Magnescale LT10A-105, LT10A-105B, LT10A-105C, LT10A-205B, LT10A-205C Instruction Manual

...
取扱説明書/InstructionManual/Bedienungsanleitung
表示ユニット/ Counter Unit / Anzeigeeinheit
LT10A / LT11A Series
お買い上げいただき、ありがとうございます。
ご使用の前に、この取扱説明書を必ずお読みください。
ご使用に際しては、この取扱説明書どおりお使いください。
お読みになった後は、後日お役に立つこともございますので、必ず保管してください。
Read all the instructions in the manual carefully before use and strictly follow them. Keep the manual for future references.
Lesen Sie die ganze Anleitung vor dem Betrieb aufmerksam durch und folgen Sie beim Betrieb des Geräts den Anweisungen. Bewahren Sie diese Bedienungsanleitung zum späteren Nachlesen griffbereit auf.
[For U.S.A. and Canada]
THIS CLASS A DIGITAL DEVICE COMPLIES WITH PART15 OF THE FCC RULES AND THE CANADIAN ICES-003. OPERATION IS SUBJECT TO THE FOLLOWING TWO CONDITIONS. (1) THIS DEVICE MAY NOT CAUSE HARMFUL
INTERFERENCE, AND
(2) THIS DEVICE MUST ACCEPT ANY
INTERFERENCE RECEIVED, INCLUDING INTERFERENCE THAT MAY CAUSE UNDERSIGNED OPERATION.
CET APPAREIL NUMÉRIQUE DE LA CLASSE A EST CONFORME À LA NORME NMB-003 DU CANADA.
(E) (1)
LT10A / LT11A Series
Safety Precautions
Magnescale Co., Ltd. products are designed in full consideration of safety. However, improper handling during operation or installation is dangerous and may lead to fire, electric shock or other accidents resulting in serious injury or death. In addition, these actions may also worsen machine performance. Therefore, be sure to observe the following safety precautions in order to prevent these types of accidents, and to read these "Safety Precautions" before operating, installing, maintaining, inspecting, repairing or otherwise working on this unit.
Warning Indication Meanings
The following indications are used throughout this manual, and their contents should be understood before reading the text.
Warning
Failure to observe these precautions may lead to fire, electric shock or other accidents resulting in serious injury or death.
Caution
Failure to observe these precautions may lead to electric shock or other accidents resulting in injury or damage to surrounding objects.
Symbols Requiring Attention Symbols Prohibiting Actions
CAUTION ELECTRICAL DO NOT
SHOCK DISASSEMBLE
(2) (E)
LT10A / LT11A Series
Warning
•Do not use this unit with voltages other than the specified supply voltage as this may result in fire or electric shock.
• Do not handle the I/O connector with wet hands as this may result in electric shock.
• Do not open the cover of the counter unit to disassemble or modify the unit, as this may result in burns or injury. These actions may also damage the internal circuitry.
• This device operates with DC power supply. Absolutely do not connect an AC power supply to the I/O connector as this may result in fire or electrical shock. Doing so may also damage the internal circuitry.
Caution
• The unit does not have an explosion-proof structure. Therefore, do not use the unit in an atmosphere charged with inflammable gases as this may result in fire.
•Be sure to turn off the power before connecting or disconnecting power and signal connectors in order to prevent damage or misoperation.
• The unit does not have an earthquake-proof structure. Therefore, do not use the unit in moving areas or areas exposed to strong shocks.
