Madur GA-21 Plus Operating Manual

FLUE GAS ANALYSER GA-21 plus
Operating manual
Version: 2.1 06/2017
www.madur.com Manual GA-21Plus
1. Introduction................................................................................................................................................4
1.1. Use of this operating manual..............................................................................................................4
2. Maintenance...............................................................................................................................................5
2.1. Gas sensors........................................................................................................................................5
2.2. Gas system.........................................................................................................................................5
2.3. Battery.................................................................................................................................................5
2.4. Service intervals..................................................................................................................................6
2.5. Errors during operation.......................................................................................................................6
2.6. Switching off after use.........................................................................................................................6
3. General information..................................................................................................................................7
4. Construction..............................................................................................................................................9
4.1. Elements of gas circuit........................................................................................................................9
4.1.1. Gas probe with in-line filter.........................................................................................................9
4.1.2. Gas pump..................................................................................................................................10
4.1.3. Gas chamber.............................................................................................................................10
4.2. Measurement system........................................................................................................................10
4.2.1. Gas sensors..............................................................................................................................10
4.2.2. Ambient temperature sensor.....................................................................................................12
4.2.3. Flue gas temperature sensor....................................................................................................12
4.2.4. Gas chamber temperature sensor............................................................................................12
4.2.5. Differential pressure sensor......................................................................................................12
4.2.6. Analogue inputs.........................................................................................................................12
4.3. Data input/output system..................................................................................................................14
4.3.1. Keyboard...................................................................................................................................14
4.3.2. Display.......................................................................................................................................16
4.3.3. Printer........................................................................................................................................16
4.3.4. Interface RS232C......................................................................................................................16
4.3.5. Power supply.............................................................................................................................17
5. Operation..................................................................................................................................................18
5.1. Use of the keyboard..........................................................................................................................18
5.1.1. Description of the keys..............................................................................................................18
5.1.2. Selecting menu options.............................................................................................................19
5.1.3. Entering numbers......................................................................................................................20
5.1.4. Editing text................................................................................................................................20
5.2. Basic operating instructions..............................................................................................................21
5.2.1. Setting-up the analyser.............................................................................................................21
5.2.2. Switching on..............................................................................................................................21
5.2.3. Switching off / standby..............................................................................................................21
5.2.4. Initial calibration........................................................................................................................23
5.2.5. Measurement parameters.........................................................................................................24
5.2.5.1. Fuel selection......................................................................................................................25
5.2.5.2. Averaging time....................................................................................................................27
5.2.5.3. Reference oxygen parameters (O2R)................................................................................27
5.2.5.4. Boiler parameters...............................................................................................................28
5.3. Displaying the measurement results.................................................................................................29
5.3.1. Current values...........................................................................................................................29
5.3.2. Averaged values........................................................................................................................30
5.3.3. The averaging process..............................................................................................................31
5.4. Pressure/Draft measurements..........................................................................................................32
5.5. Soot (smoke) Measurement..............................................................................................................33
5.6. Options..............................................................................................................................................34
5.6.1. Device info................................................................................................................................35
5.6.2. Sensors.....................................................................................................................................36
5.6.2.1. Sensor CO (only in dual CO sensor configuration)............................................................36
5.6.2.2. CO range............................................................................................................................37
2
www.madur.com Manual GA-21Plus
5.6.2.3. O2 cell calibration...............................................................................................................37
5.6.2.4. External channels...............................................................................................................38
5.6.2.5. Sensor calibration...............................................................................................................40
5.6.2.6. Calibration adjustment........................................................................................................42
5.6.3. Settings.....................................................................................................................................43
5.6.3.1. Backlighting.........................................................................................................................43
5.6.3.2. Contrast..............................................................................................................................43
5.6.3.3. Clock...................................................................................................................................45
5.6.3.4. Printer.................................................................................................................................46
5.6.3.5. Language/Country..............................................................................................................47
5.6.3.6. Acoustic warning.................................................................................................................47
5.6.3.7. Soot (1/3)............................................................................................................................47
5.6.3.8. Standard.............................................................................................................................47
5.6.4. Control list.................................................................................................................................48
5.6.5. madur Service...........................................................................................................................48
5.7. Storing of results...............................................................................................................................49
5.7.1. Storing measurement reports...................................................................................................50
5.7.1.1. Report table........................................................................................................................51
5.7.1.2. Header (Notice) editing.......................................................................................................53
5.7.1.3. Erasing reports...................................................................................................................54
5.7.2. Continuous storing of results....................................................................................................54
5.7.2.1. Banks table.........................................................................................................................56
5.7.2.2. Erasing banks.....................................................................................................................56
5.7.3. Data-logger...............................................................................................................................57
5.8. Triple XL test......................................................................................................................................57
5.9. Printer................................................................................................................................................59
5.9.1. Printing the contents of a screen..............................................................................................59
5.9.2. Printing of all results..................................................................................................................59
6. Gas block diagram..................................................................................................................................61
7. Menu structure.........................................................................................................................................62
7.1. Structure of the main menu...............................................................................................................62
7.2. Structure of the menu STORE..........................................................................................................63
3
www.madur.com Manual GA-21Plus
1. INTRODUCTION
Congratulations on your purchase of the microprocessor-controlled flue gas ana-
lyser
GA-21 plus
It is a high quality instrument produced using the latest technical discoveries. The GA-21 plus is easy to use. You also have the possibility of using the extra capabilities of this analyser, of examining measurement results, checking service intervals and many others. Please read the operating manual carefully. It contains valuable tips and informa­tion for problem-free operation. Regular maintenance will help to keep your analyser ready for use at all times.
1.1. Use of this operating manual
Keep this manual with the analyser GA-21 plus, and take it with you when carry­ing out measurements. Here is a short description of the sections of this manual.
2. Maintenance
This section contains important information designed to keep your analyser work­ing and to save you time and trouble.
3. General information
This describes the most important characteristics of your new analyser.
4. Construction
A knowledge of the hardware of the analyser will help you to use the instrument correctly and to avoid unnecessary down-time.
