Mallinckrodt INOMAX DSir Plus Operation Manual

Part No. 20003 Rev - 01
Part No. 20717 Rev-01
2014-07
Operation Manual
(800 ppm INOMAX® (nitric oxide) for inhalation)
Series 3 software
Part No. 20717 Rev-01
2014-07
This Product will perform in conformity with the description contained in this operating manual and accompanying labels and/or inserts, when assembled, operated, maintained and repaired in accordance with the instructions provided. This Product must be checked prior to use following the Pre-Use Checkout procedure described in section two. A defective Product should not be used. Parts that are broken, missing, visibly worn, distorted or contaminated should be replaced immediately.
Should such repair or replacement become necessary, the manufacturer recommends that a telephone request for service advice be made to the local distributor. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by the manufacturer or local distributor. The Product must not be altered.
The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Mallinckrodt Representatives.
Caution: U.S. Federal and Canadian law restrict this device to sale by or on the order of a licensed medical practitioner. Outside the U.S.A. and Canada, check local laws for any restrictions that may apply.
Inhaled Nitric Oxide mixtures must be handled and stored in compliance with federal, state and local regulations.
These products have unit serial numbers with coded logic which indicate the year of manufacture and a
sequential unit number for identication.
No license is conveyed, either expressed or implied, with the purchase or usage hereof under any patent or patent application covering this product. See Patents www.mallinckrodt.com/patents and any respective foreign equivalents thereof.
Open Source Software A CD-ROM is available upon request containing the full source code to the open source software used within this product. Portions of this software are copyright © 1996-2002 The FreeType Project (www.freetype.org). All rights reserved.
Korean fonts Baekmuk Batang, Baekmuk Dotum, Baekmuk Gulim, and Baekmuk Headline are registered trademarks owned by Kim Jeong-Hwan.
Mallinckrodt, the “M” brand mark and the Mallinckrodt Pharmaceuticals logo are trademark of a Mallinckrodt company. Other brands are trademarks of a Mallinckrodt company or their respective owner. ©2016 Mallinckrodt
Important: Before using the INOmax DS
IR
, read through this manual.
Read through the manuals for the ventilator, humidier and any other accessory items used. Follow the manual
instructions and obey the Warnings and Cautions. Keep this manual readily available to answer questions.
SN 20051234
The rst four numeric digits indicate the year of product manufacture, and the next
four digits are the sequential unit number produced.
Ref 10007
INOmax DSIR part number
Contents
1/ General Information ........................................................................................................................1-1
Indications for Use..........................................................................................................................1-1
Introduction to this Manual .............................................................................................................1-2
INOmeter Operation .....................................................................................................................1-18
Theory of Operation .....................................................................................................................1-22
Environmental Effects ..................................................................................................................1-26
2/ Automated Pre-Use Checkout ........................................................................................................2-1
Initial connections ...........................................................................................................................2-2
High Pressure Leak Test and Automated Purge ............................................................................2-5
Integrated Pneumatic Backup INOMAX Delivery Test ...................................................................2-7
Performance Test ...........................................................................................................................2-8
INOblender Test .............................................................................................................................2-9
Depressurizing the Regulator Supply Line ...................................................................................2-10
3/ Patient Application .......................................................................................................................... 3-1
INOblender Operation ....................................................................................................................3-4
INOblender Used as a Stand-Alone Device ...................................................................................3-5
INOblender use with the NeoPuff ...................................................................................................3-6
Integrated Pneumatic Backup NO Delivery ....................................................................................3-7
Changing INOMAX Cylinders and Purging the Regulator Assembly ...........................................3-10
Oxygen Dilution Chart ..................................................................................................................3-13
Duration Chart INOMAX Cylinder 88-Size ...................................................................................3-14
Duration Chart INOMAX Cylinder D-Size .....................................................................................3-15
Monitoring the Environment .........................................................................................................3-19
Entering Patient Information .........................................................................................................3-20
Connection to Various Breathing Systems ...................................................................................3-24
Acutronics Medical Systems AG Fabian +nCPAP Evolution ......................................................3-25
Acutronics Medical Systems AG Fabian HFO ............................................................................3-26
A-Plus Medical Babi-Plus Bubble CPAP Circuit .........................................................................3-27
Bagging Systems While Using the Injector Module ....................................................................3-28
Bunnell Life Pulse High Frequency Ventilator Circuit .................................................................3-32
Connecting INOmax DS Connecting INOmax DS
Circle Anesthesia System ........................................................................................................... 3-34
Dräger Babylog VN500/Innity Acute Care System and Heinen & Löwenstein
Leoni-plus Ventilator ................................................................................................................3-36
Fisher & Paykel Healthcare Bubble CPAP .................................................................................3-37
Fisher & Paykel Healthcare Infant Circuit Nasal Cannula ..........................................................3-38
Fisher & Paykel Healthcare Optiow Breathing Circuit ..............................................................3-39
Hamilton Arabella Nasal CPAP ..................................................................................................3-40
ICU Ventilator Circuit ..................................................................................................................3-41
Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Filtered Circuit ...............3-42
Sensormedics 3100A/B High Frequency Oscillatory Ventilator with a Rigid or Flexible Circuit .3-43