(E) i
LT10A / LT11A Series
Contents
1. Note to users ..............................................................1
1-1. General precautions............................................. 1
1-2. Handling instructions............................................ 1
1-3. Cautions on operation .......................................... 2
1-4. Instructions for connecting to the measuring
unit ....................................................................... 2
2. Summary .................................................................... 3
2-1. Features ............................................................... 3
2-2. System structure .................................................. 4
3. Connecting and installating ...................................... 5
3-1. Connecting the cables ......................................... 5
3-2. Installing the counter unit ..................................... 5
4. Name and function of each part ............................... 6
4-1. Front panel .......................................................... 6
4-2. Rear panel ........................................................... 8
4-3. Function description ............................................. 9
4-3-1. Reset key .................................................. 9
4-3-2. Preset functions ........................................9
4-3-3. Result evaluation .................................... 10
4-3-4. Peak-hold function ..................................10
4-3-5. Peak-hold pause function ....................... 11
4-3-6. Latch functions ....................................... 12
5. Operation ..................................................................13
5-1. Initial settings ..................................................... 13
5-1-1. Basic settings ......................................... 13
5-1-2. BCD model (only LT10A-105B/205B,
LT11A-101B/201B) ................................. 15
5-1-3. RS-232C model (only LT10A-105C/205C,
LT11A-101C/201C) ................................ 16
5-2. Various settings ................................................. 19
5-2-1. Setting the preset value .......................... 19
5-2-2. Setting the measuring mode ................... 20
5-2-3. Setting the comparator values ................21
5-2-4. Setting the key lock ................................ 22
6. I/O connector ............................................................ 23
6-1. Connector pin assignment ................................. 23
6-2. I/O circuitry (I/O connector area)........................ 25
6-3. Signal timing ...................................................... 26
7. BCD output (only BCD model) ................................ 27
7-1. Connector pin assignment ................................. 27
7-2. Signal timing ...................................................... 29
7-3. Interface cable ................................................... 30
7-4. BCD input/output circuits ................................... 31
ii (E)
LT10A / LT11A Series
8. RS-232C Interface (only RS-232C model) .............. 32
8-1. Terminal pin assignment .................................... 32
8-2. Connecting a personal computer ....................... 33
8-3. RS-232C Interface ............................................. 34
8-4. RS-TRG circuit ................................................... 34
8-5. Output ................................................................ 35
8-6. Commands......................................................... 36
9. Alarm display/output ............................................... 40
10. Specifications .......................................................... 41
10-1. LT10A/LT11A (Specifications) ......................... 41
10-2. Accessories ...................................................... 43
10-3. Options .............................................................43
10-4. Dimensions ...................................................... 44
11. Troubleshooting ...................................................... 45
(E) 1
LT10A / LT11A Series
1-1. General precautions
When using Magnescale Co., Ltd. products, observe the following general precautions along with those given specifically in this manual to ensure proper use of the products.
Before and during operations, be sure to check that our products function properly.
Provide adequate safety measures to prevent damage in case our products should develop a malfunction.
Use outside indicated specifications or purposes and modification of our products will void any warranty of the functions and performance as specified for our products.
When using our products in combination with other equipment, the functions and performance as noted in this manual may not be attained, depending upon the operating environmental conditions. Make a thorough study of the compatibility in advance.
Design and specifications are subject to change without notice.
1. Note to users
1-2. Handling instructions
Do not open the cover of this device or put your hand inside. Otherwise the internal circuit may be broken by static electricity.
To prevent malfunctions caused by static electricity, always turn off the power when touching other parts than the key switches.
Do not route the connecting cable through the same duct as the machine power line.
To prevent misoperation of surrounding equipment due to noise emitted from this counter unit and vice versa, connect the power cable in a twisted manner when supplying DC power.
When providing DC power, be sure to use within the specified voltage range.
Absolutely do not connect an AC power supply to the I/O connector. Doing so may damage the internal circuits.
Do not use connection prohibited I/O connector pins as relay pins.
When connecting the BCD connector, be sure to wire the connector correctly. Failure to do so may damage the internal circuits.
2 (E)
LT10A / LT11A Series
1-4. Instructions for connecting to
the measuring unit
The LT10A/LT11A is a counter unit designed for the DT series measuring unit.
When the LT10A/LT11A is connected to a DT series and turned on, the digits corresponding to the channel connected may flash. This shows that initialization is required to calibrate the new pairing of both device and unit. Here you should move the measuring unit at least 1 mm, as shown in the figure, and then press the reset key on the LT10A/LT11A corresponding to the channel. The device will now return to its normal measuring state.
Note
Once this procedure has been carried out the digits will not flash while the device is connected to that measuring unit, even if the power is turned off.
Place the counter unit more than 0.5 m (20") away from a high voltage source, large current source, large power relay, etc.
For installation of the counter unit, avoid a location exposed to chips, cutting oil, or machine oil. If unavoidable, take adequate countermeasures.
Do not put a vinyl cover directly over the counter unit or put it in a closed container.
The ambient temperature should be in the range of 0 °C to 40 °C (32 °F to 104 ° F). Avoid exposure to direct sunlight, hot air currents, or heated air.
1-3. Cautions on operation
Carry out the key operations or I/O (BCD, etc.) connections and operations in line with the explanations given in the appropriate sections. Failure to correctly operate this device may result in a malfunction.
RESET
P
1 mm or more
(E) 3
LT10A / LT11A Series
2. Summary
2-1. Features
Compact size suited to inclusion in systems. DIN size (72 × 72 mm). Can be panel-mounted.
