2.1 Intended Use .......................................................................................................................................2
3 Warnings, Cautions, and Errors .............................................................. 3
10.1 Preparing the Patient for Testing ..................................................................................................... 18
10.2 Turning On the Instrument ............................................................................................................... 18
10.3 Control Panel .................................................................................................................................... 18
Page 4
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
ii
8106556-5 02/15
10.4 Main Menu ....................................................................................................................................... 19
10.5 Selecting the Test Protocol ............................................................................................................... 19
10.6 Probe Check (Beginning a Test) ........................................................................................................ 20
10.7 Calibration and Test .......................................................................................................................... 21
10.9 Test Technique.................................................................................................................................. 22
17.8 Language ........................................................................................................................................... 42
17.9 Reset to Default ................................................................................................................................ 42
18 Advanced Options for DPOAE Testing ................................................. 43
18.1 Instructions for Customizing a Test Protocol .................................................................................... 43
18.2 Selecting the Level of Primary Tones ................................................................................................ 44
18.3 Setting the Averaging Time .............................................................................................................. 44
18.4 Setting the PASS SNR Level ............................................................................................................... 44
18.5 Setting the Number of Frequencies for PASS ................................................................................... 45
18.7 Save Protocol .................................................................................................................................. 45
19 Advanced Options for TEOAE Testing .................................................. 45
19.1 Instructions for Customizing a Test Protocol .................................................................................... 46
19.2 Selecting the Averaging Time ........................................................................................................... 47
19.3 Setting the PASS SNR Level ............................................................................................................... 47
19.4 Setting the Number of Frequencies for PASS ................................................................................... 48
19.6 Save Protocol .................................................................................................................................... 48
20 Cleaning and Maintenance ................................................................... 49
20.1 Cleaning and Disinfection ................................................................................................................. 49
Compliance
MAICO Diagnostics is an ISO 13485 certified corporation.
Page 8
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
1
8106556-5 02/15
Standard
Issue
Date
Title
ANSI/ASA 3.6
2010
Specification for Audiometers
IEC 60601-1
2007
Medical Electrical Equipment – General Requirements for
Basic Safety and Essential Performance, Ed. 3
IEC 60645-1
2004
Electroacoustics – Audiological equipment – Part 1: Puretone audiometers
IEC 60645-3
2007
Electroacoustics – Audiometric equipment – Part 3: Test
signals of short duration
IEC 60645-6
2010
Electroacoustics – Audiometric Equipment – Part 6:
Instruments for the measurement of otoacoustic emissions
ISO 14971
2007
Application of Risk Management to Medical Devices
ISO 10993
2009
Biological Evaluation of Medical Devices
EN 60601-1-2
2007
Medical electrical equipment – Part 1-2: General
requirements for basic safety and essential performance –
Collateral standard: Electromagnetic compatibility –
Requirements and Tests
MAICO Diagnostics warrants that this product is free from defects in material and
workmanship and, when properly used, will perform in accordance with applicable
specifications. If this instrument does not meet these criteria within one year of original
shipment, it will be repaired, or at our option, replaced at no charge when returned to our
service facility.
NOTE: Changes in the product not approved by MAICO Diagnostics shall void this warranty.
MAICO Diagnostics shall not be liable for any indirect, special or consequential damages,
even if notice has been given of the possibility of such damages.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED,
INCLUDING, BUT NOT LIMITED TO, ANY IMPLIED WARRANTY OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE.
For assistance with this ERO•SCAN® Otoacoustic Emission Test System contact your Special
Equipment Distributor or contact MAICO Diagnostics by phone at 888.941.4201 or by fax
at 952.903.4100.
2.1 Intended Use
The ERO•SCAN® OAE Test Instrument is indicated for testing of cochlear function in infants,
children, and adults by measuring otoacoustic emissions (OAEs). This instrument is suitable
for use in all settings, including hospitals, schools, physician's offices, and audiologist
practices. Factory defined protocols allow for simple screening measurements and user
customizable protocols allow for diagnostic evaluations.
The ERO•SCAN® is intended to be used by hearing healthcare professionals (i.e. ENT
doctors, audiologists) and/or technicians, neonatal nurses and school nurses who have been
trained by a hearing healthcare professional.
Page 10
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
3
8106556-5 02/15
The WARNING label identifies conditions or practices that may
present danger to the patient and/or user.
The CAUTION label identifies conditions or practices that could
result in damage to the equipment
The ERO•SCAN® Otoacoustic Emission Test System should be
charged using only the provided power supply. Injury to personnel
or damage to equipment can result when a three-prong to twoprong adaptor is connected between the ERO•SCAN® power supply
and an AC outlet.
No modifications of the equipment are allowed by anyone other
than a qualified MAICO representative. Modification of the
equipment could be hazardous.
The ERO•SCAN® product has been verified by an independent
laboratory to conform to international standards for EMC
(electromagnetic emissions and immunity). The user is advised to
avoid installation and use of this instrument in proximity with other
devices or equipment that may emit or be susceptible to
electromagnetic interference, including mobile phones. If the
instrument is used adjacent to other devices or equipment, the user
is instructed to verify that no disturbance is found in the operation
of this or other equipment in proximity.
This icon indicates that patient applied parts of the instrument
conform to IEC 60601-1:2005, Type B requirements.
WARNING
CAUTION
WARNING
3 Warnings, Cautions, and Errors
In this manual the following two labels identify potentially dangerous or destructive
conditions and procedures.
NOTE: Notes help you identify areas of possible confusion and avoid potential problems
during system operation.
Page 11
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
4
8106556-5 02/15
Instruments which bear the Underwriters Laboratories, Inc. label
should be interconnected with accessories that have the proper
electrical compatibility and are listed as meeting the requirements of
the UL Medical and Dental Equipment Standard. Connection of
accessories not meeting these requirements may result in electrical
leakage currents in excess of those allowed by the standard and
present a potential electrical shock hazard to the person being
tested.
Any program aimed at obtaining reliable measurements of
otoacoustic emissions should be staffed and supervised by
appropriately-trained individuals.
Page 12
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
5
8106556-5 02/15
Attach Probe
No probe is detected at the start of a test.
BT Device Not
Found
The paired Bluetooth® device cannot be detected. The device may be turned
off or too far away.
BT Error #xxx
There is an error condition with the Bluetooth® device. Check the status.
BT Not
Configured
The ERO•SCAN® instrument is not paired with any Bluetooth® device.
Device not
Responding
The printer is not responding to queries from the instrument.
Due For Service
Indicates calibration of instrument is recommended. Message will appear
upon the calibration due date set in the device. Message appears during
device startup once per day.
Fit Error
Cannot Obtain L
For a DP test, the desired level (L1 or L2) cannot be obtained within allowable
limits. User should refit the probe and retry the test.
Fit Error
Too High
For a DP test, the level of the calibration tone is too high. User should refit the
probe and retry the test.
Fit Error
Too Low
For a DP test, the level of the calibration tone is too low. User should refit the
probe and retry the test.
Limit Error
Overflow error during the calculation of the DFTs for a DP test. User should
repeat the test.
Memory Almost
Full
Saved tests are within 5 tests of the maximum limit.
Memory Full!
The maximum saved test limit is reached. The user will need to clear the
memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. The user must charge the
battery before additional tests can be performed.
Printer Error
Indicates a problem with the printer. Check the printer status.
Printer Paper
Out!
Indicates that printer paper has run out.
Time/Date Error
The clock is checked during power on to ensure it has not lost time and been
reset. In the case of clock reset, this message is shown. The user should set
the correct date/time.
NOISE / Orange
The indicator labeled ‘NOISE’ provides a visual indication (AMBER) that the
noise level measured during the test exceeds a nominal threshold.
Also used to indicate some error conditions and when the outcome of test is
REFER, NOISY, or NO SEAL.
TEST / Yellow
The indicator labeled ‘TEST’ provides a visual indication (YELLOW) that the
selected test is being performed. This indicator will remain on steady state
during the test function.
READY / Green
The indicator labeled ‘READY’ lets the user know that the instrument is not
currently performing a test function and that it is available to perform a test
function.
CHARGE / Blue
The indicator labeled ‘CHARGE’ provides a visual indication (BLUE) of the
battery recharging function and battery status. The rate of illumination of the
indicator provides a means of identifying the status of the charging function.
3.1 Status/Error Messages
Display Messages:
Indicator LEDs (lights):
Page 13
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
6
8106556-5 02/15
This product and its components will perform reliably only when
operated and maintained in accordance with the instructions
contained in this manual, accompanying labels, and/or inserts. A
defective product should not be used. Make sure all connections to
external accessories are snug and secured properly. Parts which may
be broken or missing or are visibly worn, distorted, or contaminated
should be replaced immediately with clean, genuine replacement
parts manufactured by or available from MAICO.
This product should not be used in the presence of fluid that can
come into contact with any of the electronic components or wiring.
Should the user suspect fluids have contacted the system
components or accessories, the unit should not be used until
deemed safe by a MAICO certified service technician.
Do NOT use in the presence of flammable gaseous mixtures. Users
should consider the possibility of explosions or fire when using this
device in close proximity to flammable anesthetic gases.
Do NOT use the ERO•SCAN® in a highly oxygen-enriched environment, such as a hyperbaric chamber, oxygen tent, etc.
Equipment is not user repairable. Repairs and battery replacement
must be performed by a qualified service representative only.
Use and store the instrument indoors only. It is recommended that
the instrument be operated within an ambient temperature range
of 15 °C / 59 °F to 35 °C / 95 °F and in relative humidity between
30 % and 90 % (non-condensing).
Transport and store the instrument in temperature between +5 °C /
+4 °F 1 to +40 °C / +104 °F.
Annual calibration recommended. Have an authorized service
technician perform electrical safety checks on the unit in order to
maintain continued compliance to IEC and UL 60601-1.
WARNING
CAUTION
4 Customer Responsibility
Page 14
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
7
8106556-5 02/15
If the system is not functioning properly, do not operate it until all
necessary repairs are made and the unit is tested and calibrated for
proper functioning in accordance with MAICO published
specifications.
Use only the disposable eartips designed for use with this
instrument.
Never insert the probe tube into the ear canal without affixing an
eartip.
The eartips are disposable and for single patient use only. Do not
clean or reuse eartips.
Probe tubes are disposable and should be replaced when clogged. If
a probe tube is reused after removal from the probe head, it will not
sit as tight as before.
Do not attempt to clean probe tubes. This may cause damage to
the probe.
Do not drop or otherwise cause undue impact to this device. If the
instrument is dropped or otherwise damaged, return it to the
manufacturer for repair and/or calibration. Do not use the
instrument if any damage is suspected.
CAUTION
5 Safety Precautions
The following safety precautions must be observed at all times. General safety precautions
must be followed when operating electrical equipment. Failure to observe these precautions
could result in damage to the equipment and injury to the operator or patient.
The employer should instruct each employee in the recognition and avoidance of unsafe
conditions and the regulations applicable to his or her work environment to control or
eliminate any hazards or other exposure to illness or injury.
It is understood that safety rules within individual organizations vary. If a conflict exists
between the material contained in this manual and the rules of the organization using this
instrument, the more stringent rules should take precedence.