SLE Life Support SLE5000 ........................................................................................................3-44
Spontaneous Breathing Patient on a Mask Circuit ....................................................................3-45
Spontaneous Breathing Patient on a Nasal Cannula .................................................................3-46
Teleex Medical Comfort Flo Humidication System .................................................................3-47
Vapotherm 2000i ........................................................................................................................3-48
Vapotherm Precision Flow ..........................................................................................................3-49
Viasys Infant Flow CPAP System; Cardinal Airlife nCPAP System ............................................3-50
Viasys Infant Flow SiPAP ...........................................................................................................3-52
Sample Tee to the Bunnell Life Pulse Circuit ...................................3-33
IR
Injector Module to the Bunnell Life Pulse Circuit ..............................3-33
IR
Part No. 20717 Rev-01
2014-07
i
4/ Transport ..........................................................................................................................................4-1
Transport Options ..........................................................................................................................4-1
A. Intrahospital transport (within the hospital) when moving the
INOmax DS
B. Intrahospital transport (within the hospital) when removing the INOmax DS
as a unit (cart and cylinders) .............................................................................4-1
IR
and INOblender
IR
from the cart. .............................................................................................................................4-2
C. When using the INOblender as a stand-alone device. ..............................................................4-6
INOblender Test Using the INOmax DS
to Analyze Output ....................................................4-8
IR
INOblender Stand-Alone Pre-use Checkout ...........................................................................4-10
D. InterHospital Transport (Between Hospitals) when using a separate INOmax DS
and
IR
INOblender for transport ...........................................................................................................4-11
Duration Chart ..............................................................................................................................4-12
INOMAX Cylinder D-Size .............................................................................................................4-12
Transport Regulator/Cap Assembly Application ...........................................................................4-13
Connection to a Dual-Limb Transport Ventilator Circuit .............................................................4-16
Connection to a Single-Limb Transport Ventilator Circuit ...........................................................4-17
Cylinder Leak Check ....................................................................................................................4-19
5/ Alarms ..............................................................................................................................................5-1
Alarm History ..................................................................................................................................5-3
Alarm Help......................................................................................................................................5-5
6/ Calibration ........................................................................................................................................6-1
Low Calibration...............................................................................................................................6-2
Oxygen Sensor High Calibration ....................................................................................................6-4
NO Sensor High Calibration ...........................................................................................................6-7
NO
Sensor High Calibration.........................................................................................................6-11
2
7/ Maintenance .....................................................................................................................................7-1
Cleaning the INOmax DS Replacing the O
, NO and NO2 Sensors ........................................................................................7-6
2
...........................................................................................................7-2
IR
Replacing the Water Separator Cartridge ......................................................................................7-8
Cylinder Leak Check ....................................................................................................................7-10
Preventative Maintenance .............................................................................................................7-11
Parts and Accessories .................................................................................................................. 7-12
8/ Product Specications....................................................................................................................8-1
Ventilator Compatibility ...................................................................................................................8-1
RS232 Data Output ........................................................................................................................8-7
Electromagnetic Compatibility Information .....................................................................................8-9
9/ Appendix ..........................................................................................................................................9-1
Manual Pre-Use Checkout .............................................................................................................9-1
Initial connections .........................................................................................................................9-2
High Pressure Leak Test ................................................................................................................9-5
Manual Purge and Alarm Verication ...........................................................................................9-6
Integrated Pneumatic Backup INOMAX Delivery Test .................................................................9-7
Performance Test .........................................................................................................................9-7
INOblender Test ...........................................................................................................................9-8
ii
Part No. 20717 Rev-01
2014-07
WARNING:
Warnings tell the user about dangerous conditions that can cause injury to the operator or the patient if you do not obey all of the instructions in this manual.
Caution:
Note:
Blue arrow denotes required user action.
WARNING:
Cautions tell the user about how to properly use the equipment and conditions that could cause damage to the equipment.
Read and obey all warnings and cautions.
Notes provide clarication or supplemental information.
Integrated Pneumatic Backup
• The integrated pneumatic backup is intended for short term use when the electronic delivery system fails until a replacement NO delivery device can be brought to the bedside.
• The integrated pneumatic backup delivers a variable concentration of NO to the
patient depending on the ventilator ow being used.
• When using the integrated pneumatic backup with breathing circuit gas ows of
5 L/min, the delivered NO dose will be approximately 40 ppm. Breathing circuit gas
ows less than 5 L/min will deliver an NO dose greater than 40 ppm.
