Maico MA 42, MA 41 Operation Manual

Page 1
Operation Manual MA 42
Page 2
Table of Contents
Operation Manual MA 42
Introduction ................................................................................................................. 4
1.1 General ...................................................................................................................... 4
1.2 Intended Use Statement ............................................................................................ 4
1.3 Indications for Use Statement .................................................................................... 4
1.4 Contraindications of Use Statement .......................................................................... 4
1.5 Essential Performance............................................................................................... 5
1.6 Description ................................................................................................................. 5
For your Safety ........................................................................................................... 6
2.1 How to read this Operation Manual ........................................................................... 6
2.2 Customer Responsibility ............................................................................................ 7
2.3 Manufacturer’s Liability .............................................................................................. 7
2.4 Regulatory Symbols .................................................................................................. 8
2.5 General Precautions .................................................................................................. 9
2.6 Electrical Safety and Measuring Security .................................................................. 9
2.7 Device Control ......................................................................................................... 11
2.8 Electromagnetic Compatibility (EMC) ...................................................................... 11
Warranty, Maintenance and After-Sales Service ................................................... 12
3.1 Warranty .................................................................................................................. 12
3.2 Maintenance ............................................................................................................ 12
3.3 Cleaning and Disinfection Recommendations ......................................................... 13
3.4 Accessories and Replacement Parts ....................................................................... 14
3.5 Recycling and Disposal ........................................................................................... 14
Unpacking and Hardware Orientation..................................................................... 15
4.1 Unpacking the System ............................................................................................. 15
4.2 Hardware Orientation............................................................................................... 17
Operating the Device ................................................................................................ 20
5.1 Getting started with the MA 42 ................................................................................ 20
5.2 Functionality of Operating Elements ........................................................................ 20
5.3 The Display of the MA 42 ........................................................................................ 23
5.4 Measurement Methods of Audiometry ..................................................................... 23
5.5 User Menu ............................................................................................................... 40
5.6 Managing Test Results ............................................................................................ 44
5.7 Patient Management................................................................................................ 45
Technical Data .......................................................................................................... 47
6.1 The MA 42 Hardware .............................................................................................. 47
6.2 Connections ............................................................................................................. 49
6.3 Pin Assignment ........................................................................................................ 50
6.4 Calibration Values and Maximum Levels ................................................................. 51
6.5 Electromagnetic Compatibility (EMC) ...................................................................... 58
6.6 Electrical Safety, EMC and Associated Standards .................................................. 60
6.7 Checklist for subjective Audiometer Testing ............................................................ 61
Page 3
Title: MA 42 – Operation Manual Date of issue/last revision: 24/04/2017
Operation Manual MA 42
MAICO Diagnostics GmbH Sickingenstr. 70-71 10553 Berlin Germany Tel.: + 49 30 / 70 71 46-50 Fax: + 49 30 / 70 71 46-99 E-mail: sales@maico.biz Internet: www.maico.biz
MAICO Diagnostics 10393 West 70th Street Eden Prairie, MN 55344 United States Tel.: +1 (888) 941 4201 Fax: +1 (952) 903 4100 E-mail: info@maico-diagnostics.com Internet: www.maico-diagnostics.com
FOR NORTH- AND SOUTH AMERICA
Copyright © 2017 MAICO Diagnostics.
All rights reserved. No part of this publication may be reproduced or transmitted in any form or by any means without the prior written permission of MAICO. The information in this publication is proprietary to MAICO.
Compliance
MAICO Diagnostics is an ISO 13485 certified corporation.
Caution for USA: Federal Law restricts this device to sale by or on the order of a licensed medical professional.
Trademark Notice
QuickSIN® is a U.S. registered trademark of Etymotic Research, Inc.
Page 4
Operation Manual MA 42
Introduction
This section offers you important information about:
the intended use of the device indications and contraindications of use essential performance features and benefits a description of the device
1.1 General
Thank you for purchasing a quality product from MAICO Diagnostics. The Audiometer MA 42 is manufactured to meet all quality and safety requirements, and has been certified with the CE-symbol according to Medical Directive 93/42/EEC.
In designing the MA 42 we placed particular importance on making it a user-friendly device, meaning its operation is simple and easy to understand. All functions of the MA 42 are software controlled, allowing easy upgrades of new features and function in the future.
The user manual should make it as easy as possible for you to become familiar with the functions of the MA 42.
If you have questions or ideas for further improvements, please contact us. Your MAICO Team
1.2 Intended Use Statement
The MA 42 is a portable or stand-alone audiometer intended to be used for the identification of hearing loss and the factors that contribute to the occurrence of the hearing loss in the age range of children to adults. It is intended to be used by audiologists, ENTs, hearing healthcare professionals, or other trained technicians in a hospital, clinic, healthcare facility or other suitable quiet environment as defined in ANSI S3.1/ISO 8253-1 or equivalent.
1.3 Indications for Use Statement
The MA 42 is a portable two-channel air, bone, and speech audiometer. This audiometer is used to screen, test, and diagnose hearing loss.
1.4 Contraindications of Use Statement
The patient is too young, sick or uncooperative to perform the tasks.
Page 5
Operation Manual MA 42
1.5 Essential Performance
The following is considered essential performance:
To generate and present stimulus signals in the audible range as specified in the
applicable IEC 60645 series in normal condition
Record and store a patient response
1.6 Description
The MA 42 is a portable two-channel audiometer with pure tone and speech testing. The device also includes special test function: Master Hearing Aid, Decay (Carhart), SISI, ABLB and MLB. It can be used as a portable audiometer or a desktop unit for ENT diagnostics, hearing aid fittings in the office, and for mobile audiometry in clinics and abroad.
The MA 42 audiometer delivers 11 air conduction (AC) test frequencies from 125 Hz to 8 kHz, with transducer dependent levels from -10 dBHL to 120 dBHL. Bone conduction (BC) can be tested with 10 test frequencies from 250 Hz to 8 kHz with levels from
-10 dBHL to 80 dBHL (with the standard bone conductor B71). As an upgrade option, the MA 42 is also capable of high frequency audiometry up to 16 kHz.
The large back lighted LCD color display shows level, frequency, transducer, signal type, audiograms, and other information for each channel.
The sound presented to the patient’s ear can be monitored by the examiner with an internal monitor speaker or external monitor phone.
The MA 42 performs tests using headphones, bone conduction receiver, and configuration dependent insert phones or speakers. Built-in test signals include pure tone, pulse tone, warble tone, narrow band noise, white noise, and speech noise. Inputs include ports for a live speech microphone and a CD player for speech test material. Speech tests can also be imported via a removable SD memory card. Outputs have separate jacks for air conduction headphones, bone conduction transducer, configuration dependent insert phones and sound field speakers.
Furthermore, the patient management feature provides the ability to store results in the device for further evaluation and documentation.
Results can be printed directly via a USB printer or stored as a PDF file on the included SD memory card or USB flash drive. The MA 42 can be connected to the PC via USB to track the session and store the results in NOAH or the MAICO Database.
In order to remain current with present and future technologies the MA 42 is compatible with PCs, easy to use, extremely reliable, and is adaptable to future developments. It is also designed to be easily serviced as the need arises.
The speaker outputs can also be used as line level outputs for an external amplifier or active speaker. Please, contact your authorized service center to change to line output levels.
Page 6
Operation Manual MA 42
WARNING
For your Safety
This section offers you important information about:
how to read the operation manual where to spend special attention the customer responsibility the explanation of all regulatory symbols used important cautions and warnings that have to be considered during the
whole time handling and operating your device
2.1 How to read this Operation Manual
This Operation Manual contains information pertinent to the use of the MAICO device system including safety information as well as maintenance and cleaning recommendations.
READ THIS ENTIRE MANUAL BEFORE ATTEMPTING TO USE THIS SYSTEM!
Use this device only as described in this manual.
In this manual, the following two labels identify potentially dangerous or destructive conditions and procedures:
CAUTION
NOTE: Notes help you identify areas of possible confusion and avoid potential problems during system operation.
All images and screenshots are only examples and may differ in appearance from the actual device settings.
The WARNING label identifies conditions or practices that may present danger to the patient and/or user.
The CAUTION label identifies conditions or practices that could result in damage to the equipment
Page 7
Operation Manual MA 42
WARNING
2.2 Customer Responsibility
All safety precautions given in this operation manual must be observed at all times. Failure to observe these precautions could result in damage to the equipment and injury to the operator or subject.
The employer should instruct each employee in the recognition and avoidance of unsafe conditions and the regulations applicable to his or her work environment to control or eliminate any hazards or other exposure to illness or injury.
It is understood that safety rules within individual organizations vary. If a conflict exists between the material contained in this manual and the rules of the organization using this instrument, the more stringent rules should take precedence.
This product and its components performs reliably only when operated and maintained in accordance with the instructions contained in this manual, accompanying labels, and/or inserts. A defective product should not be used. Make sure all connections to external accessories are snug and secured properly. Parts which may be broken or missing or are visibly worn, distorted, or contaminated should be replaced immediately with clean, genuine replacement parts manufactured by or available from MAICO.
NOTE: Customer responsibility includes proper maintenance and cleaning of the device (see Sections 3.2 and 3.3). Breach of the customer responsibility can lead to limitations of Manufacturer’s Liability and Warranty (see Sections 2.3 and 3.1).
2.3 Manufacturer’s Liability
Usage of the device in a way deviant from the intended use leads to a limitation or termination of the manufacturer’s liability in case of damage. Improper use includes disregarding the operation manual, the operation of the device by underqualified personnel as well as making unauthorized alterations on the device.
Page 8
Operation Manual MA 42
REP
2.4 Regulatory Symbols
The following Table 1 gives an explanation of the symbols used on the device itself, on the packaging and the accompanying documents including the Operation Manual.
Table 1 Regulatory Symbols
REGULATORY SYMBOLS SYMBOL DESCRIPTION
Serial number
Date of manufacture
Manufacturer
Caution, consult accompanying documents
Warning, consult accompanying documents
Return to authorized representative, special disposal required
Reference number
Patient applied part type B according to IEC 60601-1
Refer to instruction manual (mandatory)
Keep away from rain
Transport and storage temperature range
Transport and storage humidity limitations
Voltage transformer
Electrostatic sensitive devices
Do not reuse
Authorized representative
Conforms to European Medical Device Directive 93/42/EEC
ETL listed mark
Logo
Page 9
WARNING
WARNING
WARNING
WARNING
WARNING
2.5 General Precautions
Before starting a measurement make sure, that the device works properly. Use and store the instrument indoors only. For operation, storage and transport conditions see Table in Section 6 Technical Data. For operation in certain places, a recalibration may be necessary.