5. Operation
This section explains the use of the analyser with detailed descriptions of the vari­ous functions. It is important to read the sections 5.1 and 5.2 first. The other sections can be read when required.
4
www.madur.com Manual GA-21Plus
2. MAINTENANCE
2.1. Gas sensors
The following points are important for maximising the operational life of the cells and achieving measurements without errors:
do not exceed the rated concentration; all the gas cells may react unpredictably to the presence of substances which are
not usually found in flue gases. For this reason, the analyser may not be cleaned with chemical solvents. If organic solvents enter the gas chamber, the cells may become unstable or even suffer permanent damage;
some cells require the presence of a bias voltage when not in use. Do not allow
the battery to become fully discharged;
do not switch the analyser off before the system is purged of remaining flue gas; store the analyser in a cool place to reduce the ageing of the cells.
2.2. Gas system
The gas system of the analyser is protected by a double in-line filter. The ele­ments should be changed when necessary. Check the condensate traps regularly and empty them when needed.
2.3. Battery
The rechargeable battery used in the analyser is maintenance-free. An acoustic warning signal is given when the voltage drops below 11 V. The GA-21 plus will switch off automatically if the voltage should drop below 10.5 V. The analyser will operate for about 5 hours from the battery.
Warning:
Some parts of the analyser draw current even when the device is
switched off. If the analyser is not being used then the battery
should be charged fully at least once a month.
5
www.madur.com Manual GA-21Plus
2.4. Service intervals
The parameters of the electrochemical cells change with time, and therefore it is necessary to carry out a service periodically. This service should include re-calibration of the gas sensors using test gases. The recommended service interval is 800 operating hours or 6 months, whichever comes first. The analyser is fitted with an operating hour counter and automatically shows when a service is necessary.
The operating time and the number of hours until the next service are shown in
MENU->OPTIONS->madur SERVICE -> DEVICE INFO .
2.5. Errors during operation
The analyser checks all circuitry continuously. Should an error become evident, then this is signalled acoustically and by the message "ERROR" on the screen. The error can be found shown on the screen MENU->OP TIONS ->madur S ERVICE -> CONTROL
LIS T .
2.6. Switching off after use
The operating life of the electrochemical cells is dependent on the usage of the instrument. The wear and tear on the cells is greater when the cell is exposed to the gas for longer or the concentration is higher. The gas cell is hence "used up" during measure­ments.
Info:
For this reason, the analyser should not be switched off until all the
gas tubing has been completely cleared of flue gases and the gas
cells have been purged with fresh air for several minutes.
The GA-21 plus will not normally allow the instrument to be switched off if there are still traces of flue gas in the system. The analyser will continue to operate until the sig­nals from the gas cells are nearly zero. This activity is shown by the message "VENTILA­TION REMOVE PROBE FROM FLUE".
The [OFF] key must be pressed twice to switch the instrument off completely. Pressing the key once will put the analyser in the standby mode and the instrument will remain in this mode until it is activated again by pressing "C" or switched off completely.
6
www.madur.com Manual GA-21Plus
Warning:
Certain conditions, such as strong electro-magnetic fields can
cause the analyser to become "hung-up". The only solution to this
problem is to switch the instrument off. It is, however, possible
that the analyser will no longer react to the keyboard and can not
be switched off.
Info:
Pressing the [OFF] key for a period of about three seconds will always
switch the analyser off.
3. GENERAL INFORMATION
The GA-21 plus is a multi-functional flue gas analyser. Electrochemical sensors are used for the measurement of gas concentration. The instrument can be fitted with between 2 and 4 of these sensors. All analysers are fitted with O2 and CO sensors, fur­ther gas cells must be chosen when the instrument is ordered.
The following description is based on an analyser containing 3 cells, the third one being an NO sensor.
Oxygen, O2 Carbon monoxide CO Nitric oxide NO Carbon dioxide CO2 Nitrogen oxides NOx
The first three gases (O2, CO, NO) are measured directly using the electrochem­ical cells. The remaining components are calculated. The concentrations of oxygen and carbon dioxide are shown in percent. The concentrations of the remaining gases are shown as follows:
volume concentration in [ppm] absolute mass concentration in [mg/m3] mass concentration relative to the oxygen content in [mg/m3]
In addition, the air inlet or ambient temperature and flue gas temperature are measured. Using the measured temperatures, gas concentrations and the known fuel
7
www.madur.com Manual GA-21Plus
parameters the analyser calculates a variety of combustion parameters such as Stack
Loss – SL, Efficiency - , Excess Air - , Loss through Incomplete Combustion – IL.
The GA-21 plus also measures differential pressure.. This can be used to meas­ure the draught or pressure in the flue gas channel. The instrument is also equipped for the soot test according to Bacharach method.
The analyser is also fitted as a 3 channel data logger, capable of measuring voltage or current together with temperature using thermocouples or thermistors.
The instrument can be used at a broad range of ambient temperatures (0 - 50°C). It is fitted with a rechargeable battery making it independent of the mains power supply.
Hence, the GA-21 plus is ideally suited for all those involved in the construction, maintenance and adjusting of heating systems.
The keyboard of the instrument can be wiped clean if necessary.
The GA-21 plus is controlled by a microprocessor. The easy-to-read LCD dis­play, the comprehensive keyboard and the built-in printer allow the operator to communic­ate with the instrument easily and to document the measurements on the spot. The EEP­ROM memory used in the GA-21 plus allows large quantities of data to be stored without fear of data loss should the instrument be switched off or should the battery be accident­ally run down. The GA-21 plus comes complete with programmed data for the 18 most common fuels. A further 10 fuels can be programmed by the operator to cover all eventu­alities.
The RS232C interface fitted to the GA-21 plus allows all stored data to be read for later treatment or the instrument can be used on-line with the computer.
8
www.madur.com Manual GA-21Plus
4. CONSTRUCTION
4.1. Elements of gas circuit
4.1.1. Gas probe with in-line filter
The gas probe is used to extract the gas under investigation and to measure its temperature. The gas probe consists of two parts, the probe pipe with thermocouple (1) and the probe holder (2).