Devices such as PLCs are connectable from the counter unit. Go/No Go output is standard to all models. Models capable of BCD and/or RS-232C also outputs are available.
Resolution
LT10A series : 0.005 mm (0.0002") LT11A series : 0.001 mm (0.0001")
As well as the current value, maximum and minimum values and peak-to-peak values can also be measured.
Add/Sub calculation is standard feature (only for 2 channel models). Can measure the widths or steps.
Can carry out Go/No Go test on different lots. (BCD output models.) Four different upper and lower limits can be stored in memory for the Go/No Go comparison.
Power is compatible to DC 9 to 24V. Provided via the power input connector. Use a power cable less than 10 meters long.
The counter unit LT10A/LT11A series is designed to be incorporated into assembly lines or jigs, and to be used for measuring components or Go/No Go. Use in combination with a DT series measuring unit. Types are available to suit various uses.
Number of
input channel
1
2
1
2
Model
LT10A-105 LT10A-105B LT10A-105C LT10A-205 LT10A-205B LT10A-205C LT11A-101 LT11A-101B LT11A-101C LT11A-201 LT11A-201B LT11A-201C
Output
Go/No Go
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
RS-232C
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
BCD
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
◯◯
Resolution
(mm)
0.005
0.001
4 (E)
LT10A / LT11A Series
RESET
SET
P
MODE COMP
A
B
RESET
A
RESET
B
SET
P
MODE COMP
RESET
SET
P
MODE COMP
A
B
RESET
A
RESET
B
SET
P
MODE COMP
RESET
SET
P
MODE COMP
A
B
RESET
A
RESET
B
SET
P
MODE COMP
Go/No GoGo/No GoGo/No GoGo/No GoGo/No Go Go/No Go
PLC etc.
Computer
Counter unit
Measuring unit
LT10A-105 LT11A-101
1 CH
LT10A-205 LT11A-201
2 CH (including Add/Sub)
DT12N/512N DT32N
DT12P/512P (IP64 Compatible) DT32P (IP64 Compatible)
RS-232C RS-232CBCDBCD
LT10A-105B LT11A-101B
1 CH
LT10A-205B LT11A-201B
2 CH (including Add/Sub)
LT10A-105C LT11A-101C
1 CH
LT10A-205C LT11A-201C
2 CH (including Add/Sub)
Power source 9 to 24 V (Power input connector.)
2-2. System structure
(E) 5
LT10A / LT11A Series
3-1. Connecting the cables
Secure all connecting cables so as to prevent accidental disconnection.
Make certain the counter unit’s power is off before connecting or disconnecting the measuring unit.
3-2. Installing the counter unit
When mounting in a panel
1. Cut out an opening to match the dimensions shown
(Fig. 2).
2. Insert the counter unit into the cut-out opening in the
panel from the front.
3. Attach the supplied counter stopper from the rear.
4. Press in the counter stopper until it touches the panel.
Note
When attaching the counter stopper to the counter unit, leave enough space (min. 30 mm/1.18") between the top and bottom. (Fig. 3)
3. Connecting and installating
Fig. 2
Fig. 3
Panel thickness: max. 12/0.47"
Cut-out dimensions
Panel box
Front of counter unit
68  /2.68"
+1.0 +0.4
+0.039" +0.016"
68  /2.68"
+1.0
+0.4
+0.039"
+0.016"
4-R1 or less
Fig. 1
Counter stopper (supplied)
Front of counter unit
Counter stopper (supplied)
30/1.18"
30
6 (E)
LT10A / LT11A Series
4-1. Front panel
4. Name and function of each part
q
: Go/No Go test result indicator
Gives the result of comparing the displayed value to the comparator’s upper and lower limits. Over upper limit, \ between upper and lower limit, under lower limit.
w
RESET
RESET
A
RESET
B
: Reset key
Resets the displayed value to zero.
When a value has been preset it returns to this.
e
P
: Preset key
Enters the preset mode. (For the current value, maximum value, and minimum value.)
r
MODE
: Measuring mode setting key
Key to enter the mode for selecting one of maximum, minimum, peak-to-peak (maximum–minimum), or current values.
t
COMP
: Comparator value setting key
Key to enter the mode for setting the comparator upper or lower limit.
y
SET
: Setting key
Set a mode or a value.
u
: Number selection key
Selects the number for the digit chosen.
i
: Digit selection key
Selects the digit to change when setting numeric values.
Normally, when it is held down for 5 seconds, key lock is established; alternatively, if key lock is already established, it is released.