The ERO•SCAN® is intended to be used by hearing healthcare professionals (i.e. ENT
doctors, audiologists) and/or technicians, neonatal nurses and school nurses who have been
trained by a hearing healthcare professional.
5.1 Cautions – General
Page 15
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
8
8106556-5 02/15
Do not open the case of the ERO•SCAN® Instrument. Refer servicing to
qualified personnel.
Do not touch the contacts on the bottom of the instrument and the
patient at the same time.
Do not connect the instrument to the patient and the PC at the same
time.
WARNING
WARNING
WARNING
5.2 Warning - Electric Shock Hazards
5.3 Warning - In Case of Emergency
In case of emergency, disconnect the instrument from the supply
mains by removing the micro-USB cable from the connector as shown
in Figure 2 on page 16.
5.4 Warning - Explosion
This system is not explosion proof. Do not use in the presence of
flammable anesthetics or other gases.
5.5 Warning - Battery Safety
This instrument contains a rechargeable lithium-ion battery. The battery
is not user replaceable and must be returned to an authorized MAICO
service location for repair.
5.6 Warning - General
Proper use of this device depends on careful reading of all instructions.
Page 16
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
9
8106556-5 02/15
WARNING
6 Recycling/Disposal
Many local laws and regulations require special procedures to recycle or dispose
of electrical equipment-related waste including batteries, printed circuit boards,
electronic components, wiring and other elements of electronic devices. Follow
all your respective local laws and regulations for the proper disposal of batteries
and any other parts of this system.
Below is the contact address for proper return or disposal of electronic wastes relating to
MAICO products in in Europe and other localities.
Batteries may explode or cause burns, if disassembled, crushed or
exposed to fire or high temperatures.
Page 17
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
10
8106556-5 02/15
No.
Symbol
IEC Pub.
Description
980 & 60601-1
Serial Number
980 & 60601-1
Date of Manufacture
980 & 60601-1
Manufacturer
980 & 60601-1
Caution, Consult Accompanying
Documents
980 & 60601-1
Return to Authorized Representative,
Special Disposal Required
980 & 60601-1
Reference Number
60601-1
B Patient Applied Part According to
IEC60601-1
980 & 60601-1
Consult Operating Instructions
60601-1
Keep Dry
60601-1
Transport and Storage Temperature
range
Logo
ECREP
980 & 60601-1
EU Authorized Representative
980 & 60601-1
Conforms to European Medical
Device Directive 93/42/EEC
REF
7 Regulatory Symbols
Page 18
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
11
8106556-5 02/15
Page 19
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
12
8106556-5 02/15
User Interface
Buttons
Probe Connector
Micro-USB
Connector
Figure 1
Charge Status
Indicator
Display
Test Status
Indicators
8 Introduction
The purpose of the ERO•SCAN® test system is to provide a rapid measurement and
documentation of Distortion Product Otoacoustic Emissions (DPOAEs) or Transient Evoked
Otoacoustic Emissions (TEOAEs) at several frequencies.
The ERO•SCAN® is available as a Screening or Diagnostic version. Please see chapter 18 and
19 for further information.
8.1 How Does the ERO•SCAN
®
Device Work?
The system consists of the instrument,
Micro-Probe, single-use eartips,
replaceable probe tubes, and other
accessories. The ERO•SCAN® instrument
contains the hardware and software for
generating the test stimuli, measuring and
displaying the OAEs, and storing the
results until they are printed. The plastic
housing contains circuit boards that
provide the signal processing and
Display the test results. The instrument
also contains a rechargeable lithium-ion
battery to power the device. The
instrument uses an organic light-emitting
diode (OLED) Display screen and three
light-emitting diodes (LEDs) to provide a
visual Display of test status to the
operator. Four membrane-type push
buttons located on the keypad of the
device allow the user to control testing
and printing, and to reset test protocols.
The Micro-Probe houses speaker and
microphone which produce test stimuli
and measure the sound pressure level
(SPL) present in the sealed ear canal.
Interface of the instrument to the ear
canal is accomplished through disposable
eartips, which fit onto the probe tube.
The disposable eartips are color coded to
facilitate easy selection by size.
Page 20
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
13
8106556-5 02/15
8.2 How are the Results Stored and Reported?
When the ERO•SCAN® is set in its default settings, the instrument will store the results from
one patient (left and right ear) in its non-volatile memory for subsequent printing. However,
the ERO•SCAN® instrument is capable of storing up to 250 test results. The results are
displayed via the OLED on the front of the device and are stored in the device’s internal
memory. After testing is completed, results can be printed using the printer and/or exported
to a computer. Test results are stored in the non-volatile memory so the operator can delay
printing until a later time if desired.
8.3 Sensitivity and Specificity
Sensitivity and specificity of this type of device are based on the test characteristics defined
by the user, and may vary depending on environmental and operating conditions. The
presence of otoacoustic emissions suggests normal outer hair cell function, which in turn
correlates to normal hearing. However, a passing result using this instrument is not an
indication that the full auditory system is normal. Thus, a PASS result should not be allowed
to override other indications that hearing is not normal. A full audiologic evaluation should
be administered if concerns about hearing sensitivity persist. A REFER test result should not
be assumed to be an indicator of a lack of auditory function, however, it should be followed
with full audiologic diagnostic testing.
Page 21
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
14
8106556-5 02/15
ERO•SCAN
®
Instrument
Micro-Probe
Box of Single Use Eartips
Package of Probe Tubes
Eartip Remover
Micro-USB Power Supply for Charging the Lithium-Ion Battery
Operating Instructions
PC Database Software with Manual
Micro B to A USB Cable for PC Communication/Charging
Protective Carrying Case
Optional:
Printer
Printer Power Supply for Charging Printer
Thermal Printer Paper Roll
9 Setup
9.1 Unpacking the System
The following is a list of parts shipped with each ERO•SCAN® system:
NOTE:
We highly recommend to use SanibelTM eartips in order to guarantee optimal test results.
If any of these parts are missing or you want to order optional accessories or Sanibel
disposables, contact your special equipment distributor or MAICO Diagnostic at (+49) 30
70 71 46-50. We recommend that you save the shipping box and packing materials in case
you need to store or ship the system.
9.2 Battery Charging
The ERO•SCAN® instrument is powered by an integrated rechargeable lithium-ion battery
(1800 mAh) providing 20 hours (1000 tests, minimum) of operation between full charging.
The battery status is indicated by the battery icon shown in the upper right corner of the
Main Menu (Display 1, page 19). Full battery charge is represented by a full battery symbol
on the Display and reduces to an empty battery in increments corresponding to the
discharge of the battery.
NOTE:
Battery life will vary depending on each product configuration.
The maximum capacity of this battery will decrease with time and usage.
For maximum battery life, do not completely deplete the battery, but charge when 5 %-
10 % of the battery remains.
Page 22
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
15
8106556-5 02/15
Page 23
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
16
8106556-5 02/15
Misalignment of the plug and socket can cause damage. The
plug and socket should be visually inspected prior to each
installation of the remote probe. If damage is observed, contact
MAICO Diagnostics.
CAUTION
Figure 4
Figure 3
Figure 2
The Micro-USB port on the bottom of the instrument is used for
charging via the power supply. Connect the charger as shown
in Figure 2.
The indicator labeled ‘CHARGE’ (Figure 3) provides a visual
indication (BLUE) of the battery recharging function and battery
status during operation.
During battery charging the indicator will be lit whenever the
Micro-USB connector is engaged and powered. The rate of
illumination of the indicator provides a means of identifying the
status of the charging function, and is defined as follows:
Steady-state illumination indicates that the battery is fully
charged. This identifies that the charging cycle has been
completed or was not implemented because the battery was
already fully charged.
Slow blink illumination indicates that the charging function
is in process.
Fast blink illumination indicates a fault condition in which
the user would be directed by the Operating Instructions to
return the instrument for service.
During instrument operation, the user is warned of a low
battery condition by the following illumination of the CHARGE
indicator:
Two fast blinks followed by a pause and then repeated until
the battery is charged.
9.3 Installing the Micro-Probe
Turn off the ERO•SCAN® and insert the Micro-Probe connector
into the socket on the top of the ERO•SCAN® (Figure 4). The
plug will fit in only one direction. A MAICO logo is on the probe
connector and will align with the instrument control panel.
Page 24
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
17
8106556-5 02/15
Figure 6
Figure 7
Figure 5
9.4 Attaching Eartips
The ERO•SCAN® instrument comes with a box of disposable
eartips that fit a variety of ear canal sizes. The probe tube must
have an eartip attached before inserting it into an ear canal. The
determination of the appropriate eartip size should be made by
persons with proper training and experience. The eartip must
seal the ear canal. The best test results are obtained when the
eartip is inserted deeply into the ear canal instead of flush with
the ear canal opening. Caution must be taken, however, to
ensure that the eartip does not extend too deeply into the ear
canal. Use only the eartips approved for use with the
instrument. The eartips are disposable and should be replaced
after each patient.
After selecting an eartip, push it onto the probe tube until it is
flush against the base of the probe tube (Figure 5). Twisting the
eartip slightly while pushing it onto the probe is recommended.
Be sure the eartip is fully seated on the probe. There should be
no gaps between the eartip and the collar of the probe head
(Figure 6).
9.5 Removing Eartips
In order to remove the eartip, grasp the eartip at the base using
the eartip removal tool and pull it smoothly straight off the
probe tube (Figure 7). If you do not have a Removal tool at
hand, grasp the eartip at the base with your fingers and twist it
while pulling it off the probe tube. Grasping the base of the
eartip will prevent the probe tube from being inadvertently
pulled out of the probe head along with the eartip.
NOTE: If the probe tube becomes dirty or clogged, it must be
replaced. See the section
chapter 20.3
for further information.
Page 25
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
18
8106556-5 02/15
Figure 8
10 Operating Instructions
10.1 Preparing the Patient for Testing
Otoscopic examination of the patient’s ear canals should be performed prior to testing.
Excessive cerumen or vernix in the ear canals may interfere with the test and give invalid or
incomplete results. Patients with excessive cerumen, debris, or foreign bodies in the ear
canals should be referred to a qualified professional for removal of the blockage prior to
testing.
Place the patient in a position that will allow easy access to the ear canal. Use the shirt clip
on the remote probe to secure the probe to clothing or bedding. The patient should remain
still and quiet while the test is being performed.
10.2 Turning On the Instrument
To turn on the ERO•SCAN® instrument, press the DOWN key located below the
instrument’s Display screen (see Figure 8). The yellow ‘TEST’ light will appear briefly just
above the Display screen. The green ‘READY’ light will remain on indicating the instrument
is ready for use. The Flash Screen will appear briefly. This Display will indicate the type of
instrument – Screener (SCR), Screener Plus (SC+), Standard (STD), or Combo (CMB),
software version, serial number (for example ME1234567), and a recalibration date.
If the battery is sufficiently charged, the ERO•SCAN® will power on and automatically check
the date and time. If there are no date/time errors detected, the Main Menu (Display 1,
page 19) will appear on the display.
If this is the first time the ERO•SCAN® is being used, or if you wish to change the date or
time, see the section
follow the instructions in this section to set the correct date and time.