No modification of this equipment is allowed. Do not drop or otherwise cause undue impact to this device. If the instrument is dropped or otherwise damaged, return it
to the manufacturer for repair and/or calibration. Do not use the instrument if any damage is suspected.
Calibration of the instrument: The audiometer and the headphone complement each other and share the same serial number (i.e. 7663252). Therefore, the instrument shall not be used with any other headphone prior to recalibration. Recalibration also needs to be conducted, when a defected headphone is replaced. Uncalibrated instruments may lead to faulty measurements and sometimes even damage the hearing of the examinee.
Operation Manual MA 42
2.6 Electrical Safety and Measuring Security
This icon indicates that patient applied parts of the
In Case of Emergency
WARNING
In Case of Emergency
instrument conform to IEC 60601-1 Type B requirements.
In case of emergency, disconnect the instrument from the computer.
In case of emergency, disconnect the instrument from power supply.
Do not position the instrument in a way that it is difficult to operate the disconnection device. The supply mains and the power socket shall be accessible at all times.
Do not use the instrument if the mains cable and/or the outlet is damaged.
To transfer data to a PC, establishing a PC-connection via USB is required. See Section 4.2.4 on how to safely establish a connection with a power supplied PC or laptop (medical device/non-medical device) or to a battery-driven laptop.
Page 10
Operation Manual MA 42
WARNING
WARNING
CAUTION
This equipment is intended to be connected to other equipment thus forming a Medical Electrical System. External equipment intended for connection to signal input, signal output or other connectors shall comply with the relevant product standard e.g. IEC 60950-1 for IT equipment and the IEC 60601-series for medical electrical equipment. In addition, all such combinations – Medical Electrical Systems – shall comply with the safety requirements stated the general standard IEC 60601-1, edition 3, clause 16. Any equipment not complying with the leakage current requirements in IEC 60601-1 shall be kept outside the patient environment i.e. at least 1.5 m from the patient support or shall be supplied via a separation transformer to reduce the leakage currents. Any person who connects external equipment to signal input, signal output or other connectors has formed a Medical Electrical System and is therefore responsible for the system to comply with the requirements. If in doubt, contact qualified medical technician or your local representative. If the instrument is connect to a PC (IT equipment forming a system) ensure not to touch the patient while operating the PC.
If the instrument is connected to a PC (IT equipment forming a system) assembly and modifications shall be evaluated by qualified medical technician according to safety regulations in IEC 60601.
The instrument is not intended for operation in areas with an explosion hazard.
Never short-circuit the terminals.
In order to maintain a high level of safety and to ensure the instrument works properly, it is necessary to have the instrument and its power supply checked according to the medical electrical safety standard IEC 60601-1 by a qualified service technician at least once a year. For more information see Section 3.
The use of non-calibrated devices can lead to incorrect test results and is not advisable.
Prevent cable breakage: cables must not be bend or buckled.
8100460 Rev. 8 10 24/04/2017
Page 11
Operation Manual MA 42
WARNING
2.7 Device Control
The user of the device should perform a subjective device check once a week according ISO 8253-1. See Section 6.7 for a checklist.
For annual calibration please see Sections 2.6 and 3.1.
2.8 Electromagnetic Compatibility (EMC)
Electrostatic discharge (ESD) according to IEC 61000-4-2. Use the device only in an electrostatic controlled environment.
To avoid the risk of electric shock, this equipment must only be connected to supply mains with protective earth.
The instrument fulfills the relevant EMC requirements. Avoid unnecessary exposure to electromagnetic fields, e.g. from mobile phones etc. If the device is used adjacent to other equipment it must be observed that no mutual disturbance appears.
Please also refer to EMC consideration in Section 6.5.
8100460 Rev. 8 11 24/04/2017
Page 12
Operation Manual MA 42
WARNING
Warranty, Maintenance and After-Sales Service
This section offers you important information about:
warranty conditions maintenance cleaning and disinfection recommendations accessory and replacement parts recycling and disposal of the device
3.1 Warranty
The MAICO MA 42 is guaranteed for at least one year. This warranty is extended to the original purchaser of the instrument by MAICO through
the distributor from whom it was purchased and covers defects in material and workmanship for a period at least one year from date of delivery of the instrument to the original purchaser.
The device shall only be repaired and serviced by your distributor or by an authorized service center. Opening the instrument case voids the warranty.
No modification of this equipment is allowed.
In the event of repair during the guarantee period, please enclose evidence of purchase with the instrument.
3.2 Maintenance
In order to ensure that the instrument works properly, it has to be checked and calibrated at least once a year.
The service and calibration must be performed by your dealer or to a service center authorized by MAICO.
When returning the instrument for repairs or calibration it is essential to send the acoustic transducers with the device. Please include a detailed description of faults. In order to prevent damage in transit, please use the original packing when returning the instrument.
8100460 Rev. 8 12 24/04/2017
Page 13
Operation Manual MA 42
WARNING
3.3 Cleaning and Disinfection Recommendations
It is recommended that parts (device and accessories like headphones, ear cushions) which come in direct contact with the patient be subjected to standard cleaning and disinfecting procedure between patients.
Recommendations for cleaning and disinfection of MAICO device presented in this document are not intended to replace or contradict policies in effect or procedures required for infection control at the facility.
If there is not a high infection potential, MAICO recommends:
Before cleaning always switch off and disconnect the device from power supply. For cleaning use a lightly dampened cloth with soap water solution Disinfect the instrument and its accessories by wiping the surfaces with wet
Sani-Cloth® Active wipes or a comparable product and allow them to take effect for the duration related to the specific disinfection aim listed in product data sheet of the disinfection product. Please follow also its instructions for cleaning.
Wipe before and after each patient After contamination After infectious patients
Disinfect computer, keyboard, transport trolley etc. with Sani-Cloth® Active wipes:
once a week after contamination when polluted
CAUTION
CAUTION
To avoid damage of the device and its accessories, please mind the following:
Do not autoclave or sterilize. Do not use the device in the presence of fluid that can come into
contact with any of the electronic components or wiring. Should the user suspect fluids have contacted the system
components or accessories, the unit should not be used until deemed safe by a MAICO certified service technician.
Do not use hard or pointed objects on the device or its accessories Use 70 % isopropyl alcohol only on hard cover surfaces
Discard single-use equipment after use! In case of re-use of the single-use equipment you enhance the risk of cross contamination!
8100460 Rev. 8 13 24/04/2017
Page 14
Operation Manual MA 42
3.4 Accessories and Replacement Parts
Some reusable components are subject to wear with use over time. MAICO recommends that you keep these replacement parts available (as appropriate for your MA 42 device configuration). Ask your authorized local distributor when accessories need to be replaced.
3.5 Recycling and Disposal
Within the European Union it is illegal to dispose of
electric and electronic waste as unsorted municipal waste. According to this, all MAICO products sold after August 13, 2005, are marked with a crossed-out wheeled bin. Within the limits of Article (9) of DIRECTIVE 2002/96/EC on waste electrical and electronic equipment (WEEE), MAICO has changed their sales policy. To avoid additional distribution costs we assign the responsibility for the proper collection and treatment according to legal regulations to our customers.
Non-European countries Outside the European Union, local regulations should be
followed when disposing of the product after its useful life.
8100460 Rev. 8 14 24/04/2017
Page 15
Operation Manual MA 42
Unpacking and Hardware Orientation
This section provides information on:
unpacking the device components becoming familiar with the hardware inclusive connections how to store the device rear panel
4.1 Unpacking the System
Check Box and Contents for Damage
It is recommended that you unpack your MA 42 carefully making sure that all
components are removed from the packing materials.
Verify that all components are included as shown on the packing list included with
your shipment.
If any component is missing, contact your distributor immediately to report the
shortage.
If any component appears to be damaged in shipment, contact your distributor
immediately to report it. Do not attempt to use any component or device that appears to be damaged.
Reporting Imperfections
Notify the carrier immediately if any mechanical damage is noted. This will insure that a proper claim is made. Save all packaging material so the claim adjuster can inspect it as well.
Report Immediately any Faults
Any missing part or malfunction should be reported immediately to the supplier of the device together with the invoice, serial number, and a detailed report of the problem.
Keep Packaging for Future Shipment
Save all the original packing material and the shipping container so the device can be properly packed if it needs to be returned for service or calibration (see Section 3.2).
8100460 Rev. 8 15 24/04/2017
Page 16
Operation Manual MA 42
The availability of configurations with the following components is country specific. Contact your local distributor for more information.
Components
DD45 Headphones * TDH39 Headphones * DD450 Headphones for high frequency audiometry (with High Tone License) * B71 Bone Conductor * B81 Bone Conductor * BKH10 Bone Conductor with contra masking phone * Holmco 8103 Headphones * IP30 Insert Phones * Canton CD220 Free Field Speaker * Cable for Canton CD220 Free Field Speaker Patient Response Switch Carrying Case Gooseneck Microphone USB Cable USB Flash Drive SD Memory Card (2 GB) Sennheiser PC 131 Headset for monitoring and talk forward Stereo Adaptor, 1/8” to 1/4” Monitor Phone for examiner Talk-Back Microphone Sound Room Patch Cords (single)
Sound Room Kit CD Player incl. power supply and connecting cable Operation Manual Quick Guide
* Calibration required
8100460 Rev. 8 16 24/04/2017
Page 17
Operation Manual MA 42
WARNING
4.2 Hardware Orientation
4.2.1 Where to Setup
The MA 42 should be operated in a quiet room, so that the audiometric examinations are not influenced by outside noises. Ambient sound pressure levels in an audiometric test room shall not exceed the values specified in the norm ISO 8253-1:2010 or ANSI S3.1-1999. For use in noisier environments, headphones with optional sound insulation muffs are available.
Electronic devices, which emit strong electromagnetic fields (e.g. microwaves or radiotherapy devices), can influence the function of the audiometer. Therefore, it is not recommended to use these devices in close proximity to the audiometer as it may lead to incorrect test results.
The test room must be at a normal temperature, usually from 15 °C/59 °F to 35 °C/95 °F, and the device should be switched on approximately 10 minutes before the first measurement. If the device has been cooled down (e.g. during transport), please wait until it has warmed to room temperature before using.
External devices such as a computer, printer or Ethernet which are connected to the device must meet electrical safety requirements, such as IEC 60601-1 or UL 60601-1. This is to avoid electrical shock to the user or the patient. See Section 2 for more information
4.2.2 MA 42 Device
The ergonomic design of the MA 42 makes it easy to control the hearing level, signal presentation, and frequency with one hand.
