Magnescale MF10-P1, MF10-P2 Instruction Manual

[For U.S.A. and Canada]
Digital Tolerance Indicator
MF10-P1/MF10-P2MF10-P1/MF10-P2
Read all the instructions in the manual carefully before use and strictly follow them. Keep the manual for future references.
Instruction Manual
PRECAUTIONS ON SAFETY
Meanings of Signal Words
CAUTION
Warning Indications
Do not use the product with voltage in excess of the rated voltage. Excess voltage may result in malfunction or fire.
Never use the product with an AC power supply. Otherwise, explosion may result.
The following precautions must be observed to ensure safe operation of the product. Doing so may cause damage or re.
Installation Environment
· Do not use the product in environments subject to ammable or explosive gases.
· To secure the safety of operation and maintenance, do not install the product close to high-voltage devices and power devices.
· Do not use the product in any atmosphere or environment that exceeds the ratings.
· Do not use the product in environments subject to exposure to water, oil, chemicals, etc.
Power Supply and Wiring
· Do not impose voltage exceeding the rated voltage: 10 to 30 VDC, including 10% ripple (p-p).
· Do not apply voltages or currents that exceed the rated ranges.
· When supplying power to the product, make sure that the polarity of the power is correct, and do not connect to an AC power supply.
· Do not miswire such as the polarity of the power supply.
· Do not apply any load exceeding the ratings.
· Connect the load correctly.
· Do not short both ends of the load.
· Do not short-circuit the open collector output load.
· Be sure to turn OFF the power when you plug/unplug the connector with the measuring unit, connect/disconnect with the digital tolerance indicator, or add digital tolerance indicators.
· High-Voltage lines and power lines must be wired separately from this product. Wiring them together or placing them in the same duct may cause induction, resulting in malfunction or damage.
Installation
· Do not install the product in locations subjected to strong magnetic eld or electric eld.
Others
· Do not attempt to disassemble, repair, or modify the product in any way.
· Do not use the product if the case is damaged.
· When disposing of the product, treat it as industrial waste.
· When making setting, be sure to check safety such as by stopping the equipment.
Installation Location
· Do not install the product in the following locations.
(1) Locations subject to direct sunlight (2) Locations subject to condensation due to high humidity (3) Locations subject to corrosive gas (4) Locations subject to vibration or mechanical shocks exceeding the rated values (5) Place where there are dusts, salt contents or iron powders
Power Supply and Wiring
· It may take time for the measurement to stabilize right after the power is turned ON, depending on the environment.
· Output pulses may occur when the power supply is turned OFF. Turn OFF the power supply to the load or load line rst.
· The product is ready to operate 1.5s after the power supply is turned ON. If the digital tolerance indicator and load are connected to power supplies separately, turn ON the power supply to the product rst.
· Be sure to turn OFF the power when you plug/unplug the connector with the measuring unit, connect/disconnect with the digital tolerance indicator, or add digital tolerance indicator.
· Extended length on the digital tolerance indicator end must be up to 30 m. For extension, use a cable with 0.3 mm
Installation
· Do not apply the forces on the cord exceeding the following limits:Pull: 40 N; torque: 0.1 N·m; pressure: 20 N; bending: 3 kg
· Do not pull or twist the measuring unit connector with excessive force when it is fixed to the digital tolerance indicator. (9.8 N or less)
Others
· Always keep the protective cover in place when using the product. Not doing so may cause malfunction.
· Do not use thinner, benzine, acetone, and lamp oil for cleaning.
· Digital tolerance indicator: 1 · Instruction Sheet (this sheet)
Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury or in property damage.
PRECAUTIONS
PRECAUTIONS FOR SAFE USE
PRECAUTIONS FOR CORRECT USE
2
larger.
