
The dialogue prompts the user to fill the well. Once it’s full press the Go button. As bubbles escape the level of the dispersant may drop.
Once complete, ensure that the well is approximately a quarter full, using a pipette to remove excess dispersant if necessary.
Hydro 2000µP Page 4-9
MAN0389-1.0 Hydro 2000uP.book Page 10 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
Measuring new samples
The above procedures are ideal if the SOP is set up or all the parameters for a man­ual measurement have been specified. Users measuring a new material must either create or modify an SOP or set up the options for a new manual measurement. This section outlines the setup for measuring a new sample.
Measuring new samples - manual measurement
The basic procedure is the same as outlined above. When the user selects the Options button in the measurement display, the Measurement Options dia­logue appears. This dialogue controls all aspects of the measurement.
Use the Materials tab to select a different material and dispersant or define the
optical properties of the new material. This tab also allows users to specify options regarding the result calculation.
Note
If a material is created or modified the computer calculates a new scattering
i
model. This usually takes between one and 10 minutes, depending on the power of the computer. However, if the ratio of the refractive index of the sample to the dispersant is in the range 0.99 to 1.01, then this time may substantially increase. Under these conditions we recommend that the model is generated overnight.
Use the Measurement tab to set up measurement times, set measurement
alarms, etc.
Press the Document button in the Measurement display to open the Document dialogue. Add relevant details that will allow a user to reproduce the measurement at a later date.
All the options available are basically the same when setting up an SOP. For more details of these options see Chapter 4 of the main User Manual.
Measuring new samples - SOP measurements
Creating an SOP is easy. Select Configure-New SOP; the software will run through an SOP Wizard. This wizard asks a series of questions, such as:
what are the optical properties of the sample?
what is it suspended in?
how fast should the pump be?
how much ultrasonic dispersion is required?
do any additives need to be added to the sample to aid dispersion?
Page 4-10 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 11 Friday, March 16, 2007 4:11 PM
Making a measurement Chapter 4
how will the result be calculated and displayed?
These are all the questions that need to be answered in order to define a measure­ment.
Anyone can create an SOP but typically it is the responsibility of the system admin­istrator. An SOP can be created by a specialist and then distributed to all Master­sizer 2000 users in an organisation.
Remember that no instrument needs to be connected to the software to write an SOP. The Malvern software can be installed on a remote computer and SOPs cre­ated and edited before they are tried out on an instrument. (The Essentials Man- ual has details on installing the software on another computer.)
A completely new SOP can be created by selecting Configure-New SOP and fol­lowing the SOP Wizard. An existing SOP can be easily edited either to correct it or to create a new SOP based on it. If several slightly varying SOPs are being saved, it is useful to include the differentiating features in the SOP name e.g. “glopalene ultrasound 60”, “glopalene ultrasound 40”.
To edit an SOP select Configure-Existing SOP…. A tabbed dialogue will appear. Each tab corresponds to a screen from the SOP Wizard. Select a tab and change any of the entries as required. Click OK to save the changes.
To create a new SOP based on an existing one, first modify the existing SOP using Configure-Existing SOP. Once all the changes have been made, the Save-As dialogue will appear. Save the SOP under a different name.
Hydro 2000µP Page 4-11
MAN0389-1.0 Hydro 2000uP.book Page 12 Friday, March 16, 2007 4:11 PM
Chapter 4 Making a measurement
Page 4-12 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Changing cell holder position
Introduction
5
The Essentials Manual covers most maintenance procedures. This chapter describes one non-essential procedure, how to move the cell holder from its posi­tion on the right of the optical bench, should this be necessary.
The Hydro 2000µP is typically configured to be positioned on the right of the opti­cal bench optical bench the control unit.
. For users who wish to locate the dispersion unit on the left of the
we recommend that the cell holder be swapped to the other side of
Hydro 2000µP Page 5-1
1
2
ill 5433
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Chapter 5 Changing cell holder position
5
To swap over the cell holder:
1. Disconnect all connectors from the control unit.
2. Carefully tip the control unit and the cell holder onto its back to expose the feet
and connection plate (as shown on the illustration below).
3. Using a 4mm Allen key, remove the eight screws securing the connection
plate.
4. Swap the cell holder to the other side of the control unit.
5. Flip the connection plate over and screw it back into position.
6. Unscrew the four feet and move them to the alternative locations (now
located at the outside edge of the assembly).
4
1
4
1
4
1
2
3
3
1
ill 5434
Page 5-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Part 2 ­Appendices
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Specification
Specification
A
Dispersion type Wet.
Capacity Approx. 20ml
Dispersion mechanisms
Modes of operation Automatic via SOPs.
Weight
Cell
Control box/cell holder
Dimensions
Cell
Control box/cell holder
AWA2003 - Continuously variable and independent pump and ultrasound.
AWA2004 - Continuously variable pump.
Manual via computer on-screen operating dialogues.
4.25kg (AWA2003), 4kg (AWA2004)
8kg (AWA2003), 7.75kg (AWA2004)
Width 108mm
Height: 242mm (excluding motor and handle)
Height: 335mm (including motor and handle)
Depth: 252mm
Width: 284mm
Height: 338mm
Depth: 252mm
Particle size range 0.05-120µm, depending on particle shape and density.
Power requirement 100-120V / 200-240V AC, 48W
Hydro 2000µP Page A-1
Only use the PSU and cables provided by Malvern. Using any other PSU may compromise safety and will void any warranty.
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Appendix A Specification
A
Pump
Speed
Accuracy
Max dispersant viscosity
Safety:
Ultrasound
Frequency
Power
Safety
0 and 500-5000 rpm
500-5000 rpm, +/-5% of F.S.