Figure 1
8100460 Rev. 8 17 24/04/2017
Page 18
Operation Manual MA 42
WARNING
11
13
14
16
15
12
9
10
8 7
6 5
4 2
1 3 17
4.2.3 Rear Panel Connections
Figure 2
: Power switch : CD input
: Power socket 100-240VAC/50-60Hz : Speaker left channel
: Network socket : Speaker right channel
: USB out socket : Bone conduction receiver
: USB in socket : Insert phone left channel
: Patient response switch socket : Insert phone right channel
: Talk Back microphone socket : Phone left channel
: Mic - live voice microphone socket : Phone right channel
: Mon - monitor phone output socket
4.2.4 Establishing a Connection to a PC, USB Printer or CD Player
To transfer data to a PC, establishing a PC connection via USB is required. If the MA 42 is used with office equipment that is not a medical device itself (see Table 2, PC Connection 1), make sure to establish the PC connection in one of the following ways (see Table 2, PC Connection 2, 3 or 4).
In case a CD player shall be connected proceed accordingly to PC Connection 3 (if used with a CD player with power supply) or 4 (if used with a battery driven CD player).
Make sure you use only office equipment with the device that is a medical device itself or meets the requirements of IEC 60950. If a non-medical device is used within the patient environment (1.5 m from patient as defined in IEC 60601) a voltage transformer must be used (exception: a battery driven laptop is used).
8100460 Rev. 8 18 24/04/2017
Page 19
Table 2 PC-Connections
PC CONNECTIONS
PC Connection 1: Medical device – Medical device
Operation Manual MA 42
PC Connection 2: Medical device – Non-Medical device
PC Connection 3: Medical device – Non-Medical device
PC Connection 4: Medical device – Laptop (battery-driven)
4.2.5 Storage
When the MA 42 is not in use, store it in a location where it will be safe from damage to the screen or other sensitive components such as the acoustic transducers and cables. Store according to the recommended temperature conditions described in Section 6.1.
8100460 Rev. 8 19 24/04/2017
Page 20
Operation Manual MA 42
Operating the Device
This section offers you information about:
how to get started with the MA 42 the device layout the function keys performing the measurement methods of Audiometry patient management documentation of results changing settings in the user menu
5.1 Getting started with the MA 42
Place the MA 42 on a stable counter or table. Plug the power cord into the power socket on the rear panel. Connect all accessories with the appropriate sockets as shown in Section 4.2.3. Plug the power cord into a grounded outlet.
Turn On: The device can be turned on with the power switch, which is located on the rear panel of the MA 42.
NOTE: A device should always be off when inserting or removing an accessory from the rear panel connectors.
The device performs its initialization and boot up. Please wait until the test screen appears, this can take up to 90 seconds. If an error is detected the startup is stopped and a description of the error is shown on the display. In this case please contact your local dealer for service.
Turn Off: The operation of the device can be safely terminated by using the power switch found on the rear of the device or by disconnecting the power cable.
5.2 Functionality of Operating Elements
The main functions of the MA 42 are directly accessible by using the function buttons which are located around the display (see Figure 3). As the buttons have changing functionality, the actual function of each button is shown in the blue boxes on the screen above the button.
8100460 Rev. 8 20 24/04/2017
Page 21
Operation Manual MA 42
The following table describes the main functions of each button for the tone and speech audiometry screen:
Figure 3
(1) Level control: adjusts the hearing level for the left/right ear. (2) STIM bar: presents or interrupts the signal for the left/right ear. (3) STORE button: stores results for the left/right ear. (4) Frequency Up: changes to higher frequency for tone audiometry, enters a correct
answer for Word Recognition Score (WRS) testing, or selects the next word in the word list for Speech Recognition Threshold (SRT) testing with Wave files.
(5) Frequency Down: changes to lower frequency for tone audiometry, enters an
incorrect answer for WRS testing, or selects a previous word in the word list for SRT testing with Wave files.
(6) Function button: Monitor with options to adjust monitor and talk back settings;
for speech, the input calibration for microphone or CD player can be adjusted. (7) Function button: Function is displayed on the screen based on test screen: Tone: Select New, to delete all stored results and start a new session. Speech: Reset the result percentage counter or play Wave file. (8) Channel STIM Mode/TALK button: to change from presenter to interrupter mode,
or to talk to the patient by pressing and holding the button down
(only one STIM/TALK button should be pressed for Talk Forward capability)
8100460 Rev. 8 21 24/04/2017
Page 22
Operation Manual MA 42
(9) Function button: to select left, right or swap (10) Transducer selector button: to choose between Phones, Insert, Bone and Speaker
(only calibrated transducers are available) (11) Function button: Function is displayed on the screen based on test screen:
Tone: signal selector, to choose between tone, narrow band noise, white noise,
or to switch the channel off
Speech: Select microphone, external CD player, Wave file or speech noise
as signal source. (12) Test Signal selector button: Steady, Pulse, Warble, or P&W (pulse and warble tone) (13) Function button: Function is displayed on the screen based on test screen:
Tone: Select test for selected receiver, either the pure tone threshold, Hearing
Level (HL), Most Comfortable Loudness (MCL), Uncomfortable Loudness (UCL),
or Over Threshold (OT) tests.
Speech: Select Speech Recognition Threshold (SRT), Word Recognition Score
(WRS), Master Hearing Aid (MHA), Most Comfortable Level (MCL),
or Uncomfortable Level (UCL). (14) Select Unlock: Lock (locks the presentation of the signal in both channels), Track
(activates the masking noise to automatically increase and decrease level in
relationship to the signal), L&T (Lock and Track) (15) Function button: Function is displayed on the screen:
Tone: signal selector, to choose between tone, narrow band noise, white noise,
or to switch the channel off
Speech: Select microphone, external CD player, Wave file or speech noise
as signal source. (16) Transducer selector button: to choose between Phones, Insert, Bone and Speaker
(only calibrated receivers are available for selection) (17) Function button: to select left, right, or swap (18) Function button: to switch from tone to speech and back (19) Function button - Menu: To enter the user menu, where settings can be adjusted,
results can be printed out sor stored as PDF on SD memory card or USB flash
drive or the patient list can be entered. (20) SD memory card slot (21) VU-Level meter
NOTE: When a Function button is grey, this identifies it as an inactive button with the current test set-up (E.g. ).
8100460 Rev. 8 22 24/04/2017
Page 23
Operation Manual MA 42
5.3 The Display of the MA 42
The device is set to air conduction and pure tone by default. When starting the MA 42 the tone audiometry screen opens.
Figure 4
The default frequency is set to 1 kHz and the level of the tone is set to 30 dBHL for the right channel, and 0 dBHL noise for the left. The level limits of the chosen transducer are marked by gray shading in the lower area of the audiogram.
NOTE: Once the default settings are changed, the device applies the new settings. Reset the device to return to the default settings.
The display has an energy saving function; the backlight of the display is automatically dimmed after approximately three minutes. Any action with the MA 42, such as pressing a button or rotating the dial, immediately illuminates the backlight.
5.4 Measurement Methods of Audiometry
The patient should sit at a distance of at least 1 m from the device. Remove any obstructions that interferes with the earphone cushion placement over the
ear (e.g. hats, eyeglasses). Always use the headphones with appropriate padded ear cushions.
Ensure the headphones are placed correctly over the patient’s ears and that the red side is over the right ear and the blue side is over the left ear. Adjust the headband of the headphones so that the receivers are at the correct height (the sound output grid on the inside of the headphone should be directly over the ear canal).
Ask the patient to press the button on the patient response switch when the tone is heard, even if it is barely audible.
For hygienic reasons it is important to clean and disinfect the ear cushions on the headphone between patients (see Section 9).
8100460 Rev. 8 23 24/04/2017
Page 24
Operation Manual MA 42
5.4.1 Tone Audiometry
The MA 42 supports tone audiometric testing methods. The following testing methods can be started in the tone audiometry mode and the results can be saved to the device.
Air conduction testing Bone conduction testing High tone audiometry (configuration dependent) Sound field testing Uncomfortable Loudness (UCL) Most Comfortable Loudness (MCL) Over threshold test: Decay, SISI, ABLB, and MLB
5.4.1.1 Pure Tone Testing
During pure tone audiometry, the patient’s hearing threshold is measured in comparison with the normal hearing threshold. The examination starts with air conduction on the ear with better hearing.
Figure 5
The default setting: right channel is set to air conduction pure tone and the left channel to air conduction narrow band noise with manual masking. The frequency is automatically set to 1000 Hz.
The audiometer provides two channels, which can be separately routed to the left or right ear. Each channel can play back a signal and can be routed to one of the outputs. The signal can be set for each channel independently to tone, narrow band noise or white noise by the Function buttons (11) and (15).
Select the ear to be tested by pressing the Right or Left key (9), (17) on the control panel underneath the screen.
Next, select the transducer to be used, headphones (Phones), insert phones (Insert), bone conductor (Bone), or sound field speaker (Speaker) by pressing the appropriate button (10), (16). Only calibrated transducers are selectable.
The level and frequency is displayed as a numerical value at the top of the screen and is also indicated by the cursor within the audiogram.
8100460 Rev. 8 24 24/04/2017
Page 25
Operation Manual MA 42
The dBHL level can be changed with the Level control dial on both sides (1) of the device. Set the Level control dial to a value just below expected hearing loss and turn down the level of the other ear.
Use the Frequency Up (4) or Down (5) keys to increase or decrease the frequency. Press the STIM button (the blue part next to the Level control dial) to present or interrupt the tone. The status LED for the Stimulus Mode button (8) illuminates when the tone is presented.
Follow your preferred procedure for the hearing threshold evaluation.
NOTE: A warning prompt appears on the display in the event that the hearing level exceeds 100 dBHL. The warning prompt disappears after approximately 3 seconds. As long as the prompt is visible on the display, no further entries can be made.
Test the frequencies: Starting at 1,000 Hz, test the higher frequencies first, then the lower frequencies.
Use the Frequency Up key (4) to select the next higher frequency and use the Frequency Down key (5) to select the next lower frequency.
Once a threshold value is established at the desired frequency, press the Store button (3) to store the threshold. The appropriate symbol plots in the audiogram on the display.
To delete a stored test point, select the corresponding frequency and then press both frequency down keys (5) at the same time. The symbol of the stored test point disappears from the audiogram. Make sure, that the correct transducer and test type (HL, MCL or UCL) is selected before deleting.
If the patient is not responding, the current value can be marked as “no response” by holding the store button pressed for 3 seconds. Once all frequencies are tested, select the poorer ear and repeat the hearing threshold test.
Pulse Tone
If required, the test can also be performed with a pulsed tone. Set the Test Signal button (12) on Pulse and the pure tone is switched to a pulsating tone.
Warble Tone
If required, the test can also be performed with a warble (frequency modulated) tone. Press the Test Signal button (12) and the pure tone is frequency modulated. The warble tone can also be pulsed as described above.