Clock Settings
on page 37. If a time/date error message is indicated,
10.3 Control Panel
The ERO•SCAN® instrument uses 4 buttons to control all
functions of the instrument (Figure 8). These buttons are
arranged in a directional cursor format. The arrows on the
keypad (LEFT, RIGHT, UP, and DOWN) correspond to
the arrows that are used on the screen. The screen will indicate
which button to push by showing the appropriate arrow.
NOTE: The UP key will always bring the instrument back to
either the previous screen, menu or the main menu. The UP
key will also access the print command from the Main Menu.
Page 26
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
19
8106556-5 02/15
Battery Status
Date & Time
Selected Protocol
Start Right
Start Left
Ear Test
Change Protocol
and Settings
Display 2
10.4 Main Menu
Ear Test
Display 1
10.5 Selecting the Test Protocol
The currently selected protocol is shown on the Main Menu
(Display 1). To change the selected protocol press CHANGE at
the Main Menu. The Change Protocol Display will appear
(Display 2). Use the CHANGE arrow buttons to change the
selected protocol. Press the UP arrow to return to the Main
Menu to begin testing. Press the SETUP to enter the setup
menus.
For either DPOAEs or TEOAEs screener devices, there are two
default test protocols that vary by averaging time.
contains complete information on protocol settings. For
instructions for customizing diagnostic protocols, review
chapter 18 and 19.
NOTE: Screener default protocols cannot be customized.
Diagnostic protocols are customizable.
Appendix D
Page 27
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
20
8106556-5 02/15
Display 3
Display 4
Display 5
10.6 Probe Check (Beginning a Test)
To obtain a seal and measure emissions, gently insert the eartip
into the patient’s ear canal. It should fit snugly and comfortably.
The best test results are obtained when an eartip is inserted
deeply into the ear canal instead of flush with the ear canal. To
begin a test, insert the probe into the ear and select either the
LEFT or RIGHT arrow key to indicate which ear will be
tested.
After the test ear is selected, the Probe Check screen will be
shown. This Display shows a horizontal bar graph representing
the ear canal volume and at the same time the fitting of the
probe. When the bars are in the orange to the left (Display 3)
this represents a very large ear canal volume or indicates that
the probe is not inserted into the ear or placed properly.
When the bars are all the way to the right (Display 4), this
represents the volume is too small for the test to begin or
indicates that the probe tip is blocked. Reinsertion and/or
inspection of the probe tip is necessary.
When the probe check is anywhere in the green area
(Display 5), the ear canal volume is in the target area for testing
to begin. The test will begin automatically if the probe fit is
stable.
Appropriate adjustment of the probe/eartip position and eartip
size selection should be made until the indicator falls within the
green area and remains stable. If the test will not progress past
the Probe Check phase, change the probe tube, check that the
probe connector is fully seated in the socket and try again.
NOTE: Do not remove or insert the probe from the probe
connector unless the device is off.
When a seal is obtained, the unit will automatically begin
testing (AutoStart) and the yellow ‘TEST’ LED will illuminate
throughout the test.
Page 28
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
21
8106556-5 02/15
Display 6
Display 7
NOTE: To test children with PE tubes, the AutoStart needs to be
disabled. This is accomplished by first inserting the probe with
appropriate eartip into the ear canal and obtaining a proper
seal. To disable AutoStart at the main menu select the ear to be
tested by holding down the LEFT or RIGHT arrow keys for 3
seconds until the green ‘TEST’ light turns off. Once the key is
released, the ERO•SCAN® will calibrate and test as before.
10.7 Calibration and Test
The ERO•SCAN® instrument will automatically perform a
calibration prior to the start of each test. During calibration a
series of tones will be presented to the ear canal to calibrate the
levels of the frequencies to be tested.
Following calibration of the test tones, the test phase will begin
automatically. During the test phase, a set of bars should
appear on the Display (Display 6). These are the test results
which are displayed as the emissions are measured and can be
reviewed after the test is complete. Each column represents one
test frequency (DP) or frequency band (TE).
The ERO•SCAN® allows the user to select from two options for
viewing the results. The (SNR) graph view (Display 6) shows the
signal-to-noise ratio for each DP test frequency or TE test band.
The Value graph view (Display 7) shows the absolute emission
and noise levels for each DP test frequency or TE test band. For
additional information review the section
page 30.
Testing is complete when the green ‘READY’ light is illuminated.
Both the tester and patient should remain as still and quiet as
possible until the green light turns on.
NOTE: The UP arrow key can be used to abort a test in
progress. No record of an aborted test will be saved in memory.
Interpreting Results
on
Page 29
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
22
8106556-5 02/15
Display 8
10.8 Viewing Results
When testing is complete, a Display similar to Display 8 will
appear. The results of the test are automatically saved in
memory as soon as the test is complete. The results will be
saved even if the unit turns off or the batteries are temporarily
depleted. This screen again indicates the test ear and further
gives the results of the test.
“PASS” on the screen indicates that the patient passed the
screening.
“REFER” indicates that the patient did not pass the
screening.
“NOISY” indicates that excessive noise was present during
the test.
“NO SEAL” indicates that a seal was not maintained
throughout the test.
“FIT ERR” indicates inadequate probe placement in the ear
canal to produce target stimulus intensities.
When the test result is “NOISY,” “NO SEAL,” or “FIT ERR” the tester should reposition the
probe, selecting a different size eartip if necessary, and retest. If the test result is “REFER”
the tester can determine from the printout whether the test should be repeated. To review
the results, push the DOWN arrow key to return to the graph. After reviewing the results,
again push the DOWN arrow key to return to the Main Menu (Display 1, page 19).
NOTE: Completed tests are automatically saved. By default (Save L/R Mode), the
ERO•SCAN® instrument will save only the last test for each ear. Starting a new test for the
same ear will overwrite the existing test.
NOTE: Once the review screen is exited, results are no longer viewable within the device.
Printing or transferring to the PC is required for continued viewing options.
See the next section
ERO•SCAN® saves results.
See the section
ERO•SCAN® save mode options.
Instrument Settings – Save Mode
Managing Results
on page 24 for more information on how the
on page 40 for more information on the
10.9 Test Technique
As with other otoacoustic emission test instruments, there is a technique to learn when
using the ERO•SCAN® instrument, especially while testing infants. Experience with existing
OAE systems suggests that it may take up to 3 months to become completely proficient at
screening infants.
Page 30
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
23
8106556-5 02/15
When testing an infant with the ERO•SCAN® instrument, the infant has to be relatively
quiet and calm; it is usually preferred for the infant to be asleep. A pacifier may be used to
calm the infant, however, sucking will add noise to the test and decrease the likelihood of a
passing result. Pull gently down and back on the pinna to straighten out the ear canal.
Gently place the probe tube into the infant’s ear canal.
When testing children and adults, pull gently up and back on the outer ear during insertion
to straighten the ear canal and ensure proper placement.
10.10 Noise Sources
When the noise level exceeds the noise rejection limit of the instrument, the orange ‘NOISE’
light will appear. It is common for the ‘NOISE’ light (See Figure 1, page 12) to appear while
testing. The light will appear infrequently if the noise level in the ear canal is low, and it will
appear more often if the noise level in the ear canal is high. Otoacoustic emissions are very
low-level sounds. Any noise in the ear canal at the time of testing can mask this emission.
This noise can come from a variety of sources.
The largest source of noise can come from the patient. This is biological noise, such as
movement, coughing, sucking, talking, etc. The patient must be calm and not move or talk.
Ambient noise in the testing environment can also be a large source of noise during the
test. A properly sealed eartip can block a large amount of this noise, but performing the
testing in a relatively quiet environment is recommended.
10.11 Turning Off the Instrument
The ERO•SCAN® instrument has an automatic “shutdown” feature, designed to prolong
battery life. The unit will automatically shut down after 1 minute (default) of inactivity. To
turn it back on, simply press the large DOWN key. This feature can be re-programmed for
various periods of inactivity before “shutdown”. (See the
Auto Shutdown Time
NOTE: The UP arrow can be used to manually power off the instrument.
on page 39.)
Changing Instrument Settings –
Page 31
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
24
8106556-5 02/15
11 Managing Results
Users have the option of printing to the thermal printer or transferring results to the
ERO•SCAN® PC Database. Each ERO•SCAN® system will include one or more of these
options. Your specific options will vary depending on the configuration of the system
purchased.
11.1 Saving Results
The ERO•SCAN® instrument automatically saves the results of completed tests in the non-
volatile memory (meaning tests are saved even if the battery is temporarily discharged).
However, the ERO•SCAN® is not intended for long-term storage of test results.
NOTE: Users are strongly encouraged to print/transfer all test results at the completion of
testing to avoid potential loss of data.
When operating in the default "Save L/R" mode, the ERO•SCAN® instrument will save the
most recent test results for each ear and print/transfer only these results. This allows the
user to retest a patient after a “REFER” result and to print/transfer only the most recent test
result for each ear. It is recommended that the results be printed after each patient in the
default mode.
When operating in the "Save 250" mode, the ERO•SCAN® will save up to 250 tests. There
are two options in the “Save 250” mode:
1. The ERO•SCAN® will automatically number each test from 1 to 250. This
allows the user to save all tests for each patient (tests of the same ear are
NOT overwritten) and to test multiple patients before printing or
transferring results. In this mode, it is important to keep a record of the test
number(s) for each patient.
2. The ERO•SCAN® Database Software is used to transfer patient names to
the ERO•SCAN® and the ERO•SCAN® will display the names. When patient
names are used (patient names are uploaded from the ERO•SCAN®
Database Software to the ERO•SCAN® unit) the patient names are
displayed on the ERO•SCAN® Unit in the same order as displayed on the
Database Software. To move to a different name than the one displayed
on the ERO•SCAN® screen, use the LEFT or RIGHT arrows to cycle
through the names until the desired name is on the display. A patient
named “Unnamed” is always included at the beginning of the ERO•SCAN®
list for instances when a patient is being tested, but the patient name was
not transferred to the ERO•SCAN®. The “Unnamed” selection can be used
multiple times when transferring to the Database Software.
See
Instrument Settings – Save Mode
mode settings.
on page 40 for information on changing the save
Page 32
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
25
8106556-5 02/15
Display 9
11.2 Deleting Results
The ERO•SCAN® holds data in non-volatile memory. The data stays in the memory even
after data is printed by the thermal printer or downloaded to the ERO•SCAN® Database
Software. Data can be deleted through a few methods, depending on the Save mode.
Save L/R Mode:
A single test for the Left ear and a single test for the Right ear are held in
memory. Data is deleted when a new test for the left or right ear is
acquired.
NOTE: Following printing or data transfer to the PC software, all tests saved in memory are
marked for deletion and will be permanently deleted when a new test is started. It is not
necessary to manually clear the results.
Save 250 Mode:
Data is deleted when new Patient Names are uploaded from the Database
Software to the ERO•SCAN
deleted).
Data can be deleted using the Clear function in the System Menu
(page 39).
®
(a warning is provided that data will be
Data can be deleted in the ERO•SCAN® from the Database Software when the device is
connected to the PC and ERO•SCAN® Database Software is opened. When “Names” is
selected, the window allows data to be deleted via the Clear Instrument button. See the
ERO•SCAN® Database Software Manual
for further instruction.