Picture 1. Gas probe with hose and in-line filter
The probe pipe is typically supplies in one of four standard lengths: 150 mm, 300 mm, 450mm, 750 mm, 1000mm and 1500 mm. As standard a Ni-CrNi thermocouple is used. The probe pipe is fitted with an anchor cone to simplify fixation in the flue gas chan­nel.
The probe pipe is attached to the probe holder with a threaded connector.
The gas tubing in the GA-21 plus is protected from the dust and moisture present in the flue gases by a filter (4). The filter elements (3) should be changed when dirty to ensure a free flow of gas. The condensate should be removed from the condensate traps when necessary.
9
www.madur.com Manual GA-21Plus
The plug (6) and hose quick coupler (7) are connected to the front plate of the analyser.
4.1.2. Gas pump
The internal gas pump draws the gas into the GA-21 plus and passes it through the gas chambers. The self-cleaning pump is a high quality item and is known to be ex­tremely reliable.
The gas cells operate best with a flow rate of 90 l/h, which is set at the factory and is compensated automatically for flow restrictions or changes in battery voltage.
4.1.3. Gas chamber
The gas chamber can be fitted with minimum two and maximum six gas sensors. Oxygen and carbon monoxide cells are standard equipment. The other sensors are chosen when the instrument is ordered. The gas chamber of the GA-21 plus system is characterised by minimal dead capacity, which leads to low inertia of the analyser indica­tions. In addition to the gas sensors, there is a temperature sensor placed in the chamber for temperature compensation of the sensors.
4.2. Measurement system
4.2.1. Gas sensors
The oxygen concentration is measured using an electrochemical cell. The sensor signal is directly proportional to the oxygen volume concentration. The oxygen sensor needs no calibration with standard gases – it is fully calibrated during the initial calibration when a point corresponding with the contents of oxygen in clean air (20.95%) is determ­ined.
10
www.madur.com Manual GA-21Plus
The toxic gases are also measured using electrochemical cells.
Table 1 presents all standard gas sensors available in GA-21 plus system.
Gas Cell type Range
O2 5FO 0-20.95 %
O2 MOX-1 0-100 % CO 5F 0-20000 ppm CO 5MF 0-10 % CO A5F 2000 ppm NO 5NF 0-5000 ppm
NO2 5ND 0-1000 ppm SO2 5SF/F 0-5000 ppm
H2S 3HL/M 0-1000 ppm
H2 3HYT 0-2000 ppm
CL2 3CLH 0-250 ppm
HCl 3HL(s) 0-100 ppm
Table 1 Available gas cells and ranges
To guarantee long-lasting, trouble-free function of electrochemical sensors the fol-
lowing points are important:
the stated maximum measuring range of the sensor should not be ex-
ceeded.
all of the gas sensors can react unpredictably to certain chemicals
which are not typical for combustion gas composition. For this reason, it is absolutely forbidden to wash the parts of the gas line with chemical solvents. The vapours from such solvents can get inside the gas chamber and result in destabilisation or even permanent damage of the gas sensors.
the gas sensors are working even when the analyser is switched off -
so the battery should never be disconnected or completely dis­charged.
the analyser should not be switched off before all the flue gases have
been removed from the analyser's gas circuit.
the sensor chamber must be kept dry. the switched off analyser should be kept in a cool place, which in-
creases the lifetime of the gas sensors.
11
www.madur.com Manual GA-21Plus
4.2.2. Ambient temperature sensor
The ambient temperature is measured by the analyser for use in some of the cal­culations carried out, such as Stack Loss. This can be carried out either by the temperat­ure compensation sensor built into the plug of the probe holder (INT) or the external con­nector T3 (EXT). The temperature sensor is chosen under the suboption:
OPTIONS → SENSORS → ANALOGUE INPUTS.
4.2.3. Flue gas temperature sensor
The flue gas temperature sensor is situated in the gas probe pipe and plugged into the PROBE socket on the front panel of the analyser. This sensor is a NiCr-Ni ther-
mocouple in the standard probe version. The range of such a thermocouple is 50-800oC in the standard version (1100°C available optionally). A special gas probe with PtRh-Pt
thermocouple can also be used, with a range up to 1600oC. Proper flue gas temperature measurement is the most important factor for correct calculation of the chimney loss and other heat engineering results.
4.2.4. Gas chamber temperature sensor
To compensate for temperature drift of the gas cells, the temperature of the gas chamber is constantly monitored.
4.2.5. Differential pressure sensor
The GA-21 plus is fitted with a precise semiconductor differential pressure sensor. Stubs marked +/- PRESSURE situated on the left side of the front panel are used to measure pressure. Negative pressure in flue gas channel (chimney draught) can be measured with the one stub only; differential pressure measurement is possible using both stubs. The measuring range of the pressure sensor is +/- 25 hPa.
4.2.6. Analogue inputs
The three analogue inputs installed can be used to measure voltage or current on two channels (U and I), temperature using a thermocouple (T1 EXT) and temperature us­ing a thermistor (T3 EXT). The values measured can be viewed on the screen or stored in the EEPROM memory.
12
www.madur.com Manual GA-21Plus
Current and voltage inputs – top view (outputs - optional)
1 – GND 2 – I1 in 3 – U1 in 4 – I1 out GND (optional) 5 – I1 out (0/4-20mA, optional)
Parameter U I
Range -10V to +10V -10mA to +10mA
Input resistance 100kΩ 100Ω
Resolution 10mV 10µA
Temperature input – top view
1 – PT3 (Thermistor 1 +) 2 – TH1 (Thermocouple 1 +) 3 – GND (-)
13
www.madur.com Manual GA-21Plus
Parameter Thermocouple Thermistor
Range Ni-CrNi: 1 – 1300°C
Pt-RhPt: 0 – 1600°C
Fe-CuNi: 0 – 700°C
Pt100: 0 – 200°C Pt500: 0 – 500°C
Resolution 1°C 1°C
4.3. Data input/output system
4.3.1. Keyboard
A film keyboard is used in the GA-21 plus analyser which is durable, washable and highly resistant to external agents. The keys are modern membrane keys with per­ceptible stroke.
