The ERO•SCAN® Pro test Instrument is indicated for testing cochlear and middle ear function in infants,
children, and adults by measuring otoacoustic emissions (OAEs), tympanometry, and acoustic reflex
(with optional external Tymp•OAE Probe™). The presence of otoacoustic emissions suggests normal
outer hair cell function, which in turn correlates to normal hearing. The presence of a normal
tympanogram suggests normal middle ear function and the presence of acoustic reflex responses at or
below 95 dB HL indicates normal inner hair cell function.
The ERO•SCAN Pro is intended to be used by hearing healthcare professionals (i.e. ENT doctors,
audiologists) and/or technicians, neonatal nurses and school nurses who have been trained by a hearing
healthcare professional.
1.1 Instrument Description
What is the ERO•SCAN Pro Instrument?
The ERO•SCAN Pro test instrument consists of the handheld unit, external probe(s), printer, single-use
eartips and other accessories. The ERO•SCAN Pro instrument may be used as a screening tool, or in
conjunction with conventional tests as part of a full diagnostic evaluation.
The ERO•SCAN
hardware and software for generating the test
stimuli, measuring and displaying test results, and
storing the results until they are printed or
downloaded to the PC software. The plastic
housing contains circuit boards that provide the
signal processing and display the test results. The
instrument also contains 4 AA/UM-3/R6 alkaline
batteries to power the device. The instrument uses
an organic LED display (OLED) and light-emitting
diodes (LEDs) to provide a visual display of test data
and test conditions to the operator. Two
membrane-type push buttons and a 4-way
navigation control located on the control panel of
the device allow the user to power the instrument
on and off, control testing, and initiate printing.
The probe sections of the instrument and external
probes house a microphone, two transducers, and
two speaker tubes which produce test stimuli and measure the sound pressure level (SPL) present in the
sealed ear canal. Additionally, the combined Tymp•OAE Probe™ contains a mechanical pump system
and pressure sensor to control and monitor air pressure variation in the ear canal. Each probe houses a
keypad with two buttons (right and left) which can be used to start a test with the probe. LED’s on the
probe keypad indicate ear being tested and test status.
Interface of the instrument to the ear canal is accomplished through disposable eartips made of
industrial elastomer, which fit onto the probe tip. The disposable eartips are color coded to facilitate
easy selection by size.
H
ow are the Results Stored and Reported?
When the ERO•SCAN Pro is set in its default settings, the instrument will store the results from one
patient (most recent left and right ear test for each available protocol) in its non-volatile memory for
Pro instrument contains the
1
Page 5
Operating Instructions
subsequent printing. The results are displayed via the display on the front of the device and are stored
in the device’s internal memory. After testing is completed, results can be printed using the optional
thermal paper printer or the default PC printer via software interface. Tests can also be exported to a
computer database via optional software. Test results are stored in the non-volatile memory so the
operator can delay printing until a later time if desired.
1.2 Otoacoustic Emissions
What Are DPOAEs?
Distortion Product Otoacoustic Emissions (DPOAEs) are acoustic signals that can be detected in the ear
canal of a person with normal outer hair cell function, subsequent to stimulation of the auditory system
with a pair of pure tones at frequencies f
product tone at the frequency 2f
W
hat Are TEOAEs?
Transient Evoked Otoacoustic Emissions (TEOAEs) are
acoustic signals that can be detected in the ear canal
of a person with normal outer hair cell function,
subsequent to stimulation of the auditory system with
a series of wideband clicks.
W
hat Do Otoacoustic Emissions Results Tell Us?
Available evidence suggests that otoacoustic
emissions (OAEs) are generated by the cochlea’s outer
hair cells, and that the presence of OAEs is an
indication that the outer hair cells are normal.
Although OAE test data provide no indication of inner
hair cell function, or of hearing ability, current
research indicates that the majority of hearing-impaired individuals can be identified by a simple OAE
test. Patients who fail to generate OAEs should be rescreened and/or referred for additional audiological
testing.
6
H
ow Does the ERO•SCAN Pro Device Measure DPOAEs?
The ERO•SCAN Pro instrument generates a series of test tones, directs them into the ear canal, and
then measures the level of the DPOAE tone generated by the cochlea. By using different test
frequencies, the ERO•SCAN Pro device provides an estimate of outer hair cell function over a wide
range of frequencies.
H
ow Does the ERO•SCAN Pro Device Measure TEOAEs?
The ERO•SCAN Pro instrument generates a series of clicks, directs them into the ear canal, and then
analyzes the spectrum of the returning signal, separating the noise and emission. By using bandpass
filters, the ERO•SCAN Pro device provides an estimate of outer hair cell function over a wide range of
frequencies.
7
H
ow Does the ERO•SCAN Pro Device Work?
The digital signal processor in the instrument generates two pure tones (f
series of wideband clicks for TEOAEs through a digital-to-analog converter. These tones or clicks are
presented to the ear via speaker tubes located in the probe. A microphone in the probe measures the
sound in the ear canal and transmits the signal to the analog-to-digital converter. The digital signal
processor then uses Fast-Fourier Transforms (FFTs) to filter the signal into narrow frequency bands, and
detects any emissions present. The level of these emissions can be compared with the level of the noise.
1-f2
.
and f2. The resulting emission of interest is the distortion
1
and f2) for DPOAEs or a
1
2
Page 6
Operating Instructions
The SPL and frequencies of the test tones and the averaging time used to process the signals can be
determined by the tester through adjustable settings maintained in static memory within the ERO•SCAN
Pro instrument.
W
hat Frequency Range of Hearing is Estimated?
DPOAEs: Approximately 1.5 kHz to 12 kHz (depending on the frequency range selected). Since the
health of the hair cells in the region of the f
frequency is at about six-tenths of the f
and c) the ambient noise tends to be highest at low frequencies, the lowest f
be routinely measured is about 1 kHz. TEOAEs: Roughly 500 Hz to 4 kHz. TEOAEs can be reliably
recorded at lower frequencies than DPOAEs, but cannot be measured reliably above 4 kHz.
1.3 Tympanometry
What is tympanometry?
Tympanometry is the objective measurement of middle ear mobility (compliance) and pressure within
the middle ear system. During the test, a probe tone (226 or 1,000 Hz) is presented to the ear canal by
means of the Tymp•OAE Probe. This tone is used to measure the change in compliance in the middle
ear system while the air pressure is varied automatically from a positive value (+200 daPa) to a negative
value (-400 daPa max.).
H
ow is compliance measured?
Maximum compliance of the middle ear system occurs when the pressure in the middle ear cavity is
equal to the pressure in the external auditory canal. This is the highest peak of the curve as it is recorded
on the chart. The position of the peak on the horizontal axis and on the vertical axis of the chart will
provide diagnostic information regarding the function of the middle ear system. Examples of normal
and abnormal tympanograms can be found in a later section of this manual.
W
hat other measurements are calculated?
does not depend on the cooperation of the test person and can therefore not be falsified by the
patient.
The impedance measurement examines the acoustic resistance of the middle ear. If the eardrum is hit by
a sound, part of the sound is absorbed and sent via middle ear to the inner ear while the other part of
the sound is reflected. The stiffer the eardrum is the more sound is reflected and the less sound reaches
the inner ear. Inside the probe of the impedance measuring instrument a small loudspeaker is installed
which emits a low frequency sound through a tube (Figure 3) into the auditory canal before the
eardrum.
Together with a third tube, all three are inserted deeply into the ear canal and are made airtight against
3
Another tube is connected to the microphone inside the probe which receives the sound.
test frequency are estimated, and a) the 2f1-f2 emission
2
frequency, b) emissions tend to be weak below 600 Hz or so,
2
test frequency that can
2
Gradient calculations are reported as the tympanogram
width at half of peak compliance expressed in daPa. A
“limits” box is available on both the display and printout
to aid in diagnosis. Compliance is measured with respect
to an equivalent volume of air, with the scientific quantity
milliliter (ml). Air pressure is measured in deca-Pascals
(daPa).
What does tympanometry tell us?
The impedance measurement assists in diagnosing of the
condition of the middle ear and can therefore not be
compared directly with other audiometric tests such as
sound or speech audiometry which assists in the
measurement of hearing. Furthermore, the impedance
measurement is an objective measuring method which
Page 7
Operating Instructions
outside pressure by the ear tip. A manometer and a pump, which can produce both positive and
negative pressure, are connected with tube C
eardrum is stiff and the eardrum transmits the majority of the sound via the middle ear to the inner ear.
The highest compliance is normally reached with an air pressure corresponding to the outside pressure.
While performing tympanometry measurements,
a continuous change of positive and negative
pressure is produced by the pump of the
instrument in the outer auditory canal. The
compliance is measured simultaneously and
shown in a diagram (the tympanogram) which
illustrates the compliance in ml or mmho over the
pressure in daPa. The area for normal
tympanogram curves is hatched. Here you can
see that the highest compliance is reached with
normal pressure. When you create positive and
negative pressure the eardrum stiffens - the
compliance decreases. So you can draw
conclusions on the condition of the middle ear
from the form and the values of the
tympanogram.
N
OTE: 1.02 mm H2O = 1.0 daPa.
. Less sound is reflected to the microphone when the
1.4 Acoustic Reflex
Acoustic reflex refers to the reflexive contraction of the stapedius muscle in response to sound
stimulation (typically 70-100dB). This contraction causes reduced mobility of the ossicular chain and
reduces the compliance of the tympanic membrane which is measured by the probe as a change in
admittance. The Stapedial (acoustic) reflex is always bilateral in response to loud sound presented to
either ear, and the lowest level that causes a change in admittance is called the reflex threshold.
Typically the average threshold is around 85dB HL with normal limits falling between 70 and 95dB HL.
4
Page 8
Operating Instructions
Standard System Parts
Handheld
Eartip Kit
Cradle
589
589-7
589-2
USB Cable
Operating Manual
AA Batteries
1025-2088
1162-0802
1145-2002
Tips
589-4
589-5
589-8
Optional Parts
Roll
Calibration Cavity
Printer Kit
586-8
589-9
589-10
Probes
Standard w/ OAE Units
Standard w/ Tymp Units
589-1
589-3
2 Getting Started
2.1 Unpacking the system
Standard System Parts:
(1) ERO•SCAN Pro Hand-held
unit w/ internal probe
(1) External OAE•Probe™ or
(1) External TympOAE Probe™
(1) Cradle
(1) USB cable
(1) Box of disposable eartips in
assorted sizes
(4) AA/UM-3/R6 Alkaline
batteries
(4) Internal probe tips (4/pkg)
(4) External probe tips (4/pkg)
(1) ERO•SCAN Pro PC software
CD
(1) Operating Manual
O
ptional System Parts:
(1) Calibration cavity
(1) Thermal Paper Printer with
Lithium Ion battery installed
with:
(1) Thermal printer paper roll
(1) Printer charger
(1) Printer data cable
Internal Probe
External Probe Tips
Note: It is recommend that you
s
ave the box and packing
m
aterials in case you need to
s
tore or ship the system.
Thermal Paper
OAEProbe
TympOAE Probe
PC Software CD
5
Page 9
Operating Instructions
The
Pro instrument uses 4 AA/UM-3/R6
Alkaline batteries. Open the battery compartment by
detailed information about the
Up/down arrows change the selection
.2 Battery Installation
2
ERO•SCAN
sliding the battery panel down and install the batteries
as indicated on label inside the compartment. Once the
batteries are correctly in place, slide the panel back onto
its tracks to close the battery compartment.
3 Getting familiar with the ERO•SCAN Pro Instrument
3.1 Controls and display (Figure 1)
Power Button:
Press to power on
Press and hold to power off
I
nfo Button:
View
selection
N
avigator:
shown on the middle line
Left/right arrows initiate the action shown in
the bottom line of the display.
T
est Status Indicators:
Green – indicates the instrument is ready to
test
Yellow – indicates test is in progress
Amber – indicates an error condition
Figure 1
6
Page 10
Operating Instructions
The internal probe is located on the underside of the
and high frequency
to the ear canal where it would be measured as noise (a
nal probe assembly houses the microphone which
ear canal. Interface of the instrument to the ear canal is
disposable eartips made of industrial
3.2 Internal Probe
ERO•SCAN Pro instrument. It allows for the measurement of
DPOAEs in the range of 1.5 kHz to 6 kHz and TEOAEs in the
range of 0.7 kHz to 4 kHz.