• The integrated pneumatic backup (250 mL/min.) should not be used with the
Bunnell Life Pulse as ventilator ow rates are normally below the recommended ventilator ows.
Changing Cylinders
• Always secure a cylinder when not using it.
• Never lift a cylinder by its valve or valve protection cap or by using a chain, sling or magnet.
• Never drop a cylinder.
• Never use a hammer, pry or wedge to loosen a valve or protection cap. The valve and protection cap should operate by hand.
• Never let oil, grease or other combustible come in contact with a cylinder or valve.
• Never remove or deface cylinder labeling or markings.
• Never modify equipment without rst contacting the manufacturer.
• Never use an adaptor to connect a cylinder to the system.
• Never use equipment not designed to use INOMAX mixtures.
• Never attempt to repair a leak on a cylinder valve or its safety relief device.
• Never operate equipment that is leaking.
• Never ship a leaking cylinder.
• Never store cylinders;
- Where damage can result from the elements, such as standing water or temperatures over 125 degrees F.
- Where they can experience extreme low temperatures.
- Where they can contact corrosive substances.
- Where they can be cut or abraded by an object.
- Next to a walkway, elevator or platform edge.
- Unless they are properly secured.
Part No. 20717 Rev-01
2014-07
iii
WARNING:
High Frequency Oscillatory and Jet Ventilator Circuits
• Some high frequency ventilator circuits require a one-way valve to prevent high NO delivery.
• Place the Bunnell Life Pulse in Standby prior to suctioning the patient to avoid NO delivery transiently exceeding the set dose by up to 30 ppm. Press ENTER to reestablish ventilation as soon as the catheter is removed from the airway. This will limit the extent of over delivery above the NO set dose.
• Do not use dose settings above 40 ppm with the Acutronics Fabian HFO
ventilator. Bidirectional ow through the Injector Module may cause over-delivery
which can lead to measured NO values greater than 100 ppm
Maintenance
• For continued protection against hazard, replace the fuses only with the correct fuse type and rating.
• Handle and dispose of sensors according to facility biohazard policies. Do not incinerate.
• If the Injector Module has been used in the wet/humidied part of the breathing
circuit, it should be sterilized between each patient use.
• INOmax DS
- 4 kV input to output isolation
- 4 kV input to mains isolation, and
- an internal “reference voltage” “U” (as dened in section 20.3 of IEC60601-
1 ed. 2) of less than or equal to 50 VDC or 50 VRMS and dielectric isolation
certied in accordance with IEC 60601-1. Interface cabling must not go outside of
the room (e.g., into walls where potential isolation issues could exist). Adherence to the above provides compliance to clause 20.3 “Value of test Voltage” in edition 2 and clause(s) 8.5.4 “Working Voltage” and Clause 8.8.3 “Dielectric Strength" in edition 3.
• RS232 cable must be shielded. The RS232 cable shield shall have a minimum of 90% coverage. The shield shall only be connected at one end of the cable to minimize noise induced by ground currents.
should only be connected to RS-232 ports that have:
IR
Manually Bagging a Patient with an Injector Module
• The hyperination bag will, under some conditions, contain NO
one ppm. Use of large tidal volume breaths may expose the patients to the NO
present in the bag for part of the breath. In general, if the inspiratory ow rate induced by the manual ventilation does not exceed the fresh gas ow rate,
the patient should not be exposed to the concentrations of NO
hyperination bag.
• Adult and infant hyperination bags generate more NO
minute ventilation. If use of the bag is interrupted (for example to adjust the tracheal tube), before resuming ventilation of the patient, the user should squeeze the bag several times to empty residual gas from the bag.
• Because of the potential for inhalation of excessive concentrations of NO
the difculty in monitoring the peak inhaled NO a hyperination bag or self-inating bag is intended only for short term use.
• The monitoring system within the INOmax DS
within the hyperination bag or self-inating bag devices and the alarms for
excessive NO
iv
2
in excess of
2
present in the
2
when used at lower
2
concentrations, ventilation with
2
will not detect generation of NO2
IR
cannot warn of NO2 produced within the manual bag system.
Part No. 20717 Rev-01
, and
2
2014-07
2
WARNING:
Manually Bagging a Patient with an Injector Module continued
• To minimize the delivered concentration of NO2, the following steps should be taken for use with the manual resuscitator bags:
- Concentrations greater than 20 ppm NO should not be used because of excessive NO
generation.
2
- Use the smallest bag adequate to deliver the desired tidal volume.
- Oxygen tubing lengths greater than 72 inches should not be used (between the injector module and the bag).
- Use the highest fresh gas ow rate (up to 15 L/min) that is practical.
- Use the lowest practical inspired oxygen concentration.