Each press of a key is accompanied by an acoustic signal. The keyboard is shown in picture B (page 14) and consists of the following elements:
PROBE - Socket for the electrical connection of the probe holder
GAS -Connector for the gas hose
PRESSURE +,- -Connectors for the differential pressure sensor
RS 232 C -Socket for the RS232C cable
TEMP. 1 -Socket for the temperature channels 1 & 2
U/I 1 -Socket for the current/voltage channels
Function keys -
Arrow keys -
Printer keys -
ON/OFF keys -
Numerals - to
14
www.madur.com Manual GA-21Plus
Picture 2. Front elevation picture
15
www.madur.com Manual GA-21Plus
4.3.2. Display
The GA-21 plus has an LCD display with graphic capabilities. To improve the read-out quality, contrast regulation for the display and backlighting are used. Measure­ment results, calculation results, diagrams, text information, menu options and others ap­pears on the display.
4.3.3. Printer
The GA-21 plus is fitted with an EPSON dot matrix printer capable of printing the data as graphic or in the form of tables. All the screens seen on the analyser may be prin­ted.
Printing speed is about 2.5 lines per second. The printer uses normal paper, 57 mm wide.
4.3.4. Interface RS232C
The GA-21 plus is fitted with a serial RS232C interface. The socket is mounted on the front of the instrument and marked RS232C. The interface allows the GA-21 plus to communicate with a PC and the software madCom is available for carrying out extens­ive analysis of the stored results.
Top view
1 – +15V out 2 – RxD 3 – GND 4 – TxD
16
www.madur.com Manual GA-21Plus
4.3.5. Power supply
The GA-21 plus can be used with mains power supply or be operated from the internal battery. A soot test can only be carried out when the mains supply is connected, otherwise the heating of the probe holder will not function. The mains cable is attached to the socket on the left of the keyboard. Switching the mains supply on or off when the in­strument is in operation is not recommended. The analyser switches automatically to the internal battery when the mains supply is disconnected. The LED on the front panel shows which power supply is in use.
Red -analyser is switched on and supplied from the battery. Mains supply
not present.
Green -analyser switched off and mains supply present. The battery is char-
ging.
Orange -analyser switched on and mains supply present. The battery is char-
ging.
The maintenance-free 12 V battery has a capacity of 2 Ah. A fully charged battery will power the instrument for about five hours.
When the battery voltage drops below 11 V, three acoustic warnings and blinking of the LED on the front panel will appear. The message ERROR appears at the top of the screen and the CONTROL LIST contains the line BATT LOW.
The instrument switches off automatically if the battery voltage drops below
10.5V.
The battery is charged by connecting the instrument to the mains power supply.
The state of charge can be seen during the initial calibration or read from the screen DEVICE INFO:
17
www.madur.com Manual GA-21Plus
5. OPERATION
5.1. Use of the keyboard
5.1.1. Description of the keys
- Switches to the STORE menu
- Switches to the first results screen. From the current results, the use of this key switches between MEAS­URE and HOLD.
- Switches to the MAIN MENU
- Confirms entries and switches options
to - Alphanumeric keyboard for entering data and for fast
movement between MENU options
- Decimal point when entering data. If used when the screen is showing RESULTS, this key switches the zoom function on or off.
- starts a print-out
- moves the paper forward one line
- switches the analyser on
- switches the analyser to stand-by or off
- moves one step backwards in the menu
- moves one step forward in the menu or, during edit­ing, moves the cursor one space to the right
and - move the cursor vertically in the menu
18
www.madur.com Manual GA-21Plus
5.1.2. Selecting menu options
The menu structure is shown at the end of the manual. The various displayed
menus are leading the user through the structure.
The following picture shows an screen, which explains the different types of exist-
ing lines.
Here is an explanation of the movement between the various menu options:
and - move the cursor vertically in the menu
- if the cursor is on the left side of the Title line, the key moves one step backwards in the menu
- if the cursor is on the right side of the Option line or the Title line, the key leads to the displayed Option.
- executes the displayed function (starts data storage for example ), in the Command line starts or ends the respective edit function (setting the average time for example ) in the Data line switches (i.e. YES/NO) in the Switch line
to - for fast movement between MENU options
If a displayed line carries a function, then there will be a number (1 to 9) on the left side of the line. Press the number of the option you want to select, and the cursor will "jump" to the selected line. If number 4 is pressed in the above demo screen the beam would move to BOILER DATA.
19
www.madur.com Manual GA-21Plus
5.1.3. Entering numbers
Numbers are edited as follows:
- digital keys enter corresponding numbers
- move the cursor within the line
- erases whole number (zero on each position)
- confirms and ends the entry
5.1.4. Editing text
Text is edited as follows:
- switches the keyboard into either digital or alpha betical mode. When in digital mode the cursor has a block form on the screen, in the alphabetical mode it has the shape of an underlining
- move the cursor within the line
- erases the whole text line
- the alphanumeric keys enter numbers in digital mode or letters in alphabetical mode. Each key has a three letter designation. The first press enters the first letter, the next presses the second and the third letter. Key presses in alphabetical mode does not move the cursor.
- (SPACE) prints 0 number in digital and the space character in alphabetical mode. The space moves all the next characters one position to the right.
- (DEL) erases the character above cursor position. This moves all the following characters one position to the left.
20
www.madur.com Manual GA-21Plus
- ends the entering of a text (if only one line is edited
- e.g. the name of a fuel) or moves the cursor down to the next line (if there are several lines in the ed ited text - e.g. the report header)
- end the editing of a header or footer - it enables the storage of edited text. If you do not need to store the contents of header or footer you should leave the screen by pressing the MENU key.