Checking the Package Content
THIS CLASS A DIGITAL DEVICE COMPLIES WITH PART15 OF THE FCC RULES AND THE CANADIAN ICES-003. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS. (1) THIS DEVICE MAY NOT CAUSE HARMFUL INTERFERENCE, AND (2) THIS DEVICE MUST ACCEPT ANY INTERFERENCE RECEIVED, INCLUDING
INTERFERENCE THAT MAY CAUSE UNDERSIGNED OPERATION.
CET APPAREIL NUMÉRIQUE DE LA CLASSE A EST CONFORME À LA NORME NMB-003 DU CANADA.
Installation
1
1-1 Dimensions
10
114 (Max. with the protective
cover open)
Dimensions in parentheses ( ) indicates the ones with related components. The cover could come off if it is tilted by 170 degrees or more.
33.5
2.6
30.2
27.8
10
4.2
16
7.8 (Max. with the protective cover open)
170° (Max. with the protective cover open)
3.4
32.1 (49.5)
5.2
83.4
Ø4
6.9
12.7
Unit: mm
1-2Input/Output Circuit Diagram
MF10-P1 (NPN type) MF10-P2 (PNP type)
Overcurrent Protection Circuit
Main Circuit
3.3VDC
Load
Load
Brown
Black: Control Output 1
Orange: Control Output 2
PLC, etc.
External Input
(Operating current
Pink
1 mA max.)
Blue
10 to 30 VDC
0V
Overcurrent Protection Circuit
Main Circuit
Brown
Black: Control Output 1
Orange: Control Output 2
External Input
Pink
Blue
10 to 30 VDC
PLC, etc. (Operating current 3 mA max.)
Load
Load
0V
1-3 Mounting the Digital Tolerance Indicator
Mounting on DIN Track
1. Let the hook on the digital tolerance indicator’s measuring unit connection side catch the track.
2. Push the unit until the hook clicks into place.
Removing from DIN Track
1. Push the unit in the direction 1.
2. Lift the unit in the direction of arrow 2 while performing step (1).
Measuring Unit Connection Side Hook
2
1
When used in a row
· Up to 30 digital tolerance indicators can be installed in a row.
* Fix the cable in a suitable position to prevent possible cable breakage.
1-4 Mounting the Measuring Unit
1. Open the protection cover.
2. Insert the measuring unit, with the lock lever on its connector area facing upward, all the way into the connector port.
To remove it, press and hold the lock lever then pull
Lock Lever
the measuring unit out.
* Fix the cable in a suitable position to prevent possible cable breakage.
1
2
2
Settings
2-1 Setting and Display Overview
[NO/NC Indicator: Orange] [GO Indicator Orange]
Turns ON when judged as GO. The up/down arrows on the right indicate HIGH/LOW judgment result.
Indicates current NO/NC setting.
[Preset Indicator: Green] Turns ON when preset has been set.
[ST Indicator: Blue] Turns ON when Tolerance setting is in progress.
2-2 Switching Control Output
Press button.
NO/NC
Under NO (Normal Open) setting, the output turns on when a workpiece is within the tolerance (GO).
of [NO/NC Indicator] turns ON.
2-3 Reference Point
Tolerance Setting Button [ST] Button Allows threshold setting based on the workpiece.
Displays the measured value. Indication is in 0.1μm step.
Minute Threshold Adjustment [UP/DOWN] Button Changes a threshold value. During the area setting, H or L is displayed on the left of the threshold value you are setting, allowing identification of HIGH or LOW threshold value that is being set
Mode/OUT Switch [MODE] Button
Switches the threshold value to set by short pressing to HIGH or LOW. Pressing and holding for 3 seconds or longer switches between setting and detection modes.
Output Switch [NO/NC] Butt on
Switches the polarity of the external output.
When reference point use setting is ON (See 3 Convenient Setting Features) The measured value is not displayed until the measuring unit passes the reference point after power ON. When the reference point is used, turn on the power with the spindle extended as far as possible, and then move the spindle 1.5 mm or more.