Noise
The signal may be changed from a pure tone to narrow-band noise or white noise by pressing the signal key (11), (15) underneath the display, whichever one corresponds to the current test channel. This changes the status displayed in the test channel from “Tone” to “Noise.”
8100460 Rev. 8 25 24/04/2017
Page 26
Operation Manual MA 42
5.4.1.2 Masking
Masking is required if there is a notable threshold difference between the left and right ears. It is possible for sound to be transmitted to both ears via bone conduction while testing the poorer ear. This is called “crossover.”
Crossover occurs often while testing bone conduction, but it can also occur during air conduction testing. Relevant to crossover is the sound level received by the opposite ear. The difference between the original test signal in the test ear and the received signal in the opposite ear is called “interaural attenuation”.
For bone conduction measurements the interaural attenuation is 0 to 15 dB. Bone conduction crossover is therefore possible even with a slight difference in hearing loss between ears.
To ensure that the patient does not experience crossover, mask the opposite ear. Masking increases the hearing threshold of the opposite ear when crossover is occurring. For bone conduction the masking signal is automatically routed to the opposite output of the phones or inserts.
The masking is done with a noise signal which is transmitted by the headphone. For pure tone audiometry a narrowband noise is used. This noise changes its center frequency according to the frequency of the test signal.
Manual Masking
The masking is switched on by selecting the signal source as noise (11) or (15) and then pressing the STIM button (8) on the side presenting the noise.
NOTE: Be cautious to set the noise to an appropriate level prior to selecting STIM (8) and turning on the masking noise to the patient.
Once the threshold has been established with the introduction of noise, press the Store button on either side of the device. The hearing threshold value is stored in the audiogram with the corresponding masking symbol.
The masking sound should be continuously presented for effective masking by pressing the STIM button (2). You can interrupt or turn off the masking signal by pressing the STIM button (2) again.
Automatic Masking
With the manual masking, as described above, the masking level should be adjusted every time you change the test signal level. The MA 42 has a tracking function for easy masking.
Set the tone level and the masking level to a desired difference for effective masking. Press the TRACK button (14) to implement the automatic masking feature. The masking level is automatically changed if you adjust the test signal level (e.g. if the test level is at 30 dBHL and the masking level 50 dBHL and you change the test level to 45 dBHL the masking level adjusts automatically to 65 dBHL).
8100460 Rev. 8 26 24/04/2017
Page 27
Operation Manual MA 42
5.4.1.3 High Tone Audiometry (configuration dependent)
When available, pure tone hearing threshold of high frequencies from 8 kHz to 16 kHz can be tested. The High Tone option must be activated and a separate high tone transducer must be calibrated for all frequencies.
First select the corresponding output Phone or Insert by the button (10) or (16). The item High Tone is now available in the Function button (13). Select High Tone by pressing the Function button (13) until High Tone is highlighted. The High Tone diagram is displayed, combining the left and right ears results. Additional frequencies from 8 kHz to 16 kHz can be selected now by the Frequency Up (4) or Down (5) key. The test procedure is identical to the description in Section 5.4.1.1.
If a transducer is selected which is not calibrated for high frequencies, the standard diagram is displayed and high frequencies are not available.
On the print out, pure tone audiometry results appear in a high tone diagram combining left and right ear results.
5.4.1.4 Bone Conduction Testing
Bone conduction - the transmission of sound waves through the skull directly to the inner ear - provides information about the function of the inner ear. For a neural hearing loss the values of air conduction and bone conduction are the same. In this case a hearing loss of the middle ear can be eliminated.
Place the bone conduction oscillator so that the flat, circular side of the transducer is placed on the mastoid, at the noticeable ledge of the cranial bone behind but not touching the pinna. The other side of the headband is placed in front of the opposite ear. Set the receiver selector to Bone (10) or (16) and the signal selector to Tone (11) or (15).
Perform the test in the same way as described in the air conduction section above. For hygienic reasons it is important to clean and disinfect the bone conduction oscillator
after each patient (see Section 3.3).
5.4.1.5 Sound Field Testing (configuration dependent)
Set the transducer selector (10) or (16) to Speaker. Perform the test in the same way as described in the air conduction section above.
Warble tones should be used in the sound field as pure tones may provide inaccurate results in the typical test room.
When results are stored using sound field an ‘S’ is displayed on the audiogram. This signals sound field/speakers were used for testing and ear specific information cannot be determined.
Results may also be displayed with an ‘A’ on the tone screen, when speaker is selected as the transducer and Aided is selected from the Function button (13).
8100460 Rev. 8 27 24/04/2017
Page 28
Operation Manual MA 42
WARNING
5.4.1.6 Uncomfortable Loudness (UCL) Testing
Testing of UCL can be measured using pure tone or speech stimuli. The purpose is to determine the dBHL level at which the stimuli becomes uncomfortable to the patient. The UCL is described as the level between very loud and too loud as perceived by the patient when listening to the test signal. This information is valuable for determining the limits of a patient's dynamic range.
Because this test uses high sound pressure levels, it is extremely important to perform this test using the utmost caution to avoid causing hearing loss. To prevent the possibility of extreme discomfort by the patient, it is important to start the test at a comfortable level.
Press and release the test mode selector key (13) below the display to select UCL. The LCD display in the bottom row changes from HL to UCL. Start with a test level of 60 dB and present the tone briefly (max. 1 s.). If the signal was recognized by the patient as “not uncomfortable,” increase the level and proceed as described before. If the signal was uncomfortable for the patient store the value. Proceed accordingly with other test frequencies.
HL
5.4.1.7 Most Comfortable Loudness (MCL)
Testing of MCL can be measured using pure tone stimuli or speech. The purpose is to determine the most comfortable listening level for the patient for a given stimulus. The dB level at which the stimulus is the most comfortable is determined. This level might be described as the level at which the patient would be comfortable listening for an extended period of time. Press the test mode selector key (13) underneath the display and select MCL in order to test and store the Most Comfortable Loudness.
5.4.1.8 The SISI Test
The SISI (Short Increment Sensitivity Index) is guided by the principle that patients with cochlear impairment are hypersensitive to small intensity increments. The continuous test tone is increased by 1 dB for a period of 0.2 seconds every 4.8 seconds. Whenever the increment is heard the patient has to press the response switch. The MA 42 calculates the percentage of detected increments. The SISI test information and the test score are shown on the LCD display. The test ends automatically after 20 presented increments. The score is expressed as a percentage of ratio of the increments heard to the delivered increments (all increments heard = 100 % and no increments heard = 0 %). A high score indicates a cochlear impairment. A low score is related to normal hearing or conductive or retrocochlear pathology.
8100460 Rev. 8 28 24/04/2017
Page 29
Operation Manual MA 42
5.4.1.8.1 Preparation of the SISI Test
To start the test select SISI with the Function button (13), choose OT Tests and SISI Test.
Figure 6
Select the test frequency with the frequency keys (4), (5). For the test you should choose the frequency where the maximum bone conduction hearing loss is measured as described in Section 6.1.3.
Adjust the test intensity with the Level control dial (1) to a value of 20 dB above the individual hearing threshold.
Recognize that in the SISI mode the test is presented continuously. You can interrupt the test by pressing the Pause (9) button. Press Start (19) to resume the test.
5.4.1.8.2 Training of the Patient
Only with careful training during the following conditioning phase can a valid test result can be achieved.
The training starts with level steps of 5 dB. That means a continuous tone, e.g. 65 dBHL, is briefly increased by 5 dB, e.g. to 70 dBHL. During the increment time, a green box is shown in the middle of the LCD display. The number of detected (recognized) increments is counted and shown on the display. Furthermore, the number of presented (given) increments is shown. Instruct the patient to press the patient response button when they hear a change in loudness from the continuous presentation tone. The response button must be pressed while the green box is displayed on the screen and thus the tone is being presented to the patient for this to be counted.
When the patient has understood the procedure of the test press the “Inc” button (11) to select the increment level for the test (i.e. 1, 2, 3, 4, or 5 dB increments). Once selected the increment level is displayed at the top of the screen below the frequency.
8100460 Rev. 8 29 24/04/2017
Page 30
Operation Manual MA 42
5.4.1.8.3 Performing the SISI Test
When the patient is ready, set the increment to 1 dB and press the Start (19) button to run the SISI test.
Now 20 increments are presented to the patient. You can delay the test by pressing the Pause (9) button. This could be helpful in
determining if the patient malingers. The test ends automatically after 20 presented increments. The number of detected
increments is counted and indicated as percentage of the presented increments. The result is only stored if a minimum 10 increments has been presented.
If you are sure that the result of the test is consistent you can abort the test at any time by pressing the Stop (18) button. The result is then indicated accordingly.
Retrocochlear hearing impairment is represented by low scores (< 25 %), cochlear defects by high scores (> 70 %). Press the EXIT (6) button to leave the SISI test.
5.4.1.9 The Threshold Decay (Carhart) Test
This examination is conducted as a fatigue test in cases of sensitivity loss. To start the test select Decay with the Function button (13), choose OT Tests. When the
screen advances, select Decay Test. Make sure that the correct transducer is selected in advance as you cannot change transducers in this screen.
Select left or right ear (9) and the test frequency with the frequency keys (4), (5).
Figure 7
Adjust the test level with the intensity control dial (1). The level should be set to a value 5 dB above the individual pure tone hearing threshold at this frequency.
Instruct the patient to hold the response switch while the tone is heard and to release the switch as soon as the tone is no longer heard.
Press Start (19) to begin the test and present the test signal to the patient.
8100460 Rev. 8 30 24/04/2017
Page 31
Operation Manual MA 42
The patient response LED lights and a stopwatch counts the seconds as long as the patient response switch is pressed. As soon as the patient releases the response switch, the level is increased for 5 dB.
The test stops automatically if the patient is able to hear the tone at the same level for 60 seconds or if the level is increased for 30 dB. Press Stop (18) to terminate the test manually.
5.4.1.10 ABLB
The ABLB (Alternate Binaural Loudness Balance) test is a recruitment test. Alternating tones between the two ears with headphones are used. The intensity of the tone in the better hearing ear is varied until the loudness of the tone is judged by the patient to be equivalent in both ears. This is called loudness balance.
To start the test select OT Test with the Function button (13), choose ABLB, then the ABLB test screen is displayed:
Figure 8
The test can be performed at all frequencies between 125 Hz and 8 kHz. Select the test frequency by the Frequency Up (4) and Down (5) key. Set the level of the better hearing ear 5 dB above the threshold. Present the left or right STIM button (2). Adjust the level of the hearing-impaired ear with the Level control (1) until the loudness in both ears matches. Press the left or right Store button (3). Repeat the test for several levels.