5.2. Basic operating instructions
5.2.1. Setting-up the analyser
Connect the components of the analyser system:
- Attach the hose to the connector GAS. The plug from the probe holder connects to the socket PROBE.
- Attach the outlet hose from the filter to the connector INLET
- Ensure that the gas flow from the base of the analyser is free of ob structions.
5.2.2. Switching on
Before use, the condition of the analyser should be checked:
-Check the filter elements. Clean or replace as necessary.
-Check the state of charge of the battery if the analyser is to be operated from the internal battery. This is carried out by switching on and reading the battery voltage or charge from the initial screen, or from the screen MENU/OPTIONS/DEVICE INFO.
5.2.3. Switching off / standby
Pressing the key [OFF] once will switch the analyser to the STANDBY mode.
In STANDBY mode the power consumption is drastically reduced. Amongst oth­ers the pump and the backlight are switched off. Only the sensors remain supplied to be ready to operate immediately after the instrument is switched on again.
21
www.madur.com Manual GA-21Plus
The analyser does not switch to STANDBY mode if [OFF] is pressed in case:
-the key is being pressed during Initial Calibration. In that case the instru ment realises the action, but will switch to STANDBY only after the cal ibration is finished.
-the key is being pressed while the is still a high gas concentration detec ted in the gas chamber. In this case the instrument will automatically switch to FLOODING WITH AMBIENT AIR mode and switch to STANDBY mode later on.
The following screen is shown in the STANDBY mode.
By pressing [OFF] the instrument can be switched off completely. If [C] is pressed the instrument switches ON again. If the instrument is in STANDBY mode for more than 8.5 hours it switches off automatically. It is highly recommended to carry out a
calibration of the oxygen sensor when re-activating the instrument.
22
www.madur.com Manual GA-21Plus
5.2.4. Initial calibration
After the instrument is switched on it carries out an initial calibration for 2 minutes.
Warning: Make sure the probe is not in the stack during the initial calib
ration
Initial calibration is of basic importance for measurement correctness and it must not be interrupted before it is finished automatically.
During this time the Oxygen sensor is calibrated to 20,95% in ambient air, and the other sensors (CO, NO,...) will be zeroed.
23
www.madur.com Manual GA-21Plus
5.2.5. Measurement parameters
Directly after the calibration is finished the option PARAMETER is displayed for control reasons or in case changes need to be done. Usually the option is selected as fol­lows:
- selects the MENU screen
and - move the cursor to the PARAMETER option
- select the option. The following screen will appear:
24
www.madur.com Manual GA-21Plus
5.2.5.1. Fuel selection
This option enables the selection of a fuel (if necessary), or the programming of a customised fuel.
From PARAMETER you move to the option as follows:
and - move the cursor to FUEL
- choose SELECT FUEL
and - move the cursor to the desired fuel
- confirm the selection - an asterix (*) will apear on the left side
25
www.madur.com Manual GA-21Plus
Each fuel of this screen is defined through various parameters. To view them pro-
ceed as follows:
and - move the cursor to the desired fuel
- displays the fuel parameters , or gives the opor tunity to enter fuel data for the programmable fuels.
- back to option FUEL
The following screen appears.
The already defined fuels can only be viewed. For the user definable fuels all
parameters can be programmed.
The variables are as follows:
0000000000- Name of the fuel
DP-
CO2max - the maximum concentration of CO2 (for complete combustion,
O2=0 %)
A1, B - factors for Siegert's formula
ALPHA - fuel specific parameter
O2R - reference oxygen
Vgd - volume of the flue gas at perfect combustion of the fuel
HV - Heating value of the fuel
UNITS - unit in which the fuel is measured
26
www.madur.com Manual GA-21Plus
5.2.5.2. Averaging time
To select the average time proceed as follows:
and - move the cursor to AVERAG. TIME
- select Average Time edition
to - select the time (1 to 60 )
- confirm the number
and - select min or sec
- confirm the selection
5.2.5.3. Reference oxygen parameters (O2R)
To select the reference oxygen proceed as follows:
and - move the cursor to O2R
- switches between a pre-programmed standard value <NORM> and a free programmable one, called manual value.
to - if "manual" was selected, a number from 1 to 20
can be entered
- confirms and ends editing
27
www.madur.com Manual GA-21Plus
5.2.5.4. Boiler parameters
This option enables the user to have relevant boiler parameters printed on the
Protocol printout for exact documentation.
and - move the cursor to BOILER-PARAMETER
- select with YES/NO whether or not the Boiler Para­meters shall appear on the printout, respectively be­ing stored in a Protocol
- if YES was selected, this key moves to the Boiler Parameter menu. The boiler parameters can be entered in the screen shown below:
Boiler power
Boiler consumption
Boiler temperature
28
www.madur.com Manual GA-21Plus
5.3. Displaying the measurement results
Select the measurement screen by pressing DATA. The measurement results are
displayed on three screens, either as current values or as averaged values.
5.3.1. Current values
- select the first screen
and - move backward or forward to select a screen
29
www.madur.com Manual GA-21Plus
- switches between AVERAGED VALUES and CUR-
RENT VALUES
- switches On/Off the HOLD function ("freezing" the current values )
5.3.2. Averaged values
Select desired averaging time as described in section 5.2.5.2.
- select the screen. The title line has to read AVER-
AGED VALUES
and - backward or forward to select one of four
screens
- switches between AVERAGED VALUES and CUR-
RENT VALUES
- the display on the AVERAGED VALUE screen gets "synchronised" (a row of strokes appears instead of the values) and a new averaging session starts. The meas ured results are displayed after the whole cycle is finished.
- switches On/Off the HOLD function ("freezing" the current values)
Warning:
If a continuous storage recording is taking place the analyser won't syn-
chronise the averaging process (not to interfere with the continuity of stor-
age )
30
www.madur.com Manual GA-21Plus
5.3.3. The averaging process
The analyser is capable to display all results, measured or calculated, either as current value or as averaged value. The averaging time can be selected in a range of 10 seconds to 60 minutes. If a time of 1 min was set, the averaged results over that time would be shown on the display. During the start of the averaging process a row of strokes appears on the screen, since there are no results to display.