11.3 Printing to a Thermal Printer
Printing to an optional thermal printer is by way of Bluetooth®
connection. First establish Bluetooth® pairing between the
ERO•SCAN® instrument and the printer by following the
instructions in the section
Device Pairing
NOTE: See the printer operating manual for instructions on
using the printer or chapter
Following instructions provided with your printer, be sure the
printer is on and ready for communication/printing. From the
ERO•SCAN® instrument Main Menu (Display 1, page 19), press
the UP button to enter the device connection screen
(Display 9). Press the CONNECT button.
on page 38.
Instrument Settings - Bluetooth®
MPT-II Mobile Thermal mini printer
.
Page 33
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
26
8106556-5 02/15
Display 10
The ERO•SCAN® will search for the paired printer showing
Display 10 while searching. When the printer is found, all the
test results that are stored in memory will print out
automatically.
The ERO•SCAN® instrument will power off when printing is
complete.
NOTE: All printed test results are marked for deletion, but will
continue to be stored in memory until a new test is started at
which time all tests in the memory will be erased. This allows
the user to reprint the tests if printing is unsuccessful (for
example, the paper runs out before printing is complete).
NOTE: See the ERO•SCAN® Database Software Manual for
instructions on printing from the Database Software.
11.4 Connecting to the PC Database Software
Connection to the PC database software is achieved by using the provided Micro-USB to
USB-A cable or via Bluetooth®.
USB Connection:
Plug the correct end (USB-A) of the cable into an available USB port on the computer and
the correct end (Micro-USB) of the cable into the port found on the bottom of the
ERO•SCAN® instrument.
The ERO•SCAN® instrument will detect the connection to the PC and wait for an action or
communication from the PC database software.
Bluetooth® Connection:
Refer to the ERO•SCAN® Database Software Manual for connection via Bluetooth®.
NOTE: See the ERO•SCAN® Database Software Manual for instructions on using the
application.
Page 34
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
27
8106556-5 02/15
MPT-II printer
MPT-II rechargeable battery pack
Thermal printer paper
Printer Power supply/charger with plug adapters
CAUTION
12 MPT-II Mobile thermal mini printer
Cautions
RISK OF EXPLOSION IF BATTERY IS REPLACED BY AN
INCORRECT TYPE.
DISPOSE OF USED BATTERIES ACCORDING TO LOCAL
REGULATIONS.
Do not place the battery in fire or heat the battery.
Do not connect the positive terminal and the negative
terminal of the battery to each other with any metal
object.
Do not place the batteries in microwave ovens, high-
pressure containers, or on induction cookware.
In the event that the battery leaks and the fluid gets into
one’s eye, do not rub the eye. Rinse well with water and
immediately seek medical care. If left untreated the
battery fluid could cause damage to the eye.
This equipment is for professional use only.
Do not open the cover while printing. The printer head
might be hot.
Do not expose the printer to any liquid.
Do not disassemble the printer.
If printer is stored for a longer period of time, store the
battery separately.
For correct battery charging make sure to follow
charging instruction.
General information
MPT-II printer kit includes:
Page 35
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
28
8106556-5 02/15
Figure 10
Insert battery as shown (Figure 10).
Figure 11
Green Power indicator will be lit if printer is
powered by battery (Figure 11).
Figure 9
12.1 Operation
Paper loading:
Battery pack insertion:
Open the lid by pushing on the sides, insert paper roll as shown,
and close the lid (Figure 9).
NOTE: Reorder paper from Maico or your local distributor.
Power on:
Push POWER BUTTON for two seconds in order to power
ON or OFF.
One short beep will be heard at power ON, two short beeps at
power OFF.
Page 36
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
29
8106556-5 02/15
Green LED indicator
Blue LED indicator
Status
Sound
Note
Off
Fast flash
Charging
-
Power On
Off
On
Charging
-
Power Off
Off
Slow flash
Battery nearly
discharged
-
-
Off
On
Charging
completed
-
Power On
Off
Off
Charging
completed
-
Power Off
On
Off
Power ON,
battery
-
-
Slow flash
Slow flash
Out of paper
Beep
-
Slow flash
Off
Sleep mode
-
-
Indicators:
Self-test:
Paper feed:
Bluetooth® setup:
Pairing process:
While printer is powered OFF, press and hold PAPER FEED
button, then press and hold POWER BUTTON
simultaneously. When beep is heard after approx. 3 seconds,
release both buttons, and a test page will print with information
on current status and character samples.
When powered press PAPER FEED button. Paper will feed as
long as the button is pressed.
Pairing with the instrument is necessary before printing via
Bluetooth® is possible.
Power on printer
Let the instrument / PC search for other Bluetooth
devices.
A device called MPT-II should occur in the list of devices.
When prompted enter password “1234”.
Finish pairing.
®
Page 37
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
30
8106556-5 02/15
NOTE: Do not have several printers powered on and within
range while searching.
It is possible to pair up to 8 instruments with one printer. If
more instruments are paired the oldest will be deleted.
For detailed instructions how to initiate Bluetooth® search,
please, refer to your device’s user manual.
NOTE: See the chapter
detailed information.
Bluetooth® Device Pairing
for more
Page 38
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
31
8106556-5 02/15
Figure 12
Technical specifications
Print mode:
Thermal line dot print
Printing width: 48 mm (1.9 in.)
Resolution: 8 dots/mm (203 dots per
in. (dpi))
Dots per line: 384 dots
Thermal paper:
Paper width = 56mm +/- 1 mm (2.2 in.
+/- 0.04 in.) max. 40 mm (1.6 in.)
diameter
Battery pack:
2-cell Li-Ion battery pack
7.4 V-1500 mAh
Power supply / charger:
12 V/1 A UE15WCP1-120125SPA
Maximum current consumption 0.5 A
See also Figure 12.
Physical properties:
Size: 102 mm x 75 mm x 45 mm
(4.02 in. x 2.95 in. x 1.77 in.) (W x L x
H)
Weight: 205 g including battery,
without paper
Environmental specifications:
Operational temperature range: -10 °C
to +50 °C (+14 °F to +122 °F)
Operational humidity range: 20 % to
85 %
Storage temperature range: -20 °C to
+70 °C (-4 °F to +158 °F)
Page 39
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
32
8106556-5 02/15
Signal
-to-Noise Ratio (SNR)
**Each column represents an individual frequency.
13 Interpreting Results
13.1 Understanding the Display
The display on the ERO•SCAN® test instrument will indicate the results of the test with a
graphic display. The display will be generated and shown during the test and can be
reviewed after the test is complete. The display might differ depending on the settings used
for the test.
Frequency
Figure 133
The display above shows a 6-column graph (Figure 13). Each frequency (DPOAEs) or
frequency band (TEOAEs) is indicated by one column.
The ERO•SCAN® allows the user to select from two options for viewing the results. The SNR
graph view shows the signal-to-noise ratio for each DP test frequency or TE test band. The
Value graph view shows the absolute emission and noise levels for each DP test frequency
or TE test band.
Page 40
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
33
8106556-5 02/15
Display 11
Display 12
13.2 SNR Graph View
Display 11 shows the SNR bar graph view. These are the signalto-noise ratio (SNR) test results which are displayed as the
emissions and noise floor are measured. Each column
represents one DP test frequency or TE frequency band. The
height of each column represents the SNR measured.
When a protocol with a “PASS” criterion has been selected, the
user will see a horizontal green line at the decibel level
corresponding to the SNR required for a pass. Green vertical
bars represent a pass, a yellow bar is a non-Pass at the
frequency band.
NOTE: Diagnostic protocols can display green or purple bars.
Green is displayed when a “PASS” criterion is set in the device.
For more information on this see
page 44.
Setting the PASS SNR Level
on
13.3 Value Graph View
Display 12 shows the Value graph view for right ear test. Red
circle symbols represent the absolute emission levels at each DP
test frequency or TE frequency band. For the left ear, dark blue
“X” symbols will represent the absolute emission levels at each
DP test frequency or TE frequency band. White upside-down
triangles represent the noise floor at each DP test frequency or
TE frequency band.
Refer to page 41 for instructions to switch the default setting of
the graphs between the SNR Bar Graph and Value Graph.
Page 41
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
34
8106556-5 02/15
Figure 14
14 Interpreting Printed Results
Results from the ERO•SCAN® can be transferred to a portable thermal printer or the
ERO•SCAN® Database Software for a full page printout. The information, as specified in
section 14.1 and 14.2, is included on both printouts. For instruction on printing from the
database, see the ERO•SCAN® Database Software Manual.
14.1 Understanding the DPOAE Printout
The following information is provided for each test (Figure 14):
1) The software version number (e.g.: V100.01)
2) The time and date of the test, based on the setting of
the internal clock; if the clock is set correctly, this time
and date will be correct
3) The test number (if operating in “Save 250” mode)
(e.g.:001)
4) The protocol selected (e.g.: DP 4s)
5) The averaging time used for this test (e.g.: 4 sec avg.)
6) Instrument/Probe serial number (SN)
7) The ear selected (Right or Left)
8) A PASS/REFER indication if there is a criterion set for the
selected protocol
9) The open blocks indicate the noise floor in dB SPL
10) The solid blocks indicate the emission level in dB SPL
11) The f2 frequency in kHz (e.g.: 2.0, 3,0, 4.0, 5.0)
12) The level of the emission in dB SPL (DP)
13) The noise floor in dB SPL (NF)
14) The signal-to-noise ratio (SNR) = DP – NF
Page 42
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
35
8106556-5 02/15
Figure 15
14.2 Understanding the TEOAE Printout
The following information is provided for each test (Figure 15):
1) The software version number (e.g.: V100.01)
2) The time and date of the test, based on the setting of
the internal clock; if the clock is set correctly, this time
and date will be correct
3) The test number (if operating in “Save 250” mode)
(e.g.:001)
4) The protocol selected (e.g.: TE 32s)
5) The averaging time used for this test (e.g.: 10 sec avg.)
6) Instrument/Probe serial number (SN)
7) The ear selected (Right or Left)
8) A PASS/REFER indication if there is a criterion set for the
selected protocol
9) The open blocks indicate the noise floor in dB SPL
10) The solid blocks indicate the emission level in dB SPL
11) The frequency band center (F)
12) The level of the emission in dB SPL (TE)
13) The noise floor in dB SPL (NF)
14) The signal-to-noise ratio (SNR) = TE – NF
Page 43
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
36
8106556-5 02/15
15 Rounding Results
The user needs to be aware that the SNR and single PASS criteria are calculated from the
full internal precision of the instrument, and not from the values shown in the printout for
the emission (TE) and noise floor (NF) estimates.
This approach is used to preserve the full precision of the test results, but can result in some
apparent errors in the printout due to the effects of rounding. In the printout example
above, assume the actual values at 1.5 kHz were TE = 4.5 dB, NF = -0.4 dB, which results in
SNR = 4.9 dB. The printout values are rounded up to the nearest integer and are shown as
TE = 5, NF = 0, and SNR = 5. This can result in what appears to be an error with regard to
the pass criterion. If the pass criterion is 5 dB while the actual SNR = 4.9, the printed value
will be 5, but a “P” will NOT be printed.