Main display
1 CH input model : LT10A-105/105B/105C
LT11A-101/101B/101C
2 CH input model : LT10A-205/205B/205C
LT11A-201/201B/201C
mm
M A X M I N P — P C P H1 2 3 4
P
RESET
SET
P
MODE COMP
q
r u
i
tye
w
mm
M A X M I N P — P C P H1 2 3 4
P
RESET
SET
P
MODE COMP
mm
M A X M I N P — P C P L1 2 3 4
P
A+B
BA
A
RESET
B
(E) 7
LT10A / LT11A Series
Note
If upper display A+B and lower display A have been selected, the comparator setting values of channel A will be applied on both the upper and lower limits.
When the upper display A+B is selected, the lower display only shows the current value of the selected channel and no operation can be made on the lower display.
Preset status indicator
When
is displayed the preset value is set.
Peak-hold indicator
When MAX/MIN/P-P is displayed the data shown is the maximum/minimum/maximum–minimum value. When neither of the them is shown, the current value is displayed.
Comparator upper limit setting indicator
Displayed when the comparator value has been set. The upper digits are the upper limit.
Comparator lower limit setting indicator
Displayed when the comparator value has been set. The lower digits are the lower limit.
Up to four different comparator upper and lower limit settings (CPH1 to CPH4 and CPL1 to CPL4) can be stored in the memory for LT10A-105B/205B and LT11A­101B/201B.
LT10A-105/205/105C/205C and LT11A-101/201/101C/ 201C have only one setting each.
Key lock indicator
Lighted when key lock is established; off when key lock is released.
Comparator upper limit setting indicator
Comparator lower limit setting indicator
Peak-hold indicator
Preset status indicator
Main display
Displays the measured data, setting data for various modes, or alarms, etc.
Selected channel indicator (2 channel model)
Choose one of these four.
A:data from measuring unit, input channel A
B:data from measuring unit,
input channel B
A+B: sum of data from channel A and B
In order to carry out calculations such as A–B or –A+B,
change the direction of A or B to “+” or “–”. (Initial settings)
Upper AA+BA+BA+B Lower BA B
Selected channel indicator
Key lock indicator
mm
M A X M I N P — P C P H1 2 3 4
P
mm
M A X M I N P — P C P L1 2 3 4
P
A
B
8 (E)
LT10A / LT11A Series
q Power input connector
Connected to the external power supply (+9 V to +24 V DC). Use a power cable less than 10 meters long. Connector used : MC1.5/3-ST-3.5 (provided) made
by Phoenix Contact
w I/O connectorsSee “6. I/O connector”.
There are three kinds of I/O connectors: one type is used for channel A, one for channel B and one for both channels A and B. Input : Reset, Peak-hold start, Peak-hold pause, RS
trigger.
Output: Go/No Go output.
4-2. Rear panel
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
7 6 5 4 3 2 1 7 6 5 4 3 2 1
BCD OUT
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
RC-232C
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
7 6 5 4 3 2 1 7 6 5 4 3 2 1
BCD OUT. B
BCD OUT. A
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
RC-232C
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
AB
SIG. IN A
SIG. IN B
FG +V 0V
DC IN
1 channel-models 2 channel-models
LT10A-105, LT11A-101 LT10A-205, LT11A-201
LT10A-205B, LT11A-201BLT10A-105B, LT11A-101B
LT10A-105C, LT11A-101C LT10A-205C, LT11A-201C
Common I/O connector
q
e
t
u
r
w
y
w
FG +V 0V
DC IN
No. 1 2 3
No.
1
2
3
Signal name
FG
+ V
0 V
Signal
Frame ground
Main DC power (+9 V to +24 V) input
GND for power
(E) 9
LT10A / LT11A Series
e Measuring unit input :SIG. IN A
r Measuring unit input : SIG. IN B (2 channel models)
ty BCD Output
With the 2 channel models the upper and lower selections of the front panel’s main display correspond to the BCD OUT. A/BCD OUT. B. So if “A+B” is selected the output is to BCD OUT. A. The following operations are possible when BCD input/output terminal is used. Input : Comparator value selection (4 settings),
measuring mode (current value, maximum value, minimum value, peak-to-peak values) selection
Output: 5 digits
Outputs one of the current, maximum, minimum, and peak-to-peak values selected via the keys on the front panel and the external input. Alarm output
u RS-232C interface
(See “8. RS-232C Interface”.) Reset, peak-hold start, setting/recall of preset values, setting the comparator value, selecting and outputting the current value/maximum value/minimum value/ peak-to-peak value.