If measurement parameters are changed during the averaging process, the aver­aging cycle has to be re-started.
In case the data storage is activated it is stopped as well.
The following incidents cause an interruption of the averaging cycle.
change of the averaging time
change of the reference oxygen
change of the fuel
switching the instrument to the STANDBY mode
31
www.madur.com Manual GA-21Plus
5.4. Pressure/Draft measurements
It is possible to measure differential pressure with the GA-21 plus. To do so the stubs PRESSURE (-) and (+) on the front plate are used. The measurement is performed as follows:
- select the main MENU
and - move the cursor to PRESSURE
- select the option. The following screen appears
The current pressure and gas temperature are displayed on the screen. The last line shows the command ZEROING. By pressing this key the sensor is set to zero, e.g. the current value will be taken as zero value..
Please be aware that the sensor has to be zeroed before the measurement.
- ZEROING the sensor
Connect end of the sample hose to the PRESSURE(+) stub. The screen will dis­play the measured pressure and the measured gas temperature.
- leaves the option. The displayed value will be stored until a new one is stored or either a printout of the measured values is performed or they are stored as a report.
32
www.madur.com Manual GA-21Plus
5.5. Soot (smoke) Measurement
To perform a soot test it is essential to prevent the flue gas from condensing on the filter. Therefore it is necessary to use a heated probe holder. It is only possible to per­form a soot test if the instrument is connected to the mains.
The soot test is done by means of the Bacharach comparison method.
- select the main MENU
and - move the cursor to SOOT TEST
- select the option, the pump is switched off and the following screen appears:
Insert the soot filter in the slit of the heated probe holder, and place the probe in
the stack.
- starts the soot measurement. The pump is switched on and sucks in 60 s (according to. TÜV) the re­quired volume of 1.63 +/-0.07 l
33
www.madur.com Manual GA-21Plus
- stops the soot test
After the soot test the filter is compared with a scale, and the evaluated soot num-
ber is entered.
to - enter the detected soot number
- confirm the soot number
The stored soot number remains stored until a new one is entered, or either a
printout of the measured values is performed, or they are stored in a report.
5.6. Options
The OPTIONS can be reached from the MENU as follows:
and - move the cursor to OPTIONS
- select option. The following screen appears
34
www.madur.com Manual GA-21Plus
5.6.1. Device info
and - move the cursor to DEVICE INFO
- select option
The DEVICE INFO screen shows the serial no. of the instrument, the date of the last service, the operating time, the time until the next service, the speed of the sucked flue gas , the battery charge, the battery capacity and whether or not the communication with the madCom PC software is locked.
35
www.madur.com Manual GA-21Plus
5.6.2. Sensors
Select the option SENSORS from the screen OPTIONS as follows:
and - move the cursor to SENSORS
- select the option. The following screen appears
5.6.2.1. Sensor CO (only in dual CO sensor configuration)
In dual CO sensor configuration there is possibility to select working mode:
-Auto – This mode is designed to protect low range CO sensor. Measurement results collected using low concentration CO sensor. When the meas­ured gas concentration exceeds value inserted in CO C. range then the purging pump starts flushing the low concentration CO sensor with ambient air for its protection. Result are from this point collected using high concentration CO sensor. When the gas concentration drops be­low value inserted in CO C. range then the pugging pump stops and the measurements are again collected using low concentration CO sensor.
-Hi – In this mode purging pump is always on and results are always collec­ted using high concentration CO sensor.
36
www.madur.com Manual GA-21Plus
5.6.2.2. CO range
If the concentration of CO in the gas stream exceeds the limit set here, then the
cell is automatically purged with fresh air. This limit can be altered as follows:
and - move cursor to the line CO RANGE
- start editing of the value
to - enter a number between 0 and 20000 ppm
- close the option
5.6.2.3. O2 cell calibration
To increase the accuracy of the O2 sensor during measurements over a longer period of time the GA-21 plus has the capability of calibrating the sensor with ambient air.
Take the probe out of the stack !
and - move the cursor to O2 cell calibration
- select the option. The following screen appears
- confirms the calibration
37
www.madur.com Manual GA-21Plus
5.6.2.4. External channels
Get from the screen SENSORS to the option EXTERNAL CHANNELS as fol­lows:
and - move the cursor to EXTERNAL CHANNELS
- select option
and - move the cursor to the line to be changed
- change the setting:
TH GAS - Gas temperature sensor – default setting: NiCr-Ni. For special
applications (temperatures up to +1600°C) the setting can be changed to PtRh-Pt thermocouple. Of course that would need the use of the respective thermocouple as well.
Pt AMB - Ambient temperature sensor - default setting: Pt 500. The
ambient temperature can either be measured with an internal sensor placed in the probe plug (Pt500), or with an external sensor (Pt500), connected to the TEMP (T3 EXT.) socket.
T1/T2 EXT - External thermocouple - Select the desired temperature
sensor. The T1 EXT input can be defined as a Ni-CrNi, Pt-RhPt or Fe-CuNi thermocouple.
38
www.madur.com Manual GA-21Plus
T3/T4 EXT - External resistance thermometer - Select the desired temper-
ature sensor. The input T3 EXT can be defined as a Pt100, Pt500 or KTY-2k thermistor.
U/I1 EXT - External Voltage/Current input - Select the desired input sig-
nal . The socket U/I1 can be used to measure either current (0­10 mA) or voltage (-10V - +10V).
U/I2 EXT - External Voltage/Current input – second current or voltage in-
put
39
www.madur.com Manual GA-21Plus
5.6.2.5. Sensor calibration
Except for O2 every electrochemical sensor needs to be calibrated with calibra­tion gas. To calibrate the sensors they are fed one by one with calibration gas of a known concentration. After a few minutes the sensor supplies a stabilised signal which can be stored as calibration value. The order of calibration does not effect the calibration.