- After starting fresh gas ow, squeeze the bag several times to empty residual
gas in the bag prior to using the system to ventilate a patient.
Manually Bagging a Patient with the INOblender
The purge procedure must be followed to help ensure NO
system before the manual resuscitator bag is connected to the patient.
The manual bag should be squeezed repeatedly during use to avoid NO
up in the bag.
If the bag is not squeezed repeatedly while delivering INOMAX, the bag should
be removed from the patient and the bag purge procedure performed before continuing.
The INOblender should be upright when setting the oxygen owrate for accurate
setting.
Do not use pneumatically powered nebulizers with the INOblender. This will result
in signicant over delivery of INOMAX in excess of 80 parts per million (ppm).
- The INOblender outlet pressure has been validated for use up to 400 millibar (5.8 psig) pressure. The amount of back-pressure generated by pneumatic nebulizers
is signicantly greater (20-30 psig) and will result in over delivery of INOMAX
in excess of 80 ppm. The user adjusted dose setting on the INOblender will not correlate with, or have an effect on the actual delivered dose.
- In addition, the INOblender owmeter is not back-pressure compensated and will display a lower ow rate than actual when pressure is applied to the outlet.
is purged from the
2
building
2
Purging Cylinders
• A new INOMAX cylinder and regulator must be purged before use to ensure the patient does not receive an excess level of NO
• If the INOmax DS depressurize the regulator supply line.
• If the INOmax DS purge procedure.
• If the INOmax DS procedure.
Part No. 20717 Rev-01
2014-07
.
2
is not going to be used on a patient within 10 minutes,
IR
is not used and is pressurized for more than 10 minutes, repeat
IR
is depressurized and not used within 12 hours, repeat pre-use
IR
v
WARNING:
Use Outside of Product Labeling
• The INOmax DSIR must only be used in accordance with the indications, usage, contraindications, warnings and precautions described in the INOMAX (nitric oxide) drug package inserts and labeling. Refer to this material prior to use.
• The manufacturer does not recommend that the INOmax DS helium/oxygen mixtures in any situation. The INOmax DS
be utilized with
IR
is intended to deliver
IR
INOMAX therapy gas only in conjunction with the delivery of air and oxygen. The use of helium/oxygen mixtures will lead to over-delivery of INOMAX which may lead to interruption of therapy.
• The use of devices which radiate high-intensity electrical elds may affect the
operation of the INOmax DS
. Constant surveillance of all monitoring and life
IR
support equipment is mandatory whenever interfering devices are in operation on or near a patient.
• Do not connect items which are not specied as part of the system.
• The approved patient population for the INOmax DS
, as specied in the drug
IR
labeling for INOMAX (nitric oxide) for inhalation, is limited to neonates. The INOmax DS
is not intended to be used in other patient populations.
IR
Transport
• If the INOmax DSIR or INOblender is to be used in a transport vehicle, they should
be afxed to the transport mounting post (part number 10009), which is part of
the transport mounting bracket assembly (part number 50041).
• The transport mounting post and/or the transport mounting bracket assembly should be secured to the transport isolette/transport gurney in a manner which will secure the INOmax DS
/INOblender.
IR
Troubleshooting or Calibrating
• If an alarm occurs, safeguard the patient rst before troubleshooting or repair
procedures.
• Use caution when troubleshooting the INOmax DS in use for a patient. When possible replace the unit in question and perform troubleshooting procedure once the unit is removed from the patient.
• If changing an NO sensor while delivering NO to a patient, install the NO sensor only when the NO high range calibration screen is displayed otherwise there is a risk that the system will shut down.
• INOMAX can be administered during the sensor calibration process. However, inspired gases are not monitored and gas monitoring alarms are disabled.
• Loss of communication between the INOmax DS more than one hour will result in interruption of INOMAX delivery.
delivery system while
IR
and the INOMAX cylinder for
IR
vi
Part No. 20717 Rev-01
2014-07
WARNING:
Ventilators and Breathing Devices
• The INOmax DSIR subtracts gas from the breathing circuit via the gas sampling system at 230 mL per minute which can cause the ventilator to auto-trigger.
Adjusting the ow sensitivity may be necessary. The trigger sensitivity of the
ventilator should be checked after connecting the INOmax DS circuit.
• Set the INOmax DS
alarm thresholds for the current patient conditions to
IR
monitor any inadvertent changes in treatment.
• Be certain all cables and hoses are positioned to help prevent damaging or occluding them.
• The use of pediatric and neonatal ventilator settings with adult size breathing circuits can result in high levels of NO
. Always use the size of breathing circuit
2
that is appropriate for the patient.
• The humidier chamber volume should not be more than 480 mL to prevent
elevated NO
values.
2
• When handling any component of the patient circuit that comes in contact with
patient’s uids wear personal protective equipment (PPE).