Note: Tympanometry, acoustic reflex,
DPOAE measurements are not supported by the internal probe
of the ERO•SCAN Pro.
The internal probe consists of a patented dual isolated spring
design which prevents hand movement from being transferred
common problem with any truly handheld OAE probes). The
inter
measures the acoustic information present in the ear canal and
the tubing which carries the stimulus from the receiver(s) to the
accomplished through
elastomer, which fit onto the probe tip. A disposable internal
probe tip snaps to the shaft of the spring assembly and serves
as an attachment point for the disposable eartips.
3.3 External Probes
There are two types of external probe available for use with the ERO•SCAN Pro instrument. Each
consists of the following basic parts:
Connection cable – connects probe to ERO•SCAN Pro instrument by way of the connector
Probe enclosure – contains the circuitry, one receiver, and, in the case of the Tymp•OAE
Probe, the pump and pressure sensor
Keypad – consists of LED status indicators and Right/Left test start buttons
Probe cable – connects the probe enclosure to the probe head
Probe head – contains the microphone and one receiver
Probe tip – disposable tip onto which the eartip is affixed for sealing to the ear
OAE•Probe™
The OAE•Probe allows for the measurement of:
7
Page 11
Operating Instructions
Green:
Probe has been detected and
instrument is ready to test
Flashing Yellow:
AutoStart in process
Solid Yellow:
Testing in progress
DPOAEs in the range of 1.5 kHz to 12 kHz
TEOAEs in the range of 0.7 kHz to 4 kHz.
T
ymp•OAE Probe™
The Tymp•OAE Probe allows for measurements of:
DPOAEs in the range of 1.5 kHz to 12 kHz
TEOAEs in the range of 0.7 kHz to 4 kHz
Tympanometry with a 226 Hz or 1000 Hz probe tone
Acoustic reflexes with 226 Hz probe tone (stimulus options: 0.5, 1, 2, & 4 kHz; broadband noise,
low pass noise & high pass noise) or 1000 Hz probe tone (stimulus options: broadband
noise)
The external probe keypads consist of a left (indicated by
L
) and right (indicated by R) button to initiate the selected
test protocol or series. Below the left and right buttons
are blue (for left) and amber (for right) LEDs which
illuminate during testing to indicate the ear under test.
Each probe provides a status indictor in the form of LED(s)
located on the keypad. In the case of the OAE•Probe the
indicator is between the left and right buttons. On the
Tymp•OAE Probe the indicator is a row of LED’s above
the left and right buttons.
The LED located between the left and right buttons
indicates status of the probe:
The status indicator bar on the Tymp•OAE Probe will indicate the pressure sweep of the pump by
showing a sweeping sequence of lights.
Note: When idle (not testing) the probe status LED should be solid green. If the LED is yellow when idle
t
hen the probe has not been detected by the instrument. Power off the instrument, check the probe
c
onnection, and turn the instrument on again to detect the probe. The status LED should now be solid
g
reen.
The external probes are preferred by most users when conducting OAE measurements for infants,
young children, and other difficult to test populations such as individuals with multiple handicaps. The
external probe allows the user to insert the probe tip into the ear and then wait until the patient is quiet
before starting the test. The internal probe may be preferred by some users for fast OAE testing of
cooperative patients.
I
nstalling the External Probe
8
Page 12
Operating Instructions
Turn off the ERO•SCAN Pro and insert the external probe plug into the
socket on the top of the ERO•SCAN Pro. The plug will fit only in one
direction. The arrows on the plug should face the display on the
ERO•SCAN Pro.
Turn on the ERO•SCAN Pro. The status indicator on the external probe
keypad will be solid green indicating the ERO•SCAN Pro has detected
the presence of the remote probe. Disconnect and reinsert the
connector if the status indicator on the probe is not illuminated or is
yellow rather than green. To return to using the internal probe of the
handheld, turn the instrument off, disconnect the external probe and
power up the instrument again.
Note: Misalignment of the plug and socket when installing the remote probe can cause damage to the
p
ins in the plug and the pin receptacles in the socket. The plug and socket should be visually inspected
p
rior to each installation of the remote probe. If damage is observed, contact your Special Equipment
D
istributor or Maico Diagnostics.
9
Page 13
Operating Instructions
The
Pro instrument arrives preloaded with default protocols
or press the L or R button on the
If testing with the external probe, be sure to connect the external probe prior to powering on
4 Conducting a Measurement
4.1 Quick Start
ERO•SCAN
and ready to test. Testing with the external probe can be started with
just five easy steps.
1. Turn on the instrument.
2. From the main menu (Figure 2), select the desired protocol or
series using the up or down arrows on the navigator.
Figure 2
When Tymp•OAE Probe is connected, the pump will perform an initialization. Do NOT insert the
Tymp•OAE Probe into the ear canal while the pump is performing this initialization. The probe should
be placed in the ear canal either BEFORE or AFTER the pump initialization.
3. Select an eartip and place if fully on the probe tip.
4. Secure the probe to the patient and insert the eartip deeply
into the patient’s ear.
5. To begin a test Press the right or left arrow button on the
control panel (Figure 1)
external probe.
Additional information for each step is provided below.
4.2 Powering on the instrument
To turn on the ERO•SCAN Pro instrument, press the power button
located on the handheld control panel just below the left corner of
the display window.
A flash screen showing the firmware version (first line), serial
number (second line), and date set in the instrument (third line) will
illuminate briefly before showing the main display.
the instrument. When the probe is detected the status indicator on the external probe keypad
will be green.
If testing with the internal probe, be sure the external probe is disconnected prior to powering
on the instrument.
10
Page 14
Operating Instructions
4.3 Selecting a Protocol or Protocol Series
Protocols are individual DPOAE, TEOAE, tympanometry or acoustic
reflex measurements. Protocol series are a sequence of individual
protocols linked together. A series can be composed of any
combination of DPOAE, TEOAE, tympanometry or acoustic reflex
protocols that are loaded into the instrument. The series will run
with just one button press (left/right arrow on control panel or R/L
button on probe) to begin testing. Series are indicated by
multiplicity symbol to the left of the Series name.
Each instrument comes preloaded with protocols and protocol series. For a description of the protocols
and series included see Appendix B. For information on creating and customizing protocol series see
section 9.
The protocol selected will default to the last protocol or series used to conduct a measurement. This
makes it easy to switch to the other for testing. To select a different protocol or series use the up and
down arrows on the navigator located on the instrument control panel.
4.4 Selecting an Eartip
The ERO•SCAN Pro instrument comes with a box of disposable eartips that fit a variety of ear canal
sizes. The probe tip must have an eartip attached before inserting it into an ear canal. The eartip kit has
12 different size eartips that are color-coded for easy selection. The determination of the appropriate
eartip size should be made by persons with proper training and experience. The eartip must seal the ear
canal. The best test results are obtained when the eartip is inserted deeply into the ear canal instead of
flush with the ear canal opening. The fit should be secure so that the probe will remain in the ear canal
even with a light tug on the probe.
Caution must be taken, however, to ensure that the eartip does not extend too deeply into the ear
canal. Use only the eartips approved for use with the instrument. Contact your local special equipment
distributor or Maico Diagnostics for ordering information. The eartips are disposable and should be
replaced after each patient. Do not attempt to clean or reuse these eartips.
After selecting an eartip, push it onto the probe tip until it is flush against the base of the probe tip. The
sound outlet tubes on the probe tip are recessed to minimize the likelihood of clogging. If the probe tip
does become plugged or clogged, it must be replaced. See section 11 on care and maintenance for
further information. To remove the eartip, grasp the eartip at the base and twist it while pulling it
straight off the end of the probe tip.
4.5 Preparing the Patient
When possible, otoscopic or visual examination of the patient’s ear canals should be performed prior to
testing. Excessive cerumen or vernix in the ear canals may interfere with the test and give invalid or
incomplete results. Patients with excessive cerumen, debris, or foreign bodies in the ear canals should be
referred to an audiologist or physician for removal of the blockage prior to testing.
Place the patient in a position that will allow easy access to the patient’s ears. The patient should remain
still and quiet while the test is being performed.
Explain to the patient that the measurement is painless. The patient does not have to respond when
there are loud test sounds or when the pressure in the auditory canal changes. The patient should be
instructed not to swallow, chew or move during the measurement.
11
Page 15
Operating Instructions
Lanyard Use
Shirt Clip
Secure the probe enclosure using the neck loop or shirt clip such that there is no weight on
In addition, when performing acoustic reflex measurements explain to the patient that loud test sounds
will occur during the reflex measurement. It is very important that the patient does not move because
movements can be perceived as a false compliance change.
A
mbient Noise Levels in the Testing Environment
The ERO•SCAN Pro uses a novel noise-rejection algorithm that permits accurate DPOAE and TEOAE
measurements in background noise as high as 65 DB SPL A-weighted (typical office environment). If the
ambient noise level rises too high (and/or the eartip seal is poor), then all samples will be noisy and
accurate measurements will be impossible, in which case the test result will indicate “noisy.”
Securing the External Probe to the Patient
The external probe is positioned on the patient using the lanyard affixed to the back of the probe
enclosure. The lanyard is designed to be used as either a neck loop or a shirt clip. Selecting the style for
use will depend on patient factors and the preferences of the user
Adjusting Lanyard
The lanyard is affixed to the probe enclosure with an auto-locking mechanism. To adjust the lanyard
position, slide the locking mechanism down (toward to the connection cable) and hold the mechanism
in the downward position while pulling the lanyard in the desired direction. Pull the lanyard down
(toward the connection cable) to use the shirt clip or up (toward the probe cable) to use it as a neck
loop. Once in position, release the locking mechanism and pull up gently on the lanyard.
.
Neck Loop
the probe cable and eartip. The probe enclosure should be positioned as close to the ear as
possible.
12
Page 16
Operating Instructions
Probe not in ear or leak (AutoStart 1):
Be sure the eartip is
Probe in ear and seal detected (AutoStart 2):
not begin, that may be the result of instability (probe is moving
ive) or there is
A different eartip may be
required.
Probe blocked (AutoStart 3):
changing angle or position of the probe until AutoStart image 2
10 on
4.6 Conducting a measurement
After selecting the desired protocol or series, press the right or left arrow button on the navigator or L
or
R
button on the external probe keypad to begin a measurement. Select the left arrow or L button to
start a left ear test. Press the right arrow or
A
utoStart
The first phase in the test sequence is AutoStart which checks the fit of the probe in the ear canal. The
actual calibration and measurement will commence once an adequate probe fit has been achieved. The
sequence can be started with the probe placed in the ear or prior to positioning the probe in the ear.
This is a matter of user preference. Users of the internal probe tend to prefer to start the test before
positioning the probe in the ear.
AutoStart consists of low-frequency, alternating tones which are used to check for the following
conditions: Seal of the eartip to the ear canal, leak, blocked probe, clogged probe tip, stability of the
probe, and noise. During AutoStart the condition of the probe in the ear is represented by the following
images:
This image indicates the probe is outside the ear canal or there is no
seal. Continue to insert the probe into the ear canal. If this condition
persists, a different eartip may be required.
securely seated deep in the ear canal.
R
button to start a right ear test.
AutoStart 1
This image indicates the probe is in the ear canal. The test will start
soon as long as all conditions of AutoStart are met. If the test does
because fit is not secure or the patient is too act
excessive noise present. If this image persists and testing does not
AutoStart 2
AutoStart 3
begin, refit the probe and try again.
This image indicates the probe is blocked. The eartip or probe tip
might be blocked by ear wax (cerumen) or vernix (birth fluid) or the
probe has been pushed against the ear canal wall.
Remove the probe from the ear and check for wax/debris then reinsert
appears. If the blocked probe condition persists, see section
Troubleshooting.
Note: Do not hold the probe in the ear during OAE testing. This will introduce noise into the
m
easurement. Common sources of noise are environmental (room noise), biological (patient breathing,
m
oving, talking, chewing, etc.), or physical (probe movement).