Again, the pass/refer criterion is based on the full precision of the results, and not the
rounded values that are printed. The full precision value for the SNR must be equal to or
greater than the pass criterion (5 dB in this example) for the “P” to be printed. A similar
apparent problem can occur in which the printed SNR value appears to be incorrect. If the
actual values were TE = 4.5 dB, NF = 0.4 dB, resulting in SNR = 4.1 dB, the printed values
would be TE = 5 dB, NF = 0, SNR = 4. The printed SN value of 4 dB appears to be an error,
but is in fact correct.
Page 44
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
37
8106556-5 02/15
Display 13
Display 14
Display 15
16 Clock and Date/Time Settings
When the ERO•SCAN® test instrument is first used, the correct date and time will need to
be set on its internal clock. The date and time are listed on the test printout as day-monthyear (e.g., 07-MAR-14). The clock should be set prior to testing, as changing it after tests
are saved will not change the date on the printout (i.e., whatever date was previously in
memory will be the date on the printout).
Seasonal time changes such as Daylight Saving Time will also require resetting the clock. If
the instrument is being powered on for the first time or if the instrument’s battery is
completely discharged and the battery is not charged within approximately one hour a
“TIME/DATE ERROR” message will occur. If this message appears, reset the time and date.
16.1 Accessing the Clock Menu
To change the time and date press CHANGE at the main
menu (Display 1, page 19) and then press SETUP at the
protocol selection Display (Display 2, page 19). The current date
and time presently set in the unit will be shown (Display 13). If
the time and date are correct, press the UP key to escape
back to the main menu.
NOTE: When a Teal arrow is displayed for a DOWN arrow,
this represents an alternative menu is available. To access, hold
the DOWN arrow for 3 seconds until the ‘READY’ light (green
LED) turns off and release the key.
16.2 Changing the Date/Time
If either the date or time is incorrect, press the CHANGE key
to access the menu to change the month (Display 14). Press the
LEFT or RIGHT keys to scroll forward or backward through
the months.
You will see the abbreviated name for each month (Display 14).
When the desired month appears on the display, press the
NEXT key to enter the day selection screen. Pressing the
LEFT or RIGHT keys will scroll through the days of the
month. Repeat this process for the year, hour, and minute using
the LEFT or RIGHT arrow keys to make the selection and the
NEXT key to advance to the next display.
When the correct minute is selected, pressing the DOWN
arrow key (Display 15) will return to the Main Menu. The time
and date changes are automatically saved.
Page 45
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
38
8106556-5 02/15
Display 16
Display 17
Display 18
17 Instrument Settings
The ERO•SCAN® instrument allows the user to change many of the instrument's settings or
functions. These settings include Bluetooth® Device Pairing, Clearing Test Results, Auto
Shutdown Time, Minimum Amplitude Value, Save Mode, Clock Mode, Language, and Reset
to Default Settings.
To access the menus to change these functions, press CHANGE at the main menu
(Display 1, page 19) and then press SETUP at the Protocol Change (Display 2, page 19) to
enter the Clock menu (Display 13, page 37). At the Clock menu, hold down the CHANGE
key for 3 seconds until the ‘READY’ light (green LED) turns off and release the key.
17.1 Bluetooth
®
Device Pairing
The Bluetooth® pairing menu (Display 16) allows the user to
pair the ERO•SCAN® unit with a Bluetooth® device, such as a
thermal printer or personal computer, for printing test results
and data transfer.
The ERO•SCAN® unit can be paired with only one device at any
time. To establish Bluetooth® pairing, turn on the device that
will be paired with the ERO•SCAN® unit (e.g.: thermal printer).
Then select DISCOVER to initiate discovery of available
Bluetooth® devices. The ERO•SCAN® will search for available
Bluetooth® devices for approximately 15 seconds. During this
time the user will see the message "Please Wait" on the
Display and the yellow ‘TEST’ LED will flash. Discovery can be
canceled by pressing the CANCEL button.
When discovery is complete, all discovered devices will be
shown in the order in which they were found (Display 17). A
compatible thermal printer will appear as "PRT-##-##" (e.g.:
PRT-54-81) and other devices will be shown by their name
which can vary depending on the device. Use the CHANGE
buttons to select the desired device and then use the PAIR
button to pair the ERO•SCAN® to the selected device.
Pairing will be confirmed (Display 18). The pairing process is
complete. Select Main Menu to exit the Bluetooth® pairing
menu.
NOTE: See the
Bluetooth® pairing is unsuccessful or if any error messages are
displayed.
Troubleshooting
section on page 51 if
Page 46
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
39
8106556-5 02/15
Display 19
Display 20
When pairing to a computer for use with the Database Manager Software, confirm the
computer has Bluetooth® availability or a Bluetooth® dongle will be required. Confirm the
Bluetooth® settings of the computer ‘Allow Bluetooth® devices to find this PC (See
Figure 16).
Figure 16 Bluetooth® Settings
Select the Bluetooth® pairing Menu (Display 16, page 38) and press DISCOVER to start
the pairing process. When the computer is seen on the device, select PAIR. While the
device is pairing to the computer a notification will Display on your computer “Add a
device“. Select this notification on your computer and enter PIN 1234. Once paired, the
instrument will show Display 18 on page 38. The first time the ERO•SCAN® Database
Manager Software is opened, select ‘Detect Com Port’ to finalize the ERO•SCAN®
connection.
17.2 Clearing Test Results
The Test Results Clear menu (Display 19) allows the user to clear
the test results stored in the unit without printing them. Select
the LEFT or RIGHT arrow key to clear the results and select
Yes or No to verify clearing or to cancel. To advance to the
next menu without clearing the results, press NEXT.
NOTE: Following printing or data transfer to the PC software, all
tests saved in memory are marked for deletion and will be
permanently deleted when a new test is started. It is not
necessary to manually clear the results using this menu.
17.3 Auto Shutdown Time
The Power Off menu (Display 20) refers to the Auto Shutdown
time which controls how long the ERO•SCAN® instrument
waits before shutting itself off after a period of inactivity. It is
not necessary to manually turn off the ERO•SCAN® unit. The
Automatic Shutdown feature is designed to prolong the battery
life of the instrument when it is not in use. By default, the
instrument automatically shuts off after 1 minute has elapsed.
The Auto Shutdown time may be increased or decreased by
pressing the CHANGE keys. The times available are 30
Page 47
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
40
8106556-5 02/15
Display 21
Display 22
seconds, 1, 2, or 4 minutes. Once you have made your
selection, press NEXT.
17.4 Save Mode/Storing Test Results
The ERO•SCAN® unit automatically stores only the most recent
test result for each ear L/R (Display 21), but has the capacity to
store 250 individual tests. To change the mode to save up to
250 tests, press the LEFT or RIGHT arrow keys to change the
menu to 250. Once you have made your selection, press
NEXT.
There are two options in the Save 250 mode (Display 22):
1. The ERO•SCAN® will automatically number each test
from 1 to 250
2. The ERO•SCAN® Database Software is used to transfer
patient names to the ERO•SCAN® and the ERO•SCAN®
will display the names. Up to 50 names can be stored in
the device and 250 tests.
When numbers are used (no patient names are uploaded from
the ERO•SCAN® Database Software to the ERO•SCAN®), each
test is automatically incremented, starting with test number 1.
When patient names are used (patient names are uploaded from the ERO•SCAN®
Database Software to the ERO•SCAN® unit) the patient names are displayed on the
ERO•SCAN® Unit in the same order as displayed on the Database Software. To move to a
different name than the one displayed on the ERO•SCAN® screen, use the left or right
arrows to cycle through the names until the desired name is on the display. A patient
named “Unnamed” is always included at the beginning of the ERO•SCAN® list for
instances when a patient is being tested, but the patient name was not transferred to the
ERO•SCAN®.
It is recommended that you go to the Test Results Clear screen (Display 19, page 39) to
clear any previous memory locations after you changed the save mode setting.
NOTE: When using the “250 test” mode, it is important to keep a record of the test
number for each patient. When 245 tests have been saved, the user will be warned that
the memory is almost full. When the ERO•SCAN® unit reaches 250 saved tests, it will not
allow any further testing. At this point either the results must be printed, transferred to the
PC software, or they must be cleared from memory.
Page 48
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
41
8106556-5 02/15
Display 23
Display 24
Display 25
17.5 Minimum Amplitude
The Minimum Amplitude setting allows the user to set the unit
to include minimum amplitude values in the pass/refer criterion
(Display 23). The ERO•SCAN® is set with this feature turned OFF
when it is shipped from the factory. If the MIN VALUE is set to
“ON”, a result is not considered a pass unless the amplitude at
each frequency is equal to or greater than the minimum value
programmed into the unit. This is in addition to meeting the
other pass criteria including the minimum SNR and the number
of passing frequencies for overall test “Pass”.
To change the mode to Minimum Amplitude setting, press the
LEFT or RIGHT arrow keys to make a selection. Once you
have made your selection, press NEXT.
The minimum DP amplitude when ON is -5 dB SPL.
The minimum TE amplitude selections are -5 dB SPL. and -10
dB SPL.
17.6 Clock Mode
17.7 Graph Style
The Clock Mode menu (Display 24) allows the user to change
the clock from a 24 hour mode to a 12 hour mode. To change
the clock mode, press the CHANGE keys. Press the NEXT
to exit this menu.
The Graph Style menu (Display 25) allows the user to select
from two options for viewing the results. The SNR graph view
shows the signal-to-ratio for each DP test frequency or TE test
band. The Value graph view shows the absolute emission and
noise levels for each DP test frequency or TE test band.
See the section
information.
Interpreting Results
on page 32 for further
Page 49
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
42
8106556-5 02/15
Display 26
Display 27
17.8 Language
17.9 Reset to Default
The Language setting (Display 26) allows the user to select
among several languages. To change the language, press the
CHANGE keys until the desired language is shown. Press
NEXT to exit this menu.
The Reset to Default menu (Display 27) will return all instrument
settings and protocol settings to their original factory defaults.
Select the LEFT or RIGHT arrow keys to reset and select
YES or NO to verify reset. To exit the System menu without
resetting the instrument, press NEXT to return to the Main
Menu.
NOTE: This will un-pair the Bluetooth® device, clear the test
results, and reset ALL system and protocol settings.
Page 50
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
43
8106556-5 02/15
18 Advanced Options for DPOAE Testing
NOTE: The Advanced Options for DPOAE Testing section is reserved for those units
purchased as a diagnostic device. Instruments purchased as a screener can skip to Section
20.
The Advanced Options menus permit modification of the test parameters and pass criteria
for the customizable DP protocols. Changes to the protocol should be made only by
qualified personnel, usually the administrator. If you are not familiar with the use of these
variables, do not attempt to change the protocols. Changes to any of these characteristics
may yield test results that differ from those obtained in other test modes.
The ERO•SCAN® instrument comes with pre-programmed protocol settings. See Appendix
D for the manufacturer settings of these protocols. Test protocol changes are saved in the
non-volatile memory so the settings will be retained even when the battery is discharged
temporarily.