• Patient disconnect and high pressure alarms are required for the ventilator.
• Patient circuit pressure and gas loss will result if cap is not in place (secured).
• Use only “Latex-Free” breathing circuits and ventilators when using the INOmax DS
• If the INOmax DS
.
IR
is to be used in a transport vehicle, it should be afxed to the
IR
transport mounting post.
• Avoid recirculation of gases. Undesired recirculation of gases will occur if fresh
gas ows are less than the patient minute volume and may result in:
- Higher NO remove NO
levels due to the limited ability of the carbon dioxide absorbent to
2
.
2
- Higher NO concentrations than those set due to NO recirculated through the absorber.
- Reduction in O
concentration because nitrogen is the balance gas for nitric
2
oxide and will be present in the re-circulated gases.
• Only use parts/accessories designated for use with this system.
to the breathing
IR
Part No. 20717 Rev-01
2014-07
vii
(Intentionally left blank)
viii
Part No. 20717 Rev-01
2014-07
Part No. 20003 Rev - 01
Part No. 20717 Rev-01
2014-07
1/ General Information
General
Information
Part No. 20717 Rev-01
2014-07
1/ General Information
General
Information
1/ General Information
Indications for Use
• The INOmax DSIR (delivery system) delivers INOMAX (nitric oxide for inhalation) therapy gas into the inspiratory limb of the patient breathing circuit in a way that provides a constant concentration of nitric oxide (NO), as set by the user, to the patient throughout the inspired breath. It uses a specially designed injector module, which enables tracking of the ventilator waveforms and the delivery of a synchronized and proportional dose of NO. It may be used with most ventilators.
• The INOmax DS comprehensive alarm system.
• The INOmax DS the absence of an external power source.
• The INOmax DS
which along with user supplied 10 L/min of oxygen provides 20 ppm in the gas ow to a patient's breathing
circuit. It may also use the INOblender for backup.
• The target population is controlled by the drug labeling for INOMAX which is currently neonates; refer to the
drug label for specic information. The primary targeted clinical setting is the Neonatal Intensive Care Unit
(NICU) and secondary targeted clinical setting is the transport of neonates.
provides continuous integrated monitoring of inspired O2, NO2 , and NO and a
IR
incorporates a battery that provides up to six hours of uninterrupted INOMAX delivery in
IR
includes a backup NO delivery capability that provides a xed ow of 250 mL/min of NO
IR
Part No. 20717 Rev-01
2014-07
1-1
Introduction to this Manual
Denitions and abbreviations
Term Denition
% v/v % volume/volume Breathing circuit Part of ventilator or breathing system that connects to the INOmax DS Breathing system Non-invasive breathing devices. Control wheel Rotary control used to change and conrm settings.
®
Cylinder Aluminum cylinder containing INOMAX HFOV High frequency oscillatory ventilation.
therapy gas.
®
.
IR
INOblender
®
Back up to the INOmax DSIR. Allows manual ventilation of the patient, providing
uninterrupted delivery of INOMAX. INOMAX Nitric oxide for inhalation. INOmeter
®
Counter mounted on a cylinder that records the amount of time the INOMAX cylinder
valve is open. Infrared (
) Infra-red technology by which the INOmax DSIR communicates with the INOmeter
IR
mounted on each cylinder. N
2
Nitrogen. NO Nitric oxide. NO O
2
2
Nitrogen dioxide.
Oxygen. ppm Parts per million. Pre-use circuit Connectors and tubing assembly required for INOmax DS psig Pounds per square inch gauge. Set NO Dose of INOMAX set by the user.
for pre-use checkout.
IR
This manual shows the Set NO displays associated with the 0-80 ppm range.
1-2
Part No. 20717 Rev-01
2014-07
1
2
3
4
11
12
13
14
15
16
10
9
8
7
6
1. Sample Line Inlet
2. Main Power Indicator
3. Display Screen
4. Alarm Speaker (under front label)
5. Integrated Pneumatic Backup Switch
6. Control Wheel
7. Injector Module Tubing Outlet
8. Injector Module Cable Inlet
5
9. Water Trap Bottle
10. Purge Port
11. INOMAX Gas Inlets
12. Blender Gas Outlet
13. Ethernet Port
14. Infrared Connector
15. USB Port
16. Water Separator Cartridge
23
22
21
20
19
17. Water Trap Bottle
18. Sample Gas Outlet Port
19. Clamp Assembly
20. Electrical Cord Inlet
21. Equipotential T erminal
22. ON/Standby Switch
23. RS232 Port
Figure 1-1 INOmax DSIR Front View Figure 1-2 INOmax DSIR Rear View
18
17
Part No. 20717 Rev-01
2014-07
1-3
1
2
3
5
6
8
10
4
7
9
11
13
1. INOmax DS
IR
2. INOmax DSIR Mounting Post
3. Clamp Assembly
4. INOMAX Regulator (2)
5. INOblender
6. Small Part Bin
7. INOmeter
8. INOMAX Cylinder
9. Cylinder Holding Bracket
10. Cylinder Mounting Strap
11. Oxygen Cylinder Bracket
12. Caster Lock Lever
13. Caster (4)
Figure 1-3 INOmax DSIR and Cart
1-4
12
Part No. 20717 Rev-01
2014-07
(Intentionally left blank)
Part No. 20717 Rev-01
2014-07
1-5
Navigating the Display Screens
Note:
The specic level is identied by the highlighted card on the Menu Button. The red arrows indicate going back to a previous screen.