4-3. Function description
4-3-1. Reset key
Counter unit
Measuring mode (Current value, maximum value, minimum value, peak-to-peak value)
Preset mode (
and the selected digit
will flash.)
“Error” is displayed
All the numbers for the channel with an error flash.
Operation performed when reset key is pressed
Display is set to zero. When a preset value has been set, it is recalled.
Preset value is set to zero.
“Error” is canceled and then returnes to the measuring state.
Automatically carries out initialization to calibrate the device to a new measuring unit. (Before resetting the measuring unit must be moved at least 1 mm.)
4-3-2. Preset functions
It is possible to set preset values for each of the current, maximum, and minimum value measuring modes.
For instructions on setting the preset value, please refer to P18, “5-2-1. Setting the preset value.”
10 (E)
LT10A / LT11A Series
4-3-4. Peak-hold function
Stores the maximum, minimum, and peak-to-peak (maximum–minimum) values of the measured values.
The above mentioned measuring mode is set using the keys on the front panel.
The device starts storing values either when a start signal is supplied to the start/latch pins (channel A: pin w, channel B: pin r) of the lower 7-pin I/O connector (common) (See “6. I/O connector”.) or when the reset key has been pressed.
4-3-3. Result evaluation
Go/No Go test is carried out by comparing data from the current measuring mode (current value, maximum value, minimum value, or peak-to-peak value) to the comparator upper and lower limits.
This result is displayed on the front panel and output from the I/O connector (See “6. I/O connector”.).
Result
High
Go
Low
Display
>
\
?
=
Condition
Data > upper limit
Upper limit > data > lower limit
Lower limit > data
=
Operation
I/O connector (common) A CH: starts on “L” (ON) signal from pin w; I/O connector (common) B CH: starts on “L” (ON) signal from pin r.
Reset key is pressed
Result
Starts storing from the current value.
Starts storing from zero. When a preset value is set the device starts storing from the preset value.
(E) 11
LT10A / LT11A Series
4-3-5. Peak-hold pause function
Pause the storing of the maximum, minimum and peak­to-peak (maximum-minimum) values of the measured values.
When the peak-hold function is to be paused, set the pause pin of the lower 7-pin I/O connector (common) to ON; to resume storing, set it to OFF.
Maximum value
Current value
Pause ON Pause OFF
Peak value data update stop (pause)
When pause is ON
The peak value updating is stopped. The current value is constantly updated. If the maximum value mode, minimum value mode or peak­to-peak value mode is set as the measuring mode, the Go/ No GO judgment output and the output data are not updated even when the measuring unit is operated.
When pause is OFF
The peak values are constantly updated.
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
DC IN
PAUSE (B) START (B) PAUSE (A)
START (A)
Common I/O connector
w
e
r
t
12 (E)
LT10A / LT11A Series
4-3-6. Latch functions
In the current value mode, this function holds output data and Go/No GO output for that value.
[Latch conditions]
The start input signal is set as the latch input by the parameter setting.
Current value mode
Note
This function does not operate when the measuring mode is peak value mode.
Current value
Latch OFF
Latch ON
LT10A / LT11A Series
(E) 13
5. Operation
This sections uses the 2 channel model in its explanations. The 1 channel model is the same as the 2 channel model without the B channel. BCD and RS-232C models are noted in the text.
5-1. Initial settings
An initialization is carried out at the time of shipping; however it is possible to make the following selections depending on intended use. Details of the settings at the time of shipping are given in each section.
Changing between inches/mm
Turn on power while holding down
RESET
A
and press
MODE
key.
Press
to change between inches/mm.
Press
SET
to set and return to the measuring state.
Device is set to mm at the time of shipping.
To change the initial settings hold down the
SET
key and
press the
MODE
key for approximately 2 seconds.
Basic operation
MODE
: to the next item. : Select item.
SET
: Set item.
Note
Even if you select an item with the
key, no changes
will be mode until you press the
SET
key.
Once the initial setting mode has been entered it is not
possible to return to the measuring state partway through. Press the
MODE
key repeatedly to skip the items.
5-1-1. Basic settings
1. Setting the display (2 channel models)
One of the following may be chosen:
When the upper display A+B is selected, the lower display only shows the current value of the selected channel and no operation can be made on the lower display.
2. Setting the display resolution or direction (channel A)
0.001/0.005/0.01/–0.001/–0.005/ –0.01 mm
With the measuring unit’s
spindle pushed in:
+: positive direction –: negative direction
When set to inches, values are chosen from 0.0002,
0.0005, –0.0002, and –0.0005.