2-4 Tolerance Judgment
* Also see “5. Detailed Settings” when making the settings.
Setting for sensing within the range of the upper and lower limits (Threshold 2-point)
2-point area setting
1. Select [Area detection Mode] after the process of following [Setting Mode][Judgment Output Mode]
2. Press and hold the [MODE] button for 3 seconds or longer to exit the Setting Mode.
3. Follow the procedure below and make the settings.
Threshold value HIGH : Upper Limit Workpiece Height Threshold value LOW : Lower Limit Workpiece Height
Workpiece setting
Upper Limit
Workpiece
ST
Moving the finger off the button displays "2Pnt".
Measuring a workpiece with ± tolerance (Threshold 2-point)
± tolerance setting
1. Select [Setting Mode] [Tolerance Setting : HIGH] and configur e the tolerance value on the High end.
2. Select [Tolerance Setting : LOW] and configure the tolerance value on the Low end.
3. Select [Area detection Mode] in [Judgment Output Mode]
4. Press and hold the [MODE] button for 3 seconds or longer to exit the Setting Mode.
5. Follow the procedure below and make the settings.
Threshold value HIGH : Preset value + Tolerance setting (HIGH) Threshold value LOW : Preset value - Tolerance setting (LOW)
Standard
Workpiece
ST
Hold for 3 seconds or longer
Measuring for one reference (Threshold 1-point)
2-point setting
1. Select [Normal Detection Mode] after the process of following [Setting Mode][Judgment Output Mode]
2. Press and hold the [MODE] button for 3 seconds or longer to exit the Setting Mode.
3. Follow the procedure below and make the settings.
Workpiece setting
Workpiece
1
Threshold value setting: Set the value in the middle between the measured values for the 1st and 2nd points.
ST
Measuring for standard workpiece as reference (Threshold 1-point)
1-point setting
1. Select [Normal detection Mode] after the process of following [Setting Mode][Judgment Output Mode]
2. Press and hold the [MODE] button for 3 seconds or longer to exit the Setting Mode.
3. Follow the procedure below and make the settings.
Standard
Workpiece
Setting Error
Error / Display / Cause RemedyError Origin Tuning Type
Tolerance Judgment Error
The 1st and the 2nd measuring points are close, or tolerance setting is too small.
Near Error
The difference between the 1st and the 2nd measured values is too small.
Overflow Error
The preset or tolerance setting value is too large.
Underflow Error
The preset or tolerance setting value is too small.
Threshold setting: Set the workpiece value as the threshold.
ST
· Ensure the wider distance between the 1st and the 2nd measuring points.
· Set the larger difference between the tolerance settings of HIGH and LOW.
· For hysteresis setting, congure a smaller setting value.
· Congure the preset value again.
· Congure the tolerance setting again.
· For hysteresis setting, congure a smaller setting value.
· Congure the preset value again.
· Congure the tolerance setting again.
Hold for 3 seconds or longer
· Congure the preset value again.
· Congure the tolerance setting again.
HIGH
GO
Upper Limit
Workpiece
LOW
Workpiece setting
Lower Limit Workpiece
HIGH
GO
LOW
Tolerance setting (HIGH)
Lower Limit Workpiece
ST
Workpiece
Workpiece
Release the button when "toL" flashes
GO
ST
Standard
Workpiece
1
NOGO
Workpiece setting
Workpiece 2
GO
NOGO
Release the button when "1Pnt" flashes
2-5 Minute Adjustment of Threshold Level
Under [Detection Mode], press and hold button.
1.
The threshold value blinks.
2.
<<For Area Detection Mode Setting>>
· HIGH Threshold Value Display
<<For Normal Detection Mode Setting>>
Press button to adjust
3.
UP/DOWN
the threshold level.
Pressing and holding the button allows quick setting. To manually set threshold values, always congure them so that "HIGH threshold value > LOW threshold value". If they are configured as "HIGH threshold value < LOW threshold value";
- GO judgment is not given regardless of a measured value.