Recent Alarms Screen (second level)
Main Screen (rst level)
Patient Information Screen (second level)
Menu Screen (second level)
1-6
Navigating the Menu Screen (see page 1-8)
Part No. 20717 Rev-01
2014-07
Menu Screen (second level)
Low Calibration Screen Settings Screen
Pre-Use Wizard Alarm History Screen
High Calibration Screen Automated Purge Screen
Part No. 20717 Rev-01
2014-07
1-7
19
1
3
2
4
5
6
Main Display Screen
• On the main screen the user can
7
view alarm messages, monitored values and graphical information.
• By pressing the “Menu Button” on the touch screen (top right hand corner), the user can access the menu screen (see Figure 1-5 ).
8
10
1718
16
15
14
13
12
11
9
1. Alarm Silence Button
2. Upper Alarm Limit Button
3. Lower Alarm Limit Button
4. Monitored Value
5. Menu Button
6. Text Message Area
7. Monitor Area
8. Graphical Area
9. Patient Information Button
10. Sample Line Icon
11. Water Trap Bottle Icon
12. Inspiratory Limb Icon
Figure 1-4 Main Display Screen
1
8
910
1. Pre-Use Checkout Button
2. Auto Purge Button
3. Return to Previous Level Button
4. Monitor Area
5. Alarm History Button
2
7
6. Settings Button
7. High Calibration Button
8. High Calibration Due Date
9. Low Calibration Button
10. Last Low Calibration Date
Figure 1-5 Menu Screen (second level)
13. Injector Module Icon
14. Delivery Line Icon
15. Integrated Pneumatic Backup Line Icon
16. Integrated Pneumatic Backup Switch Icon
17. Delivery Setpoint Display
18. NO Delivery Setpoint Button
19. Cylinder Icon
Menu Screen (second level)
3
• On the menu screen the user can access the Pre-Use Checkout (#1) and the Auto Purge (#2) wizards.
4
Note:
The Pre-Use Checkout and Auto Purge buttons are inactive (greyed out) if a dose is set.
5
• To review the complete alarm history, press the Alarm History button (#5), (refer to Section 5/ Alarms).
• To initiate a low (room air) or high calibration, press either the Low Cal (#9) or High Cal (#7) buttons.
6
(refer to Section 6/ Calibration).
• Press the Settings button (#6) to
view circuit ow and calculated
delivery graphs, change display brightness, change alarm volume, change time zone and view software
revision (see
F
igure 1-
6).
1-8
Part No. 20717 Rev-01
2014-07
8
1. Return to Previous Level Button
2. Monitor Area
3. Display Brightness Button
4. Time Adjust Button
7
Figure 1-6 Settings Screen
1
2
3
4
6
5. Software Revision
6. Alarm Volume Button
7. Calculated Delivery Graph
8. Circuit Flow Rate Graph
5
Settings Screen (third level)
• The circuit ow graph, combined with
calculated delivery graph, is a user level tool to ascertain NO delivery system limitations in the context of mechanical ventilation.
• The circuit ow rate graph displays the real time peak and average ow rate in the
breathing circuit over a 10 second time period, as measured by the injector module. The
area in green represents the circuit ow range
where the INOmax DS
system is rated to
IR
deliver NO from 1-80 ppm. (see maximum NO delivery graph page 1-25). Display graphic areas in yellow represents where some inaccuracy of NO delivery is to be expected.
• The calculated delivery graph displays the delivery subsystem calculated % error in NO delivery compared to the desired set dose over a 10 second time period. The green
to yellow transition points are dened to be
NO delivery error of +20% (over-delivery) to -20% (under-delivery) compared to the desired NO set dose. High set dose levels,
high peak inspiratory ows and proportional NO ow control inaccuracies can contribute
to the indicating arrow varying from the ideal midpoint.
Note:
• If the NO dose is not set, both graphs will remain inactive.
• See page 1-25 for further explanation of maximum deliverable NO concentration.