Note
With the LT10A series, 0.001 mm and –0.001 mm (0.0001" and –0.0001") are not avaible.
Upper AA+BA+BA+B Lower BA B
factory-set
factory-set (LT11A)
A
A
B
LT10A / LT11A Series
14 (E)
4. Selecting the start input terminal (I/O connector) function
(See “6. I/O connector”.)
:
Start function Setting this terminal to “L” (ON) sets the peak-hold value to the current value and restarts the storing procedure.
: Latch function
When using the current value measuring mode, setting this terminal to “L” (ON) stores the output and display of the Go/No Go comparison at that point in time.
Note
When the latch is ON, display and Go/No Go output storage by the DRQ input for the BCD model and RS-TRG input for the RS-232C model is invalidated.
Initial settings are now complete for the standard model.
Pressing
MODE
.. Standard model Returns to the measuring
state.
BCD model Go to section 5-1-2. RS-232C model Go to section 5-1-3.
3. Setting the display resolution or direction (channel B, 2 channel models)
0.001/0.005/0.01/–0.001/–0.005/–0.01 mm.
With the measuring unit’s spindle pushed in:
+: positive direction –: negative direction
When displaying A+B:
If you set the direction of A to “–” the data displayed is the calculation “–A+B”. The same can be done with B.
Note
With the LT10A series, 0.001 mm and –0.001 mm
(0.0001" and –0.0001") are not avaible.
When the addition A+B is chosen the direction for B can
be selected, but its resolution will be the same as that of A.
factory-set
LT10A / LT11A Series
(E) 15
5-1-2. BCD model
(only LT10A-105B/205B, LT11A-101B/ 201B)
Proceeds to the next setting mode from “5-1-1. Basic settings” step 4.
1. BCD logic
Setting the BCD output logic. “+” is true logic. “–” is false logic.
Exception : Logic for the DRQ,
READY, and alarm terminals cannot be changed. (See “7. BCD output”)
2. BCD output format
Setting the BCD output format.
:BCD is output
according to the DRQ signal input, and the resulting status is held even if the DRQ signal goes off.
:BCD is output
according to DRQ signal input, and assumes high­impedance status when there is no DRQ signal input.
: BCD is output at all times at the set time interval
without the input of the DRQ signal.
factory-set
factory-set
factory-set
3. Automatic BCD output time interval
This mode is established when
has been set in step 2. Select one of the eight time intervals listed below. 1/2/4/8/16/32/64/128 ms (See “7-2. Signal timing” for the BCD input/output timing)
Initial settings are now complete for the BCD model.
Press
MODE
to return to the measuring state.
LT10A / LT11A Series
16 (E)
: Outputting with measuring mode information
and comparator Go/No Go result
1st byte : Channel name (A or B) 2nd byte : Current mode
(N: Current value,
P: Peak-to-peak value,
I : Minimum value,
A: Maximum value) 3rd byte : Unit (M: mm, I: inch) 4th byte : Comparator Go/No Go
result U : Upper limit over G : Within range L : Lower limit under E : When an alarm has
occurred
5th byte : Sign (“+”
or “–”)
6th to 11th bytes
: Numerical data
(ex.00.000)
: Outputting according to the mode 1 format
(statistical calculations) of the digital printer P40 (End of sales).
Whether, with the 2 channel model, to output B channel data following a space or to divide it with the delimiter is selected by step 9. (except for P40 mode)
Note
When set to mode even the 2 channel model only outputs the A channel.
: (“+” or space)
factory-set
5-1-3. RS-232C model
(only LT10A-105C/205C, LT11A-101C/ 201C)
Proceeds to the next setting mode from “5-1-1. Basic settings” step 2.
1. Setting the output data format
: Normal output
1st byte : Channel name (A or B) 2nd byte : Sign (“+”
or “–”)
3rd to 8th bytes: Numerical data
(ex.12.345)
: Outputting with measurement mode
information 1st byte : Channel name (A or B) 2nd byte : Current mode
(N: Current value,
P: Peak-to-peak value,
I : Minimum value,
A: Maximum value) 3rd byte : Unit (M: mm, I: inch) 4th byte : Sign (“+”
or “–”)
5th to 10th bytes
: Numerical data
(ex.00.000)
LT10A / LT11A Series
(E) 17
factory-set
factory-set
factory-set
2. Setting the data signalling rate
is displayed and the data
signalling rate can be selected.
600/1200/2400/4800/9600/ 19200/38400 bps
3. Setting the data length
is displayed and the
data can be set to 7 or 8 bits.