To perform the test at another frequency, press the Frequency Up (4) or Down (5) key. A new ladder diagram appears. Use the Function button (9) or (11) to jump to the previous or next diagram. Select 500 ms or 1000 ms as modulation time by the Function button (13). To exit the ABLB test screen press the EXIT button (6).
8100460 Rev. 8 31 24/04/2017
Page 32
Operation Manual MA 42
5.4.1.11 MLB
The MLB (Monaural Level Balance) test is used to detect perceived loudness differences between two frequencies in the same ear. A tone with two alternating frequencies is presented to the same ear. The level of the frequency with normal hearing threshold is adjusted until it matches the loudness of the other frequency. The patient has to indicate if the loudness of both frequencies is equal.
To start the test select OT Tests with the Function button (13), choose MLB (15), then the MLB test screen is displayed:
Figure 9
Select the left or right ear by the Function button (9). Then select left and/or right Frequency Up (5) or Down (5) buttons for selection of the frequencies to test. Adjust the level by the Level controls (1). Present the stimulus using the STIM button (2). Press the Store button (3) to save the level of equal loudness. Repeat the procedure for several different levels.
To compare the loudness of further frequencies press the Frequency Up (4) or Down (5) button, a new ladder diagram appears. Use the Function button (9) or (11) to jump to the previous or next diagram. Select 500 ms or 1000 ms as modulation time by the Function button (13). To exit the MLB test screen press the Exit button (6).
8100460 Rev. 8 32 24/04/2017
Page 33
Operation Manual MA 42
5.4.1.12 Stenger
The Stenger test is used to confirm the presence of pseudohypacusis. During this test, two tones of the same frequency are presented simultaneously to both ears, and only the louder tone is perceived. Select HL to perform the Stenger test and setup the audiometer to present a tone in both ears (9, 11, 15, 17). Instruct the patient to press the response button when the tone is heard. Present a tone to the better hearing ear 10 dB above threshold and wait for the patient to indicate the tone has been heard. Now present the tone to the poorer ear 10 dB below the indicated threshold (the patient may “ignore” this tone). Present the tones simultaneously by pressing the lock Function button (14) and the STIM Mode button (8) to set it to interrupter mode. If the patient responds it is a negative. If the patient does not respond it is a positive Stenger, indicating that the tone is heard in the poorer ear and the patient is ignoring the stimulus.
5.4.1.13 Langenbeck
The Langenbeck noise audiometry investigates the hearing threshold of pure tones while masking noise is presented simultaneously to the same ear with a constant level for all frequencies. The tone in noise hearing threshold is typically as high as the level of the noise. The noise has a higher influence to neural hearing losses and the tone in noise hearing threshold is increased. Set the signal of the first channel to tone and the signal of the second channel to noise and route both channels to the same ear. Set the level of the noise to a fix value and test the hearing threshold.
5.4.2 Speech Audiometry
The MA 42 supports speech audiometry. To conduct speech tests using speech test material you can use a CD player, Wave files from the SD memory card, or a microphone.
If you are using a CD player powered by electrical current, the player must meet electrical safety requirements, such as IEC
CAUTION
60601-1 or UL 60601-1. This is to avoid electrical shock of either the patient or you. If you are not sure if your player meets these requirements it is safer to use battery power.
5.4.2.1 Input Calibration
The MA 42 must be calibrated to the particular speech test to ensure valid test levels. That means every time you change the speech test CD you must recalibrate the device.
To calibrate the CD speech input, select CD with the signal selector key (11), (15). Press the Monitor button (6) and then InCal (19) and the calibration screen appears (see Input Calibration screen in Figure 11).
8100460 Rev. 8 33 24/04/2017
Page 34
Operation Manual MA 42
Figure 10
Figure 11
On every CD with speech test material there is a reference signal, such as a reference tone or speech simulating noise. Play back the reference signal from the CD. Use the left or right Level control (1) and adjust the levels until the VU-meter (19) shows all yellow lights and one green light.
If one or more red lights are on, reduce the level using the Level control dial (1). To calibrate the microphone for live voice testing, select MIC with the Signal selector key
(11), (15). Press the Monitor button (6) and then InCal (19) and the calibration screen appears. Use the left or right Level control (1) and adjust the levels until the VU-meter (21) shows all yellow and one green light.
NOTE: When multiple individuals use the device, input calibration is required when switching to a new user. The device stores that last input calibration for ease of use.
Store the calibration and leave the calibration mode by pressing the OK button on the left side of the display.
The Wave files included with the device have been pre-calibrated and no input calibration is required.
5.4.2.2 Performing Speech Testing
Use the Function button for Speech on the right side of the tone screen (18) to switch to speech testing. The speech test screen opens, the right ear is selected, and the level is set to the default value.
The Speech Recognition Threshold (SRT) is the hearing threshold for speech. It is the lowest level at which the patient correctly recognizes 50 % of the speech. Speech testing can be done via recorded speech test material from CD, Wave files, or with the microphone and live voice using standardized word lists.
Ask the patient to repeat each word. Often times a carrier phrase such as, “Say the word _______” may be used. The patient should sit at a distance of at least 1 meter from the device. Additionally, any obstructions which may interfere with the placement of the earphone cushions on the ear (i.e. hair, eyeglasses) should be removed.
8100460 Rev. 8 34 24/04/2017
Page 35
Operation Manual MA 42
Figure 12
Ensure the headphones are put on correctly. Adjust the headband of the headphones so that the receivers are at the correct height (the sound output grid should be placed directly over the ear canal).
NOTE: When the patient is sitting in the same room as the tester, CD or Wave files is recommended for stimulus presentation.
The signal can be set to microphone, CD, Wave file, Speech Noise (SN), or switched off, for each channel independently by the Function buttons (11) and (15). Each channel can be routed to the left or right ear by the Function buttons (9) or (17), this enables flexible mixing.
Select Phones, Insert, Bone, or Speaker (only calibrated receivers are available) by the receiver Select button (10), (16). Toggle to the required signal by the Selector button (11) or (15) to MIC/CD/Wave/SN.
5.4.2.3 Speech Audiometry with Microphone or CD player
Connect the microphone or CD player to the corresponding input on the rear side of your MA 42. Set the signal of the test channel to microphone or CD by the Function button (11). Make sure that the input signal is calibrated correctly, as described above. Select the test SRT, WRS, MCL, MHA or UCL by the Function button (13).
For the SRT test familiarize the patient with a closed set of spondaic words at a level loud enough for them to hear and understand. Begin the test and decrease the level as the patient repeats the word. Once the threshold has been found press the Store button (3) to save the result.
For the WRS test, the level remains fixed and correct or incorrect answers can be entered by the Frequency Plus (4) and Frequency Minus (5) button. Once the word list has been completed by the patient, press the Store button to save the established WRS score. To clear the word counter, press the Reset key (7) on the left side of the display. If no counter is set the Reset key (7) is displayed grey.
If MCL is selected, the level is increased until the patient indicates the level is comfortable to listen to. Store the value by pressing the Store button (3).
8100460 Rev. 8 35 24/04/2017
Page 36
Operation Manual MA 42
For the UCL testing the level is increased until the patient indicates the level is uncomfortable and the result can be saved by pressing the Store button (3).
To delete a stored test result, select the corresponding level, then press both frequency down keys (5) at the same time. The test result disappears from the speech audiogram or table. Make sure, that the correct transducer and test type (SRT, WRS, MCL or UCL) is selected before deleting.
5.4.2.4 Speech Audiometry with Wave Files
If Wave is selected by the Speech Signal selector (11) or (15), a menu pops up with available word lists which are stored on the SD memory card. A word list can be selected by using the Level control (1) to scroll through the list. A list can be loaded by pressing the Stimulus button (2). The word list is displayed on the speech audiometry screen.
The level is displayed as a numerical value at the top of the screen. The level can be changed with the Level controls (1) on both sides of the device. Before starting the playback of the Wave files, the first word can be selected by the Frequency Up (4) and Down (5) buttons. Press the Function button Play (7) to start. Once the speech test is started the test can only be stopped by pressing the Cancel (18) or Store (3).
Figure 13
The procedure for the SRT, WRS, MCL, and UCL test is the same as the procedure for CD or microphone testing. For the SRT test, a word needs to be selected in the word list by the Frequency up (4) or Down (5) key. When the Play button (7) is pressed, the selected word is presented. Increase or decrease the level when the patient has a correct or incorrect response until the SRT is estimated and store the result by pressing the Store button (3).
For the WRS test score, tally the correct words by pressing the Frequency Up (4) key and the incorrect words by pressing the Frequency Down (5) key. The next word is played back automatically. The correct word is displayed green, while incorrect word is displayed red. By pressing Repeat (7), the word is repeated. Once a Wave file has been selected, the intensity cannot be changed. Select the presentation level before selecting Play (7).
The percentage of the speech discrimination score is displayed and stored in the speech table or audiogram as soon as the Store button (3) is pressed.
8100460 Rev. 8 36 24/04/2017
Page 37
Operation Manual MA 42
Press the Function button List (6) to load another word list. MCL and UCL measurements are best performed with the Passage Wave file included
with the device. Select the ear for testing and Wave as the speech signal. A menu pops up with available word lists which are stored on the SD memory card, select Passages. Press the Function button Play (7) to start the presentation. Rotate the Level control (1) to increase the volume until the patient reports this is uncomfortably loud. Press Store (3) to save the results.
5.4.2.5 Masking
Speech audiometry masking can be used in the same manner as described in Section
5.4.1.1 for pure tone audiometry. The signal of the non-testing ear is set to speech noise (SN), instead of narrowband noise by the Function button (11) or (15). This channel needs to be set to interrupter mode by pressing the channel STIM button (8) for continuous presentation of the masking noise. Adjust the level of the masking channel by the corresponding Level control (1) for effective masking.
5.4.2.6 Speech in Noise Tests - QuickSIN® (configuration dependent)
When performing speech in noise tests, speech and noise are given simultaneously in two channels but in the same ear. The speech source can be provided by a microphone, Wave files or an external CD player. Select the signal source on one channel (button 11 or 13) and speech noise (SN) on the other channel.
To present the speech in noise in the same ear, the noise channel has to be routed to the same ear the speech channel (button 9 or 17). Please make sure that the speech is presented in interrupter mode (8) in order to present both signals simultaneously.
Speech-in-noise test, the QuickSIN®, was developed to provide a quick estimate of signal-to-noise ratio (SNR) loss. Configuration dependent this test is incorporated into the MA 42, with selection made in the speech testing screen under the Wave file lists (11) or (15). For more information refer to Etymotic Research’s QuickSIN® Speech-in-Noise Test manual. MA 42 does not incorporate the entire QuickSIN® lists but provides Standard QuickSIN® lists 1-12 (i.e. Track 3-14 from the CD recording).