Part No. 20717 Rev-01
2014-07
1-9
Display and user controls
The INOmax DSIR has a color touch screen display and a control wheel for adjusting and entering user settings. The buttons on the touch screen and the control wheel perform a variety of functions using a three­step procedure (see “Setting and making changes on the INOmax
The touch screen buttons and control wheel are used to:
• Set the concentration of delivered NO
• Adjust alarm limits
• Silence alarms
• Calibrate the sensors
• Review alarm history
• Dene setup options
• Enter patient information
DSIR” see page 1-12).
Note:
• If a button has been selected and no activity has been sensed within 20 seconds, the display will return to its previous condition. If a button is greyed out, it is not active.
• Position delivery system so user screen is unobstructed and the speaker is not covered.
When a value is being changed, pressing the "Cancel Active Status" button during editing will stop the change and return the parameter to its original value (similar to the escape key on a computer).
1-10
Part No. 20717 Rev-01
2014-07
Main Screen
Cylinder icons are not visible and the NO delivery setpoint button will remain inactive until the
INOmax DS
recognizes an INOMAX cylinder.
IR
Caution:
High frequency and/or high intensity light emission, in the area of the INOmeter, may interfere with communication between the INOmax DS
and the INOmeter on the
IR
INOMAX cylinder (see page 1-15).
The cylinder icons will appear on the main screen in relation to their position on the cart when the user is
facing the INOmax DS
Note:
When using the transport regulator/cap
.
IR
assembly (PN 10022) only one cylinder will be displayed.
Part No. 20717 Rev-01
2014-07
When an INOMAX cylinder valve is opened, the cylinder handle graphic will turn green representing an open INOMAX cylinder valve.
1-11
Setting and making changes on the INOmax DS
IR
Dose settings
Displayed dose settings are 1, 5, 10, 20, 40, 60 and 80 ppm. Each click on the control knob corresponds to a known change in dose. The incremental dose per click corresponds to a value dependent upon the dose range in which the change is made, as illustrated in the table at right.
Dose Settings Adjustments
Dose Setting Range Dose Change Per Click
< 1 ppm 0.1 ppm
1 to 40 ppm 1 ppm
40 to 80 ppm 2 ppm
1. SELECT (press) a button on the touch
screen associated with the desired function. (An audible beep will sound when a button is selected, and the button will be displayed in inverse video.)
1-12
2. ROTATE the control wheel clockwise or
counterclockwise to adjust the value.
Part No. 20717 Rev-01
2014-07
3. CONFIRM the selection by pressing the control wheel or the button associated with the desired function again.
Note:
• After conrming a desired
dose, the NO alarm setting (high and low) will automatically be set for the
rst setting only.
• Any other changes will require the high and low alarm settings to be adjusted.
• Also a two minute lockout period will prevent the low NO monitoring alarm from occurring while the measured values stabilize.
The Monitor Alarm Delay
Active indicator is not active following a dose change from zero.
Part No. 20717 Rev-01
2014-07
1-13
Settings Screen Adjustments
Access the settings screen (third menu level).
Display Brightness setting
1. Push the display brightness button on the touch screen.
2. Turn the control wheel to indicate the display brightness level desired. Choices range from one (darkest) to 10 (brightest).
3. Push the control wheel to conrm the selection.
4. When nished with the menu screen, push the return to previous level button on the touch screen.
Alarm Volume setting
1. Push the alarm volume button on the touch screen.
2. Turn the control wheel to indicate the volume level desired. Choices
range from one (softest) to ve (loudest).
3. Push the control wheel to conrm the
selection.
4. When nished with the menu screen push the return to previous level button on the touch screen.
1-14
Time Adjust setting
1. If the "Time" button is pressed the Time Adjust screen will appear.
2. Press the Hour or Minute button on the touch screen.
3. Turn the control wheel to adjust the displayed hour or minute.
4. Push the control wheel to conrm the
selection.
5. When nished with the menu
screen push the return to previous level button on the touch screen.
Note:
Changing the displayed time does not impact the time written to the INOmeter since the time written to the INOmeter is the GMT time, not the displayed time.
Part No. 20717 Rev-01
2014-07
Infrared Communication between the INOMAX Cylinders and the INOmax DS
IR
WARNING:
Loss of communication between the INOmax DSIR and the INOmeter for more than one hour will result in interruption of INOMAX delivery.
The INOmax DSIR has an interface using infrared (IR) technology which allows the INOmax DSIR to communicate with the INOmeter INOMAX cylinder for the correct expiration date and cylinder concentration. The INOmax DS
(which is mounted to each INOMAX cylinder). The INOmax DSIR checks the
also transmits
IR
a conrmed patient identier to the INOmeter on any open INOMAX cylinder. The INOmax DS
cart (PN 10018) has a cover (see Figure 1-7, ) with an infrared transceiver mounted
IR
1
directly above each INOMAX cylinder. When INOMAX cylinders are loaded, communication will take place between the INOmax DS INOmax DS
is complete. A cylinder icon will be displayed on the main screen when an INOMAX cylinder is
IR
recognized by the INOmax DS
and the INOmeter (see Figure 1-7, ) after the boot up phase of the
IR
IR
(see “Loading INOMAX Cylinders onto the INOmax DS
Caution:
Nothing should be placed between the INOmeter and the cart to which it is attached.