18.1 Instructions for Customizing a Test Protocol
To enter the DPOAE Menu:
1) Press CHANGE at the main menu (Display 1, page 19).
2) Using the CHANGE buttons, select the DPOAE protocol you want to
customize (the “DP 4s” protocol is not customizable).
3) Press SETUP at the Protocol menu (Display 2, page 19).
4) At the Clock menu (Display 13, page 37) the teal arrow represents a
custom protocol menu is available. Hold down CHANGE key for 3
seconds until the ‘READY’ light (green LED) turns off.
5) At the New BT Device menu (Display 16, page 38) the teal arrow
represents a custom protocol menu is available. Hold down the
CHANGE key for 3 seconds until the ‘READY’ light (green LED) turns off.
You will now see the Level L1 screen (Display 28, page 44). You are in the DPOAE menu
and will be able to scroll the available protocol parameters with the NEXT button and
make changes by using the LEFT or RIGHT arrow keys to CHANGE the selection.
NOTE: If you push the DOWN arrow key without holding it for 3 seconds, you will scroll
through date and time, etc., rather than accessing the custom protocol menus.
Page 51
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
44
8106556-5 02/15
Display 29
Display 30
Display 28
Selecting the Level of Primary TonesThe intensity of the
primary tones (L1, L2) may be changed to any level between 40
dB SPL and 70 dB SPL. The level L1 will change in 1 dB
increments by pushing the LEFT or RIGHT arrow keys
(Display 28) and press NEXT to move to the L2 screen.
Press the NEXT key to set the level of L2 as with L1 above.
18.2 Setting the Averaging Time
The Averaging Time can be changed to one of four settings.
The Averaging Time will have a large impact on the time
required to perform the test and on the signal-to-noise ratio
(SNR). A 2 second average for 6 frequencies would produce a
test in about 18 seconds, which includes the Probe Check
sequence. The possible settings for the Averaging Time are as
follows:
0.5 sec., 1.0 sec., 2.0 sec., or 4.0 sec.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 29).
NOTE: Longer averaging times help to reduce the noise floor
which can improve the likelihood of obtaining a passing result,
particularly with a noisy patient (like a baby sucking a pacifier)
or in a noisy environment. However, shorter averaging times
may be preferred for young children and/or uncooperative
patients.
18.3 Setting the PASS SNR Level
In order to provide a PASS/REFER determination for each test,
the PASS SNR must be set. This number refers to the number of
decibels that the DPOAE signal must be above the noise to be
considered a PASS at that frequency. The limits for the PASS
SNR are 3 dB to 10 dB. Pressing the LEFT or RIGHT arrow
keys will increase or decrease the requirement. This requirement
is used in combination with the number of frequencies
(discussed below) to determine an overall PASS/REFER for each
test.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 30).
Page 52
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
45
8106556-5 02/15
Display 31
Display 32
Display 33
18.4 Setting the Number of Frequencies for PASS
The number of frequencies required for determining a PASS can
be set from 0 to 12. If the setting is on 0, then no indication of
PASS/REFER will be made. This setting is used in conjunction
with the PASS SNR (Display 30, page 44) to set the criteria for
the overall test PASS/REFER indication. For example, if the PASS
SNR is set to 5 dB and the number of frequencies for PASS is set
to 3, then the test must contain at least 3 frequencies where
the emission is at least 5 dB above the noise to indicate a PASS.
The number of frequencies for PASS should also be based on
the number of frequencies being tested. Setting the number of
frequencies for PASS to 5 when only 4 frequencies are being
tested would result in every test being labeled as a REFER.
NOTE: To disable the PASS/REFER indication set the number of
frequencies for pass to 0.
NOTE: Diagnostic protocols Display the SNR graph with purple
bars. Once the “Number of Frequencies for PASS” is set above
zero, green bars Display to identify this setting.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 31).
18.5 Reset Protocol
Selecting the RESET arrow key in the Reset Protocol menu
(Display 32) will return the selected protocol settings to their
original factory settings. Press the NEXT key to exit.
NOTE: This does not affect the instrument settings or the
settings of any other protocol.
18.6 Save Protocol
Once all of the settings have been selected for the protocol,
these settings can be saved by selecting the SAVE keys
(Display 33). Press the DONE key to exit.
Selection of the DONE before saving will not save the
selection made to the protocol.
19 Advanced Options for TEOAE Testing
Page 53
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
46
8106556-5 02/15
NOTE: The Advanced Options for TEOAE Testing section is reserved for those units
purchased as a diagnostic device. Instruments purchased as a screener can skip to Section
20.
The Advanced Options menu permits modification of the test stimuli and measurement
values for the TE Custom Protocol. Changes to the protocol should be made only by
qualified personnel, usually the administrator. If you are not familiar with the use of these
variables, do not attempt to change the protocols. Changes to any of these characteristics
may yield test results that differ from those obtained in other test modes.
The ERO•SCAN® instrument comes with pre-programmed protocol settings. See Appendix
D for the manufacturer settings of these protocols. Test protocol changes are saved in the
non-volatile memory so the settings will be retained even when changing the batteries.
19.1 Instructions for Customizing a Test Protocol
To enter the TEOAE Menu:
1) Press CHANGE at the main menu (Display 1, page 19).
2) Using the CHANGE buttons, select the TEOAE protocol you want to
customize (the "TE 64s" protocol is not customizable).
3) Press SETUP at the Protocol menu (Display 2, page 19)
4) At the Clock menu (Display 13. page 37) the teal arrow represents a
custom protocol. Hold down the CHANGE key for 3 seconds until the
‘READY’ light (green LED) turns off.
5) At the New BT Device menu (Display 16, page 38) the teal arrow
represents a custom protocol menu is available. Hold down the
CHANGE key for 3 seconds until the ‘READY’ light (green LED) turns off.
You will now see the Averaging Time screen (Display 34, page 47). You are in the TEOAE
menu and will be able to scroll the available protocol parameters with the NEXT button
and make changes by using the LEFT or RIGHT arrow keys to CHANGE the selection.
NOTE: If you push the DOWN arrow key without holding it for 3 seconds, you will scroll
through date and time, etc., rather than accessing the custom menus.
Page 54
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
47
8106556-5 02/15
Display 34
Display 35
19.2 Selecting the Averaging Time
The Averaging Time can be changed to one of five settings. The
Averaging Time will have a significant impact on the time
required to perform the test and on the signal-to-noise ratio
(SNR). A 4 second average would produce a test in about 4
seconds. A 32 second average would produce a test in about
32 seconds. This timing does not reflect the Probe Check
process. The possible settings for the Averaging Time are as
follows:
4, 8, 16, 32 or 64 seconds.
The instrument will automatically stop the test when the pass
criterion is met prior to the averaging time. Press the
CHANGE keys to select an option and the NEXT key to
exit (Display 34).
19.3 Setting the PASS SNR Level
In order to provide a PASS/REFER determination for each test,
the PASS SNR must be set. This number refers to the number of
decibels that the TEOAE signal must be above the noise to be
considered a PASS at that frequency. The limits for the PASS
SNR are 3 dB to 10 dB. Pressing the LEFT or RIGHT arrow
keys will increase or decrease the requirement. This requirement
is used in combination with the number of frequencies
(discussed below) to determine an overall PASS/REFER for each
test.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 35).
Page 55
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
48
8106556-5 02/15
Display 36
Display 38
Display 37
19.4 Setting the Number of Frequencies for PASS
The number of frequencies for determining a PASS can be set
from 0 to 6. If the setting is on 0, then no indication of
PASS/REFER will be made. This setting is used in conjunction
with the PASS SNR to set the criteria for the overall test
PASS/REFER indication. For example, if the PASS SNR is set to 4
dB and the number of frequencies for PASS is set to 3 then the
test must contain at least 3 frequencies where the emission is at
least 4 dB above the noise to indicate a PASS.
Press the CHANGE keys to select an option and the NEXT
key to exit (Display 36).
NOTE: Diagnostic protocols display the SNR graph with purple
bars. Once the “Number of Frequencies for PASS” is set above
zero, green bars display to identify this setting.
19.5 Reset Protocol
Selecting the RESET arrow key in the Reset Protocol menu
(Display 37) will return the selected protocol settings to their
original factory settings. Press the NEXT key to exit.
NOTE: This does not affect the instrument settings or the
settings of any other protocol.
19.6 Save Protocol
Once all of the settings have been selected for the protocol,
these settings can be saved by selecting the SAVE keys
(Display 38). Press the DONE key to exit.
Selection of DONE before saving will not save the selections
made to the protocol.
Page 56
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
49
8106556-5 02/15
Important:
• Do not immerse the instrument or probe in fluids or attempt
to sterilize the instrument or any of its accessories.
• Do not allow any fluid to enter the device.
• Do not use autoclave sterilization.
• Do not use alcohol-based disinfectants.
• Take care not to put excessive pressure on the clear display
window or allow any utensil to puncture the display window
or control panel.
CAUTION
20 Cleaning and Maintenance
20.1 Cleaning and Disinfection
Use new eartips for each patient. Eartips are for single patient use only. The probe tube,
which does not make direct contact with the patient, should be replaced if there is any sign
of contamination or if the test will not progress past the Probe Check phase. Disinfection of
the probe tube between patients is not required.
External parts of the instrument/probe can be cleaned to remove visible particulate
contamination. Do not attempt to insert any object into probe.
This instrument is not designated as a ‘sterile’ device. Wiping with a clean cloth or towel
and a mild non alcohol-based disinfecting solution, provides a suitable form of cleaning and
low-level disinfection of the housing and probe exterior. Repeat this weekly, or as often as
conditions warrant, to prevent a build-up of grime from normal handling and use.
We believe low-level disinfection is appropriate for this type of instrument. This may not
conform to the infection control guidelines of the user’s facility. The disinfection materials
and procedures applied in the users’ facility may be more appropriate for their
circumstances than the methods outlined above (see cautions below). The frequency of
cleaning and disinfecting is dependent on the facility’s risk assessment, usage, and test
environment.
NOTE: Long-term exposure to any disinfecting agents has the potential to alter the material
properties of the plastic housing and labeling of the device.
Always follow the safety and disposal guidelines given by the manufacturer of cleaning and
disinfectant chemicals.
20.2 Maintenance
This instrument requires no regular maintenance beyond routine cleaning and annual
calibration. The probe tube requires replacement only when it becomes clogged.
Page 57
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
50
8106556-5 02/15
To replace the probe tube, use the eartip to grasp the probe tube
(the clear plastic tube) and twist slightly while pulling the probe
tube straight out of the probe head (Figure 17).
Figure 17
Dispose of the used probe tube immediately to avoid confusing
used tubes and new tubes. Take a new probe tube from the
package and insert the tube into the probe head until it is fully
seated (Figure 18). A properly inserted probe tube will snap
securely into place when it is fully seated in the probe head.
Figure 18
NOTE: If the probe tube is re-used after it was removed from the
probe head once it will not sit as tight as before.
CAUTION
A defective product should not be used. Make sure all connections to external accessories
are snug and secured properly. Parts which may be broken or missing or are visibly worn,
distorted or contaminated should be replaced immediately with clean, genuine replacement
parts manufactured by or available from MAICO.
Equipment is not user repairable. Repairs and battery replacement must be performed by a
qualified service representative only.