NOTE: Pre-calibration of the QuickSIN® lists was implemented into the recording and installation of the Wave files on the MA 42. No input calibration is required.
8100460 Rev. 8 37 24/04/2017
Page 38
Operation Manual MA 42
5.4.2.7 Master Hearing Aid (MHA)
After you have completed the speech audiometry testing you may wish to demonstrate to the patient the difference a hearing aid can make. Use the MA 42 Master Hearing Aid (MHA) to adjust various frequency slopes for speech signals.
Figure 14
The MHA feature utilizes input signals from the Live Voice (Mic), external CD/MP3 player or Wave files. These signals are then filtered to simulate the benefit of hearing aids in the case of a high frequency hearing loss. The level of the low frequencies can be reduced by 6 dB, 12 dB, and 18 dB per octave.
Begin by selecting a signal source in the speech mode with the Function button (11) or (15). Start the MHA function by pressing the Test Selection button (13). When using Wave files begin testing by pressing the Play button (7) to play back the imbedded Wave files. Press the Stimulus mode (11) or (15) indicator to alternate between signal sources like the microphone or the CD. Change the dB presentation levels with the Level control dials (1). The influence of the filters can be changed by using the Frequency Up (4) and Down (5) keys. The MHA function can be performed using monaural or binaural presentations.
8100460 Rev. 8 38 24/04/2017
Page 39
Operation Manual MA 42
5.4.3 Monitoring
All signals presented to the patient can be monitored by the examiner via a monitoring headset or the internal speakers. For this purpose, press the Monitor button (6) and the monitor screen appears. Then the monitor level of the left and right channel can be adjusted with the corresponding Level control dial (1). Enable monitoring by pressing the Function button Monitor (9), while Mic or CD is selected. Monitoring by the integrated speaker is switched on by pressing the button (13) and the external headset is activated for monitoring via button (11). In order to hear the signal given to the patient, make sure to activate the corresponding channel for monitoring (9 and 17).
Figure 15
The talk back microphone is activated by the button (15) and its level adjusted by the Frequency Up (4) and Down buttons (5).
5.4.4 Talk Forward
Connect the microphone headset (or optional gooseneck microphone) to the microphone socket (8) on the rear side of the device.
NOTE: Only connect the microphone when the unit is off.
To talk to the patient press and hold the STIM/TALK button (8) and speak into the microphone. Pressing only one STIM/TALK button is required to activate this feature. Adjust the level by turning the left or right Level control (1) while pressing and holding the STIM/TALK button.
8100460 Rev. 8 39 24/04/2017
Page 40
Operation Manual MA 42
5.5 User Menu
The User Menu enables the user to customize the device to meet their specific needs. Additionally, the menu allows the user to print out the results via USB printer, store the results as a PDF on an SD memory card or USB flash drive, and the ability to enter the patient list. To enter the User Menu press the Menu button (19) on the right side of the display.
Figure 16
To choose an item from the menu use the Level control dials (1). A short description of the selected setting item are displayed below the user menu list.
To display the sub items or change the setting of the selected item press the STIM Presenter button (2).
To confirm the changes press the OK button (7) on the left side of the display, or press Cancel (6) to return without any change.
NOTE: When changes have been made to the menu and saved, restart the device to confirm all changes are implemented.
8100460 Rev. 8 40 24/04/2017
Page 41
Back
: Frequency control jumps to
Level
Steps
5 dB
;
2 dB; 1 dB
Change effect on dialing an
D
efines what happens if Swap
Pulse
500 ms, slow pulsing
250 ms, fast pulsing
These menu items are available:
Operation Manual MA 42
Setup Menu
Frequency Settings Default Frequency
Frequency Roll
Standard Frequencies
Level Default Level
Inverse Dialing
Speech Level Unit
Signal Default Signal Types
Swap Function
Display of the Left
and Right ear
Presenter Duration
Interrupter/Presenter
Description
Default Frequency Set (On/Off): set default frequency to 1,000 Hz if side, transducer or signal type has changed.
1,000 Hz if the highest and lowest frequency was reached. Stop: Frequency control function stops at highest and lowest frequency. Wrap: Jumps to the lowest/highest frequency when the highest/lowest frequency is reached.
Select/Deselect lists of available frequencies in device.
On/Off: Set default level after changing signal type. Set Default Levels: Set the default level for tone, noise and speech signals.
encoder: This changes the direction of rotation of the dial required for increasing or decreasing the level of the signal (Dial Up; Dial Down).
dBSPL/dBHL: Select the level unit for speech signals.
defines which signals are selected by default: Signal/Noise or Signal/Signal.
is selected: Swap Ears, the left and right ear is interchanged, the signals of the channels remain. Swap signals, only the signals are interchanged and the ears remain.
Same, left dial controls level of the left ear, right dial controls the level of the right ear Interchanged, left and right dial controls the level of the opposite ear
Unlimited, signal is presented as long as the STIM bar is pressed:
1.5 seconds, signal is switched off after 1.5 seconds. User defined duration, user can define a maximum presentation duration.
Default Presenter or Interrupter mode
8100460 Rev. 8 41 24/04/2017
Page 42
Operation Manual MA 42
Change the level after storing a
(10, 20, or 30
dB).
Stores PDF files to a USB flash drive
Change the display brightness
by pressing the Store button (3)
.
Setup Menu
Noise Interrupter/Presenter Pure Tone Average
(PTA)
CD Channel Select the CD Channel Store Properties
Change Frequency after Store
Change Level after Store
Default Transducer
Selection of Default Transducer
Monitoring Monitoring Tests Start Test
Printer Settings Set Printer Settings
PDF Save Option Save on SD Card
Save on USB Flash Drive
Device Information Show Information Date/Time Set Date/Time
Display Brightness
Description
Default presenter or interrupter mode Select/Deselect frequencies for the
calculation of the PTA value for the default transducer (500, 1000, 2000).
Both / Channel A / Channel B Next: Move to next test frequency
after storing a threshold. Remain: Stay on same frequency after storing (Remain).
threshold. Stay at the same test level (0 dB) or increase/decrease by
Headphones (on) or Inserts (off)
Monitor Speech only or All Signals Tone/Speech defines test which
is loaded after start-up. Opens a dialog to select a printer
and configure its settings Stores PDF files to SD memory card
Shows device information Opens a dialog to change the date,
time, and the date format to US or Int
from 1 % to 100 % by using the Level controls (1), store new value
Language
Diagram Settings Display in Speech Test
Bone Lines
Symbol Setting
Import/Export User Settings
8100460 Rev. 8 42 24/04/2017
Number of Diagrams in Tone Test
English/Deutsch/Italiano/Français/ Español, etc. Language change requires a restart of the device.
Diagram or Table None: No audiogram displayed,
only the level and frequency. One: one combined audiogram Two: two separate audiograms for left and right
On/Off, displays a dotted line, connecting the bone conduction results
Int Symbols / US Symbols / UK Symbols
Export User Settings to SD Card or Import User Settings from SD Card. Provides the possibility to store the user settings on a SD Card.
Page 43
Operation Manual MA 42
5.5.1 Setup Date and Time
Select Date/Time in the User Menu by scrolling down with the left or right Level control (1) and select Set Date/Time by using the Stimulus Presenter bar (2). The following screen appears:
Figure 17
Set the date format to International or US by the Function button (9). Jump to the required position of the date or the time by the left Frequency Up/Down button (4) or (5) and change the value by the right Frequency Up/Down button (4) or (5) or left Level control (1). Press the Function button OK (6) to store the changes or Exit (7) to leave the Date/Time setting screen without saving the changes.
5.5.2 Set Printer Settings
Select the printer by turning the left or right Level control (1) down. The color mode is automatically adjusted. If the color mode is wrong adjust the color mode as well. Jump to the field paper format by pressing the Stimulus Presenter bar (2) several times and select A4 or Letter format by using the Level controls. If the printer is connected to your Ethernet network select Ethernet as port. Additionally, the IP address of the printer needs to be entered in the field “IP Address.” Select the number of the IP address by rotating the Level controls and press the Store button to enter the selected number.
Save the settings and return to the User Menu by pressing the Function button OK (7).
Figure 18
8100460 Rev. 8 43 24/04/2017
Page 44
Operation Manual MA 42
Test your printer settings by a sample print before starting the audiological assessment. The wrong settings may require a restart of the device.
Enter the field clinic name by using the STIM button (2). Select letters by rotating the Level control (1) and enter the selected letter by pressing the Store button (3). Save the settings and return to the User Menu by pressing the Function button OK (7).
5.6 Managing Test Results
All stored results can be directly printed via the USB printer. Make sure that a compatible printer is connected via the USB port (4) and the device is configured according to the connected printer settings; refer to menu settings in Section 5.5.2.
The results can also be stored as a PDF file on a SD memory card or USB flash drive to be later transferred to a PC for further usage. The PDF file contains the measurement results. An SD memory card needs to be inserted in the SD card slot (20) or a USB flash drive connected to the USB socket (4) on the rear side of the device.
NOTE: It is recommended using the USB flash drive for storing PDF results.
NOTE: Please refer to the Audiometry Module operation manual for transferring
of results to the PC. This is located on the included CD and USB. Before transferring data to a PC make sure that you have installed the PC software
properly according to the separately delivered operation manual.
When the examination is complete, press the Menu button (19) in the tone or speech test mode. The user menu is opened and the functionality of the Function buttons (11), (13) and (15) changes to PDF, Print, and Patients, respectively.
To print out the results press the Print button (13). Make sure that a compatible printer is connected and the printer settings are correct.
To store the results on the SD memory card or USB flash drive, press the PDF button (11). A PDF is created and stored on for further transfer to a PC or print out via a PC connected printer. Make sure that the setting PDF Save Option is set to correct storage medium SD memory card or USB flash drive.
Enter the patient list by pressing the Patient button (15) to store the results. After printing or creating a PDF you return automatically to the tone or speech test mode. To store the results within the Patient Management, review Section 5.7.
8100460 Rev. 8 44 24/04/2017
Page 45
Operation Manual MA 42
5.7 Patient Management
The patient management option allows the results of the audiological tests to be stored on the SD memory card. The results can be reloaded at a later time to be reviewed, edited, or printed. The patients can be stored by a number ID or by entering the name and birth date. The demographic patient information can be entered using the Level controls (1) or a connected USB keyboard.
Enter the User Menu by pressing the Menu button (19) in the tone or speech audiometry screen. Press the button Patients (15) to display the patient list.
To comply with patient privacy laws the option to add a PIN code for entering to this list has been offered when the language is set to English. Enter the Patient List and select New Patient by pressing the New Patient button (15). Select Login On by pressing the functional button Login On/Off (15). A screen appears to set your personal four digit login PIN. Select four numbers between 0 and 9 by the level dial and enter each by pressing the STIM button (2) and confirm your personal login PIN by pressing the button OK (9).