13
Page 17
Operating Instructions
Proceeding after a single protocol measurement
Proceeding after a Series measurement
Proceeding following a Series is slightly different than what was
erent set of
Testing ears with PE tubes
To test OAE’s of individuals with PE tubes or middle ear perforations, the AutoStart may need to be
disabled. This is accomplished by first inserting the probe with eartip attached into the ear canal. Be
sure the fit is deep and secure to obtain a proper seal, To disable AutoStart at the main menu select the
ear to be tested by holding down the right or left arrow key for 3 seconds until the green light turns off.
Once the key is released, the EROSCAN Pro will calibrate and test as usual.
Calibration
The ERO•SCAN Pro will automatically perform a calibration prior to each frequency tested (DPOAE) or at
the start of each test (TEOAE).
T
est Phase
During the test phase a flashing indicator will appear to the right of the display. Test results are
displayed as they are collected. For more information regarding test results see section 5.
Testing is complete when the green “READY” light is illuminated. Both the tester and patient should
remain as still and quiet as possible until the green light turns on.
4.7 Proceeding to the next test
After testing has been completed the selection menu will automatically
appear in about 5 seconds. The selection menu will offer the following
options:
*Bring the selection menu up by pressing any arrow key
described above. The selection menu will offer a diff
options:
Left – starts a left ear test of the selected protocol
Right – starts a right ear test of the selected protocol
Done – return to the main menu
Review – return to the result screen*
Left Series – starts the complete series for the left ear
Right Series – starts the complete series for the right ear
Done – Returns to the main menu
Repeat Last – starts the series from the last test performed
14
Page 18
Operating Instructions
DP/TE-Gram
displays the absolute values of the signal
the test
8.10 for more information regarding Boys Town norms.
values on a simple to read display.
Value Graph
and obtain a better probe fit.
5 Understanding the Test Results
5.1 Understanding the OAE Results Display
During OAE testing the results will appear on the screen as the test progresses. The display shows a
graph with up to 12 columns. Each f2 frequency (DPOAEs) or frequency band (TEOAEs) is indicated by
one column. The number of columns shown will vary depending on the number of frequencies being
tested with the selected protocol. The test frequency is shown along the horizontal axis at the bottom
of the display. Three different OAE viewing options are described below:
The
DP/TE-Gram graph
(emission) as a block plotted at the intersection point of
frequency (Hz) along the horizontal axis and the amplitude (dB SPL)
along the vertical axis.
Note: Boys Town normative data may optionally be shown on the
display for qualifying DPOAE measurements when the DP/TE-Gram
graph is used as shown in the display to the left. See Setup section
A bright solid block indicates that the SNR is at least 6 dB for DPOAE
and 4 dB for TEOAE. A hashed block indicates that the SNR is less
than these specified values. Examples are shown in the display to the
left.
This allows the user to judge the quality of the measurement and the
possible influence of noise while still viewing the absolute emission
The
Value graph
and noise floor. The noise is represented by an open bar. The signal is
a represented by a solid bar when it meets the pass criteria and as a
dashed bar when it fails to meet the pass criteria for the protocol.
When viewing the value graph the user can immediately see if noise
was a contributing factor in obtaining a Refer.
For example on the test result shown to the left we can see that the
measurement at 2 kHz (the bar farthest to the right) is completely
obscured by noise (no measurable signal). At 3 kHz there is some noise
in the measurement, but the measured signal at that frequency meets
or exceeds the pass criteria. At 4 kHz the signal emerging above the
noise does not meet the pass criteria. At 5 kHz there is no noise in the
measurement. This is an example of a test that should be repeated
after an attempt is made to reduce noise (environmental or patient)
displays the absolute values of the signal (emission)
15
Page 19
Operating Instructions
SNR Graph
for each test
Test Details
data) press the info button located on the
Tympanogram
of the tympanogram. The area surrounded by the box is valid for
Test Details
The
SNR graph
frequency. SNR is the difference between the measured emission and
measured noise floor. The SNR is shown on the vertical axis so the
height of each column represents the SNR for that test frequency. For
example, if the column goes to the top of the display then the SNR is
15 dB or greater.
The bright solid yellow columns meet the SNR pass criteria for that
frequency. The hashed columns do not meet the pass criteria set for
that frequency.
Note: if the protocol does not have a pass criteria then all the columns
will appear hashed.
To view test details such as the protocol name, date/time of test, test
number, serial number, OAE signal, noise floor value and signal-tonoise ratio (numerical
instrument control under the bottom right corner of the display. Use
the up and down arrow buttons to scroll through the details screen. If
any details are out of the viewable area select SHIFT to extend the
viewable area of the display. When finished viewing the details, select
EXIT to return to the results screen.
a
rrow key to bring up the selection box.
shows signal-to-noise ratio (SNR)
From the results screen press any
5.2 Understanding the Tympanometry Results Display
After having completed a tympanometry measurement you can see the
results on the display. On the left side of the display you see a graph
“normal” tympanograms. The curve that appears on the graph
represents the movement of the ear drum. When the peak of the
curve appears inside the box, the screening is generally considered a
PASS. The ear canal volume is represented on the display by the arrow
on the graph directed toward the vertical axis of the graph.
To view test details such as the ear canal volume, gradient, compliance,
and peak pressure (numerical data) press the info button located on
the instrument control under the bottom right corner of the display.
16
Page 20
Operating Instructions
5.3 Interpreting the Tympanometric test result
As a general rule, values for ear canal volume should be between 0.2 and 2.0 ml (children and adults).
A variance will be seen within this range depending on the age and ear structure of the person. For
example, a 2.0 ml or larger reading in a small child could indicate a perforation in the tympanic
membrane, while it may be a normal reading in an adult. You will become more familiar with the
normal ranges when you use the instrument.
The normal range for compliance is 0.2 ml to approximately 1.8 ml. A compliance peak within the range
indicates normal mobility of the middle ear system. A peak found outside of these limits may indicate
one of several pathologies.
Middle ear pressure should be equivalent to ambient air pressure (0 daPa on an air pressure scale).
Minor shifts of the peak compliance to the negative may occur with congestion and are rarely to the
positive side. Establish criteria for abnormal negative pressure when you become more familiar with
using the equipment. It is generally accepted that negative pressure of greater than -150 daPa indicates
a referral for medical evaluation.
5.4 Abnormal Tympanometric Values
It is the purpose of this section to provide samples of tympanograms which reflect abnormal states of
the middle ear mechanism. It is not the intention of this section to provide you with a complete guide to
interpreting results. Complete information regarding pathologies and abnormal impedance testing can
be found in published audiology literature.
A perforation in the tympanic membrane will cause a high ear canal volume measurement because the
instrument will measure the volume of the entire middle ear space. The EROSCAN Pro may refuse to
run the test, with the instrument indicating a seal problem, or a flat tympanogram will be recorded
since no movement will occur with a change in air pressure.
17
Page 21
Operating Instructions
An extremely flaccid tympanic membrane or an ossicular chain discontinuity will yield a very high peak
compliance in the presence of normal middle ear pressure and ear canal volume will be normal.
A fixation of the ossicular chain, as in otosclerosis, will produce a tympanogram with very low
compliance in the presence of normal middle ear air pressure while ear canal volume is normal.
Middle ear fluid such as in serious otitis media will yield a very flat tympanogram with no definite peak
and negative air pressure. A resolving case or beginning case may produce a reduced peak in the
presence of severe negative middle ear pressure. The ear canal volume is normal and the reflex is either
absent or at an elevated level.
Eustachian tube dysfunction in the absence of fluid will show a normal compliance curve, but it will be
displayed to the negative side of the tympanogram. Ear canal volume will be normal.
18
Page 22
Operating Instructions
During the measurement of the acoustic reflex the change of the
reflex occurs, the
When the test is finished a summary display will show the final
.
The stimulus type is followed by a yes/no indication that the
present. A compliance change meeting or exceeding the selected
measured. A compliance change failing to meet the criteria is
See Appendix B for additional
5.5 Understanding the Acoustic Reflex Display & Result
compliance is represented on the instrument display as shown to the
left. The zero-line at the bottom of the graph indicates the measured
compliance without a test sound. Changes in compliance are shown
as deviations from the zero-line. If a Stapedial
compliance increases and the curve rises. The change in compliance is
shown for each intensity level for the stimulus being presented.
compliance change curve for each stimulus presented in the protocol
Above each curve the test stimulus type is indicated by the following:
compliance change reached the selected criteria to consider the reflex
criteria is indicated by a symbol (default criteria is 0.05 ml) followed
by the stimulus intensity (in dB HL) where the Stapedial reflex was
indicated by a symbol followed by the maximum intensity level.
The overall test outcome is indicated by PASS or REFER in the upper
right corner of the display. This is determined by the pass criteria
established for the protocol.
information on default protocols.
A correct interpretation of the measuring results can only follow in connection with the tympanogram,
the graphic reflex display (to rule out artifact) and other actual data. But in principle a Stapedial reflex
indicates that the patient hears on the “stimulus ear” and that the sound lead functions.
Note: the acoustic reflex measurement is automatically conducted with pressure held at peak
compliance (compensated).
19
Page 23
Operating Instructions
To review test results stored in the instrument scroll down the main
From the
Review Results
screen select
Review Results
using the right
Review Results
will display all tests saved in the instrument. For each
From the test result display, press any arrow button to bring up the
6 Managing Test Results
The ERO•SCAN Pro saves one right and one left ear test for each protocol. Once a new test for that ear
and protocol is started, the previous results are overwritten. W
s
hould be printed before a new patient is tested. When the test results are printed (to PDF, PC printer
or thermal printer) they are marked for deletion and will be erased when a new test is started.
Note: the ERO•SCAN Pro can be configured to save up to 350 tests organized by patient numbers. For
more information on this option see section 8.7.
6.1 Reviewing Test Results
hen testing is completed, the results
menu until
Show Results
is selected and press the right arrow button.
arrow button.
saved test the ear under test, the test outcome, and the protocol name
are shown.
L
or R indicates the ear tested (right or left)
indicates a PASS result
indicates a REFER result
A dash (-) in the result box indicates no pass criteria was established for
the protocol or the test outcome was No Seal, Noisy, or Fit Error
To view a specific test, scroll up or down to select the test and press
the right arrow button to
Select
to view the results.
selection box with the following options:
Press the right arrow button to delete the test result
Press the up arrow button to return to the list of saved tests
Press the down arrow button to return to the test result
Press the info button to view the test details
20
Page 24
Operating Instructions
6.2 Deleting Test Results
D
eleting a single test
To delete a single test(s) follow the steps outlined in the previous section (above) and choose
from the options provided on the selection menu.
D
eleting all tests:
To delete all saved tests select
From the
buttons on the control panel to select
action by pressing the right arrow button. When prompted,
confirm selection by pressing the down arrow button for
return to the
press the up arrow for
Review Results
Review Results
Show Results
screen, use the up and down arrow
from the main menu.
Delete All
and select the
Yes
screen without deleting saved tests,
No
.
delete
. To
21
Page 25
Operating Instructions
The
PC software must be installed prior to connecting the cradle to the
7 Printing Test Results
There are three options for printing tests results from the ERO•SCAN Pro:
1. Quick Print to PDF
2. Quick Print to the default PC printer
3. Fast and portable printing is an option with the thermal paper printer
These printing options are explained in this section.
7.1 Connecting the cradle to a computer
Connecting to the computer will allow access to additional features such as Quick-print to PDF or the
default 8.5 x 11 PC printer and customization of protocols and series. See section 9 for more
information on customizing protocols and series. See Appendix A for software installation instructions.
ERO•SCAN Pro
computer. See appendix A for software installation instructions.
Use the included USB cable to connect the cradle to the computer by connecting one end of the cable
to the USB port on the underside of the cradle and the other end to the desired USB port on the
computer. Place the two-position button located on the upper right of the top side of the cradle in the
down position for computer use.
22
Page 26
Operating Instructions
Open the
Print Results
application
by double clicking on the
Transfer of test data will occur
file or print to the default PC
The first time test data is transferred to the
printing preferences. You can also set your
and then
: the test
Print test results to a PDF file in the default
that can be named and saved for import into
7.2 Quick-Print to PDF or to the default PC printer
application icon.
Place the ERO•SCAN Pro gently in
the cradle.
automatically when the instrument
is placed in the cradle.
This application will print to a PDF
printer.
Note: During installation of the application a folder will be created at C:\_\My Documents\My Test
Results. The default directory can be changed in the printing preferences.