Annual calibration is recommended. Have an authorized service technician perform electrical
safety checks on the unit in order to maintain continued compliance to IEC and UL 60601-1.
20.3 Probe Tube Replacement
Probe tubes are disposable and should be replaced when they become clogged. A package
of replacement probe tubes is included with this instrument. Do not attempt to clean the
probe tube.
DO NOT ATTEMPT TO CLEAN PROBE TUBES. THIS MAY CAUSE
DAMAGE TO THE PROBE.
Page 58
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
51
8106556-5 02/15
Problem
Solutions
Instrument does
not turn on
The DOWN arrow must be pressed for a full second (the
Yellow ‘TEST’ LED will illuminate).
Connect the charger as shown in Figure 2 on page 16. Confirm
that the blue "Charging" LED is illuminating in a slow blink
pattern. Wait at least 10 minutes and then attempt to turn on
the instrument
Contact MAICO for service if the problem persists.
The test will not
start
Select a different sized eartip.
Reposition the probe.
Change the probe tube.
Verify that the eartip is sealed in the ear canal via feedback
from the PROBE CHECK screen.
Check that the instrument will start in your own ear with the
proper eartip for testing yourself. If the test will not start or if
the AutoStart tones sound unusual, replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
The results will not
print
Check the printer status. Turn the printer on (wake from sleep
mode) by pressing the large button.
If the printer does not turn on, plug in the power supply to
charge the battery.
Be sure the printer has paper.
If paper comes out of the printer but there is no text on the
paper then the paper is in backwards.
Press the large printer button twice rapidly to run demo print.
Contact MAICO for service if the problem persists.
Display is frozen
and instrument will
not respond to
button presses
Press and hold the DOWN arrow button for 10 seconds to
force the instrument to power off. Powering the instrument
back on again should reset/restore normal function.
Contact MAICO for service if the problem persists.
21 Troubleshooting
Page 59
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
52
8106556-5 02/15
Message
Solutions
Attach
Probe
Probe not detected. Check that the probe connector is fully
seated in the socket.
Disconnect and reconnect the probe.
Cycle instrument power.
Contact MAICO for service if the problem persists.
BT Device
Not Found
Paired to Printer:
o Check that the printer is turned on.
o Move closer to the printer.
o Try again.
Paired to PC Computer or Bluetooth
®
dongle:
o Check that the serial port is open. Establishing the serial
port is handled by the PC and/or the software, not by
the ERO•SCAN® instrument.
BT Error
#xxx
Check BT device (printer or PC) status.
Attempt to connect to BT device again.
Contact MAICO if the problem persists.
BT Not
Configured
Printing has been attempted, but no BT device is paired with
the ERO•SCAN® Instrument. Establish Bluetooth® pairing.
Device Not
Responding
The printer is not responding to queries from the instrument.
Check printer status.
Awaken printer from sleep mode.
Charge printer battery if necessary.
Due for Service
Indicates calibration of instrument is recommended. Message
will appear upon the calibration due date set in the device.
Message appears during device startup once per day.
Fit Error
Cannot Obtain L
For a DP test, the desired level (L1 or L2) cannot be obtained
within allowable limits. User should refit the probe and retry
the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
Fit Error
Too High
For a DP test, the level of the calibration tone is too high. User
should refit the probe and retry the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
Fit Error
Too Low
For a DP test, the level of the calibration tone is too low. User
should refit the probe and retry the test.
Replace the probe tube.
Contact MAICO for service if the problem persists across
several patients.
Page 60
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
53
8106556-5 02/15
Message
Solutions
Limit
Error
Overflow error during the calculation of the DFTs for a DP test.
User should repeat the test.
Cycle instrument power.
Contact MAICO for service if the problem persists.
Memory almost full
Saved tests are within 5 tests of the maximum limit. Print or
transfer test result to avoid interruption in testing.
Memory Full!
The maximum saved test limit is reached. The user will need to
clear the memory before any additional tests can be performed.
Power Low!
The battery charge level is too low for operation. The user must
charge the battery before additional tests can be performed.
Printer Error
Indicates a problem with the printer. Check the printer status.
Reset the printer or cycle the printer power.
Printer Paper Out!
Replace the paper roll.
Time/Date Error
The clock is checked during power on to ensure it has not lost
time and been reset. In the case of clock reset, this message is
shown. The user should set the correct date/time.
Screener version:
DPOAE: 2.0 kHz to 5.0 kHz
TEOAE: 1.5 kHz to 4.0 kHz
Diagnostic version:
DPOAE: 1.5 kHz to 12.0 kHz
TEOAE: 0.7 kHz to 4.0 kHz
Stimulus Intensity Range:
DPOAE: 40 dB SPL to 70 dB SPL
TEOAE: 83 dB SPL peak equivalent (±3 dB)
Microphone System Noise:
-20 dB SPL at 2 kHz (1 Hz bandwidth) / -13 dB SPL at 1 kHz
(1 Hz bandwidth)
Dimensions and Weight:
Length: 1.0 meter (40 in.)
Weight: 28 g (1.00 oz.)
Connector:
HDMI
INSTRUMENT SPECIFICATIONS
Power Supply:
Lithium-Ion rechargeable
Battery Life:
1000 tests per charge, minimum
20 hours on-time
Dimensions and Weight:
Instrument Dimensions: 66 mm x 31 mm x 145 mm
(2.25 in. x 1.23 in. x 5.78 in.)
Instrument Weight: 180 g (6.4 oz.)
User Interface:
OLED Display to provide user information and progress
of measurement
4-button keypad to control instrument functions
Connectors / Communications:
Integrated USB communication capability for battery
charging and communication with PC-based database
programs or an optional printer
HDMI Connector for connection to the Micro-Probe
Integrated Bluetooth® Class 2 + EDR with SPP Protocol
for communication with optional printer
22 Specifications
Page 62
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
55
8106556-5 02/15
POWER SUPPLY SPECIFICATIONS (use only approved power supply)
Model No:
UE08WCP-050160SPA
Output:
5.0 V DC, 1.6 A
Input
100 V-240 V AC, 50 Hz - 60 Hz, 400 mA
ENVIRONMENTAL REQUIREMENTS
Operating Temperature:
15 ° C to 35 ° C (59 ° F to 95 ° F)
Operating Relative Humidity:
30 % to 90 % (non-condensing)
Maximum Operating Altitude:
2000 meters (6000 feet)
Transport and Storage
Temperature:
5 ° C to 40 ° C (41 ° F to 104 ° F)
Page 63
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
56
8106556-5 02/15
Appendix A: Flowcharts
A.1 Test Operation Flowchart
Page 64
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
57
8106556-5 02/15
A.2 Setup Menu Flowchart
NOTE: DPOAE Menu and TEOAE Menu are not accessible in the ERO•SCAN® Screening
version.
Page 65
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
58
8106556-5 02/15
Appendix B: Test Sequence
A complete test sequence consists of a Probe Check, calibration, and test phase. The Probe
Check phase determines when the calibration phase should proceed, while the calibration
phase calibrates the level of the tones that will be applied during the actual test phase.
Artifact rejection is employed during the test phase to reduce the effect of transient noise
bursts.
Immediately after the test button is pressed, the Probe Check phase of the test begins.
Probe Check phase checks both the quality and stability of the seal by measuring the
response obtained from a sequence of test tones. The stability of the seal is determined by
comparing the responses obtained over time. When the level of the response is within an
acceptable range and is stable over time, the unit proceeds to the calibration phase.
FOR DPOAE
The calibration phase automatically measures the response obtained from a sequence of
calibration tones and calculates the voltage needed to obtain the desired pressures. If the
desired peak pressure cannot be obtained, the unit will use the maximum voltage. A
successful calibration leads to the actual test phase.
The test phase consists of measuring the response obtained from the pairs of test
frequencies (f1, f2) applied to the receivers. Two receivers are used, with each receiver
generating one frequency in order to reduce intermodulation distortion. Frequency domain
estimates of the actual L1, L2, distortion (DP) and noise floor (NF) are obtained via the
discrete Fourier Transform, with a bin resolution of approximately 31 Hz. The NF estimate is
obtained by averaging the power in the 4 closest (+/-2) bins to the DP bin.
FOR TEOAE
The calibration phase automatically measures the peak pressure obtained from a sequence
of clicks and calculates the voltage required to obtain the target peak pressure. If the
desired peak pressure cannot be obtained, the unit will use the maximum voltage.
The test phase consists of measuring the response obtained from repeated sequences of
clicks applied to the receivers. The click sequence is 3-1-1-1 repeated twice. Signal and noise
floor estimates are obtained by adding/subtracting the two response sequences,
respectively. The energy of the signal and noise floor estimates in various frequency bands is
obtained in real time and displayed once per second. The average peak pressure of the
stimulus is calculated after completion of the test.
Artifact rejection is employed during the test phase to reduce the effect of transient noise
bursts by the use of an adaptive rejection threshold. The unit attempts to accept the quieter
sections of the test, while rejecting the noisier portions of the test. When the noise level is
approximately constant during the test, the instrument will tend to accept most of the data
in the test. However, as the level of the noise becomes more variable over time, the
instrument will attempt to accept the quieter portions of the recording. Noise estimates are
obtained approximately 32 times per second and a suitable threshold is estimated from the
data. Data segments with a noise floor above this threshold are rejected, which tends to
Page 66
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
59
8106556-5 02/15
lower the noise floor of the test. In order to reduce the possibility of obtaining an artificially
low noise floor, the minimum threshold level is limited.
Comment about Variations in the SNR Estimate
The user needs to be aware that the SNR estimate has an inherent statistical variation due to
the effects of random noise, especially when no emission is actually present. If a test is
performed with the instrument’s probe placed in a test cavity, it can be shown theoretically
that the SNR will be greater than 6 dB approximately 7 times out of 100. This is not a
limitation of the instrument, but a fundamental property of the method used to estimate
the SNR in all emission testing. In order to reduce the occurrence of this “false” emission,
the instrument limits the minimum value of NF, which has the effect of reducing the SNR for
tests that have a low noise floor. As the noise level of the test increases, the user will notice
that more “false” emissions will appear, which is to be expected.
Page 67
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
60
8106556-5 02/15
Appendix C: Pass/Refer Criteria
Pass/Refer Criteria for DPOAE
The decision that a DPOAE exists is based on detecting a signal whose level is significantly
above the background noise level. This requires a statistical decision, since the random noise
level in the DPOAE filter channel can be expected to exceed the average of the random
noise levels in the four adjacent filter channels — used as the reference for comparison —
roughly half the time.
Extended measurements of the noise distributions in both the DPOAE filter channel “DP
level” and the rms average of the 4 adjacent channels “N level” indicate that the signal-to-
noise ratio (the difference between DP and N) has a standard deviation of 5.5 dB. As shown
in Diagram 1 (page 60), this implies a 10 % probability of seeing a 7 dB SNR simply from
the variability of the noise levels in the 2 filter sets.
Requiring an SNR of 6 dB in three out of four frequencies drops the probability of passing
an ear with significant hearing loss to 1 % or less.
NOTE: By the binomial distribution, two of three frequencies at >8.4 dB or three of six
frequencies at >7 dB should also ensure less than 1 % probability of passing a moderatelysevere hearing-impaired infant.