- HIGH and LOW indicators turn ON at the same time and error output is provided if the judgment result is other than HIGH/LOW.
MODE
MODE
Under NC (Normal Close) setting, the output turns on when a workpiece is outside the tolerance (NoGO).
of [NO/NC Indicator] turns ON.
Threshold Value HIGH Threshold Value LOW
*Under normal direction
Setting is Completed
Threshold value HIGH
Setting as preset value
Threshold value LOW
Tolerance setting (LOW)
*Under normal direction
Setting is Completed
Threshold value = Middle of the workpieces 1 and 2
Workpiece 2
Setting is Completed
Threshold value
Setting is Completed
MODE
· LOW Threshold Value Display
· Measured Value Display· Threshold Value Display
The threshold level becomes higher.
· Measured Value Display
MODE
1.5 mm or more
The threshold level becomes lower.
3
Convenient Setting Features
Initializing Settings
Setting Reset
Initialize all settings to the factory-set defaults.
[ ] [ ]
Hold both for 3 sec. or longer
UP/DOWN
MODE MODE
Press UP button once
or press DOWN button twice
Saving/Reading Settings
User Save Function/User Reset Function
User Save: The current settings are saved. User Reload: The saved settings are loaded.
Hold both for 3 sec. or longer
User Save Function
[ ] [ ]
UP/DOWN UP/DOWN
MODE MODE
User Reload Function
[ ] [ ]
MODE MODE
UP/DOWN
Using the measuring unit reference point/Setting the point at power ON as origin
Reference Point Use Setting
1. Select [Setting Mode ] [Reference Point Use Setting]. ON : The unit automatically waits for the reference point signal. When the reference point is used, turn on the power with thespindle extended as far as possible, and then move the spindle 1.5 mm or more. A measured value is displayed. OFF : The reference point is set as the position of the measuring unit at power ON, and the measured value is displayed. The displayed value is the preset value. * After the setting, turning the power OFF then ON, or searching the reference point again, reects the reference point use setting to measurement. * When the reference point use setting is ON, a hyphen mark (-) is displayed until the measuring unit passes the reference point.
4
Maintenance
4-1 Troubleshooting
Troubleshooting
Phenomena
Nothing is shown on the indication.
The digital tolerance indicator restarts during operation.
Nothing is shown on the digital indication.
Input signal is not received.
The measured value is not displayed in 0.0001 step
The judgment output is not properly provided
Lost tracking of the settings made.
Error Display
Load short circuit detection error
Overcurrent protection error
Digital tolerance indicator EEPROM error
Measuring unit communications time-out error
Measuring unit memory error
Measuring unit speed error
Measuring unit signal level error
Is the power supply ON? Are the cables not broken?
Is the Eco Function not turned ON ?
Are the external input settings OFF?
Have the display digits configuration properly ?
Have the tolerance setting and hysteresis properly configured?
The judgment output line is short circuited.
A connection error is found in the measuring unit.
An error is found in the digital tolerance indicator setting memory.
A communications error is found between the measuring unit and the digital tolerance indicator.
An error is found in measuring unit memory.
The speed of passing the reference point was too high.
A measuring unit circuit failure
Cause Remedy
Check the wiring and measuring unit, the power supply voltage and capacity.
Refer to "1-2 Input/Output Circuit Diagram".
Turn OFF the Eco function.
Refer to "5 Detailed Settings".
Check the wiring and external input settings. Refer to "1-2 Input/Output Circuit Diagram".
Configure it properly.
Refer to "5 Detailed Settings".
Configure the tolerance setting and hysteresis properly.
Refer to "5 Detailed Settings".
Reset the settings.
Refer to "5 Detailed Settings".
Turn off the power supply, check whether the output line is short circuited or not, and then turn on the power supply again.
Check if the measuring unit is correctly mounted, and turn ON the power supply again.