23
application you will be prompted to establish your
preferences by selecting
Printing Test Results
from the menu.
Do not automatically print test results
will be taken when he instrument in placed in the
cradle
Preferences
: no action
Print test results to the default printer
results will be sent to the default PC printer and
and an 8.5 x 11 page will be printed. There will
be no electronic copy saved.
directory
electronic medical records systems or for printing
in the future.
: the test results will be sent to a PDF file
Page 27
Operating Instructions
The application will give the user the
skip this step if no patient name is
Click [Cancel] to abort the
To view a saved PDF file, click the
in the main window or navigate to
option to add the patient's name to
the print out. Enter the patient's
name and click [OK]. Click [Cancel] to
desired.
You will be prompted to save the file.
The software will suggest a file name
or you may use a file name of your
choice.
file save.
View Test Results in PDF files
the directory where the file is saved
using Windows Explorer.
button
24
Page 28
Operating Instructions
OAE PDF or PC Printout
D
ata table:
F2 = the f2 frequency
P1 = the sound pressure level of f1
P2 = the sound pressure level of f2
DP = the level of the emission in dB SPL
NF = the noise floor in dB SPL
SNR = the signal-to-noise ratio (DP level minus the noise floor)
P = indicates that the pass criteria has been met for the indicated frequency
S
ignal to Noise Graph:
Vertical axis = SNR (dB)
Horizontal axis = f2 frequency tested (Hz)
Green bars indicate that the pass criteria have been met. Red bars indicate that SNR and/or DP
amplitude have not been met.
D
P-Gram Graph
Vertical axis = indicates the absolute value of the signal or noise in dB
Horizontal axis = f2 frequency tested (Hz)
The green line is the signal. The red line is the noise. The difference between the two lines is
the signal-to-noise ratio.
25
Page 29
Operating Instructions
Tympanometry PDF or PC Printout
D
ata table:
Frequency = probe tone frequency (226 or 1,000 Hz)
Ear volume = indicates the volume of the external ear canal
Gradient = indicates graph width in daPa value (tympanometric width at 50% of the peak)
Compliance = displays the peak compliance
Peak Pressure = displays the pressure corresponding to peak compliance
G
raph:
Vertical = relative canal volume (ml) indicated by the arrow and dynamic compliance indicated
by the peak of the curve (ml)
Horizontal = pressure corresponding to peak compliance (daPa)
If the tympanogram is within the preset limits, the peak will be within the box displayed on the
screen and the test result will be
Pass
.
26
Page 30
Operating Instructions
Acoustic Reflex PDF or PC Printout
D
ata table:
Pressure = pressure compensation to match point of peak compliance from preceding
tympanometry measurement
Number of stimuli = number of test stimuli included in the protocol (up to four allowed)
Minimum for a pass = number of passing stimuli required to meet the compliance change
criteria
G
raphical table:
Each horizontal row shows the compliance changes for one stimulus type plotted as a function
of stimulus intensity which is indicated in the header for each vertical column.
The solid line shows the measured compliance change for a particular combination of stimulus
and intensity
Green = compliance change criteria has been met
Yellow = compliance change criteria has not been met
Red = the maximum intensity has been reached without meeting the compliance change criteria
Pass/refer is indicated at the end of each horizontal row
27
Page 31
Operating Instructions
7.3 Thermal printer set up
Use the included data cable to connect the cradle to the printer. First, insert the appropriate
end of the data cable into the port located on the underside of the cradle. Connect the cradle
and printer by inserting the appropriate end of the data cable into the data port on the back of
the printer. Next, connect the 2.5mm jack on the power supply to the power connector
located on the back of the printer. Finally, connect the appropriate ends of the power cord
into the power adapter and an outlet.
The printer is permanently powered up and ready to print once the power supply is connected
to the printer and an outlet. The light on the face of the printer will illuminate solid green
when the printer is powered up and ready to print. A flashing green light indicates the printer
is not ready (paper low, no paper loaded, lid open, etc…). No light indicates the printer is off.
Paper:
Paper rolls must be 57.5 (±0.5) mm wide, 55 mm maximum diameter, and have thermally
sensitive coating on the outside. The printer will accept rolls which are coreless or wound on a
core.
ote: To order more paper, contact your local special instrument distributor or Maico
N
Diagnostics.
Procedure for loading paper:
Slide the lid release button forward until the lid springs open.
Unwind a small amount of paper from the roll and insert paper roll into the printer with the
paper coming up from under the roll.
Close the lid.
28
Page 32
Operating Instructions
Use the included printer connection cable to connect the cradle
rear of the printer.
Place the two-position button located on the upper right of the
After loading, check that the paper is straight and advances properly. Tear off any excess
paper by pulling the paper sharply towards you across the serrated tear bar. In the event of a
jam or other paper loading problem, release the lit and straighten the paper before closing
again.
Pressing the paper feed button when the printer is idle advances the paper. “Double-clicking”
the button (pressing and releasing the button twice in quick succession) prints a demo/test
message showing the printer firmware version and settings.
.4 Connecting the cradle to the Printer
7
to the printer. Connect the appropriate end of the connection
cable into the serial data port at the
Connect the opposite end of the data cable into the printer
connector on the underside of the cradle.
top side of the cradle in the raised position for printer use.
29
Page 33
Operating Instructions
7.5 Printing with the Thermal Paper Printer
Be sure the printer power is on and the button on the cradle is in the printer position. Place the
instrument gently in the cradle. A printer icon should appear on the display and the test results should
begin printing immediately.
DPOAE Printout:
To the left is a sample DPOAE printout from the thermal paper
printer.
The header shows the protocol name in the first line followed by the
date/time of the test, test number, instrument serial number, probe
serial number, and firmware version.
Patient number is indicated or a blank line is provided to write in the
name. (See section 8.7 for information on using numbered patients).
The ear (Right or Left) and the test result (Pass or Refer) will be
indicated on the printout.
Data table:
F2 = the f2 frequency
P1 = the sound pressure level of f1
P2 = the sound pressure level of f2
DP = the level of the emission in dB SPL
NF = the noise floor in dB SPL
SNR = the signal-to-noise ratio (DP level minus the noise floor)
P = indicates that the pass criteria has been met for the indicated
frequency
Solid bars indicate that the pass criteria have been met. Hashed bars
indicate that SNR and/or DP amplitude have not been met.
30
Page 34
Operating Instructions
TEOAE Printout
To the left is a sample TEOAE printout from the thermal paper
printer.
The header shows the protocol name in the first line followed
by the date/time of the test, test number, instrument serial
number, probe serial number, and firmware version.
Patient number is indicated or a blank line is provided to write
in the name. (See section 8.7 for information on using
numbered patients).
The ear (Right or Left) and the test result (Pass or Refer) will be
indicated on the printout.
Data Table:
F = the frequency band
P = peak pressure level of the click stimulus
TE = the level of the emission in dB SPL
NF = the noise floor in dB SPL
SNR = the signal-to-noise ratio (TE level minus the noise floor)
P = indicates that the pass criteria has been met for the
indicated frequency
Graph:
Vertical axis = SNR (dB)
Horizontal axis = frequency band (Hz)
Solid bars indicate that the pass criteria have been met. Hashed
bars indicate that SNR and/or DP amplitude have not been met.
Figure 48
31
Page 35
Operating Instructions
Tympanometry Printout
To the left is a sample tympanometry printout from the thermal
paper printer.
The header shows the protocol name in the first line followed
by the date/time of the test, test number, instrument serial
number, probe serial number, and firmware version.
Patient number is indicated or a blank line is provided to write in
the name. (See section 8.7 for information on using numbered
patients).
The ear (Right or Left) and the test result (Pass or Refer) will be
indicated on the printout.
Data table:
Frequency = probe tone frequency (226 or 1,000 Hz)
Ear volume = indicates the volume of the external ear canal
Gradient = indicates graph width in daPa value (tympanometric
width at 50% of the peak)
Compliance = displays the peak compliance
Peak Pressure = displays the pressure corresponding to peak
compliance
Graph:
Vertical = relative canal volume (ml) indicated by the arrow and
dynamic compliance indicated by the peak of the curve (ml)
Horizontal = pressure corresponding to peak compliance (daPa)
If the tympanogram is within the preset limits, the peak will be
within the box displayed on the screen and the test result will be
Pass
Figure 49
.
32
Page 36
Operating Instructions
coustic Reflex Printout
A
To the left is a sample acoustic reflex printout from the thermal
paper printer.
The header shows the protocol name in the first line followed
by the date/time of the test, test number, instrument serial
number, probe serial number, and firmware version.
Patient number is indicated or a blank line is provided to write
in the name. (See section 8.7 for information on using
numbered patients).
The ear (Right or Left) and the test result (Pass or Refer) will be
indicated on the printout.
Data table:
Pressure = pressure compensation to match point of peak
compliance from preceding tympanometry measurement
Graph:
See section 5.5 for explanation
33
Page 37
Operating Instructions
To enter the
Set Up
menu, scroll down the main menu to
Set Up
and
To change the language, select
Language
from the
Set Up
menu.
To set the instrument time and date, from the set up menu select the
Use the left and right arrow buttons to select a portion of the time
When the time and date are set as desired, press the right arrow
8 Set Up
select by pressing the right arrow button.
Scroll through the set up menu options using the up and down
arrow buttons on the control panel. To change the settings, select
item to be changed by pressing the right arrow button.
Note: the current setting for each option is displayed after the colon.
8.1 Language
Scroll through the language options using the up and down
arrow buttons on the control panel. To change the settings, select
another language and press the right arrow. The languages available
will vary by instrument.
8.2 Time/Date
item showing the current date and time settings. Use the right
arrow button to enter the
or date. The selected portion of the date will appear highlighted. Use
the up and down arrow buttons to change the selected portion
of the Time/Date.
The time can be showing using a 12 hour (am & pm) or 24 hour
clock. To use the 12 hour clock, select
the time. For the 24 hour clock, select
time.
button until
of the display. Press the right arrow to
Save
appears in the bottom right corner of the display
Time/Date
menu.
am
or pm in the last portion of
24h
in the last portion of the
save
the time/date.
34
Page 38
Operating Instructions
To change the contrast of the display, select
Display
from the set up
Use the up and down arrows to adjust the contrast. When done
To view the instrument details, select
The user can select how the instrument will operate when performing
, testing will be
DP Early
the measurement will stop as soon as the
.
8.3 Display Contrast
menu.
8.4 Instrument Details
8.5 DP Early Stop
adjusting the contrast press the right arrow button to
up menu. Instrument details will show th
software) version and a summary of other instrument settings.
a DPOAE test. With the
completed at all frequencies included in the selected protocol. The test
is complete when all frequencies have been measured. If the
Stop
feature is
instrument can determine that the criteria for a PASS or REFER result
has been met. In this case some frequencies may not be measured,
but testing will be completed faster.
To change this setting, select
Using the up and down arrow buttons on the control panel select
off
or on and then press the right arrow button to
on,
DP Early Stop feature off
DP Early Stop
from the
Save
Set Up
select
and exit.
menu
and exit.
35
Page 39
Operating Instructions
8.6 OAE Minimums
This setting allows the user to select if minimum amplitude values
will be used as part of the pass/refer criterion. If
on
, a result is not considered a pass unless the OAE amplitude is
equal to or greater than the minimum value set in the protocol.
This is in addition to meeting the other pass criteria including the
minimum SNR and the number of passing frequencies for overall
test “Pass.”
The default minimum DP amplitude is -5 dB SPL.
The default minimum TE amplitude is -12 dB SPL.
Users of diagnostic instruments can set the minimum amplitude value when creating custom protocols.
Note: if
have a value established within the protocol. For more information regarding minimum amplitudes in
custom protocols, see section 9.
To change this setting, select
arrow buttons on the control panel select
and exit.
Use Minimums
is on then the minimum amplitude values are applied to all OAE protocols that
Use Minimums
from the
on
or
Set Up
off
and then press the right arrow button to
menu. Using the up and down
Use Minimums
select
is
8.7 Save Mode (Patients)
In the default operation mode, the ERO•SCAN Pro saves the most
recent right and left ear test for each protocol. Once a new test for
that ear and protocol is started, the previous results are
overwritten. When the test results are printed they are flagged for
deletion and will be erased when a new test is completed.