Preliminary ERO•SCAN® trials with infants indicate that the tester’s technique is the single
most important variable in the pass rate on normal-hearing infants. Some testers pick up the
technique (see
producing pass rates comparable to those for other DPOAE equipment they have used for
months; other testers take longer.
Occasional claims of extraordinarily low probabilities of missing an ear with hearing loss
appear to be based on poor statistics. As discussed by Gorga (Mayo Clinic Teleconference,
1998), since the incidence of significant hearing loss is roughly 2 per 1,000, verifying a 99.7
% accuracy would require testing hundreds of thousands of babies with a given system.
Thus to demonstrate that only 3 babies out of 1,000 with hearing loss were missed would
require follow-up testing on 500,000 babies. To our knowledge, no one has performed
such tests to date.
Operating Instructions
section, page 18) with only a couple of day’s practice,
Page 68
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
61
8106556-5 02/15
Diagram 1
Pass/Refer Criteria for TEOAE
The same basic principles that underlie DPOAE Pass/Fail criteria underlie TEOAE Pass/Fail
criteria. In the case of transients, requiring SNR of 4 dB at any three out of the six test
frequencies drops the probability of passing an ear with a significant hearing loss to less
than 1 %.
NOTE: The SNR limits for transients are lower than the corresponding limits for distortion
products primarily because the traditional noise calculation used in TEOAE measurements
(and in the ERO•SCAN® instrument) gives a 3 dB lower SNR than the calculation used for
DPOAEs. Without that difference, the numerical SNR value for a PASS with the two
methods would be quite similar.
The ERO•SCAN® uses a novel noise-rejection algorithm (patent pending) that permits
accurate DPOAE and TEOAE measurements in background noise and babble as high as 55
dB SPL to 65 dB SPL (A-weighted). Briefly explained, use of available memory in the
ERO•SCAN® processor permits a post-hoc statistical analysis that identifies those samples
whose retention would improve the overall accuracy. Those samples are included in the final
analysis; the noisier samples are rejected.
The improved operation in noise with the new algorithm was so substantial that we
conducted a complete replica of our original validation tests in "fully impaired ear" cavities
and were able to verify that no increase in false negatives (false passes) was introduced.
Under no test conditions was any such degradation uncovered.
Page 69
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
62
8106556-5 02/15
The artifact rejection can only reject the noisiest samples in a measurement period. If the
ambient noise level rises too high (and/or the eartip seal is poor), then all samples will be
noisy and accurate measurements will be impossible, in which case the test result will
indicate “noisy”.
Diagram 2
Page 70
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
63
8106556-5 02/15
Name
# of
Freq.
F2 Freq. [kHz]
L1/L2
Averagin
g Time
Pass
SNR
# Passing Freq.
for Test Pass
Screening
DP 4s
4
2, 3, 4, 5
65/55
4 sec
6 dB
3
DP 2s
4
2, 3, 4, 5
65/55
2 sec
6 dB
3
Diagnostic
DP
2.0-5.0
4
2, 3, 4, 5
65/55
**
4 sec**
6 dB**
3**
DP
1.5-6.0
6
1.5, 2, 3, 4, 5, 6
65/55
**
4 sec.**
6 dB**
0**
DP
1.6-8.0
12
1.6, 2, 2.5, 3.2,
3.6, 4, 4.5, 5,
5.6, 6.3, 7.1, 8
65/55
**
4 sec.**
6 dB**
0**
DP
1.5-12
12
1.5, 2, 3, 4, 5, 6,
7, 8, 9, 10, 11,
12
65/55
**
4 sec.**
6 dB**
0**
Name
# of Freq.
Bands
Freq. center
bands [kHz]
Averaging
Time (max)
Pass SNR
# Passing Freq.
for Test Pass
Screening
TE 64s
6
1.5, 2, 2.5,
3, 3.5, 4
64
4 dB
3
TE 32s
6
1.5, 2, 2.5,
3, 3.5, 4
32
4 dB
3
Diagnostic
TE1.5-
4.0
6
1.5, 2, 2.5,
3, 3.5, 4
64 sec**
4 dB**
3**
TE0.7-
4.0
6
0.7, 1, 1.4,
2, 2.8, 4
64 sec.**
4 dB**
0**
Appendix D: Configurations and Test Protocols
DPOAE Protocols
(Diagnostic version also includes DP 4s screening protocol)
TEOAE Protocols
(Diagnostic version also includes TE 64s screening protocol)
** Customizable fields:
L1/L2 : DP: 40 to 70 dB SPL
Average time : DP: 0.5, 1.0, 2.0 or 4.0 sec, TE 4, 8, 16, 32 or 64 sec.
Pass SNR : DP and TE: 3 dB to 10 dB
Passing Freq. for Test Pass : DP and TE: 1 to 12
Page 71
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
64
8106556-5 02/15
Appendix E: EMC Compatibility
Portable and Mobile RF communications equipment can affect the ERO•SCAN®. Install and operate
the ERO•SCAN
The ERO•SCAN
not use the ERO•SCAN
stacked use is necessary, the user should verify normal operation in the configuration.
The use of accessories, transducers and cables other than those specified, with the exception of
servicing parts sold by MAICO Diagnostics as replacement parts for internal components, may result
in increased EMISSIONS or decreased IMMUNITY of the device. Anyone connecting additional
equipment is responsible for making sure the system complies with the IEC 60601-1-2 standard.
Electromagnetic Compatibility
Although the instrument fulfills the relevant EMC requirements precautions should be taken to avoid
unnecessary exposure to electromagnetic fields, e.g. from mobile phones, etc. If the device is used
adjacent to other equipment it must be observed that no mutual disturbance appears.
Electrical Safety, EMC and Associated Standards
1. UL 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety
2. IEC/EN 60601-1: Medical Electrical Equipment, Part 1 General Requirements for Safety
3. CAN/CSA-C22.2 No. 60601-1: Medical Electrical Equipment, Part 1 General Requirements for
Safety Electrical Equipment for Laboratory Use
4. IEC/EN 60601-1-1: Collateral Standard, Safety Requirements for Medical Electrical Systems
5. IEC/EN 60601-1-2: Medical Electrical Equipment, Part 1 - Electromagnetic Compatibility -
Requirements and Tests
6. Essential Requirements of the current European Union Medical Device Directive 93/42/EEC
®
according to the EMC information presented on this page and the next 4 pages.
®
has been tested for EMC emissions and immunity as a standalone instrument. Do
®
adjacent to or stacked with other electronic equipment. If adjacent or
7. RoHS (Restriction of the use of certain Hazardous Substance)
Guidance and Manufacturer’s Declaration - Electromagnetic Emissions
The ERO•SCAN
®
is intended for use in the electromagnetic environment specified below. The customer or
the user of the ERO•SCAN
®
should assure that it is used in such an environment.
Emissions Test
Compliance
Electromagnetic environment - Guidance
RF Emissions
CISPR 11
Group 1
The ERO•SCAN
®
uses RF energy only for its
internal function. Therefore, its RF emissions are
very low and are not likely to cause any
interference in nearby electronic equipment.
RF Emissions
CISPR 11
Class B Limits
The ERO•SCAN
®
is suitable for use in all
commercial, industrial, business, hospital, and
residential environments.
Harmonic Emissions
IEC 61000-3-2
Class A Category
Voltage Fluctuations /
Flicker Emissions
IEC 61000-3-3
Complies
Recommended Separation Distances between Portable and Mobile RF
Communications Equipment and the ERO•SCAN®
The ERO•SCAN
®
is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the ERO•SCAN
®
can help prevent electromagnetic interferences by
maintaining a minimum distance between portable and mobile RF communications equipment (transmitters)
and the ERO•SCAN
®
as recommended below, according to the maximum output power of the
communications equipment.
Rated Maximum
Output Power of
Transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.17√𝑃
80 MHz to 800 MHz
d
= 1.17√𝑃
800 MHz to 2.5 GHz
d
= 2.23√𝑃
0.01
0.12
0.12
0.23
0.1
0.37
0.37
0.74
1
1.17
1.17
2.33
10
3.70
3.70
7.37
100
11.70
11.70
23.30
For transmitters rated at a maximum output power not listed above, the recommended separation distance d
in meters (m) can be estimated using the equation applicable to the frequency of the transmitters, where P is
the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
Page 73
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
66
8106556-5 02/15
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ERO•SCAN
®
is intended for use in the electromagnetic environment specified below. The customer or
the user of the ERO•SCAN
®
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance
Electromagnetic
Environment-Guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material the
relative humidity should
be greater than 30%.
Electrical Fast
Transient/Burst
IEC 61000-4-4
±2 kV for power supply
lines
±1 kV for input/output
lines
±2 kV for power supply
lines
±1 kV for input/output
lines
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment.
Surge
IEC 61000-4-5
±1 kV differential mode
±2 kV common mode
±1 kV differential mode
±2 kV common mode
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment.
Voltage Dips, Short
Interruptions and
Voltage Variations on
Power Supply Lines
IEC 61000-4-11
<5 % UT
(>95 % dip in UT) for
0.5 cycle
40 % UT
(60% dip in UT) for 5
cycles
70 % UT
(30% dip in UT) for 25
cycles
5 % UT
(>95% dip in UT) for 5
sec
<5 % UT
(>95 % dip in UT) for
0.5 cycle
40 % UT
(60% dip in UT) for 5
cycles
70 % UT
(30% dip in UT) for 25
cycles
5 % UT
(>95% dip in UT) for 5
sec
Mains power quality
should be that of a typical
commercial, hospital, or
residential environment. If
the user of the
ERO•SCAN
®
requires
continued operation
during power mains
interruptions, it is
recommended that the
ERO•SCAN
®
be powered
from an uninterrupted
power supply.
Power Frequency (50/60
Hz)
IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should be
at levels characteristic of
a typical location in a
typical commercial or
hospital environment.
Note: UT is the a.c. mains voltage prior to application of the test level.
Page 74
Operating Instruction
ERO•SCAN®
Screening and Diagnostic Version
operating_instructions_8106556-6_EROSCAN_e_15a
67
8106556-5 02/15
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The ERO•SCAN
®
is intended for use in the electromagnetic environment specified below. The customer or
the user of the ERO•SCAN
®
should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Electromagnetic Environment-Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150 kHz
to 80 MHz
3 V/m 80 MHz to
2.5 GHz
3 Vrms
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any part
of the ERO•SCAN®, including cables than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
Recommended separation distance
d
= 1.17√𝑃
d
= 1.17√𝑃 80 MHz to 800 MHz
d
= 1.17√𝑃 800 MHz to 2.5 GHz
where P is the maximum output power rating of
the transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in meters (m).
Field Strengthens from fixed RF transmitters, as
determined by an electromagnetic site survey
(a*), should be less than the compliance level in
each frequency range (b*).
Interference may occur in the vicinity of
equipment marked:
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply to all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
(a*) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength in the location in which the
ERO•SCAN
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as
re-orienting or relocating the ERO•SCAN®.
(b*) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
®
is used exceeds the applicable RF compliance level above, the ERO•SCAN® should be observed to