Magikitchn MKG60 Service Manual

Solid State Control Calibration Procedure
MKG Solid State Griddles - All Options
The following procedure describes the steps to calibrate the Solid State
Temperature Control for any of the MKG-ST Griddle Models.
L80-131 Rev 01
type.
Figure 2: To validate that the unit is operational, turn on one set of burners by pushing the button (circled
above). Pilot should light and indicator light should illuminate (indicated by the arrow).
Page | 2 L80-131 Rev 01
Figure 3: To gain proper access to the Thermostat wires, some disassembly of the unit is required. Start by
removing the grease pan, indicated above.
Figure 4: Next remove the four (qty. 4) Philips head screws, which will allow you to remove the front Left and
front Right panels.
Page | 3 L80-131 Rev 01
Figure 5: Remove front Left and Front Right panels.
Figure 6: Remove four (qty. 4) Phillips head screws from the front the Solid State Control Panel, indicated with
arrows, as well as four (qty. 4) 3/8" hex head self-tapping screws from front panel (indicated with circles).
Page | 4 L80-131 Rev 01
Loading...
+ 8 hidden pages