The user can choose to enable the
be automatically populated with two selections,
Patient 1
will appear. Select the desired protocol and conduct a measurement as described previously in this
manual.
Once a test has been completed for
Patient Names menu. The automatic numbering will continue until the test results are printed, at which
point all the numbered patients and test results will be flagged for deletion and then erased when a
new test is started. The new test will become the first test for Patient 1 and the automatic numbering
sequence will begin again. The instrument can hold up to 350 test results. The number of patients is
not limited.
NOTE: Users may also use the “Delete All” option found at the bottom of the Patient menu. Upon
confirming the selection, all tests and all patient placeholders will be deleted. An empty placeholder for
Patient 1 will appear in the Patient menu.
Use the
, select Patient 1 using the right arrow button. The main menu with all available protocols
No Name
place holder for tests that are not associated with a specific patient.
When
automatically add a new menu,
that appears when the instrument is powered on. This menu will
Patient 1
Numbered Patients
Patient 1
, the instrument will automatically add
and
are enabled the instrument will
No Name
Numbered Patients
Patient Names
. To perform and save a test for
, as the first screen
Patient 2
feature.
to the
36
Page 40
Operating Instructions
The
Pro provides audible feedback to the user to indicate
There are three options for viewing the DPOAE test results: SNR
Gram. These are explained in
To change this setting, select
buttons on the control panel select
button to
select
and exit.
Patients
No Patients
from the
or
Set Up
menu. Using the up and down arrow
Numbered Patients
and then press the right arrow
Note: The ERO•SCAN Pro also supports the upload of patient names by way of the optional Patient
Management software. See Patient Management Manual (1162-0806) for more information.
8.8 Sounds
ERO•SCAN
when testing is complete.
To change this setting, select
the up and down arrow buttons on the control panel select
off
and then press the right arrow button to
Sounds
from the
select
Set Up
and exit.
menu. Using
on
or
8.9 Graph Style
graph, Value graph, and DP/TEsection 5.1.
37
Page 41
Operating Instructions
When viewing DPOAE results using the DP/TE-Gram option (section
8.10 Norms
8.9), the user may optionally show the Boys Town normative data
on the instrument display.
The user can select to show the 95
range. The authors of the referenced normative study (Gorga et al.
1997) suggest the 95
normal hearing and hearing impaired populations.
The Boys Town Norms template is explained in the following window. Users should refer to the
reference article for additional information.
th
to 5th range as most accurately separating
th
to 5th range or the 90th to 10th
8.11 Reset
To restore the instrument back to the manufacturer default settings, select
and confirm the selection when prompted.
38
Reset
from the
Set up
menu
Page 42
Operating Instructions
CAUTION: Protocols should only be created or modified by hearing healthcare professionals, such
hearing healthcare professional.
9 Customizing Protocols and Series
Users of any instrument type can create series of protocols. Users of diagnostic instruments can also
customize individual protocols. The
custom protocols and series. This section will explain how to customize protocols and series. For
software installation instructions see Appendix A.
Protocols are files which contain the specific test parameters to conduct a DPOAE, TEOAE, or
tympanometry test and assign a Pass or Refer to the test result. A series is a set of linked protocols that
run in sequence by pressing one button to begin the series. Each ERO•SCAN Pro comes preloaded with
protocols and series.
as audiologists and ENT physicians, with adequate education and experience in the areas of
o
tacoustics emissions, tympanometry and acoustic reflexes. Individuals without adequate education
a
nd experience should use the provided default protocols or seek supervision from a qualified
9.1 Creating/Editing Protocol Files
Protocol files originate on the computer and are then loaded to the ERO•SCAN Pro instrument. To
create or edit a protocol file, click on the button, Edit Protocol Files on the Computer, located in the
upper left corner of the main window.
Print Results
software application is used to create and manage
39
Page 43
Operating Instructions
The Protocols dialog to the left shows the protocol files stored on the computer. This list can consist of
protocols loaded into the instrument and protocol files which are currently located only on the
computer.
Manufacturer default protocols appear in blue font. User created or modified protocols appear in black
font.
The right panel of the Protocols window contains a number of buttons to perform various functions
related to creating, editing, and managing protocol files saved on the computer. These functions are:
View – displays the settings and parameters for the selected protocol.
Create New – begins the process to create a new DPOAE, TEOAE or Tympanometry protocol file
Modify – opens the protocol file for editing
Remove – deletes the protocol file from the
Rename File – this allows the user to change the file name for the protocol
Copy – creates a duplicate of the selected protocol which can then be used as the starting point
for a user defined custom protocol
Choose Directory – allows the user to select change the location where custom protocol files are
saved
Done – exits the protocol window
40
Page 44
Operating Instructions
Creating a DPOAE Protocol
To create a new protocol file, click the
window. A new dialog will appear, select
Create New
button located on the right panel of the protocol
DPOAE Protocol
.
The New DPOAE Protocol dialog above shows the window for creating or modifying a DPOAE Protocol.
Protocol Name: the text entered into the Protocol Name field is what will appear in the instrument when
the protocol is loaded. This should be kept to approximately 14-16 characters in length so that the full
protocol name can be viewed on the instrument display.
Evaluation Criteria: check the box if a Pass or Refer outcome should be determined for this protocol.
Indicate the number of passing frequencies required for the overall test result to be
passing frequencies for a test pass can be set to any value between 1 and 10, but cannot exceed the
total number of frequencies tested.
F2: use the slider bars or type the numerical value into the text box to set the f2 frequency. The f2
frequency is customizable in 100 Hz increments between 1.5 and 12 kHz.
Frequency Ratio: the frequency separation between the two primary tones (f2/f1). The f2/f1 ratio is
adjustable between 1.1 to 1.4.
P1/P2: these settings control the intensity of the two primary tones (f1 and f2). P1 and P2 are
adjustable between 40 and 70 dB SPL. The maximum paired setting is 70/60. The minimum paired
setting is 40/40.
Averaging Time: established the length of time the OAE will be measured and averaged at each test
frequency. This is adjustable between 1 and 4 seconds. Note: Longer averaging time generally results
in better test outcomes particularly when measuring in noisy environments, but shorter averaging times
may be desirable for pediatric populations.
SNR to Pass: defines the minimum SNR required for that frequency to be considered a passing
frequency. SNR to Pass is adjustable between 3 and 10 dB
Minimum DP: defines the minimum DPOAE (signal) amplitude required for that frequency to be
considered a passing frequency. Minimum DP is adjustable between -20 and 0 dB SPL. If the user does
not wish to use minimum amplitude as part of the Pass criterion, n/a can be selected for this parameter
and/or the instrument setting for
Note: when creating or modifying a protocol, the desired setting can be established for the first test
frequency (top row) and then applied to all frequencies by clicking the
button.
C
reating or Modifying TEOAE Protocols
To create a new protocol file, click the
window. A new dialog will appear, select
DP Minimums
Create New
TEOAE Protocol
There are three frequency range options when
new TEOAE protocol.
Narrow Band –
includes: 1500, 2000, 2500, 3000, 3500, and 4000 Hz
Wide Band – 6 Frequencies (700-4000 Hz) which includes:
700, 1000, 1400, 2000, 2800, and 4000 Hz
Wide Band – 10 Frequencies (750-4100 Hz) which includes:
can be set to
off
.
Apply Values to All Frequencies
button located on the right panel of the protocol
.
and 4100 Hz
Note: all TEOAE test frequencies refer to the center of the
frequency band or filter.
42
Page 46
Operating Instructions
Above is the TEOAE protocol editing window.
Protocol Name: the text entered into the Protocol Name field is what will appear in the instrument when
the protocol is loaded. This should be kept to approximately 14-16 characters in length so that the full
protocol name can be viewed on the instrument display.
Target Peak Pressure: sets the target intensity for the click stimulus. Target peak pressure is adjustable
between 70 and 85 dB SPL.
Averaging Time: the maximum length of time the TEOAE measurement will run before displaying a Pass
or Refer result. The test will stop automatically once a Pass result is obtained. The averaging time can
be adjusted between 8 and 64 seconds.
Evaluation Criteria: check the box if a Pass or Refer outcome should be determined for this protocol.
Indicate the number of passing frequencies required for the overall test result to be
passing frequency bands for a test pass can be set to any value between 1 and 10, but cannot exceed
the total number of frequency bands tested.
SNR to Pass: defines the minimum SNR required for that frequency to be considered a passing frequency
band. SNR to Pass is adjustable between 3 and 10 dB
Minimum TE: defines the minimum TEOAE (signal) amplitude required for that frequency band to be
considered a passing frequency. Minimum DP is adjustable between -20 and 0 dB SPL. If the user does
not wish to use minimum amplitude as part of the Pass criterion, n/a can be selected for this parameter
and/or the instrument setting for
Use Minimums
can be set to
off
.
Pass.
Number of
43
Page 47
Operating Instructions
To the left is the Tympanometry protocol editing window.
what will appear in the instrument when the protocol is loaded.
pass: selects the low peak
Maximum peak compliance for a pass: selects the high peak
Creating a Tympanometry Protocol
To create a new protocol file, click the
window. A new dialog will appear, select
Create New
Tympanometry Protocol
Protocol Name: the text entered into the Protocol Name field is
This should be kept to approximately 14-16 characters in length so
that the full protocol name can be viewed on the instrument
display.
Frequency: sets the probe tone frequency for the tympanometry
measurement. The selections for probe tone frequency are 226
and 1000 Hz.
Evaluate Pass or Refer: A within limits box will be shown on the
tympanometry display and print out by default for the 226 Hz
probe tone measurement. The Pass or Refer outcome will is based
on the presence or absence of the tympanometric peak within the
limits box. If no limits box or Pass/Refer outcome is desired, check
the box
Setting the within limits box:
Minimum peak pressure for a pass: selects the low pressure limit
Maximum peak pressure for a pass: selects the high pressure limit
Minimum peak compliance for a
Do not evaluate Pass or Refer
for the box (-300 to 400)
for the box (-300 to 400)
compliance limit for the box (0.0 to 2.0)
button located on the right panel of the protocol
.
.
compliance limit for the box (0.0 to 2.0)
Modifying a Protocol
To change the settings of an existing protocol, click the
protocol window (Figure 68). The protocol editing window for that protocol will open. Make the
desired changes and click
Manufacturer default protocols cannot be modified by the user. They can, however, be copied and
used as the starting point for user defined custom protocols. See instructions below for copying a
protocol.
R
emoving a Protocol
To permanently delete a protocol from the directory, click the
of the protocol window (Figure 68).
Removing a protocol file from the directory does NOT remove it from the instrument. To remove a
protocol from the instrument, follow the instructions provided below in section 9.2.
Save
.
44
Modify
button located on the right panel of the
Remove
button located on the right panel
Page 48
Operating Instructions
Creating an Acoustic Reflex Protocol
Above is the Acoustic Reflex protocol editing window.
Protocol Name: the text entered into the Protocol Name field is what will appear in the instrument when
the protocol is loaded. This should be kept to approximately 14-16 characters in length so that the full
protocol name can be viewed on the instrument display.
Evaluation Criteria: check the box if a Pass or Refer outcome should be determined for this protocol.
Indicate the number of passing frequencies required for the overall test result to be
passing frequency bands for a test pass can be set to any value between 1 and 4, but cannot exceed the
total number of frequency bands tested.
Using the check boxes under the
Using the check boxes 60-95 dBHL, select up to 6 intensities to be presented
Green check marks and/or red anti symbols will be shown in the area corresponding to each stimulusintensity combination. The green check mark indicates an allowed combination, whereas the red anti
symbol indicates a conflict between the stimulus type and the intensity.
The compliance change criterion determines the amount of compliance change required to consider the
acoustic reflex present. This is selectable between 0.03 ml and 0.09 ml.
Note: when using the 1kHz probe tone, broadband noise is the only allowable stimulus type. Use the
check box to the left of the broadband noise stimulus row to select this option.
Reflex Stimulus
column, select up to 4 stimuli to be presented.
Pass.
Number of
45
Page 49
Operating Instructions
Renaming a Protocol File
The protocol name that appears in the instrument is a parameter established within the protocol file.
The Windows file name for the protocol can and may be different than the protocol name. For ease of
managing protocols you may want to rename protocol files so that the protocol name and file name
match. To rename a protocol file, click the
window.