0.54 Poise (0.054 Pa.s)
Over current and over temperature protected.
48 kHz (nominal)
0 - 20W (dependent upon dispersant)
Over current protected.
Page A-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Chemical compatibility
Introduction
B
The Hydro 2000µP is manufactured from materials that give the widest protection from chemical attack. However, it is important to check that any sample or disper­sant used is chemically compatible with the materials it will come into contact with in the dispersion unit.
This appendix lists all materials that come into contact with the sample and disper­sant during normal operation.
Component Materials
Cell, pump cavity, drain, well, drain tube 316 stainless steel
Internal pipes FEP
Impeller, input syringe tube, non-aqueous dispersant inlet valve
Drain sump Polypropylene
Windows Glass
Internal seals Perlast™ (G70G)
Drain and overflow pipes Tygon™ (MH2075)
Non-return valve, well lid PEEK
Beaker Pyrex
PTFE
Hydro 2000µP Page B-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Appendix B Chemical compatibility
B
Page B-2 MAN 0389
MAN0389-1.0 Hydro 2000uP.book Page 1 Friday, March 16, 2007 4:11 PM
Regulatory statements
This appendix presents regulatory information.
C
Hydro 2000µP Page C-1
MAN0389-1.0 Hydro 2000uP.book Page 2 Friday, March 16, 2007 4:11 PM
Appendix C Regulatory statements
C
CE Declaration of Conformity
The CE badge on this product, shown below, signifies conformance to European Commission Directives.
Page C-2 MAN 0389
ill 8005
MAN0389-1.0 Hydro 2000uP.book Page i Friday, March 16, 2007 4:11 PM
Index
A
Accessory comms in connector
2-8
Accessory comms out connector 2-8 Accessory control dialogue 3-2 Anaerobic fill 3-4, 4-9 Anaerobic fill dialogue 4-9 Anaerobic fill sequence 4-4 Aqueous valve 2-7, 4-8
B
Background measurements
2-3
Bubbles 4-4
C
CE Declaration
C-2
Cell
filling
4-8
Cell unit
details
2-10
function 2-3, 2-5 Cell unit holder 2-6 Cell windows 2-3, 2-12 Changing dispersants 4-3 Chemical compatibility B-1 Cleanliness
guidelines
4-3
Cleanliness critical 4-2 Control cable 2-9, 2-11 Control unit 2-3, 2-5 Controller unit 2-7 Controlling the dispersion unit
manually
3-2
D
Degassing dispersant
4-5
Degassing water 4-5 Dimensions A-1 Dispersant
changing
4-3
degassing 4-5
overview 2-3
Dispersant inlet port 2-11 Dispersant input/syringe 2-5 Dispersant temperature 3-3 Dispersant valve
function
2-11
two types 4-8
types 2-7 Dispersion unit operation 2-3 Drain 2-8 Drain beaker 2-5 Drain beaker holder 2-6 Drain port 2-11 Drain valve 3-3 Draining 2-4, 4-4 draining 4-4
E
Error indicators
3-4
F
Filling the cell
4-5, 4-8
Filling the syringe 4-6 Front panel (controller unit) 2-9
G
Getting help
1-3
H
Hardware features
2-4
Hazardous dispersants/samples 2-6
I
Instrument preparation
4-5
L
Lid locking lever
2-6, 2-12
Low volume use 4-2 Lower/upper limit (temperature) 3-8
M
Main system components
2-4
Hydro 2000 µP Page i
MAN0389-1.0 Hydro 2000uP.book Page ii Friday, March 16, 2007 4:11 PM
Index Hydro 2000 µP
Making a measurement 3-2 Manual control 3-2 Manual drain 2-8 Measurements
guidelines
4-4
Menu commands 1-3 Model number 1-1
N
Non-aqueous valve
2-7, 4-8
O
Operator functions
1-2
Overflow 2-11
P
Particle size range
A-1
Pause pump 3-3 Power input 2-7 Power requirement A-1 Power supply 2-4 Power switch 2-9 Powering up 4-5 Preparing for measurement 4-5 Pump - what it does 2-3 Pump speed 2-2, 3-6, 4-4
Anaerobic fill 4-9
changing 2-2 Pump speed control 3-3 Pump speeds 4-4
R
Sample well 2-5 Sampler Selection dialogue 3-5 Sampler Settings dialogue 3-6 Scattered laser light 2-3 Slurry 4-3 SOP control 3-5 Specification A-1 Stabilising temperature 3-8 Status indicator 2-9 Status indicators 3-4 Supervisor functions 1-2 Swapping cell holder position 5-2 Syringe
capacity to use
4-6
filling 4-6
Syringe capacity 2-11
T
Temperature monitoring
3-8
Temperature sensing 2-2 Time-out period (temperature) 3-8
U
Ultrasonic control cable
2-10, 2-11
Ultrasonic power
changing
2-2
Ultrasonic probe 2-4 Ultrasonic system 2-2 Ultrasonics settings 3-7 Ultrasound
optimum level duration 3-7
Ultrasound control 3-3
Rear panel (controller unit) Regulatory statements C-1 Representative sample 4-3 Rinsing system 4-3
S
Sample handling unit options Sample handling unit selection 3-5 Sample loss
minimising
4-2
Sample preparation 4-3
Page ii MAN 0389
2-7
3-6
V
Var i ab le s pe ed pum p
2-2
Vari-flow options 3-6
W
Water
degassing
4-5
Well cone 2-10 Well lid 2-4, 2-6, 2-11 Window ring lock screws 2-12
Loading...