Turn ON the power again. Reset the settings if the error is not corrected.
Turn OFF the power supply and check if the measuring unit and digital tolerance indicator are correctly connected, and then turn ON the power supply again. If the error persists, the measuring unit or digital tolerance indicator is broken. Replace the measuring unit or digital tolerance indicator.
Turn OFF the power supply and check if the measuring unit is correctly connected, and then turn ON the power supply again.
setting
If the error persists, the measuring unit is broken. Replace the measuring unit.
Check that excessive impact is not applied to the measuring unit. Turn ON the power supply again or perform the reference point research.
Refer to "3 Convenient Setting Features"
Check if the measuring unit is correctly mounted, and then turn ON the power supply again. If the error persists, the measuring unit is broken. Replace the measuring unit.
RemedyCauseError Name / Display
Preventing Malfunction
Key Lock Function
Enable/Cancel (The same procedure)
Reference point search again
(to capture the measuring unit reference point again)
1. Press and hold [MODE] and [NO/NC] buttons for 3 seconds or longer. * Reference Point Use Setting is ON: The reference point is not acquired yet (hyphen). Pass the measuring unit reference point. Reference Point Use Setting is OFF: Set the position at execution to the preset value.
Preset Function
Set any preset value for the criteria position and perform measurement and judgment output. The preset value on factory shipment is 0, which can be used for zero-resetting.
Enable Cancel
Hold both for 3 sec. or longer Hold both for 3 sec. or longer
Enable
1. Select [Setting Mode] [Preset Input Value] and set any value. Press and hold the [MODE] button for 3 seconds or longer to exit the Setting Mode.
2. Under the [Detection Mode], press and hold [ST] and [UP] buttons for 3 seconds or longer.
Cancel
1. Under the [Detection Mode], press and hold [ST] and [DOWN] buttons for 3 seconds or longer. * When the reference point use setting is ON, the reference position information is saved and can be recovered after power OFF.
* A preset value can be congured within a range from -1999.9999 to 9999.9999. (in 0.0001 step with initial value of 0) * To prevent EEPROM to reach its life for writing (100,000 times), it is recommended that writing to EEPROM should be turned OFF by selecting [Setting Mode] [Writing to EEPROM from External Input] if presetting is performed for each measurement by the external input.
Status Display
Lock ON
Measured value upper limit error
Measured value lower limit error
Moving average count unreached
Reference point not acquired
Disables all the button operations.
The key lock function enabled
The measured value is over the display upper limit (9999.9999).
The measured value is under the display lower limit (-1999.9999).
The measured values for the number of moving average count is being acquired from the measuring unit.
The measuring unit did not pass the reference point.
Cancel the key lock function.
Refer to "3 Convenient Setting Features"
Review the preset value.
Review the preset value.
Please wait until the moving average result is calculated
Have the measuring unit pass the reference point (the point the measuring unit is pressed in by 1.5 mm or more from where it is fully extended).
Hold both for 3 sec. or longer
* Press either of UP/DOWN.
Hold both for 3 sec. or longer
RemedyCauseError Name / Display
Function Selection: [ ]
4-2 Ratings and Specifications
MF10-P1
NPN output
Model
Control output External input*4 Minimum display unit Power supply voltage Power consumption*1
Control output*2
Protection circuit
Number of banks Ambient temperature range*3 Ambient humidity range Mass Cable lengs
*1. Power supply voltage 10 V to 30 V:
Normal mode: 2250 mW max.(Power supply voltage 30 V: Power consumption 75 mA max./Power supply voltage 10 V: Power consumption 155 mA max.) Power saving ECO: 2100 mW max. (Power supply voltage 30 V: Power consumption 70 mA max./Power supply voltage 10 V: Power consumption 135 mA max.)