Copying a Protocol File
To copy a protocol, select the protocol to be copied and click the
panel of the protocol window (Figure 68). The copied protocol file will appear in the list of protocols
with the same protocol name, but the file name will be “Copy of
Select the new copy of the existing protocol and click the
the protocol window (Figure 68). To modify the protocol follow the instructions provided above and
save changes.
Rename
button located on the right panel of the protocol
Copy
button located on the right
selected protocol
Modify
button located on the right panel of
” .
46
Page 50
Operating Instructions
To load new or modified protocols into the
This dialog allows the user to manage the
right panel of the window are the following
View: displays the selected protocol or series
Remove: removes the selected protocol or
s the "Create New
Series" dialog shown in figure 85.
Instructions for creating a new series of
can select a protocol file to load in the
9.2 Managing Protocols in the Instrument
instrument or to remove unused protocols from
the instrument, click on the button located in
the upper right corner of the main window
Manage Protocols and Protocol Series
protocols and series in the instrument. On the
buttons:
settings
series from the instrument
Create New Series: open
.
R
emove Protocol or Series
To remove protocols or series from the instrument, select the protocols or series to be removed and click
the
Remove
shift+click to select multiple protocols to remove or load.
M
anufacturer default protocols cannot be removed – use the hide button instead.
W
hen series are removed from the instrument they are DELETED permanently.
button located in the right panel of the Protocols in Instrument window. Use ctrl+click or
protocols can be found in section 9.3 of
this manual.
Load: opens the directory from which the user
instrument
Done: exit
47
Page 51
Operating Instructions
To load a protocol into the instrument, click the
button located in the right panel of the
Load Protocols
Load
Protocols in Instrument window.
When prompted (Figure 80) select the appropriate
directory for the protocol being loaded.
Factory = manufacturer created protocols
User = user modified or created protocols
Select the protocol or protocols you wish to load and click
N
ote: Custom protocols will display an unlock symbol in front of the protocol name on the instrument
d
isplay.
Each protocol must have a unique name. When attempting to reload a protocol file from the computer
directory that has been modified from the version currently loaded in the instrument, you must first
remove the old protocol by that name and then load the modified protocol. If the protocol being
removed and reloaded was used in any series then the series will need to be removed and recreated
using the new protocol file.
9.3 Creating Series
Protocol series are a sequence of individual protocols which are linked together. A series can be
composed of any combination of DPOAE, TEOAE, or Tympanometry protocols that are present in the
instrument. Series are indicated on the instrument main menu by the multiplicity symbol to the left of
the series name. Series help to streamline the testing process by minimizing the number of button
presses required to run multiple protocols on a single patient. When using series, the probe can be
placed in the ear canal and a group of sequenced protocols for that ear can be started and completed
by pressing only one button to begin the right or left series.
Each instrument comes preloaded with protocols and protocol series. For a description of the protocols
and series included see Appendix B.
The users of all instruments can create custom series.
To create a new series, click on the
right corner of the main window.
Then click on the button located in the right column of the Manage Protocols
dialog.
Open
.
Manage Protocols and Protocol Series
button located in the upper
48
Page 52
Operating Instructions
Always proceed = testing will proceed to the
Step 1: Type the series name into the text box provided. Series names should be kept to approximately
12 characters in length to ensure the full protocol name will display on the instrument.
Step 2: From the list of
series and then click
Step 3: You will be prompted to determine when the testing sequence should proceed to the next
protocol. Make a selection and click
Available Protocols
Add
.
on the left of the window, select the first protocol for the
Ok
.
These rules can help improve the efficiency of testing by applying the logic you may use typically in your
practice setting. For example, you may want to start with an OAE test and only proceed with the
tympanometry test when there is a Refer outcome on the OAE test. Another option might be to start
with a tympanometry test and only proceed with the OAE test when the tympanometry result is within
normal limits (Pass). Finally, you may choose to establish a complete test battery series that runs a
several protocols in sequence regardless of the outcome of any single protocol.
Continue repeating step 2 (selecting from available protocols) and step 3 (determining how to proceed)
as described above until the series contains the desired protocols. When finished, click
Create New Series of Protocols
instrument is provided. When the instrument is removed from the cradle the new series will appear in
the list of protocols.
Note: if the series contains any OAE frequencies above 6 kHz or a tympanometry protocol then the
appropriate external probe must be connect in order for the series to be shown in the instrument.
window. Confirmation that the series was successfully loaded into the
protocol in the series regardless of test outcome
Proceed on Refer = testing will proceed to the next
protocol in the series only when the outcome of
selected test is
Proceed on Pass = testing will proceed to the next
protocol in the series only when the outcome of
selected test is
Refer
Pass.
.
Save
on the
49
Page 53
Operating Instructions
Problem
Possible Causes
Resolution
Test will not progress past
Probe fit is poor and/or noise level
Reposition or refit eartip (trying a different
or patient noise if possible. If problem
cavity or in your own ear.
Clogged probe tip.
Replace probe tip.
Probe tip is not fully attached.
Press firmly on the probe tip tabs until
secure.
External probe has not been
detected.
Check connector, power off the instrument
and then power on again.
Equipment malfunction.
Contact your special instrument distributor
or Maico Diagnostics.
Protocols or series are missing
The external probe which
supports the test function (high
DPOAE or
tympanometry) is not connected.
Turn off the instrument, connect the probe
The protocols were removed from
the instrument using the PC
software (possible only for
diagnostic instruments).
Reload protocols using PC software.
long enough.
full second. Try again
Batteries are dead or installed
improperly.
in the battery compartment.
No communication between
Button on cradle is the printer
position.
Check that the button on the cradle is the
computer position (down).
Cradle is not connected to the
Check USB connector on the underside of
the cable is fully seated in both connectors.
Communication settings within
PC software are incorrect.
Check and correct communication settings.
See Appendix A for instructions.
Thermal printer does not print.
Button on cradle is in the
computer position
Check that the button on the cradle is in
the printer position (up).
Cradle is not connected to the
Check the printer connection cable on the
both connectors.
The printer is in sleep mode.
When the printer is ready to print the
wake the printer from sleep mode.
Printer battery is not charged.
Plug the printer in to the charger. When
be green. Press the button to wake the
printer from sleep mode.
Paper is not installed correctly.
An amber light will appear when the paper
instructions in section 7.3 to replace or
reposition the paper roll.
There are no tests in the memory.
Confirm there are tests saved in the
instrument. Follow the instructions in
section 6.1 for reviewing test results.
Printer is stuck in error mode
Reset the printer following instructions
provided in section7.3
10 Troubleshooting
AutoStart
from the menu
Instrument does not turn on Power button was not depressed
instrument and computer
is too high.
frequency
computer.
size if necessary) and reduce environmental
persists, check that the test will start in a
and turn the instrument on again to detect
the probe.
The power button must be pressed for one
Install new batteries according to the label
the cradle and on the computer. Be sure
50
printer.
underside of the cradle and the back of the
printer. Be sure the cable is fully seated in
button will be green. Press the button to
the printer is ready to print the button will
is low or incorrectly installed. Follow the
Page 54
Operating Instructions
11 Care and Maintenance
11.1 Cleaning and Disinfecting the Instrument
This instrument and its accessories may be wiped clean with a damp cloth using a mild antiseptic
solution (e.g., cetylcide). Take care not to put excessive pressure on the clear display window or allow
any utensil to puncture the display window or control panel. D
Do not immerse the instrument in fluids or attempt to sterilize the instrument or any of its accessories
11.2 Maintenance & Calibration
This instrument should be calibrated annually by your special equipment
distributor or by Maico Diagnostics. Beyond that, it requires no regular
maintenance other than routine cleaning and battery replacement. The probe
tip requires replacement only when it becomes clogged.
11.3 Probe Tip Replacement
Probe tips are disposable and should be replaced when they become clogged.
Four replacement probe tips of each type (internal and external) are included
with this instrument. Do not attempt to clean the probe tip.
I
nternal Probe Tip Replacement
To replace the probe tip, squeeze the tabs as shown in the picture to the
right. The tabs should audibly snap off the probe assembly. Pull the probe tip
directly off the probe and discard it.
Obtain a replacement probe tip and orient the tip with the arrows on the face
of the probe tip directed toward the top of the instrument. The probe tip will
only fit on one way; be careful not to force the tip in place. Push the tip
directly down onto the probe. Once the probe tip is in place on the probe,
push firmly downward on the top of the tabs one at a time until a click is
heard. Tug lightly on the probe tip to verify that the tip is securely attached.
E
xternal Probe Tip Replacement
To remove:
Using a small pointed object, such as a pen or small screwdriver, push in the notches on the left and
right sides of the rear of the external probe until each tab is released (Figure 11A).
Slide the probe tip off the front of the probe and discard (Figure 11B).
To replace:
Align a replacement tip with the front of the probe (Figure 11C). Align the tab on the external probe
tip with the notch on the probe body. T
securely on the probe, remove the probe tip and reorient it.
Press firmly on the tabs to snap them into place (Figure 11D).
N
OTE: If the probe tip is not inserted completely, the ERO•SCAN Pro will not perform a test.
he tip will only fit in one direction. If the tip does not fit
o not allow any fluid to enter the device.
.
Figure 11A
51
Figure 11B
Figure 11C
Figure 11D
Page 55
Operating Instructions
time without a disposable eartip
12 Important Safety Precautions
The ERO•SCAN Pro Test System should be used only by those individuals trained to perform the testing
for which it has been designed. No person should attempt to use this instrument without the necessary
knowledge and training to understand how this equipment is to be properly utilized and interpreted.
The ERO•SCAN Instrument probe tip must not be inserted into an ear at any
properly affixed.
IMPORTANT
In the event of a critical system failure, the message shown to the left
w
ill be displayed. Discontinue use of the instrument and contact your
S
pecial Equipment Distributor or contact Maico Diagnostics by phone
a
t (888) 941-4201 or by fax at (952) 903-4100.
12.1 Precautions
READ THIS ENTIRE MANUAL BEFORE ATTEMPTING TO USE THIS SYSTEM.
Use this device only as described in this manual.
Use only the disposable eartips designed for use with this instrument.
Never insert the probe tip into the ear canal without affixing an eartip.
The eartips are disposable and for single patient use only. Do not clean or reuse eartips.
Use only disposable 1.5v AA/UM-3/R6 Alkaline batteries for the test instrument. Do not use
rechargeable cells in this device; do not mix battery types and do not mix old and new batteries.
Remove the batteries from the instrument if it will not be used for 4 weeks or more.
Do not immerse the unit in any fluids. See the Care and Maintenance section of this manual for
proper cleaning procedures.
Do not drop or otherwise cause undue impact to this device. If the instrument is dropped or
otherwise damaged, return it to the manufacturer for repair and/or calibration. Do not use the
instrument if any damage is suspected.
Use and store the instrument indoors only. Do not use this instrument or its accessories in
temperatures below 40°F (4°C) or above 100°F (38°C), or in relative humidity of more than
90%.
Do not attempt to open or service the instrument. Return the instrument to the manufacturer
for all service. Opening the instrument case will void the warranty.
Do not operate the printer if the power supply has a damaged cord or plug. See the instructions
on the following page.
Do not expose the printed results to sunlight or heat. Printing on thermal paper fades with
exposure to light or heat.
Photocopies of test results should be made if the records are to be kept indefinitely.
52
Page 56
Operating Instructions
13 System Specifications
DPOAE SYSTEM
PRIMARY TONES: Frequency: F2 from 1.5 kHz to 12 kHz
Intensity: Up to 6 kHz: 40/40 to 70/60 dB SPL
Over 6 kHz: 40/40 to 65/55 dB SPL
MIC SYSTEM NOISE: <=-20 dB SPL @ 2 kHz (1 Hz Bandwidth)
ARTIFACT: <-20 dB SPL @ 2F1-F2 Frequency
F1/F2 RATIO: 1.2 (default) adjustable 1.1 to 1.4
F1/F2 DIFFERENTIAL: 0 to 30 dB SPL
TEOAE SYSTEM
STIMULUS: Adjustable up to 83 dB SPL Broadband Click
CLICK BANDWIDTH: 500 Hz to 4000 Hz
ANALYSIS BANDS: 6 or 10 Bands
ARTIFACT: <-10 dB SPL
TYMPANOMETRY SYSTEM
PROBE TONE: Frequencies: 226 Hz, 1000 Hz
Level: 85 dB SPL with in-ear calibration
AIR PRESSURE: Range: 0.05cc to 3.0cc: +300 to -400 daPa
3.0cc or greater: at least +200 to -200 daPa
Capacity: 0.05cc to > 5.0 cc
Speed: 60 daPa/Sec. nominal
500 to 700 daPa/sec (depending on ear volume).
