Note: Due to Malmet’s policy of continuous product improvement;
design and technical specifications are subject to change without notice.
It is important that the name from whom you purchased your device and the name of the installer are recorded above. The installer is
responsible for the correct installation, start up and demonstrating the operation of this device. They are also responsible for issuing relevant
certificates of compliance (these may differ from state to state).
To obtain maximum life and efficiency from your Malmet Bedpan / Urinal Bottle Washer Disinfector and to
ensure safe operation, please read this manual thoroughly and follow all instructions before operating the
device.
This manual provides information on the operation of the device. It is recommended that all persons
operating the device have access to this manual for training purposes.
This unit is not intended for use by any person without the proper training, experience or knowledge.
The specifications supplied in this manual were in effect at the time of publication. However, owing to
Malmet (Australia)’s policy of continuous improvement, changes to these specifications may be made at
any time without notice on the part of Malmet (Australia).
Quality Policy
Malmet’s quality management system is certified to ISO 9001:2015 and ISO 13485:2016 and guarantees
the quality of this product.
Certifications
Electrical Safety: Cert No. CS10847N to IEC 61010-2-040, IEC 61010-1
EMC Compliance: Report T130119 to IEC 60601-1-2
Watermark: Cert No. WMKA21156
Important Warranty Reminder
Should you have any problems with your device, contact the company from whom you purchased it, or
Malmet (Australia) Pty Ltd.
It is important that the name from whom you purchased your device and the name of the installer are
recorded on the front page of this manual. The installer is responsible for the correct installation, start up
and demonstrating the operation of this device. They are also responsible for issuing relevant certificates
of compliance (these may differ from state to state).
Please read and understand this manual before using this device, if this device is used in a manner not
specified by the manufacturer protection by the device may be impaired.
Please refer to this manual for information wherever this warning symbol is displayed -
Be aware of 240V Voltage
Disconnect power when servicing
Mains power ISO switch or circuit-breaker must be in an
accessible position, easily reached and not obstructed so device
can be isolated from mains power during service
Only process items listed as per the design parameters in this
manual
Be aware of steam discharge
Goods and racks are hot to handle
Safety gloves and goggles must be worn when changing detergent
Safety clothing with reflective tape can activate the hands free
sensor when device is in standby mode
Be aware of hot surface, pipes and hoses from steam and hot
water
Install temperature probes and element over temperature
protection thermal cut-outs correctly
Plumbing service connection must comply with AS/NZS 3500
The Malmet Bedpan / Urinal Bottle Washer Disinfector has been designed within the following parameters:
a) A single bedpan with lid and two urinal bottles can be emptied, cleaned and disinfected during each
automatic cycle.
The utensils that can be cleaned in the device are:-
i) Standard size bed pans
ii) Standard commode bowl
iii) Standard male and female urinal bottles
iv) Most plastic urinal bottles including male non-spill and female cervec
v) Small slipper pan
vi) Large slipper pan
b) The cradle is designed to ensure that utensils are not dislodged during the cleaning cycle; the contents
are emptied during door closure.
c) The chamber and door are self-cleaning and do not permit water or soil to remain after a properly
completed cycle. Steam disinfection ensures all internal surfaces are totally clean and safe.
d) The flush and clean stage:
i) Removes the soil
ii) Clears the trap
e) A complete cycle is completed in approximately four (4) minutes for the 20 Amp Bedpan /Urinal Bottle
Washer Disinfector and six (6) minutes and thirty (30) seconds for the 10 Amp Bedpan / Urinal Bottle
Washer Disinfector
Note: Dependant on ambient air temperature.
Note: The first cycle will take longer as the device has to reach the operating water temperature of
90˚C.
f) Malmet Door Obstruction Feature
This is factory set to 12Nm ±10%; causing the door to re-open when an obstruction is detected. This is
designed to protect the integrity of the device and the articles within.
For the safe use of this device the responsible body should ensure that all operators are adequately trained
to operate and maintain the device in its safe use.
Hands Free Operation
Zero Contamination – Totally Hands Free
The Malmet Bedpan / Urinal Bottle Washer Disinfector is fully automatic.
Door open, door close and cycle start can be activated without touching the device.
Front View
Manual operation of the door and a manual start button are available.
To Operate
To Open Door – Break Sensor Beam
To Close Door – Break Sensor Beam
To Start Cycle – Break Sensor Beam once when green LED’s flash within 8 seconds after door
closure
Note: If door is opened then closed and cycle has not commenced the beam must be broken twice.
The detergent system delivers 30ml of concentrated detergent in the initial wash stage.
Current Safety Data Sheet for Malmet Device Detergent is available in pdf format from Malmet’s website.
www.malmet.com.au
Detergent Out Condition - F 30 Fault
If insufficient detergent is available to deliver the required quantity, the device will display an F 30 dEt fault.
Access to the load will be restricted until the condition is rectified.
To rectify this fault, install a new detergent bottle as per the procedure in this manual “Changing the Detergent Bottle”.
On completion of changing the detergent bottle, press either the “Manual Door” or “Manual Start” button
to initiate a priming sequence. If priming is successfully the fault will clear and door will open. Inspect the
items, ensuring they have been cleaned sufficiently before unloading. It is recommended the items are
reprocessed in the event of an F 30 dEt Fault.
If priming is unsuccessful, access to the load will continue to be restricted and F 30 dEt fault displayed.
To avoid the detergent out condition check the low level
mark on detergent door, replace detergent when the
detergent level reaches this mark.
Push down slipper pan
until it locks into
position
Detergent Model with Hands Free Operation
Operation, Maintenance and Installation Manual
Loading Configurations
The device is intended to be loaded as per the below configurations, please refer to the design parameters
section of this manual for items that this device is designed to process. Part loads or singular items can be
processed but must be loaded in their designated positions as below.
6. Once the device is “loaded” the operator breaks the infra-red beam to close the door.
The display shows Shut
7. When the door is fully closed 2 green LED’s will flash in the infra-red sensor area for 8 seconds to advise
the user to break the beam again to start a cycle.
The display shows rdY
If the 8 seconds has elapsed the flashing LED’s will stop and the device will not start for a cycle. The
operator has to repeat the process from Step 5 (or press manual start).
Note: If this occurs the beam must be broken twice.
Note: On 10A models, if the steam generator temperature is not at 90°C after the beam has been
broken or the manual start button pressed, the display will show “CYCL”. The Green LED’s
will stop flashing and stay on. When the steam generator has reached temperature the
display will show rinS and the wash cycle will start automatically.
8. Once a cycle has been initiated the elements are turned on in the steam tank to start heating the
water.
The wash pump is turned on for 10 seconds for the initial wash.