*2. When lining up 4 or more digital tolerance indicators, the 2 output total is 20 mA or less. *3. When used in a row, operating ambient humidity ranges are: 3 to 10: 0 to +50℃,11 to 16: 0 to +45, 17 to 30: 0 to +40 *4. Details on inputs are as follows:
NPN output
PNP output
MF10-P2
PNP output
2 1
0.1 μm 10 to 30 VDC, including ripple (p-p) 10% Power supply voltage 24 V:Normal mode: 2040 mW max.(Power consumption 85 mA max.) Power saving ECO: 1920 mW max.(Power consumption 80 mA max.) Load voltage: 30 VDC max., open collector output type Load voltage: The total of the two outputs must be 100 mA max. Residual voltage and load current less than 10 mA: 1 V max.
(
Load current 10 to 100 mA: 2 V max. Off state current: 0.1 mA max. Power supply reverse polarity protection, output short-circuit protection and output incorrect connection protection 4 Operating: When lining up 1 or 2 digital tolerance indicators: 0°C to 55°C Storage: –10°C to 60°C (with no icing or condensation) Operating and storage: 35% to 85% RH (with no condensation) Approx. 75 g 2 m
Contact input (Relay or switch)
ON: Short circuit to 0 V (Outflow current: 1 mA max.)
OFF: Open or short circuit to Vcc
ON: Short circuit to Vcc (Sink current: 3 mA max.)
OFF: Open or short circuit to 0 V
Non-contact input (Transistor)
ON: 1.5 V max. (Outflow current: 1 mA max.)
OFF: Vcc-1.5 V to Vcc (Leakage current: 0.1 mA max.)
ON: Vcc-1.5 V to Vcc (Sink current: 3 mA max.) OFF: 1.5 V max. (Leakage current: 0.1 mA max.)
)
Input time
ON: 2 ms min. OFF: 20 ms min.
Detailed Settings
5
Hold button for 3 seconds or longer to enter Setting mode.
MODE
Setting mode provides the function settings described hereafter. The initial display shown after transition from one function to another represents the factory default.
A
1. Function Selection Enabling 4 to 15
Basic setting
MODE
2. Detection Function Changing Response Time
STND Standard Mode
MODE
3. Output Mode Selection Specify signal assignment to two output wires.
Normal Output Mode
MODE
A
Function Selection: [ ]
UP/DOWN
Detailed setting
UP/DOWN
GIGA Giga Mode HS High-speed
UP/DOWN
Hybrid Output Mode
Normal Output Mode
Output line
Control Output 1 OFF OFFON Control Output 2
Hybrid Output Mode
Output line
Control Output 1 OFF OFFON Control Output 2
* If the judgment output mode is [Normal Detection Mode ], the output is provided in the [Normal Output] pattern regardless of the setting.
When the value flickers due to vibration or other factors, the value can be stabilized by averaging.
SHS Super High-speed Mode
SHS
HS
STND
GIGA
GO Judgment
HIGH Judgment
ON OFF
Response time
Mode
3 ms
10 ms
100 ms
1000 ms
*For NO (Normal Open)
NoGO
Error Judgment/
Judgment
Undetermined
OFF ONOFF
LOW
GO
Judgment
Judgment
ON OFFON
Average Count
98
998
Error Judgment/ Undetermined
Measurement Cycle 1 8
The external output is reversed when NC (Normal Close). The indicator is not reversed.
1 ms 1 ms 1 ms 1 ms
4. BANK Switching Set values are saved for each configured bank.
*The preset value configured in BANK is reflected when BANK is switched.
BANK1
MODE
UP/DOWN
BANK2 BANK3 BANK4
5. Tolerance setting (HIGH) Configure HIGH tolerance value for tolerance judgment.
The value is used to calculate the HIGH tolerance value for performing tolerance
Tolerance setting (HIGH)
MODE
Enter tolerance HIGH value
MODE
6. Tolerance setting (LOW) Configure LOW tolerance value for tolerance judgment.
Tolerance setting (LOW)
MODE
Enter tolerance LOW value
MODE
7. Reference Point Use Setting
Using measuring unit
MODE
Judgment.