Cart”, page 1-17).
IR
2
IR Communication Interference
The INOmax DSIR transceiver is located under the cart cover and should be protected from outside IR sources. The INOmax DS The INOmax DS
transceiver transmits via a 30 degree transmission cone projecting towards the oor (see
IR
dotted lines in Figure 1-7). The specications of the IR beam call for it to have a range of 20 cm (7.9 in). Based
on these specications it sould not affect other devices in the vicinity of the INOmax DS
cart was designed to protect the INOmeter from external light/IR energy sources.
IR
.
IR
The INOmeter uses a lower energy source which results in a lower IR beam range than the INOmax DS The INOmeter does not transmit IR signals unless it is mounted on the INOmax DS
Caution:
A strong magnetic eld could affect the ability of the INOmeter to detect if the cylinder valve is opened or closed. This may affect the ability of the INOmax DS or closed) of the cylinder valve.
If there is interference with the INOmax DS will not be displayed and a “Cylinder Not Detected” alarm will activate if there is a set INOMAX dose.
If IR communication interference occurs, we recommend taking the following actions:
• Move the external IR source
• Move the INOmax DS
cart to reduce the external IR source in the area of the INOmeter
IR
• Shield the INOmeter from the suspect IR source
If the actions listed above do not remedy this issue, the transport regulator/cap assembly (PN 10022) may be utilized.
Part No. 20717 Rev-01
2014-07
cart.
IR
to detect the position (open
IR
/INOmeter communication, the cylinder icon on the user screen
IR
cart.
IR
1-15
External Light Interference
Caution:
High frequency and/or high intensity light emission, in the area of the INOmeter, may interfere with communication between the INOmax DS
and the INOmeter on the INOMAX cylinder.
IR
If there is interference with the INOmax DSIR/INOmeter communication, the cylinder icon on the user screen will not be displayed and a “Cylinder Not Detected” alarm will activate if there is a set INOMAX dose.
Test results have demonstrated susceptibility to unintended infrared energy from articial light sources. Most notably, various compact uorescent lighting xtures that focus or reect light, increasing the light intensity in
the vicinity of the INOmax DS
cart, could affect INOmeter communications.
IR
If external light interference occurs, we recommend taking the following actions:
• Move the interfering light source
• Move the INOmax DS
cart to reduce the high intensity light in the area of the INOmeter
IR
• Shield the INOmeter from the suspect light source
If the actions listed above do not remedy this issue, the transport regulator/cap assembly may be utilized.
Figure 1-7
1-16
1
2
Part No. 20717 Rev-01
2014-07
Loading INOMAX Cylinders Onto the INOmax DSIR Cart
Note:
2
Check the INOMAX gas cylinders for the correct product identity labels, cylinder concentration and expiration date.
4
Loading the rst
INOMAX cylinder on the cart will result in a cylinder icon displayed on the screen .
1
2
1
3
Loading a second INOMAX cylinder onto the cart will result in a second cylinder icon displayed on the screen .
3
4
Part No. 20717 Rev-01
2014-07
1-17
INOmeter Operation
Removal of the INOmeter from the cylinder must be performed by authorized personnel. Do not dispose of the INOmeter.
INOmeter battery replacement is only to be performed by authorized personnel.
The INOmeter replaces the standard rubberized cylinder valve handle on the INOMAX cylinders and is used to open and close the cylinder valve. The INOmeter is a time-metric device which records the amount of time the INOMAX cylinder valve is opened.
The INOmeter is designed for use with INOMAX cylinders and with the INOvent, INOmax DS and INOmax DS
When used with INOmax DS INOmax DS and the expiration date to the INOmax DS communicated from the INOmax DS
delivery systems.
IR
, two-way infrared (IR) communication occurs between the
IR
and the INOmeter. The INOmeter communicates the INOMAX cylinder concentration
IR
. Patient ID (when conrmed) and dose information are
IR
to the INOmeter.
IR
Figure 1-8
Note:
• Cylinders are shipped with the INOmeter covered in a tamper-proof seal.
• A valve lock is secured to the cylinder by a lanyard.
• The lock must be removed to open the cylinder valve for use.
1. Remove and properly dispose of tamper-proof seal or covering (see Figure 1-8).
2. The lock is secured to the cylinder by a lanyard (see Figure 1-9).
Figure 1-9
1-18
Part No. 20717 Rev-01
2014-07
Loading...
+ 176 hidden pages