Warning:
Any calibration screen value change is followed by loss of the original cal-
ibration settings. Thus the CALIBRATION option should be selected only when it is
to be carried out.
The standard gas concentration should be similar to that usually met during measurement and it should not be less than 10% of the required measuring range of a given sensor
The time the sensor is fed with standard gas should not be less than 4 min. After each calibration the sensors should be flooded with ambient air for at least 15 minutes.
Calibration procedure:
and - move the cursor to CALIBRATION W. GAS
- select the option
- carry on with the calibration
- abort the calibration, move back to SENSORS
40
www.madur.com Manual GA-21Plus
and - move the cursor to the sensor to be calibrated
- select the gas type
Calibration screen will appear:
All currently installed electrochemical sensors are visible in this screen
The individual rows have the following meanings:
Cell – Sensor type
Signal – Current value of sensor's signal
Stored – Value of stored signal
- in the GAS [ppm] line, starts editing of the standard gas concentration
to - enter the concentration of the standard gas [ppm] in
the line GAS
41
www.madur.com Manual GA-21Plus
- in the GAS line. Confirm standard gas editing. Sup­ply the instrument with standard gas for about 4 min.
- in the STORE line, as soon as the value has stabil­ised, stores the measured value as calibration value. Disconnect the standard gas and flood the instru­ment for at least 15 min with ambient air.
- return to select the next gas, or end the calibration
5.6.2.6. Calibration adjustment
Calibration adjustment is a new feature (introduced in firmware version 2.07). Al­lows to perform a quick correction to a factory calibration using one calibration mixture (mixture of gases in N2 balance).
Calibration adjustment does not substitute factory calibration with a single gas in N2 balance. I .e. calibration adjustment cannot create matrix of cross-sensitity signals – this is created only during standard calibration.
- means that correction for a selected gas is applied.
If all sensor are unmarked: , it means no adjustment to factory calibration is ap­plied.
To perform calibration adjustment:
Enter the concentration for each gas present in mixture. If any of listed gas is not present in mixture set its concentration to 0 → analyser will not include this sensor in ad­justment.
42
www.madur.com Manual GA-21Plus
Apply reference mixture for few minutes and allow the signals to stabilise.
Finish the adjustment by pressing * ADJUST * button.
Adjustment is remembered, even after powering ON/OFF the analyser! To re­move this calibration adjustment, one need to manually perform REMOVE ADJUST-
MENT.
5.6.3. Settings
This option allows the user to adjust some of the standard settings.
and - move the cursor to the option SETTINGS
- select the option
5.6.3.1. Backlighting
- pressing the key switches the BACKLIGHTING on and off
5.6.3.2. Contrast
and - move the cursor to CONTRAST
- select option
Follow the explanations on the screen to adjust the contrast
43
www.madur.com Manual GA-21Plus
44
www.madur.com Manual GA-21Plus
5.6.3.3. Clock
This option allows the internal clock to be adjusted.
and - move the cursor to CLOCK
- select option
and - move the cursor to desired line
- opens editing
to - set date and time. They are entered in the sequence
TIME - HH:MM.SS and DATE as given by the format line below.
- confirms settings
45
www.madur.com Manual GA-21Plus
5.6.3.4. Printer
This option allows the printer (printout) settings to be defined.
and - move the cursor to PRINTER SETTINGS
- select option
SHORTFORM - switches between YES/NO. If the option is activ-
ated, then a print-out of the current values will show only the volume concentrations of the gases, not the mass concentrations in mg/m³. A stored report will still have all the information as usual.
AUTO PRINTING - switches between YES/NO. If the option is activ-
ated, automatic printouts will be carried out ac­cording to the programmed average time, provided the time is not less than 2 min. If the av­erage time is shorter than 2 min. only one prin­tout will be performed.
No OF COPIES - switches between 1/2/3. FOOTER No - the instrument can store up to 50 footers. Num-
ber one is already stored and can not be changed. If address 0 is selected, no footer will be on the printout. The footers 2 to 50 are freely programmable.
46
www.madur.com Manual GA-21Plus
5.6.3.5. Language/Country
The instrument enables the user to select between languages.
and - move the cursor to LANGUAGE
- confirms the selection
and - select language
- confirms the language
5.6.3.6. Acoustic warning
- enables/deisables acoustic warning
5.6.3.7. Soot (1/3)
5.6.3.8. Standard
- sets the instrument to default settings
47
www.madur.com Manual GA-21Plus
5.6.4. Control list
The option CONTROL LIST displays a similar list to the one above. The list
shows which subassembly functions correctly (OK), or which is defective (*E*).
and - move the cursor to CONTROL LIST
- select option
5.6.5. madur Service
This option is for service engineers only. A code is needed to enter the menu -
exit with [MENU] key.
48
www.madur.com Manual GA-21Plus
5.7. Storing of results
Since the measured results can be stored in the EEPROM memory, it is possible to read them out later for treatment using the madCom software. Data can be stored in two separate forms. Up to 30 measurement reports and up to 10 banks from continuous measurements containing a total of 1024 sets of results. Using an averaging time of 30 minutes, this would allow the results of over 21 days of measurements to be stored (please be aware that for measuring sessions of that length a proper gas conditioning system is absolutely necessary)
- selects the menu STORE
49
www.madur.com Manual GA-21Plus
5.7.1. Storing measurement reports
The measurement report contains all the results obtained in a single measure­ment. It contains either current or averaged values and other user-defined factors such as fuel, footer, header etc. If the report should contain soot measurement results or pressure data, then these measurements must be carried out before storing the report. The report will contain either the current or the averaged results depending on which screen was vis­ible when the storing process was started.
Store REPORTS as follows
- the measured values are displayed. Do you want to print the displayed valuesproceed as follows
- select option. The measured values are stored in a buffer memory
and - move the cursor to REPORTS
- select option REPORTS
and - move the cursor to SAVE REPORT No. XX
- the "buffered" data are stored in a Report with the displayed number.