4. Setting the stop-bit
is displayed and the stop
bit can be set to one or 2 bits.
5. Setting the parity
is displayed and the
parity can be switched on or off.
: No parity : Parity
Note
When the data length is set to 7 bits in step 3, select “Parity”.
6. Selecting odd or even parity
Switches to this mode when
is chosen above.
: Odd parity : Even parity
7. Selecting the function for the RS-TRG terminal
:Mode for inputting
mechanical contact outputs (such as relay and switch outputs).
:Mode for inputting
electronic circuit outputs (such as transistor outputs).
:To output at a set
interval.
Note
The RS-TRG input cannot be used when
has been selected.
factory-set
factory-set
factory-set
LT10A / LT11A Series
18 (E)
8. Selecting the output time interval.
This mode is selected when
has been chosen in step 7. One of the following eight intervals can be chosen.
0.2/0.5/1.0/5.0/10/30/60/300 s
9. Selecting the data transmission format (2 channel models)
is displayed and one of
the following is chosen.
: format (a) given below : format (b) given below
To output from channel A –12.345, and from channel
B 67.891: (a) A–12.345MB+67.891
C
R
L
F
(b) A–12.345
CR L
F B+67.891 CR LF
Note
M means a space.
factory-set
factory-set
10. Plus sign setting of the output data
When the output data is positive, output characters will be set as a plus sign. This setting will also be reflected in the input data (preset value, comparator value).
: Outputs plus as a sign. : Outputs space as a sign. (compatible to
previous model)
Initial settings are now complete for the RS-232C model.
Press
MODE
to return to the measuring state.
factory-set
LT10A / LT11A Series
(E) 19
5-2. Various settings
In the setting modes there is always an indicator flashing in the main display.
5-2-1. Setting the preset value
1 Push
P
to select.
The A channel preset value setting mode.
SET
P
MODE COMP
1, 4, 7 3, 6 2, 5
RESET
B
RESET
A
P
A
2 Push to select a sign.
Push
to select a digit to be set. The selected digit flashes. Push
to select a numeral.
3 Push
SET
to set.
for A channel flashes.
4 Push
P
to select B channel preset value setting mode. B channels sign flashes “+”. Single channel models return to the measuring state.
5 Same as 2.
6 Push
SET
to set.
Both A/B channel
flashes.
7 Push
P
to returns to the measuring state.
Both A/B channel
lights.
Note
If
SET
is not pressed the previous setting is kept.
If the maximum value mode, minimum value mode or
peak-to-peak value mode is set as the measuring mode, the peak value of each mode has been set is set to the preset value when the preset value is set to the value which has more than the peak value.
LT10A / LT11A Series
20 (E)
3 Push
SET
to set. Set mode flashes. The other channel A modes disappear.
4 Push
MODE
to select the setting mode for channel B.
Currently set channel B mode flashes.
5 Same as 2.
6 Push
SET
to set. Modes set for channels A and B flash.
7 Push
MODE
to return to the measuring state.
Modes set for channels A and B light.
Note
If you do not press
SET
the previous settings will be kept.
M A X AM A X
P ―P
A
B
5-2-2. Setting the measuring mode
1 Push
MODE
to select the mode for setting a measuring mode (current value, maximum value, minimum value, or peak-to peak value) for channel A. Each of there modes are indicated “A”, “MAX”, “MIN”
and “P-P”, respectively.
2 Push
to select a measuring mode.
The selected mode’s indicator flashes.
3, 6 2, 5
1, 4, 7
SET
P
MODE COMP
RESET
B
RESET
A
M A X M I N P ―P
A
LT10A / LT11A Series
(E) 21
5-2-3. Setting the comparator values
1 Push
COMP
to select the mode for setting the channel A comparator values. Start from the CPH (comparator upper limit) setting.
2 Push
to select a sign.
Push
to select digits. A selected digit flashes. Push
to select a number.
RESET
SET
P
MODE COMP
A
RESET
B
3, 6, 9, 12 2, 5, 8, 11
1, 4, 7, 10, 13
C P H
C P L
A
3 Push
SET
to set.
“CPH” flashes.
4 Push
COMP
to select the CPL comparator lower limit
setting mode. Sign of the CPL value flashes.
5 Same as 2.
6 Push
SET
to set.
“CPH” and “CPL” flash.
7 Push
COMP
to select the mode for setting the channel B
comparator values. Single channel models return to the measuring state.
8 9 10 Same as 2, 3, 4, 5, 6. 11 12
13 Push
COMP
to return to the measuring state.