The tolerance value can be set by buttons.
(from -199.9999 to 999.9999, in 0.0001 step, with initial value of 0.1)
The value is used to calculate the LOW threshold value for performing tolerance judgment.
The tolerance value can be set by buttons.
(from -199.9999 to 999.9999, in 0.0001 step, with initial value of -0.1)
Select whether using the measuring unit reference point or setting the point at power ON as origin.
* Pay attention that this setting is not reflected until the power is
UP/DOWN
turned OFF then ON again or the origin is searched again.
Origin at Power On
B
UP/DOWN
UP/DOWN
Refer to "3 Convenient Setting Features"
B
8. Direction Specify a direction to increment/decrement the measured value.
For normal (nor) case: The pressing-in direction is +.
UP/DOWN
Normal
Reverse
MODE
For reversed (rEv): The pressing-in direction is -.
9. Judgment Output Mode Change the judgment output mode.
Area Detection Mode
MODE
10. External Input A type of external input is changed.
Preset
Execute: ON for less than 3 seconds Cancel: ON for 3 seconds or more
MODE
11. Display Digits Set the number of digits to display under the detection mode.
0.1 μm step
MODE
UP/DOWN
Normal Detection Mode
Area Detection Mode: The threshold values are set as HIGH/LOW. Judgment (display/output) is done with three levels; (HIGH) if the value exceeds the HIGH threshold value, (LOW if the value falls under the LOW threshold value, and (GO) if the value is within the tolerable range. Normal Detection Mode: The threshold value is set and the judgment is done for non-defect (GO) or defect (NoGO).
When judgment setting [in tUnE] is selected, short circuit time must be the same as the key input time.
UP/DOWN
BANK Switching BANK 1 if OFF, BANK2 if ON. Switching to BANK3 or 4 by external input is not available.
1μm step 10 μm step
Input OFF
* Judgment is made for measured value including non-display digit s (in 0.1μm step).
UP/DOWN
Tuning
2-point setting 2-point area setting
1-point setting ± tolerance setting
100 μm
step
1st point Less than
3 seconds 3 sec min.
2st point Less than
3 seconds –
12. Preset Input Value
Preset Input Value
MODE
Preset Input Value
MODE
13. Eco Function
Eco function OFF
MODE
14. Hysteresis width
Standard setting
MODE MODE
Standard setting
MODE
15. Writing to EEPROM of External Input
ON
MODE
See Preset Function in "3 Convenient Setting Features"
The preset value can be set by button.
(from -1999.9999 to 9999.9999, in 0.0001 step, with initial value of 0) Pressing and holding the button allows quick setting.
Saving Power Consumption
UP/DOWN
Eco function ON
UP/DOWN
User setting
MODE
UP/DOWN
The indicators (white digital) turn OFF. They turn ON for approx. 10 seconds and then turn OFF by button operation.
Set the hysteresis width . Hysteresis width is provided for threshold to prevent the judgment output from becoming unstable near the boundaries.
If chattering occurs, check the output stability while
*
setting the hysteresis width.
*
Usually, set the value to zero,
The hysteresis width can be set by button.
User setting input
UP/DOWN
OFF
A
Move to Detection Mode by holding the button for 3 seconds or longer.
(from 0.0000 to 9999.9999, in 0.0001 step, with initial value of 0.0001)
The settings that have been changed by an external input with "oFF" will not be overwritten to prevent EEPROM from reaching its lifespan (100,000 writings).
UP/DOWN
Note) After finish the setting, if the measuring unit which measuring length differs is re-connected, the setting value will be initialized.
Shinagawa Intercity Tower A-18F, 2-15-1, Konan, Minato-ku, Tokyo 108-6018, Japan
MF10-P1/MF10-P2 2-A01-184-1A
©2013 Magnescale Co., Ltd.
Printed in Japan
2013.3
Loading...