Caution: Remember the PIN very well, otherwise it is not possible to enter the patient list anymore!
Each time you enter the patient list this PIN must be entered by selecting the digit by the Level control dial and entering it by pressing the STIM bar (2). Moving forward, the PIN is required to enter the patient list. To deactivate the PIN, select Login Off after following the same steps as listed above.
Figure 19
Select a patient using the Level controls and press the Stimulus button (2) to display the stored sessions. Select a session and press the PDF button (11) to save the PDF on the SD memory card or USB flash drive. To print the results to a connected printer press Print (13). The patient information is only included on the print out, if it is done in the patient list, or if the PDF is created in the patient list screen.
Current results can be stored to a numbered patient or to a named patient. Entering the Patient List, a patient with a new number is selected automatically. Just press the Save button (9) to save the current session to the new patient number. To save the results to an existing patient, select a patient by the Level controls and press the Save button.
8100460 Rev. 8 45 24/04/2017
Page 46
Operation Manual MA 42
To save the current results to a new patient with a new patient name, press the button New Patient (15) and a screen appears to enter the patients last name, first name, ID and date of birth.
Figure 20
Enter the characters of the name by scrolling through the alphabet with the Level controls (1) and enter the selected character by using the Stimulus button (2). Jump to the next or previous field by using the Frequency Up (4) or Down (5) button. The date can be entered in the same manner with the Level controls. Delete the last character or number by pressing the Delete button (19).
A USB keyboard can also be utilized to enter the patient information. Connect it to the USB connector (4) and type in the characters. Jump to the next field with the tab key. Press the Function button OK (6) to save the new patient and go back to the patient list. The new patient is selected and the current measurement results can be saved to this new patient by pressing the Function button Save (9).
Press the button No Name (9) to store the results only by a patient number, without entering a name or use Cancel (6) to go back to the Patient List without saving.
8100460 Rev. 8 46 24/04/2017
Page 47
Operation Manual MA 42
Technical Data
This section offers you important information about
the MA 42 hardware specifications connections the pin assignment calibration values electromagnetic compatibility (EMC) electrical safety, EMC and associated Standards
6.1 The MA 42 Hardware
The MA 42 audiometer is an active, diagnostic medical
product according to the class IIa of the EU medical directive 93/42/EEC.
General Information About Specifications
The performance and specifications of the device can only be guaranteed if it is subject to technical maintenance at least once per year.
MAICO Diagnostics puts diagrams and service manuals at the disposal of authorized service companies.
STANDARDS
Medical CE mark Standards: ANSI S3.6: Type 2A
IEC 60645-1: Type 2 IEC 60645-2: Type A ISO 389 Acoustics – Reference zero for calibration of audiometric equipment IEC 60601-1 class I, protection class B, International Standards for Medical Electrical Equipment Medical Device Directive 93/42/EEC
DEVICE SPECIFICATIONS
Power supply: 100 - 240 V~ 50/60 Hz ±10 % Power Consumption: Approximately ~60 VA Environmental
conditions:
Operation: +15 °C … +35 °C / +59 °F … +95 °F
Humidity: 30 % … 90 %, non-condensing
Storage: 0 °C … +50 °C / +32 °F … +122 °F
Humidity: 10 % … 95 %, non-condensing
Transport: -20 °C … +50 °C / -4 °F … +122 °F
Humidity: 10 % … 95 %, non-condensing
8100460 Rev. 8 47 24/04/2017
Page 48
Operation Manual MA 42
Weight: 1.5 kg/2.7 lbs Dimensions: W x D x H: 34.5 x 20 x 8 cm / 13.4” x 7.9” x 3.2” Monitor: Build in monitor speaker, headset Device Fuses: 2 x 1A slow blow Warm-up Time: Less than 10 min after power on Communication: Talk forward and talk back Mode of Operation: Continuous Stimulus Functions: Tone Presenter/Interrupter
Interlock (tone presentation of both channels simultaneously) Tracking (fixed level difference between both channels) Masking
Data Connection: USB, LAN Ethernet External Devices: CD Player, USB Printer, USB Keyboard Supported Printers: HP (PCL 3 and PCL 5e)
Epson (ESC/P2, LQ, Stylus Color)
AUDIOMETRY
Test signals: Pure tone, Pulse tone, Warble tone Level Steps: 5 dB, 2 dB or 1 dB level steps (user selectable) Test Frequencies: 125 Hz to 8,000 Hz
9,000 Hz to 16,000 Hz optional*
Masking Signals:
Narrow Band Noise: With the same center frequency resolution as pure tone
White Noise:
Speech Noise:12 dB/octave above 1 kHz (+/-5 dB)
Speech Signals:
External CD Player, Microphone, WAV file from SD memory card.
Maximum Sound Pressure Level:
AC with Headphone: -10 dBHL to 120 dBHL BC with Bone Conduction
-10 dBHL to max. 80 dBHL
Oscillator: Insert earphones: -10 dBHL to 120 dBHL Sound field speaker from
Amp:
Tone: -10 dBHL to 100 dB Speech: -10 dBHL to 90 dB
HL
HL
Modulation: Pulse Tone: 0.25/0.5 s on time Warble Tone: 5 % sinus frequency modulation, repetition rate 5 Hz
Tests:
Tone: HL, MCL, UCL Speech: SRT, WRS, MCL, UCL, MHA Over Threshold Test Decay, SISI, ABLB, MLB Patient Response: Handheld response switch
8100460 Rev. 8 48 24/04/2017
Page 49
7 6 5 4 2
6.2 Connections
Operation Manual MA 42
Figure 21
1
9
3
8
10
11
13
12
Table 3 Connections on Backside
CONNECTIONS
No Connection socket Specification
1 On/Off Power 2 Power (100 ... 240 V~ 50/60 Hz) 3 Network Ethernet 4 USB Out USB 2.0 5 USB In USB 2.0 6 Patient Response Switch 7 Talk Back Microphone 8 Mic (Microphoney) 9 Monitor Phone 10 CD Input 11 Speaker Left 12 Speaker Right 13 Bone (Bone Conductor) 14 Insert Phone Left 15 Insert Phone Right 16 Phone Left (Headphone) 17 Phone Right (Headphone)
RI= 500 ZI= 1 k, UI= 0.38 - 500 mV ZI= 1 k, UI= 0.38 - 500 mV ZA= 250 , UA= 8 V
eff
ZI= 47 k, UI= 0.04 - 5 V ZA= 4 , UA= 8 V ZA= 4 , UA= 8 V ZA= 4 , UA= 8 V ZA=10 , UA=1 V ZA=10 , UA=1 V ZA=10 , UA=1 V ZA=10 , UA=1 V
eff
eff
eff
eff
eff
eff
eff
14
15
eff
eff
eff
17
16
8100460 Rev. 8 49 24/04/2017
Page 50
6.3 Pin Assignment
Operation Manual MA 42
Socket No. Connector Pin L
(left)
2
DC socket
L (Live)
Rated current international: 250 V/2,5 A
Socket No. LAN Ethernet
3
RJ45 Socket
RJ45 Cable Plug
4
Pin G (top)
G (Ground)
Pin N (right)
N (Neutral)
TX + Transmit Data TX- Transmit Data­RX+ Receive Data+ Not connected RX- Receive Data
1. +5 VDC
2. Data -
3. Data +
4. Ground
5
Socket No. Connector Pin 1 Pin 2 Pin 3
6; 7; 8; 11; 12; 13; 14;
Ground DC bias Signal
15; 16; 17
6.3 mm Stereo
-
9; 10
Ground DC bias Signal
Ground Right Left
3.5 mm Stereo
1. +5 VDC
2. Data -
3. Data +
4. Ground
8100460 Rev. 8 50 24/04/2017
Page 51
Operation Manual MA 42
6.4 Calibration Values and Maximum Levels
Calibration values and Max Levels: Headphone DD45
Coupler IEC 60318-3, Force 4-5 N, PTB-DTU Report 2009-2010
Frequency [Hz]
Tone RETSPL dB re 20µPa
NBN RETSPL dB re 20µPa
Max Tone
[dBHL]
Max NBN
[dBHL]
125 47.5 51.5 90 75 250 27.0 31.0 110 95 500 13.0 17.0 120 110 750 6.5 11.5 120 110 1000 6.0 12.0 120 110 1500 8.0 14.0 120 110 2000 8.0 14.0 120 110 3000 8.0 14.0 120 110 4000 9.0 14.0 120 110 6000 20.5 25.5 120 110 8000 12.0 17.0 110 100
Signal IEC 60645-2
RETSPL
IEC Max
Level [dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
Speech 20.0 110 18.5 100 SN 20.0 110 18.5 110 WN 0.0 110 0.0 110
8100460 Rev. 8 51 24/04/2017
Page 52
Operation Manual MA 42
Calibration values and Max Levels: Headphone TDH39
Coupler IEC 60318-3, Force 4-5 N, ANSI S3.6 and ISO 389-1
Frequency [Hz]
Tone RETSPL dB re 20µPa
NBN RETSPL dB re 20µPa
Max Tone
[dBHL]
Max NBN
[dBHL]
125 45.0 49.0 90 75 250 25.5 29.5 110 95 500 11.5 15.5 120 110 750 8/7.5 13/12.5 120 110 1000 7.0 13.0 120 110 1500 6.5 12.5 120 110 2000 9.0 15.0 120 110 3000 10.0 16.0 120 110 4000 9.5 14.5 120 110 6000 15.5 20.5 120 110 8000 13.0 18.0 110 100
Signal IEC 60645-2
RETSPL
IEC Max
Level [dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
Speech 20.0 110 19.5 110 SN 20.0 110 19.5 110 WN 0.0 110 0.0 110
8100460 Rev. 8 52 24/04/2017
Page 53
Operation Manual MA 42
Calibration values and Max Levels: Headphone Holmco 8103
Coupler IEC 60318-3, Force 4-5 N, PTB
Frequency [Hz]
Tone RETSPL dB re 20µPa
NBN RETSPL dB re 20µPa
Max Tone
[dBHL]
Max NBN
[dBHL]
125 39.5 43.5 90 80 250 25.0 29.0 105 95 500 18.5 22.5 110 100 750 13.5 18.5 120 105 1000 12.0 18.0 120 110 1500 10.0 16.0 120 110 2000 9.5 15.5 120 110 3000 9.0 15.0 115 110 4000 9.0 14.0 110 110 6000 19.5 24.5 100 110 8000 20.0 25.0 100 110
Signal IEC 60645-2
RETSPL [dBHL]
IEC Max
Level [dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
Speech 20.0 110 24.5 110 SN 20.0 110 24.5 110 WN 0.0 110 0.00 110
8100460 Rev. 8 53 24/04/2017
Page 54
Operation Manual MA 42
Calibration values and Max Levels: Headphone DD450
Ear simulator IEC60318-1 with adapter, Force 8.8 N ± 0.5 N, ANSI 3.6 and ISO 389-8/ISO 389-5
Frequency [Hz]
Tone RETSPL [dB
SPL
]
NBN RETSPL [dB
SPL
]
Max Tone
[dBHL]
Max NBN
[dBHL]