        
LT10A / LT11A Series
22 (E)
5-2-4. Setting the key lock
Set the key lock to invalidate the key operations. This stores the displayed values or set values should a key is hit by accident. The key lock can be set during normal operations.
Setting
1. Press and hold
(for about 5 seconds). Keep pressing until the key lock indicator in the main display changes from flashing to on.
Note
The setting is not performed unless the key is pressed while the indicator is flashing.
Release
1. Press and hold
(for about 5 seconds). Keep pressing until the key lock indicator in the main display changes from flashing to off.
Note
The release is not performed unless the key is pressed while the indicator is flashing.
Key lock indicator
mm
M A X M I N P — P C P L1 2 3 4
P
B
Note 1
With the BCD models four different settings can be stored in memory. The operations are as follows.
A CH CPH1 CPH2 CPH3 CPH4
CPL1 CPL2 CPL3 CPL4
↓↓↓↓
B CH CPH1 CPH2 CPH3 CPH4
CPL1 CPL2 CPL3 CPL4
When the 4 settings are not required press
COMP
repeatedly.
Changing the 4 different setting values is carried out via
the BCD connector input terminal.
When the BCD connector is not connected the setting is
CPH1 and CPL1.
Note 2
If
SET
is not pressed the previous setting is maintained.
Note 3
When the
SET
key is pressed to set the CPL the sign (“+” or “–”) of the CPH value may flash. This is because the CPH (upper limit) is less than the CPL (lower limit), and the device is in the CPH setting mode. In this case, return to and start from the CPH setting.
LT10A / LT11A Series
(E) 23
6. I/O connector
The I/O connector on the rear panel of the counter unit has functions for Go/No Go output based on the comparator function, start input, pause input, RS-232C trigger input and reset input.
6-1. Connector pin assignment
Rear of counter unit
Use a shielded cable for connection to the FG pin on the
rear of the counter unit. (Prepare a shield cable by yourself.)
Connector used : MC1.5/7-ST-3.5 (provided) made
by Phoenix Contact
Outer cover
Knitted shield
Cross section of the cable
Signal
(See “4-3. Function description”.) I/O connector A
Pin Signal
IN/OUT Signal
No. name
1 GND
2NC – Connection prohibited
3 RESET (A) IN Reset input (A CH)
4 LO (A) OUT Go/No Go output Low (A CH)
5 GO (A) OUT Go/No Go output Go (A CH)
6 HI (A) OUT Go/No Go output High (A CH)
7 GND
I/O connector B (not provided for 1-channel models)
Pin Signal
IN/OUT Signal
No. name
1 GND
2NC – Connection prohibited
3 RESET (B) IN Reset input (B CH)
4 LO (B) OUT Go/No Go output Low (B CH)
5 GO (B) OUT Go/No Go output Go (B CH)
6 HI (B) OUT Go/No Go output High (B CH)
7 GND
7 6 5 4 3 2 1 7 6 5 4 3 2 1
1 2 3 4 5 6 7
AB
FG +V 0V
DC IN
I/O connector BI/O connector A
Power input connector
Common I/O connector
LT10A / LT11A Series
24 (E)
Start/latch input
The “L” (ON) signal sets the maximum, minimum, and
peak-to-peak values to the current value and restarts their storing. (Start function)
If, when the initial setting of
which was set at the time
of shipping is changed to
, the current value mode serves as the measuring mode, the “L” (ON) signal will hold the Go/No Go output (I/O connector) and display. (Latch function)
Note
While the Go/No Go output is at the “L” level, reset/preset value recall cannot be effected by reset key or external reset/preset value recall input.
Reset input
“L” (ON) sets the measured value to zero. When there is a preset value this is recalled.
Note
Even when the “L” level is held, the Go/No Go output (I/O connector) and the display are not held.
I/O connector (common)
Pin Signal
IN/OUT Signal
No. name
1 GND
2 START (A) IN Start/latch input (A)
3 PAUSE (A) IN Pause input (A)
4 START (B) IN Start/latch input (B)
∗1
5 PAUSE (B) IN Pause input (B)
∗1
6 RS-TRG IN
RS-232C data output/trigger input
2
7 GND
1 : The connection of this pin is prohibited for 1-channel
models.
2 : The connection of this pin is prohibited except in RS-
232C models.
Go/No Go output
High: displayed value > upper limit → “L” (ON) Go : upper limit
displayed value lower limit → “L” (ON)
Low : lower limit > displayed value → “L” (ON)
Note
All Go/No Go outputs are “H” (OFF) when an alarm is set.
Loading...
+ 78 hidden pages