TEST TIME: 2.5 seconds nominal (depending on ear volume).
STIMULUS: 500, 1000, 2000, 4000 Hz Pure tone
Broadband noise, low pass noise, high pass noise
PROBE TONE: 226 or 1000 Hz
STIMULUS INTENSITY: 500-4,000 Hz 70-95 dB HL, in 5 dB steps
Noise stimulus 60-90 dB HL, in 5 dB steps
STANDARDS: IEC60645-5
ANSI S3.39-1987
53
Page 57
Operating Instructions
14 Warranty
Maico Diagnostics warrants that this product is free from defects in material and workmanship and,
when properly used, will perform in accordance with applicable specifications. If this instrument does
not meet these criteria within one year of original shipment, it will be repaired, or at our option,
replaced at no charge when returned to our service facility. Changes in the product not approved by
Maico Diagnostics shall void this warranty. Maico Diagnostics shall not be liable for any indirect, special
or consequential damages, even if notice has been given of the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING, BUT
NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR
PURPOSE.
For assistance with this ERO•SCAN Pro
Distributor or contact Maico Diagnostics by phone at (888) 941-4201 or by fax at (952) 903-4100.
The serial number of your instrument can be found inside the battery compartment and the printer
serial number is under the rechargeable battery. Please write the instrument and printer serial numbers
below for future reference.
ERO•SCAN Pro Instrument Serial Number:
Printer Serial Number:
Date Purchased:
Purchased From:
S
AVE THIS MANUAL FOR FUTURE REFERENCE
®
Test System contact your Special Equipment
54
Page 58
Operating Instructions
5
6
Appendix A: ERO-SCAN Pro Software Installation Instructions
1 Application Installation
To begin installation double click on the installer file: EroscanPro_installer_user_p-120508.exe
Then follow the steps as shown below (from left to right by row):
1
3
2
4
55
Page 59
Operating Instructions
If the installed cradle driver is an older version,
13
14
it will be uninstalled and the new driver will be
installed.
7
9
8
10
11
56
12
Page 60
Operating Instructions
The first time test data is transferred to the
2 Setting PC Software Preferences
PC software preferences can be set using the following in the Preferences menu
P
rinting Test Results
application you will be prompted to establish your
printing preferences. You can also set your
preferences by selecting
Printing Test Results
Do not automatically print test results
will be taken when he instrument in placed in the
cradle
Print test results to the default printer
results will be sent to the default PC 8.5 x 11
printer. There will be no electronic copy saved.
Print test results to a PDF file in the default
directory
that can be named and saved for import into
electronic medical records systems or for printing
in the future.
: the test results will be sent to a PDF file
Preferences
from the menu.
and then
: no action
: the test
57
Page 61
Operating Instructions
Report Header
Users can optionally choose to customize the printed report header to include practice information
and/or logo as desired. Select
For Text only in the report header, enter the desired information into the four text lines available in the
Report Header dialog.
To add a custom logo to the report header, use the
logo file. Finish by selecting the position of the logo using the
dialog.
Preferences > Report Header
Change
.
button to set the path to a large or small
Header Options
at the bottom of this
C
ommunications
Select the communication port for the cradle with
Preferences > Communications
58
Page 62
Operating Instructions
Boys Town Norms Template
Users may optionally show the Boys Town Norms template on the DP-Gram of the PDF or printed test
report by selecting that option with
The Boys Town Norms template is explained in the following window. Users should refer to the
reference article for additional information.
Preferences > Show Boys Town Norms Template with DP Graphs.
59
Page 63
Operating Instructions
Screener
DP
TE
Combo
Protocol Name
Protocols Included
X X
Screening (DP)
Tymp 226 Hz > DP QuickScreen
X X
Screening (TE)
Tymp 226 Hz > TE QuickScreen
X X
Infant (DP)
DP QuickScreen > Tymp 1000 Hz
X X
Infant (TE)
TE QuickScreen > Tymp 1000 Hz
Diagnostic
DP
TE
Combo
Protocol Name
Protocols Included
X X
Diagnostic
Tymp 226 Hz > DP Dx 12
X X
Evaluation (DP)
Tymp 226 Hz > DP Eval 6
X X
Evaluation (TE)
Tymp 226 Hz > TE Eval 10
X X
Screening (DP)
Tymp 226 Hz > DP QuickScreen
X X
Screening (TE)
Tymp 226 Hz > TE QuickScreen
X X
Infant (DP)
DP QuickScreen > Tymp 1000 Hz
X X
Infant (TE)
TE QuickScreen > Tymp 1000 Hz
Screener w/ Ipsi Reflex
DP
TE
Combo
Protocol Name
Protocols Included
X X
Screening (DP)
Tymp 226 Hz > DP QuickScreen
X X
Screening (TE)
Tymp 226 Hz > TE QuickScreen
X X
Infant (DP)
DP QuickScreen > Tymp 1000 Hz
X X
Infant (TE)
TE QuickScreen > Tymp 1000 Hz
X X
Screening+ (DP)
Tymp 226 Hz > Ipsi 1F 90 dB > DP QuickScreen
X X
Screening+ (TE)
Tymp 226 Hz > Ipsi 1F 90 dB > TE QuickScreen
X X
Infant+(DP)
DP QuickScreen > Tymp 1000 Hz > Ipsi BN (1k)
X X
Infant+(TE)
TE QuickScreen > Tymp 1000 Hz > Ipsi BN (1k)
Diagnostic w/ Ipsi Reflex
DP
TE
Combo
Protocol Name
Protocols Included
X X
Diagnostic
Tymp 226 Hz > DP Dx 12
X X
Evaluation (DP)
Tymp 226 Hz > DP Eval 6
X X
Evaluation (TE)
Tymp 226 Hz > TE Eval 10
X X
Screening (DP)
Tymp 226 Hz > DP QuickScreen
X X
Screening (TE)
Tymp 226 Hz > TE QuickScreen
X X
Infant (DP)
DP QuickScreen > Tymp 1000 Hz
X X
Infant (TE)
TE QuickScreen > Tymp 1000 Hz
X X
Diagnostic+
Tymp 226 Hz > Ipsi 4F Auto > DP Dx 12
X X
Evaluation+ (DP)
Tymp 226 Hz > Ipsi 4F@90dB > DP Eval 6
X X
Evaluation+ (TE)
Tymp 226 Hz > Ipsi 4F 90 dB > TE Eval 10
X X
Screening+ (DP)
Tymp 226 Hz > Ipsi 1F 90 dB > DP QuickScreen
X X
Screening+ (TE)
Tymp 226 Hz > Ipsi 1F 90 dB > TE QuickScreen
X X
Infant+(DP)
DP QuickScreen > Tymp 1000 Hz > Ipsi BN (1k)
X X
Infant+(TE)
TE QuickScreen > Tymp 1000 Hz > Ipsi BN (1k)
Appendix B: Default Test Protocols and Protocol Series
60
Page 64
Operating Instructions
Appendix C: Test Technique and Sequence
Test Technique
As with any other OAE or Tympanometry instrument, there is a technique to learn when using the
ERO•SCAN Pro instrument, especially for infants and young children. Experience with existing systems
suggests that it may take up to 3 months to become completely proficient at selecting the proper eartip
and positioning the probe.
When testing an infant, the following suggestions might be helpful: The infant has to be relatively quiet
and calm; it is usually preferred for the infant to be asleep. A pacifier may be used to calm the infant;
however, sucking will add noise to the test and decrease the likelihood of a passing result. When
testing an infant, gently pull down and back on the pinna to straighten out the ear canal. Generally the
external probe is preferred for testing newborns. Position the probe pre-amp enclosure so that there is
no weight on the probe cable and it is directed away from the infant's hands. Place the probe in the
ear canal and, if necessary, wait for the infant to calm before beginning the test. Swaddling the infant
in a blanket can help prevent him/her from moving and pulling the probe out of the ear during testing.
When testing a young child, the following suggestions might be helpful: The child has to be relatively
quiet and calm. Quiet distracting toys should be used to distract the child during testing. Ideally these
toys are something the child can hold during testing. This will keep his/her hands occupied preventing
the child from pulling the probe out of the ear. Generally, the external probe is preferred for testing
young children. Secure the probe to the patient using the lanyard neckloop or the clip being sure to
keep any weight off the probe cable. Although it is a matter of preference, many users prefer to start
the test prior to inserting the probe in the ear. This allows the tester to focus on the patient.
Test Sequence
A complete test sequence consists of an AutoStart, calibration and test phase. The AutoStart phase
determines when the c
the level of the tones that will be applied during the actual t
during the t
Immediately after the test button is pressed, the A
both the quality and stability of the seal by measuring the response obtained from a sequence of test
tones. The stability of the seal is determined by comparing the responses obtained over time. When the
level of the response is within an acceptable range and is stable over time, the unit proceeds to the
c
alibration and test phase.
est phase to reduce the effect of transient noise bursts.
DPOAE
The calibration phase automatically measures the response obtained from a calibration tone that is
present prior to each set of test tones and calculates the voltage needed to obtain the desired pressures.
If one or more of the desired pressures cannot be obtained, the instrument continues with
measurement, but will display an error message at completion of the test if appropriate.
The t
est phase consists of measuring the response obtained from the pairs of test frequencies (f1, f2)
applied to the receivers. Two receivers are used, with each receiver generating one frequency in order to
reduce intermodulation distortion. Frequency domain estimates of the actual P1, P2, distortion (DP) and
noise floor (NF) are obtained via the discrete Fourier Transform, with a bin resolution of approximately
31 Hz. The NF estimate is obtained by averaging the power in the 4 closest (+/-2) bins to the DP bin.
alibration and test phase should proceed, while the calibration phase calibrates
est phase. Artifact rejection is employed
utoStart phase of the test begins. Autostart checks
61
Page 65
Operating Instructions
TEOAE
The calibration phase automatically measures the peak pressure obtained from a sequence of clicks and
calculates the voltage required to obtain the target peak pressure. If the desired peak pressure cannot
be obtained, the unit will use the maximum voltage.
The test phase consists of measuring the response obtained from repeated sequences of clicks applied
to the receivers. The click sequence is 3-1-1-1 repeated twice. Signal and noise floor estimates are
obtained by adding/subtracting the two response sequences respectively. The energy of the signal and
noise floor estimates in various frequency bands is obtained in real time and displayed once per second.
The average peak pressure of the stimulus is calculated after completion of the test.
Artifact rejection is employed during the t
use of an adaptive rejection threshold. The unit attempts to accept the quieter sections of the test, while
rejecting the noisier portions of the test. When the noise level is approximately constant during the test,
the instrument will tend to accept most of the data in the test. However, as the level of the noise
becomes more variable over time, the instrument will attempt to accept the quieter portions of the
recording. Noise estimates are obtained approximately 32 times per second and a suitable threshold is
estimated from the data. Data segments with a noise floor above this threshold are rejected, which
tends to lower the noise floor of the test. In order to reduce the possibility of obtaining an artificially low
noise floor, the minimum threshold level is limited.
est phase to reduce the effect of transient noise bursts by the
62
Page 66
Operating Instructions
Note: All manufacturer authorized protocols which provide a Pass/Refer outcome
Appendix D: Pass/Refer Criteria
Pass/Refer Criteria for DPOAE
The decision that a DPOAE exists is based on detecting a signal whose level is significantly above the
background noise level. This requires a statistical decision, since the random noise level in the DPOAE
filter channel can be expected to exceed the average of the random noise levels in the four adjacent
filter channels — used as the reference for comparison — roughly half the time.
Extended measurements of the noise distributions in both the DPOAE filter channel “DP level” and the rms average of the 4 adjacent channels “N level” indicate that the signal-to-noise ratio (the difference
between DP and N) has a standard deviation of 5.5 dB. This implies a 10% probability of seeing a 7 dB
SNR simply from the variability of the noise levels in the 2 filter sets.