50
www.madur.com Manual GA-21Plus
5.7.1.1. Report table
This option allows one to view, print and erase Report contents.
and - move the cursor to REPORTS TABLE
- select option
Reports without [] are empty.
- erases the selected Reports
- shows the selected Reports, the following screen REPORT NO.: XX is opened
51
www.madur.com Manual GA-21Plus
52
www.madur.com Manual GA-21Plus
5.7.1.2. Header (Notice) editing
This option allows to edit a notice which can be stored and printed with a Report.
The HEADER may have 4 lines of 20 characters.
Measurement site and/or short description of the site are commonly stored in
such notes.
and - move the cursor to HEADER EDITING
- select option
The pictures show screens without text. The screen appears as it was when the
report was stored. How the text should be edited is described in section 5 .1.4 .
This screen can be left at any time using [MENU] and [DATA].
53
www.madur.com Manual GA-21Plus
5.7.1.3. Erasing reports
All protocols can be erased in that option.
and - move the cursor to DELETE ALL REPORTS
- select option
- leaves the screen without erasing the Reports
- erases all reports
5.7.2. Continuous storing of results
This type of storing involves the continuous storing of a set of up to 10 measure­ment results (ONE BLOCK) during a measurement session. The EEPROM can store up to a total of 1024 blocks freely dividable between 10 measuring sessions. These 10 sets of results are known as BANKS. The size of a single bank is only limited by the free memory available. Each bank also contains all necessary information about the measure­ment (fuel, time, O2R, averaging time etc.). All the results stored are average results for the time set. If the analyser is switched off, and under certain circumstances, the storing of averaged results will stop automatically.
This is signalled by a single long warning tone.
54
www.madur.com Manual GA-21Plus
Starting the continuous storing of results ( in Banks )
- call the STORE menu
check whether enough free memory is available for the measurement. If not, one or more banks must be erased. The option DELETE ALL BANKS can be used.
and - move the cursor to the option BANKS
- open the option BANKS
and - move the cursor to the line STORAGE
Set or check the averaging time.
- start the storing, the word ACTIVE appears
- the measured/calculated values are shown, storage is active
Stopping the continuous storing of results
- call the STORE menu
and - move the cursor to the option BANKS
- open the option BANKS
and - move the cursor to the line STORE
55
www.madur.com Manual GA-21Plus
- storage is stopped, the word ACTIVE disappears
5.7.2.1. Banks table
The option BANKS TABLE provides information about the stored Banks.
and - move the cursor to the option BANKS TABLE
- select the option BANKS TABLE.
The last bank can be erased in the BANKS TABLE screen (the Bank must not be
active)
and - move the cursor to the last stored Bank
- erase bank
5.7.2.2. Erasing banks
The option DELETE ALL BANKS allows to erase all Banks.
and - move the cursor to DELETE ALL BANKS
- select the option
- leaves the screen without erasing BANKS
- erases all BANKS
56
www.madur.com Manual GA-21Plus
5.7.3. Data-logger
GA-21plus analyser can be optionally equipped with data-logger. Such device has additional microSD card slot, located beneath the printer cover. Data-lgger options are available from Store menu:
To use data-loger, insert SD card to the slot and, if necessary, format it from analyser’s menu. Then thick Recording option to start collecting measurements to csv file (comma separated) on SD card.
5.8. Triple XL test
for
2 seconds.
Before executing the test, set necessary parameters:
1. Header (Notice) editing
2. Fuel selection
3. Averaging time
4. Reference oxygen parameters (O2R)
57
www.madur.com Manual GA-21Plus
5. Boiler parameters When all is set, execute 3-XL by pressing ***** START **** - analyser will perform test ac­cording to the set averaging time. Each test will start immediately after previous one has ended. Currently performed test is indicated with its number in top left corner. Results presented on screen are for the previous period, i.e. during the first test, instead of results ‘----’ are presented. During the second test, results from the first period are presented. During the third test, results from the second test are presented:
3-XL test is stored along with reports. 3-XL test occupy memory of 2 reports, and
are marked in Report table with +:
Reports from 3XL test contains 4 pages:
Average from all 3 tests Average from the 1st test Average from the 2nd test Average from the 3rd test
58
www.madur.com Manual GA-21Plus
5.9. Printer
5.9.1. Printing the contents of a screen
This is not possible for the screen (D1). A screen may not be printed whilst data is
being entered using the alphanumeric keyboard until this activity has been ended.
- prints the contents of a screen
5.9.2. Printing of all results
- open the first results screen (D1)
- start a print-out containing all measured results
If the current values are displayed they will be printed; if the averaged values are
displayed then these will be printed.
59
www.madur.com Manual GA-21Plus
60
www.madur.com Manual GA-21Plus
6. GAS BLOCK DIAGRAM
1. Gas inlet
2. Filter
3. Main gas pump
4. Flow control sensor
5. Pressure sensor
6. Electrochemical cells in the main gas channel
7. Check valve
8. Purging pump
9. Electrochemical cell in the secondary gas channel
10. Gas outlet 1
11. Gas outlet 2
12. Differential pressure inlets for the duct's flow velocity measurement
13. Second pressure sensor
61
www.madur.com Manual GA-21Plus
7. MENU STRUCTURE
7.1. Structure of the main menu
62
MAIN
Parameter Fuel selection Choose fuel View/define
Average time
O2R
Boiler data
Pressure Pressure
Soot test Soot test
Options Device info. View data
Sensors CO range
O2 cell calibr. Calibrate
External channels Choose input TH Gas
PT Amb.
T1ext. / T3ext.
U/I1
Calibration w. gas Choose cell Calibrate
CO2 zero
Settings Backlighting
Contrast Set contrast
Clock Set date/time
Printer settings Shortform
Automatic printing
No. of copies
Choose footer
Language/country Language
Standard
Control list
Service
www.madur.com Manual GA-21Plus
7.2. Structure of the menu STORE
STORE
Reports Save Report No.
Reports Table View Report
Header Editing Edit
Banks Activate/Lock
List of Banks View List
Delete All Reports Erase All
Delete All Banks Erase All
63