125 30.5 34.5 100 75 250 18.0 22.0 110 85 500 11.0 15.0 115 95 750 6.0 11.0 120 100 1000 5.5 11.5 120 100 1500 5.5 11.5 115 100 2000 4.5 10.5 115 100 3000 2.5 8.5 115 100 4000 9.5 14.5 115 100 6000 17.0 22.0 105 90 8000 17.5 22.5 105 90 9000 19.0 24.0 100 85 10000 22.0 27.0 100 85 11200 23.0 28.0 95 80 12500 27.5 32.5 90 75 14000 35.0 40.0 80 70 16000 56.0 61.0 60 60
Signal IEC 60645-2
RETSPL
IEC Max
Level [dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
Speech 20 90 19 90 SN 20 85 19 85 WN 0.0 115 0.0 115
8100460 Rev. 8 54 24/04/2017
Page 55
Operation Manual MA 42
Calibration values: Insert phone Eartone 3A and IP30
Reference equivalent threshold sound pressure level Coupler IEC 60318-5, ANSI 3.6 and ISO 389-2
Frequency Hz Tone IEC
60318-5 RETSPL dB re 20µPa
NBN IEC 60318-5 RETSPL dB re 20µPa
Tone
Max Level
dBHL
Max Level
dBHL
125 26.0 30.0 90 90 250 14.0 18.0 105 105 500 5.5 9.5 110 110 750 2.0 7.0 115 110 1000 0.0 6.0 120 110 1500 2.0 8.0 120 110 2000 3.0 9.0 120 110 3000 3.5 9.5 120 110 4000 5.5 10.5 115 110 6000 2.0 7.0 100 100 8000 0.0 5.0 95 95
Signal IEC 60645-2
RETSPL
IEC Max Level
[dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
NBN
Speech 20.0 100 12.5 100 SN 20.0 100 12.5 100 WN 0.0 100 0.0 100
8100460 Rev. 8 55 24/04/2017
Page 56
Operation Manual MA 42
vcvcvc
Calibration values: Bone conductor Radioear B71/B71W/B81/BKH10 Force: 4.9 ... 5.9 N Mastoid placement
Coupler IEC 60318-6, ANSI 3.6-2010 and ISO 389-3
Frequency [Hz] Reference
Air radiation Max level
equivalent threshold force level for tone
[dB] (re 1µN) Min - max
[dB]
(ONLY for BKH10)
125
82.5 - -
Tone
[dBHL]
250 67 - 45 500 58 - 65 750 48.5 - 70 1000 42.5 - 70 1500 36.5 - 70 2000 31 - 75 3000 30 4-18 80 4000 35.5 - 80 6000 40 10.5-31 50 8000 40 - 50
Signal IEC 60645-2
RETSPL
IEC Max
Level [dBHL]
ANSI S3.6
RETSPL
ANSI Max
Level [dBHL]
Speech 55.00 60 55.00 65 SN 55.00 60 55.00 65 WN - - - -
8100460 Rev. 8 56 24/04/2017
Page 57
Operation Manual MA 42
Calibration values: Sound field (0 degree incidence)
Reference equivalent threshold sound pressure level and maximum hearing levels
For Canton CD220
ISO 389 – 7 and ANSI S3.6-1996
Frequency [Hz]
Tone RETSPL [dB
SPL
]
NBN RETSPL [dB
SPL
]
Max level
Tone [dBHL]
CD220
Max level
NBN [dBHL]
CD220
125 22.0 22.0 65 55 250 11.0 11.0 75 70 500 4.0 4.0 85 75 750 2.0 2.0 85 75 1000 2.0 2.0 85 75 1500 0.5 0.5 90 80 2000 -1.5 -1.5 90 85 3000 -6.0 -6.0 90 85 4000 -6.5 -6.5 95 85 6000 2.5 2.5 85 80 8000 11.5 11.5 80 70
Signal IEC60645-2
RETSPL [dBSPL]
IEC Max
Level [dBHL]
CD220
ANSI S3.6
RETSPL
[dBHL]
ANSI Max
Level [dBSPL]
CD220
Speech 0.0 90 14.5 75.5 SN 0.0 90 14.5 75.5 WN 0.0 90 0.0 75.5
8100460 Rev. 8 57 24/04/2017
Page 58
Operation Manual MA 42
Guidance and manufacturer’s declaration
- electromagnetic emissions
The
MA 42
is intended for use in the electromagnetic environment specified below. The customer or the user of the
MA 42
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment
- guidance
Recommended separation distances between portable and mobile RF communications equipment and the
MA 42
.
Rated Maximum output
Separation distance according to frequency of transmitter
[m]
0.01
0.1
1
10
100
Guidance and Manufacturer’s Declaration
- Electromagnetic Immunit
y
Immunity Test
IEC 60601 Test
Compliance
Electromagnetic
< 5 %
UT
6.5 Electromagnetic Compatibility (EMC)
Portable and mobile RF communications equipment can affect the MA 42. Install and operate the MA 42 according to the EMC information presented in this section. The MA 42 has been tested for EMC emissions and immunity as a standalone MA 42. Do not use the MA 42 adjacent to or stacked with other electronic equipment. If adjacent or stacked use is necessary, the user should verify normal operation in the configuration. The use of accessories, transducers and cables other than those specified, with the exception of servicing parts sold by MAICO as replacement parts for internal components, may result in increased EMISSIONS or decreased IMMUNITY of the device. Anyone connecting additional equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
RF emissions CISPR 11
RF emissions CISPR 11 Harmonic emissions IEC 61000-3-2 Voltage fluctuations / flicker emissions IEC 61000-3-3
The MA 42 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the MA 42 can help prevent electromagnetic interferences by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the MA 42 as recommended below, according to the maximum output power of the communications equipment.
power of transmitter [W]
0.12 0.12 0.23
0.37 0.37 0.74
1.17 1.17 2.33
3.70 3.70 7.37
11.70 11.70 23.30 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1 At 80 MHz and 800 MHZ, the higher frequency range applies. Note 2 These guidelines may not apply to all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Group 1 The MA 42 uses RF energy only for its internal function.
Class B The MA 42 is suitable for use in all commercial, industrial,
Complies Class A Category Complies
150 kHz to 80 MHz d = 1.17
Therefore, its RF emissions are very low and are not likely to cause any interference in nearby electronic equipment.
business, and residential environments.
80 MHz to 800 MHz d = 1.17
800 MHz to 2.5 GHz d = 2.23
The MA 42 is intended for use in the electromagnetic environment specified below. The customer or the user of the Instrument should assure that it is used in such an environment.
Electrostatic Discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC61000-4-4 Surge
IEC 61000-4-5
Voltage dips, short interruptions and voltage variations on power supply lines
IEC 61000-4-11
level
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
(>95 % dip in UT) for 0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT (>95 % dip in UT) for 5 sec
+6 kV contact
+8 kV air
+2 kV for power supply lines
+1 kV for input/output lines
+1 kV differential mode
+2 kV common mode
< 5 % UT (>95 % dip in UT) for
0.5 cycle
40 % UT (60 % dip in UT) for 5 cycles
70 % UT (30 % dip in UT) for 25 cycles
<5 % UT
8100460 Rev. 8 58 24/04/2017
Environment-Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be greater than 30%.
Mains power quality should be that of a typical commercial or residential environment.
Mains power quality should be that of a typical commercial or residential environment.
Mains power quality should be that of a typical commercial or residential environment. If the user of the MA 42 requires continued operation during power mains interruptions, it is recommended that the MA 42 be powered from an uninterruptable power supply or its battery.
Page 59
Operation Manual MA 42
Note:
Guidance and manufacturer’s declaration
— electromagnetic immunity
(a)
Power frequency (50/60 Hz)
IEC 61000-4-8
UT is the A.C. mains voltage prior to application of the test level.
The MA 42 is intended for use in the electromagnetic environment specified below. The customer or the user of the MA 42 should assure that it is used in such an environment, Immunity test IEC/EN 60601
Conducted RF IEC/EN 61000-4-6
Radiated RF IEC/EN 61000-4-3
3 A/m
test level
3 Vrms 150kHz to 80 MHz
3 V/m 80 MHz to 2,5 GHz
3 A/m
Compliance level Electromagnetic environment –
3 Vrms
3 V/m
Power frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or residential environment.
guidance Portable and mobile RF communications equipment should be used no closer to any parts of the MA 42, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Pd 2,1 Pd 2,1
80 MHz to 800 MHz
Pd 3,2
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, (a) should be less than the compliance level in each frequency range (b)
Interference may occur in the vicinity of equipment marked with the following symbol:
800 MHz to 2,5 GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the MA 42 is used exceeds the applicable RF compliance level above, the MA 42 should be observed to verify normal operation, If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the MA 42.
(b)
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
8100460 Rev. 8 59 24/04/2017
Page 60
Operation Manual MA 42
6.6 Electrical Safety, EMC and Associated Standards
1. UL 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety
2. IEC/EN 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety
3. CAN/CSA-C22.2 No. 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety Electrical Equipment for Laboratory Use
4. IEC/EN 60601-1-1: Collateral Standard, Safety Requirements for Medical Electrical Systems
5. IEC/EN 60601-1-2: Medical Electrical Equipment, Part 1 - Electromagnetic Compatibility - Requirements and Tests
6. Essential Requirements of the current European Union Medical Device Directive 93/42/EEC
7. 2011/65/EU of 8 June 2011 on the restriction of the use of certain hazardous substances in electrical and electronic equipment (RoHS)
8. Directive 2002/96/EC of the European Parliament and of the Council of 27 January 2003 on waste electrical and electronic equipment (WEEE)
8100460 Rev. 8 60 24/04/2017
Page 61
Operation Manual MA 42
6.7 Checklist for subjective Audiometer Testing
8100460 Rev. 8 61 24/04/2017
Page 62
Internet:
www.maico.biz
Specifications are subject to change
Operation Manual MA 42
MAICO Diagnostics GmbH Sickingenstr. 70-71 10553 Berlin Germany Tel.: + 49 30 / 70 71 46-50 Fax: + 49 30 / 70 71 46-99 E-mail: sales@maico.biz
8100460 Rev. 8 62 24/04/2017