Requiring an SNR of 6 dB in three out of four frequencies drops the probability of passing an ear with
moderate-to-severe hearing loss to less than 1%. Note: three of six frequencies at >7 dB SNR will also
ensure less than 1% probability of passing a moderate-to-severe hearing-impaired ear.
P
ass/Refer Criteria for TEOAE
The same basic principles that underlie DPOAE Pass/Fail criteria underlie TEOAE Pass/Fail criteria. In the
case of transients, requiring SNR of 4 dB at any three out of the six test frequencies drops the
probability of passing an ear with a moderate-to-severe hearing loss to less than 1%.
The SNR limits for transients are lower than the corresponding limits for distortion products primarily
because the traditional noise calculation used in TEOAE measurements (and in the ERO•SCAN Pro
instrument) gives a 3 dB lower SNR than the calculation used for DPOAEs. Without that difference, the
numerical SNR value for a PASS with the two methods would be quite similar.
have been verified in a metal cavity (the equivalent of an ear with moderate-to-severe
hearing loss) to have a less than 1% probability of passing the DPOAE or TEOAE test.
Pass/Refer criteria based on probability statistics of a cavity are independent of age
related normative data (Christensen & Killion, 1999).
Users with the goal of detecting mild hearing loss should ensure that the minimum
amplitude setting is turned on (see section 8.6) and should collect normative data
from the target patient population with the ERO-SCAN Pro to verify the protocol being
used meets the screening goal.
Users creating custom protocols should collect normative data to validate any custom
Pass/Refer criteria.
Preliminary ERO•SCAN Pro trials with infants indicate that the tester’s technique is the single most
important variable in the pass rate on normal-hearing infants. Some testers pick up the technique (see
Operating Instructions
comparable to those for other DPOAE equipment they have used for months; other testers take longer.
The ERO•SCAN Pro uses a novel noise-rejection algorithm (patent pending) that permits accurate
DPOAE and TEOAE measurements in background noise as high as 55-65 dB SPL (A-weighted). Briefly
explained, use of available memory in the ERO•SCAN Pro processor permits a post-hoc statistical
analysis that identifies those samples whose retention would improve the overall accuracy. Those
samples are included in the final analysis; the noisier samples are rejected.
Appendix C) with only a couple of days’ practice, producing pass rates
63
Page 67
Operating Instructions
Note: The box shown on the tympanogram indicates the normative area where the peak of the
related norms are used in setting Pass/Refer criteria.
The improved operation in noise with the new algorithm was so substantial that we conducted a
complete replica of our original validation tests in "fully impaired ear" cavities and were able to verify
that no increase in false negatives (false passes) was introduced. Under no test conditions was any such
degradation uncovered.
The artifact rejection can only reject the noisiest samples in a measurement period. If the ambient noise
level rises too high (and/or the eartip seal is poor), then all samples will be noisy and accurate
measurements will be impossible, in which case the test result will indicate “noisy.”
Occasional claims of extraordinarily low probabilities of missing an ear with hearing loss appear to be
based on poor statistics. As discussed by Gorga (1999), since the incidence of significant hearing loss is
roughly 2 per 1000, verifying a 99.7% accuracy would require testing hundreds of thousands of babies
with a given system. Thus to demonstrate that only 3 babies out of 1000 with hearing loss were missed
would require follow-up testing on 500,000 babies. To our knowledge, no one has performed such
tests to date.
Pass/Refer Criteria for Tympanometry:
Tympanometry measures sound reflection from the tympanic membrane, while air pressure is
varied in the ear canal. The tympanogram is a quick measurement providing a snapshot of the
overall status of the middle ear. The primary features of a tympanogram are the ear canal
volume, peak amplitude of the tympanogram (compliance), pressure point of the peak
(pressure), width of the tympanogram (gradient), and overall shape of the curve (Hall &
Mueller, 1997). Reported normative values for these measurements vary considerably
depending on subject population, methodology, and criteria for the study. For example, the
reported limits for pressure can vary by 50 dapa and compliance limits can vary by 0.30 ml
depending on the study cited (Gelfand, 2001; Hall & Mueller, 1997). Additionally, the range of
values found with various pathologies overlaps the range found with normal ears (Gelfand,
2001; Harris et al., 2005).
tympanogram is expected under normal conditions. The Pass/Refer result is based on presence or
absence of the peak within the bounding box. The limits selected are intended to separate normal
ears from significantly abnormal ears across most populations.
Users with the goal of detecting mild conditions of the middle ear should ensure that appropriate age
It is for these reasons that tympanograms are not used in isolation to diagnose or screen for
pathologies of the middle ear system, but rather are considered with additional audiological
measures, such as Otoacoustic Emissions and Acoustic Reflex measurements as part of a
diagnostic or screening battery.
ass/Refer Criteria for Acoustic Reflex:
P
An acoustic reflex, or contraction of the Stapedius muscle, occurs under normal conditions
when a sufficiently intense sound is presented to the auditory pathway. This contraction of the
muscle causes a stiffening of the ossicular chain which changes the compliance of the middle
ear system. For best results, this reflex measurement is automatically conducted at the air
pressure value where the compliance peak occurred during the tympanometric test. Stimulus
tones of varying intensities at 500, 1000, 2000 or 4000 Hz are presented as short bursts. If a
change in compliance greater than 0.05 ml is detected, a reflex is considered present.
64
Page 68
Operating Instructions
Note: The limits selected are intended to separate normal ears from significantly
norms are used in setting Pass/Refer criteria.
The required compliance change of 0.05 ml is an industry standard default for impendence
systems with automated acoustic reflex measurement and has been verified via real-ear testing
with the ERO-SCAN Pro to result in comparable pass/refer results to existing clinical standard
equipment (Grason-Stadler, 2005; Interacoustics, 2007; Interacoustics, 2011; Maico, 2005; and
Maico, 2011).
A correct interpretation of the measuring results can only follow in connection with the
tympanogram, the graphic reflex display and other actual data. But in principle a Stapedial
reflex indicates that the patient hears on the “stimulus ear” and that the sound lead on the “probe ear” functions.
abnormal ears across most populations.
Users with the goal of detecting mild conditions should ensure that appropriate age related
It is for these reasons that acoustic reflex measurements are not used in isolation to diagnose or
screen for pathologies of the middle ear system, but rather are considered with additional
audiological measures, such as Otoacoustic Emissions and Tympanometry measurements as
part of a diagnostic or screening battery.
65
Page 69
Operating Instructions
Figure 1
Figure 3 (see next page)
Appendix E: High Frequency Measurements
High Frequency DPOAE Measurements
In healthy young ears, distortion product otoacoustic emissions are normally present in the 6-12 kHz
region. Figure 1 shows ERO•SCAN measurements on a 12 year old. Figure 2 shows ERO•SCAN
measurements obtained on an adult male in his 60s.
The results in the figures below are consistent with our findings in testing 8 children (age 5-13 years)
and 12 adults (age 50-78 years): normal emissions to 12 kHz in children (Figure 1), and no response
above 8 or 10 kHz in older adults (Figure 2).
Although sufficient data to establish age-related PASS/REFER norms are not available, high-frequency
DPOAE measurements may still prove useful in:
Ototoxic drug monitoring in young patients (and older patients who have measurable OAEs)
Prospective studies of noise-induced hearing loss, especially in young subjects
Obtaining scientific data on the effect of aging on DPOAEs39
66
Figure 2
Page 70
Operating Instructions
Number of tests averaged
1 2 3 4 6
10
16
95% C.D.
15.2
10.7
8.8
7.6
6.2
4.8
3.8
80% C.D.
12.4
8.8
7.2
6.2
5.1
3.9
3.1
Important Considerations When Monitoring
Typical of all DPOAE and TEOAE measurements, the lower limit of measurement is determined by noise,
most of which has a Gaussian distribution. In the absence of an emission, both the signal and reference
channels contain nothing but noise. Just like tossing 10 coins simultaneously where the long-term
average will be five heads, it is not unusual to see seven or eight heads in any one toss. Similarly, an
apparent SNR of 5 dB or even 10 dB can occasionally appear even though the long-term average SNR in
a cavity is 0 dB. In the absence of an emission, a single SNR reading will exceed 5.5 dB one time in six
on the average. This is just as true at high frequencies (8, 10, and 12 kHz) as it is at lower frequencies
(2, 3, 4, 6 kHz). However, the noise floor is slightly higher at frequencies above 6 khz. To safeguard
against mistaking noise for the signal, it is recommended that minimum DPOAE amplitudes be used as
part of any protocol measuring frequencies above 6 khz.
To give an actual example: Figure 2 on the previous page indicates the absence of emissions for an
adult. Figure 3, a subsequent test on the same subject, shows apparent emissions at 8, 10, and 12 kHz,
with displayed SNRs of 7, 5, and 4 dB, respectively. Figure 3 was selected from a large number of tests
on that ear to illustrate that such readings can occur by chance.
There appear to be no published norms for the use of high-frequency DPOAEs for ototoxic monitoring,
but we can determine the significant difference between the averaged test results from one session to
the averaged test results of another session based on known statistical variation. Averaging the results
of several DPOAE test improves the reliability compared to a single test. This is particularly important
when DPOAE test results are used to monitor changes in hearing sensitivity at specific frequencies over
two or more different sessions.
Table 1, below, gives the number of tests required for the comparison between two test sessions.
The critical differences at two different confidence levels (80% and 95%) are shown. An 80%
confidence level is normally adequate for clinical testing while a 95% confidence level is common for
research reporting, where a reduced risk of error is normally required.
The
critical difference
chance. An example: You want to be reasonably (80%) certain that any shift beyond 5 dB will be
statistically significant. How many tests must be averaged in each session to obtain this result? The table
below indicates that the average of 6 tests in each session will give the desired critical difference of 5.1
dB. Since a 5 dB change is normally considered clinically significant, we suggest the use of at least six
tests during each session.
If a more sensitive test is required, more tests must be averaged. For example, averaging 10 tests in each
condition gives a 3.9 dB critical difference at the same 80% confidence level. In contrast, a 12.4 dB
change would be required at the 80% confidence level when comparing a single test from each session.
Alternately, a 50% increase in testing can be used to improve from an 80% to a 95% confidence level
at a given criterion. Example: The average of 4 tests will give a 6.2 dB critical difference at the 80%
confidence level; the average of 6 tests will provide the same 6.2 dB at the 95% confidence level.
Table 1
is the difference between two measurements that probably did not occur by
67
Page 71
Operating Instructions
Instructions for Averaging Results
For monitoring purposes, it is the DP level itself that should be averaged. The DP level indicates the
health of the outer hair cells and the middle ear, when testing in relatively quiet conditions. The SNR is
the measurement of choice in screening, where varying noise levels are present.
Average the DP level at each frequency of interest for the total number of tests performed during one
session. Repeat the same procedure for each subsequent session. Compare the frequency-specific
average from the prior session to the average from the current session using the critical differences
provided in Table 1.
Producing the Desired Eardrum SPL
Siegel (1994) reported that large differences between eardrum SPL and the SPL measured by the
microphone in an OAE probe could occur at high frequencies. In our own measurements, we have seen
occasional differences as large as 15 dB from the combination of wavelength effects and improper
eartip seating.
Following Harris et al (1989), Siegel recommended that at high frequencies the receiver drive required to
produce the desired eardrum SPL be predicted from a low-frequency probe measurement of ear canal
SPL (to adjust for individual ear canal volume differences) and a previously determined receiver-toaverage-eardrum calibration.
The ERO•SCAN follows Siegel’s recommendations, additionally imposing an upper limit of 70 dB SPL
measured at the microphone of the probe in order to minimize spurious distortion products. The P1 and
P2 SPL values printed by the ERO•SCAN unit are those measured by the microphone in the probe in
order to minimize spurious distortion products. These provide a check that neither probe is blocked, but
as described above may differ from actual eardrum SPLs.
68
Page 72
Operating Instructions
References
Christensen, L.A. & Killion, M.C. (1999). A Pass/Refer criterion for screening newborns using DPOAEs.
Paper presented at the International Evoked Response Audiometry Study Group, XVI Biennial
for a priori criteria and for multifrequency audiometric standards.
Grason-Stadler (2005). Reference Instruction Manual: FSI TympStar Version 2
Hall, J.W., & Mueller, H.G. (1997).